Liko M230 Mobile Patient Lifts Installation Guide
- June 16, 2024
- Liko
Table of Contents
- Liko M230 Mobile Patient Lifts
- Product Information
- FAQ
- Product Description
- Safety Instructions
- Definitions
- Technical Data
- Dimensions
- Assembly
- Operation
- Position of the Lift when Lifting
- Charging the Batteries
- Alternative Charging Procedures
- Maximum Load
- Recommended Lifting Accessories
- Simple Troubleshooting
- Inspection and Maintenance
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Liko M230 Mobile Patient Lifts
Product Information
Specifications
- Model: Liko M220, Liko M230
- Product Numbers: 2050010 (Liko M220), 2050015 (Liko M230)
- Compliance: Medical Device Directive 93/42/EEC, EN ISO 10535:2006, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ANSI/AAMI ES60601-1, CAN/CSA C22.2 no 60601-1
Product Description
The LikoTM M220 and LikoTM M230 are patient lifts designed to assist caregivers in lifting and transferring patients. The person being lifted is referred to as the patient, while the person assisting is referred to as the caregiver.
The lifts feature a warning triangle that indicates situations requiring extra care and attention. It is important to read and understand the instruction guide for both the patient lift and lifting accessories before use. The equipment should only be used by trained personnel. If there are any uncertainties or questions, please contact a Liko/Hill-Rom representative.
Assembly
Follow these steps to assemble the Liko M220/M230:
- Ensure all parts are present and undamaged.
- Attach the lift mast to the base using the provided screws.
- Securely tighten the screws to ensure stability.
- Attach the handles to the lift mast.
- Check that all connections are secure before use.
Operation
To operate the Liko M220/M230:
- Make sure the patient is positioned properly on the underlying surface before lifting.
- Use the handles to maneuver the lift. Do not apply force to the lift arm or directly to the lift mast, as this may cause a tilting hazard.
- Avoid unbalanced lifting, as it poses a tipping risk and may damage the lift equipment.
- Never leave a patient unattended during a lifting situation.
Charging the Batteries
Follow these steps to charge the batteries:
- Connect the charger to a power source.
- Connect the charger cable to the lift’s battery charging port.
- Allow the batteries to charge fully before use.
Maximum Load
The maximum load capacity of the Liko M220/M230 is determined by its specifications. Ensure that the weight of the patient and any additional equipment does not exceed this limit.
Recommended Lifting Accessories
When using the Liko M220/M230, it is recommended to use compatible lifting accessories. These accessories are designed to enhance safety and assist in different lifting scenarios. Refer to the instruction guide for a list of recommended lifting accessories.
Simple Troubleshooting
If you encounter any issues with the Liko M220/M230, refer to the troubleshooting section of the instruction guide. It provides simple solutions for common problems that may arise during operation.
Inspection and Maintenance
Regular inspection and maintenance are essential for ensuring the proper functioning and longevity of the Liko M220/M230. Followthe guidelines provided in the instruction guide for inspection schedules and maintenance procedures.
FAQ
Q: Can the lift be modified?
A: Under no circumstances should the lift be modified. Any modifications may compromise safety and void the warranty. For more information, please contact Liko/Hill-Rom.
Q: What precautions should I take when using powerful sources of electromagnetic disturbance?
A: When using powerful sources of electromagnetic disturbance, such as diathermy cables, ensure they are not positioned on or near the lift. If you have any questions or concerns, consult a responsible assistive device technician or the supplier.
7EN150106-06
2013-12-18
Applies to the following models:
Liko M220 Prod. no. 2050010
Liko M230 Prod. no. 2050015
Product Description
- Liko M220 and M230 are easy-to-use mobile lifts primarily intended for use in nursing homes and home environment. Both models are excellent aids for daily transfers of adults and children; for instance, for transfers to and from a wheelchair, toilet and floor.
- The models differ in the way the base width is adjusted while both lifts feature an electric lift mechanism. Liko M230 has an electric base and Liko M220 a manual base for opening and closing the legs.
- An individually fitted sling and other accessories are of utmost significance for performance and safety when using Liko’s lifts.
- In this document, the person being lifted is referred to as the patient, and the person helping them is referred to as the caregiver.
- is a warning triangle used for situations which require extra care and attention.
IMPORTANT!
Read the instruction guide for both the patient lift and lifting accessories
before use. Lifting and transferring a person always involves a certain level
of risk. It is important to completely understand the contents of the
instruction guides. The equipment should only be used by trained personnel.
Please contact your Liko/Hill-Rom representative in the event of any
uncertainties or questions.
NOTE!
This instruction guide contains important information for the user of the
product.
All who use the product should review and completely understand the contents
of the instruction guide. Remember to keep the instruction guide in a place
where it is always available to those using the product.
WARNING
Certain factors may limit the ability of the mobile lift to perform correctly,
which may include the following: floor thres-holds, multiple heights of the
floor surfaces, obstacles, and extra thick pile carpet. These factors can lead
to the mobile lift not rolling as intended, possible imbalance of the mobile
lift, and additional strain placed on the caregiver. If you are uncertain if
your care environment meets the intended requirements please contact your Liko
/Hill-Rom representative for further assistance.
Use the handles to maneuver the lift. Do not apply force to the lift arm or
directly to the lift mast to maneuver; this may cause a tilting hazard.
Safety Instructions
Before using, make sure that:
- the lift is assembled in accordance with the assembly instructions;
- the lifting accessories are properly attached to the lift;
- charge the batteries for at least 6 hours;
- you have read the instruction guides for the lift and lifting accessories;
- personnel using the lift are informed of the correct operation and use of the lift.
Before lifting, always make sure that:
- the lifting accessories are not damaged;
- the lifting accessories are correctly attached to the lift;
- the lifting accessories hang vertically and can move freely;
- the lifting accessories are selected appropriately, in terms of type, size, material and design, with regard to the patient’s needs;
- the lifting accessories are correctly and safely applied to the patient in order to prevent injuries;
- the sling bar latches are intact. Missing or damaged latches must always be replaced with new ones;
- the sling’s strap loops are correctly connected to the sling bar hooks when the sling straps are stretched up but before the patient is lifted from the underlying surface.
Unbalanced lifting poses a tipping risk and may damage the lift equipment!
Never leave a patient unattended during a lifting situation!
Liko M220 and M230 have been tested by an accredited testing institute and
fulfill the requirements specified in the Medical Device Directive 93/42/EEC,
for Class I products.
Liko M220 and M230 comply with the requirements in EN ISO 10535:2006, IEC
60601-1, IEC 60601-1-2, IEC 60601-1-11, ANSI/AAMI ES60601-1, and CAN/CSA C22.2
no 60601-1.
Under no circumstances must the lift be modified. Please contact Liko/Hill-Rom
for more information.
Particular care must be taken when using powerful sources of electromagnetic
disturbance, so that, for example, diathermy cables are not positioned on or
near the lift. If you have any questions, please consult the responsible
assistive device technician or the supplier.
Definitions
Technical Data
Dimensions
Table of measurements
(mm)
A max | A min | B | B1 | B2* | B3 | D1 | *D2 | E | F | F1 | Lmax | Lmin | M | N | O | ||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
max | min | max | min | ||||||||||||||
1895 | 1330 | 1160 | 892 | 645 | 605 | 1020 | 682 | 970 | 555 | 900 | 210 | 1035 | 105 | 25 | |||
1720 | 570 | 455 | 245 | 455 |
(inch.)
A max | A min | B | B1 | B2* | B3 | D1 | *D2 | E | F | F1 | Lmax | Lmin | M | N | O | ||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
max. | min. | max. | min. | ||||||||||||||
74.6 | 52.4 | 45.7 | 35.1 | 25.4 | 23.8 | 40.2 | 26.9 | 38.2 | 21.9 | 35.4 | 8.3 | 40.7 | |||||
4.1 | 1.0 | 67.7 | 22.4 | 18.0 | 9.6 | 18.0 |
Note: When changing to other lifting accessories check that the lift still achieves desired lifting height.
- Reference measurement according to Standard EN ISO 10535:2006
Assembly
Remove the two transport safety straps, on the sling bar and on the lift mast.
Before assembly, make sure you have the following parts:
- lift mast with lift arm and sling bar with latches, lift motor (incl. actuator);
- handle
- base with electric (M230) or manual (M220) adjustment;
- control box, charger cable, hand control with cable;
- instruction guide and quick reference guide with chain;
- 4 screw M10x25,
- 1 screw M5,
- 2 screw M10x16,
- 1 casing (M10),
- 1 Allen wrench 3 mm,
- 2 Allen wrench 6 mm.
-
Place the base on floor with wheels contacting the floor. Lock both rear wheels.
Place the lift mast in the foot of the base. -
Secure the lift mast, using two of the M10x25 screws and the 6 mm Allen wrench provided. Make sure screws are tight.
-
Place the control box in the bracket on the lift mast. Lock the control box in place using the M5 screw and Allen key 3 mm provided. Do not over tighten the screw.
-
Install handles at the lift mast:
A: attach in the lift mast with two M10x25 screws and the 6 mm Allen wrench provided.
(Remove decal)
B: insert the casing (M10) in the forward attachment of the handles, use two M10x16 and the two 6 mm Allen wrenches provided Make sure the handles all four screws are tight! -
Place the quick reference guide (use the provided chain) and the hand control on the handles.
-
Connect cables to the control box, see illustration. Make sure plugs are fully seated.
-
Connect the charger cable to the socket under the control box. Make sure the plug are fully seated.
Attach the cable, see illustration. -
Connect the battery and secure it to the control box bracket. A click sound can be heard when the battery is installed correctly.
-
9. Reset the emergency stop by turning the button in the direction of the arrows on the button. Charge the battery, see “Charging the Batteries” at page 8.
After assembly, make sure that:
- the battery of the lift is charged;
- the motion of the lift arm corresponds to the buttons on the hand control;
- the manual and electrical emergency lowering works;
- the adjustment of the base width is functional;
- the rear wheels brakes is functional.
Operation
-
Emergency Stop
Use in urgent need to stop all of the lifts electrical functions.
Activate: Press the red button on the control box.
Reset: Turn the button in the direction of the arrows until the button springs out. -
Manual Emergency Lowering
Turn in the direction of the arrows on the red cylinder. -
Electrical Emergency Lowering
Push a narrow object into the hole marked on the control box (marked Emergency).
The object used to press must not be sharp, since this may cause damage to the control box!
Hand control
The lifting motion is operated with the push buttons on the hand control. The
direction of the arrows applies when the hand control is held as shown in the
picture. To raise or lower the lift arm, press .
The lifting motion stops when the push button is released.
For electrical adjustment of the base width (Liko M230), push .
Manual Adjustment of the Base (Liko M220) Press the raised pedal to adjust the base width.
Never move the lift by pulling on the actuator!
Wheels with brakes
During the lifting of a patient, the wheels should be unlocked in order for
the lift to be able to center around the patient’s center of gravity The rear
wheels can be locked against rotation and turning. To lock the wheel, press
down the lock pedal. To release the brake, lift up the lock pedal with your
foot.
Installation of Latches
After installation, ensure that the spring loaded latches is taut against the
sling bar and moves freely in the sling bar hook.
Lift correctly!
Before each lift, make sure that:
- the Sling loops at opposite sides of the Sling are at the same height
- all the Sling loops are fastened securely in to the Sling bar hooks
- the Sling bar is level during the lift, see Figure 1.
If Sling bar is not level (see Figure 2), lower the user to a firm surface and adjust according to the Sling in use Instruction Guide.
An improper lift can be uncomfortable for the user and cause damage to the lift equipment! (see Figure 2).
Position of the Lift when Lifting
Charging the Batteries
Indications for Charging the Battery
In the event of low battery voltage, a signal from the control box will sound.
The sound will occur only when one button on the hand control is pushed in.
When this happens, the battery must be charged as soon as possible. However,
there is still sufficient power for additional lifts.
Battery Charging and Maintenance
-
To get maximum battery life, it is important to charge the battery regularly. We recommend charging after use of the lift or every night. Batteries are fully charged after approx. 6 hours. When the battery is fully charged, the charger is switched off automatically. If the lift is not used every day, we recommend connecting it to the charger, or pressing the emergency stop after use, in order to turn off the power and save the battery. Ensure that the battery is fully charged before pressing the emergency stop.
NOTE: The lift cannot be charged with the emergency stop engaged.
NOTE: When charging, a yellow indicator light on the control box will illuminate. The yellow light turns off when the battery is fully charged. If the light has not turned off after 8 hours of charging, the battery may need to be replaced. Stop charging and replace the battery. -
Never charge batteries in a wet area.
Alternative Charging Procedures
Built-in Charger (standard):
Connect the charger cable to an electrical outlet (100-240 V AC). Make sure
that both lights on the control box are lit. The yellow light indicates
charging and the green light indicates that power is being supplied to the
charger. If the charger cable is beginning to stretch, it should be replaced
in order to minimize the risk of the cable getting caught and breaking.
NOTE: The lift cannot be used when the charger cable is plugged into an electrical outlet.
Wall-Mounted Charger or Table Charger:
Remove the battery from the control box by releasing the locking device on top
of the battery.
- Alt. A. (accessory) Place the battery on the wall-mounted charger. Plug the charger into an electrical outlet (100-240 V AC). Make sure that both lights on the charger are lit. The yellow light indicates charging and the green light indicates that power is being supplied to the charger.
- Alt. B. (accessory) Place the battery on the charger in the table charger housing. Plug the charger into an electrical outlet (100-240 V AC). Make sure that both indicator lights on the charger are lit. The yellow light indicates charging and the green light indicates that power is being supplied to the charger.
Liko M220 and M230 comply with the Directive 2002/96/EC on waste electrical and electronic equipment. Old batteries are to be deposited at the nearest recycling facility in accordance with local regulations or given to personnel authorized by Liko/Hill-Rom.
Maximum Load
Different maximum loads may apply to different products on the assembled lift unit: lift, sling bar, sling and any other accessories used. For the assembled lift unit, the maximum load is always the lowest maximum load rating for any of the components. For example, a Liko M220/Liko M230 which is approved for 182 kg (400 lbs) can be equipped with a sling which is approved for 200 kg (440 lbs). In this case, the maximum load of 182 kg (400 lbs) applies to the assembled lift unit. Check the markings on the lift and lifting accessory or contact your Liko/Hill-Rom representative if you have any questions.
Recommended Lifting Accessories
Using lifting accessories other than those approved can entail a risk.
When changing lifting accessories, the highest possible lifting height of the
lift is affected. Therefore, before changing lifting accessories you should
always ensure that the lift, after change, can fulfil the desired lifting
height in order
to manage the lifting situations for which the lift is to be used.
For Liko M220 and Liko M230 all slings compatible with Universal SlingBar 450
are recommended. To select suitable slings and other lifting accessories,
refer to the brochure ”Lifting accessories”. For additional guidance in
selecting a sling, refer to the instruction guides for the respective sling
models. Here, you will also find guidance for combining Liko’s sling bars with
Liko’s slings.
Contact your Liko/Hill-Rom representative or visit
www.liko.com for advice and information on Liko’s
product range.
Simple Troubleshooting
The lift doesn’t work up/down.
Electrical adjustment of the base width does not work, in/out (Liko M230
only).
- Check that the emergency stop has not been engaged.
- Make sure that the cables to the control box are connected correctly.
- Make sure that the charging cable is not connected to an electric outlet.
- Check that the battery is charged.
- Check that the battery’s contact plates are not defective or broken.
- If the problem persists, please contact Liko/Hill-Rom.
The charger doesn’t work.
- Check that the emergency stop has not been engaged.
- Check that the battery’s contact plates are not defective or broken.
- If the problem persists, please contact Liko/Hill-Rom.
The lift is stuck in the high position.
- Check that the emergency stop has not been activated.
- Use the electrical emergency lowering device to lower the patient onto a firm surface.
- Use the manual emergency lowering device to lower the patient onto a firm surface.
- Check the battery voltage.
- If the problem persists, please contact Liko/Hill-Rom.
If you hear unusual sounds.
Contact Liko/Hill-Rom.
Inspection and Maintenance
Care and Maintenance
For trouble-free use, certain details should be checked each day the lift is
used:
- Inspect the lift and check to make sure that there is no external damage.
- Check the functionality of the latches.
- Check the integrity of the lifting motion and the base-width adjustment.
- Check to make sure that the emergency lowering (both electrical and manual) works.
- Charge the batteries each day the lift is used and make sure the charger works.
Please contact your Liko/Hill-Rom representative in the event of any uncertainties or questions.
When necessary, clean the lift with a moist cloth and check that the wheels
are free from dirt. Find more
detailed information regarding cleaning and disinfection of your Liko/Hill-Rom
product in the document Care and Maintenance at our website: www.liko.com.
The lift should not be exposed to running water.
Service
Liko M220/Liko M230 should be periodically inspected at least once a year.
Periodic inspection, repair and maintenance should be performed only in accordance with the Liko Service Manual and by personnel authorized by Liko and using original Liko spare parts.
Service Agreement
Liko/Hill-Rom offers the opportunity to enter into service contracts for the
maintenance and regular inspection of your Liko product.
Expected Life Time
The product has an expected life time of 10 years when correctly handled,
serviced and inspected in accordance with Liko’s instructions.
Transport and Storage
During transportation, or when the lift is not to be used for a long time, the
emergency stop should be engaged. The environment where the lift is
transported and stored should have a temperature of -20 to +40 °C
(-4 °F to + 104 °F) and a relative humidity of 10 to 95 %. The air pressure
should be 700–1060 hPa.
Recycling
For instructions on how to recycle your Liko product, please visit our
website: www.liko.com.
Product Changes
Liko/Hill-Rom reserves the right to make product changes without prior notice.
Contact your Liko/Hill-Rom representative for advice and information about
product upgrades.
Design and Quality by Liko in Sweden
Hill-Rom´s Management system are certified in accordance with ISO 9001 and its
equivalent for the medical device industry, ISO 13485. Hill-Rom´s Management
system is also certified in accordance with environmental standard ISO 14001.
Manufacturer
Liko AB
Nedre vägen 100 SE-975 92 Luleå Sweden