KLEIN TOOLS CL1100 AC Clamp with Temperature Voltage Tester Instruction Manual
- June 17, 2024
- KLEIN TOOLS
Table of Contents
Instruction Manual
CL1100
CL1100 AC Clamp with Temperature Voltage Tester
For Professionals…Since 1857®
www.kleintools.com
- DATA HOLD
- BACKLIGHT
- WORKLIGHT
- BAR GRAPH
- MAX / MIN
- 3-3/4 DIGIT 3999 COUNT LCD
- AUTO / MANUAL RANGE
- NON CONTACT VOLTAGE TESTER
- TEMPERATURE
GENERAL SPECIFICATIONS
The Klein Tools CL1100 is an auto-ranging clamp meter. It measures AC / DC voltage, AC / DC current, resistance, capacitance, frequency, duty cycle, and temperature. It can also test non-contact voltage, diodes, and continuity.
- Operating Altitude: 2000 meters
- Humidity: 80% max
- Operating Temperature: 0°C / 32°F to 45°C / 113°F
- Storage Temperature: 0°C / 32°F to 60°C / 140°F
- Accuracy Temperature: 18°C / 64°F to 28°C / 82°F
- Temperature Coeffi cient: 0.1*(specifi ed accuracy) / °C
- Dimensions: 8.875″ x 3.375″ x 1.5″
- Weight: 10.7oz.
- Calibration: Accurate for one year
- CAT Rating: CAT III 600V
- Accuracy: ± (% of reading + # of least signifi cant digits)
WARNINGS
To ensure safe operation and service of the tester, follow these instructions.
Failure to observe these warnings can result in severe injury or death.
- Before each use, verify meter operation by measuring a known voltage or current.
- Never use the meter on a circuit with voltages that exceed the category based rating of this meter.
- Do not use the meter during electrical storms, or in wet weather.
- Do not use the meter or test leads if they appear to be damaged.
- Ensure meter leads are fully seated, and keep fi ngers away from the metal probe contacts when making measurements.
- Do not open the meter to replace batteries while the probes are connected.
- Use caution when working with voltages above 60V DC, or 25V AC RMS. Such voltages pose a shock hazard.
- To avoid false readings that can lead to electrical shock, replace batteries if a low battery indicator appears.
- Unless measuring voltage or current, shut off and lock out power before measuring resistance or capacitance.
- Always adhere to local and national safety codes. Use individual protective equipment to prevent shock and arc blast injury where hazardous live conductors are exposed.
SYMBOLS
**** | AC Alternating Current |
---|---|
**** | DC Direct Current |
DC/AC Voltage or Current | |
Ground | |
AC Source | |
**** | Warning or Caution |
**** | Dangerous levels |
Double Insulated Class II
| Safe for disconnect from live conductors
FEATURE DETAILS
A. Use CAT III rated leads or higher.
Do not attempt to measure more than 1000V DC or 750V AC or 2000µA.
B. Keep hands below line when measuring high current levels.
C. Auto Power-Off (Apo)
- Device will power off after 30 minutes non-use.
- Turn the dial or press a button to wake.
- Disabled during Max / Min function.
- Holding Select button while turning on disables Auto Power-Off.
D. Select Functionality Button
- Switch between AC and DC.
- Switch between and .
- Switch between ˚F and ˚C.
E. Hold / Backlight / Worklight
- Press to hold the current input on the display.
- Press again to return to live reading.
- Press for 2 seconds to enable / disable lights.
- Using lights drains the battery signifi cantly.
F. Auto / Manual Range
- Press repeatedly to cycle through manual ranges.
- Press for 2 seconds to return to auto ranging mode.
- AT is displayed on LCD only during auto ranging mode.
G. Max / Min Hold
- Press to enter Max / Min mode; the largest and smallest values will be saved while in this mode.
- Press repeatedly to alternate between the maximum and minimum readings.
- Press for 2 seconds to return to live reading and clear the stored maximum and minimum values.
H. I. Battery Replacement
- When indicator is displayed on the LCD, batteries must be replaced.
- Remove the back screw and replace 2 x AAA batteries.
J. Magnetic Back
- Attach instrument to metal for hands-free use.
K. Bar Graph
- The bar graph shows an approximate analog representation of a measurement.
- The bar graph responds much faster than the digital display.
- The scale of the bar graph is zero to the maximum reading of the selected range.
L. Temperature Input Jack
- Input jack for K-type thermocouple probe.
M. Temperature Switch
- Move the switch down to measure temperature, and up to measure all other functions that require standard test leads. Test leads must be removed before using the temperature function.
N. Temperature Adjustment
- Remove battery cover.
- Place temperature probe in distilled water with stirred crushed ice (a standard for 0°C and 32°F).
- Use a fine tip screwdriver to adjust the screw recessed in the bottom right corner of the battery cavity as shown until the display reads 32°F.
FUNCTION INSTRUCTIONS
-
AC Current (large): < 400A
Features: • Center wire in guides for best accuracy.
• Opposing currents cancel (use line-splitter when necessary). -
AC / DC Current (small): <2000µA
Features: • **** Select AC or DC current source.
• Current above 2000µA will damage instrument. -
AC / DC Voltage: <750V AC or 1000V DC
Features: • Select AC or DC voltage source. -
Frequency (Hz) / Duty Cycle (See Feature Details)
Features: • Select , ,or setting.
• Frequencies greater than 500 kHz will display “0.000Hz”. -
Capacitance: < 4000µF
Features: • Safely discharge capacitor before measurement.
• Reading may take up to 60 seconds for large capacitors. -
Resistance: <40MΩ
Features: • **** Do not measure resistance on a live circuit. -
Diode / ContinuityDiode / Continuity Features: Continued on Reverse
Diode Features:
Display shows:
• Forward voltage drop if forward biased.
• “O.L.” if reverse biased.
Continuity Features: • Display shows resistance.
• Buzzer sounds if less than 30Ω. -
Non Contact Voltage (NCV): > 25V AC
-
Temperature
Features: • **** Do not apply voltage to thermocouple.
• Fahrenheit range: -22° to 752°F
• Celsius range: -30° to 400°C
SYMBOLS USED ON LCD
**** | AC Measurement | **** | DC Measurement |
---|---|---|---|
Negative DC Value | AT | Auto Range Active | |
O.L. | Overload: Range Exceeded | Apo | Auto Power-Off Active |
Low Battery | Hold Active | ||
MIN | Minimum Reading | MAX | Maximum Reading |
% | Duty Cycle Mode | Hz | Frequency Mode |
V | Voltage Measurement | A | Current in Amps |
Resistance in Ohms | Diode Test | ||
F | Capacitance in Farads | Continuity Test | |
n | Nano 10-9 | µ | Micro 10 -6 |
m | Milli 10-3 | k | Kilo 10 3 |
M | Mega 10 6 | °C | Degrees Celsius |
°F | Degrees Fahrenheit |
ELECTRICAL SPECIFICATIONS
DC Voltage Measurement
Range | Resolution | Accuracy |
---|---|---|
400mV ∼ 400V | 0.1mV ∼ 0.1V | ± (0.5% + 4 digits) |
1000V | 1V | ± (0.8% + 10 digits) |
Overload Protection: 1000V
AC Voltage Measurement
Range | Resolution | Accuracy |
---|---|---|
400mV ∼ 750V | 0.1mV ∼ 1V | ± (2.0% + 5 digits) |
Overload Protection: 750V RMS
Frequency Response: 40 to 400 Hz
Minimum Voltage for Frequency Measurement: 200mV
Response: Averaging
DC Current Measurement
Range | Resolution | Accuracy |
---|---|---|
400µA | 0.1µA | ± (1.2% + 3 digits) |
2000µA | 1µA |
Overload Protection:
- Voltage: 600V RMS
- Current: 2000µA
AC Current Measurement
Range | Resolution | Accuracy |
---|---|---|
400µA | 0.1µA | ± (2.0% + 5 digits) |
2000µA | 1µA | ± (1.5% + 5 digits) |
40A | 0.01A | ± (2.9% + 15 digits) |
400A | 0.1A | ± (1.9% + 8 digits) |
Overload Protection:
- Voltage: 600V RMS
- Current: 2000µA
Frequency: 45 to 400Hz
Minimum Current for Frequency Measurement: 400µA or 20A
Response: Averaging
Resistance Measurement
Range | Resolution | Accuracy |
---|---|---|
400Ω ∼ 4MΩ | 0.1 Ω ∼ 0.001MΩ | ± (1.0% + 4 digits) |
40MΩ | 0.01MΩ | ± (2.0% + 4 digits) |
Overload Protection: 600V RMS
Capacitance Measurement
Range | Resolution | Accuracy |
---|---|---|
40nF ∼ 4000µF | 0.01nF ∼ 1µF | ± (3.5% + 6 digits) |
Overload Protection: 600V RMS
Frequency Measurement
Range | Resolution | Accuracy |
---|---|---|
99.99Hz ∼ 499.9kHz | 0.01Hz ∼ 0.1kHz | ± (0.1% + 4 digits) |
Overload Protection: 600V RMS
Sensitivity: 1.8V RMS
Duty Cycle Measurement
Range | Resolution | Accuracy |
---|---|---|
0.1 ∼ 99.9% | 0.1% | ± (0.2% per kHz + 0.1% + 5 digits) |
Overload Protection: 600V RMS
Diode Test
Overload Protection| Range| Test Current| Open Circuit
Voltage
---|---|---|---
600V RMS| 2.0V| Appx. 0.25mA| 1.6V DC
Continuity Test
Overload Protection | Open Circuit Voltage |
---|---|
600V RMS | 0.44V |
Temperature Measurement
Range | Resolution | Accuracy |
---|---|---|
-22°~14°F (-30°~-10°C) | 0.1°F (0.1°C) | ±5.4 F (±3.0 C) |
15°~752°F (-9°~400°C) | 0.1°F (0.1°C) | ±(1.0% + [3.6 F or 2.0 C]) |
Overload Protection: 30V RMS
Non Contact Voltage Detector
On Voltage
Appx. 25V AC
WARRANTY
See www.kleintools.com , contact Klein Tools at 1-877-775-5346, or visit a distributor for warranty information.
CLEANING
Turn instrument off and disconnect test leads. Clean the instrument by using a damp cloth. Do not use abrasive cleaners or solvents.
STORAGE
Remove the batteries when instrument is not in use for a prolonged period of time. Do not expose to high temperatures or humidity. After a period of storage in extreme conditions exceeding the limits mentioned in the Specifi cations section, allow the instrument to return to normal operating conditions before using it.
DISPOSAL / RECYCLE
Caution: This symbol indicates that equipment and its accessories shall be subject to a separate collection and correct disposal.
CUSTOMER SERVICE
KLEIN TOOLS, INC.
450 Bond Street
Lincolnshire, IL 60069
1-877-775-5346
www.kleintools.com
139660
Documents / Resources
|
KLEIN TOOLS CL1100 AC Clamp with Temperature Voltage
Tester
[pdf] Instruction Manual
CL1100 AC Clamp with Temperature Voltage Tester, CL1100, AC Clamp with
Temperature Voltage Tester, Clamp with Temperature Voltage Tester, Temperature
Voltage Tester, Voltage Tester, Tester
---|---
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>