SARTORIUS Picus 2 Electronic Pipettes Installation Guide

June 16, 2024
SARTORIUS

SARTORIUSPicus 2 Electronic Pipettes

Specifications

  • Pipette Type: Sartorius Pipettes
  • Cleaning Method: Cleaning and Decontamination
  • Cleaning Frequency: Daily for outer surface, every three months for lower part
  • Cleaning Reagents: 70% (v/v) Ethyl alcohol

Product Usage Instructions

Cleaning and Decontamination in Pipette Maintenance
Pipettes are precision instruments that require regular maintenance to ensure optimal performance. Adequate cleaning and decontamination are necessary to maintain the purity required for laboratory applications.

  • Cleaning refers to the removal of any foreign materials such as dust, nucleic acids, proteins, buffers, and salts from the pipette.
  • Decontamination refers to the removal or killing of microorganisms like viruses, bacteria, and fungi. It also includes the removal or neutralization of chemicals and radioactive materials.

Cleaning Guide

Cleaning the Outer Surface of the Pipette (Daily)

  1. Prepare a disinfectant liquid or mild detergent, such as 70% ethanol, and a soft, lint-free cloth.
  2. Gently clean the outer surface of the pipette, including the tip cone, using the moistened cloth.
  3. Wipe the surface dry with a clean cloth.
  4. Regularly change the Safe-Cone Filter using the provided tweezers.
  5. Ensure that no liquids enter the inner parts of the pipette.

Daily Cleaning Quick Guide

Pipette Cleaning Reagent
Upper part 70% (v/v) Ethyl alcohol
Lower part 70% (v/v) Ethyl alcohol

Cleaning the Lower Part of the Pipette (Every Three  Months)

Note: Always ensure that the pipette materials are chemically compatible with the detergent, disinfectant, or decontaminant liquids you plan to use. Refer to the pipette user manual for information on pipette materials and chemical compatibility.
Note: Using tip cone filters may lengthen the service interval. Change filters regularly.
Note: It is recommended to always use gloves when cleaning the pipette.

Multichannel Pipettes
Cleaning and maintaining the lower parts of all models: Disassembling of the lower part of a multichannel pipette should only be done by an authorized Sartorius service provider. Please contact your nearest Sartorius service provider or distributor.

FAQ

  • Q: How often should I clean the outer surface of the pipette?
    A: The outer surface of the pipette should be cleaned daily using a disinfectant liquid or mild detergent.

  • Q: What cleaning reagent should I use for the upper and lower parts of the pipette?
    A: For both the upper and lower parts, use 70% (v/v) Ethyl alcohol as the cleaning reagent.

  • Q: Can I disassemble the lower part of a multichannel pipette for cleaning?
    A: No, the disassembling of the lower part of a multichannel pipette should only be done by an authorized Sartorius service provider.

Cleaning and Decontamination in Pipette Maintenance

Pipettes are precision instruments whose performance can be significantly impacted if not adequately maintained. To maintain the level of purity that is critical in many laboratory applications, adequate cleaning and decontami- nation of pipettes is necessary.

  • Cleaning refers to removal of any foreign materials from the pipette such as dust, nucleic acids, proteins, buffers and salts.
  • Decontamination refers to the removal or killing of microorganisms such as virus, bacteria and fungi. It also refers to the removal or neutralization of chemicals and radioactive materials.

Sartorius pipettes are designed with the end-user applications in mind and therefore are designed to be very easy to clean, have as many autoclavable products as possible, and to tolerate a variety of cleaning reagents. When selecting solutions for cleaning or decontamination of the pipette, consider the reagent or liquid that has been pipetted. Additionally, check that the cleaning solution is compatible with the pipette materials and would not have any adverse effect. Information about pipette materials are available in the pipette user manuals which can be downloaded from www.sartorius.com or hardcopies ordered by email from lhinfo.finland@sartorius.com.

Cleaning Guide

Cleaning the Outer Surface of the Pipette (Daily)

  • To clean and decontaminate the outer surface of the pipette, use a disinfectant liquid or mild detergent, for example 70% ethanol, and a soft, lint-free cloth. Gently clean the surface of the pipette with the moistened cloth and wipe it dry. Pay special attention to the tip cone.
  • Always check the chemical compatibility between the pipette materials and the disinfectant or decontaminant liquid.
  • Change the Safe-Cone Filter regularly with the tweezers provided with the pipette.
  • Do not let liquids enter the inner parts of the pipette.

Daily Cleaning Quick GuideSARTORIUS-Picus-2-Electronic-Pipettes-
\(1\) Cleaning the Lower Part of the Pipette (Every Three Months)

The frequency of pipette cleaning and maintenance depends on how heavily the pipette is used and which reagents it is used for – whether corrosive or biohazardous. For a pipette that is used daily, we recommend you clean and decontaminate it and check its performance every three months, however cleaning and decontamination should always be done after over-aspiration or when the pipette has clearly been contaminated. Check the cleaning | decontamination protocol in the pipette user manual and chemical compatibility of the selected cleaning solution for your pipette. When opening the pipette lower parts, as a matter of routine the parts should be inspected for wear and tear. Always check the pipette performance after opening the lower parts. Pipette calibration is a fundamental part of Good Laboratory Practice (GLP) and we recommend that this should be done at minimum once a year. Sartorius also provides a complete repair and calibration service, including a service report and performance certificate.

Disinfectant Liquid
Always ensure the pipette materials are chemically compatible with the detergent, disinfectant, or decontaminant liquids you plan to use. In the pipette user manual, see the section on pipette parts and materials for information on pipette materials. For information on chemical compatibility, please send a request by email to lhinfo.finland@sartorius.com.

NOTE
Using tip cone filters may lengthen the service interval. Change filters regularly.

NOTE
We recommend always using gloves when cleaning the pipette.

Multichannel Pipettes
Cleaning and maintaining the lower parts of all models: Disassembling of the lower part of a multichannel pipette should only be done by an authorized Sartorius service provider. Please contact your nearest Sartorius service provider or distributor.SARTORIUS-Picus-2-Electronic-Pipettes-
\(2\) Pipette Parts (For All Pipettes Except for Proline®)

  1. Tip ejector collar
  2. Locking ring
  3. Tip cone
  4. Piston
  5. Piston seal

To Disassemble and Clean the Pipette (Except for Proline®):

  1. Eject the tip cone filter (if fitted)

  2. Unscrew the tip ejector collar (1) counterclockwise and remove it.

  3. Depending on the pipette volume:

    • With all pipettes other than the 5000 μl or 10 ml models, unscrew the locking ring (2) counterclockwise and carefully remove it together with the tip cone (3).
      or

    • With a 5000 μl pipette, hold the locking ring (2) firmly with your fingers and turn the tip cone cylinder counterclockwise with the other hand (see image below). Do not turn the locking ring as you turn the tip cone (3) as this will break the pipette. or

    • With a 10 ml pipette hold the tip cone cylinder firmly and turn the locking ring (2) clockwise with your fingers (see image below). Do not turn the tip cone (3) as this will break the pipette.

  4. Clean the tip ejector collar (1), the tip cone holder, the tip cone cylinder, and the piston (4) with a disinfectant liquid or mild detergent and a soft, lint-free cloth.

  5. Clean the interior of the tip ejector collar (1) and the tip cone cylinder with a cotton swab. Be careful with 3, 10, 20, and 100 μl pipettes to ensure the seal inside the tip cone is not damaged.

  6. Rinse the parts with pure (Type 3) water or distilled water if necessary and allow them to dry.

  7. Depending on the pipette volume:

    • With a 3, 10, 20, or 100 μl pipette, apply a thin layer of grease on the piston (4).
      or

    • With a 200 or 1000 μl pipette, apply a thin layer of grease around the seal (5).

    • With a 5000 μl or10 ml pipette, apply a thin layer of grease on the interior of the tip cone cylinder and around the seal (5)

NOTE
Avoid excess grease. Only use the grease provided with the pipette.

NOTE
Before reassembling, check that there is no lint or particles on the surface of the piston.

To Reassemble the Pipette

  1. Depending on the pipette volume:

    • With a 3, 10, 20, 100, 200, or 1000 μl pipette, carefully place the tip cone (3) on the piston (4) and attach it by screwing the tip cone holder clockwise.
      or

    • With a 5000 μl pipette, carefully place the tip cone cylinder on the pis-ton (4) and screw clockwise. Make sure the tip cone cylinder is properly tightened. Avoid over-tightening.

    • With a 10 ml pipette, carefully place the tip cone cylinder on the pis-ton (4) and screw the locking ring (2) counterclockwise. Make sure the tip cone cylinder is properly tightened. Avoid over-tightening.

  2. Attach the tip ejector collar (1) by screwing it clockwise.

  3. Insert a new tip cone filter.

  4. Press the operating button several times to ensure that the grease has spread evenly.

  5. Check the performance of the pipette.

NOTE
Always check the performance of the pipette after in-house service or maintenance.

Picus® and Picus® Nxt Parts:
Lower parts of Picus® or Picus® NxT single channel pipette:

  • Tip cone, tip cone holder, tip ejector collar, piston and spring

Tacta® Parts:
Lower parts of a Tacta® single channel pipette:

  • Tip ejector collar, tip cone holder, tip cone, piston

To Disassemble and Clean the Proline® Pipette

  1. Hold down the tip ejector.
  2. Place the tooth of the opening tool between the tip ejector and the tip ejector collar to release the locking mechanism.
  3. Carefully release the tip ejector and remove the tip ejector collar.
  4. Place the wrench end of the opening tool over the tip cone, turning it anticlockwise. Do not use any other tools. The 5 ml tip cone is removed by turning it anticlockwise without any tools.
  5. Remove the tip cone, piston, and spring. Remove the filter if fitted.
  6. Place the tip cone, tip ejector, tip ejector collar, piston, O-ring, and spring into a beaker containing Proline Biocontrol and leave for at least 30 minutes to secure complete decontamination.
  7. Remove the components from the beaker and rinse them with distilled water, then dry, preferably with warm air, for at least one hour.
  8. Before replacing the tip cone it is recommended to grease the piston and O-ring slightly by using the silicone grease provided.
  9. After reassembling press the operating button several times to ensure that the grease has spread evenly. 10. Check the pipette calibration.

Proline® Pipette Parts

  1. Operating button
  2. Tip ejector collar
  3. Tip ejector

NOTE
On models 720005, 720000, 722001 and 722004 (≤ 10 µl pipettes) the O-ring cannot be accessed for maintenance.

NOTE
Excessive use of grease may jam the piston.

Application-Specific Cleaning and Decontamination of Pipettes

Selection of disinfectant or decontaminant liquid: Always ensure the pipette materials are chemically compatible with the detergent, disinfectant, or decontaminant liquids you plan to use. Check the pipette user manual for information on pipette materials. For information on chemical compatibility, please send a request by email to lhinfo.finland@sartorius.com.

When Cleaning Sartorius Pipettes for Specific Applications

  1. The outside surface of the upper part (body) of pipettes should be cleaned by wiping with cleaning reagent or decontaminant solution using a soft, lint-free cloth, based on chemical compatibility with the pipette. The upper part (body) of Sartorius pipettes should never be immersed (or soaked) in the cleaning agent.
  2. Use the same procedure described for cleaning and maintaining the lower parts of all models as described in Section 2. Cleaning and decontamination of the lower part of the pipette can be done by wiping or immersion based on required cleaning or decontamination procedure for the contaminant. Lower parts of the pipette may be soaked in cleaning reagent or decontaminant solution, based on chemical compatibility with the pipette.
  3. Always remove the cleaning agent thoroughly with pure (Type 3) water or distilled water after decontamination is completed. Allow parts to dry before greasing and reassembly.
  4. Re-greasing and checking of pipette performance should be done after opening the lower parts.

The upper and lower parts of the pipette are indicated in each pipette user manual. Additionally, disassembly procedures of single channel and multi- channel pipettes are clearly explained in each pipette user manual.

Multichannel Pipettes
Cleaning and maintaining the lower parts of all models: Opening the lower part of a multichannel pipette should only be done by an authorized Sartorius service provider. Please contact your nearest Sartorius service provider or distributor.

Cleaning and Decontamination Reagent Quick-Guide

Reagent Proline® Proline® Plus mLine® Tacta® Picus® Picus® NxT
Detergent
Radioactivity cleaning reagent
Nucleic Acid Nuclease removal agent
DEPC (Diethyl pyrocarbonate)
Ethanol or Isopropanol
Sodium hypochlorite
Hydrogen peroxide
Vapor Hydrogen Peroxide
Quaternary ammonium disinfectant
UV (ultraviolet radiation)

NOTE
The display on all Sartorius pipettes and the soft keys on Picus® and Picus® NxT electronic pipettes are chemically resistant to the cleaning agents list- ed above.

Cleaning Agents and Their Activities

Cleaning reagent| Radioactive solutions| Acids & Alkalis| Organic solvents| DNA | RNA| RNase | DNase| Proteins
---|---|---|---|---|---|---
Pure (Type 3) water or Distilled water| | √| | | |
Detergent| √| | √| | | √
Radioactivity cleaning reagent| √| | | | |
Sodium hypochlorite| | | | √| |
Nucleic Acid removal reagent| | | | √| |
Nuclease removal reagent| | | | | √|
DEPC (Diethyl pyrocarbonate)| | | | | √|
UV (ultraviolet radiation)| | | | √| |

Cleaning Procedures

  • Examples : Examples of cleaning products are provided here only as an aid. Sartorius does not endorse any particular brand of product. Examples of commonly used cleaning reagents: Detergent:

  • Decon® or Deconex®, Radioactivity cleaning reagent : Decon 90®, Decon NoCount®, Nucleic Acid removal reagent: DNA Away®, RNA Away®, Nuclease removal reagent: RNaseZAP®, RNase | DNase Away® Detergents, Radioactivity cleaning reagents, Nucleic Acid removal re-agents and Nuclease removal reagents: Follow the cleaning product manufacturer’s instructions for use. After cleaning with detergents or radioactivity cleaning reagents, remove the cleaning agent from the pipette thoroughly with pure (Type 3) water or distilled water.
    Sodium hypochlorite: Wipe the outside of the entire pipette or soak lower parts in 0.5 – 3% (w/v) sodium hypochlorite for 15 minutes. Remove the cleaning agent from the pipette thoroughly with pure (Type 3) water or dis- tilled water.

  • DEPC (Diethyl pyrocarbonate): Soak the lower autoclavable parts of the pi-pette in 0.1% DEPC for 2 hours. Rinse thoroughly with sterile DEPC-treated water, then autoclave the DEPC-treated pipette parts to deactivate DEPC. Use DEPC decontamination only with autoclavable parts of the pipette since DEPC is toxic and must be deactivated by autoclaving after use.

  • Ethanol | Isopropanol : Wipe the outside of the entire pipette or soak lower parts in 70% (v/v) Ethanol or 60% (v/v) Isopropanol for 30 minutes.

  • Hydrogen peroxide: Wipe the outside of the entire pipette or soak lower parts in 3% Hydrogen peroxide for 10 minutes. Remove the cleaning agent from the pipette thoroughly with pure (Type 3) water or distilled water.

  • UV radiation : Expose pipette for 30 – 60 minutes to UV. The degree of inactivation by ultraviolet radiation is dependent on the UV dose applied. Sartorius pipettes have been tested and shown to have durability and no changes in performance after temporary exposure. Take note that prolonged or frequent exposure to UV radiation may cause yellowing and bristling of the pipette. We do not recommend the use of UV as an effective way to remove contamination on pipettes.

Decontamination Agents and Their Activities

Decontamination reagent| Bacteria spores| Vegetative Bacteria| Virus Enveloped| Virus
Non-enveloped| Fungi
---|---|---|---|---|---
Ethanol | Isopropanol| | √| √| |
Sodium hypochlorite| (poor)| √| √| √| √
Hydrogen peroxide| √| √| √| √| √
Vapor Hydrogen Peroxide| √| √| √| √| √
Quaternary ammonium disinfectant| | √| √| |
UV (ultraviolet radiation)| (poor)| √| √| √| √

Decontamination Procedures

  • Examples : Examples of decontamination products are provided here only as an aid. Sartorius does not endorse any particular brand of product. Examples: Quaternary ammonium disinfectant: Aniospray, Phagoclean.

  • Ethanol | Isopropanol: Disinfection – Wipe the outside of the entire pipette or soak lower parts in 70% (v/v) Ethanol or 60% (v/v) Isopropanol for
    30 minutes.

  • Sodium hypochlorite: Disinfection – Wipe the outside of the entire pipette or soak lower parts in 0.5 – 3% (w/v) sodium hypochlorite for 15 minutes. Remove the cleaning agent from the pipette thoroughly with pure (Type 3) water or distilled water.

  • Hydrogen peroxide : Disinfection – Wipe the outside of the entire pipette or soak lower parts in 3% Hydrogen peroxide for 10 minutes. Sterilization – soak lower parts in 7.5% Hydrogen peroxide for 10 minutes. Remove the cleaning agent from the pipette thoroughly with pure (Type 3) water or distilled water.

  • Vapor Hydrogen peroxide (VHP): Disinfection and sterilization: Follow the VHP provider’s instructions for use. A minimum treatment of 0.1 – 3 mg/L vapor hydrogen peroxide at 25°C for 2 minutes has been shown to have broad-spectrum efficacy.

  • Quaternary ammonium disinfectants: Follow the disinfectant manufacturer’s instructions for use. After disinfection, remove the cleaning agent from the pipette thoroughly with pure (Type 3) water or distilled water.

  • UV radiation : Expose pipette for 30 – 60 minutes to UV. The degree of inactivation by ultraviolet radiation is dependent on the UV dose applied. Sartorius pipettes have been tested and shown to have durability and no changes in performance after temporary exposure. Take note that prolonged or frequent exposure to UV radiation may cause yellowing and bristling of the pipette. We do not recommend the use of UV as an effective way to remove contamination on pipettes.

Autoclaving Guide

Mechanical Pipettes
All Sartorius mechanical pipettes except Proline® are fully autoclavable. The entire mechanical pipette can be sterilized by steam, autoclaving at 121°C, (252°F), 1 atm for 20 minutes.

Autoclaving Instructions

  1. Remove the Safe-Cone Filter if attached.
  2. The single channel pipettes can be autoclaved without special preparations. For multichannel pipettes, unscrew the lower part of the multichannel pipettes by holding the connecting collar and turning the lower part 360° counterclockwise.
  3. Put the pipette into the sterilization bag, and place the bag into the autoclave.
  4. Sterilize at 121°C and 1 bar overpressure for 20 minutes.
  5. After autoclaving, the pipette must be cooled down and dried before reassembly. Drying can be done overnight at room temperature or at 60°C for 3 hours then cooled down.
  6. To reassemble the multichannel pipette, reattach the lower part to the pipette by holding the connecting collar and turning the lower part 360°clockwise. With a single-channel pipette, check if the lower parts have become loose during autoclaving, and tighten them if necessary.
    It is recommended to check the performance of the pipette after every autoclaving. It is also recommended to grease the piston/seal of the pipette after every 10 autoclaving.

Electronic Pipettes
The lower parts of Sartorius electronic pipettes are autoclavable, excluding the 1200 μl multichannel models. Please see the autoclaving symbol printed on the lower part of your multichannel pipette to ensure the section is autoclavable.

Autoclaving Instructions

  1. Remove the Safe-Cone Filter if attached.
  2. Disassemble the lower part: For single channel electronic pipettes, unscrew the tip ejector collar, tip cone and piston as instructed in user manual by turning them counter-clockwise and place these parts in a sterilization bag. For multichannel electronic pipettes, unscrew the connecting collar counterclockwise to remove the tip cone housing, and place it in a sterilization bag.
  3. Sterilize at 121°C and 1 bar overpressure for 20 minutes.
  4. After autoclaving, the autoclaved pipette parts must be dried and cooled down before reassembly. Drying can be done overnight at room temperature or at 60°C for 3 hours then cooled down.
    It is recommended to check the performance of the pipette after every auto claving. It is also recommended to grease the piston | seal of the single channel electronic pipette after every 10 autoclaving. We recommend sending multi channel pipettes to your local Sartorius service center for greasing after every 10 autoclaving.

Autoclaving Quick-Guide

Pipette Model Autoclavability Notes
Proline® Plus, Single channels Fully autoclavable No need to dis-
mLine®, Tacta® assemble before
autoclaving
Proline® Plus, Multi-channels Fully autoclavable Unscrew lower
mLine®, Tacta® part before
autoclaving
Picus®, Single channels Only lower parts Autoclave tip
Picus® NxT ejector collar, tip
cone and piston
Picus®, Multi-channels Only lower parts Autoclave the tip
Picus® NxT cone housing
Picus®, Multi-channel Not autoclavable Not autoclavable
Picus® NxT 1200 µl

Drying of Pipettes
Drying of Sartorius pipettes can be done overnight at room temperature or at 60°C for 3 hours. Pipettes must always be cooled down after drying in the oven before reassembly. The upper parts (body) of electronic pipettes should not be dried in a drying oven.

Summary of Pipette Parts

Mechanical Pipettes Parts And Materials

Tacta® Mechanical Pipette
Pipette Parts and Materials

  1. Operating button (polyamide (PA), silicone (SI), stainless steel (SS))
  2. Finger support (polypropylene (PP))
  3. Tip ejector (PA)
  4. Handle (PP)
  5. Volume lock (SI)
  6. Display (polycarbonate (PC))
  7. Tip ejector collar (PP)
  8. Tip cone (polyvinylidene fluoride (PVDF) in 3, 10, 20, and 100 μl pipettes, polyetherimide (PEI) in 200 and 300 μl pipettes, polyphenylene sulphide (PPS) in 1000 μl, 5000 μl, and 10 ml pipettes)
  9. Safe-Cone Filter (polyethylene (PE))
  10. Adjustment window (PC)
  11. Tip cone housing (PA)
  12. Tip ejector bar (PA)
  13. Stainless steel, corrosion-resistant metal ring (SSt EN 1.4404 | AISI 316L)

SARTORIUS-Picus-2-Electronic-Pipettes- \(14\) mLine® Mechanical Pipette

Proline® Plus Mechanical Pipette

Proline® Mechanical Pipette

Electronic Pipette Parts and Materials: Picus and Picus® NxT

Contact

  • Sartorius Biohit Liquid Handling Oy
  • Laippatie 1
  • 00880 Helsinki
  • Finland
  • Phone +358.9.755.951
  • lhinfo.finland@sartorius.com
  • Sartorius Lab Instruments GmbH & Co. KG
  • Otto-Brenner-Strasse 20
  • 37079 Göttingen
  • Germany
  • Phone +49.551.308.0
  • www.sartorius.com

Further Information About Pipette Cleaning, Decontamination and Maintenance

Kindly email to lhinfo.finland@sartorius.com

Pipette Maintenance, Servicing or Calibration

Visit https://www.sartorius.com/sartorius/en/EUR/applications/laboratory /liquid-handling/liquid-handling-services

SARTORIUS-Picus-2-Electronic-Pipettes- \(19\)Sartorius Lab Instruments GmbH & Co. KG Otto-Brenner-Strasse 20

The information and figures contained in these instructions correspond to the version date specified below.
Sartorius reserves the right to make changes to the technology, features, specifications and design of the equipment without notice. Masculine or feminine forms are used to facilitate legibility in these instructions and always simultaneously denote all genders.

Copyright notice

  • These instructions, including all components, are protected by copyright.
  • Any use beyond the limits of the copyright law is not permitted without our approval.
  • This applies in particular to reprinting, translation and editing irrespective of the type of media used.

Last updated
07 | 2023

© 2023
Sartorius Lab Instruments GmbH & Co. KG
Otto-Brenner-Str. 20
37079 Goettingen, Germany
Cleaning-and-Decontamination-Guide-For-Pipettes-Manual-en-Sartorius