CA-MI NEW ASKIR 230/12V BR Aspirator Instruction Manual
- June 16, 2024
- CA MI
Table of Contents
CA-MI NEW ASKIR 230/12V BR Aspirator
Product Information
Specifications:
- Electrical features: 14V 4A
- AC/DC adapter models: UE60-140429SPA1 or UES65-140429SPA1 of FUHUA (input: 100-240V~ – 50/60Hz – 100VA)
- Internally powered equipment: Pb Battery 12V 4A
- Cigarette lighter adapter: 12V 4A
- Type: Surgical aspirator
- Usage: Desk-type electric suction unit for the aspiration of body liquids, oral, nasal, and tracheal aspiration in adults or children
- Power management system: Operated using an electronic power management system
- Battery operation and status monitoring: Equipped with an internal battery operation and status monitoring system
- Alarm: Visual and acoustic alarm indicating low battery level
- Vacuum meter: Equipped with a vacuum meter (bar & kPa)
- Vacuum regulator: Vacuum regulator located on the front panel
Product Usage Instructions
General Warning:
Read the instruction manual carefully before use. This device is for use by
qualified personnel (surgeon/professional nurse/assistant). The use of the
device at home is restricted to an adult in full possession of mental
faculties and/or home carers. The instrument must not be disassembled. For
technical services, always contact CA-MI SRL.
Important Safety Rules:
- Check the condition of the unit before each use. Carefully inspect the surface of the unit for visual damage. Check the mains cable and do not connect to power if damage is apparent.
- Before connecting the appliance, always check that the electric data indicated on the data label and the type of plug used correspond to those of the mains electricity to which it is to be connected.
- For repairs, exclusively contact technical service and request the use of original spare parts. Failure to comply with the above can jeopardize the safety of the device.
- Use only for the purpose intended. Do not use for anything other than the use defined by the manufacturer. The manufacturer will not be responsible for damage due to improper use or connection to an electrical system not complying with current regulations.
- The medical device requires special precautions regarding electromagnetic compatibility and must be installed and used by the information provided in the accompanying documents. The NEW ASKIR 230/12V BR device must be installed and used away from mobile and portable RF communication devices (mobile phones, transceivers, etc.) that may interfere with the said device.
- Instrument and accessory discharging must be done according to current regulations in the country of use.
- WARNING: Do not change this equipment without the permission of the manufacturer CA-MI Srl. None of the electric or mechanical parts have been designed to be repaired by customers or end-users. Do not open the device, and do not mishandle the electric/mechanical parts. Always contact technical assistance.
- Using the device in environmental conditions different than those indicated in this manual may seriously harm the safety and technical characteristics of the device.
- Do not perform repairs or maintenance operations on the device when it is being used on a patient.
FAQ:
- Q: Can I use this device at home?
- A: The use of the device at home is restricted to an adult in full possession of mental faculties and/or home carers.
- Q: Can I repair the device myself?
- A: No, none of the electric or mechanical parts have been designed to be repaired by customers or end-users. Always contact technical assistance.
- Q: Can I use the device near mobile phones or other RF communication devices?
- A: No, the device must be installed and used away from mobile and portable RF communication devices that may interfere with the device.
NEW ASKIR 230/12V BR is a surgical aspirator with the following electrical features: 14V 4A with AC/DC adapter mod. UE60-140429SPA1 or UES65-140429SPA1 of FUHUA (input: 100-240V~ – 50/60Hz – 100VA) or Internally powered equipment(Pb Battery 12V 4A) or with a cigarette lighter adapter (12V 4A). NEW ASKIR 230/12V BR is a desk-type electric suction unit for the aspiration of body liquids, oral, nasal, and tracheal aspiration in adults or children. The appliance is designed for easy transport and almost continuous use. The unit is operated using an electronic power management system, also monitoring the internal battery operation and status. Provided with visual and acoustic alarms indicating low battery levels, the main unit is also equipped with a vacuum meter (bar & kPa) and a vacuum regulator on the front panel.
Thanks to these characteristics and its functions, this device is particularly suitable for different applications: utilization in hospital wards, for tracheotomy, for suction of body liquids, and for minor surgery. A foot switch for switching the device on remotely is available on request.
GENERAL WARNING
READ THE INSTRUCTION MANUAL CAREFULLY BEFORE USE. THE DEVICE IS FOR USE BY QUALIFIED PERSONNEL (SURGEON / PROFESSIONAL NURSE / ASSISTANT)THE USE OF THE DEVICE AT HOME IS RESTRICTED TO AN ADULT IN FULL POSSESSION OF MENTAL FACULTIES AND / OR HOME CARERS THE INSTRUMENT MUST NOT DISASSEMBLED. FOR TECHNICAL SERVICES ALWAYS CONTACT CA-MI SRL
IMPORTANT SAFETY RULES
- Check the condition of the unit before each use. The surface of the unit should be carefully inspected for visual damage. Check the mains cable and do not connect to power if damage is apparent;
- Before connecting the appliance always check that the electric data indicated on the data label and the type of plug used, correspond to those of the mains electricity to which it’s to be connected;
- Respect the safety regulations indicated for electrical appliances particularly:
- Use original components and accessories provided by the manufacturer to guarantee the highest efficiency and safety of the device;
- The device can be used only with the bacteriological filter;
- Never immerse the appliance in the water;
- Do not place or store the aspirator in places where it may fall or be pulled into the bathtub or washbasin. In the event it is accidentally dropped, do not attempt to remove the device from the water whilst the plug is still connected: disconnect the mains switch, remove the plug from the power supply, and contact the CA-MI technical service department. Do not attempt to make the device work before it has been thoroughly checked by qualified personnel and/or the CA-MI technical service department.
- Position the device on stable and flat surfaces in a way that the air inlets on the back aren’t obstructed;
- To avoid incidents, do not place the aspirator on unstable surfaces, which may cause it to accidentally fall and lead to a malfunction and/or breakage. Should there be signs of damage to the plastic parts, which may expose the inner parts of the energized device, do not connect the plug to the electrical socket. Do not attempt to make the device work before it has been thoroughly checked by qualified personnel and/or the CA-MI technical service department.
- Don’t use in the presence of inflammable substances such as anesthetic, oxygen, or nitrous oxide;
- Don’t touch the device with wet hands and always prevent the appliance from coming into contact with liquids;
- Don’t leave the appliance connected to the power supply socket when not in use;
- Don’t pull the power supply cable to disconnect the plug remove the plug from the mains socket correctly;
- Store and use the device in places protected against the weather and far from any sources of heat. After each use, it is recommended to store the device in its box away from dust and sunlight.
- In general, it is inadvisable to use single or multiple adapters and/or extensions. Should their use be necessary, you must use ones that comply with safety regulations, however, taking care not to exceed the maximum power supply tolerated, which is indicated on the adapters and extensions.
- For repairs, exclusively contact technical service and request the use of original spare parts. Failure to comply with the above can jeopardize the safety of the device;
- Use only for the purpose intended. Don’t use for anything other than the use defined by the manufacturer. The manufacturer will not be responsible for damage due to improper use or connection to an electrical system not complying with current regulations.
- The medical device requires special precautions regarding electromagnetic compatibility and must be installed and used according to the information provided with the accompanying documents: the NEW ASKIR 230/12V BR device must be installed and used away from mobile and portable RF communication devices (mobile phones, transceivers, etc.) that may interference with the said device.
- Instrument and accessory discharging must be done according to current regulations in the country of use.
- WARNING: Do not change this equipment without the permission of the manufacturer CA-MI Srl. None of the electric or mechanical parts have been designed to be repaired by customers or end-users. Don’t open the device, do not mishandle the electric/mechanical parts. Always contact technical assistance
- Using the device in environmental conditions different than those indicated in this manual may harm seriously the safety and the technical characteristics of the same.
- Do not perform repairs or maintenance operations on the device when it is being used on a patient
- The medical device is in contact with the patient using a disposable probe (not supplied with the device). Suction tubes for insertion in the human body purchased separately from the machine should comply with ISO 10993-1 standards on material biocompatibility.
- The product and its parts are biocompatible by the requirements of regulation EN 60601-1.
- Operation of the device is very simple and therefore no further explanations are required other than those indicated in the following user manual.
- The lead battery integrated into the device is not to be considered ordinary domestic waste. Such a component must be disposed of in a specific collection center to be recycled.
- Use in Home-Care: Keep all accessories of the device out of reach of children under 36 months of age since they contain small parts that may be swallowed.
- Do not leave the device unattended in places accessible to children and/or persons not in full possession of mental faculties as they may strangle themselves with the patient’s tube and/or the power cable.
The manufacturer cannot be held liable for accidental or indirect damages should the device be modified, or repaired without authorization or should any of its components be damaged due to accident or misuse. Any minimal modification/repair on the device voids the warranty and does not guarantee compliance with the technical requirements provided by the MDD 93/42/EEC (and subsequent changes) and its normative.
CONTRAINDICATIONS
- Before using the NEW ASKIR 230/12V BR, consult the instructions for use: failure to read all the instructions in this manual can be harmful to the patient.
- Don’t use the device for thoracic drainage.
- The device must not be used for suction of explosive, corrosive, or easily flammable liquids.
- NEW ASKIR 230/12V BR is not suitable for MRI. Do not introduce the device in MRI environments.
DISPOSAL OF WASTE BATTERIES – (Directive 2006/66/EC)
This symbol on the battery or the packaging indicates that the battery
provided with this product shall not be treated as household waste. By
ensuring these batteries are disposed of correctly, you will help prevent
potentially negative consequences for the environment and human health which
could otherwise be caused by inappropriate waste handling of the battery. The
recycling of the materials will help to conserve natural resources.
At the end of their life hand the batteries over to the applicable collection points for the recycling of waste batteries. For more detailed information about recycling this product or battery, please contact your local Civic Office, your household waste disposal service, or the shop where you purchased the product.
TECHNICAL CHARACTERISTICS
Model | NEW ASKIR 230/12V BR (REF RE 310211) |
---|---|
Typology (MDD 93/42/EEC) | Class IIa Medical device |
UNI EN ISO 10079-1 Classification | HIGH VACUUM / LOW FLOW |
Power Feeding | 14V 4A with AC/DC adapter mod. UE60-140429SPA1 or |
UES65-140429SPA1 of FUHUA (input: 100-240V ~ –
50/60Hz – 100VA) or Internally powered equipment (Pb Battery 12V 4A) or with cigarette lighter adapter (12V
4A)
Maximum Suction Pressure (without jar)| -75kPa (- 0.75 bar)
Minimum Suction Pressure (without jar)| Less -25kPa (-0.25 bar)
Maximum Suction Flow (without jar)| 16 l /min
Weight| 3.50 Kg
Insulation Class (when used with the AC/DC adapter mod. UE60-140429SPA1 or
UES65-140429SPA1)| Class II
Insulation Class (when used with an Internal battery)| Internally Powered
Equipment
Insulation Class (when used with a car cigarette lighter cable)| Class II
Size| 350 x 210 x 180 mm
Battery Holding Time| 80 minutes
Battery Time Charge| 180 minutes
Accuracy of Vacuum Indicator| ± 5%
Working Condition| Room temperature: 5 ÷ 35 °C Room humidity percentage:
10 ÷ 93 % RH
Atmospheric pressure: 800 ÷ 1060 hPa
Conservation Condition and Transport| Room temperature: – 25 ÷70 °C Room humidity percentage: 0 ÷ 93% RH
Atmospheric pressure: 500 ÷ 1060 hPa
Please note that the maximum suction value can vary according to reduced atmospheric pressure if the device is used at high altitude. The operator must take use at high altitudes into account. In such conditions, the vacuum produced by the internal pump can drop considerably due to the reduction of atmospheric pressure.
SYMBOLS
CARE AND MAINTENANCE
- If the device was stored at a temperature below the minimum value specified by the manufacturer, keep it in a place with a room temperature of 20°C for at least 1 hour before use.
- If the device was stored at a temperature above the maximum value specified by the manufacturer, keep it in a place with a room temperature of 20°C for at least 30 minutes before use.
CLEANING DEVICE
Use a soft dry cloth with not–abrasive and not–solvent detergents. To clean the device’s external parts always use a cotton cloth dampened with detergent. Don’t use abrasive or solvent detergents.
PARTICULAR CARE SHOULD BE TAKEN TO ENSURE THAT THE INTERNAL PARTS OF THE EQUIPMENT DO NOT GET IN TOUCH WITH LIQUIDS. NEVER CLEAN THE EQUIPMENT WITH WATER.
During all clearing operations use protection gloves and apron (if need be, also wear a face mask and glasses) to avoid getting in contact with contaminating substances (after each utilization cycle of the machine).
ACCESSORIES SUPPLIES
DESCRIPTION | CODE |
---|---|
COMPLETE ASPIRATION JAR 1000ml | RE 210001/02 |
CONICAL FITTING | RE 210410 |
TUBES SET 6 mm x 10 mm | 51100 |
HYDROPHOBIC AND ANTIBACTERIAL FILTER | SP 0046 |
AC/DC ADAPTER (UE60-140429SPA1 or UES65-140429SPA1) | SP 0208/01 |
POWER SUPPLY CORD FOR AC/DC ADAPTER | SP 0020/03 |
CIGARETTE LIGHTER CABLE | SP 0007/02 |
CA-MI FOOT SWITCH (on request) | Code 52130 |
The filter is produced with (PTFE) hydrophobic material to prevent fluids from entering the pneumatic circuit. It should be changed immediately if it becomes wet or if there is any sign of contamination or discoloration. It should also be changed if the unit is used with a patient whose risk of contamination is unknown. Don’t use the suction unit without the protection filter. If the suction unit is used in an emergency or in a patient where the risk of contamination is not known the filter must be changed after each use. Available under request with different versions with complete jar 2000ml.
Suction catheter
Single-use device to be used on a single patient. Do not wash or re-
sterilize after use. Reuse may cause cross-infections. Don’t use after lapse
of the sell-by date Check the expiry date on the original packaging of the
suction catheter and check the integrity of the sterile packaging. CA-MI
declines any liability for injury to the patient correlated to the
deterioration of the above-mentioned sterile packaging due to the handling of
the original packaging by third parties.
WARNING:
Suction tubes for insertion in the human body purchased separately from the
machine should comply with ISO 10993-1 standards on material biocompatibility.
-
Aspiration jar:
The mechanical resistance of the component is guaranteed up to 30 cycles of cleaning and sterilization. -
Beyond this limit, the physical-chemical characteristics of the plastic material may show signs of decay. Therefore, we recommend that you change it.
-
Silicone tubes:
The number of cycles of sterilization and/or cleaning is strictly linked to the employment of the said tube. Therefore, after each cleaning cycle, it is up to the final user to verify whether the tube is suitable for reuse. The component must be replaced if there are visible signs of decay of the material constituting the said component. -
Conical fitting:
The number of cycles of sterilization and the number of cleaning cycles is strictly linked to the employment of the said component. Therefore, after each cleaning cycle, it is up to the final user to verify whether the fitting is suitable for reuse. The component must be replaced if there are visible signs of decay in the material constituting the said component. -
Service life of the device:
More than 1000 hours of operation (or 3 years) by the standard conditions of testing and operation. Shelf life: maximum 5 years from the date of manufacture.
CLEANING OF ACCESSORIES
Before using the device, the manufacturer advises you to clean and/or sterilize the accessories. Washing and/or cleaning the autoclavable jar is to be carried out as follows:
- Wear protective gloves and an apron (glasses and a face mask if necessary) to avoid contact with contaminating substances;
- Disconnect the tank from the device and remove the said container from the support of the device.
- Separate all the parts of the cover (overflow device, washer).
- Disconnect all tubes from the jar and the protection filter
- Wash each part of the container from secretions under cold running water and then clean every single part in hot water (temperature not exceeding 60°C)
- Once again, carefully wash every single part using, if necessary, a non-abrasive brush to remove any deposits.
- Rinse with hot running water and dry all parts with a soft cloth (non-abrasive). It is possible to wash with commercial disinfectants by carefully following the instructions and dilution values supplied by the manufacturer. After cleaning, leave the parts to dry in an open, clean environment.
- Dispose of the aspiration catheter according to that provided by local laws and regulations.
The silicone aspiration tubes and the conical fitting may be carefully washed in hot water (temperature must not exceed 60°C). After cleaning, leave the parts to dry in an open, clean environment. When cleaning is complete, reassemble the container for liquid aspirations according to the following procedure:
- Place the overflow valve into its seat in the cover (under the VACUUM connector)
- Insert the floating valve keeping the o-ring towards the opening of the cage
- Place the o-ring into its seat around the cover
- After completing assembling operations always make sure that the covers seal perfectly to avoid vacuum leakages or liquid exit
After disposing of disposable parts and disassembling the jar wash in running cold water and rinse thoroughly. Then soak in warm water (temperature shall not exceed 60°C). Wash thoroughly and if necessary use a non-abrasive brush to remove incrustations. Rinse in running warm water and dry all parts with a soft cloth (non-abrasive).
The jar and the cover can be autoclaved by placing the parts into the autoclave and running one sterilization stem cycle at 121°C (1 bar relative pressure – 15 min) making sure that the jar is positioned upside down Mechanical resistance of the jar is guaranteed up to 30 cycles of sterilization and cleaning at the indicated conditions (EN ISO 10079-1). Beyond this limit, the physical-mechanical characteristics of the plastic may decrease and replacement of the part is therefore recommended. After sterilization and cooling at the environment temperature of the parts make sure that these are not damaged.
The aspiration tubes can be sterilized on an autoclave using a sterilization cycle at 121°C (1 bar relative pressure – 15 min). The conical connector can be sterilized on an autoclave using a sterilization cycle at 121°C (1 bar relative pressure – 15 min).
DO NOT WASH, STERILIZE, OR PUT IN AUTOCLAVE THE ANTIBACTERIAL FILTER
PERIODICAL MAINTENANCE CHECKS
The NEW ASKIR 230/12V BR suction equipment does not need maintenance or lubrication.
It is, however, to inspect the unit before each use. About the training, given the information contained in the user manual and since it is easy to understand the said device, it doesn’t appear to be necessary. Unpack the instrument and always check the integrity of plastic parts AC/DC switching adapter, and feeding cable, they might have been damaged during previous use. Connect the cable to the electrical network and turn the switch on. Close the aspirator outlet with your finger and with the suction regulator in maximum vacuum position check that the vacuum indicators reach – 75kPa (-0.75 bar) minimum (internal battery). Rotate the knob from right to left and check the aspiration regulating control. The vacuum indicator should go down – 25kPa (-0.25 bar). Verify that loud noises are not present, these can indicate wrong functioning. The device is protected by a safety fuse (F 10A L 250V) situated in the cigarette lighter cable. When replacing, always check the type and value as indicated.
Internally, the device is protected (see electrical specifications) by two fuses (T 15A L 125V) that cannot be reached from the outside. Therefore, contact the manufacturer to request the assistance of an authorized and qualified technician when they need to be replaced. If it’s replaced make sure that its replacement is always the same type and value, as indicated.
The device is made up of a lead battery which cannot be accessed by outside. To replace it, consult the technical staff authorized by the manufacturer.
Fault type | Cause | Solution |
---|---|---|
1. Red light on | Battery run down | Hook up the power cord to the electricity |
mains,
positioning the equipment power switch on 0.
2. No light| Defective AC/DC adapter or technical internal problem| Contact
the technical service
3. No aspiration| Jar Cap badly screwed down| Unscrewed the cap, then re-
screw it correctly
4. No aspiration| The lid seal is not in its seat| Unscrew the cap and insert
the seal properly in its seat
5. The float doesn’t close| If the cap has been washed, ensure that the float
is not
partially detached
|
Insert the float into it’s place
6. The float doesn’t close| The float is covered by dirty material| Unscrewed
the cap, left the, and put it in on the autoclave
7. Low suction| Foam inside the jar| Fill the jar to 1/3 full of ordinary
water
8. No aspiration due to flow leakage of mucus| Filter blocked| Replace filter
9. The Vacuum power on the patient side is either very low or absent| ·
Vacuum regulator set to minimum
· Protection filter blocked or damaged
· Connection tubes blocked, kinked, or disconnected
· Shut-off valve blocked or
| · Turn the vacuum regulator clockwise and check the value of the vacuum on the gauge
· Replace the filter
· Replace or reconnect the tubes, check the jar connections
· Empty the jar, or disconnect the tube from the jar and unblock the shut- off valve. The unit will only work in
| damaged
· Pump motor damaged
| the upright position
· Contact the technical service
---|---|---
10. Noisy| Technical internal problem| Contact the technical service
Faults 1 – 2 – 3 – 4 – 5 – 6 – 7 – 8 – 9 – 10| None of the remedies has
achieved the desired results| Contact the seller or CA-MI After-sales
Assistance Service
The device is made up of a lead battery which cannot be accessed by outside. To replace it, consult the technical staff authorized by the manufacturer.
USE ONLY THE RECOMMENDED BATTERIES FROM CA-MI. THE USE OF OTHER BATTERIES IS NOT RECOMMENDED AND INVOLVES THE CANCELLATION OF THE WARRANTY
If the service personnel has to replace the internal battery, pay special attention to the polarity of the same component. The + / – polarities are indicated directly on the battery. If the overfill security system is activated, don’t proceed with the liquid aspiration. If the overfill security system doesn’t work there are two cases:
- 1° case – If the overfill security system doesn’t work the aspiration will be stopped by the bacteriological filter who avoids the liquid penetration inside the device.
- 2° case – If both the security systems don’t work, there is the possibility that liquid comes inside the device, in this case, return the device to CA-MI technical service.
CA-MI Srl will provide upon request electric diagrams, components lists, descriptions, setting instructions, and any other information that can help the technical assistance staff for product repair.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT CA-MI TECHNICAL SERVICE. CA-MI DOES NOT GIVE A GUARANTEE IF THE INSTRUMENT, AFTER THE TECHNICAL SERVICE CHECKING, APPEARS TO TAMPER. IMPORTANT INFORMATION FOR CORRECT DISPOSAL OF THE PRODUCT BY EC DIRECTIVE 2012/19/UE-WEEE:
The symbol on the device indicates the separate collection of electric and electronic equipment. At the end of the life of the device, don’t dispose of it as mixed solid municipal waste, but dispose of it by referring to a specific collection center located in your area or returning it to the distributor, when buying a new device of the sample type to be used with the same functions. This procedure of separated collection of electric and electronic devices is carried out forecasting a European environmental policy aiming at safeguarding, protecting, and improving environment quality, as well as avoiding potential effects on human health due to the presence of hazardous substances in such equipment or to improper use of the same or parts of the same Caution: The wrong disposal of electric and electronic equipment may involve sanctions.
INSTRUCTION FOR USE
- The device must be checked before each use to detect malfunctions and/or damage caused by transport and/or storage.
- The working position must be such as to allow one to reach the control panel and to have a good view of the empty indicator, the jar, and the antibacterial filter.
- It is recommended not to keep the device in your hands and/or to avoid prolonged contact with the body of the apparatus.
WARNING : For proper use, place the aspirator on a flat, stable surface to have the full volume of use of the jar and better efficiency of the overflow device.
Operation with AC/DC power supply:
The symbol near the instructions before each use identifies the model that is
indicated on the Rating Label and in the instructions for use. 12V jack on the
casing requires the user to read the type of power supply to be connected in
line with
- Place the unit on a flat, horizontal surface.
- Connect the short silicon tube, with an antibacterial filter, to the suction connector. The other tube, with one end connected to the antibacterial filter with the other end to the jar’s lid connector where has fixed the red float. When 90% of the volume of the jar is reached there is the activation of the security float (the float closes the aspiration connector on the jar) to avoid liquid penetration inside the device.
- Connect the long silicon tube to the other jar’s lid connector
- Connect the other end of the long silicon tube to the probe plastic connector then connect the suction probe to it.
- Connect the switching adapter to the device with the appropriate connector and insert the power cable plug into the power socket. To start the treatment press the I switch to turn it on
- Set the desired vacuum value (mmHg / kPa) with the appropriate vacuum regulator. Turning the handle clockwise increases the vacuum value: these values can be read on the “vacuum indicator” instrument.
- To suspend and/or terminate the treatment, press the switch again and pull the plug out from the power socket.
- Unscrew the jar’s lid and fill the jar 1/3 full of ordinary water (this is for easy cleaning operations and rapid reaching of the functional vacuum) then re-screw the lid on the jar correctly.
- To extract the accessories and start with cleaning.
WARNING : The power supply cable plug is the element of separation from the electrical mains system: even if the units are equipped with a special on / off switch button, the power supply plug must be kept accessible once the device is in use to allow a further method of disconnection from the mains supply system.
WARNING : Ensure that the filter is fitted with the Arrows at the patient side.
WARNING : The inside of the medical device must be regularly checked for
the presence of liquids or other visible contamination (secretions). If
liquids or other visible contamination are present, the medical device must be
replaced immediately due to the risk of insufficient vacuum flow.
These products have been designed, tested, and manufactured exclusively for
“single use and for a period of use not exceeding 24 hours unless stated
below.
WARNING : Before using the device, check the battery power status. Before each use proceed with charging the battery. To maintain the device in good condition, recharge the battery every 3 months (when not in use).
Recharging operations:
To be able to charge the internal battery it is necessary to connect the
universal switching adapter (mod. UE 60140429SPA1 or UES65-140429SPA1 of
FUHUA) to the electric network for approx. 180 minutes with the main switch to
position 0. The battery’s autonomy when fully charged is approx. 80 minutes
with continued operations.
TAB. I – INDICATOR LIGHTS DURING OPERATIONS
When an external power supply (regardless of the state of the battery charger)
and when the device is working (after having turned it on), the LED stays in a
FIXED GREEN position.
LED Signal | Phase | Problem / Cause | Solution |
---|---|---|---|
Flashing Green Led | During recharge | Battery recharge running | Wait |
Steady Greed Led | During recharge | Recharging cycle complete | Remove power |
supply
Steady Red Led| During battery operation| Flat battery| Start recharging cycle
WARNING : During this signal, you will hear a long, continuous beep
(duration of sound 0.8 sec
/ sound frequency: every 8.5 sec, which notifies the user
regarding the battery discharge.
Flashing Red Led| The device automatically turns off when the battery is flat| Battery completely flat| When the device is restarted the LED will flash red: begin the battery recharge cycle
immediately
Steady Orange Led| During battery operation| Intermediate status| Guaranteed battery function / Recharge when the red LED
the signal comes on.
Operation using cigarette lighter DC 12V
- Connect the device’s external plug 12V to the lighter plug with the cigarette lighter cable. Check the battery power status of the vehicle before the cigarette lighter cable. Press the switch to start the suction
- Press the switch to the I position to turn it on.
WARNING : Only use the originally supplied or recommended replacement cigarette lighter cables (view the chapter “Important Safety Rules”)
ATTENTION : The on/off button (used as a network switch) is the element that separates the mains; in any case, even if the device is equipped with a specific on/off button, the 12Vdc power plug must remain accessible once the device is in use to allow the simultaneous disconnection of all the poles from the mains.
Foot control (where applicable):
Plug the foot switch power cable into the socket marked with the label on the
back of the device. The foot switch can replace the ON/OFF button for
switching the machine on or off.
NEVER USE THE DEVICE WITHOUT JAR AND / OR PROTECTION FILTER
RISK OF ELECTROMAGNETIC INTERFERENCE AND POSSIBLE REMEDIES
This section contains information regarding the conformity of the compliance with the EN 60601-1-2 (2015) Standard. CISPR group and category classification: group 1, category B. The NEW ASKIR 230/12V BR surgical aspirator is an electro-medical device that requires particular precautions regarding electromagnetic compatibility and must be installed and commissioned according to the electromagnetic compatibility information supplied. Portable and mobile radio communication devices (mobile phones, transceivers, etc.) may interfere with the medical device and should not be used near, adjacent to, or on top of the medical device. If such use is necessary and unavoidable, special precautions should be taken so that the electro-medical device functions properly in its intended operating configuration (for example, constantly and visually checking for the absence of anomalies or malfunctions).
The use of accessories, transducers, and cables different from those specified, except transducers and cables sold by the appliance and system manufacturer as spare parts, can lead to an increase in emissions or i decrease in the immunity of the device or system. The following tables supply information regarding the EMC (Electromagnetic Compatibility) characteristics of the electro-medical device.
Guidance and manufacturer’s declaration – electromagnetic Emissions
The surgical aspirator NEW ASKIR 230/12V BR is intended for use in the electromagnetic environment specified below. The customers or the users of the surgical aspirator NEW ASKIR 230/12V BR should ensure that it’s used in such an environment.
Emissions test
Irradiated / Conducted emissions CISPR11
Irradiated / Conducted emissions CISPR11
Harmonic emissions EN 61000-3-2
Voltage fluctuations/flicker emissions
EN 61000-3-3
Compliance
Group 1
Class [B]
Class [A]
Complies
Electromagnetic environment – guidance
- The surgical aspirator NEW ASKIR 230/12V BR only used RF energy only for its internal functioning. Therefore, its RF emissions are very low and do not cause interference in the proximity of any Electronic appliances.
- The surgical aspirator NEW ASKIR 230/12V BR can be used in all environments, including domestic and those connected directly to the public mains distribution that supplies power to environments used for domestic scopes.
Guidance and manufacturer’s declaration – Immunity Emissions
The surgical aspirator NEW ASKIR 230/12V BR is intended for use in the electromagnetic environment specified below. The customers or the users of the surgical aspirator NEW ASKIR 230/12V BR should ensure that it’s used in such an environment.
Immunity Test| The level indicated by the
EN 60601-1-2
| Compliance Level| Electromagnetic environment – guidance
---|---|---|---
Electrostatic discharge (ESD)
IEC/EN 61000-4-2
| ± 8kV on contact| The device doesn’t change its state| Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the
relative humidity should be at least 30%.
± 2,4,8,15kV in air
Electrical fast transient / burst EN 61000-4-4| ± 2kV power supply lines
± 1kV for input/output lines
| The device doesn’t change its state| The main power quality should be that
of a typical commercial environment or hospital.
Surge
EN 61000-4-5
| ± 1kV differential mode| The device doesn’t change its state| The main power
quality should be that of a typical commercial environment or hospital.
Loss of voltage, brief voltage interruptions, and variations
EN 61000-4-11
| <5%UT (>95% dip UT)
for 0.5 cycle 40%UT (>60% dip UT)
for 5 cycle 70%UT (>30% dip UT)
for 25 cycle
<5%UT (>95% dip UT)
for 5 sec
| –| Mains power quality should be that of a typical commercial environment or
hospital If the user of the surgical aspirator NEW ASKIR 230/12V BR requests
that the appliance operate continuously, the use of a continuity unit is
recommended.
Magnetic field EN 61000-4-8| 30A/m| The device doesn’t change its state| The
power frequency magnetic field should be
measured in the intended installation location to ensure that it’s sufficiently low.
Nota UT is the value of the power supply voltage
Conducted Immunity EN 61000-4-6| 3Vrms 150kHz to 80Mhz (for nonlife-
supporting devices)
| V1 = 3 V rms| The portable and mobile RF communication devices, including cables, must not be used closer to the NEW ASKIR 230/12V BR device, than the separation distance calculated by the equation applicable to the transmitter frequency.
Recommended separation distance d = [3.5 / V1] P
d = [12 / E1] P from 80 MHz to 800MHz d = [23 / E1] P from 800 MHz to 2.5 GHz
Where P is the maximum nominal output voltage of the transmitter in Watt (W) depending on the manufacturer of the transmitter and the recommended separation distance in meters (m). The intensity of the field from the fixed RF transmitters, as determined by an electromagnetic study of the site), could be lower than the level of conformity of each frequency interval b). It is possible to check for interference in proximity to devices identified by the following symbol:
---|---|---|---
Note 1 : At 80 MHz and 800 MHz the interval with the highest
frequency is applied
Note 2 : These guidelines may not be applicable in all situations.
Electromagnetic propagation is influenced by the absorption and reflection of
buildings, objects, and people.
The field intensity for fixed transmitters such as the base stations for radiotelephones (mobile and cordless) and terrestrial mobile radio, amateur radio devices, radio AM and FM transmitters, and TV transmitters can not be theoretically and accurately foreseen. To establish an electromagnetic environment generated by fixed RF transmitters, an electromagnetic study of the site should be considered. If the field intensity measured in the place where the device will be used surpasses the above-mentioned applicable level of conformity, the normal functioning of the device should be monitored. If abnormal performance arises, additional measures such as changing the device’s direction or positioning may be necessary.
The field intensity on an interval frequency of 150 kHz to 80 MHz should be less than 10 V/m.
Recommended separation distance between portable and mobile radio-
communication devices and the monitor
The NEW ASKIR 230/12V BR surgical aspirator is intended to operate in an
electromagnetic environment where RF irradiated interferences are under
control. The client or operator of the NEW ASKIR 230/12V BR device can help
prevent electromagnetic interference by keeping a minimum distance between the
portable and mobile RF communication devices (transmitters) and the NEW ASKIR
230/12V BR device, as recommended below, about the radio-communication maximum
output power.
The maximum nominal output power of the Transmitter W| Separation
distance from the frequency transmitter (m)
---|---
150 kHz to 80 MHz
d = [3.5 / V1] P
| 80 MHz to 800 MHz
d = [12/E1] P
| 800 MHz to 2.7 GHz
d = [23/E1] P
0,01| 0,12| 0,12| 0,23
0,1| 0,38| 0,38| 0,73
1| 1,2| 1,2| 2,3
10| 3,8| 3,8| 7,3
100| 12| 12| 23
For transmitters with a maximum nominal output power not shown above, the recommended separation distance in meters (m) can be calculated using the equation applicable to the transmitter frequency, where P is the maximum nominal output power of the transmitter in Watt (W) depending on the transmitter’s manufacturer.
Note 1 : At 80 MHz and 800 MHz the interval with the highest
frequency is applied
Note 2 : These guidelines may not be applicable in all situations.
Electromagnetic propagation is influenced by the absorption and reflection of
buildings, objects, and people.
WARRANTY CONDITIONS
This product is guaranteed for 24 months from the date of purchase. The warranty includes the repair or replacement of defective spare parts free of charge if the defect has been clearly described by the customer and determined by technical service. Inspections on the part of the seller, performed at the request of the customer and intended to determine whether the device is fully functional, are not covered by the free-of-charge warranty service. This service will be charged to the customer depending on the effort required. The consumables components are not subject to warranty. Consumable components are silicon tubes, filters, seals, conical adaptors, and suction catheters. Also excluded from the warranty is all damage resulting from improper handling, wilful damage, or improper care of the device.
The warranty shall expire if repairs and servicing are not carried out by technical service.
RULES FOR RETURNING AND REPAIRING
UNDER NEW EUROPEAN RULES, CA-MI REQUIRES THE FOLLOWING PROCEDURES TO BE CARRIED OUT TO PROTECT THE INSTRUMENT AND THE SAFETY OF ALL WHO COME IN CONTACT WITH IT.
Before returning an instrument for repair, the external surfaces and all accessories MUST be carefully disinfected with a cloth soaked in methylated spirits or hypochlorite-based solution. The instrument and accessories should then be placed in a bag with a note outlining the disinfection undertaken. Failure to follow this procedure will result in the instrument being returned to the purchaser unrepaired. Instruments returned for repair MUST be accompanied by a description of the problem. CA-MI will not be responsible for damage caused through improper use. To avoid such damage, please read the instructions carefully. Where CA-MI determines that an instrument is faulty, a replacement will be provided only if a SALES RECEIPT and STAMPED GUARANTEE are provided. CA-MI will not be responsible for damaged accessories. These may be replaced at the customer’s expense.
Warranty Certificate
CA-MI Srl Via Ugo La Malfa 13 – Frazione Pilastro – 43013 Langhirano (PR)
Italia
Tel. +39 0521 / 637133 – 631138
Fax. +39 0521 / 639041
Registro A.E.E : IT8020000000264
Registro Pile e Accumulatori: IT09060P00000971
E-mail: vendite@ca-mi.it / export@ca-
mi.it www.ca-mi.it