ACCURIS INSTRUMENTS E3100 MyView UV Transilluminator Instruction Manual
I. Overview
The Accuris E3100 My View UV Transilluminator is a laboratory instrument
designed to provide a uniform source of UV light for the excitation of
fluorescent dyes such as Ethidium Bromide, SmartGlowTM and other nucleic acid
stains. The E3100 is ideally suited for observation, analysis and photo
imaging of stained electrophoresis gels that are positioned on the lighted
viewing surface.
The midrange wavelength of the E3100 Transilluminator is 302nm. (Other
manufacturers of transilluminators may refer to the same wavelength range as
300nm, or 312nm, but the spectral range is the same.)
II. Warnings
⚠ UV Transilluminators are powerful sources of UV radiation, which is
hazardous to the skin and eyes.
⚠ Always make sure the protective UV blocking cover is positioned to prevent
UV exposure to the user, and wear protective clothing and protective glasses
to prevent UV exposure to the skin and eyes.
⚠ Although the UV blocking cover will effectively block UV radiation, it is
always recommended to wear protective UV blocking glasses.
⚠ Use caution when cutting gels on the glass surface. Although this surface is
scratch resistant, use minimal pressure when working with metal blades or
tools.
⚠ To ensure long life of your transilluminator and the UV bulbs, always turn
off the instrument when not in use. It is recommended not to leave the power
on for periods longer than 15 minutes
III. Installation
Install the E3100 Transilluminator on a level and stable surface such as a
laboratory counter, close to an available electrical outlet. The instrument
should be oriented with the power switch and hinges facing the user. Leave
sufficient space around the instrument to allow proper airflow, and take care
not to block the cooling fan on rear side of the housing. Make sure that the
front power switch is set to the off position, and then connect the included
DC power adapter plug to the back of the instrument and to an appropriately
rated electrical outlet.
Note: The E3100 accepts 12V DC input from the included power adapter. This
power adapter accepts 100V to 240V AC input, and different plug cords are
available to match your local outlet type
IV. Operation
- Place a gel or sample on the viewing surface of the transilluminator.
- Position the UV blocking cover to prevent exposure.
- Turn on power to the UV light by pressing the front switch.
V. Imaging using the Accuris SmartDocTM Enclosure
To image gels on the E3100 Transilluminator, it is recommended to use an
imaging enclosure.
The Accuris SmartDocTM enclosure, for imaging using a smart phone camera, is
designed to fit the viewing surface of the E3100. Filters are available for
blocking the UV wavelength to allow image capture with minimal background
fluorescence. Visit: www.accuris-usa.com for
details
VI. Specifications
VII. Cleaning and Care
The glass surface should be cleaned regularly with a cloth dampened with
water, a mild soap solution, or ethanol. In order to extend the working life
of light tubes and UV glass filter, power off the instrument when not in use.
For service, contact Benchmark Scientific’s Service Department at 908-7695555
© Copyright, 2020, Benchmark Scientific
PO Box 709
Edison, NJ 08818
USA
Phone: 908-769-5555
Email: info@accuris-usa.com
www.benchmarkscientific.com / www.accuris-usa.com