The SIL O 2 Oxygen Analyser
User Operation Manual Rev. 1.2
19/06/2023
SIL O2 Oxygen Analyser
Revision History
Revision No.| Change Description| Date
0 1| First Issue| 19-10-2018
1| Note added for Non-ATEX Sensor usage| 22-10-2018
1.1| Reference to 15% O2 in trouble shooting section removed.| 30-11-2018
| Calibration Span instruction amended and note added|
1.2| SIL2 capability specified| 19-06-2023
Introduction
The Ntron SIL O2 Analyser is a ATEX approved SIL 2 capable Oxygen measuring
instrument. It contains three programmable alarm settings, industry standard
Analogue Output(active source) and a non-programmable safety trip relay. The
range of operation is 0 to 25% Oxygen. The Resolution is to 0.01%.
Together with one of the range of Ntron ATEX approved Oxygen Sensors, it forms
a SIL2 capable measurement unit which can operate on its own or as part of a
larger installation.
Note. Non-Atex (Ex) Sensors variants of the models listed in the manual
may also be used with the SIL O2 Analyser as long as they are installed in a
Non-ATEX (Ex) or‚ Safe area or zone.
The Alarm settings are typically factory set according to the end user’s
requirements. Interface software and hardware is available for the user to
perform such functions themselves. Contact Ntron for further details.
1.1 The purpose of this User and Operation Manual
This Manual is limited to providing the user with necessary details on
installation, interface wiring, operation and maintenance of the SIL O2
Analyser and Oxygen Sensor, to ensure safe and reliable operation of this
Analyser and Sensor unit.
Note: This manual does not cover the design or application of a Safety
Instrumented System (SIS) into which the SIL O2 Analyser unit may be
incorporated. Further details of this can be obtained in the Ntron SILO2
Technical and Safety Manual.
The main user interface features
Installation and Operation
Note: For system application information, consult the SIL O2 Analyzer
Technical User Manual
2.1 Notes for a safe installation
Please do not short-circuit the + and -ve sensor cables while connecting to
the SIL O2 Analyzer. This could damage the sensor.
This system can also be used in a non-Ex area application. In all
applications, the SIL O2 Analyzer Din rail module is always located in the
non-Ex area!
Assignment of the mounting rail housing terminals with intrinsically safe
circuits and non-intrinsically safe circuits is indicated on the nameplate and
is clearly indicated. In addition, the 4-pin terminals of intrinsically safe
circuits follow the standard color protocol for intrinsically safe terminal
connections and are coloured Blue. Connections for intrinsically safe external
circuits are arranged in such a way that, in accordance with EN 60079-11, the
exposed parts are at least 50 mm from the exposed connections and parts of the
non-intrinsically safe circuits.
For safe operation, a protective ground/earth connection to terminal 13 or 15
should be made. Assembly / disassembly, installation, operation and
maintenance may only be performed by qualified personnel in accordance with
regulatory requirements and the SIL O2 Oxygen Analysis Manual.
During installation, the technical data and the electrical values of the
connected circuits must be respected.
Electrical supply. The SILO2 Analyser operates with an Extra low Voltage (ELV)
supply. (See specification section in this manual)No special precautions are
necessary but the ELV supply to the SIL O2, if generated from a Mains power
source, that mains power source and associated wiring must be in accordance
with IEC/EN standards and is the responsibility of the user to provide and to
ensure correct and safe installation of such.
When the SIL O2 analyzer is integrated into another system, also refer to the
system manual for additional operational information.
The Ntron SILO2 Analyser is designed for use with a range of Ntron ATEX
certified Oxygen Sensors to form a measurement system. These Sensor options
are detailed in the Specification section of this manual.
If the SIL O2 Analyser has been supplied with a Sensor, the two units will
have been calibrated at the Ntron factory prior to delivery. Otherwise, if a
Sensor is supplied separately from the SIL O2 Analyser, a calibration will be
required prior to putting the system into operational service.
The Ntron Sensors suitable for use with the SIL O2 analyzer are listed in the
Installationn section which follows.
The SIL O2 oxygen analyzer is an associated explosion-proof [Ex ia] IIC or [Ex
ib] IIC electrical device and should always be used outside potentially
explosive areas (in a safe area). Only electrical circuits, certified as
intrinsically safe, can be connected to other intrinsically safe circuits in
the Ex zone.
Before operation, the intrinsic safety must be verified for the SIL O2 oxygen
analyzer circuit connected to the circuit of other equipment, including the
interconnecting cabling.
The data contained in the EC test certificate and the regulations of EN
60079-14: 2011-10 must be observed.
The Sensors are ATEX rated devices and have certain conditions of use
assigned.
Note. Non-Atex (Ex) Sensors variants of the models listed in the manual
may also be used with the SIL O2 Analyser as long as they are installed in a
Non-ATEX (Ex) or‚ Safe area or zone.
For Acetal-bodies Sensors, the following instructions apply; The Sensor should
be installed in such a way that it is not subject to impact by other objects
and should not be located close to additional heat sources.
Note the ambient temperature range as it appears on the Sensor labelling.
Care must be taken when installing equipment with plastic enclosures or
plastic parts of enclosures to ensure that the equipment is protected from any
situations that could cause a build up of static charge. The equipment must
not be installed into locations in which it could come into contact with,
through normal or abnormal circumstances, fast moving dust laden air/gas or
non-conductive fluids. The equipment must be cleaned only with a damp cloth.
2.2 Installation
The installation must be in accordance with the local electrical codes and
taking into account the details in the Specification section of this manual.
The SIL O2 analyzer is designed for DIN rail mounting and can be supplied
mounted in an additional enclosure or provided without an additional enclosure
for customer mounting in a system enclosure or control panel.
The Sensor option chosen will determine the process connection/installation
for that Sensor type. The Sensor types and related process connections are
given below.
Sensor Type: Process Connection:
Model OC-25 Ntron Sensor base, Flow Through or Tri-Clamp
Model OC-26 KF40 Flange
Model OC-200’Oxyprobe‘ series Bushing insert or Extract Probe Holder
The Sensor is typically supplied with a corresponding connection cable of
either standard length or to a customer specified length.
The Ntron OxyExtract Manual or Automatic probe insertion mechanism.
2.2.1 Mounting the SIL O2 Analyser
Mount the SIL O2 Analyser onto 35mm DIN rail as shown. Any Trunking/wireways
above and below the Analyer when in situ, must be positioned so that there is
a minimum of 30mm clearance between such trunking and the Analyser connection
terminals.
2.3 Electrical and Interface Connections
2.3.1 Sensor Input and Customer Interface Wiring
The Ntron ATEX Sensor selected for use with the SIL O2 Analyser as listed in
the Installation section previously, has intrinisically safe parameter values
which match the Isolation Barrier output which is built into the SIL O2
Analyser. The Sensor connects to the Blue terminals as shown below.
The dark Grey terminals are for safe (Non-Ex area) customer interface
connections. See the configuration section on the following page.
2.3.2
Operational Settings and Wiring Configuration
The SIL O2 Analyser has two presentable alarm levels and associated Relays,
referred to as RL1 and RL2. It also has a third presentable alarm level with
an associate Transistor Digital Output (DO) The programmed setting of these
outputs can vary depending on application. The Inspection and Calibration
certificate supplied with each SIL O2 Analyser and Sensor unit will detail
these settings. See the Ntron SIL O2 Technical and Safety manual for further
details.
The SIL O2 Analyser has a main trip Relay referred to as RL3. This is not
programmable and is normally open (contacts)/de-energised when the Analyser is
not under power. The Relay energises and its associate contacts close when
power is applied and the Analyser and Sensor are healthy and ready to operate.
The Analogue output is an industry standard 4-20mA active source output. The
connected circuit should be of maximum resistance 420 ohm.
To meet the SIL 2 operational requirements as an Analyser and Sensor unit, the
main trip Relay is required to be connected in series with other elements
within the Analyser as shown in the diagrams below. For any other
configurations, See the Ntron SIL O2 Technical and Safety manual for further
details.
*Special case for use of the 4-20 mA analogue output within a safety
system
When using the 4-20 mA analogue output from the SILO2 Oxygen Analyser within
the safety system, this signal would typically be required to be processed by
a safety PLC which would then activate the Final Trip Element. See the Ntron
SIL O2 Technical and Safety manual for further details.
2.3.3 Typical Relay Configuration
These relays are programmed to respond to the measured Oxygen level setpoints
(Alarms), rising or falling in the range 0-25% Oxygen. These are pre-set to
the customer’s requirements.
They operate in ‘Fail Safe’ mode which means that they are energized (under
current), presenting a closed contact to the customer/user when the SIL O2
module is under power and the status is healthy/good.
When at a level setpoint (Alarm) or when the SIL O2 module is not-powered, the
relays are de-energised (not under current), presenting an open contact to the
customer/user. This is also known as ‘Fail Open’ (FO) condition.
Transisitor (DO)
This is an NPN device that can switch 24VDC and can be configured to be
normally open or normally closed.
Analogue 4-20mA
This is an active output and requires connection to a passive external circuit
suitable for proper operation.
It is set to the range 0-25% Oxygen / 4-20mA (0% Oxygen = 4mA, 25% Oxygen =
20mA.)
This range can not be adjusted.
If the 4-20 mA output is not used, the SIL O2 terminals T11 and T12 must be
wired by the user. This output is internally connected in series with the SIL
O2 Analyzer digital display, which will not work if the aforementioned
terminals are in open circuit!
2.3.4 Sensor Process Connection options
The Sensors are suitable for mounting directly onto process
Lines via a selection of process fittings. The electrical signal connectorsare
rated to IP67.
2.4 Operation
Before turning on the SIL O2 analyzer for the first time or after
disconnecting the sensor for maintenance / replacement, make sure that a
working sensor is connected. Otherwise, the SIL O2 analyzer will go into
failure mode, with the red error indicator on and the RL3 relay disabled. As
the fault circuit performs a cyclic check every 30 seconds or so. When the
sensor is connected, resetting the red indicator and relay 3 occurs
automatically, but it is advisable to power off the system to allow the sensor
to be disconnected / reconnected under normal circumstances, if possible.
Turn on/apply power to the SIL O2 Analyser.
When operating under normal process/system conditions, the SIL O2 Analyser
will give a digital readout of the measured Oxygen level (from the Sensor in
the process).
This measurement is converted into a 4-20mA output signal proportional to the
Oxygen being measured. If an Alarm level even occurs, the relevant programmed
output device will operate (Relay or Transistor) and the ineterface circuit to
the user equipment will be interrupted. For further alarm or fault events and
their conditions, please see the troubleshooting section at the end of this
Manual.
2.4.1 Calibration Procedures
The SIL O2 Analyser and Sensor unit requires periodic calibrations performing,
the timing between such calibrations being determined by the application and
process requirements.
Typically, a periodic calibration check would reveal if any calibration
adjustment is necessary. The SPAN calibration point of 20.9% O2 (Air) is the
important setting and typically, if calibration adjustment is required, this
is the value that would be adjusted. This adjustment can be made ussing
ambient air as the calibration standard or, for greater accuracy, certified
cylinder gas at 20.9% O2.
It is possible to also adjust the Zero point of the SIL O2 Analyser. Typically
this would only be required when fitting a new/replacement Sensor. A Complete
calibration procedure is given below.
Note: Calibration Adjustment procedure.
The displayed number change starts with the digit to the far right of the
decimal point. Press and hold the Up or Down buttons to change the reading.
This gives an accelerated adjustment. More accurate adjustment is achieved by
pressing the Up or Down button at one press (press release, release of press,
etc.). When the display reads 20.9%, the calibration of the range is complete
2.4.2 Setting the Zero point.
This must be done each time a new sensor is installed before a span
calibration is performed.
Subsequently, a zeroing can be performed periodically as needed.
A zeroing operation requires the application of a zero oxygen gas to the
sensor, typically nitrogen (minimum quality <100 ppm). Note: Accessing the
Zero function when the sensor is in ambient air (20.9% O2) will result in an
error / fault condition.
Depending on the type of sensor installed, zero gas must be applied through a
process fitting or calibration adapter, and be sure to leave a free exhaust at
a suitable location so that the sensor does not become pressurized.
The sensors to be used with the SIL O2 unit will generally give a small output
at zero oxygen levels. The zero adjustment function allows the user to adjust
this small signal so that the SIL O2 analyzer displays 0.0 when no oxygen is
present in the measured gas.
- ZERO mode is activated after pressing the UP and DOWN button simultaneously for more than one second. As an acknowledgment of receipt, the A3 LED flashes. An adjustment of the indicated value is made by pressing the UP or DOWN buttons individually as required.
- The change starts with the digit to the right of the decimal point. (00.0) Press and hold the Up or Down buttons to change the displayed reading. This allows an accelerated adjustment. Accurate adjustment is achieved by pressing the Up or Down button at one press (press release, release of press, etc.). When the display reads 0.0% O2, the setting is complete. By simultaneously pressing the two buttons or a delay of 20 seconds, the ZERO operating mode ends.
2.4.3 SPAN Calibration
This can be done periodically or as needed in accordance with the requirements
of the calibration protocol.
Calibration can be performed by applying either clean ambient air or a 20.9%
certified oxygen rate to the sensor.
Note:
-
The SIL O2 analyzer should only be calibrated at approximately 20.9% oxygen.
-
If an ambient air calibration is performed, it is recommended to confirm the oxygen level with a certified portable oxygen analyzer prior to calibration.
-
If a zero setting is made first, then after the end of the zero operation mode, the Span mode is automatically activated after a delay of 20 seconds.
-
If A Span only calibration is required, then just press the UP or DOWN buttons individually to increase or decrease the displayed value of oxygen to read 20.9.
Note: Ensure the Sensor is in 20.9% oxygen when you do this! 20.90 . -
Calibration complete.
Operational safety and maintenance instructions.
If it is assumed that safe operation is no longer possible, the device must be
taken out of service and protected against accidental use. The reasons can be:
visible damage to the device failure of the electrical function long shelf
life at temperatures above 85 ° C
Transport damage
Before the device can be put back into service, a professional routine check
must be carried out in accordance with DIN EN 61010, part. 1. This examination
must be carried out by the manufacturer. Repair work on Ex devices may only be
carried out in accordance with § 9 of Ex. (Elex V).
3.1 Fault conditions
The SIL O2 Analyzer failure operation is not locked out by a fault. When a
fault condition is repaired, the trip relay RL3 and the red fault indicator
will return to their “healthy” or “operational” state after a few seconds
following an automatic internal cyclic check.
3.2 Wire break
When manually disconnecting the sensor (for example for maintenance) or
following a break in the sensor connection cable, the SIL O2 Analyzer RL3
fault relay switches off and interrupts the safety circuit.
The red fault indicator will illuminate on the SIL O2 analyzer and the 4-20 mA
analog output will reach a constant high value of 22.0 mA. At the same time,
the setpoint relays RL1 and RL2 will be deactivated.
When the sensor is reconnected or the cable fault is corrected, the RL3 relay
automatically resets after about 30 seconds and the red LED goes out. At the
same time, RL1 and RL2 are reset according to their setpoints and the measured
gas level. See also note on page 13 of this manual for reset requirements
under specific conditions.
3.3 System faults
A number of internal faults in the SIL O2 analyzer generate a fault output and
interrupt the safety circuit. In some cases, cycling the power supply of the
SIL O2 analyzer may be sufficient to remedy the problem. Otherwise, please
contact Ntron for assistance.
3.4 Troubleshooting
Possible Faults and their solutions
The following possible conditions are applicable to a system (SIL O2 Analyser
and Sensor) already installed and commissioned. Some conditions below may also
be applied to new systems not yet commissioned.
-
Action: Performing a Zero calibration when the Oxygen Sensor is in ambient air. Problem: An incorrect Zero level input to the Analyser will result in an overrange fault. This can occur if a Zero calibration is performed when the Sensor is in ambient air. If this occurs, the Red fault LED will illuminate and the Analyser display and Analogue output will read an over-range value. The Alarm level LED’s may also illuminate. This may also occur if a genuine overrange event takes place, with gas containing more than 25% O2 being applied. (25%O2 displayed) Solution: re-calibrate with Nitrogen gas (Zero) and then re-calibrate at the Span point of 20.9%.
-
Action: Powering on the Analyser with no sensor connected or a wire break to the sensor. Problem: If the sensor is disconnected, this would ordinarily cause relays R1, R2 and R3 to deenergise. Analyser display and Analogue output will read a fault value. (28% O2 displayed and 22mA output) Solution: Check Sensor wiring and connect the sensor to the Analyser. After approximately 30 seconds, the Analyser will reset and the Red LED should extinguish.
-
Action: Unable to perform a Span calibration. Problem: The Analyser cannot be adjusted to display 20.9% O2 during Span calibration.
Solution: The Sensor may be approaching end of life or has been damaged and cannot generate sufficient output. Replace the sensor Alarm level LED’s may or may not be illuminated depending on configuration. -
Action: Alarm Level LED’s (Yellow) illuminated. Problem: Alarm level LED’s illuminated.
Solution: Genuine Oxygen level alarm event.
Incorrect sample gas levels being applied to the Sensor or incorrect alarm level settings are configured within the Analyser. If this is not expected then check sample gas with independent instrument gas/and/or re-configure Analyser Alarm setpoints. This will require interface with PC based software. -
Action: Red Fault LED illuminated when none of the above conditions are present. Problem: Analyser in fault condition. Analyser display and Analogue output will read a fault value.
(28% O2 displayed and 22mA)
Solution : Internal Analyser electronics or programming fault. Try Power On/Off cycle to reset.
If fault does not clear then seek further assistance.
Problem: No Analyser display (Blank)
Solution: Analogue output circuit is broken or interrupted. Investigate
and repair.
3.4.1 Possible Sensor Faults
Applicable to all sensor types (OC-25, OC-26, OC-200)
Problem:. SIL O2 Analyser will not calibrate (Cannot be set to 20.9%)
Solution: Sensor is at end of life. Sensor has been wetted or
contaminated. The Sensor output signal can be checked at the SIL O2 Analyser
terminals 14 and 16. See specification section for details of a healthy output
signal.
Problem:. SIL O2 Analyser reads Zero (0.00)
Solution: Actual low Oxygen measurement. Sensor wires have been shorted
together, damaging the Sensor.
Verify the Oxygen level by other means. Check the Sensor output signal.
Replace Sensor as necessary.
3.5 General Maintenance
Establish a periodic checking and maintenance routine in line with the
requirements for Safety Instrumented Systems. See the Ntron SIL O2 Analyser
Technical and Safety manual for further details.
The SIL O2 Analyser requires little physical maintenance. The user Enclosure
into which it is housed should provide protection against a build up of dust
or other contaminants on the surface of the SIL O2 Analyser housing. If such
contaminants are seen during regular inspection, such can be removed by gentle
suction device or by wiping with a damp but not wet, cloth. The ingress of
such contaminants should be investigated.
Ensure all wiring to the SIL O2 Analyser is secure and in good condition,
paying particular attention to the security of the connection terminals if the
user enclosure is subject to vibrations.
There are no user serviceable parts within the SIL O2 Analyser. If any
malfunction is detected, mechanical or electrical, the SIL O2 Analyser should
be immediately removed from service following the correct protocols.
The Sensor used has a finite life span. The Performance of the Sensor is
verified by calibration check and this should be performed according to the
protocols required by the safety system. Replacement of the Sensor is
necessary when it does not meet the required performance levels.
During operation life, the Sensor should be kept clean of contaminants. It can
be wiped with a soft damp cloth. Observe the restrictions regarding potential
static charges as detailed on the Sensor Installation
Instruction Documents found in the appendix to this manual.
When removing Sensors from their process installation for replacement or other
service requirements, observe the following points.
- Ensure any and all system shut down protocols are followed as applicable.
- Do not disconnect the Sensor from the connecting cables whilst the system is operating or is ‘Live’.
- Ensure the process being measured is shut down or the Sensor connection is isolated by any intermediate valve mechanism if fitted.
- Always remove the Sensor connector before unscrewing the OC-25/OC-200 Sensors or unclamping the OC-26 Sensors from their process fittings
- Block off the exposed process connection if required during the period the Sensor is disconnected.
- Protect the disconnected Sensor cable from damage during the period the Sensor is disconnected.
Specifications
SIL 02 Analyser | Specifications-Electrical |
---|---|
Supply Power | 24 V DC +/- 10% |
Power Consumption | 1.5 Watt |
Analogue Output | 4-20mA active source. 22mA Max output . Load 390 Ohm@22mA, |
Max 420 Ohing20mA constant current.
Communications| RS232
Relay Contact outputs RL 1 /RL2/RL3| Urn 125 V AC / 110 V DC Min Current 10 uA
DC. Min. Voltage 10 mV DC. Type According to IEC 947-5-1 reap. EN
60947)
Transistor Output (Do)| Switching parameters: < 28 V @ < 50 mA
Intrinsically Safe Connection| voltage Uo DC 6 V current Intensity to 0.2
mA power Po 0.3 mW max. outer InductIvity Lo 1000 m1-1 max. outer capacity
Co 10 OF
| Specifications-Mechanical
Terminal /Wire Size| Pluggable / Quick Release terminals, capacity 2.5nunl
Mounting| 35 mm Din rail
Housing Material| PBT
Ingress Protection| 1P20
Combustibility Class| VO according to UL
Weight| 300g
Operating Temperature| -20…+50°C, 0…95% Humidity, non-condensing.
Dimensions| 70 mm x 145.5 mm x 100 mm
Indicators| LCD Display and LED alerts on analyser.
Certification/Standards| CE, ATEX 2014/34/EU/ EN 60079-0:2012 + A11:2013, EN
60079-
11: 2012, EN 60079-26: 2015 0158 II (1) G II (1) D ; [Ex ia Ga] IIC; [Ex ia
Da] IIIC
Functional safety: SIL2 Capable according to IEC 61508/61511
EMC 2014/30/EU
EN 61326-3-2:2008; EN 61000-6-3: 2007 + A1: 2011
EN 61000-4-2:2009; EN 61000-4-3: 2006 + A1: 2008 + A2: 2010
EN 61000-4-4:2012
2006 + A1: 2008 + A2: 2010
EN 61000-4-5:2014
EN 61000-4-6:2014
Oxygen Sensor| Specifications-Electrical
---|---
Models| OC-25, OC-26, OC-200 series
Range| 0-25% Oxygen
Signal Output| 300-375mV in Air OC25, OC-26. 135-160mV in Air OC-200
Technology/Lifespan| Solid State Long Life Up to 24 months application
dependent.
Storage Life 12 months.
Process Connections| 2″ tri-clamp / flow-through / KF40 flange / via Oxy
Extract
Intrinsically Safe Sensor
Input Connection
Parameters.| voltage Uo DC 12 V current intensity Io 120 mA
power Po 0.55 W
max. outer inductivity Lo 0 mH
max. outer capacity Co 1.2 μF
| Specifications-Mechanical
Dimensions OC-25| 100mm High x 50mm Diameter
Dimensions OC-26| 75mm High x 55 Diameter (KF40)
Dimensions OC-200| Variable length to order, typically 200mm long x 12mm
Diameter probe.
Max. Diameter is 30mm.
Protection Class| IP67 When inserted into process fitting with Connector
fitted.
Weight| OC-25= 250g; OC-26=150g, OC-200=180g
Environmental| Temperature: -20 to +45-50°C, 10-95% Humidity, no condensation.
Certification/Standards| CE, ATEX 2014/34/EU/ 1180 II 1 GD
EN60079-0:2012 + A11:2013
EN60079-11:2012
ATEX II 1 GD; Ex ia IIC T6 Ga(-20°C≤Ta≤+55°C)
Ex ia IIIC T90°C Da( -20°C≤Ta≤+55°C)
IECEx BAS 09.0148X
Appendices
- SIL O2 Analyser and Sensor Connection Diagrams
- SIL O2 Analyser CE and ATEX Certificates
- Sensor CE and ATEX Certificates
5.1 SIL O2 Analyser and Sensor Connection Diagrams
- E364 SIL O2 with OC-25 Sensor
- E511 SIL O2 with OC-26 Sensor
- E395 SIL O2 with OC-200 ‘Oxyprobe’
5.2 SIL O2 Analyser CE and ATEX Certificates
- CE Certificate CETX003
- ATEX Certificate Dekra BVS 13 ATEX E 010
EU Declaration of Conformity
We declare, under our full responsibility, that we believe the products
identified in this declaration, and to which this declaration relates are in
conformity with the requirements of Council Directives:
Directive 2014/34/EU- Equipment and protective systems intended for use in
potentially explosive atmosphere.
Directive 2011/65/EU and of Council of 8th June 2011 on the restriction of the
use of certain hazardous substances in electrical and electronic equipment
(RoHS2)()
OJ L 174, 1.7.2011, p.88 & P106
Description of Equipment
SilO2 Oxygen Analyser.
EU Declaration of Conformity
Standards and marking
Standards applied:| ATEX:
EMC:
EN 60079-0: 2012 + A11: 2013
Explosive atmospheres – Part 0: Equipment – General requirements.
EN 60079-11:2012 Explosive atmospheres – Part 11: Equipment protection by
intrinsicsafety “i”.
EN61326-3-2:2008 Electromagnetic compatibility – immunity requirements for
safety related systems and for equipment intended to perform safety related
functions in industrial applications with specified electromagnetic
environment.
EN 61000-6-3: 2007 + A1: 2011, Emissions Light Industrial
EN61000-4-2:2009, Electrostatic Discharge
EN61000-4-3: 2006 + A1: 2008 + A2: 2010, Immunity Heavy Industrial
EN61000-4-4:2012, Burst Immunity
EN61000-4-5:2014, Surge Immunity
EN61000-4-6:2014, Conducted Immunity
---|---
Equipment marking:| II (1)G (Ex ia Ga) IIC
II (1)D (Ex ia Da) IIIC
Notified body:| DEKRA Exam GmbH.
Dinnendahlstrasse 9
44809 Bochum
Germany.
BVS 13 ATEX E 010
The Authorised Signatory to this declaration, on behalf of the manufacturer,
is identified below:
Name
David
Title/Position;
Managing Director
Address
Ntron Ltd, Mullaghboy Industrial Park, Navan, Co. Meath, Ireland.
5.3 Sensor CE and ATEX Certificates
- CE Certificate SUII01, SUII02
- ATEX Certificate BAS02ATEX1230X-11
EU Declaration of Conformity
We declare, under our sole responsibility, that we believe the products
identified in this declaration, and to which this declaration relates are in
conformity with the requirements of the EU Council Directives as stated below;
Directive 2014/34/EU equipment and protective systems intended for use in
potentially explosive atmospheres.
Directive 2011/65/EU and of Council of 8th June 2011 on the restriction of the
use of certain hazardous substances in electrical and electronic equipment
(RoHS2)()
OJ L 174, 1.7.2011, p.88 & P106 Annex IV 1b exemption.
Description of Equipment: | OC-Type Oxygen Sensors For Use in Hazardous Areas. |
---|---|
Standards applied: | EN60079-0:2012 + A11:2013 |
EN60079-11:2012
Marking:| II II 1 GD Ex ia IIC T6 Ga(-20°C≤Ta≤+55°C)
Ex ia IIIC T90°C Da( -20°C≤Ta≤+55°C)
Certificate No: BAS02ATEX1230X
(Other markings: IECEx BAS 09.0148X)
Notified Body:| SGS Baseefa
Rockhead Business Park
Staden Lane, Buxton, Derbyshire,
SK17 9RZ, UK.
Notified Body Number: 1180
The Authorised Signatory to this declaration, on behalf of the manufacturer,
is identified below:
Name: David : Managing Director
Address : Ntron Ltd Mullaghboy Industrial Estate Navan Co. Meath Ireland, C15
XD61
Signature:
Ph: 00353469071333
Fx: 00353469071331
email: info@ntron.com
Web: www.ntron.com
Date: 12/02/2019
SENSOR USE/INSTALLATION INSTRUCTIONS
To ensure security of personnel, goods and plant it is important that the
national standards laid down for the installation of electrical equipment are
followed, that instructions on the nameplate are achieved, and that the work
is carried out by trained personnel. Furthermore, it is necessary to use the
oxygen sensor in accordance with the advice and data given in this text.
Oxygen sensors for hazardous areas are especially designed to conform to
national and international standards governing risks of explosion and
certificates for each type are issued by an approved organisation or notified
body. Be careful ! A certification number followed with “X” or a “U” requires
specific conditions for use.
- Care must be taken when installing equipment with plastic enclosures or plastic parts of enclosures to ensure that the equipment is protected from any situations that could cause a build up of static charge. The equipment must not be installed into locations in which it could come into contact with, through normal or abnormal circumstances, fast moving dust laden air/gas or non-conductive fluids. The equipment must be cleaned only with a damp cloth.
- When installing equipment with Metal enclosures, earthing of the metal enclosure must be ensured to avoid the potential of electrostatic discharge.
STORAGE
The sensor should be stored in a cool dry place.
HANDLING WARNINGS
Do not expose open end of sensor housing to liquids, particulates, grease, or
oil.
Do not touch top of sensor within open end of sensor housing.
Do not allow vacuum grease to coat open end of sensor.
Do not expose sensor to sudden mechanical shocks.
INSTALLATION & CALIBRATION
Remove the sensor connector plug from the old sensor.
Remove the old sensor by uscrewing counterclock wise or by removing from
housing.
Remove and dispose of the old sealing O-Ring where fitted.
Fit the new sealing O-Ring if required.
Install the replacement sensor threading on clockwise or replacing in housing.
Fully hand tighten the sensor housing to ensure O-Ring compression.
Re-connect the sensor plug.
Calibrate the analyser to the new sensor.
Note: The system must not be put back on-line until the analyser has been
calibrated to the new sensor.
ELECTRICAL CERTIFICATION
Sensor Type’s OC-2, OC-7, OC-16, OC-17, OC-18, OC-19, OC-20, OC-21, OC-22,
OC-23, OC-24, OC-28, OC-28M & OC-47 carry the following markings.
BAS02ATEX1230X
Ex ia IIC T6 Ga(-20°C≤Ta≤+55°C)
Ex ia IIIC T90°C Da( -20°C≤Ta≤+55°C)
IECEx BAS 09.0148X
Ui:30v, Ii:120mA, Pi: 0.55W, Ci:0, Li:0
Sensor types OC-19, OC-20, OC-25, OC-25M, OC-26, OC-26M carry the following
markings.
BAS02ATEX1230X
Ex ia IIC T6 Ga(-20°C≤Ta≤+55°C)
Ex ia IIIC T90°C Da( -20°C≤Ta≤+55°C)
IECEx BAS 09.0148X
Ui:12v, Ii:120mA, Pi: 0.55W, Ci:1.2uf, Li:0
Application Note For All Sensor Types:
The Oxygen Sensors may only be used within a -20 to +55 Deg. C temperature
range and will withstand a 500V isolation test between the input connection
and frame of the apparatus.
Application Note for OC-25M, OC-26M and OC-28M Sensors.
See Sensor use/Installation Instructions note 2 above.
MAINTENANCE
At the end of the sensor’s life, the output of the oxygen cell will fall to zero in less than a day. If the sensor will no longer re-calibrate, it should be replaced. II 1GD II 1GD
Certificate Number
BAS02ATEX1230X
Issue 11
Issued 3 September 2018
Page 1 of 5
-
EU – TYPE EXAMINATION CERTIFICATE
-
Equipment or Protective System Intended for use in Potentially Explosive Atmospheres Directive 2014/34/EU
-
EU – Type Examination Certificate BASO2ATEX1230X — Issue 11 Number:
3.1 In accordance with Article 41 of Directive 2014/34/EU, EC-Typc Examination Certificates referring to 94/9/EC that were in existence prior to the date of application of 2014/34/EU (20 April 2016) may be referenced as if they were issued in accordance with Directive 2014/34/EU. Supplementary Certificates to such EC-Type Examination Certificates, and new issues of such certificates, may continue to bear the original certificate number issued prior to 20 April -
Product: A Series of Oxygen Sensors
-
Manufacturer: Ntron Limited
-
Address: Mullaghboy Industrial Estate, Navan, County Meath, Ireland
-
This re-issued certificate extends EC – Type Examination Certificate No. BASO2ATEX1230X to apply to product designed and constructed in accordance with the specification set out in the Schedule of the said certificate but having any variations specified in the Schedule attached to this certificate and the documents therein referred to
-
The original certificate was issued by The Electrical Equipment Certification Service, Notified Body Number 0600, which retains responsibility for its original documentation. SGS Baseman, Notified Body Number 1180, in accordance with Article 17 of Directive 2014/34/EU of the European Parliament and of the Council, dated 26 February 2014, is responsible only for the additional work relating to this re-issued certificate and any other supplementary certificate it has issued. The examination and test results are recorded in confidential Report No. See certificate history
-
Compliance with the Essential Health and Safety Requirements has been assured by compliance with: EN 60079-0: 2012 + All: 2013 EN 60079-11: 2012
except in respect of those requirements listed at item 18 of the Schedule. -
If the sign “X” is placed after the certificate number, it indicates that the product is subject to the Specific Conditions of Use specified in the schedule to this certificate.
-
This EU – TYPE EXAMINATION CERTIFICATE relates only to the design and construction of the specified product. Further requirements of the Directive apply to the manufacturing process and supply of this product. These are not covered by this certificate.
-
The marking of the product shall include the following : II 1GD Ex is IIC T6 Ga (-20°C < Ta < +55°C) Ex is IIIC T90°C Da (-20°C Ta <+55°C)
SGS Baseefa Customer Reference No. 2144
Project File No. 18/0344
This document is issued by the Company subject to its General Conditions for
Certification Services accessible at http://wwwcgc.com/enaerms-and-
Conditions.asox and the Supplementary Terms and Conditions accessible at
http://Anvw.sits.com/Magicelarlennt and-Conditions. Attention is drawn to
the limitation of liability, indemnification and jurisdiction issues defined
therein. Any holder of this document is advised that information contained
herein reflects the Company’s findings at the time of its intervention only
and within the limits of Client’s instructions, if any. It does not
necessarily indicate that the equipment may be used in panicular industries or
circumstances. The Company’s sole responsibility is to its Client and this
document does not exonerate parties to a transaction from exercising all their
rights and obligations under the transaction documents. This document cannot
be reproduced except in full, schedule included, without prior ‘Witten
approval of the Company. Any unauthorized alteration, forgery or falsification
of the content or appearance of this document is unlawful and offenders may be
prosecuted to the fullest extent of the law.
SGS Baseefa Limited
Rockhead Business Park, Staeden Lane, Buxton, Derbyshire SK17 9RZ
Telephone +44 (0) 1298 766600
Fax +44 (0) 1298 766601 e-mail baseefae@sgs.com web
site www.sgs.co.uk/sgsbaseefa Registered in
England No. 4305578. Registered address: Rossmore Business Park, Ellesmere
Port, Cheshire, CH65 3EN
BAS-CERT-037
n2 0 BREARLEY noncircadian R S SINCL IR Manager TECHNICAL MANAGER On behalf of
SGS Baseefa Limited
SENSOR USE/INSTALLATION INSTRUCTIONS
PRELIMINARY ADVICE
To ensure security of personnel, goods and plant it is important that the
national standards laid down for the installation of electrical equipment are
followed, that instructions on the nameplate are achieved, and that the work
is carried out by trained personnel. Furthermore, it is necessary to use the
oxygen sensor in accordance with the advice and data given in this text.
Oxygen sensors for hazardous areas are especially designed to conform to
national and international standards governing risks of explosion and
certificates for each type are issued by an approved organisation or notified
body. Be careful ! A certification number followed with “X” or a “U” requires
specific conditions for use.
The Specific Condition of Use for the sensor types listed below is; Possible
Electrostatic Risk-Do not charge by Rubbing and do not clean with Solvent
STORAGE
The sensor should be stored in a cool dry place.
HANDLING WARNINGS
Do not expose open end of sensor housing to liquids, particulates, grease, or
oil.
Do not touch top of sensor within open end of sensor housing.
Do not allow vacuum grease to coat open end of sensor.
Do not expose sensor to sudden mechanical shocks.
INSTALLATION & CALIBRATION
Remove the sensor plug from the old sensor.
Remove the old sensor by unscrewing counter clock wise.
Install the replacement sensor threading on clockwise.
Fully tighten the sensor housing.
Re-connect the sensor plug.
Calibrate the analyser to the new sensor.
Note: The system must not be put back on-line until the analyser has been
calibrated to the new sensor.
ELECTRICAL CERTIFICATION
Sensor Type’s OxyProbe 200 Sensor Probe carry the following markings.
OXYGEN SENSOR
BAS02ATEX1230X
Ex ia IIC T6 Ga(-20°C≤Ta≤+55°C)
Ex ia IIIC T90°C Da( -20°C≤Ta≤+55°C)
IECEx BAS 09.0148X
Ui:12v, Ii:120mA, Pi: 0.55W, Ci:1.2uf, Li:0Application Note For All Sensor
Types:
The Oxygen Sensors must only be used within the temperature range quoted in
the certification details, and will withstand a 500V isolation test between
the input connection and frame of the apparatus.
MAINTENANCE
If the sensor will no longer re-calibrate, it should be replaced. It is
recommended that sensors be replaced after 36 months.
II 1GD
EU Declaration of Conformity
We declare, under our sole responsibility, that we believe the products
identified in this declaration, and to which this declaration relates are in
conformity with the requirements of the EU Council Directives as stated below;
Directive 2014/34/EU equipment and protective systems intended for use in
potentially explosive atmospheres. Directive 2011/65/EU and of Council of 8th
June 2011 on the restriction of the use of certain hazardous substances in
electrical and electronic equipment (RoHS2)()
OJ L 174, 1.7.2011, p.88 & P106 Annex IV 1b exemption.
Description of Equipment:| OxyProbe200 – Oxygen Sensors For Use in Hazardous
Areas.
OxyProbe200 –
---|---
Standards applied:| EN60079-0:2012 + A11:2013
EN60079-11:2012
Marking:| II II 1 GD Ex ia IIC T6 Ga(-20°C≤Ta≤+55°C)
Ex ia IIIC T90°C Da( -20°C≤Ta≤+55°C)
IECEx BAS 09.0148X
Certificate No: BAS02ATEX1230X
Notified Body:| SGS Baseefa
Rockhead Business Park
Staden Lane, Buxton, Derbyshire,
SK17 9RZ, UK.
Number: 1180
The Authorised Signatory to this declaration, on behalf of the manufacturer,
is identified below:
Name: David Address: Ntron Limited
Mullaghboy Industrial Estate
Navan
Co. Meath
Ireland, C15 XD61
Signature:
15 Description of Product
The Series of Oxygen Sensors is designed to provide an electrical signal in
proportion to the percentage oxygen present in the measured atmosphere.
Alternative sensor types are as follows:
Model OC-2/202/224 | Standard Oxygen Sensor |
---|---|
Model OC- 7 /207 | In-situ Oxygen Probe |
Model OC-16/216 | Oxygen Sensor |
Model OC-17/18/19/20 | Oxygen Sensor |
Model OC-21/22/23/24/221 | ingle or Dual Sensor |
Model Oxyprobe 200 | Oxygen Probe |
Model OC-25 | Oxygen Sensor |
Model OC-47 | Oxygen Sensor |
Model OC-26 | Oxygen Sensor |
Model OC-26M | Oxygen Sensor (Metal Enclosure) |
Model OC-28 | Oxygen Sensor |
Model OC-28M | Oxygen Sensor (Metal Enclosure) |
Model OC-25M | Oxygen Sensor (Metal Enclosure) |
Each Oxygen Sensor circuit consists of an electrochemical cell together with a
resistor encapsulated within a plastic or metal enclosure. For the Dual Sensor
versions two circuits are incorporated in one enclosure. An aperture at the
bottom of the enclosure allows for gas transfer. External connections are made
via an integral plug and socket.
This apparatus is not designed for use in oxygen enriched atmospheres greater
than 21 % oxygen.
Input Parameters (Models OC-2, OC-7, OC-16 to OC-18, OC-21 to OC-24, OC-28,
OC-28M, OC-47, OC-202, OC-207, OC-216, OC-221 and OC-224):
Ui = 30V, Ii= 120mA, Pi= 0.55W, Ci= 0, Li= 0
Input Parameters (Oxyprobe 200, Models OC-19, OC-20, OC-25, OC-25M, OC-26 and
OC-26M): Ui = 12V, Ii= 120mA, Pi= 0.55W, Ci= 1.2µF, Li= 0 16 Report Number
GB/BAS/ExTRl 8.0198
17 Specific Conditions of Use
- Care must be taken when installing equipment with plastic enclosures or plastic parts of enclosures to ensure that the equipment is protected from any situations that could cause a build-up of static charge. The equipment must not be installed into locations in which it could come into contact with, through normal or abnormal circumstances, fast moving dust laden air/gas or non-conductive fluids. The equipment must be cleaned only with a damp cloth.
- The metal enclosure of the OC-25M, OC-26M and OC-28M Oxygen sensors are a potential source of electrostatic discharge. The OC-25M, OC-26M and OC-28M Oxygen sensors must be installed in its end-use application in a manner where the metal enclosure is earthed.
18 Essential Health and Safety Requirements
In addition to the Essential Health and Safety Requirements (EHSRs) covered by
the standards listed at item 9, the following are considered relevant to this
product, and conformity is demonstrated in the report:
Clause Subject
1 .2. 7 L VD type requirements
1.2.8 Overloading of equipment (protection relays, etc.)
1.4.1 External effects
1.4.2 Aggressive substances, etc.
19 Drawings and Documents
New drawings submitted for this issue of certificate:
Number Sheet Issue Date Description E165 1 of I 4 03-07-2018
Sensor Certification Drawing Model OC-17 & 18 I OC-19 & 20
El95 1 of 1 5 15-08-2018
Sensor Certification Drawing Model OC-21, OC-221, OC-22, OC-23, OC-24, OC-25 &
OC-47 E523 1 of 1 0 09-08-2018
Model OC-26 and OC-28 Sensors Certification Drawing
Current drawings which remain unaffected by this issue:
Number Sheet Issue Date Description
El48 1 of 1 G 23/10/2014 Standard Oxygen Sensor Model OC-2, OC-202 & OC- 224
El20 1 of 1 5 30/10/14 Model OC-16/0C-216 Oxygen Sensor Certificate Drawing
E292 1 of 1 4 30/10/2014 Oxyprobe 200 Oxygen Sensor El49 1 of 1 G 30/10/2014
In-Situ Oxygen Probe with Model OC-7, OC-207 Oxygen Sensor Certification
Drawing
Certificate History
Certificate No.| Date|
Comments
---|---|---
BASO2ATEX1230| 3 September 2002| The release of the prime certificate. The
associated test and assessment against the requirements of EN50014:1997 +
Amendments 1 & 2, EN50020:1994 and EN50284 is documented in Test Report No.
02(C)0079.
BASO2ATEX1230/1| 26 November 2003| To permit the inclusion further sensor
types, a change of the first part of the marking to 0 II 1GD and a change
to input parameters for certain types, documented in Test Report No.
03(C)0659.
BASO2ATEX1230/2| 3 March 2005| To permit minor changes to the labelling and
the inclusion of further sensor types, documented in Test Report No.
05(C)0111.
Certificate No.| Date|
Comments
---|---|---
BAS02ATEX1230X Issue 3| 14 May 2010| This issue of the certificate
incorporates previously issued primary & supplementary certificates into one
certificate and confirms the current design meets the requirements of
81460079-0:2004, E1460079- I 1:2007, EN6 I 241-0:2006 & 81461241-11:2006, and
includes revision of the equipment marking in accordance with EN60079-0:2009.
At the same time Pi for the unit has been reduced from 1W to 0.75W. Also, in
order to comply with the later standards, suffix ‘X’ has been added to the
Certificate number. These changes are documented in Test Reports
GB/BAS/ExTR09.0232/00 & GB/BAS/ExTR09.0233/00.
BASO2ATEX1230X Issue 4| 29 July 2010| This issue introduces a new sensor model
known as OxyExtract. This model is documented in Test Report
GB/BAS/ExTRI0.0156/00 and has been added to the list in the main schedule.
BASO2ATEX1230X Issue 5| I October 2010| To permit minor changes to
encapsulation in the OxyExtract model and to confirm that all sensors comply
with EN60079-0:2009. Changes are documented in Test Report
GB/BAS/ExTR10.0222/00
BASO2ATEX1230X Issue 6| 23 November 2010| To permit the name OxyExtract to be
changed to Oxyprobe 200. No Test Report required for this change.
BASO2ATEX1230X Issue 7| 15 July 2011| To permit changes to drawing E148,
including the removal of some internal oxygen cell types and the addition of
Sensor type OC-224. At the same time the Certificate title and the description
in the Schedule have been modified to reflect the way in which type numbers
appear on the certification labels. Changes are documented in Test Report
GB/BAS/ExTR11.0171/00
BASO2ATEX1230X Issue 8| 18 March 2013| To permit various changes to the
Oxyprobe 200 including; Alternative electrochemical cell, minor changes to the
enclosure, the inclusion of additional optional components and to confirm the
new input parameters. To permit additional equipment, 0C-25 to be incorporated
in to this certificate To add an additional of condition of safe use stating
that “Models OC-2, OC-17, 0C-18, OC-19 and OC-20 must be housed inside an
IP-54 rated enclosure.” Changes are documented in Test Report
GB/BAS/ExTR13.0051/00
BASO2ATEX1230X Issue 9| 6 January 2015| To permit modification of the input
parameters and extension of the rated ambient to -20°C to +55°C. To allow the
inclusion of an alternative oxygen sensor forming new models 0C-202, OC-207,
OC-216 and 0C-221 To confirm that the Series of Oxygen sensors has been
assessed against IEC 60079-0:2011 and IEC 60079-11:2011 with respect to the
differences from IEC 60079-0:2004, IEC 60079-0:2007, IEC 60079-11:2006, IEC
61241-0:2004 and IEC 61241-11:2005 and that none of the differences affect the
equipment. Changes are documented in Test Report GB/BAS/ExTR14.0565/00 Change
of Company Name
Certificate No.
| Date|
Comments
BA SO2 ATEXI230X Issue 10| 3 December 2015| To permit the inclusion of an
alternative Oxygen Sensor forming new Model 0C-47. Changes are documented in
Test Report GB/BAS/ExTR15.0350/00
BA SO2ATEX1230X Issue 11| 3 September 2018| To permit the inclusion of an
alternative Oxygen Sensor forming new Models 0C-25M, OC-26, 0C-26M, OC-28 and
0C-28M. Certificate also permits new constructions for sensors OC-19 and
OC-20. Changes are documented in Test Report GB/BAS/ExTRI 8.0198/00. Project
18/0344.
For drawings applicable to each issue, see original of that issue.
Ntron Ltd.
Mullaghboy Industrial Estate,
Navan, County Meath, Ireland.
00353 46 9071333
www.ntron.com
Documents / Resources
|
NTRON SIL O2 Oxygen
Analyser
[pdf] Instruction Manual
SIL O2 Oxygen Analyser, SIL O2, Oxygen Analyser, Analyser
---|---
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>