CLARIO 2AAUFISPIRO1 iSpiro Ultrasonic Sensor Portable Spirometer Instruction Manual

June 1, 2024
CLARIO

CLARIO-logo

CLARIO 2AAUFISPIRO1 iSpiro Ultrasonic Sensor Portable Spirometer

CLARIO-2AAUFISPIRO1-iSpiro-Ultrasonic-Sensor-Portable-Spirometer-product-
image

Product Information

Specifications

  • Model Number: 782200USEN
  • Version: 01.00
  • Date: 28 FEB 2024

Introduction
The product is a medical device designed for measuring and monitoring respiratory parameters.

What’s in the Box?

  • Main device
  • User manual
  • Batteries
  • Cleaning and disinfection agents

Indications for Use
The product is intended for use by healthcare professionals to assess respiratory function in patients.

Operation

  • Operation Conditions
    Ensure the device is operated within the specified temperature and humidity range.

  • Setting Up Your Device
    Follow the step-by-step instructions in the user manual to set up the device correctly.

  • Device Indicators
    Familiarize yourself with the different indicators on the device and their meanings as described in the manual.

  • Turn On and Off
    Press and hold the power button to turn on/off the device.

  • Understanding Test Quality
    Refer to the user manual for guidelines on interpreting test results and ensuring quality measurements.

Maintenance

  • Performance and Calibration
    Regularly calibrate the device as per the recommended schedule to maintain accuracy.

  • Cleaning and Disinfection Procedure
    Follow the specified cleaning and disinfection procedure to ensure proper hygiene standards.

  • Cleaning the SpiroWay
    Use the recommended cleaning agents to clean the SpiroWay component of the device.

  • Batteries
    Replace batteries following the instructions provided in the manual when needed.

  • Troubleshooting
    If you encounter any issues with the device, refer to the troubleshooting section of the manual for guidance.

FAQ

  • Q: How often should I calibrate the device?
    A: It is recommended to calibrate the device at least once every six months or as per specific guidelines provided in the manual.

  • Q: Can I use any cleaning agent for disinfecting the device?
    A: No, only use the recommended cleaning and disinfection agents specified for this product to ensure proper maintenance and safety.

INTRODUCTION

Product Description
The iSpiro Ultrasonic Sensor is a portable spirometer that is capable of pairing via Bluetooth with smart devices to perform a spirometry test. The iSpiro Ultrasonic Sensor detects the volume and flow moving through the device to record spirometry data. The device is powered by standard 2 x AAA Alkaline batteries. The iSpiro Ultrasonic Sensor is used in combination with the SpiroWay® mouthpiece.

What’s in the box?
Your iSpiro Ultrasonic Sensor box contains:

  • iSpiro Ultrasonic Sensor Device (b)
  • SpiroWay® (a)
  • iSpiro Ultrasonic Sensor Cap (c)
  • Carrying pouch (d)
  • Instructions for use
  • Quick-Start Guide
  • AAA BatteriesCLARIO-2AAUFISPIRO1-iSpiro-Ultrasonic-Sensor-Portable-Spirometer-fig- \(1\)

Caution
Please check that there is no visible damage on any device components. If any damage is present, do not use or attempt to repair the device but contact the manufacturer directly.

Indications for Use
The iSpiro Ultrasonic Sensor is intended to be used by adults and children over 5 years old in physician’s offices, clinics and home setting to conduct basic lung function and spirometry testing.

Restrictions, Contraindications and Possible Adverse Effects

  • Diagnosis of medical conditions or prescription of treatments can only be made by a qualified healthcare professional who may use results obtained with the iSpiro Ultrasonic Sensor as adjunct information when performing a full medical examination that has taken into consideration your clinical history and other test results.
  • The iSpiro Ultrasonic Sensor must be used by a single user at any one time. If the iSpiro Ultrasonic Sensor will be transferred to and used by a new user, it must be cleaned and disinfected according to the information given in this user manual before use.
  • A SpiroWay® mouthpiece must not be shared between users, including family members. Spirometry tests should only be performed if you are not experiencing any shortness of breath, are in good health and capable of performing a lung function test. Test results may otherwise be unreliable.
  • Failure to perform the required breathing  maneuver correctly during a test may lead to inaccurate and unacceptable results. More information about how to perform a spirometry test correctly is described in this user manual. The device should not be used if test accuracy and/or reliability is jeopardized by these or other external factors.
  • Spirometry tests can be physically demanding. The forced expiratory maneuver used in spirometry increases intrathoracic, intraabdominal, and intracranial pressures. Potential risks of spirometry are primarily related to maximal pressures generated in the thorax and their impact on abdominal and thoracic organs, venous return and systemic blood pressure, and expansion of the chest wall and lung. The physical effort required can increase myocardial demand. Caution must be used if you  have medical conditions that could be adversely affected by these physiological consequences. Although such risks are likely to be minimal for spirometry in most patients, the potential risks associated with testing should always be weighed against the benefit of obtaining information about lung function. Spirometry should be discontinued if you experience pain during the maneuver. If you have any of these potential contraindications, please seek spirometry testing in primary care settings or pulmonary function laboratories where you will be under the supervision of healthcare professionals and there may be access to emergency care if needed.
  • Patients aged 5 years old and over and physicians are the intended operators of the iSpiro Ultrasonic Sensor device. A competent adult should assist patients (children or older patients) who may need assistance.
  • Relative Contraindications for Spirometry.
  • Due to increases in myocardial demand or changes in blood pressure;
    • Acute myocardial infarction within 1 week Systemic hypotension or severe hypertension Significant atrial/ventricular arrhythmia Non compensated heart failure Uncontrolled pulmonary hypertension Acute core pulmonale
    • Clinically unstable pulmonary embolism History of syncope related to forced expiration/cough
  • Due to increases in intracranial/intraocular pressure;
    • Cerebral aneurysm
    • Brain surgery within 4 weeks
    • Recent concussion with continuing symptoms
    • Eye surgery within 1 week
  • Due to increases in sinus and middle ear pressures;
    • Sinus surgery or middle ear surgery or infection within 1 wk
  • Due to increases in intrathoracic and intraabdominal pressure;
    • Presence of pneumothorax
    • Thoracic surgery within 4 weeks
    • Abdominal surgery within 4 weeks
    • Late-term pregnancy
  • Infection control issues;
    • Active or suspected transmissible respiratory or systemic infection, including tuberculosis
    • Physical conditions predisposing to transmission of infections, such as hemoptysis, significant secretions, or oral lesions or oral bleeding
    • If you have or suspect having any of the conditions above, consult your healthcare professional before using the iSpiro Ultrasonic Sensor.

OPERATION

Mode of Operation: There are no minimum or maximum limits to the period of device use (‘on’ time) or disuse (‘off’ time). The device will time-out and automatically switch off if it is not actively used for a period of time. Batteries of the device should be removed if it will not be used for more than a month.

Operation Conditions
The required operation conditions for the iSpiro are:

  • Temperature: +15°C to +35°C
  • Relative Humidity: 30% to 85%
  • Pressure: 700 hPa to 1060 hPa

The iSpiro should only be used within the ambient temperature, relative humidity and ambient pressure ranges given above. The device should remain within this range for at least 1 hour before use.

Storage / Transport Environment
The required storage conditions for the iSpiro are:

  • Ambient temperature: -20°C to +60°C
  • Relative Humidity: 5% to 85%
  • Pressure: 700 hPa to 1060 hPa
  • The required transport conditions for the iSpiro Ultrasonic Sensor are:
  • Ambient temperature: -20°C to +60°C
  • Relative Humidity: 5% to 85%
  • Pressure: 700 hPa to 1060 hPa

The iSpiro should not be used in the presence of inflammable liquids or detergents, nor in the presence of inflammable anesthetic gases (oxygen or nitrogen).

Setting Up Your Device
ASSISTANCE: If you need assistance setting up, using or maintaining your iSpiro please contact your doctor, study investigator or Customer Care.

  1. Remove the cap from the iSpiro Ultrasonic Sensor device. 
  2. Slide open the battery cover, insert the AAA alkaline batteries in the correct orientation, slide the battery cover back to the closed position.CLARIO-2AAUFISPIRO1-iSpiro-Ultrasonic-Sensor-Portable-Spirometer-fig- \(3\)
  3. Take the SpiroWay out of the bag.
  4. Insert the SpiroWay® mouthpiece into the spirometer as shown below and keep pushing until you hear a click.
  5. Open the device by pressing the button.CLARIO-2AAUFISPIRO1-iSpiro-Ultrasonic-Sensor-Portable-Spirometer-fig- \(6\)

Device Indicators
There are 3 LED lights located on the front of the device. The LED lights may be turned on or flashing various colors in various patterns. The LED lights indicate the current status of the device. Please see the following information for guidance on LED light indications.

Table 2: Device Led Indicators

LED Display| Indication/s
None of the LEDs are on| The device is switched off
LED indicators are consecutively flashing green.| The device is switching on
LED number 3 is constantly flashing green.| The device is switched on
LED number 2 is fading blue.| The device is connected to the app. Bluetooth connection has been established.
LED number 2 and LEDs 1 and 3 together are flashing yellow in turn.| The zero flow level adjustment is in progress.
LED number 1 is constantly blue.| The device is ready for a test.
During a test, LED number 2 is constantly flashing blue.| The test has timed- out (there has been no inhalation/exhalation over a while)
All LEDs are flashing red.| The zero flow level adjustment setup has been unsuccessful. (Mouthpiece is removed)
All LEDs are flashing red.| There is a foreign object between the sensors. (Check device error in troubleshooting section)
LEDs are consecutively flashing yellow.| Over-the-air connection is being established.
LED number 3 flashes red three times.| Battery low warning.
LEDs flash in reverse order and remain switched off.| The device is switching off.

Turn on and off

  1. To connect the iSpiro Ultrasonic Sensor with your device press the power key button on the iSpiro Ultrasonic Sensor.
  2. Before starting a spirometry test with a patient, ensure that the spirometer is connected to the smart device.
  3. After completing your spirometry session, switch off the device by pressing the power key. Use the cap to protect the mouthpiece from contamination when the device is not in use.

How to perform a Lung function test
For instructions on how to perform the maneuver, please see the related Instructions for Use of the mobile application. Lung function test should be conducted according to ATS standards (see Graham B, Steenbruggen I, Miller M, et al. Standardization of spirometry 2019 update. An official American Thoracic Society and European Respiratory Society technical statement. Am J Respir Crit Care Med. 2019; 200:e70-e88).

MAINTENANCE

Handle the iSpiro Ultrasonic Sensor and SpiroWay mouthpiece with care. Do not use the device or its accessories if they are visibly damaged, particularly if there is damage to the filters on the mouthpiece or deformation of the mouthpiece itself. Store the iSpiro Ultrasonic Sensor and SpiroWay in dust-, dirt-, and moisture-free conditions. Before each use, always check that the device is free from contaminants and does not have any visible damage.

Performance and calibration
The iSpiro Ultrasonic Sensor is provided to the user factory-calibrated and does not require routine re-calibration. If you suspect performance errors in the iSpiro Ultrasonic Sensor, cease use of the device and test your lung function in a clinical setting to verify the results . Devices with confirmed performance errors should be returned to Clario.

Cleaning and disinfection procedure
You should clean the iSpiro Ultrasonic Sensor body and cap at least once a week or whenever the device is visibly contaminated.
You MUST perform the cleaning step before performing the disinfection step. Cleaning will prevent the physical buildup of contaminants on device surfaces and remove larger debris. Disinfection kills and destroys pathogens such as bacteria, viruses, or other microorganisms which might still be present on device surfaces after initial cleaning.

  1. Wash Hands
    Before beginning the procedure, wash hands thoroughly with soap and water .CLARIO-2AAUFISPIRO1-iSpiro-Ultrasonic-Sensor-Portable-Spirometer-fig-
\(11\)

  2. Perform Cleaning
    First, remove the SpiroWay® from the iSpiro Ultrasonic Sensor device. Using a disinfectant wipe, wipe all accessible surfaces of the device and cap using moderate pressure (as shown) for at least 30 seconds to remove contaminants. Be gentle and use care when wiping the sensors to avoid damaging them.

  3. Caution
    Prevent any excess liquids contained within the wipes from entering the components of the iSpiro Ultrasonic Sensor. Never immerse the product in water or any other liquid solution.

  4. Perform disinfection
    To disinfect the device and cap use a fresh disinfectant wipe to wipe over all accessible surfaces again using moderate pressure and for the contact time recommended by the wipe manufacturer.

  5. Wash Hands
    Wash hands thoroughly after performing a cleaning or disinfection procedure, and before handling the cleaned or disinfected components again for packing and storage.

Cleaning and disinfection agents
The following cleaning/disinfectant agent/s may be used to clean and disinfect this device: CaviWipes™ disinfecting towelettes (Active ingredients: Isopropanol (17.2%) and Ammomium Chloride), which can be purchased at:

Cleaning the SpiroWay
To clean the SpiroWay® Mouthpiece once a week and whenever visibly soiled;

  • Add dishwashing detergent (e.g. those containing 5-15% anionic surfactant, 5% nonionic surfactant) to warm water to create a soapy solution.
  • Shake the mouthpiece gently in the soapy solution.
  • Hold the mouthpiece under running tap water to rinse, do not rub
  • Leave the mouthpiece upright on a clean lint-free cloth at room temperature until it is completely dry

Do not insert the SpiroWay® mouthpiece into your iSpiro Ultrasonic Sensor device until it is completely dry
The SpiroWay® should be replaced every 3 months.

  • Danger

    • The SpiroWay® mouthpiece must be replaced and the device thoroughly disinfected if user has used the mouthpiece whilst having or suspected of having a dangerous infectious disease.
    • The SpiroWay® must be replaced immediately if the filters are damaged or whenever a risk of contamination is suspected.
  • Warning
    Risk of Cross-Contamination
    The SpiroWay® mouthpiece is indicated for single-patient-use only to prevent any potential of cross-contamination.

  • Danger
    Thorough cleaning and disinfection of the iSpiro Ultrasonic Sensor device must be performed prior to the dedication of the iSpiro Ultrasonic Sensor device to a new user. A new mouthpiece must be used by the new user.

Batteries
The iSpiro Ultrasonic Sensor device operates with 1.5V AAA Alkaline batteries. The battery life of the iSpiro Ultrasonic Sensor is approximately 12-18 months, assuming daily use of the device. The battery charge level is continuously monitored by the device. The device will not turn on if the battery charge level is low and will make a beeping sound to notify you.
The batteries of the device should be removed if the device is not going to be used for more than a month.

Instructions for battery replacementCLARIO-2AAUFISPIRO1-iSpiro-
Ultrasonic-Sensor-Portable-Spirometer-fig- \(15\) CLARIO-2AAUFISPIRO1-iSpiro-Ultrasonic-Sensor-
Portable-Spirometer-fig- \(16\) Disposal of iSpiro Ultrasonic Sensor
This product is not to be discarded as regular household waste but should be discarded as electronic waste in accordance with local regulations and returned to a collection point of recycling for electric and electronic devices.
Used batteries should be disposed of in designated battery recycling containers in accordance with local laws and regulations.

TROUBLESHOOTING

Problem Cause Solution
Device not turning on Multiple possible causes Check battery orientation and

correct polarities
Remove the AAA batteries, wait 30 seconds and reinstall AAA batteries
Replace AAA batteries
Check that battery cap is in the lock position, or if the cap is broken, contact manufacturer
iSpiro Ultrasonic Sensor cannot connect to

a smart device via Bluetooth®

| The smart device is out of range| Bring your smart device closer to the iSpiro® device
Smart device Bluetooth® is disabled| Enable Bluetooth® on your smart device
Bluetooth® connection not working properly| Your smart device will need Bluetooth® version 4.0 or higher. Find and select iSpiro Ultrasonic Sensor from the list of detected devices.
Test results are inconsistent| SpiroWay® mouthpiece is dirty| Clean SpiroWay® to ensure that the lumen is not obstructed or replace with a new mouthpiece
SpiroWay® mouthpiece is damaged| Replace SpiroWay®
SpiroWay® mouthpiece is installed incorrectly| Refer to the user manual for proper installation of SpiroWay®
The test does not start – Cannot set up zero flow level adjustment| Direct air flow through on the iSpiro Ultrasonic Sensor device| Close the cap of the iSpiro Ultrasonic Sensor to avoid effects of environmental flow
Place the device on a flat surface
Remove causes of direct air flow e.g. air conditioner, opened window, fan, etc.
Test starts before you start blowing| Rough handling of the device| Keep the device as stable as possible after starting a test
Device disconnected during test| The device is turned off accidentally or due to rough handling during use| Switch the device on again and proceed with a new test
Bluetooth® connection disrupted| Reconnect the device and proceed with a new test

For any other technical queries please call Clario´s Customer Care.

SIGNS AND SYMBOLS

The following signs and symbols provided for the safe use and storage of your iSpiro Ultrasonic SensorCLARIO-2AAUFISPIRO1-iSpiro-Ultrasonic-Sensor-
Portable-Spirometer-fig- \(17\) CLARIO-2AAUFISPIRO1-iSpiro-Ultrasonic-Sensor-
Portable-Spirometer-fig- \(18\)

TECHNICAL FEATURES

Summarized technical data

Flow / Volume Measurement Method Ultrasonic pulse transit-time measurement
Power Supply 2 x 1.5V AAA Alkaline or rechargeable batteries
Dimensions 110 x 63 x 41 mm
Weight (With batteries) 90 g
Weight (Without batteries) 67 g
Flow range 0 – 14 L/s
Maximum volume measured 10 L
Volume accuracy (Average) ± 2.50 % or 0.05 L
Flow Accuracy (Average) ± 10.00 % or 170 mL/s
Highest Expiratory Impedance* 48.54 Pa*s/L
Volume resolution 1 mL
Flow resolution 1 mL/s
Medical device class Class IIa
Wireless connection BLE 4.2

*Tested according to ISO26782 Annex B

Device Parameters
The iSpiro Ultrasonic Sensor records the following spirometry data:

Parameters Definition Unit
FVC Forced Vital Capacity — The volume of air that can forcibly be blown out
after full inspiration L
FEV0.75 Forced Expiratory Volume within 0.75 seconds: The volume of air that
can forcibly be blown out within 0.75 seconds, after full inspiration. L
FEV1 Forced Expiratory Volume within 1 second L
FEV3 Forced Expiratory Volume within 3 seconds L
FEV6 Forced Expiratory Volume within 6 seconds L
FEV0.75/FVC The ratio of FEV0.75 to FVC
FEV1/FVC The ratio of FEV1 to FVC
FEV3/FVC The ratio of FEV3 to FVC
FEV6/FVC The ratio of FEV6 to FVC
PEF Peak Expiratory Flow — The maximal flow rate achieved during the
maximally forced expiration initiated at full inspiration. L/s
MMEF Mean Mid-Expiratory Flow — synonymous with FEF25-75 L/s
FEF25 Forced Expiratory Flow at 25% of vital capacity — synonymous with
MEF75 L/s
FEF50 Forced Expiratory Flow at 50% of vital capacity — synonymous with
MEF50 L/s
FEF75 Forced Expiratory Flow at 75% of vital capacity —synonymous with MEF25

L/s
FEF25-75| Forced Expiratory Flow from 25% to 75% of vital capacity — synonymous with MMEF| L/s
MET25-75| Mid-Expiratory Time — synonymous with FET25-75| s
FEV0.75/FEV6| The ratio of FEV0.75 to FEV6| —
Parameters| Definition| Unit
---|---|---
FEV1/FEV6| The ratio of FEV1 to FEV6| —
FEF50/FVC| The ratio of FEF50 to FVC| 1/s
MMEF/FVC| The ratio of MMEF to FVC| 1/s
FET| Forced Expiratory Time| s
BEV| Back extrapolated volume| L
FIV1| The forced inspiratory volume within 1 second| L
FIVC| Forced inspiratory vital capacity| L
PIF| Peak inspiratory flow| L/s
FIF25-75| Forced inspiratory flow at 25% of vital capacity — synonymous with MIF75| L/s
FIV1/FIVC| The ratio of FIV1 to FIVC| —
R50 (FEF50/FIF50)| The ratio of flow at 50% of expiration and flow at 50% of inspiration — synonymous with FEF50/FIF50| —
VC| Vital capacity, from slow expiration| L
VCin| Inspiratory vital capacity, from slow inspiration| L
VCex| Expiratory vital capacity, from slow expiration| L
ERV| Expiratory reserve volume| L
IRV| Inspiratory reserve volume| L
IC| Inspiratory capacity from end of tidal breathing| L
Rf| Respiratory frequency| 1/min
VT| Tidal Volume| L
MVV| Maximum voluntary ventilation| L/min
MVV6| Maximum plat voluntary ventilation for 6 seconds| L/min
MVVtime| Duration of the trial in seconds| s

SAFETY WARNINGS AND PRECAUTIONS

Notes on Safety in this Instruction Manual
Following the ANSI (American National Standards Institute) recommendations for safety notes, specific passages of this instruction manual are clearly marked as safety notes.

Degree of Danger| Injury to persons| Damages to property| Meaning of Indicator
---|---|---|---
|

X

| | DANGER indicates an immediate hazardous situation, which, if not avoided, will result in serious injury or death. Limited to extremely dangerous situations.
|

X

| | WARNING indicates a potential hazardous situation, which, if not avoided, may result in serious injury or death.
|

X

|

(X)

| Caution indicates a potential hazardous situation, which, if not avoided, may result in minor or slight injury. Also used to indicate precarious procedures.

Additional safety symbols used in this user manual.

**** General Warning Sign
**** General mandatory action sign

List of safety warnings and precautions

| Special warning should be given by handlers of the device to elderly, pediatric or differently-abled users prior to use of the device.
---|---
| Do not use this device for any other purpose than its intended use to avoid potential dangers or harm to users. This device is not recommended for children under the age of 5.
| If any damage is present on the device or its components upon initial unboxing of the product then do not use the device as this may affect device performance and return it to the supplier.
| Take care when inserting the mouthpiece into the device to avoid physical injury and/or damage to the device.

| Do not use this device or its accessories if any parts are damaged, detached or blocked to avoid risk of choking or asphyxiation during use.
| In case of adverse events related to device use, immediately cease further use to avoid harm or damage and notify relevant local authorities and the manufacturer.
| Maximal inflation is unnatural; you may not have achieved it before, and it may seem somewhat uncomfortable.
| Regardless of the data presented on this device, cease testing if you feel unwell or have respiratory illness symptoms and contact your healthcare provider immediately.
---|---
| Medications and treatments based on device results must be directed only by a qualified doctor.
| For your safety, if there is an excessive decrease in your FEV1 value then cease use of the device and inform your healthcare provider.
| Do not perform more than 8 spirometry trials in one spirometry session. If you experience pain or sensations of dizziness during testing, cease device use and rest and inform your healthcare provider.
| Do not walk or run whilst performing a lung function test with the iSpiro Ultrasonic Sensor for your safety and to avoid result inaccuracies.
| Do not perform a spirometry test with food or objects in your oral cavity as this may lead to risk of choking.
| Do not share your iSpiro Ultrasonic Sensor or SpiroWay® with any other users, including family members, to prevent the transmission of infectious material. The device must be cleaned and disinfected, a new mouthpiece must be used, and a new account must be created for a new user of the device.
| Remove all batteries when the device will not be used for prolonged periods of time to prevent damage to the device through battery leakage or oxidation.
---|---
| To protect the environment, dispose of the product only through local EPA or WEEE collection facilities.
| Check the mouthpiece airway for foreign materials before each use to prevent risk of choking and to avoid measurement inaccuracies.
| Coughing or spitting into the device may cause incorrect readings.
| Keep the device dry. The limited Ingress Protection (IP22) rating of the iSpiro Ultrasonic Sensor device case will not prevent damage from water leaking into the case and damaging electronics.
| Do not use accessories not described in this user manual to prevent damage to the device, measurement inaccuracies or risk of affecting electromagnetic performance.
| Do not touch the filters on the mouthpiece and do not use if filters have been damaged. Damage to filters may result in measurement inaccuracies.
| Device accuracy can be affected by extremes of temperature, humidity and altitude. Use, store and transport your device only as specified in this user manual.
| The device may give inaccurate readings if operated in the presence of strong electromagnetic sources, such as electrosurgical equipment, magnetic resonance imaging devices or computed tomography (CT) equipment.
| Do not use the device in the presence of direct air currents (e.g. wind), sources of heat or cold, direct sun rays, or other sources of light or energy (such as heaters or radiators), dust, sand, or any other chemical substances as these conditions can affect the performance or the expected life of the iSpiro Ultrasonic Sensor device.
---|---
| If your device is damaged or malfunctioning, contact the manufacturer or distributor if purchased from a reseller. Unauthorized repairs will terminate the product warranty and may result in a faulty device.
| Follow all data security warnings and recommendations for your personal smart device as per it’s manufacturer’s instructions to protect your personal data.
| Do not share your iSpiro Ultrasonic Sensor account information with unauthorized parties. This will help protect your personal data.
| iSpiro Ultrasonic Sensor devices may be affected by electromagnetic interference. Please read EMC information and precautions provided in this user manual before use.
| Do not use iSpiro Ultrasonic Sensor devices in the presence of external portable and mobile radio frequency (RF) communications equipment as this may affect data transmission.
| Keep your iSpiro Ultrasonic Sensor device away from strong sources of magnetic and RF fields such as large electric motors, amateur radio transmitters, radar, anti-theft systems, stereo speakers, cell phones, and radio frequency identification (RFID). Television and radio transmitters could cause interference if the device is used close to them.
| Federal Law (USA) restricts this device to sale by or on the order of a physician.

Safety information regarding electromagnetic compatibility

  • The iSpiro Ultrasonic Sensor should not be operated at the same time as electrical devices with a high RF power output (e.g. HF surgical equipment) during intended use.
  • The iSpiro Ultrasonic Sensor may be affected by portable wireless communications equipment such as antennas, wireless home network devices, mobile phones, cordless telephones and their base stations, walkie-talkies, etc. A minimum distance of 30 cm (12 inches) should be kept between these and any part of the iSpiro Ultrasonic Sensor.
  • Use of accessories or components (e.g transducers) other than those specified by the manufacturer may result in increased emissions or decreased immunity of the iSpiro Ultrasonic Sensor. The iSpiro Ultrasonic Sensor should not be used adjacent to or stacked with other equipment. If this is unavoidable the configuration in which it is used should be monitored closely to ensure that the device continues to function normally.
  • If there are electromagnetic fields from other electrical devices nearby whilst the iSpiro Ultrasonic Sensor is being used, the device may fail to respond or not function as normal. To restart the iSpiro Ultrasonic Sensor, turn the device off and on again. If the problem persists, move the device away from sources of electromagnetic interference as described in the EMC compatibility section of this user manual.

IT Networks

  • Connection of the mobile application to an IT network that includes other equipment could result in previously unidentified risks to patients, operators, or third parties.

  • Subsequent changes to the IT network could introduce new risks and require additional analysis.

  • Changes to the IT network include:

    • changes in the IT network configuration,
    • connection of additional items to the IT network,
    • disconnecting items from the IT network,
    • update of equipment connected to the IT network, and
    • upgrade of equipment connected to the IT network.
  • It is the user’s responsibility to identify, analyze, evaluate, and control these risks.

  • IEC 80001-1:2021 provides guidance to address these risks.

ORDERABLE ACCESSORIES

  • SpiroWay® 4-Pack (Ref number: 892115)
  • iSpiro Ultrasonic Sensor Cap (Ref number: 707014)
  • iSpiro Ultrasonic Sensor Pouch (Ref number: 707013)

TERMS OF WARRANTY

  • The SpiroWay Mouthpiece is the applied part of the iSpiro Ultrasonic Sensor device. To purchase these accessories please contact your local distributors or Clario.

  • The iSpiro Ultrasonic Sensor hardware is guaranteed against manufacturing defects for a period of 24 months effective from the date of purchase, upon the provision of an invoice or sales receipt. The service life of the iSpiro Ultrasonic Sensor (includes spirometer, cap and carry pouch) and is 5 years from purchase. The service life of the Spiroway mouthpiece is 3 months from first use. There are no user serviceable parts in the iSpiro Ultrasonic Sensor product.

  • The customer must return goods for replacement or repair at the customer’s expense to the authorized supplier or manufacturer. The product must be returned with a clear written explanation of the fault or problem.

  • This warranty does not apply, at the discretion of the manufacturer, in the following cases:

  • Improper installation or operation of the device

  • Use of the product for purposes other than those specified in this user manual

  • Damage due to failure to follow instructions

  • Damage due to unauthorized repair,

  • modification or reconfiguration performed on the device
    Damage caused by falls, hit, lack of proper care or maintenance

  • Damage caused by abnormal physical or electrical stress or defects of the electric supply (battery cell) or of equipments

  • If the serial number is altered, deleted, removed or rendered illegible

  • In any case, the entire liability of İnofab Sağlık Teknolojileri A.Ş. under the provision of this agreement shall be limited to the amount paid by the customer for the product.

ELECTROMAGNETIC COMPATIBILITY

Meeting the requirements for EMC
(electromagnetic compatibility) and preventing the unsafe use of the device, medical devices including the iSpiro Ultrasonic Sensor manufactured by Clario conform to the EN60601-1-2 standard which defines the levels of immunity to electromagnetic interference as well as maximum levels of electromagnetic emissions for medical devices. For details, please see the following tables:

Table 1: Emission table for IEC 60601-1-2

Guidance and manufacturer’s declaration – electromagnetic emissions

iSpiro Ultrasonic Sensor battery-operated spirometer devices are intended for use in the electromagnetic environments specified below. Users of these devices should assure that it is used in such environments.
Emission Test| Compliance| Electromagnetic environment – guidance
RF emissions CISPR 11| Group 1| iSpiro Ultrasonic Sensor uses RF energy for its internal function. Its Radio Bluetooth, BLE module also complies with the national regulations. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. However, a separation distance of 30 cm shall be maintained.
RF emissions CISPR 11| Class B| The iSpiro Ultrasonic Sensor devices are suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2| Not applicable
Voltage fluctuations / flicker emissions IEC 61000-3-3| Not applicable| Emissions are not applicable because iSpiro Ultrasonic Sensor does not connect to mains supply but operates with AAA batteries.

Table 2: Immunity (Stimulation mode) table according to IEC 60601-1-2

Guidance and manufacturer’s declaration – electromagnetic immunity

iSpiro Ultrasonic Sensor battery-operated spirometer devices are intended for use in the electromagnetic environment specified below. Users of these devices should assure that it is used in such an environment.
Immunity Test Standard| IEC 60601 test level| Compliance level| Recommended separation distance
Electrostatic discharge (ESD) IEC 61000-4-2| ±8 kV contact

±2, ±4, ±8 kV, ±15 kV Air

| ±8 kV contact

±2, ±4, ±8 kV,

±15 kV Air

| Floor should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8| 30 A/m| 30 A/m| Power frequency magnetic fields should be at levels characteristic of a typical location in a typical domestic, commercial or hospital environment.
Guidance and manufacturer’s declaration – electromagnetic immunity

iSpiro Ultrasonic Sensor battery-operated spirometer devices are intended for use in the electromagnetic environment specified below. Users of these devices should assure that it is used in such an environment.
Immunity Test Standard| IEC 60601 test level| Compliance level| Recommended separation distance
|

  • 385 MHz
  • 27 V/m
  • 450 MHz
  • 28 V/m

| 27  V/m

28  V/m

| Portable and mobile RF communications equipment should be used no closer to any part of the iSpiro Ultrasonic Sensor devices including cables, than the recommended separation distance calculated from the equation appropriate to the frequency of the transmitter.
| 710, 745, 780 MHz
9 V/m| 9 V/m| Recommend separation distance d = 1.2 √P 80 MHz to 800 MHz
Radiated RF IEC 61000-4-3| 810, 870, 930, 1720, 1845, 1970, 2450 MHz 28 V/m| 28 V/m| d = 2.3 √P 800 MHz to 2.7 GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
| 5240, 5500, 5785 MHz 9 V/m| 9 V/m| Field strengths from fixed RF transmitters as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.b
| 80 MHz to 2.7 GHz 10 V/m| 10 V/m| Interference may occur in the vicinity of equipment marked with the following symbol:

Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: T hese guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
a Field strength from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones and land mobile radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the iSpiro Ultrasonic Sensor devices are used exceeds the applicable RF compliance level above, the iSpiro Ultrasonic Sensor device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the iSpiro Ultrasonic Sensor device.
b Over the frequency range 150 kHz to 80MHz, field strengths should be less than 10 V/m.

Recommended separation distances between portable and mobile RF communications equipment.

iSpiro Ultrasonic Sensor devices are intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. Users of iSpiro Ultrasonic Sensor devices can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the iSpiro Ultrasonic Sensor device as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter| Separation distance according to frequency of transmitter
80 MHz – 800 MHz d = 0.35 √P| 800 MHz – 2500 MHz d = 0.7 √P
0.01 W| 0.035 m| 0.07 m
0.1 W| 0.11 m| 0.22 m
1 W| 0.35 m| 0.7 m
10 W| 1.11 m| 2.21 m
100 W| 3.5 m| 7 m
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

Note: At 80MHz and 800MHz, the separation distance for the higher frequency range applies Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

BLUETOOTH WIRELESS COMMUNICATIONS

Bluetooth is used to pair the iSpiro Ultrasonic Sensor with smart devices and transfer test data between them. Bluetooth data transfer involves the widely trusted use of pre-shared key authentication and encryption algorithms. The strength of Bluetooth security depends heavily on the length and randomness of the passkey used for Bluetooth pairing. Initial pairing involves mutual authentication between devices where a link key is set up between them for later authentication and encryption. Encryption of data sent between devices prevents unintended or malicious interference. Each iSpiro Ultrasonic Sensor will connect to the smart device it is paired with and will not be confused with other Bluetooth radio communications.

Frequency Band: 2400 Mhz
Transmission Frequency Range: 2402 – 2480 Mhz
Max Output Power: 0.7 dBm
Antenna gain: 5.3 dB

FCC / IC NOTICE

USA
iSpiro Ultrasonic Sensor
FCC ID: 2AAUFISPIRO1

This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:

  1. This device may not cause harmful interferences, and
  2. This device must accept any interference received, including interferences that may cause undesired operation.

Canada
iSpiro Ultrasonic Sensor
Sensor Unit IC: 11335A-ISPIRO1
This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions:

  1. This device may not cause interference, and
  2. This device must accept any interference, including interference that may cause undesired operation of the device.

MANUFACTURER INFORMATION

  • UK Responsible Person: Exco Intouch Limited,
    6th Floor City Gate East, Tollhouse Hill, Nottingham, NG1 5FS UK

  • Swiss Responsible Person: PHT Corporation SarL, Chemin Louis-Hubert 2, 1213 Petit-Lancy, Geneva, Switzerland

©2024 eResearchTechnology GmbH or one of its affiliates All rights reserved. eResearchTechnology GmbH Sieboldstraße 3
97230 Estenfeld, Germany Tel: +49 9305 720-9891 Fax: +499305720-7891

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals