Andesfit Health UFR103 Infrared Forehead Thermometer Instruction Manual

June 1, 2024
Andesfit Health

Andesfit Health UFR103 Infrared Forehead Thermometer

Andesfit-Health-UFR103-Infrared-Forehead-Thermometer-
product

Instruction manual Infrared Forehead Thermometer
Model:UFR103

Introduction

Your new Infrared forehead thermometer uses advanced infrared (IR) technology to measure temperature instantly and accurately on the forehead or object.
Easy to use and less measurement time This thermometer does not need to contact body or object to ensure the safety- and hygiene. Its ergonomic design makes this thermometer be simple and very easy to check the temperature. It only takes 1 second to take measurement and reading.

  • Body and object modes
    This thermometer supports to measure temperature of body and object. The measuring range of object mode is 0.0 to 100.0°C (32.0°F~199.9°F ). That means except to take body temperature,it also supports to take the surface temperature of object as below, Surface temperature of milk in the bottle;

    • Surface temperature of baby bath;
    • Environment temperature.
  • Alarm sounds
    When body temperature is over 37.5°C, an alarm sounds for 10 times. A Memory function It stores up to 20 sets recent measurement data.

  • Auto power off
    Automatically power off if left idle for Approx 60$5s.

  • Measurement
    Measurement time interval in 1s and the measurement distance is within 1~3cm. Please read the manual carefully before you use the unit,and keep for future reference.

Intended use
The Infrared Thermometer is intended for the intermittent measurement and monitoring of human body temperature from forehead or object. The device is indicated for use by people of all ages at homecare and in hospital.

Safety Information

To assure the correct use of the product, basic safety measures should always be followed including the warning and the caution listed in the instruction manual

Symbol descriptions
The following symbols may appear in this manual, on the label, on the device, or on it’s accessories. Some of the symbols represent standards and compliances associated with the device and its use.

Andesfit-Health-UFR103-Infrared-Forehead-Thermometer-
\(1\)

  • This thermometer is not intended to substitute for a consultation with your physician.The Forehead scan Temperature serves as a reference only. It cannot be a judgment on fever.
  • Basic safety precautions should always be observed, especially when the thermometer is used on or near children and disabled persons.
  • Please place the device unreachable by young.
  • Avoid direct sunlight.
  • Do not touch the lens.
  • No modification of this device is allowed.
  • The swallowing of small parks like packing bag, battery,battery cover and so on may cause the suffocation.
  • Please do not use a dilution agent, alcohol or petrol to clean the unit.
  • Please A treat is gently and prevent the falling from a high place.
  • Please do not immersed it in liquid. Never leave battery in the battery compartment for a long time without use, as they may leak and cause damage to the unit.
  • Please take off the battery if you do not intend to use within 3 months. Replace with new batteries if the unit display a low battery symbol.
  • Do not mix the old and new batteries together.
  • Do not use during the transportation.

WARNING
Do not dispose of electrical appliances as unsorted municipal waste, use separate collection facilities. Contact you local government for information regarding the collection systems available. If electrical appliances are disposed of in landfills or – dumps, hazardous substances can Teak into the groundwater and get into the food chain, damaging your health and well-being.

Classification

  1. Internally powered equipment;
  2. Type BF applied part;
  3. Protection against ingress of water or Particulate matter: IP21;
  4. Not category AP /APG equipment;
  5. Mode of operation:Continuous operation.

The user must check that the equipment functions safely and see that it is in proper working condition before being used.

Product Structure

Display

  1. Object mode
  2. Forehead temperature mode
  3. Ambient temperature mode
  4. Temperature unit (°C/°F)
  5. Temperature reading
  6. Memory symbol
  7. Battery symbol
  8. Horn symbol
  9. Number of memory groups
  10. Bluetooth symbol

Body

Andesfit-Health-UFR103-Infrared-Forehead-Thermometer-
\(4\)

Battery Installation

Remove the battery cover from the battery compartment, insert the battery.

1. Remove the battery cover from the battery compartment as the arrow direction accordingly.
2. Insert 2 AAA powerful batteries into the compartment and ensure each battery is in the proper direction, Positive (+) and Negative (-) are displayed on the back of battery cover.
3. Close the battery cover.

Low battery and replacement
When power on, the low battery symbol will display once the unit start to work, and you must replace with new batteries, otherwise the unit can’t work.

Battery type and replacement
Please use 2pcs AAA identical 1.5V alkaline batteries. Do not use the batteries beyond their expiry date. Please remove the batteries if you do not need to use for long time.

WARNING
Dispose of the battery in accordance with all federal, state and local laws. To avoid fire and explosion hazard, do not burn or incinerate the battery.

Setting Mode

How To Set

1. Mode setting: The display all shows for one second after turning on the On-off key, the default setting is “Body mode”after turning on . Short press the setting button once to switch to “Object mode”, and then press the setting button again to switch to “Ambient temperature mode”. If you press the setting button again, it will return to the “Body mode”.Andesfit-Health-UFR103
-Infrared-Forehead-Thermometer- \(6\) Note: The Body Mode is used to measure the forehead temperature, and the Object Mode is used to measure the object temperature, then the Ambient mode is used to measure the Ambient temperature.
2. Unit setting and sound switch settings: Long press the setting button for 3 seconds to enter the °C/°F and sound switch settings.
Step 1 : Enter °C / °F switch: the interface displays °C or °F, press the memory key to switch the temperature unit, press the set key to confirm the selection, at the same time enter the sound setting selection mode (Restart, the unit will default to “°C”). Step 2 : Enter the sound switch setting: the interface displays On or OFF, press the memory key to switch. If you select on, the horn symbol is displayed on the screen. If you select OFF, the horn symbol is not displayed on the side. Press the set button to confirm the selection. The setting is completed automatically. After setting, it will automatically return to the measure mode and wait for measurement. (Restart, the sound will default to on)

Proper Use Of The Unit

Pre-measurement
About Normal Body Temperature&Fever The temperature in the forehead and temple area differs from the internal temperature, which is taken orally or rectally. Vasoconstriction, an effect that constricts the blood vessels and cools the skin, can occur during the early stages of a fever.In this case, the temperature measured by the Infrared thermometer may be unusually low. If the measurement therefore does not match the patient’s own perception or is unusually low, re peat the measurement every 15 minutes. As a reference, you can also measure the internal body temperature using a conventional oral or rectal thermometer. Body temperature can vary from one individual/person to next. It also varies by location on the body and time of day. Below shows the statistical normal ranges from different sites. Please keep in mind that temperatures measured from different sites, even at the same time, should not be directly compared. Fever indicates that the body temperature is higher than normal. This symptom may be caused by infection, overdressing or immunization. Some people may not experience fever even when they are ill.
These include, but are not limited to, infants younger than 3 months old, persons with compromised immune systems, persons taking antibiotics, steroids or antipyretics (aspirin, ibuprofen, acetaminophen), or persons with certain chronic illnesses. Please consult your physician when you feel ill even if you do not have fever.

*Table1 Body Site Normal Temperature Range**

Andesfit-Health-UFR103-Infrared-Forehead-Thermometer-
\(19\)

Note: Body Temperature at WebMD; website: http://firstaid.webmd.com /body-temperature;retrieved at 2010 Jan 7.

As a Body Thermometer
Press the On-off button ,after the display is fully displayed for 1 second, the measurement preparation screen is displayed. The C icon lights up. Align the unit with the center of the forehead and let the infrared thermometer probe be 1cm away from the forehead,measurement completion prompts measurement results.(Buzzer: One long beep Voice)

Andesfit-Health-UFR103-Infrared-Forehead-Thermometer-
\(9\)

Note:

  • If the reading is≥37.5°C(99.5°F)and <43°C(109.4°F), the buzzer ten short beeps.
  • After the measurement is completed, the measurement can be continued after 1 second.
  • As the forehead measurement temperature is likely to be affected by sweat, oil and the surroundings, the reading shall be taken as a reference only.
  • If the probe is placed at an angle close to the forehead measurement, the reading will be affected by surrounding temperature.
  • Babies’ skin reacts very quickly in the ambient temperature. Therefore, do not take their temperature with the non-contact thermometer during/after breastfeeding, because the skin temperature maybe lower than the internal body temperature.
  • The user and the infrared thermometer should be quiescence for at least 30 minutes under similar indoor conditions.

As a Object Thermometer
Press the On-off button, after the display is fully displayed for 1 second, switch to the object mode and display the measurement preparation screen. The °C icon will light up, Point the infrared thermometer at the object to be measured, and press On-off button to measure, measurement completion prompts measurement results.(Buzzer: One long beep Voice)

Andesfit-Health-UFR103-Infrared-Forehead-Thermometer-
\(9\)

Note:

  • As the object scan temperature is likely to be affected by the surroundings, the reading shall be taken as a reference only.
  • If the probe is placed at an angle close the object scan, the reading will be affected by surrounding temperature.

As a Ambient Thermometer
Press the On-off button, after the display is fully displayed for 1 second, switch to the ambient temperature mode, display the measurement preparation screen, the C icon lights up, press the On-off key measurement, measurement completion prompts measurement results. (Buzzer: One long beep Voice)

Andesfit-Health-UFR103-Infrared-Forehead-Thermometer-
\(11\) Memory-recall of measurements
When the device is off, press the “memory button” to enter the read memory mode,the memory icon M and”the first group of memory values”appear, Each time you press the key in sequence, 20 groups of memories are read in sequence, and the first group of memories The value is the latest measurement reading, and so on. When reading the 20th group of memory, if you press the memory key again, the first group of readings will be re-read.

Memory-clear Of Measurements
In the shutdown state, long press the memory button for about 8 seconds, the LCD screen displays CL, press the On-off key to confirm, CL flashes 3 times to clear all memories, and the buzzer beeps 3 times continuously

Exceptional Situation

Andesfit-Health-UFR103-Infrared-Forehead-Thermometer-
\(14\) Andesfit-Health-UFR103-Infrared-
Forehead-Thermometer- \(15\)

Please contact the distributor if you can not solve the problem, do not disassemble the unit by yourself!

Care And Maintenance

Care for the main unit
Keep the unit in the storage case when not in use.

  • Clean the unit with a soft dry cloth. Do not use any abrasive or volatile cleaners.
  • Never immerse the unit or any of its’ component in water.

Maintenance

  • Do not clean the unit with naphtha, thinner or gasoline etc.
  • Store the unitin a clean and dry location, not subject the unit to extreme hot or cold temperature, humidity and direct sunlight.Andesfit-Health-UFR103-Infrared-Forehead-Thermometer- \(16\)
  • Remove the batteries if the unit will notbe used in 3 months or longer.
  • Do not use the unit under the influence of electromagnetic interference (nearby cellphones, microwave etc.) Andesfit-Health-UFR103-Infrared-Forehead-Thermometer- \(17\)

Note: We will not be responsible for any quality problem if you do not care and maintain the product as instructed. Specification

Specification

Andesfit-Health-UFR103-Infrared-Forehead-Thermometer-
\(20\)Andesfit-Health-UFR103-Infrared-
Forehead-Thermometer- \(21\)

Clinical Measurement Accuracy And Safety Verification
The product has passed clinical trials. The measured results of the infrared forehead thermometer was compared with the measured results of mercury thermometers, the deviation average At, = 0.011°C not exceeding 0.3°C; the clinical repeatability of the infrared forehead thermometer SR =0.100°C, not exceeding 0.3°C. The measured results up to the laboratory standard and the clinical standard.Therefore, the deviation average and the clinical repeatability of the infrared forehead thermometer are complied with the regulatory requirement ISO 80601-2-56. The conclusions are drawn from the clinical trials, the accuracy and safety are complied with the regulatory requirement.

Warranty Information

  • The unit is guaranteed to be free of defects in workmanship and materials under normal use for a period of 1 Year from the date of purchase.
  • For repair under this warranty. Our authorized service agent must be advised of the fault within the period of the warranty. This warranty only covers parts and labor service under normal operations. Any defect resulting from natural causes, eg. flood, hurricane etc, is not covered in this guarantee. This guarantee also does not cover damage incurred by use of the unit notin accordance with the instructions, accidental damage, or being tampered or serviced by unauthorized service agents.
  • The fllowing will be excluded from this warranty-if the thermometer has been misused, abused, or neglect in following the manual’s instructions on purpose and unauthorized repair or modifications.
  • The device requires no calibration.
  • The device is not repairable and contains no user serviceable parts.

EMC

IEC 60601-1-2: 2014 ME EQUIPMENT and ME SYSTEMS identification, marking and documents for Class B product Instructions for use The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environments and so on.

Warning: Don’t near active HF surgical equipment and the RF shielded room of an ME system for magnetic resonance imaging, where the intensity of EM disturbances is high.

Warning: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally. Warning: Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and resuit in improper operation.”
Warning: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Infra-red Ear Thermometer (TE-66, TE-68, TE-82), including cables specified by the manufacturer. Otherwise, degrada- tion of the performance of this equipment could result.

If any : a list of all cables and maximum lengths of cables (if applicable), transducers and other ACCESSORIES that are replaceable by the RESPONSIBLE ORGANIZATION and that are likely to affect compliance of the ME EQUIPMENT or ME SYSTEM with the requirements of Clause 7 (EMISSIONS) and Clause 8 (IMMUNITY). ACCESSORIES may be specified either generically (e.g. shielded cable, load impedance) or specifically (e.g. by MANUFACTURER and EQUIPMENT OR TYPE REFERENCE). If any: the performance of the ME EQUIPMENT or ME SYSTEM that was determined to be ESSENTIAL PERFORMANCE and a description of what the OPERATOR can expect if the ESSENTIAL PERFOR- MANCE is lost or degraded due to EM DISTURBANCES (the defined term “ESSENTIAL PERFOR- MANCE” need not be used).

Technical description

  1. all necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFORMANCE with regard to electromagnetic disturbances for the excepted service life.
  2. Guidance and manufacturer’s declaration -electromagnetic emissions and Immunity

Table 1Andesfit-Health-UFR103-Infrared-Forehead-Thermometer-
\(22\)

Table 2Andesfit-Health-UFR103-Infrared-Forehead-Thermometer-
\(23\) Table 3Andesfit-Health-UFR103
-Infrared-Forehead-Thermometer- \(24\) Andesfit-Health-UFR103-Infrared-Forehead-Thermometer-
\(25\)

Manufacturer
Shenzhen Urion Technology Co.,Ltd. Floor 4-6th of Building D, Jiale Science&Technology Industrial Zone, No.3, ChuangWei Road, Heshuikou Community, MaTian Street, GuangMing New District, 518106 ShenZhen, PEOPLE’S REPUBLIC OF CHINA
Tel: (86) -755-29231308
E-mail: urion@urion.com.cn
MADE IN CHINA

Eu representative Shanghai
International Holding Corp. GmbH (Europe) Eiffestrasse 80, 20537 Hamburg, Germany
Tel: +49-40-2513175
E-mail: shholding@hotmail.com

AndesFit Itd
Unit 513 Lakeside 1, No.8 Science Park West Ave., Hong Kong Science Park, Hong Kong Customer Services: (852) 35951880 www.andesfit.com