Andesfit Health UFR103 Infrared Forehead Thermometer
Instruction manual Infrared Forehead Thermometer
Model:UFR103
Introduction
Your new Infrared forehead thermometer uses advanced infrared (IR) technology
to measure temperature instantly and accurately on the forehead or object.
Easy to use and less measurement time This thermometer does not need to
contact body or object to ensure the safety- and hygiene. Its ergonomic design
makes this thermometer be simple and very easy to check the temperature. It
only takes 1 second to take measurement and reading.
-
Body and object modes
This thermometer supports to measure temperature of body and object. The measuring range of object mode is 0.0 to 100.0°C (32.0°F~199.9°F ). That means except to take body temperature,it also supports to take the surface temperature of object as below, Surface temperature of milk in the bottle;- Surface temperature of baby bath;
- Environment temperature.
-
Alarm sounds
When body temperature is over 37.5°C, an alarm sounds for 10 times. A Memory function It stores up to 20 sets recent measurement data. -
Auto power off
Automatically power off if left idle for Approx 60$5s. -
Measurement
Measurement time interval in 1s and the measurement distance is within 1~3cm. Please read the manual carefully before you use the unit,and keep for future reference.
Intended use
The Infrared Thermometer is intended for the intermittent measurement and
monitoring of human body temperature from forehead or object. The device is
indicated for use by people of all ages at homecare and in hospital.
Safety Information
To assure the correct use of the product, basic safety measures should always be followed including the warning and the caution listed in the instruction manual
Symbol descriptions
The following symbols may appear in this manual, on the label, on the device,
or on it’s accessories. Some of the symbols represent standards and
compliances associated with the device and its use.
- This thermometer is not intended to substitute for a consultation with your physician.The Forehead scan Temperature serves as a reference only. It cannot be a judgment on fever.
- Basic safety precautions should always be observed, especially when the thermometer is used on or near children and disabled persons.
- Please place the device unreachable by young.
- Avoid direct sunlight.
- Do not touch the lens.
- No modification of this device is allowed.
- The swallowing of small parks like packing bag, battery,battery cover and so on may cause the suffocation.
- Please do not use a dilution agent, alcohol or petrol to clean the unit.
- Please A treat is gently and prevent the falling from a high place.
- Please do not immersed it in liquid. Never leave battery in the battery compartment for a long time without use, as they may leak and cause damage to the unit.
- Please take off the battery if you do not intend to use within 3 months. Replace with new batteries if the unit display a low battery symbol.
- Do not mix the old and new batteries together.
- Do not use during the transportation.
WARNING
Do not dispose of electrical appliances as unsorted municipal waste, use
separate collection facilities. Contact you local government for information
regarding the collection systems available. If electrical appliances are
disposed of in landfills or – dumps, hazardous substances can Teak into the
groundwater and get into the food chain, damaging your health and well-being.
Classification
- Internally powered equipment;
- Type BF applied part;
- Protection against ingress of water or Particulate matter: IP21;
- Not category AP /APG equipment;
- Mode of operation:Continuous operation.
The user must check that the equipment functions safely and see that it is in proper working condition before being used.
Product Structure
Display
- Object mode
- Forehead temperature mode
- Ambient temperature mode
- Temperature unit (°C/°F)
- Temperature reading
- Memory symbol
- Battery symbol
- Horn symbol
- Number of memory groups
- Bluetooth symbol
Body
Battery Installation
Remove the battery cover from the battery compartment, insert the battery.
1. Remove the battery cover from the battery compartment as the arrow
direction accordingly.
2. Insert 2 AAA powerful batteries into the compartment and ensure each
battery is in the proper direction, Positive (+) and Negative (-) are
displayed on the back of battery cover.
3. Close the battery cover.
Low battery and replacement
When power on, the low battery symbol will display once the unit start to
work, and you must replace with new batteries, otherwise the unit can’t work.
Battery type and replacement
Please use 2pcs AAA identical 1.5V alkaline batteries. Do not use the
batteries beyond their expiry date. Please remove the batteries if you do not
need to use for long time.
WARNING
Dispose of the battery in accordance with all federal, state and local
laws. To avoid fire and explosion hazard, do not burn or incinerate the
battery.
Setting Mode
How To Set
1. Mode setting: The display all shows for one second after turning on the
On-off key, the default setting is “Body mode”after turning on . Short press
the setting button once to switch to “Object mode”, and then press the setting
button again to switch to “Ambient temperature mode”. If you press the setting
button again, it will return to the “Body mode”. Note: The Body Mode is used to measure the forehead
temperature, and the Object Mode is used to measure the object temperature,
then the Ambient mode is used to measure the Ambient temperature.
2. Unit setting and sound switch settings: Long press the setting button for
3 seconds to enter the °C/°F and sound switch settings.
Step 1 : Enter °C / °F switch: the interface displays °C or °F, press the
memory key to switch the temperature unit, press the set key to confirm the
selection, at the same time enter the sound setting selection mode (Restart,
the unit will default to “°C”). Step 2 : Enter the sound switch setting:
the interface displays On or OFF, press the memory key to switch. If you
select on, the horn symbol is displayed on the screen. If you select OFF, the
horn symbol is not displayed on the side. Press the set button to confirm the
selection. The setting is completed automatically. After setting, it will
automatically return to the measure mode and wait for measurement. (Restart,
the sound will default to on)
Proper Use Of The Unit
Pre-measurement
About Normal Body Temperature&Fever The temperature in the forehead and temple
area differs from the internal temperature, which is taken orally or rectally.
Vasoconstriction, an effect that constricts the blood vessels and cools the
skin, can occur during the early stages of a fever.In this case, the
temperature measured by the Infrared thermometer may be unusually low. If the
measurement therefore does not match the patient’s own perception or is
unusually low, re peat the measurement every 15 minutes. As a reference, you
can also measure the internal body temperature using a conventional oral or
rectal thermometer. Body temperature can vary from one individual/person to
next. It also varies by location on the body and time of day. Below shows the
statistical normal ranges from different sites. Please keep in mind that
temperatures measured from different sites, even at the same time, should not
be directly compared. Fever indicates that the body temperature is higher than
normal. This symptom may be caused by infection, overdressing or immunization.
Some people may not experience fever even when they are ill.
These include, but are not limited to, infants younger than 3 months old,
persons with compromised immune systems, persons taking antibiotics, steroids
or antipyretics (aspirin, ibuprofen, acetaminophen), or persons with certain
chronic illnesses. Please consult your physician when you feel ill even if you
do not have fever.
*Table1 Body Site Normal Temperature Range**
Note: Body Temperature at WebMD; website: http://firstaid.webmd.com /body-temperature;retrieved at 2010 Jan 7.
As a Body Thermometer
Press the On-off button ,after the display is fully displayed for 1 second,
the measurement preparation screen is displayed. The C icon lights up. Align
the unit with the center of the forehead and let the infrared thermometer
probe be 1cm away from the forehead,measurement completion prompts measurement
results.(Buzzer: One long beep Voice)
Note:
- If the reading is≥37.5°C(99.5°F)and <43°C(109.4°F), the buzzer ten short beeps.
- After the measurement is completed, the measurement can be continued after 1 second.
- As the forehead measurement temperature is likely to be affected by sweat, oil and the surroundings, the reading shall be taken as a reference only.
- If the probe is placed at an angle close to the forehead measurement, the reading will be affected by surrounding temperature.
- Babies’ skin reacts very quickly in the ambient temperature. Therefore, do not take their temperature with the non-contact thermometer during/after breastfeeding, because the skin temperature maybe lower than the internal body temperature.
- The user and the infrared thermometer should be quiescence for at least 30 minutes under similar indoor conditions.
As a Object Thermometer
Press the On-off button, after the display is fully displayed for 1 second,
switch to the object mode and display the measurement preparation screen. The
°C icon will light up, Point the infrared thermometer at the object to be
measured, and press On-off button to measure, measurement completion prompts
measurement results.(Buzzer: One long beep Voice)
Note:
- As the object scan temperature is likely to be affected by the surroundings, the reading shall be taken as a reference only.
- If the probe is placed at an angle close the object scan, the reading will be affected by surrounding temperature.
As a Ambient Thermometer
Press the On-off button, after the display is fully displayed for 1 second,
switch to the ambient temperature mode, display the measurement preparation
screen, the C icon lights up, press the On-off key measurement, measurement
completion prompts measurement results. (Buzzer: One long beep Voice)
Memory-recall of measurements
When the device is off, press the “memory button” to enter the read memory
mode,the memory icon M and”the first group of memory values”appear, Each
time you press the key in sequence, 20 groups of memories are read in
sequence, and the first group of memories The value is the latest measurement
reading, and so on. When reading the 20th group of memory, if you press the
memory key again, the first group of readings will be re-read.
Memory-clear Of Measurements
In the shutdown state, long press the memory button for about 8 seconds, the
LCD screen displays CL, press the On-off key to confirm, CL flashes 3 times to
clear all memories, and the buzzer beeps 3 times continuously
Exceptional Situation
Please contact the distributor if you can not solve the problem, do not disassemble the unit by yourself!
Care And Maintenance
Care for the main unit
Keep the unit in the storage case when not in use.
- Clean the unit with a soft dry cloth. Do not use any abrasive or volatile cleaners.
- Never immerse the unit or any of its’ component in water.
Maintenance
- Do not clean the unit with naphtha, thinner or gasoline etc.
- Store the unitin a clean and dry location, not subject the unit to extreme hot or cold temperature, humidity and direct sunlight.
- Remove the batteries if the unit will notbe used in 3 months or longer.
- Do not use the unit under the influence of electromagnetic interference (nearby cellphones, microwave etc.)
Note: We will not be responsible for any quality problem if you do not care and maintain the product as instructed. Specification
Specification
Clinical Measurement Accuracy And Safety Verification
The product has passed clinical trials. The measured results of the infrared
forehead thermometer was compared with the measured results of mercury
thermometers, the deviation average At, = 0.011°C not exceeding 0.3°C; the
clinical repeatability of the infrared forehead thermometer SR =0.100°C, not
exceeding 0.3°C. The measured results up to the laboratory standard and the
clinical standard.Therefore, the deviation average and the clinical
repeatability of the infrared forehead thermometer are complied with the
regulatory requirement ISO 80601-2-56. The conclusions are drawn from the
clinical trials, the accuracy and safety are complied with the regulatory
requirement.
Warranty Information
- The unit is guaranteed to be free of defects in workmanship and materials under normal use for a period of 1 Year from the date of purchase.
- For repair under this warranty. Our authorized service agent must be advised of the fault within the period of the warranty. This warranty only covers parts and labor service under normal operations. Any defect resulting from natural causes, eg. flood, hurricane etc, is not covered in this guarantee. This guarantee also does not cover damage incurred by use of the unit notin accordance with the instructions, accidental damage, or being tampered or serviced by unauthorized service agents.
- The fllowing will be excluded from this warranty-if the thermometer has been misused, abused, or neglect in following the manual’s instructions on purpose and unauthorized repair or modifications.
- The device requires no calibration.
- The device is not repairable and contains no user serviceable parts.
EMC
IEC 60601-1-2: 2014 ME EQUIPMENT and ME SYSTEMS identification, marking and documents for Class B product Instructions for use The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environments and so on.
Warning: Don’t near active HF surgical equipment and the RF shielded room of an ME system for magnetic resonance imaging, where the intensity of EM disturbances is high.
Warning: Use of this equipment adjacent to or stacked with other
equipment should be avoided because it could result in improper operation. If
such use is necessary, this equipment and the other equipment should be
observed to verify that they are operating normally. Warning: Use of
accessories, transducers and cables other than those specified or provided by
the manufacturer of this equipment could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment and resuit
in improper operation.”
Warning: Portable RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the Infra-red Ear Thermometer (TE-66, TE-68,
TE-82), including cables specified by the manufacturer. Otherwise, degrada-
tion of the performance of this equipment could result.
If any : a list of all cables and maximum lengths of cables (if applicable), transducers and other ACCESSORIES that are replaceable by the RESPONSIBLE ORGANIZATION and that are likely to affect compliance of the ME EQUIPMENT or ME SYSTEM with the requirements of Clause 7 (EMISSIONS) and Clause 8 (IMMUNITY). ACCESSORIES may be specified either generically (e.g. shielded cable, load impedance) or specifically (e.g. by MANUFACTURER and EQUIPMENT OR TYPE REFERENCE). If any: the performance of the ME EQUIPMENT or ME SYSTEM that was determined to be ESSENTIAL PERFORMANCE and a description of what the OPERATOR can expect if the ESSENTIAL PERFOR- MANCE is lost or degraded due to EM DISTURBANCES (the defined term “ESSENTIAL PERFOR- MANCE” need not be used).
Technical description
- all necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFORMANCE with regard to electromagnetic disturbances for the excepted service life.
- Guidance and manufacturer’s declaration -electromagnetic emissions and Immunity
Table 1
Table 2 Table 3
Manufacturer
Shenzhen Urion Technology Co.,Ltd. Floor 4-6th of Building D, Jiale
Science&Technology Industrial Zone, No.3, ChuangWei Road, Heshuikou Community,
MaTian Street, GuangMing New District, 518106 ShenZhen, PEOPLE’S REPUBLIC OF
CHINA
Tel: (86) -755-29231308
E-mail: urion@urion.com.cn
MADE IN CHINA
Eu representative Shanghai
International Holding Corp. GmbH (Europe) Eiffestrasse 80, 20537 Hamburg,
Germany
Tel: +49-40-2513175
E-mail: shholding@hotmail.com
AndesFit Itd
Unit 513 Lakeside 1, No.8 Science Park West Ave., Hong Kong Science Park, Hong
Kong Customer Services: (852) 35951880
www.andesfit.com