FilmArray RFIT-ASY-0094 Bio Threat Panel User Guide
- June 15, 2024
- FilmArray
Table of Contents
- FilmArray RFIT-ASY-0094 Bio Threat Panel
- Product Information
- Product Usage Instructions
- FilmArray® BioThreat Panel Quick Guide For use with BioFire® FilmArray®
- STEP 1: Prepare Pouch
- STEP 2: Hydrate Pouch
- STEP 3: Prepare Sample Mix
- STEP 3: Prepare Sample Mix (cont.)
- STEP 4: Load Sample Mix
- STEP 5: Run Pouch
- Step 6: Review Report
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
FilmArray RFIT-ASY-0094 Bio Threat Panel
Product Information
- Product Name: Wells RFIT-ASY-0094
- Intended Use: Environmental Surveillance and Research
- Storage: Components from this kit should not be stored or used with any other kit. Discard any extra components from the kit after all pouches have been used
Safety Precautions
- To avoid contamination, clean work area and always wear gloves.
- Work according to appropriate biosafety conditions while manipulating potentially infectious samples.
Product Usage Instructions
Step 1: Prepare Pouch
- Insert the pouch into the Pouch Loading Station.
- Place the Sample Injection Vial into the red well.
- Place the Hydration Injection Vial into the blue well.
Step 2: Hydrate Pouch
- Twist off the Hydration Injection Vial, leaving the blue plastic cover in the Pouch Loading Station.
- Insert the Hydration Injection Vial into the pouch hydration port.
- Forcefully push down to puncture the seal and verify that the Hydration Solution has been drawn into the pouch.
Step 3: Prepare Sample Mix
Add Sample Buffer:
- Hold the Sample Buffer Tube so that the tip is facing up.
- Gently pinch the textured plastic tab on the side of the tube until the seal snaps. Do not touch the tip of the tube.
- Invert the tube over the Sample Injection Vial.
- Dispense Sample Buffer by squeezing the liquid-filled end using a slow, forceful squeeze, followed by a 2nd squeeze. Avoid excessive foaming by not squeezing the tube additional times.
For Liquid Samples (including Air-into-PBS / Dry Filter / Liquid)
- After collecting the liquid sample, use the Transfer Pipette to draw filter wash or liquid to the 3rd line (approx. 300 uL) and add it to the Sample Injection Vial.
- Tightly close the lid of the Sample Injection Vial and mix the sample by gently inverting it 3 times.
- Return the Sample Injection Vial to the red well of the Pouch Loading Station.
Step 4: Load Sample Mix
- With the Sample Injection Vial in the Pouch Loading Station, twist to loosen it from the red plastic cover and pause for 5 seconds.
- Lift the Sample Injection Vial from the red plastic cover and insert it into the Sample port.
- Forcefully push down to puncture the seal and verify that the Sample Mix has been drawn into the pouch.
Step 5: Run Pouch
Follow the instructions on the screen for starting a test. Select BT PBS
protocol.
Step 6: Review Report
The test results will be displayed on the screen, showing information about
the sample and the detected organisms.
Specifications
- Bacteria:
- Bacillus anthracis
- Yersinia pestis
FAQ
-
Q: Can the components from this kit be used with other kits?
A: No, the components from this kit should not be stored or used with any other kit. -
Q: How should I avoid contamination while using this product?
A: Clean your work area and always wear gloves to avoid contamination. -
Q: What should I do with extra components from the kit?
A: Discard any extra components from the kit after all pouches have been used.
FilmArray® BioThreat Panel Quick Guide For use with BioFire® FilmArray®
2.0 System
- For Environmental Surveillance and Research Use Only. Not for use in diagnostic procedures
- NOTE: To avoid contamination, clean work area and always wear gloves.
- Work according to appropriate biosafety conditions while manipulating potentially infectious samples.
- NOTE: Components from this kit should not be stored or used with any other kit.
- Discard any extra components from the kit after all pouches have been used.
- Materials required for powder and surface swab but not provided
- NOTE: BioFire® FilmArray® 2.0 Instrument should be powered on and ready for use prior to pouch preparation.
STEP 1: Prepare Pouch
- Insert pouch into Pouch Loading Station.
- Place Sample Injection Vial into red well.
- Place Hydration Injection Vial into blue well.
STEP 2: Hydrate Pouch
- Twist off Hydration Injection Vial, leaving blue plastic cover in Pouch Loading Station, and insert into pouch hydration port.
- Forcefully push down to puncture seal and verify that Hydration Solution has been drawn into pouch.
STEP 3: Prepare Sample Mix
Add Sample Buffer
-
Hold the Sample Buffer Tube so that the tip is facing up.
-
Gently pinch the textured plastic tab on the side of the tube until the seal snaps
NOTE: Do not touch the tip of the tube. -
Invert the tube over the Sample Injection Vial.
-
Dispense Sample Buffer by squeezing the liquid-filled end using a slow, forceful squeeze, followed by a 2nd squeeze.
NOTE: Squeezing the tube additional times will generate excessive foaming, which should be avoided.
For Liquid Samples (including Air-into-PBS / Dry Filter / Liquid)
- After collecting the liquid sample, use Transfer Pipette to draw filter wash or liquid to the 3rd line (approx. 300 uL) and add to Sample Injection Vial.
- Tightly close the lid of the Sample Injection Vial and mix sample by gently inverting 3 times.
- Return the Sample Injection Vial to the red well of the Pouch Loading Station.
STEP 3: Prepare Sample Mix (cont.)
For Surface Swab Samples (including Powder / Surface):
-
After collecting sample, place the tip of the swab into the Sample Injection Vial and swirl for 5 seconds.
-
Break the swab handle at the molded break point, leaving the tip of the swab in the Sample Injection Vial.
NOTE: If the swab handle does not have a molded break point, cut the tip of the swab with clean scissors, leaving the tip of the swab in the Sample Injection Vial. -
Discard the handle in an appropriate waste container.
-
Tightly close lid of Sample Injection Vial and mix sample by gently inverting 3 times.
-
Return Sample Injection Vial to red well of Pouch Loading Station.
STEP 4: Load Sample Mix
- With Sample Injection Vial in the Pouch Loading Station, twist to loosen it from the red plastic cover and pause for 5 seconds.
- Lift Sample Injection Vial from red plastic cover and insert into Sample port.
- Forcefully push down to puncture seal and verify that Sample Mix has been drawn into pouch.
STEP 5: Run Pouch
- Follow instructions on screen for starting a test.
- NOTE: Select BT PBS protocol.
Step 6: Review Report
- Run Summary – Displays information about the sample. Internal Controls:
- If ‘Pass’, results are valid.
- If ‘Fail’ or ‘Invalid’, RETEST SAMPLE.
- Result Summary – Displays the test results for each organism.
- ‘Detected’: Lists names of any detected organisms.
- ‘Not Detected’: Lists names of organisms that were not detected.
- ‘N/A’, pathogen was not reported. Refer to Instruction Booklet for more information.
- Run Details – Displays information about the pouch, protocol, run status, operator, instrument, serial number, and lot number. Run Status:
- If ‘Completed’, run is complete.
- If ‘Incomplete’, ‘Aborted’, or any other error message, RETEST SAMPLE.
- NO TE: Refer to the FilmArray BioThreat Panel Instructions for Use for reporting information. If repeated error messages are obtained, contact BioFire Defense Technical Support.