Welch-Ilmvac 2522 Vacuum Pump Owner’s Manual

June 16, 2024
WELCH-ILMVAC

Welch-Ilmvac-LOGO

Welch-Ilmvac 2522 Vacuum Pump

Welch-Ilmvac-2522-Vacuum-Pump-PRODUCT

Product Information

Specifications

  • Product Name: WELCH -I LM V AC
  • Not recommended for pumping acid, base or organic vapors or gases
  • Motor includes a self-resetting thermal cutout

FAQ

  • Q: Can I use this pump to pump acid or base?
    • A: No, this pump is not recommended for pumping acid or base.
  • Q: Can I use this pump to pump organic vapors or gases?
    • A: No, this pump is not recommended for pumping organic vapors or gases.
  • Q: Where can I find contact information for Gardner Denver sales office outside of the U.S. and Canada?
    • A: You can find contact information for your local Gardner Denver sales office on the back page of the user manual or on the Welch-Ilmvac website.
  • Q: What is the phone number for technical support?
    • A: The phone number for technical support is 847-676-8800.
  • Q: What is the email address for technical support?

Product Usage Instructions

Safety Information

Warning to prevent injury

  • Never operate this product if it has a damaged cord or plug. If it is not working properly, has.

Warning to reduce risk of electrical shock

  • Do not disassemble. Disassembly or attempted repairs if accomplished incorrectly can create electrical shock hazard.
  • Refer servicing to qualified service agencies only.

Warning to reduce risk of electrocution

  • Do not use this product in or near an area where it can fall or be pulled into water or other liquids.
  • Do not reach for this product if it has fallen into liquid. Never operate this product outdoors in the rain or in a wet area.

Warning to reduce risk of explosion or fire

  • Do not use this pump in or near explosive atmospheres or where aerosol (spray) products are present.

Introduction

  • This section provides an introduction to the installation process.

Unpacking

  • This section provides instructions on how to unpack the product.

Pump mounting

  • This section provides guidance on how to mount the pump.

Pump location

  • This section provides recommendations on where to place the pump.

Intake and discharge provisions

  • Refer to the accompanying documents for instructions on how to properly identify the intake and discharge of the pump.

Electrical power

  • This section provides information on the electrical power requirements for the product.

Vacuum connections

  • This section provides instructions on how to make vacuum connections.

Vacuum and pressure gauges

  • This section provides information on using vacuum and pressure gauges with the product.

Traps

  • This section provides guidance on using traps with the product.

Operation

Starting procedures

  • This section provides instructions on how to start the product.

Leak detection

  • This section provides guidance on detecting leaks in the system.

Pressure range

  • This section provides information on the pressure range of the product.

Precautions

WARNING

  • Not recommended for pumping acid, base or organic vapors or gases.
  • Be sure to properly identify intake and discharge before using the pump. See Section 2.50

For outside of U.S. and Canada, contact your local Gardner Denver sales office.

INSTRUCTION WARNING AND CAUTION

PLEASE READ BEFORE OPERATION

While reading your manual, please pay close attention to areas labeled: WARNING AND CAUTION.
The description of each is found below.

WARNING: Warnings are given where failure to observe instruction could result in injury or death to people.

CAUTION: Cautions are found where failure to observe the instruction could result in damage to the equipment, associated equipment and process.

These units confi rm to the SI International system of units of measurement.
The following symbols (with recommendation of IEC1010) of warning will be found on the pump.

  • Caution – Refer to accompanying documents
  • Caution – Risk of electrical shock
  • Caution – Hot surface

WARNING

  • Motor includes a self resetting thermal cutout and the pump could restart without actuation under fault condition.

SAFETY INFORMATION

WARNING: TO PREVENT INJURY…

  1. Never operate this product if it has a damaged cord or plug. If it is not working properly, has been dropped, damaged or has fallen into water, please return the product to a Welch service center for examination and repair.
  2. Keep the cord away from heated surfaces. All electrical products generate heat. To avoid serious burns never touch unit during or immediately after operation.
  3. Never block any air openings or place it on a soft surface where the openings may be blocked. The air openings are for ventilation of the motor inside the housing. Keep all air openings free of lint, dirt and other foreign objects.
  4. Never drop or insert fi ngers or any other object into any openings.
  5. Do not operate this product where oxygen is being administered.
  6. This pump is thermally protected and can automatically restart when the protector resets. Always disconnect power source before servicing.
  7. Wear safety glasses and goggles when operating this product. Never point any air nozzle or air sprayer toward another person or any part of the body.
  8. Use only in well ventilated areas. The motor on all pumps are totally enclosed fan cooled
  9. Do not use any tools or attachments without fi rst determining maximum air pressure for that tool or attachment. Be sure to properly identify intake and discharge before using pump. See Section 2.50
  10. WARNING: TO REDUCE RISK OF ELECTRICAL SHOCK…
  11. Do not disassemble. Disassembly or attempted repairs if accomplished incorrectly can create electrical shock hazard. Refer servicing to qualified service agencies only.
  12. Unit is supplied with a three pronged plug. Be sure to connect pump to a properly grounded outlet only.
  13. WARNING: TO REDUCE RISK OF ELECTROCUTION…
  14. Do not use this product in or near area where it can fall or be pulled into water or other liquids.
  15. Do not reach for this product if it has fallen into liquid.
  16. Never operate this product outdoors in the rain or in a wet area.
  17. WARNING: TO REDUCE RISK OF EXPLOSION OR FIRE…
  18. Do not use this pump in or near explosive atmospheres or where aerosol (spray) products are being used.
  19. Do not pump anything other than atmospheric air.
  20. Do not pump combustible liquids or vapors with this product or use in or near an area where flammable or explosive liquids or vapors may exist.
  21. Do not use this product near flames.

WARNING

  • Failure to observe the above safety precautions could result in severe bodily injury, including death in extreme cases.

INSTALLATION

INTRODUCTION

  • This manual has been compiled not only for the care and maintenance of the Welch Dry Dual Funktion Pressure/Vacuum Pump now in your possession, but as a helpful reference and guide to prevent many problems which occur if used improperly

UNPACKING

  • Carefully remove the Pressure/Vacuum Pump from the shipping case and unfasten and remove the wooden skid.
  • Preserve all paperwork for future reference. If damage has occurred from shipment, a claim must be filed with the carrier immediately; preserve the shipping carton for inspection by the carrier.
  • If you are required to communicate with your dealer or Welch Vacuum Technology, be sure to include your order numbers for quick identifi cation. Do not return the pump to the factory without fi rst calling for a return goods number.

PUMP MOUNTING

  • Rubber feet are attached to the pump.
  • Rubber feet are excellent for applications involving a semi-flexible surface such as a bench top; they help to isolate noise and eliminate creeping.
  • All Pressure/Vacuum Pumps should be mounted on a horizontal plane.

PUMP LOCATION

The Pressure/Vacuum Pump should be located preferably in a clean, dry, and well ventilated area. Please be sure not to block the ventilation holes located on the motor housing. The pump should be placed where the surrounding temperature remains between 10°C and 40°C (50°F and 104°F). Always check to insure the location chosen is protected from direct or indirect moisture contact. Welch recommends that the pump be installed at the highest point within the system to prevent possible water condensate from entering the pump. The pump should be located as closely to its system in order to utilize it most effi ciently.

WARNING

  • The motor is thermally protected and will automatically restart unexpectedly when the overload device resets.
  • Don’t pump flammable or explosive gases or vapors or operate this pump in an atmosphere containing flammable or explosive gases or vapors.

INTAKE AND DISCHARGE PROVISIONS

Welch-Ilmvac-2522-Vacuum-Pump-FIG-4

The Two regulators and two gauges make it easy to meet your vacuum and pressure requirements. The vacuum regulator and vacuum dial gauge is attached to the intake port. The vacuum regulator allows the vacuum level to be set between roughly atmospheric pressure and the maximum vacuum allowed for the pump model. A moisture trap keeps water from accidentally being drawn into the unit. The pressure regulator and pressure dial gauges are attached to the discharge port. The pressure regulator allows discharge pressure to be set between atmospheric and the maximum possible for the pump model. Discharge air is filtered and noise is muffled with a filter.

ELECTRICAL POWER

Power Source Review

Review the power source and the motor rating to be sure they agree in voltage, phase, and frequency. Serious damage may occur to the motor if it is connected to an improper voltage. All Welch pumps must be grounded. Grounding reduces the risk of electric shock in the event of an electrical short circuit. The plug must be plugged into a properly grounded outlet. Consult your local electrical codes if you have doubts.

Overload Protection

Motor thermal overload protection is made available by the motor manufacturer as an aid to minimize motor failure. Overload protection is a standard feature on both 50 and 60 Hz singlephase- motors. The motors have automatic overload protection. Automatic reset protection is designed to reset itself after a predetermined cooling period. If the fault to the drive remains unaltered, the motor will cycle on and off until the fault is corrected.

VACUUM CONNECTIONS

All pressure/vacuum pumps come with intake and discharge hose barbs which accept 1/4” ID rubber pressure/vacuum hose. Hose clamps should be used to hold the hose in place. Since all three models operate in the viscous fl ow regime, the small diameter of the hose will generate minimal conductance loss. For best results, Welch recommends the length of the tubing between the pump and the chamber be kept as small as possible.

VACUUM AND PRESSURE GAUGES

All models of pressure/vacuum pumps come with dial gauges mounted on the regulator assemblies. The vacuum gauge gives negative pressure that is pressure below atmospheric. The reference point for the vacuum gauge is atmospheric pressure. The pressure gauge gives pressure above atmospheric. The reference point for the pressure gauge is atmospheric pressure. Please keep in mind that atmospheric pressure tends to vary from day to day. As a result of this variability, the dial vacuum gauge will indicate slightly different maximum vacuum readings from day to day.

TRAPS

The need for a Trap

Welch-Ilmvac-2522-Vacuum-Pump-FIG-5

The pumps will handle humid air. All wetted aluminum parts are treated for corrosion protection from moisture. All other wetted parts are stainless steel. A plastic trap with a ball check valve is attached to the regulator assembly to prevent water condensate from accidentally being ingested into the pump. If there is a chance liquid may be drawn from the process under evacuation, Welch recommends a liquid trap be placed between the process and the pump. A simple liquid trap is a filtering flask. See figure below.

When a heavy load of water vapor is evolved from the vacuum process, a cold trap is recommended to help prevent damage to the pump mechanism. The cold trap, immersed in a suitable Dewar flask, is installed so that the water vapors may come in contact with the surfaces of the trap and condense. Commonly used refrigerants are liquid nitrogen or dry ice and acetone or alcohol. Dry ice provides sufficient cooling to freeze out most heavy water vapor loads. A variety of cold traps are available from Welch. Please call our customer service department for additional information at 847-676-8800.

WARNING

  • The pump is not recommended for pumping acid, base or organic vapors or gases. Serious damage to the pump will shorten the pump’s service life.
  • In addition, pumping flammable vapors or gases can lead to serious safety hazard leading to fire or explosion.

The Care of a Trap

When using a cold trap the refrigerant should be maintained at a high level in the flask to keep the trap at a uniformly low temperature. If the trap is rewarmed it may allow re-evaporation of the condensate. The refrigerant add tube on the liquid nitrogen trap should not be obstructed as the refrigerant boil off can produce dangerously high pressures. If the trap becomes saturated it should be disconnected from the system, drained, and cleaned. An increase in pressure in the vacuum system will normally indicate that the trap has become saturated. To clean the trap, remove the trap from the system and allow the trap to warm up and rinse off the condensate with a suitable solvent in a fume hood. Thoroughly clean and dry the trap before reinstalling into the system.

OPERATION

STARTING PROCEDURES

Starting A Welch Pressure/Vacuum Pump

Before attaching the pump to a system, familiarize yourself with the function and action of the pressure/vacuum pump that you have acquired. Review the power requirements as described in Section 2.60. Welch recommends running the pump for a few minutes to warm it up, before use. The warm-up improves the pumps ability to handle humid air.

Cleanliness

Take every precaution to prevent foreign particulates from entering the pump. Particulates will damage the pump’s performance. If you fi nd that particulates come off the process during evacuation, a particulate trap in the fireline will work. A simple, inexpensive trap may be made by placing glass wool in a glass or plastic tube. Screens must be inserted to hold the glass wool in place.

LEAK DETECTION

The importance of eliminating all leaks in a vacuum system is obvious. The pump must remove this added volume of leaked gas to maintain the desired vacuum. Leaks for these pump can be located by slightly pressuring the system and painting the suspected area with a thick soap solution. Escaping air will produce soap bubbles.

PRESSURE RANGE

Pressure/Vacuum pumps are designed to be run from slightly below atmospheric to their maximum vacuum level on the intake side. The pumps also may be run from atmospheric to their maximum rated pressure rating. Consult the Specifi cation Table (See Section 6, Specifi cations) for the ratings of your specific model.

SHUTDOWN PROCEDURES

After use, Welch recommends the pump be run for about 2 minutes disconnected from the vacuum process. The air pumped through the mechanism will purge out water vapor or droplets of water condensate that may have formed on the inside of the pump. This purge of the pump mechanism helps prevent corrosion.

MAINTENANCE

GENERAL MAINTENANCE

Welch dry pressure/vacuum units are 100% oil free. The pump employs a non-lube piston and cylinder. No maintenance is necessary for the bearings. All bearings are sealed and permanently lubricated. Lubrication should not be attempted. The units are built for continuous duty operation with the quietness and durability of a diaphragm, but with piston performance.

TROUBLESHOOTING

TROUBLESHOOTING TABLE

Poor Pumping Speed| Poor Vacuum| Loud Unit|

Possible Cause

|

Corrective Action

---|---|---|---|---
X| X| X| Damaged valves| Replace flapper valves
X| X| X| Debris in valves| Remove debris and check for valve damage
X| X| X| Damaged gasket| Replace gasket
X| X| X| Loose head screw| Tighten head screw
X| X| X| Loose fitting| Tighten fitting

SPECIFICATION

SPECIFICATION CHART

Welch Model 2522 2534

Free Air Displacement
CFM (L/min.) @60Hz| 0.76 (22)| 1.20 (34)
M3 (L/min.) @50Hz| 0.66 (19)| 1.0 (28)
Maximum Continuous Pressure
PSIG (Pascal)| 100| 50
Maximum Vacuum
In.Hg. (Torr)| 25.6 (110)| 26.2 (950
Motor Horsepower
Horsepower| 1/8| 1/7
Tubing needed
I.D. in Inches| 1/4| 1/4
Weight lbs| 11.7| 11.7
(Kg.)| (5.3)| (5.3)
Catalog Number
Wired for 115V, 60Hz,1Phwith North American 115V Plug| 2522B-01| 2534B-01
Wired for 220V, 50Hz, 1Ph with European (Schuko) Plug| 2522C-02| –
Wired for 230V, 60Hz, 1Ph with Cont. Euro 230V Plug| 2522C-10| –

NOTE: *230V 60Hz Models have a maximum continuous pressure of 50PSIG (5×105 Pascal)
**50Hz models are provided with CE agency mark.

WARRANTY

UNPACKING

  • Inspect the pump carefully. If any damage has occurred, fi le claim with the carrier immediately.
  • Save the shipping container for carrier to inspect.

OPERATING PUMP

  • Refer to the enclosed Instruction/Operation Manual for all information to properly operate and maintain the pump.

WARRANTY

This Welch Vacuum product is warranted to be free from defects in material and workmanship. The liability of Gardner Denver Welch Vacuum Technology, Inc. under this warranty is limited to servicing, adjusting, repairing or replacing any unit or component part which in the judgment of Gardner Denver Welch Vacuum Technology, Inc. has not been misused, abused or altered in any way causing impaired performance or rendering it inoperative. No other warranties are expressed or implied. The method of executing this warranty: servicing, adjusting, repairing or replacing shall be at the discretion of Gardner Denver Welch Vacuum Technology, Inc. Vacuum pumps that have been used for any period, however short, will be repaired under this warranty rather than replaced.

The warranty is effective for one year from the date of original purchase when:

  1. The warranty card has been completed and returned.
  2. The product is returned to the factory or other designated service centers, freight prepaid.
  3. The product in our judgment is defective through no action or fault of the user.

If the product has become defective through misuse, abuse, or alteration, repairs will be billed regardless of the age of the product. In this event, an estimate of the repair costs will be submitted and authorization of these charges will be required before the product is repaired and returned.

To reduce additional charges and delays either within or outside of the warranty period, contact Gardner Denver Welch Vacuum Technology, Inc. @ 847-676-8800 for a return authorization number. Products without a return authorization number will be refused by our receiving department. Before shipping, properly pack the pump, insure it against loss or damage, and on the outside of the pump packaging and the packing slip write in the return authorization number. Pumps damaged due to improper packaging are the customer’s responsibility.

PARTS LIST AND EXPLODED VIEW

Model 2522B-01, 2522C-02

Welch-Ilmvac-2522-Vacuum-Pump-FIG-6

Model 2522B-01 (115V 60Hz)

Item| Qty.| Part No.| Description
9| 1| 726025| Pressure Gauge
11| 1| 726021| Vacuum Gauge
14| 1| 2522K-06| Pressure Regulator Ass’y
15| 1| 2522K-05| Vacuum Regulator Ass’y

Model 2522C-02 (230V 50Hz)

Item| Qty.| Part No.| Description
9| 1| 726026| Pressure Gauge
11| 1| 726024| Vacuum Gauge
14| 1| 2522K-06| Pressure Regulator Ass’y
15| 1| 2522K-05| Vacuum Regulator Ass’y

Item       Qty.    Part No.      Description| Service Kit

2522K-03

| Seal Kit

2522K-04

| Foot Kit

2500K-04

---|---|---|---
1| 1| 2522K-01| Connecting Rod Ass’y| 1| –| –
2| 1| Note 1| Screw, Handle| 2| –| –
3| 1| Note 1| Screw, Head| 4| –| –
4| 1| Note 1| O-ring, Head| 1| 1| –
5| 1| 2522K-02| Valve Plate Ass’y| 1| –| –
6| 1| Note 1| O-ring, Cylinder| 1| 1| –
7A| 1| Note 1| Jar – Plastic| 1| 1| –
7B| 1| Note 1| Jar – Plastic| 1| 1| –
8| 1| Note 1| Filter Element| 1| 1| –
10| 1| Note 1| Ball| 1| 1| –
12| 1| Note 1| Gasket| 2| 2| –
13| 1| Note 1| Rubber Suction Feet| –| –| 4

Note 1: These parts are not available separately, but are supplied in service kits.

Model 2534B-01

Welch-Ilmvac-2522-Vacuum-Pump-FIG-7

Model 2534B-01 (115V 60Hz)

Item| Qty.| Part No.| Description
9| 1| 726025| Pressure Gauge
11| 1| 726021| Vacuum Gauge
14| 1| 2522K-06| Pressure Regulator Ass’y
15| 1| 2522K-05| Vacuum Regulator Ass’y

Item    Qty.      Part No.      Description| Service Kit

2534K-03

| Seal Kit

2522K-04

| Foot Kit

2500K-04

---|---|---|---
1| 1| 2534K-01| Connecting Rod Ass’y| 1| –| –
2| 1| Note 1| Screw, Handle| 2| –| –
3| 1| Note 1| Screw, Head| 4| –| –
4| 1| Note 1| O-ring, Head| 1| 1| –
5| 1| 2565K-02| Valve Plate Ass’y| 1| –| –
6| 1| Note 1| O-ring, Cylinder| 1| 1| –
7A| 1| Note 1| Jar – Plastic| 1| 1| –
7B| 1| Note 1| Jar – Plastic| 1| 1| –
8| 1| Note 1| Filter Element| 1| 1| –
9| 1| Note 1| Ball| 1| 1| –
12| 1| Note 1| Gasket| 2| 2| –
13| 1| Note 1| Rubber Suction Feet| –| –| 4
16| 1| Note 1| Replacement Filters| 3| 3| –
17| 1| Note 1| O-ring, Cylinder| 1| 1| –

Note 1: These parts are not available separately, but are supplied in service kits.

Contact

WELCH-ILMVAC

This document contains confidential and trade secret information which is the property of Welch Vacuum Technology and receipt or possession does not convey any rights to loan, sell, or otherwise disclose said information, reproduction or use of said information for any purpose other than that in connection with said information was supplied may not be made without express written permission of Welch. This document is to be returned to Welch upon request and in all events upon completion of the purpose for which it is loaned.

Copyright© 2008 Gardner Denver Thomas, Inc.
Welch Vacuum Technology
Welch is a registered trademark of Gardner Denver Thomas, Inc.

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals