Bioptic Qsep100 Advance Capillary Gel Electrophoresis Analyzer Instruction Manual

June 17, 2024
Bioptic

Bioptic Qsep100 Advance Capillary Gel Electrophoresis Analyzer

Bioptic Qsep100 Advance Capillary Gel Electrophoresis
Analyzer

Getting start with Qsep100

Getting start with Qsep100
Getting start with Qsep100

Install Qsep100

  1. Unpack Mailbox and tighten the air tube with connector
    Getting start with Qsep100

  2. Plug the power cord into Qairbox
    (* Please check the Qairbox label on the power cord)
    Getting start with Qsep100

  3. Plug the power cord and USB cord into Qsep100
    (Please make sure the label on the power cord is labeled Qsep100)
    Getting start with Qsep100

  4. Connect the other side of USB cord to the operation computer.
    Getting start with Qsep100

  5. Plug the other side of the air tube into Qsep100 and turn on the Mailbox
    Getting start with Qsep100

  6. Remove the protecting foam before turning on Qsep100

Your Qsep100 is READY TO USE
Getting start with Qsep100

*DO NOT switch on the instrument immediately after powering off Wait at least 5 seconds

Operation Quick Start

  1. Turn on the power.
    Operation Quick Start

  2. Double click the icon Q-Analyzer for Qsep100.

  3. Q-Analyzer for Qsep100 software user interface:
    Operation Quick Start

  4. Click “Connect” to link the instrument.
    Operation Quick Start
    Packing list of Cartridge Kit
    Cartridges Alignment Marker Mineral Oil Dilution Buffer Separation Buffer Buffer Tray Droppers 0.2 ml tubes.
    Operation Quick Start

  5. Buffer and Alignment Marker preparation:

  6. Add 1X Separation Buffer into “S” well and add diH2O into the others.

  • Buffer height should reach the groove of wells.
    Operation Quick Start

    1. Alignment Marker preparation:
      Add ≥ 20 μl Alignment Marker (20 bp & 1000 bp, C109100) into the 0.2 ml tube, and then add 10 μl “Mineral Oil” on top of it.

    2. Click “Change Buffer”.
      Operation Quick Start

    3. Place Alignment Marker at MA-1 position and the buffer tray on the tray holder.
      For other Alignment Marker sizes, please place them in corresponding positions.

  • Make sure the buffer tray has been placed correctly on the holder.

  • Centrifuge the Alignment Marker to the bottom of the tube and make sure no air bubble appears.
    Operation Quick Start

    1. Hold the tray holder and use the thumb to press the Alignment Marker tube tightly down into the well.
      Operation Quick Start

    2. Click “Change Sample” and place the samples (≥ 20 μl).

  • Centrifuge the samples to the bottom of the tubes and make sure no air bubble appears.
    Operation Quick Start

    1. Unpack Cartridge:
      Please follow the steps of “Unpacking Guide” in the cartridge box.

    2. Open the cartridge door and insert the cartridge with guiding groove facing the front.
      Operation Quick Start

    3. Close the cartridge door.

  • Press the door and make sure to hear the squeeze sound .
    Operation Quick Start

    1. Click “Latch”.
      Operation Quick Start
      Cartridge information will be presented on the main window after the cartridge is latched.
      Operation Quick Start
      New Cartridge Calibration:
      New cartridge needs to be calibrated before use.
      Please proceed with the following steps.

    2. Click “OK”.
      Operation Quick Start

    3. Click “HV Check”

  • The storage and transportation condition may influence the gel matrix and cause unstable current.
    Check current and see if it is stable or not. After HV Check, if the current is still unstable, please repeat this step 2-3 times.
    Operation Quick Start

    1. Click “Calibrate”
  • Make sure the Alignment Marker has been placed in correct position.
    Operation Quick Start
    During Calibration
    Software will recognize peaks from Alignment Marker signal. DO NOT use Size Marker or DNA sample to perform “Calibrate”.
    Operation Quick Start
    • For troubleshooting, please refer to cartridge unpacking guide for more details.

    1. Click on the blank column and designate ① the sample locations, ② test method, sample duration, runs, ③ result name and ④ Para by following steps 14-1, 14-2, 14-3 and 14-4.
      Operation Quick Start

    2. Click “Sample Position” and mark the position of sample on the plate and then press “OK”.
      Operation Quick Start

    3. Click “Method” to select the analysis method.
      Operation Quick Start

  • If you want to analyze the samples with Alignment Marker, please choose the appropriate Alignment Marker and then place it on the corresponding position. (Check ✓ the box) 3. Enter the result name.
    Operation Quick Start

    1. Click the icon “Para” and set the parameters. (Baseline Factor, Peak Threshold, Calculate, etc.)
      Operation Quick Start

    2. Click “Run” to start the analysis.
      Operation Quick Start
      Brief introduction of the Signal Chart:
      Operation Quick Start
      Please refer to operation manual for more details.

Introduction of the Control Panel

Introduction of the Control Panel

Customers Support

Contact Information:
Company Name: BiOptic Inc.
Office Address: (23141) 5F., No.6, Ln. 130, Minquan Rd., Xindian Dis., New Taipei City, Taiwan
Factory Address: (23141) 4F., No.108-3,Minquan Rd., Xindian Dist., New Taipei City, Taiwan
Tel: +886-2-2218-8726, Fax: +886-2-2218-8727, E-mail: service@bioptic.com.tw

Logo

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals