Solmetex DS-PSK-002 DryShield Pediatric Starter Kit User Guide
- June 16, 2024
- Solmetex
Table of Contents
Quick Start Guide
Isolation. Protection.
Workflow. Simplified.
INSTALLATION
Dedicated
Replace the existing HVE with the DS1, attaching it to the tubing. Use the Y
Connector if there is only one HVE.
|
---|---
Twist off existing HVE (with attached coupler) from tubing.| Attach DS1 to
existing tubing by twisting into place.
Portable
Insert the DS1 into the end of the Portable Operatory Kit, and plug the kit
into your HVE tip.
|
---|---
Attach one end of portable kit to HVE.| Attach DS1 to other end of portable
kit.
PATIENT INTRODUCTION
Please prep the patient by telling them:
- DryShield was designed for safety and comfort.
- This shield will protect your tongue and cheek.
- Rest your jaw on the bite block.
- Relax, breathe through your nose, and we’ll be done in no time.
PRE-INSERTION
|
---|---
Spray air and water inside the mouthpiece to lubricate.| Slide mouthpiece onto
the DS1, using a rocking motion until it locks in place.
|
Fold tail in towards the center of the mouthpiece.| Then fold outside edges of
the mouthpiece towards the center of the mouthpiece.
MOUTHPIECE INSERTION
|
---|---
Slowly insert, lining up the bite block’s anterior edge to distal of the
canine.
(For kids, midline of the primary lateral.)| Release the mouthpiece and let
fall, using finger pressure to adapt it to the inner cheek,
retromolar pads, palate and floor of the mouth.
|
Position the bite block distal to the maxillary canine, then allow patient to
rest their jaw on it.| Make sure the mouthpiece is fully adapted to the mouth,
and reposition lips as needed. Slide the
bite block distally if you need more vertical opening.
Sterilization Guide
|
---|---
Detach bite block from mouthpiece.| Scrub both with small brush (e.g.
toothbrush) using soap and water, including inside the neck and slit of
mouthpiece. Invert under water and check for removal of all debris.
|
Place mouthpiece and bite block in ultrasonic solution for 15 minutes.| Rinse
off ultrasonic solution under running water and lay onto cloth to dry
|
Place bite block and mouthpiece into autoclave bag and seal.| Steam autoclave
for min 121 C (250 F) for at least 30 min, followed by a 30-min drying cycle.
As a rule of thumb, when the mouthpiece begins to take on a perceptible tint, it is an indicator that it may be time for disposal and replenishment
WWW.DRYSHIELD.COM
877-207-1551WWW.SOLMETEX.COM