mpowerinc UNI 321 Maintenance Free Single Gas Detectors User Guide
- June 1, 2024
- mpowerinc
Table of Contents
UNI 321 Maintenance Free Single Gas Detectors
“`html
Product Information
Specifications
-
Product Name: UNI 321 Maintenance-Free Single Gas
Detectors -
User’s Guide Version: 1.61 (May 2024)
-
Gas Name: CO and H2S
-
Operating Time Units: Months, Days, Hours
-
Features: LED Alarm, Alligator Clip, Remaining Operating Time
Display, Sensor, Buzzer
Product Usage Instructions
Read Before Operating
This manual must be carefully read by all individuals who have
or will have the responsibility of using, maintaining, or servicing
this product. The product will perform as designed only if it is
used, maintained, and serviced in accordance with the
manufacturer’s instructions.
General Information
User Interface: The device features LED alarms,
alligator clip, remaining operating time display, and various keys
for interaction.
Normal Operation: Full bump test or calibration
and setting other functions can only be performed using a UNI
Docking Box MP100T or the CaliCase MP300T1 by a qualified service
technician.
Normal Mode Operation
Turning the Unit On: Press and hold the Right
Key for 3 seconds until the LCD displays beeps and the green LED
flashes. Once turned on, the battery life countdown timer
begins.
Warm-up Sequence: After powering on, the unit
enters a warm-up and self-test sequence displaying firmware
version, high alarm, and low alarm limits.
Remaining Time Display: After the start-up
sequence, the unit enters normal mode and displays the remaining
operating time in months.
FAQ
Proper Product Disposal at End of Life
The product should be disposed of following local regulations
for electronic waste disposal.
“`
UNI 321
Maintenance-Free Single Gas Detectors
User’s Guide
Ver 1.61 May 2024
UNI 321 User’s Guide
Contents
Read Before Operating ………………………………………………………………………………………………..2 General
Information…………………………………………………………………………………………………….3
User Interface………………………………………………………………………………………………………3 Display
……………………………………………………………………………………………………………….3 Normal Operation
………………………………………………………………………………………………..3 Normal Mode Operation
………………………………………………………………………………………………4 Turning the Unit On
……………………………………………………………………………………………..4 Warm-up Sequence
………………………………………………………………………………………………4 Remaining Time Display
………………………………………………………………………………………4 Alarms
……………………………………………………………………………………………………………….. 4 Hibernation Mode for RT
Models………………………………………………………………………………….5 Configuration
Mode…………………………………………………………………………………………………….6 Entering and Exiting
Configuration Mode ………………………………………………………………6 Zero (fresh air) Calibration
……………………………………………………………………………………6 Set High and Low Alarm
Levels…………………………………………………………………………….6 mPower Suite
Software…………………………………………………………………………………………7 Maintenance
………………………………………………………………………………………………………………. 8 Replacing the Sensor
Filter……………………………………………………………………………………8 Specifications
…………………………………………………………………………………………………………….. 9 Sensor
Configurations…………………………………………………………………………………………..9 Alarm Signal
Summary……………………………………………………………………………………….10 Troubleshooting
……………………………………………………………………………………………………….. 11
1
UNI 321 User’s Guide
Read Before Operating
This manual must be carefully read by all individuals who have or will have
the responsibility of using, maintaining or servicing this product. The
product will perform as designed only if it is used, maintained and serviced
in accordance with the manufacturer’s instructions.
WARNING · Never operate the monitor when the cover is removed. · Remove the
monitor cover and battery only in an area known to be non-hazardous. · Use
only mPower’s lithium battery part number M500-0001-000 [1.17.02.0002] (3.6V,
2700mAH, AA size) or part No. ER14505 cell manufactured by EVE Energy Co.,
LTD. · This instrument has not been tested in an explosive gas/air atmosphere
having an oxygen
concentration greater than 21%. · Substitution of components will impair
suitability for intrinsic safety. · Substitution of components will void
warranty. · It is recommended to bump test with a known concentration gas to
confirm the
instrument is functioning properly before use. · Before use, ensure that the
colorless ESD layer on the display is not damaged or peeling.
(The loose blue or colorless protective film used for shipment may be
removed.) AVERTISSEMENT
· N’utilisez jamais le moniteur lorsque le couvercle est enlevé. · Retirer le
couvercle du moniteur et la batterie uniquement dans une zone connue comme
non dangereuse. · Utilisez uniquement la batterie au lithium de mPower, pièce
No. M500-0001-000
[1.17.02.0002] (3.6V, 2700mAH, taille AA) ou celle de EVE Énergie Cie., Lté,
pièce No. ER14505. · Cet instrument n’a pas été testé dans une atmosphère
explosive gaz / air ayant une concentration en oxygène supérieure à 21%. · La
substitution de composants compromettra l’aptitude à la sécurité intrinsèque.
· La substitution des composants annulera la garantie. · Il est recommandé de
tester avec un gaz de concentration connu pour confirmer que l’ins trument
fonctionne correctement avant de l’utiliser. · Avant l’utilisation, assurez-
vous que la couche ESD incolore de l’écran n’est pas endommagée ou épluchée.
(Le film protecteur bleu peut être enlevé.)
Proper Product Disposal at End of Life
The Waste Electrical and Electronic Equipment (WEEE) irective (2002/96/EC) is
intended to promote recycling of electrical and electronic equipment and their
components at end of life. This symbol (crossed-out wheeled bin) indicates
separate collection of waste electrical and electronic equipment in the EU
countries. This product may contain one or more nickel-metal hydride (NiMH),
lithium-ion, or alkaline batteries. Specific battery information is given in
this user guide. Batteries must be recycled or disposed of properly. At the
end of its life, this product must undergo separate collection and recycling
from general or household waste. Please use the return and collection system
available in your country for the disposal of this product.
2
UNI 321 User’s Guide
General Information
The UNI 321 is a disposable or maintenance-free version of the UNI single gas
detector. It detects H2S or CO in a selection of models for an operation life
of either 24 months (MP102) or 36 months (MP103). The detector has a large LCD
providing maximum readability in the field. The display shows the reamining
operating time (for concentration readout use UNI MP100). Six bright red LEDs
allow for quick alarm notification. Constructed of strong and durable
material, the UNI 321 is designed to be comfortable, yet drop-resistant.
User Interface
LED Alarm
Alligator Clip
Remaining Operating Time
Display
LED Alarm
Confirm Action/ Number Increase
Key
Power On-Off/ Move Cursor Key
Sensor
Buzzer
Display
1. Gas name, includes: CO and H2S 2. Remaining time unit: Months, Days, Hours
3. Question mark: Daily alarm test needed; Bump test
due; Perform menu function 4. OK: Accept menu entry; Unit status indicator 5.
Gas concentration unit (for alarm setting display) 6. Remaining operating time
(Months/Days/Hours) 7. Low battery warning
Normal Operation
Normal operation is limited to the following functions · Displaying the
remaining operating time (the unit cannot be turned off) · Displaying (and
logging) the Alarm Level if this concentration of gas is exceeded · Entering
Configuration Mode to perform a fresh air zero calibration · Replacing the
sensor filter if used in especially highly dusty or moist environments
Full bump test or calibration, and setting other functions can only be
performed using a UNI Docking Box MP100T or the CaliCase MP300T1 by a
qualified service technician.
3
UNI 321 User’s Guide
Normal Mode Operation
Turning the Unit On
NOTE: ONCE THE UNIT IS TURNED ON IT CANNOT BE TURNED OFF AND THE BATTERY LIFE COUNTDOWN TIMER BEGINS.
Press and hold the Right Key for 3 seconds, until the LCD displays beeps, and the green LED flashes.
, the buzzer
Warm-up Sequence
After powering on, the unit enters a warm up and self-test sequence, showing
the firmware
version
followed by the the High Alarm
and Low Alarm
limits. If a sensor is not installed or cannot be identified by the instrument, the screen
flashes between
and
.
Remaining Time Display
After the start-up sequence, the unit enters normal mode and displays the
remaining operating
time in months
. When <3 months remain, the display switches to remaining days, and
when <1 day remains it switches to remaining hours. Once the unit is running, it cannot be turned off manually and it stays on until the battery loses power or the unit life expires.
Alarms
Daily Alarm Test (Flashing “?”) Before each day’s use, the ? question mark
will be flashing to remind the user to press the Left [OK] Key once to test
the function of the audio, visual and vibration alarms.
WARNING: IF ANY OF THE NEEDED ALARMS ARE NOT FUNCTIONING, DO NOT USE THE
INSTRUMENT AND RETURN IT FOR SERVICE.
Normal Alarm Function If the gas concentration exceeds an alarm limit, the
display shows the alarm value being
exceeded
and gives audio, visual and vibration alarms according to the table at the
end of this manual (see Alarm Signal Summary). Once the concentration is no longer in an alarm condition, the unit reverts back to the time display, but logs the alarm event in memory. The alarm condition must persist for more than a few seconds for it to be registered; shorter alarms may not be logged.
Calibration Fail Alarm (For Service Use Only) If the instrument fails calibration, it will alternately display
and the remaining run
time
, once per second. The user can perform a manual Zero Calibration (see below)
on the instrument alone, but a full span calibration requires a Docking Box or CaliCase.
4
UNI 321 User’s Guide
Bump Fail Alarm (For Service Use Only) If the instrument fails a Bump test using the Docking Box or CaliCase, it will alternately
display
and the remaining run time
once per second.
Bump Due Alarm (For Service Use Only) If the Bump Due setting is enabled and the due date has passed, the question mark will flash
twice per second while the display continues to show the remaining run time
.
Hibernation Mode for RT Models
Once the unit is turned on, it runs continuously until the remaining lifetime
ends. To save battery power when the unit is not used, it can be put into
Hibernation Mode by placing it into the cradle of the UNI docking box and
pressing the Bump key for 5 seconds. To exit from Hibernation Mode, press and
hold the Right Key on the unit for 3 seconds to wake it up again. Hibernation
Mode is for UNI 321 RT models MP101 and MP102 only.
5
UNI 321 User’s Guide
Configuration Mode
In the configuration mode the user can do a zero (fresh air) calibration and
set the high and low alarm thresholds. In general, use the Left Key to
increase the number or confirm, use the Right Key to move the cursor or move
to the next programming item. Detailed configuration settings can be performed
using the mPower Suite software through the UNI Docking Box MP100T or CaliCase
MP300T1 (see UNI manual or CaliCase manual).
Entering and Exiting Configuration Mode
Press and hold the Left Key and the Right Key simultaneously for 3 seconds.
The unit
prompts for a password by displaying
and
, with the first digit flashing. The
UNI 321 preset password is 0000. Use the Left [OK] Key to increase the number, and the Right [ ] Key to move the cursor. Once all four digits are entered, the cursor will move to “OK”, causing it to flash. Press the Left Key to accept and enter the Configuration Mode.
To Exit Configuration Mode, scroll through the menu using the Right [ ] Key until EXIT? is shown, and press Left [OK].
Zero (fresh air) Calibration
Zero (Air) calibration is to set the base line for the sensor. It is done in
fresh air or other
source known to be free of detectable gas (For an O2 unit, “zeroing” sets the
value to 20.9%,
so air must be used). When the LCD displays
, press the Left Key to begin the air
calibration. The unit will start a 15 second count-down, and then display the calibration result
as either pass
or fail
. The user can abort the zero calibration during the 15
seconds count-down by pressing the Right Key, after which
is displayed.
Set High and Low Alarm Levels
After Air calibration is completed or skipped, the unit displays
to set the High Alarm
limit.
· Press Left to enter the menu. The current setting value is displayed,
with the first digit flashing:
· Use the Left Key to increase the current digit, cycling from 0 to 9:
· Use the Right Key to move the cursor to the next digit:
· After all digits are entered, use the Right Key to move to the “OK” symbol,
and press the Left Key to save the entry. The unit will display
SAVE for a few seconds while storing the value but it is not necessary to
press OK to initiate saving.
· The display automatically moves to set the Low Alarm limit
Adjust the Low Alarm in the same way as for the High Alarm.
6
UNI 321 User’s Guide
4-bay CaliCase
UNI Docking Box
mPower Suite Software
mPower Suite is the PC software used to read logged alarm events and set up
extra configuration parameters of the UNI 321. To communicate with the UNI
321, an IR reader, UNI Single Docking Box MP100T or CaliCase System MP300T1
are required. For operation details see the UNI MP100 Manual.
Configuration Set-up Using mPower Suite 7
UNI 321 User’s Guide
Maintenance
NOTE: The UNI 321 is designed as a disposable instrument and does not need
maintenance under normal circumstances. However, in highly dusty or wet
environments, it may be necessary to replace the sensor filter as described
below.
Maintenance should be performed only by a qualified person who has proper
training and fully understands the contents of the manual.
Replacing the Sensor Filter
The filter may need replacement under special circumstances such as use high-
dust or condensing environments. External filter clips are available to reduce
the need to replace the internal filter. Sheets of 5 “peel-and-stick” internal
filters are available for this purpose: 1) Turn off the UNI 321 and place it
face down on a soft surface. 2) Use a T10 Torx screwdriver to loosen
(counterclockwise) each of the four screws. 3) Remove the top cover after
carefully unplugging the buzzer connector. 4) Peel a filter from the sheet and
center it over the sensor. Gently press down. 5) Re-connect in the buzzer
connector and re-install the top cover. 6) Re-install the screws in the back
cover. Be careful to not overtighten the screws.
External Filter Clip
Sensor
Filter
Battery CAUTION · Change battery only in area known to be non-hazardous. · Use
only mPower battery, PN: 1.17.02.0002 or PN ER14505 cell manufactured by EVE
Energy Co., LTD. ATTENTION · Changez la batterie uniquement dans une zone
connue pour être non dangereuse. · Utilisez uniquement la batterie au lithium
de mPower, pièce No. 1.17.02.0002 (3.6V, 2700 mAH, taille AA) ou celle deEVE
Énergie Cie., Lté, pièce No. ER14505.
8
Specifications
UNI 321 User’s Guide
Sensor Configurations
Sensor
Range (ppm)
Resolution (ppm)
t90(s)
CO 0-500
1
15
H2S 0-100
0.1
15
Span (ppm)
50
25
Low Alarm (ppm) 35
10
High Alarm (ppm) 200
15
CAUTION: Use only mPower sensors.
9
UNI 321 User’s Guide
Alarm Signal Summary
LCD Display
Reason/Alarm Signals
Low alarm Buzzer 2 beeps per second LED 2 flashes per second 1 vibration per
second
“LOW” 2 flashes per second High alarm Buzzer 3 beeps per second LED 3 flashes
per second 1 vibration per second
“HIGH” 2 flashes per second Over-range Buzzer 3 beeps per second LED 3 flashes
per second 1 vibration per second “OVER” and “500” 1 flash per second
Daily Alarm Check needed or Bump Due alarm: ? 2 flashes per second
Bump Fail alarm: Buzzer 1 beep per minute LED 1 flash per minute
1 vibration per minute “Bump” and “Remain Time” alternate display per second
Calibration Fail alarm: Buzzer 1 beep per minute LED 1 flash per minute
1 vibration per minute “CAL” and “Remain Time” display alternately once per
second
10
UNI 321 User’s Guide
Battery Low alarm Battery Empty prompt
Buzzer 1 beep per minute LED 1 flash per minute
1 vibration per minute Battery fail alarm Buzzer 1 beep per second LED 1 flash
per second “bAT LoW”1 flash per second
Sensor error Buzzer 1 beep per second LED 1 flash per second “SEN Err”1 flash
per second
Troubleshooting
Problem
Cannot turn on unit Weak or No Response
Possible reason
Depleted or defective battery Filter(s) plugged
Buzzer, LED, or vibration Bad buzzer, LEDs, or vibration alarm. alarm
inoperative
Blocked alarm port
Solution
Call authorized service center Replace external filter clip or internal filter
Call authorized service center
Clean or unblock alarm port
11
UNI 321 User’s Guide
mPower Electronics, Inc. 2910 Scott Boulevard, Santa Clara, CA 95054
Phone: (408)320-1266 Fax: (669)342-7077 info@mpowerinc.com www.mpowerinc.com
12
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>