trafag 8236 FPT Flush Membrane Transmitter Instruction Manual

June 15, 2024
trafag

**trafag 8236 FPT Flush Membrane Transmitter Instruction Manual

**

Instructions

Warning: Before installation, commissioning, and operation, ensure that the pressure transmitter is suitable for the application in terms of measuring range, design and environmental conditions. Non-observance can result in serious injury and/ or damage to equipment.

Intended use: This instrument converts pressure into an electrical signal.

The instrument has been designed and built solely for the intended use described here and may only be used accordingly. If the equipment is used in a different  manner, the protection provided by the equipment may be impaired and Trefah shall not be liable for any claims at all.

Datasheets

FPT 8236

www.trafag.com/H72343

Type label description (Example)

Electrical connections

**Designation| ** EN175301-803A (DIN43650-A)| Cable (PVC) (PUR)4 x 0.25mm2| ** M12x15–pol.| Packard Metri Pack 3–pol.
---|---|---|---|---
Type code| 8236.XX.XXXX .05| 8236.XX.XXXX .08/22/24| 8236.XX.XXXX .35| 8236.XX.XXXX .51
****
Pin configuration| **| ****| |
Output| Load resistance| I SUPPLY| U SUPPLY
---|---|---|---
4 … 20 mA| (USupply-9V)/ 20mA| (=Signal output)| 9 … 32 VDC
0 … 5 VDC| > 2.5 kΩ| ≤ 20 mA| 9 … 32 VDC
1 … 6 VDC| > 5.0 kΩ| ≤ 20 mA| 9 … 32 VDC
0 … 10 VDC| > 5.0 kΩ| ≤ 20 mA| 15 … 32 VDC
0.5 4.5 VDC ratiom.| ≥ 5.0 kΩ| ≤ 20 mA| 5 (4.75… 5.25) VDC

Connection of the measuring equipment

Current output 2-wires

**Voltage output 3-wires

**

Pressure connection

**Mounting thread G1/2“ DIN EN ISO 1179-1

**

Adjustment of connector EN 175301-803 A

Only versions without Loctite securing of housing nut