Energys DTF-102 Receiver Instruction Manual

June 15, 2024
Energys

Energys DTF-102 Receiver

Energys-DTF-102-Receiver-image

Product Information

Specifications

  • Product Name: DTF-102 Receiver
  • Model: TypeDTF-102
  • Product Number: No.-3243E R2 Product
  • Part Number: KX-621038-XXXXXX

Product Applications

Please consult with us in advance regarding use for the following purposes:

  • Radiation-related equipment
  • Nuclear applications
  • Medical use
  • Uses that have a large impact on important property or rights
  • Other uses related to life and body

Scope of Liability and Warranty

Warranty Period

The warranty period is 1 year after delivery. If a separate warranty period is stipulated in a document such as a delivery specification, the contract shall take precedence.

Warranty Conditions

During the warranty period, if a failure occurs due to our responsibility, we will provide a substitute or replacement part or repair the delivered product free of charge. However, the following cases are not covered by the warranty:

  • When used in environments other than those specified in this manual
  • Failure to follow the instructions in this manual or use in a manner other than that described in this manual
  • In case of failure due to incorrect operation
  • If the product is repaired, processed, remodeled, or disassembled by anyone other than us or someone authorized by us
  • When used in combination with parts other than our company
  • In the event of a disaster beyond our responsibility, such as a natural disaster
  • In case of damage, breakage, or failure due to dropping or transportation after purchase
  • Failure due to corrosion, rust, etc. Deterioration of appearance
  • Consumable parts

Warranty Scope

The scope of warranty is limited to our products. Regardless of the warranty period, our company does not take any responsibility for indirect damage due to failure of our products or unforeseen defects by our company (Loss of profit, opportunity loss, compensation for damage or failure to products other than those supplied by us, compensation for other operations, compensation for accidents, etc.).

About This Instruction Manual

  • Deliver this manual to the final user.
  • The contents of this manual are subject to change without notice for improvement.
  • Unauthorized reproduction or duplication of part or all of the contents of this manual is strictly prohibited.
  • If you find any unclear points, errors, omissions, etc. regarding the contents of this manual, please contact us or our distributors.

About Related Documents

Related documents:

  • TF-10 Type Probe Transmitter – Instruction Manual (No.-3071-*)
  • TF- Type Probe Transmitter – Instruction Manual (No.-3064-, 3144-, 3231-, 3223-)
  • TF- Type Probe Transmitter – Instruction Manual (No.-3066-, 3068-)

Notation of the Figure

The figures in this manual may be emphasized, simplified, or partially omitted for convenience of explanation. Pictures such as screens may differ from the actual display. In addition, the contents described may be display examples.

Product Usage Instructions

Note

This section interprets the necessary points for correct operation and notifies the important points for handling the product.

Refer

This section indicates the parts where you can refer for information.

Tip

This section provides reference information.

Frequently Asked Questions

Q: What is the warranty period for the DTF-102 Receiver?

A: The warranty period for the DTF-102 Receiver is 1 year after delivery.

Q: What is not covered by the warranty?

A: The following cases are not covered by the warranty: – When used in environments other than those specified in this manual – Failure to follow the instructions in this manual or use in a manner other than that described in this manual – In case of failure due to incorrect operation – If the product is repaired, processed, remodeled, or disassembled by anyone other than us or someone authorized by us – When used in combination with parts other than our company – In the event of a disaster beyond our responsibility, such as a natural disaster – In case of damage, breakage, or failure due to dropping or transportation after purchase – Failure due to corrosion, rust, etc. Deterioration of appearance – Consumable parts

Instruction Manual
DTF-102 Receiver
TypeDTF-102
No.-3243E R2 Product
DTF-102 Receiver KX-621038-XXXXXX

Preface

This manual is written for those who handle DTF-102 type receiver. Be sure to read this manual before using the product to ensure proper and safe operation of the product. In particular, be sure to read “For Your Safety” and use the product correctly. And keep it in a place where anyone using this product can view it at any time. The contents of this manual are subject to change without notice for improvement. Please refer to the latest instruction manual when using. If you find any unclear points, errors, omissions, etc. regarding the contents of this manual, please contact us or our distributors.
Product Applications
Consult with us in advance regarding use for the following purposes. Radiation-related equipment Nuclear applications Medical use Uses that have a large impact on important property or rights Other uses related to life and body
Scope of Liability and Warranty
[ Warranty Period ] 1 year after delivery If a separate warranty period is stipulated in a document such as a delivery specification, the contract shall take precedence.
[ Warranty Conditions ] During the warranty period, if a failure occurs due to our responsibility, we will provide a substitute or replacement part or repair the delivered product free of charge. However, the following cases are not covered by the warranty.
If a separate warranty period is stipulated in a document such as a delivery specification, the contract shall take precedence.
When used in environments other than those specified in this manual Failure to follow the instructions in this manual or use in a manner other than that described in this
manual In case of failure due to incorrect operation If the product is repaired, processed, remodeled, or disassembled by anyone other than us or
someone authorized by us When used in combination with parts other than our company In the event of a disaster beyond our responsibility, such as a natural disaster In case of damage, breakage, or failure due to dropping or transportation after purchase Failure due to corrosion, rust, etc. Deterioration of appearance. Consumable parts
I

[Warranty Scope] The scope of warranty is limited to our products. Regardless of the warranty period, our company does not take any responsibility for the following. Indirect damage due to failure of our products or unforeseen defects by our company (Loss of profit, opportunity loss, compensation for damage or failure to products other than those supplied by us, compensation for other operations, compensation for accidents, etc.).
II

About This Instruction Manual
Deliver this manual to the final user. The contents of this manual are subject to change without notice for improvement. Unauthorized reproduction or duplication of part or all of the contents of this manual is strictly
prohibited. If you find any unclear points, errors, omissions, etc. regarding the contents of this manual, please
contact us or our distributors.

About Related document

Related documents are below. TF-10 Type Probe Transmitter TF- Type Probe Transmitter TF- Type Probe Transmitter

Instruction Manual Instruction Manual Instruction Manual

No.-3071- No.-3064-, 3144-, 3231-, 3223- No.-3066-, 3068-*

Description in This Manual
Note
This interprets the necessary points for correct operation and notifies the important points for handling the product.

Refer
This indicates the part where to refer for information.

Tip
This indicates reference information.

Notation of the Figure
The figures in this manual may be emphasized, simplified, or partially omitted for convenience of explanation. Pictures such as screens may differ from the actual display. In addition, the contents described may be “display examples”.
Trademarks
Company names and brand names are either registered trademarks or trademarks of the respective companies. (R), (TM) symbols may be omitted in this manual.
Original language
This is the English translation of an original Japanese document.
Disposal of the Product
When disposing of the product, comply with local regulations.

III

For Your Safety

Warning messages given here are for safe and correct use of the product and for prevention of harm or damage. These are important safety note, so be sure to read it carefully before use and be sure to observe it.
The meanings of the symbols are as follows.

DANGER

This indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. This is to be limited to the most extreme situations.

WARNING This indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.

CAUTION

This indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices.

IV

When deciding on the “Installation Location”

DANGER

Explosion or Fire
This product is not explosion-proof. Do not use in an explosive gas atmosphere.
When using a standard gas cylinder, install or store the standard gas cylinder in a location with an ambient temperature of 40°C or less and out of direct sunlight.

CAUTION

Injury, Malfunction or Damage
Install in a location that satisfies the installation location conditions described in “Table 3-1 Installation location”.

When “Installing” or “Transporting”

CAUTION

Fall, Damage or Injury
Installation work correctly according to the instruction manual.
Tighten screws with the appropriate tightening torque. Insufficient tightening may cause damage or drop. Also, if it is tightened too much, it may not be possible to remove it.

When “Wiring”

WARNING

Fire, Electric shocks, Malfunction
Wiring work should be done by a person with appropriate technical training and experience.
Use the power supply at the rated voltage.
Be sure to turn OFF the main power supply before performing wiring connection work or wiring check.
Connect grounding.
Do not apply an excessive load to the cables and cord, such as bending and stretching them repeatedly, putting a heavy thing on them.
Use appropriate wiring material according to the rating of the device.

V

When “Starting measurement” or “Maintenance work”

WARNING

Fire, Electric shocks, Malfunction
Be sure to turn OFF the main power supply before performing wiring connection work or wiring check.
Before turning ON the power, make sure that the power wiring is properly connected.
If the fuse blows, check the cause and replace it with one of the same capacity and type. Also, when replacing the fuse, be sure to turn OFF the main power supply before starting work.
Be sure to use parts specified by our company for replacement parts and consumables.

Other CAUTION

Fire, Electric shocks, Injury or Damage
In the event of a failure that cannot be determined by looking at the instruction manual, be sure to contact our company or the agency where you purchased the product, and request repairs if necessary.

VI

Product Handling Information
If this product is used in a manner not specified by us, it may impair the protective functions and performance provided by this product. It is strictly prohibited to modify the product by the customer. Observe the following precautions.
General
Use in a manner not specified in this manual may impair the protection provided by this product. Do not disassemble the parts other than those specified in this manual. Do not modify by the customer. Do not use accessories other than this product. Even if the temperature is within the range of the ambient temperature specification of this product,
the life of the product may be shortened if it is used in a constantly high ambient temperature environment. This product is intended for industrial environments. In a domestic environment this product may cause radio interference in which case the user may be required to take adequate measures.
Precautions during construction
Install the product in a location where the required space can be secured. If space is not secured, daily inspections and maintenance will not be possible, leading to equipment stoppages and product damage.
Notes on settings
When using the product for the first time or after replacing the sensor, always setting the sensor parameters to receiver. The parameter settings to be used, see the inspection report.
VII

VIII

Overview

1.1 About This Product
This product is a control device that drives and controls the oxygen sensor in an oxygen analyzer that uses a limiting current type zirconia oxygen sensor, and is responsible for concentration calculation and output. It is used in combination with our oxygen sensor (transmitter). Since the oxygen sensor of this analyzer does not require pretreatment of the measurement gas, it is smaller and has a faster response speed than conventional sampling-type devices equipped with a pretreatment device. In addition, since calibration can be performed using only “instrumentation air” using compressed air, which is commonly found in factories and ships, there is the advantage of not needing to install or manage standard gas cylinders.

1.2 Products and Accessories

When you receive the product, open the package and check for any shipping damage. Also, make sure that the delivered items and accessories are correct.

Table 1-1 Supply Parts List (one set)

Name

No.

Q’ty

Remarks

Receiver

KX-621038-XXXXXX

1

If Receiver is delivered in a unit case or mounted on an independent panel, refer to the supply parts list of the delivery specifications.

Table 1-2 Option Parts List

Name

No.

Q’ty

Remarks

Stand mounting bracket KS-224193

1

For detail, see “3.3.3 Stand Mount”.

1.3 Temporary Storage
When storing the product temporarily prior to installation, observe the following conditions. Away from direct sunlight. The ambient temperature is between -10 and 50, with little variation in temperature. There is little humidity and dust. The location is not exposed to rainfall. There is little mechanical vibration. There are no corrosive gases or dangerous gases.

5

2 Part Names
2 Part Names
2.1 Receiver
Dimensions
250

Window

Knurled fastener 156 135

300 270

Front view

Door

Right side view

Bracket 2

200
Bottom view

Mounting hole 10
Bracket 1 Wiring hole 7-27 With grommets

*Dimensional tolerances without individual tolerance indications are according to JEM 1459.
6

Inside

2 Part Names

Door

Display panel For detail, see below fig.

Display Panel
Power lamp Lights when the power is on.

Display panel

Operation key Used for setting sensor parameters, changing/checking various settings,
and releasing errors.

Fuse
Power switch
Terminal block
Display 1 “Oxygen concentration”, “Setting data”, “Error code” and “Countdown” are displayed.
lamp Lights when the concentration unit is %.
ppm lamp Lights when the concentration unit is ppm. Display 2 “Output range”, “CH No.”, “States of Analyzer”, “Concentration upper limit alarm”, “Concentration lower limit alarm” are displayed.
7

Installation

3 Installation
This product is a device that controls Oxygen Sensor and calculates and outputs concentration. In order to measure the oxygen concentration, connect our company’s zirconia Oxygen Sensor and use it by constructing piping related parts. A typical system configuration example is shown below.
3.1 System Configuration Example
Note
Confirm the delivery specifications for scope of delivery and actual system configuration.

3.1.1 System Configuration Example 1
This system is an example of using “TF-10 Probe Transmitter” as an oxygen sensor to measure the oxygen concentration in flue gas from a boiler, etc. Manually operated valves such as needle valve and ball valve are used for calibration gas lines. At the time of calibration, after manually operating the valve to supply the calibration gas, calibration is executed by key operation on the receiver. Use instrument air or atmospheric gas for calibration. Calibration using standard gas cylinders such as zero gas and span gas is not required.

TF-10 Probe Transmitter

DTF-102 Receiver

Check valve

Control/signal wires for sensor
4 cores shielded×2

Power supply 100 to 240 VAC
Contact input
Analog output (DC4 to 20mA) Contact output

Flow meter

Needle valve

Pressure reducing valve

Calibration gas (Instrument air or atmospheric gas

Fig. 3-1 System Configuration Example 1

8

3 Installation
3.1.2 System Configuration Example 2
This system is an example of using “TF-10 Probe Transmitter” as an oxygen sensor to measure the oxygen concentration in flue gas from a boiler, etc. A solenoid valve is installed in the calibration gas line, and the contact output of the receiver is used to open and close the solenoid valve. Calibration can be performed by methods such as “receiver key operation”, “contact input”, and “receiver internal timer”. Use instrument air or atmospheric gas for calibration. Calibration using standard gas cylinders such as zero gas and span gas is not required.
DTF-102 Receiver
TF-10 Probe Transmitter

Check valve

Control/signal wires for sensor
4 cores shielded×2
Contact output

Power supply 100 to 240 VAC
Contact input
Analog output (DC4 to 20mA) Contact output

Flow meter

Solenoid valve

Calibration gas Pressure (Instrument air or atmospheric gas reducing valve

Fig. 3-2 System Configuration Example 2

3.1.3 System Configuration Example 3

This system uses a TF-III type probe transmitter as an oxygen sensor, to measure the oxygen

concentration in high-temperature flue gas in a heating furnace. In addition to calibration gas piping,

ejector air piping and purge air piping are used. Since the calibration gas and purge air are operated by

opening and closing the solenoid valve using the contact output of the receiver, calibration and purge

can be performed by methods such as “receiver key operation”, “contact input”, and “receiver internal

timer”. Use instrument air gas for calibration. Calibration using standard gas cylinders such as zero

gas and span gas is not required. The TF-III type probe transmitter uses a separate power supply

and temperature control device for heating and keeping warm in order to prevent dew condensation in

the sampling section.

K thermocouple compensating lead

Temperature control device

DTF-102 Receiver

Heater power

TF- Probe Transmitter

Calibration gas piping Ejector air piping Purge air piping

Control/signal wires for sensor
8 cores shielded×1
Contact output

Power supply 100 to 240 VAC Contact input Analog output (DC4 to 20mA) Contact output
Instrument air

Flow meter
Solenoid Pressure valve reducing valve
Fig. 3-3 System Configuration Example 3

9

3 Installation

3.2 Place Conditions for Installation

When deciding on the “Installation Location”

DANGER

Explosion or Fire
This product is not explosion-proof. Do not use in an explosive gas atmosphere.
When using a standard gas cylinder, install or store the standard gas cylinder in a location with an ambient temperature of 40°C or less and out of direct sunlight.

CAUTION

Injury, Malfunction or Damage
Install in a location that satisfies the installation location conditions described in “Table 3-1 Installation location”.

Note
Even if the temperature is within the range of the ambient temperature specification of this product, the life of the product may be shortened if it is used in a constantly high ambient temperature environment.
Install the product in a location where the required space can be secured. If space is not secured, daily inspections and maintenance will not be possible, leading to equipment stoppages and product damage.
This product is intended for industrial environments. In a domestic environment this product may cause radio interference in which case the user may be required to take adequate measures.

For safe, correct use of your oxygen analyzer, install the analyzer in a location with the following conditions to provide the best possible installation conditions.
Not caused condensation by sudden temperature fluctuations. Not directly exposed to much radiated heat. Not significantly affected by electromagnetic fields. Not exposed to much moisture and/or dust. Little voltage fluctuation. Little power frequency fluctuation. Meeting the following conditions “Table 3-1 Installation Conditions”:

Item Installation location

Table 3-1 Installation Conditions Spec.
Indoor

Ambient temperature

-10 to +50

Ambient humidity

90RH or less (Not condensation)

Environmental protection

No waterproof structure

For the installation location conditions of the oxygen sensor (transmitter), check the instruction manual of the oxygen sensor (transmitter) to be used.

10

3 Installation

3.3 Installing the Receiver

When “Installing” or “Transporting”

CAUTION

Fall, Damage or Injury
Installation work correctly according to the instruction manual.
Tighten screws with the appropriate tightening torque. Insufficient tightening may cause damage or drop. Also, if it is tightened too much, it may not be possible to remove it.

3.3.1 Panel Mount
Fit the receiver into the panel cutout. Attach Bracket 1 and Bracket 2 in the direction shown in Fig. 3-5 to hold the panel down. Panel thickness range1.6~3.2 mm

Panel cutout

Bracket 1

Bracket 2

+ 1 0

273

215

+1 0

Fig. 3-4 Panel Cutout Dimensions

Hex. bolt M8 Plane washer M8 Spring washer M8
Fig. 3-5 Panel Mount

3.3.2 Wall Mount

Attach Bracket 1 and Bracket 2 in the direction shown in Fig. 3-6, fix them to the structure on the rear

side of the receiver.

*Bolts for mounting to the rear side structure are not supply. Prepare by yourself.

2-M8 *When fixing with a screw hole

Note
Install on a structure with sufficient strength to support the receiver. Receiver weightApprox. 6.5 kg

Mounting hole 10 Bolts for mounting are not supply. Fig. 3-6 Wall Mount
11

3 Installation
3.3.3 Stand Mount
An optional stand mounting bracket is required to mount the receiver on a 50A pole. Attach Bracket 1 and Bracket 2 in the direction shown in Fig. 3-7. And attach to the pole with the Stand Mounting Brackets 1 and 2.

Bracket 1

50A Pole (1)

(2) (4), (5), (6)

(3), (5), (6)

Fig. 3-7 Stand Mount

Stand Mounting Bracket (Option Parts)

(1)

(2)

(3) (4) (5) (6)

Table 3-2 Parts List of Stand Mounting Bracket

Symbol

Parts Name

Q’ty Remarks

(1) Stand mounting bracket 1 1

(2) Stand mounting bracket 2 1

(3)

Hex. bolt

2 M8 L10

(4)

Hex. bolt

2 M8 L15

(5)

Plane washer

4

M8

(6)

Spring washer

4

M8

Fig. 3-8 Stand Mounting Bracket

12

3 Installation

3.4 Connecting Wires

When “Wiring”

WARNING

Fire, Electric shocks, Malfunction
Wiring work should be done by a person with appropriate technical training and experience.
Use the power supply at the rated voltage.
Be sure to turn OFF the main power supply before performing wiring connection work or wiring check.
Connect grounding.
Do not apply an excessive load to the cables and cord, such as bending and stretching them repeatedly, putting a heavy thing on them.
Use appropriate wiring material according to the rating of the device.

3.4.1 Connecting to Terminal Block

Pass wires through the wiring hole on the bottom of the receiver. Connect to the terminal block by referring to “Fig. 3-9 Terminal Block Arrangement Diagram” and “Table 3-3 Terminal Block Compatibility”.
-Terminal sizeM4 -Applicable crimp terminal outer diameterMax.8.5 mm

Note
The wiring port is a 27 hole with a grommet. Implement the dust-proof and water-proof treatment after wiring by yourself.

Terminal Block

IN1 IN2 VS+ IP+ H+ S+ OUT+ RY1 RY2 RY3 RY4A RY4B

POWER POWER

COM IN3 VS- IP- H- S- OUT- RY1 RY2 RY3 RY4A RY4B

E

Fig. 3-9 Terminal Block Arrangement Diagram

Item Contact
Input
Sensor Control /
Signal

Symbol IN1 IN2 IN3 COM VS+ VSIP+ IPH+ HS+ S-

Table 3-3 Terminal Block Compatibility

Contents

Item

Symbol

Input No-voltage contact Operating current 5 mA DC 24 V is applied) Common for contact input Sensing voltage
Sensing current
Heater voltage application
Heater voltage measurement

Analog Output

OUT+ OUT-

Contact Output
Power Grounding

RY1 RY2 RY3 RY4A RY4B POWER POWER
E

Contents
DC 4-20 mA Isolated output Load resistance 550 or less
No-voltage relay contact Contact capacity
AC 250 V 1A Resistive load DC 30 V 1A Resistive load
100-240 VAC ±10 max.50 VA Grounding terminal

13

3 Installation
3.4.2 Wiring Connection Example
Fig. 3-10 shows a wiring connection example.
Note
Wire according to the delivery specifications or the customer’s usage.

Oxygen Sensor (Transmitter)

Shielded Cable *The round-trip wiring
resistance 5 or less.
Recommendation cable CVV-S 2 sq×8 cores
(Temperaturemax.60) HCVV-S 2 sq×8 cores
(Temperaturemax.75)

Power Supply AC 100 to 240 V±10 max50 VA
Circuit breaker or Circuit protector

VS+ VS IP+ IP H+ H S+ S

POWER POWER
Power

E
Grounding

DTF-102 Receiver

Contact Input
COM IN1 IN2 IN3 0V +24V +24V +24V

Analog Output
OUT+ OUT

Contact Output
RY1 RY1 RY2 RY2 RY3 RY3 RY4A RY4A RY4B RY4B

S.V.

S.V.

for Air

DC 24 V

Input No-voltage contact Operating current 5 mA DC 24 V is applied)
Shielded Cable

Analog Output DC 4 to 20 mA Load Resistance 550 or less

for Purge

During Maintenance
PLC ect.

Analyzer Error

No-voltage relay contact Contact capacity
AC 250 V 1A Resistive load DC 30 V 1A Resistive load

Fig. 3-10 Wiring Connection Example

14

3 Installation

3.4.3 Wiring Method When Replacing from DTF-101 Receiver
DTF-101 receiver and DTF-102 receiver differ in the number of contact input/output points and the content corresponding to each terminal symbol. When replacing DTF-101 with this product(DTF-102), connect the wiring referring to the following.
Power line / Analog output line
There is no change. Connect to terminals with the same terminal symbol as the DTF-101 receiver.
Contact input line / Contact output line
Use “3.4.4 Wiring Confirmation Sheet When Replacing from DTF-101 Receiver” to connect contact input and contact output wiring. Here are the steps:
1. Check the input/output settings of the existing receiver
Check the contact input and contact output details from t product number (CAT NO.) of DTF-101 receiver. The product number is indicated on the rating plate of the receiver. (The rating plate is pasted on the inside of the door.)
2. Check the input/output you are actually using
Check the terminal symbols that are actually wired to the terminal block of the DTF-101 receiver. Enter the input/output items in the columns of “Used contact input items” and “Used contact output items” only for the terminals that are wired. Even if contact input/output is set, there is no need to wire the replacement receiver (DTF-102) because the terminals that are not actually wired are not used.
3. Check the input/output settings of the DTF-102 receiver
Check the contact input and contact output contents from the replacement receiver (DTF-102) product number (CAT NO.). Described in the “DTF-102 contact input setting” and “DTF-102 contact output setting” columns.
4. Relocate the wiring to the corresponding terminal on the DTF-102
receiver
Move the wiring to the terminals of the DTF-102 receiver that have the same items as the “contact input items in use” and “contact output items in use”.

DTF-101 Receiver

IN1 IN2 IN3 VS+ IP+ H+ S+ OUT+ RY1 RY2 RY3 RY4 RY5 POWER POWER

COM IN4 IN5 VS-

IP-

H-

S- OUT- RY1 RY2 RY3 RY4 RY5

E

DTF-102 Receiver

IN1 IN2 VS+ IP+ H+ S+ OUT+ RY1 RY2 RY3 RY4A RY4B

COM IN3 VS-

IP-

H-

S- OUT- RY1 RY2 RY3 RY4A RY4B

POWER POWER E

Fig. 3-11 Terminal block arrangement diagram of DTF-101 receiver and DTF-102 receiver Wiring Connection Example

15

3 Installation

3.4.4 Wiring Confirmation Sheet When Replacing from DTF-101 Receiver

Existing Receiver (DTF-101)

Product number of receiver

KX-621034 –

KX-621036

KX-621028

Contact Input

Symbol

IN1

IN2

IN3

IN4

IN5

Contact input setting table of existing receiver

0

OFF

OFF

OFF

OFF

OFF

1 REM RANGE OUT HOLD REM CAL REM PURGE MFT IN

2 REM RANGE OUT HOLD REM CAL REM PURGE RESET

Z

Wiring to the terminal block Yes None Yes None Yes None Yes None Yes None

Contact Output

Symbol

0

1

2

3

4

5

Contact output

6

setting table of

existing receiver

7

8

A

B

C

D

E

Z

Wiring to the terminal block

RY1 OFF RANGE H RANGE H RANGE L AIR AIR ZERO SPAN AIR AIR AIR AIR AIR AIR

RY2

RY3

OFF

OFF

MAINTE ALARM H

MAINTE ALARM L

MAINTE ALARM H&L

ZERO

RANGE H

ZERO

SPAN

SPAN

RANGE H

ALARM H&L RANGE H

ALARM H&L RANGE H

PURGE ALARM H&L

ALARM H&L PURGE

ZERO

PURGE

PURGE RANGE H

PURGE

ALARM H

RY4 OFF PURGE PURGE PURGE MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE

RY5 ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR

Yes None Yes None Yes None Yes None Yes None

IN1

IN2

IN3

IN4

IN5

Used contact input items

DTF-102 Receiver

Relocate the wiring to the corresponding terminal on the DTF-102

DTF-102 contact input

IN1

IN2

IN3

setting

Contact Input
Contact input setting table of
DTF-102

Symbol

IN1

IN2

IN3

A

OFF

OFF

OFF

B REM RANGE OUT HOLD SLEEP

C

REM CAL REM PURGE SLEEP

D

REM CAL REM PURGE REM RANGE

E

REM CAL REM PURGE OUT HOLD

F

REM CAL REM RANGE OUT HOLD

Z

Product number of receiver KX-621038 – A

RY1

RY2

RY3

RY4

RY5

Used contact output items

Relocate the wiring to the corresponding terminal on the DTF-102

DTF-102 contact output

RY1

RY2

RY3

RY4A

RY4B

setting

Contact Output

Symbol

RY1

RY2

A

OFF

OFF

B

RANGE H PURGE

Contact Output setting table of
DTF-102

C ALARM H&L PURGE

D

AIR

PURGE

E

AIR

PURGE

F AIR or PURGE ALARM H&L

G AIR or PURGE RANGE H

H AIR or PURGE RANGE H

Z

RY3 OFF MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE

RY4A

RY4B

ERROR

ERROR

ERROR

ERROR

ERROR or ALARM H&L

ERROR

ERROR

ERROR or ALARM H&L

Fig. 3-12 Wiring Confirmation Sheet When Replacing from DTF-101 Receiver
“MFT IN” of DTF-101 receiver corresponds to “SLEEP” of DTF-102 receiver. “RESET” of the DTF-101 receiver is not supported by the DTF-102 type receiver. If “Z” is used in the product number, check the contact input/output items in the delivery specifications. If the same solenoid valve is driven using the contact output of “AIR” and “Purge” in the existing
equipment, With the DTF-102 receiver, it is possible to aggregate into one contact output by using “AIR or PUREG”. If you are using the contact output of “ERROR” and “ALARM H&L” in the existing equipment, by using “ERROR or ALARM H&L” on the DTF-102 receiver, it is possible to aggregate into one contact output.

16

3 Installation

Wiring confirmation sheet usage example 1

Usage example conditions Product No. of existing receiver KX-621034-A3D180 Wiring location of existing receiver IN3RY1RY4RY5 Product No. of DTF-102 receiver KX-621038-ACFADC
Existing Receiver (DTF-101)

Product number of receiver
KX-621034 – A 3 D 1 8 0
KX-621036 KX-621028

Contact Input

Symbol

IN1

IN2

IN3

IN4

IN5

Contact input setting table of existing receiver

0

OFF

OFF

OFF

OFF

OFF

1 REM RANGE OUT HOLD REM CAL REM PURGE MFT IN

2 REM RANGE OUT HOLD REM CAL REM PURGE RESET

Z

Wiring to the terminal block Yes None Yes None Yes None Yes None Yes None

Contact Output
Symbol

0

1

2

3

4

5

Contact output

6

setting table of

existing receiver

7

8

A

B

C

D

E

Z

Wiring to the terminal block

RY1 OFF RANGE H RANGE H RANGE L AIR AIR ZERO SPAN AIR AIR AIR AIR AIR AIR

RY2

RY3

OFF

OFF

MAINTE ALARM H

MAINTE

ALARM L

MAINTE ALARM H&L

ZERO

RANGE H

ZERO

SPAN

SPAN

RANGE H

ALARM H&L RANGE H

ALARM H&L RANGE H

PURGE ALARM H&L

ALARM H&L PURGE

ZERO

PURGE

PURGE RANGE H

PURGE

ALARM H

RY4 OFF PURGE PURGE PURGE MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE

RY5 ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR

Yes None Yes None Yes None Yes None Yes None

IN1

IN2

IN3

IN4

IN5

Used contact input items

REM CAL

Used contact output items

RY1 AIR

RY2

RY3

RY4 MAINTE

RY5 ERROR

DTF-102 Receiver

DTF-102 contact input setting

IN1

IN2

IN3

REM CAL REM PURGE SLEEP

Contact Input
Contact input setting table of
DTF-102

Symbol

IN1

IN2

IN3

A

OFF

OFF

OFF

B REM RANGE OUT HOLD SLEEP

C

REM CAL REM PURGE SLEEP

D

REM CAL REM PURGE REM RANGE

E

REM CAL REM PURGE OUT HOLD

F

REM CAL REM RANGE OUT HOLD

Z

Product number of receiver
KX-621038 – A C F A D C

DTF-102 contact output setting

RY1 AIR

RY2 PURGE

RY3 MAINTE

RY4A

RY4B ERROR

Contact Output

Symbol

RY1

RY2

A

OFF

OFF

B

RANGE H PURGE

Contact Output setting table of
DTF-102

C ALARM H&L PURGE

D

AIR

PURGE

E

AIR

PURGE

F AIR or PURGE ALARM H&L

G AIR or PURGE RANGE H

H AIR or PURGE RANGE H

Z

RY3 OFF MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE

RY4A

RY4B

ERROR

ERROR

ERROR

ERROR

ERROR or ALARM H&L

ERROR

ERROR

ERROR or ALARM H&L

DTF-101
IN1 IN2 IN3 VS+ IP+ H+ S+ OUT+ RY1 RY2 RY3 RY4 RY5 POWER POWER

COM IN4 IN5 VS-

IP-

H-

S- OUT- RY1 RY2 RY3 RY4 RY5

E

DTF-102

IN1 IN2 VS+ IP+ H+ S+ OUT+ RY1 RY2 RY3 RY4A RY4B

COM IN3 VS-

IP-

H-

S- OUT- RY1 RY2 RY3 RY4A RY4B

POWER POWER E

Fig. 3-13 Wiring for usage example 1 (contact input/output only)

17

3 Installation

Wiring confirmation sheet usage example 2

Usage example conditions Product No. of existing receiver KX-621034-A3D1A2 Wiring location of existing receiver IN3IN5RY1RY2RY4RY5 Product No. of DTF-102 receiver KX-621038-ACFAFC

Existing Receiver (DTF-101)

Product number of receiver
KX-621034 – A 3 D 1 A 2
KX-621036 KX-621028

Contact Input

Symbol

IN1

IN2

IN3

IN4

IN5

Contact input setting table of existing receiver

0

OFF

OFF

OFF

OFF

OFF

1 REM RANGE OUT HOLD REM CAL REM PURGE MFT IN

2 REM RANGE OUT HOLD REM CAL REM PURGE RESET

Z

Wiring to the terminal block Yes None Yes None Yes None Yes None Yes None

Contact Output

Symbol

0

1

2

3

4

5

Contact output

6

setting table of

existing receiver

7

8

A

B

C

D

E

Z

Wiring to the terminal block

RY1 OFF RANGE H RANGE H RANGE L AIR AIR ZERO SPAN AIR AIR AIR AIR AIR AIR

RY2

RY3

OFF

OFF

MAINTE ALARM H

MAINTE ALARM L

MAINTE ALARM H&L

ZERO

RANGE H

ZERO

SPAN

SPAN

RANGE H

ALARM H&L RANGE H

ALARM H&L RANGE H

PURGE ALARM H&L

ALARM H&L PURGE

ZERO

PURGE

PURGE RANGE H

PURGE ALARM H

RY4 OFF PURGE PURGE PURGE MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE

RY5 ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR ERROR

Yes None Yes None Yes None Yes None Yes None

IN1 Used contact input items

IN2

IN3

IN4

IN5

REM CAL

MFT IN

Used contact output items

RY1 AIR

RY2 Purge

RY3

RY4 MAINTE

RY5 ERROR

DTF-102 Receiver

DTF-102 contact input setting

IN1

IN2

REM CAL REM PURGE

IN3 SLEEP

Contact Input
Contact input setting table of
DTF-102

Symbol

IN1

IN2

IN3

A

OFF

OFF

OFF

B REM RANGE OUT HOLD SLEEP

C

REM CAL REM PURGE SLEEP

D

REM CAL REM PURGE REM RANGE

E

REM CAL REM PURGE OUT HOLD

F

REM CAL REM RANGE OUT HOLD

Z

Product number of receiver
KX-621038 – A C F A F C

DTF-102 contact output

RY1

RY2

RY3

setting

AIR or PURGE ALARM H&L MAINTE

RY4A

RY4B ERROR

Contact Output

Symbol

RY1

RY2

A

OFF

OFF

B

RANGE H PURGE

Contact Output setting table of
DTF-102

C ALARM H&L PURGE

D

AIR

PURGE

E

AIR

PURGE

F AIR or PURGE ALARM H&L

G AIR or PURGE RANGE H

H AIR or PURGE RANGE H

Z

RY3 OFF MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE MAINTE

RY4A

RY4B

ERROR

ERROR

ERROR

ERROR

ERROR or ALARM H&L

ERROR

ERROR

ERROR or ALARM H&L

18

DTF-101

IN1 IN2 IN3 VS+ IP+ H+ S+ OUT+ RY1 RY2 RY3 RY4 RY5 POWER POWER

COM IN4 IN5 VS- IP-

H-

S- OUT- RY1 RY2 RY3 RY4 RY5

E

DTF-102

IN1 IN2 VS+ IP+ H+ S+ OUT+ RY1 RY2 RY3 RY4A RY4B

COM IN3 VS-

IP-

H-

S- OUT- RY1 RY2 RY3 RY4A RY4B

POWER POWER E

Fig. 3-14 Wiring for usage example 2 (contact input/output only)

Measurements

4 Measurements

When “Starting measurement”

WARNING

Fire, Electric shocks, Malfunction
Before turning ON the power, make sure that the power wiring is properly connected.

Note
Be sure to check the safety precautions for the oxygen sensor (transmitter) to be used before starting to use it.
4.1 Necessary Steps before Starting
When using the product for the first time or after replacing the sensor, always perform following before starting measurement.
Setting the sensor parameters Air Point Calibration
Refer
For Air Point Calibration method, see “5 Calibration”.

4.1.1 Setting Items of Sensor Parameters

Set the sensor parameters of the sensor connected to the receiver. There are two sensor parameters that need to be set: “Linearizer No.” and “Heater normal temp resistance”. For setting value of sensor parameters, see inspection report that attached the product.

Table 4-1 Sensor Parameters that Need to be Set

Item

CH No.

Set value

Remarks

Linearizer No.

143

See inspection report

Heater normal temp resistance

192

See inspection report

If “Heater normal temp resistance” is not listed in “Inspection report”, no setting is required.

Refer
Detail of setting method, see “4.3.5 Data Setting/Checking Method”.

19

4 Measurements
4.2 Starting and Stopping Measurement Starting Measurement
After turn the power supply ON, this analyzer goes through 3-minutes countdown (temperature rise) to enter measurement state. And display the oxygen concentration, output the analog output (4-20 mA). However, in order to make stable measurements, warm up the instrument sufficiently before using it. warm-up time as a guideApprox. 2 hours
Note
If the ambient temperature is lower than -10°C, turn ON the power after the ambient temperature reaches -10°C or higher.
Stopping Measurement
The measurement stops when the power supply to the analyzer is turned OFF.
20

4 Measurements
4.3 Key Operation
The key operation is used to setting sensor parameters, calibration operations, recover from error conditions, and change other settings. This is a very important operation, so be sure to check it.
Note
Key operation may change the oxygen concentration output. Therefore, when using the oxygen concentration output for control purposes, release the control functions before operating the keys.
4.3.1 Explanation of Keys
Keys
Key Switch from “Normal Display” to “CH No. selection state”. Change the number of CH No. and setting data.
Key Switch from “Normal Display” to “CH No. selection state”. Moves the CH No. and setting data digit to the left. (The digit that can be changed blinking.)
ENT Key Decide CH No. CH data setting state. Decide CH data CH No. selection state.
Key Switch from “CH No. selection state” to “Normal Display”. Cancel the CH data setting and move to “CH No. selection state”. Clear error when an error occurs. Insert a decimal point in the blinking digit. Switch between plus and minus of setting data.
21

4 Measurements

4.3.2 Display Transition
After the power is turned ON, the display of the receiver changes as shown in Figure 4-1, depending on the state change of the receiver and key operation.
Power ON

Temp. Rising

or
Switching Display

Clear

3-minutes

Measurement

or
Switching Display

CH No. selection state

Error detection

Clear

Error Detection

or
Switching Display

ENT
ENT Decision Cancel

CH data setting state

States

Fig.4-1 Display Transition
Table 4-2 Display by State of Receiver Contents

Temp. Rising

3 minutes countdown to completion of temperature rise on display 1

Measurement

Display 1Oxygen concentration measurement
Display 2Symbol of Analog output range ” – H – ” : HIGH range ” – L – ” : LOW range

Display example
2.59
– – –
20.90
– H –

Error Detection Error code on display 1

CH No. selection state
CH data setting state

Display 1CH data of selected CH No. Display 2CH No. Change the number of CH No. by Key Move to CH data setting state by ENT key
Display 1CH data of selected CH No. Display 2CH No. Change CH data by Key Decide CH data by ENT key Cancel CH data setting by Key

E-05
– – –
1
0 0 0
1 1
0 0 0

22

4 Measurements

4.3.3 Display by Operating State

Operating states such as Sleep mode, Calibrating, Purging, etc. are indicated by symbols on display 2. In any operating state, the display can be switched to “CH No. selection state” with the or keys.

Operating states

Table 4-3 Display by Operating Sates

Contents

Display 2

Display 1

Display example

Sleep Mode During Sleep Mode

SLP —–

– – – – S L P

Air Point Calibrating

During Air supply at Semi Auto / Auto Calibration

CAA

Measured value

Zero Point Calibrating

During Zero gas supply at Semi Auto / Auto Calibration

CA 1

Measured value

20.92
C A A
2.02
CA 1

Span Point Calibrating

During Span gas supply at Semi Auto / Auto Calibration

CA2

Measured value

9.02
C A 2

Purging ON

During purge air supply at Semi Auto / Auto Purge

P6H

Measured value

Purging OFF

During purge air not supply at Semi Auto / Auto Purge interval

P6F

Measured value

Recovery Time

During Recovery Time at Semi Auto / Auto Calibration or Semi Auto / Auto Purge

FEC

Measured value

20.92
P 6 H
20.92
P 6 F
1 5.60
F E C

Manual

During Manual Output Holding

Output Holding

H6L

Measured value

20.92
H 6 L

Concentration Upper Limit Alarm

Detecting Concentration Upper Limit Alarm

AH

Measured value

20.92
A H

Concentration Lower Limit Alarm

Detecting Concentration Lower Limit Alarm

AL

Measured value

2.00
A L

Note
If the operating states in the table overlap, the contents listed in the upper row of the table take precedence. e.g.) “Concentration Upper Limit Alarm” and “Manual Out Holding”indicate “HoL”

23

4 Measurements

4.3.4 Data Settings Overview
This product has various setting items assigned to each 3-digit CH number. To change or check the setting value, select the CH No. corresponding to the setting item, and change/check the setting value of that CH No.
Refer
Detail of CH data, see “7.6 CH Setting Data Table”.

4.3.5 Data Setting/Checking Method

The basic operations for setting and checking CH data are as follows.

1. Turn ON the power supply
When the receiver is OFF, turn it ON. *Data setting and checking are possible even
when the sensor is not connected.

Display 1

2.59
– – –

Display 2

2. Switch the display to “CH Data Display”

Press

key or

key to enter “CH No. selection state”.

3-digit number (CH No.) will be displayed on Display 2.

*Immediately after the power is turned ON,

it is in Temp. Rising state, but you can switch to the display.

3. Select CH No.

Use

key and

key to change to the CH No. to set

the number on Display 2.

The data of the selected CH No. is displayed on the Display 1.

4. Set CH data
– Set the Display 2 to CH No. to be set and press ENT key to move “CH data setting state”. (Data of Display 1 blinks by one digit)

1
0 0 0

CH No. select

3.00
1 9 2

Decision
ENT

Cancel

3.00
1 9 2

– Use

key,

key to change the number on Display 1

to the set value.

– Press ENT key to decide the set value. The set value will be reflected and return to “CH No. selection state”. To continue setting data, select CH No. and set the value.

Change setting value
ENT Decision
3.22
1 9 2

– Use

key When you cancel CH data setting and return to

“CH No. selection state”.

*sample value

5. Return to “Normal Display”

Press

key to return “Normal Display”.

*When the sensor temperature rising

is not completed, it returns to Temp. Rising state.

20.90
– H –

24

Calibration

5 Calibration
5.1 About Calibration
In order to correctly measure oxygen concentration, perform gas calibration using the reference gas (hereinafter referred to as calibration) in following cases.
When using analyzer for the first time. After replacing the oxygen sensor. Periodic inspections.
Recommended calibration cycle Air-Point Calibration1/month

Remark Warms up sufficiently before Calibration.
Warm-up time as a guideApprox. 2 hours

5.2 Calibration Point and Calibration Type

5.2.1 Calibration Point Type
Oxygen analyzers using our limiting current type zirconia oxygen sensor can perform measurement only by calibrating the air point. Normally, it is used for 1-Point Calibration with the calibration point as the air point. For the air point, use “instrument air”, “atmosphere”, or “standard gas with atmospheric equivalent concentration”. Other calibrations are performed to further improve accuracy for specific gas concentrations, or when accuracy confirmation and calibration using standard gases are required.

Cal. points Air-point

Table 5-1 Calibration Point Type

Contents

Examples of reference gases

Calibrate the entire measuring range at the Air-point
(Normally measured by Air-point calibration only)

Instrument air ISO 8573-1:2010[3:4:3] Calibration gas concentration :”20.90vol.%”
Atmosphere Atmospheric oxygen concentration calculated from temperature and relative humidity. Or typical value “20.60vol.%”
Standard gas cylinder Standard gas with oxygen concentration equivalent to atmospheric air (N2 base)

Zero-point

Calibrate the Zero-point of High-range and Low-range

Standard gas equivalent to 5 to 10% of measuring range e.g.0 to 25vol.% range2.0vol.%O2/N2 base

Spanpoint

Calibrate the Span-point of Standard gas equivalent to 85 to 90% of measuring range

Low-range

e.g.0 to 10vol.% range9.0vol.%O2/N2 base

Minus Calibrate the Minus spanspan-point point of Minus range

3 kinds of mixed gas e.g.-8vol.% (CO:9vol.%, H2:7vol.%, CO2:9vol.% /N2 base)

25

5 Calibration

5.2.2 Calibration method Type

This product can be calibrated by the following three methods. When changing the calibration method, it is necessary to change the CH data setting. Normally, we set the calibration method according to the customer’s usage at the time of shipment.

Cal. method Manual
Calibration
Semi Auto Calibration
Auto Calibration

CH 270 setting
0
1
2

Table 5-2 Calibration Method Type

Contents

Required plumbing system

A calibration method in which the operation of supplying and stopping the calibration gas to the oxygen sensor, waiting for the indicated value to stabilize, and the operation of loading the calibration value are all performed manually.
A calibration method in which the “calibration start operation” is performed by key operation of the receiver or contact input.
The operation of feeding and stopping the calibration gas to the oxygen sensor, waiting for stabilization, and loading the calibration value are performed automatically.
A calibration method in which the “calibration start operation” is performed automatically using the receiver’s internal timer.
The operation of feeding and stopping the calibration gas to the oxygen sensor, waiting for stabilization, and loading the calibration value are performed automatically.
Calibration start operation by receiver key operation and contact input can be used together.

Calibration gas plumbing system with manual valve operation e.g.) 3.1.1 System Configuration
Example 1
Plumbing system that can open and close solenoid valves with receiver contact output. e.g.) 3.1.2 System Configuration
Example 2 3.1.3 System Configuration Example 3

26

5 Calibration

5.3 Manual Calibration

5.3.1 Manual Calibration (Air Point)

When performing Air-point calibration by Manual Calibration, set the table 5-3 in advance. After setting it once, it is necessary to reset only when the set value is changed.

Table 5-3 CH Setting for Air-point Calibration by Manual Calibration

CH No.

Contents

Set value

Remarks

123

Air-point calibration concentration [vol.%]

180

Calibration point selection by Manual Calibration

276

Calibration point setting used for calculation

20.90 20.60 or calculated value Standard gas concentration
8
0

Instrument air Atmosphere Standard gas cylinder 8Air-point When used only for Air-point

Air-point calibration by Manual Calibration is performed in the following procedure.
1. Set CH setting
Set ” Table 5-3″ Omit it When it has already been set.
2. Send calibration air to the oxygen sensor
Send calibration air to the oxygen sensor by operating the valve, etc.

3. Check the flow rate, wait for the
measurement value is stable

Flow Valve meter

Check that the flow rate with the flowmeter.

And wait for the measurement value is stable.

Check the flow rate to be sent in the instruction manual of the oxygen sensor (transmitter).

Instrument air or atmospheric gas

4. Perform Air-point calibrationset “1” in CH No.181
After confirming the stability of the measurement value, set “1” in “CH No.181”. Calibration is executed and the displayed value becomes the value set in “Air-point calibration concentration”.

5. Stop the calibration air
Stop the sending calibration air by operating the valve, etc.

Tip
If you want to hold the analog output during Manual Calibration, you can hold the analog output by using CH No.029 Manual output hold. For details, see “7.1.3 Output hold function”.

27

5 Calibration

5.3.2 Manual Calibration (Air Point & Other Points)

When performing other points calibration by Manual Calibration, set the table 5-4 in advance. After setting it once, it is necessary to reset only when the set value is changed.

Table 5-4 CH Setting for Air-point Calibration and Other Points Calibration by Manual Calibration

CH No.

Contents

Set value

Remarks

120

Zero-point calibration concentration [vol.%]

121

Span-point calibration concentration [vol.%]

122

Minus span-point calibration concentration [vol.%]

Standard gas concentration
Standard gas concentration
Standard gas concentration

20.90

Instrument air

123

Air-point calibration concentration [vol.%]

180

Calibration point selection by Manual Calibration

20.60 or calculated value
Standard gas concentration
Select from right

Atmosphere
Standard gas cylinder
5Zero-point 6Span-point 7Minus span-point 8Air-point

276

Calibration point setting used for calculation *1

Select from right

0Air 1Air and Zero 2Air and Span 3Air and Zero and Span 4Span 5Zero and Span

277 Span-point calibration plus / minus *2

Select from right

0Plus() 1Minus()

*1 Calibration point setting used for calculation Calibration point used for oxygen concentration calculation

*2 Span-point calibration plus / minus

CH No.277 is data for manufacturer setting. Please contact us if any changes are required.

28

5 Calibration

Air-point calibration, Zero-point calibration and Span-point calibration by Manual Calibration are performed in the following procedure.
1. Set CH setting
Set the calibration point to be calibrated in “CH No.276”. Set the gas concentration of the calibration point to be calibrated to “CH No.120-123”.
2. Perform Air-point calibration
Perform Air-point calibration in the same way “5.3.1 Manual Calibration (Air Point)”.

3. Send Zero-gas to the oxygen sensor
Send Zero-gas to the oxygen sensor by operating the valve, etc.
4. Check the flow rate, wait for the measurement value is stable
Check that the flow rate with the flowmeter. And wait for the measurement value is stable. Check the flow rate to be sent in the instruction manual of the oxygen sensor (transmitter).
5. Perform Zero-point calibration
After confirming the stability of the measurement value, set “5” in “CH No.180”. Next, “1” in “CH No.181”. Calibration is executed and the displayed value becomes the value set in “Zero-point calibration concentration”.
6. Stop the zero gas
Stop the sending Zero-gas by operating the valve, etc.

Zero-point calibration only

7. Send Span-gas to the oxygen sensor

Send Span-gas to the oxygen sensor by operating the valve, etc.

8. Check the flow rate, wait for the measurement value is stable

Check that the flow rate with the flowmeter. And wait for the measurement value is stable.

Check the flow rate to be sent in the instruction manual of the oxygen sensor (transmitter).

9. Perform Span-point calibration
After confirming the stability of the measurement value, set “6” in “CH No.180”.

Span-point calibration only

Next, “1” in “CH No.181”. (When calibrating Minus span-point, set “7” in “CH No.180”.)

Calibration is executed and the displayed value becomes the value set in “Span-point calibration

concentration”.

10. Stop the Span-gas

Stop the sending Span-gas by operating the valve, etc

Flow meter

Valve

Instrument air or atmospheric gas
Zero-gas or Span-gas

29

5 Calibration
5.4 Semi Auto / Auto Calibration
Semi-auto and Auto calibration require a piping system that drives the solenoid valve for calibration gas using the contact output of the receiver.

Contact output

Flow Valve meter

Instrument air or atmospheric gas

5.4.1 Semi Auto / Auto Calibration (Air Point) – advance setting ­

When performing Air-point calibration by Semi Auto Calibration or Auto Calibration, set the table 5-5 in advance. After setting it once, it is necessary to reset only when the set value is changed.

Table 5-5 CH Setting for Air-point Calibration by Semi Auto Calibration or Auto Calibration

CH No.

Contents

Set value

Remarks

123

Air-point calibration concentration [vol.%]

208 Using contact input for calibration

270 Calibration method type

271 Air-point gas sending time [min]

273 Recovery-time [min] *3

274

Calibration cycle time [XX day ­ XX hour] *4

275

Calibration delay time [XX day ­ XX hour] *4

276

Calibration point setting used for calculation

279

Calibration point selection by Semi Auto Cal. or Auto Cal.

20.90 20.60 or calculated value Standard gas concentration
0 or 1
1 or 2
Any setting (Default5) Any setting (Default3) Any setting (Default30-00)
Any setting (Default00-07)
0
0

Instrument air
Atmosphere
Standard gas cylinder 0No use 1Use 0Manual Calibration 1Semi Auto Calibration 2Auto Calibration
Setting range1 – 99
Setting range1 – 99
Setting range[00-01] ­ [99-23] Time until first calibration start after power ON or timer reset Setting range[00-01] ­ [99-23] When only use Air-point calibration
When only use Air-point calibration

3 Recovery-time The grace period until the calibration gas returns to the sample gas after the calibration gas supply is stopped. The contact output “Maintenance” is closing until the Recovery-time ends.
4 Setting required only when using Auto Calibration

Note
Set CH271, 273, 274, 275 arbitrarily by the customer before starting use.

30

5 Calibration

5.4.2 Semi Auto Calibration (Air-Point) – Key Operation or Contact Input –

Air-point calibration by Semi Auto Calibration is performed in the following procedure.

1. Set CH setting
Set ” Table 5-5 ” Omit it When it has already been set.
2. Air-point calibration start operation
(Set “1” in “CH No.278” or Contact input close)
Using key on receiverset “1” in “CH No.278” Using contact inputCorresponding contact input terminals
close for 0.1 seconds or more. For detail of contact input see “7.3 Contact Input”.
3. (Check the Air-point calibration starting)
When Air-point calibration starts, display 2 switches to “CAA”. (Switching is about 5 seconds after calibration start operation)
After the completion of loading the calibration value, it moves to the Recovery-time (Display 2 shows “rEC”).
After the recovery time ends, the display returns to normal.
Tip
CH No.278 is a switch that starts/stops Semi-Auto Calibration. Start calibrationset “1” Stop calibrationset “2”

Indication during Air-point Calibration using key on receiver
1
2 7 8
Cal. start operation ENT Air supply is started
19.52
C A A
Loading the calibration value
20.60
C A A
Move to Recovery-time
15.60
F E C
Calibration complete Return to normal display
12.60
– H –

Operation of contact input/output, display, and analog output during Air-point calibration by Semi Auto Calibration are as shown in “Fig. 5-1 Semi Auto Calibration (Air-point) Timing Chart”.

Cal. Start Operation

Items

Key operation Contact input

Contact in/output DisplayAnalog output

ON

Set “1” in “CH No.278” or Contact input ON

about 0.1sec.

Conact output Display

“Maintenance”
“Air” “Air or Purge”
Display 2 Display 1

CLOSE
( In “NO” setting )
5sec. CH 271Air-point gas sending time CH 273Recovery-time

CLOSE
( In “NO” setting )

loading cal. value

CAA

fEC

Measurement value (before cal.)

Measurement value (after cal.)

Analog output
4-20A
By out hold setting

No hold
Setting value (CH 024) Value 5 sec. before
event occurrence (Defalut)

Measurement value (before cal.)

Measurement value (after cal.)

Measurement v alue (bef ore cal.)
Measurement v alue (bef ore cal.)

Output hold (setting value) Output hold (just before starting cal.)

Measurement v alue (af ter cal.)
Measurement v alue (af ter cal.)

Fig.5-1 Semi Auto Calibration (Air-point) Timing Chart

31

5 Calibration

5.4.3 Auto Calibration (Air-point)
Air-point calibration by Auto Calibration is performed in the following procedure.
1. Set CH setting
Set ” Table 5-5 ” Omit it When it has already been set.
2. Reset calibration cycle
Turn OFF the power switch of receiver and turn ON the power again. Or set “1” in “CH No.314”. Calibration is started automatically after “Calibration delay time”. After the first calibration is started, the calibration is automatically started at the cycle set in “calibration cycle time”.

Refer
Detail for operation of calibration cycle reset see “6.4 Reset of Cycle Timer and Countdown”.

Operation of contact output, display during Air-point calibration by Auto Calibration is as shown in “Fig. 5-2 Auto Calibration (Air-point) Timing Chart”. Operation of display 1 and analog output same as Semi Auto Calibration.

Items Cal. cycle Cal. cycle timer

CH275Calibration delay time

Contact output Display

CH 274Calibration cycle time

CH 274Calibration cycle time

Conact output

“Maintenance” Reset of cal. cy cle timer
5sec.

CLOSE
( In “NO” setting )

CH 271

CH 273

5sec.

CLOSE
( In “NO” setting )

CH 271

CH 273

“Air.” “Air or Purge”

CLOSE
( In “NO” setting )

l oa ding cal. va lue

CLOSE
( In “NO” setting )

Display

Display 2

– H –

CAA fEC – H – CAA fEC

5sec.
– H –

Fig.5-2 Auto Calibration (Air-point) Timing Chart

32

5 Calibration

5.4.4 Semi Auto/Auto Calibration (Air Point & Other Points) – advance setting –

When performing other point calibration by Semi Auto Calibration or Auto Calibration, set the table 5-6 in advance. After setting it once, it is necessary to reset only when the set value is changed.

Table 5-6 CH Setting for other point Calibration by Semi Auto Calibration or Auto Calibration

CH No. 120 121 122

Contents
Zero-point calibration concentration [vol.%] Span-point calibration concentration [vol.%] Minus span-point calibration concentration [vol.%]

Set value
Standard gas concentration Standard gas concentration Standard gas concentration

Remarks

20.90

Instrument air

123

Air-point calibration concentration [vol.%]

208 Using contact input for calibration

20.60 or calculated value
Standard gas concentration
0 or 1

Atmosphere
Standard gas cylinder
0No use 1Use

270 Calibration method type

1 or 2

0Manual Calibration 1Semi Auto Calibration 2Auto Calibration

271 Air-point gas sending time [min]

272

Zero-point and Span-point gas sending time [min]

273 Recovery-time [min]

274

Calibration cycle time [XX day ­ XX hour]

Any setting (Default5)
Any setting (Default5)
Any setting (Default3)
Any setting (Default30-00)

Setting range1 to 99 Setting range1 to 99 Setting range1 to 99 Setting range[00-01] to [99-23]

275

Calibration delay time [XX day ­ XX hour]

Any setting (Default00-07)

Time until first calibration start after power ON or timer reset Setting range[00-01] to [99-23]

276

Calibration point setting used for calculation *5

Select from right

0Air 1Air and Zero 2Air and Span 3Air and Zero and Span 4Span 5Zero and Span

279

Calibration point selection by Semi Auto Cal. or Auto Cal. *6

Select from right

0Air 1Air and Zero 2Air and Span 3Span

5 Calibration point setting used for calculation Calibration point used for oxygen concentration calculation
6 Calibration point selection by Semi Auto Cal. or Auto Cal. Calibration points that operated by Semi Auto Calibration or Auto Calibration. When calibrate Airpoint and other point, set “1” or “2”.

Note
CH271,272,273,274 and 275 must be set by the customer before use.

33

5 Calibration

5.4.5 Semi Auto Calibration (Air Point & Other Points) – Key Operation or Contact Input –

Air-point calibration, Zero-point calibration and Span-point calibration by Semi Auto Calibration are performed in the following procedure.

1. Set CH setting
Set ” Table 5-6 ”
Omit it When it has already been set

2. Calibration start operation
(Set “1” in “CH No.278” or Contact input close)
Using key on receiverset “1” in “CH No.278” Using contact inputCorresponding contact input terminals
close for 0.1 seconds or more. For detail of contact input see “7.3 Contact Input”.

3. (Check the Air-point calibration starting)
When calibration starts, display 2 switches to below. (Switching is about 5 seconds after calibration start operation) Air-point & Zero-pointCAACA 1rEC Air-point & Span-pointCAACA2rEC (Switching is about 5 seconds after calibration start operation)

After the completion of loading the calibration value, it moves to the Recovery-time (Display 2 shows “rEC”).

After the recovery time ends, the display returns to normal.

Operation of contact input/output, display, and analog output during Air-point and Zero-points calibration by Semi Auto Calibration are as shown in “Fig. 5-3 Semi Auto Calibration (Air-point & Zeropoint) Timing Chart”. For Air-point and Span-point calibration, contact output “Zero” operation changes to “Span” operation.

Cal. start operation

Items

Key operation Contact input

Contact in/output DisplayAnalog output

ON

Set “1” in “CH No.278” or Contact input ON

about 0.1sec.

Conact output
Display Analog output 4-20A
By out hold setting

“Maintenance”
“Air” “Air or Purge”
“Zero” Display 2

CLOSE
( In “NO” setting )

5sec.

CH 271

CLOSE
( In “NO” setting )

l oa ding cal. va lue

CH 272

CH 273Recovery-time

CLOSE
( In “NO” setting )

loading cal. value

CAA CA 1 fEC

Display 1

Measurement value (before cal.)

Measurement value (after Air-point cal.)

Measurement value (after cal.)

No hold
Setting value (CH 024) Value 5 sec. before
event occurrence (Defalut)

Measurement value (before cal.)

Measurement value (after Air-point cal.)

Measurement value (after cal.)

Measurement v alue (bef ore cal.)

Output hold (setting value)

Measurement v alue (af ter cal.)

Measurement v alue (bef ore cal.)

Output hold (just before starting cal.)

Measurement v alue (af ter cal.)

Fig.5-3 Semi Auto Calibration (Air-point & Zero-point) Timing Chart

34

5 Calibration

5.4.6 Auto Calibration (Air Point & Other Points)
Air-point & Zero-point calibration or Air-point & Span-point calibration by Auto Calibration are performed in the following procedure.
1. Set CH setting
Set ” Table 5-6 ” Omit it When it has already been set
2. Reset calibration cycle
Turn OFF the power switch of receiver and turn ON the power again. Or set “1” in “CH No.314”. Calibration is started automatically after “Calibration delay time”. After the first calibration is started, the calibration is automatically started at the cycle set in “calibration cycle time”.
Refer
Detail for operation of calibration cycle reset see “6.4 Reset of Cycle Timer and Countdown”.

Operation of contact output during Air-point and Zero-points calibration by Auto Calibration are as shown in “Fig. 5-4 Auto Calibration (Air-point & Zero- point) Timing Chart”. For Air-point and Span-point calibration, contact output “Zero” operation changes to “Span” operation. Operation of display 1 and analog output during calibration are same as Semi Auto Calibration.

Items Display Cal. cycle timer

CH275Calibration delay time

Contact output Display
CH 274Calibration cycle time

CH 274Calibration cycle time

“Maintenance” Reset of cal. cy cle timer
5sec.

CLOSE
( In “NO” setting )
CH 271

Conact output

“Air”

“Air or Purge”

CLOSE
( In “NO” setting )

loading cal. value

CH 272

CH 273

5sec.

CLOSE
( In “NO” setting )
CH 271

CLOSE
( In “NO” setting )

CH 272

“Zero”

CLOSE loading cal. value
( In “NO” setting )

CLOSE
( In “NO” setting )

Display

Display 2 – H – CAA CA 1 fEC – H – CAA CA 1

Fig.5-4 Auto Calibration (Air-point & Zero-point) Timing Chart

5.4.7 Calibration Stop Operation
During Semi Auto Calibration or Auto Calibration, you can stop calibration by key operation on the receiver. If “2” is set in “CH No.278” while the calibration gas is being supplied, the calibration gas is stopped and the display returns to normal after the Recovery-time. The calibration data of the calibration points that have already been loaded will be reflected in the measured values. Calibration data are loaded approximately 5 seconds before the end of calibration gas supply.

CH No.

Contents

278 Start / Stop of Semi Auto Cal.

Default 0

Remarks
1Start Calibration 2Stop Calibration

35

Purging

6 Purging

6.1 About Purging
Purging is the process of blowing compressed air to remove dust and other foreign matter (hereafter referred to as dust) that has entered or deposited in the measurement gas flow path around the oxygen sensor or inside the sampling probe. If dust is contained in the sample gas, purge as necessary to reduce dust accumulation and clogging. In addition, continuous purging prevents dust from entering the measurement gas flow path, so it is possible to protect the oxygen sensor and measurement gas flow path during periods when measurements are not performed. This product can configure a system that performs semi-automatic or automatic purging by opening and closing the solenoid valve installed in the flow path of compressed air for purging with contact output.

6.2 Purge Method Type

This product can purge by the following two methods. When changing the purge method, it is necessary to change the CH data setting. Normally, we set the purge method according to the customer’s usage at the time of shipment.

Purge method

CH 280 setting

Table 6-1 Purge Method Type Contents

Required plumbing system

No use

0

No use the purging

Semi Auto Purge
Auto Purge

A purge method in which the “purge start

operation” is performed by key operation

1

of the receiver or contact input.

The operation of feeding and stopping the Plumbing system that can open

purge air are performed automatically.

and close solenoid valves with

A purge method in which the “purge start receiver contact output.

operation” is performed automatically

e.g.) 3.1.2 System Configuration

using the receiver’s internal timer.

Example 2

The operation of feeding and stopping the

2

purge air are performed automatically.

3.1.3 System Configuration Example 3

Purge start operation by receiver key operation and contact input can be used together.

36

6 Purging

6.3 Semi Auto Purge / Auto Purge

6.3.1 Semi Auto Purge / Auto Purge – advance setting –

When performing purge by Semi Auto or Auto, set the table 6-2 in advance. After setting it once, it is necessary to reset only when the set value is changed.

CH No.

Table 6-2 CH Setting for Semi Auto Purge or Auto Purge

Contents

Set value

Remarks

209 Using contact input for calibration

0 or 1

0No use 1Use

280 Purge method type

1 or 2

0No use 1Semi Auto purge 2Auto purge

282 Purge ON time [sec] 283 Purge OFF time [sec] 284 Purge ON/OFF repetition count

Any setting (Default10)
Any setting (Default10)
Any setting (Default5)

One purge ON time Setting range1 to 99
Time between purge ON and purge ON Setting range1 to 99
Number of repetitions when “purge ON to OFF” is set to 1 time Setting range1 to 99

285 Purge recovery-time [min] *7

Any setting (Default3)

Setting range1 to 99

286

Purge cycle time [XX day ­ XX hour] *8

Any setting (Default01-00)

Setting range[00-01] to [99-23]

287

Purge delay time [XX day ­ XX hour] *8

Any setting (Default00-02)

Setting range[00-01] to [99-23] Time until first purge starting after power ON or timer reset

7 Purge recovery-timeThe grace period until the purge air returns to the sample gas after the purge air supply is stopped. The contact output “Maintenance” is closing until the purge recovery-time ends.
8 Setting required only when using Auto Purge

Note
For CH282, 283, 284, 285, 286, and 287, the customer should change the setting values according to the actual use such as dust adhesion.

37

6 Purging

6.3.2 Semi Auto Purge – Key Operation or Contact Input –
Semi Auto Purge is performed in the following procedure.
1. Set CH setting
Set “Table 6-2”. Omit it When it has already been set.
2. Purge start operation
(Set “1” in “CH No.281” or Contact input close)
Using key on receiverset “1” in “CH No.281” Using contact inputCorresponding contact input terminals
close for 0.1 seconds or more. For detail of contact input see “7.3 Contact Input”.
3. (Check the purge starting)
When purge starts, display 2 switches to “Pon”. (Switching is about 5 seconds after purge start operation)
After that, “Pon” “PoF” is repeated for the set number of times. After the completion of purge, it moves to the purge recovery-time
(Display 2 shows “rEC”) After the purge recovery-time ends, the display returns to normal.
Tip
CH No.281 is a switch that starts/stops Semi-Auto Purge. Start purgeset “1” Stop purgeset “2”

Indication during purge using key on receiver
1
28 1
ENT Purge start operation
20.92
P 6 H
Repeated for the set number of CH No.284
20.92
P 6 F
Move to Purge recovery-time
15.60
F E C
Purge complete Return to normal display
12.60
– H –

Operation of contact input/output, display, and analog output during Semi Auto Purge are as shown in

“Fig. 6-1 Semi Auto Purge Timing Chart”.

Items

Contact in/output DisplayAnalog output

Purge Start Operation

Key operation Contact input

ON

Set “1” in “CH No.281” or Contact input ON

about 0.1sec.

Conact output Display

“Maintenance”
“Purge” “Air or Purge”
Display 2 Display 1

CLOSE

1st

2nd

5 CH 282

CH 283

Purge ON time Purge OFF time

CH 282

CH 283

Purge ON/OFF times N times (CH 284)
CH 282 CH 283

CH 285 Purge Recoverytime

CLOSE OPEN CLOSE OPEN

CLOSE OPEN

p6h p6F p6h p6F p p6F p6h p6F fEC

Measurem ent value

Analog output 4-20A
By out hold setting

No hold
Setting value (CH 024) Value 5 sec. before
event occurrence (Defalut)

Meas urem ent v alue
Meas urem ent v alue

Measurem ent value Output hold (s etting value) Output hold (just before s tarting purge)

Fig.6-1 Semi Auto Purge Timing Chart

Meas urem ent v alue
Meas urem ent v alue

38

6 Purging

6.3.3 Auto Purge
Auto Purge is performed in the following procedure.
1. Set CH setting
Set ” Table 6-2 ” Omit it When it has already been set.
2. Reset purge cycle
Turn OFF the power switch of receiver and turn ON the power again. Or set “1” in “CH No.314”. Purge is started automatically after “Purge delay time”. After the first purge is started, the purge is automatically started at the cycle set in “purge cycle time”.

Refer
Detail for operation of purge cycle reset see “6.4 Reset of Cycle Timer and Countdown”.

Operation of contact output during Auto Purge is as shown in “Fig. 6-2 Auto Purge Timing Chart”. Operation of display 1, display 2 and analog output during purge are same as Semi Auto Purge.

Items

Purge cycle

Purge cycle timer

CH287Purg e delay time

Contact output
CH 286Purge cycle time

CH 286Purge cycle time

“Maintenance” Reset of cal. cy cle timer

Conact output

5sec. 1st

CLOSE 2nd

N times

Purge Recov ery -time

“Purge” “Air or Purge”

CLOSE OPEN CLOSE OPEN CLOSE

OPEN CLOSE OPEN

Fig.6-2 Auto Purge Timing Chart

5sec.

CLOSE

1st

2nd

CLOSE OPEN CLOSE OPEN

6.3.4 Purge Stop Operation

During Purge, you can stop purge by key operation on the receiver. If “2” is set in “CH No.281” during purge, purge operation is stopped and the display returns to normal after the Purge Recovery-time.

CH No.

Contents

281 Start / Stop of Purge

Default 0

Remarks
1Start Purge 2Stop Purge

39

6 Purging

6.4 Reset of Cycle Timer and Countdown

6.4.1 Reset Method of Cycle Timer

Resetting the Auto Calibration and Auto Purge cycle timer restarts the delay time count. Table 6-3 shows how to reset the cycle timer.

Timer reset method

Table 6-3 Reset Method of Cycle Timer

Detailed procedure

Reset target

Operation after reset

Setting CH No.314

Set “1” in CH No.314

Calibration delay time Calibration cycle time
Purge delay time Purge cycle time

Recount starts from calibration delay time and purge delay time

Power restart

Turn OFF the power of the receiver and turn it ON after a few seconds.

Calibration delay time Calibration cycle time
Purge delay time Purge cycle time

After Temp. Rising, recount starts from calibration delay time and purge delay time

Setting CH No.270

After setting CH No.270 to something other than “2” once, set it to “2” again.

Calibration delay time Recount starts from calibration Calibration cycle time delay time

Setting CH No.280

After setting CH No.280 to something other than “2” once, set it to “2” again.

Purge delay time Purge cycle time

Recount starts from purge delay time

Note
The delay time and cycle time of Auto Calibration and Auto Purge are not applied immediately even if the set value is changed without resetting the cycle timer. Reset the cycle timer as necessary. In addition, if the reset operation is not performed after changing the set value, the count will start with the changed set value after the currently counting timer finishes counting. Example) – During cycle time counting : After the count of cycle time with the old setting value ends, the count of cycle time starts with the new setting value. – During delay time counting : After counting with the delay time of the old setting, counting starts with the cycle time of the new setting
During error detection, the reset operation is disabled. Recover from the error and perform the reset operation.

40

6 Purging

6.4.2 Cycle Timer Countdown
The time until the next Auto Calibration and Auto Purge indicate in “CH No.288” and “CH No.289”.

CH No.

Contents

288 Next Auto Calibration countdown

289 Next Auto Purge countdown

Data XX-XX XX-XX

Remarks
Indicate the time until the next Auto Calibration starts as “XX day – XX hour” Indicate the time until the next Auto Purge starts as “XX day – XX hour”

Tip
By checking this CH data after the timer reset operation, it is possible to check whether the reset operation has been completed. When the reset operation is completed, this CH data has changed to the delay time setting value.

41

Other Settings/Functions

7 Other Settings/Functions

7.1 Analog Output

7.1.1 Analog Output Specifications

Output number One

Item

O2 concentration

Output

DC 4 to 20 mA Isolated output Load resistance 550 or less

7.1.2 Setting, Checking and Switching of Output Range

Two ranges (HIGH range and LOW range) can be used for the output range corresponding to 4 to 20 mA DC. The setting and checking of the output range are performed with “CH No.30, 31, 34, 35”.

CH No. 030
031 034 035

Contents
HIGH range span concentration LOW range span concentration
HIGH range unit
LOW range unit

Default 25
10 2 2

Remarks
Setting range1 to 99999 Default0 to 25 vol.%
Setting range1 to 99999 Default0 to 10 vol.%
0No use 1vol.ppm 2vol.%

There are two ways to switch the output range. Key operation on receiver Contact input

Switching the output range by key operation of receiver Select the output range to be used in “CH No.20”. In the case of “output range switching state by contact input”, range selection by “CH No.20” is invalid.

CH No.

Contents

020

Output range selecting

Default 1

Remarks
1HIGH range 2LOW range

Switching the output range by contact input If “3” is set to any of CH No.227~229, it be output range switching state by contact input.

CH No.

Contents

227

Contact input 1 (IN1)

228

Contact input 2 (IN2)

229

Contact input 3 (IN3)

Conditions under which output range switching is possible using contact input

Remarks

Set “3” to either

COMCorresponding contact * ON LOW range OFFHIGH range

  • The corresponding contact is the contact input terminal corresponding to the CH No. set to “3”.

42

7 Other Settings/Functions

7.1.3 Output Hold Function

This function holds the analog output at a constant value when the sample gas is not being measured, such as during Temp. Rising, Calibration, Purging, Error occurrence, or Sleep Mode. The output hold operation is set by “CH No.23” and “CH No.24”. Alternatively, You can hold the analog output by key operation or contact input regardless of the receiver status using “CH No. 029”.

CH No.

Contents

023

Output hold setting

024

Output hold value setting(%FS)

029

Manual output holding

Default 2
100 0

Remarks
0No hold 1Setting value (CH 024) 2Value 5 seconds before event
occurrence
0OFF 1Hold ON 2Using contact input

Operating states
Temp. Rising
During Calibration
During Purge ON/OFF
During Recovery Time
During Error Detecting
During Sleep Mode
During Concentration Upper Limit Alarm/ Lower
Limit Alarm During
Manual Output Holding

Table 7-1 Operation of Output Hold by States

Setting value of CH No.023

0No hold

1Setting value (CH 024)

2Value 5 seconds
before event occurrence (Default)

Setting value (CH 024)

Setting value (CH 024)

No hold

Setting value (CH 024)

Value 5 seconds before event occurrence

Value 5 seconds before event occurrence

No hold

No hold

No hold

Value 5 seconds before event occurrence

Setting value (CH 024)

Value 5 seconds before event occurrence

“Output hold value setting [%FS]” of “CH No.024” sets the output in %FS.
e.g. CH No.024 setting is “0” Output 4 mA CH No.024 setting is “50” Output 12 mA CH No.024 setting is “100” Output 20 mA

The output hold status by error detection is released when the error is restored.

43

7 Other Settings/Functions

7.1.4 Simulated Output

This is a function to output an arbitrary value for loop check of analog output. Set the output value to be output in “CH No.225”. And switch the simulated output stop/output in “CH No.224″.

CH No. 224 225

Contents
Simulated output
Simulated output value setting [%FS]

Default 0 0

0Stop 1Output

Remarks

Setting range0.0 to 100.0

” Simulated output value setting [%FS]” of “CH No.225” sets the output in %FS.
e.g. CH No.225 setting is “0” Output 4 mA CH No.225 setting is “50” Output 12 mA CH No.225 setting is “100” Output 20 mA

7.1.5 Adjustment of Analog Output
When there is an error between the analog output value of the receiver side and the input value of the customer side device (recorder, control device), the analog output adjustment function at “CH No.027” of the receiver can reduce the error.
1. Connect your device
Connect your device (recorder, control equipment, ammeter, etc.), make sure that the input current of the customer’s device can be confirmed.
2. Select CH No. 027 (Output Adjustment)
Select “CH No. 027”. Display 1 displays “cAL 1”.

3. Adjust Zero output (4 mA)

When the ENT key is pressed, move to the output adjustment status

and “0” is displayed on the Display 1.

When the

key is pressed, the Display 1 will change to “100”.

When “0” is displayed, it is the zero side (4 mA) adjustment state,

and when “100” is displayed, it is the span side (20 mA) adjustment state.

Adjust the zero side (4 mA) using the

and

keys

while “0” is displayed on the Display 1.

KeyIncrease the output.

KeyReduce the output.

4. Adjust Span output (20 mA)

Press

key to switch the Display 1 to “100”.

Adjust the span side (20 mA) using the

and

keys

while “100” is displayed on the Display 1.

5. Finish the adjustment
After the adjustment, press ENT key to return to the CH No. selection state.

44

7 Other Settings/Functions

7.2 Contact Output

7.2.1 Contact Outputs Specifications

Output number 4 points

Output

No-voltage relay contact capacity AC 250 V 1A Resistance load DC 30 V 1A Resistance load

Terminal

RY1,RY2,RY3,RY4A,RY4B

7.2.2 Contact Outputs Items and Setting

For contact output, set the output item of each contact in “CH No.200 to 203”, Also, set the operation of each contact “normally open: NO/normally closed: NC” in “CH No.204 to 207”. See Table 7-2 and Table 7-3 for contact output details and operation.

CH No. 200 201 202

Contents Contact output item RY1 Contact output item RY2 Contact output item RY3

203

Contact output item RY4A, RY4B

Default
(1 or 16)

Remarks
0No use 1Error (Analyzer error) 2Range H 3—-(No use at this product) 4Temp. rising 5Alarm H
(Concentration upper limit alarm) 6Alarm L
(Concentration lower limit alarm) 7—-(No use at this product) 8Ready 9Maintenance 10Air 11Span 12Zero 13Purge 14Air or Purge 15Alarm H&L 16Error or Alarm H&L

204

Contact output operation RY1

205

Contact output operation RY2

206

Contact output operation RY3

207

Contact output operation RY4A, RY4B

0

0 0NO

0

1NC

1

Note
RY4A and RY4B are dedicated terminals for “Error” or “Error or Alarm H&L”. Set “1” or “16” in “CH No.203”, and use “1” in “CH No.207”.

Refer
When using “Alarm H”, “Alarm L”, or “Alarm H&L”, It is also necessary to set “CH No.220 to 223”. For details, see “7.5.1 Concentration upper/lower limit alarm”.

45

7 Other Settings/Functions

Table 7-2 Details of Contact Output Contents and Operationfor RY1 – RY3

Selection

number in “CH 200 to

Item

202”

Contents (purpose)

Contact operation

Receiver state

CH 204 – 206 setting 0NO 1NC

0

No use

OPEN CLOSE

Outputs a contact according to

the output range in use.

HIGH range in use

CLOSE OPEN

2

Range H

(If set to NC, the contact will

be CLOSE while using the LOW range.)

LOW range in use

OPEN CLOSE

Temp. rising

OPEN OPEN

5

Alarm H (Concentration

Judges whether the measured Measured value is higher

value is “higher than the set

than set value

CLOSE OPEN

upper limit alarm) value/or less”.

Measured value is set value or less

OPEN CLOSE

Temp. rising

OPEN OPEN

6

Alarm L

Judges whether the measured Measured value is set value

(Concentration lower value is “lower than the set

or more

OPEN CLOSE

limit alarm)

value/or more”.

Measured value is lower than set value

CLOSE OPEN

Power OFF

OPEN OPEN

Temp. rising

OPEN CLOSE

During error detecting

OPEN CLOSE

8

Ready

Judges “during normal measurement / otherwise”.

During calibration During purge

OPEN OPEN

CLOSE CLOSE

During recovery-time

OPEN CLOSE

During sleep mode

OPEN CLOSE

During measurements other than the above

CLOSE

OPEN

Power OFF

OPEN OPEN

Temp. rising

CLOSE OPEN

During error detecting

OPEN CLOSE

9

Maintenance

Judges “during maintenance / otherwise”.

During calibration During purge During recovery-time

CLOSE CLOSE CLOSE

OPEN OPEN OPEN

During sleep mode

CLOSE OPEN

During measurements other than the above

OPEN CLOSE

10

Air

Used to open/close the solenoid valve for Air-point calibration.

When opening the solenoid valve for Air

CLOSE

OPEN

11

Span

Used to open/close the solenoid valve for Span-point calibration.

When opening the solenoid valve for Span gas

CLOSE

OPEN

12

Zero

Used to open/close the solenoid valve for Zero-point calibration.

When opening the solenoid valve for Zero gas

CLOSE OPEN

13

Purge

Used to open/close the solenoid valve for Purge.

When opening the solenoid valve for Purge

CLOSE

OPEN

14

Air or Purge

Used when sharing the solenoid valve for Air-point calibration and Purge.

When opening the solenoid valve for Air or Purge

CLOSE

OPEN

Power OFF

OPEN OPEN

Measured value is higher than upper limit setting

CLOSE OPEN

“Applicable/not applicable to

The measured value is

15

Alarm H&L

either Alarm H or Alarm L” to judge.

between the upper limit setting value and the lower

OPEN CLOSE

limit setting value.

Measured value is lower than lower limit setting

CLOSE OPEN

46

7 Other Settings/Functions

Table 7-3 Details of Contact Output Contents and Operationfor RY4A, RY4B

Selection number in “CH 203”
1
16

Item

Contents (purpose)

Error (Analyzer error)

Detects power failure and error occurrence.

Error or Alarm H&L

“Applicable/not applicable to either Error or Alarm H&L” to judge.

Contact operation

Receiver state

RY4A

Power OFF Temp. rising During measuring During error detecting Power OFF Temp. rising During measuring During error detecting During Alarm H&L detecting

OPEN CLOSE CLOSE OPEN OPEN CLOSE CLOSE OPEN
OPEN

RY4B
CLOSE OPEN OPEN CLOSE CLOSE OPEN OPEN CLOSE
CLOSE

47

7 Other Settings/Functions

7.3 Contact Input

7.3.1 Contact Inputs Specifications

Input number 3 points

Input

Input no-voltage contact Operating current 5 mA contacts close for 0.1 seconds or more (DC 24 V is applied between these terminal and COM)

Terminals

(IN1, IN2, IN3) COM

Note
Be sure Input no-voltage contact. Connecting a contact with voltage may cause a malfunction.
Do not use the input contact in parallel connection or series connection with other devices. (Figure 7-1) It may damage the receiver or other equipment.

IN1 IN2
DTF-102
COM

IN1 IN2
DTF-102
COM

Other device

[ Correct connection ] [ Incorrect connection ] Input contacts cannot be connected in parallel with other devices

Fig.7-1 Wiring Example for Contact Input

7.3.2 Contact Inputs Items

For contact input, set the input items for each contact with “CH No.227 to 229”. See Table 7-4 for detail and actions of contact input.

CH No.

Contents

227

Contact input item IN1

228

Contact input item IN2

229

Contact input item IN3

Default

Remarks
0No use 1Calibration start 2Purge start 3Output range switching 4Manual output hold 5Sleep mode

CH 227 to 229
1

Table 7-4 Contact Input Contents and Actions

Contents

Detail

Calibration start Contact CLOSE : Semi Auto Calibration start

CH No. and setting values that need to be set together
CH No.208= “1”

2

Purge start

Contact CLOSE : Semi Auto Purge start

3

Output range switching

Used to switch the output range. Contact OPEN : HIGH range Contact CLOSE : LOW range

4

Manual output hold

Contact CLOSE : Analog output is held The concentration value on display 1 is not held.

5

Sleep mode

Contact CLOSE : Switch to sleep mode

CH No.209= “1”
CH No.29= “2” CH No.197= “2”

48

7 Other Settings/Functions

7.4 Sleep Mode

7.4.1 About Sleep Mode

Sleep mode is a function that protects the oxygen sensor and measurement gas flow path from exhaust gases by continuously supplying purge air while the oxygen sensor is not driven. There are cases that non-driving oxygen sensor, measurement gas flow path and transmitter are exposed to exhaust gas and residual gas in the furnace, when combustion equipment is stopped for a long period of time due to installation, test, inspection or repair. While the sensor is in driving, the zirconia oxygen sensor is durable against dirty gases such as exhaust gas. However, when the sensor is not driven (low temperature), If exposed to dirty gas such as exhaust gas or residual gas in the furnace, the sensor will deteriorate in a short period of time and may not be able to measure. In addition, dust accumulation and clogging may occur in the sample gas flow path. In sleep mode, the oxygen sensor is not driven, and purge air is continuously supplied to protect the oxygen sensor and measurement gas flow path from exhaust gas. When performing continuous purging automatically using the sleep mode, a system is required to operate the solenoid valve as shown in “3.1.2 System Configuration Example 2” or “3.1.3 System Configuration Example 3”.

7.4.2 Switch to Sleep Mode and Operation

There are two ways to switch to “Sleep mode” : “receiver key operation” and “contact input”. It set in “CH No.197”.

CH No.

Contents

197

Sleep mode

Default 0

Remarks
0No use 1Sleep mode ON 2Using contact input

The operation when changing to sleep mode and returning is as shown in “Fig. 7-2 Timing Chart in Sleep Mode”.

Sleep mode operation

Items

Key operation Contact input

Contact in/output DisplayAnalog output
CHNo.197 = “1” or CH No. 197 = “2” & contact input ON
about 0.1sec.

Contact output DIsplay

“Maintenance” “Purge”
“Air or Purge”
Display 2
Display 1 Measurement v alue

CLOSE
( In “NO” setting )

CLOSE
( In “NO” setting )

Temp. rising (3 minutes)

SLP

—–

2.59, 2.58, … Measurement v alue

Analog output 4-20A
By out hold setting

No hold Measurement v alue

Setting value (CH 024)
Value 5 sec. before event occurrence (Defalut)

Measurement v alue Measurement v alue

Output hold (just before starting sleep mode)
Output hold (setting value) Output hold (just before starting sleep mode)

Fig. 7-2 Timing Chart in Sleep Mode

Output hold (setting value)
Output hold (setting value)

Measurement v alue Measurement v alue Measurement v alue

During sleep mode, the contact output “Purge” and “Air or Purge” are CLOSE, and display 2 changes to “SLP”. Also, the analog output is held.
When the sleep mode is canceled, it always starts with 3 minutes of Temp. rising.

49

7 Other Settings/Functions

7.5 Other Functions

7.5.1 Concentration Upper / Lower Alarm (Alarm H, Alarm L, Alarm H&L)

This function judges whether the measured value is “higher than the set value” or “lower than the set value” and outputs an alarm or contacts. When using this function, set “CH No.220 to 223”. Also, when outputting with contact output, setting of contact output is also required.

CH No. 220 221 222 223

Contents
Concentration upper limit alarm setting value Concentration lower limit alarm setting value Concentration upper limit value unit Concentration lower limit value unit

Default 90 10 0 0

Remarks
Set about upper limit value, lower limit value Setting range0.0 to 9990.0
0No use 1vol.ppm 2vol.%

20.92
A H
[ Concentration upper limit alarm detected ] “AH” is displayed on display 2

2.00
A L
[ Concentration lower limit alarm detected ] “AL” is displayed on display 2

Fig. 7-3 Display Example of Concentration Upper Limit Alarm and Concentration Lower Limit Alarm

7.5.2 Primary Delay Calculation Function

This function calculates primary delay of measurement value, to display and analog output. With the value (seconds) set in “CH No.016”, primary delay calculation processing for display and output is performed.

CH No.

Contents

016

Primary delay time [sec]

Default 0

Remarks
Unitsec.( 063%) Setting range0 to 99

7.5.3 Moving Average Function

This function calculates moving average of measurement value, to display and analog output. With the value (seconds) set in “CH No.017”, moving average processing for display and output is performed.

CH No.

Contents

017

Moving average time [sec]

Default 0

Remarks
Unitsec. Setting range0 to 30

7.5.4 Protection Function of Set Value

This function prohibits changing CH data by key operation. Whether or not you can change CH data is set in “CH No. 310”.

CH No.

Contents

310

Data setting change password

Default 102

Remarks
0Cannot be changed 102Changeable

50

7 Other Settings/Functions

7.6 CH Data Table

Table 7-5 CH Data Table1/3

CH No.

Contents

Setting range

000 Normal display setting 001 Sensor Vs monitor [mV]

0—-1Oxygen concentration

002 Sensor Ip1 monitor [mA]

003 Sensor Ip2 monitor [µA]

004 Sensor Vp monitor [V]

005 Sensor heater voltage monitor [V]

006 Sensor heater current monitor [A]

010 Measuring mode setting

2Normally measuring mode [] 3Reducing atmosphere
measurement mode [-%]

016 Primary delay time [sec]

0 to 99

017 Moving average time [sec] 020 Output range selecting
023 Output hold setting 024 Output hold value setting [%FS]

0 to 30
1HIGH range 2LOW range 0No hold 1Setting value (CH 024) 2Value 5 seconds before event
occurrence
0 to 100

027 Output adjustment

Adjustment for DC 4 to 20 mA

029 Manual output holding

0OFF 1Hold ON 2Using contact input

030 HIGH range span concentration

0 to 99999

031 LOW range span concentration

034 HIGH range unit

035 LOW range unit

120

Zero-point calibration concentration [vol.%]

121

Span-point calibration concentration [vol.%]

122

Minus span-point calibration concentration [vol.%]

123

Air-point calibration concentration [vol.%]

0 to 99999 0No use 1vol.ppm 2vol.% 0.00 to 99.90
0.00 to 99.90
-99.90 to 99.90
0.00 to 99.90

143 Linearizer No.

0 to 3

180

Calibration point selection by Manual Calibration

181 Calibration by Manual Calibration

5Zero-point 6Span-point 7Minus span-point 8Air-point
0OFF 1Perform calibration

e.g. Default 1

Remarks

Monitor

2

,

0 0 1

2
100 –
0
25 10 2 2 5.00 15.00 -5.00 20.60 See inspection report
8

0

– CH No. marked with “” in the Remarks has different default values depending on the part number of the receiver.
– CH No. marked with “” in the Remarks cannot be changed by the customer. Please contact us if you would like to make any changes.
– CH numbers not listed in the “CH Data Table” are manufacturer setting data. Do not change it by yourself.

51

7 Other Settings/Functions

Table 7-6 CH Data Table2/3

CH No.

Contents

190 Heater control mode

191 Heater voltage setting [V] 192 Heater normal temp resistance [] 193 Heater reference resistance ratio

197 Sleep mode

Setting rangeRemarks 0Heater OFF 1Constant voltage control 2Constant resistance control 1 3Constant resistance control 2 5.00 to 11.00
0.01 to 99.90
1.00 to 99.90
0No use 1Sleep mode ON 2Using contact input

e.g. Default Remarks

TF-103 TF-2

,

10.50

See inspection report

TF-102.80 TF-2.70

0

200 Contact output item RY1 201 Contact output item RY2 202 Contact output item RY3
203 Contact output item RY4A, RY4B

0No use

10

1Error (Analyzer error) 2Range H

13

3—- (No use at this product) 4Temp. rising

9

5Alarm H

(Concentration upper limit alarm)

6Alarm L

(Concentration lower limit alarm)

7—- (No use at this product)

8Ready

9Maintenance 10Air

1

11Span

12Zero

13Purge

14Air or Purge

15Alarm H&L

16Error or Alarm H&L

204 Contact output operation RY1

0

205 Contact output operation RY2 206 Contact output operation RY3

0NO 1NC

0

0

207 Contact output operation RY4A, RY4B

208 Using contact input for calibration 209 Using contact input for calibration

0No use 1Use
0No use 1Use

220 Concentration upper limit alarm setting value 0.0 to 9990.0

221 Concentration lower limit alarm setting value 0.0 to 9990.0

1

1

1

90 10

222 Concentration upper limit value unit 223 Concentration lower limit value unit 224 Simulated output 225 Simulated output value setting [%FS] 227 Contact input item IN1 228 Contact input item IN2 229 Contact input item IN3

0No use 1vol.ppm 2vol.%
0Stop 1Output
0.0 to 100.0
0No use 1Calibration start 2Purge start 3Output range switching 4Manual output hold 5Sleep mode

0

0

0

0

1

2

3

– CH No. marked with “” in the Remarks has different default values depending on the part number of the receiver.
– CH No. marked with “” in the Remarks cannot be changed by the customer. Please contact us if you would like to make any changes.
– CH numbers not listed in the “CH Data Table” are manufacturer setting data. Do not change it by yourself.

52

7 Other Settings/Functions

Table 7-7 CH Data Table3/3

CH No.

Contents

Setting rangeRemarks

270 Calibration method type

0Manual Calibration 1Semi Auto Calibration 2Auto Calibration

271 Air-point gas sending time [min]

1 to 99

272 Zero-point and Span-point gas sending time [min] 1 to 99

273 Recovery-time [min]

1 to 99

274 Calibration cycle time [XX day – XX hour] [00-01] to [99-23]

275 Calibration delay time [XX day – XX hour] [00-01] to [99-23]

276 Calibration point setting used for calculation

0Air 1Air and Zero 2Air and Span 3Air and Zero and Span 4Span 5Zero and Span

277 Span calibration value polarity

0Puls (+) 1Minus(-)

278 Start / Stop of Semi Auto Cal.

279

Calibration point selection by Semi Auto Cal. or Auto Cal.

280 Purge method type

0OFF 1Start Calibration 2Stop Calibration
0Air 1Air and Zero 2Air and Span 3Span
0No use 1Semi Auto purge 2Auto purge

281 Start / Stop of Purge

0OFF 1Start Purge 2Stop Purge

282 Purge ON time [sec]

1 to 99

283 Purge OFF time [sec]

1 to 99

284 Purge ON/OFF repetition count

1 to 99

285 Purge recovery-time [min]

1 to 99

286 Purge cycle time [XX day – XX hour] [00-01] to [99-23]

287 Purge delay time [XX day – XX hour] [00-01] to [99-23]

288 Next Auto Calibration countdown

289 Next Auto Purge countdown

300 HIGH range minus span concentration [%]

-99 to 99

301 LOW range minus span concentration [%]

-99 to 99

310 Data setting change password

0Cannot be changed 102Changeable

314 Timer reset

0OFF 1Timer reset

e.g. Default Remarks

2

5 5 3 30-00 00-07

0

0

,

0

0

2

0
10 10 5 3 01-00 00-02 Monitor Monitor 0 0 102
0

, ,

– CH No. marked with “” in the Remarks has different default values depending on the part number of the receiver.
– CH No. marked with “” in the Remarks cannot be changed by the customer. Please contact us if you would like to make any changes.
– CH numbers not listed in the “CH Data Table” are manufacturer setting data. Do not change it by yourself.

53

Maintenance

8 Maintenance

The following maintenance and inspection procedures are important in order to maintain normal functioning and accurate measurement. Make sure you thoroughly understand the procedure before performing maintenance.

When “Maintenance work”

WARNING

Fire, Electric shocks, Malfunction
Be sure to turn OFF the main power supply before performing wiring connection work or wiring check.
Before turning ON the power, make sure that the power wiring is properly connected.
If the fuse blows, check the cause and replace it with one of the same capacity and type. Also, when replacing the fuse, be sure to turn OFF the main power supply before starting work.
Be sure to use parts specified by our company for replacement parts and consumables.

Note
Be sure to check the safety precautions for the oxygen sensor (transmitter) to be used before starting to maintenance it.
Do not disassemble the parts other than those specified in this manual. Do not modify by the customer.

8.1 Daily / Periodic Inspection

Daily/periodic inspection items is nothing for this product (receiver). Below is a list of standard daily and periodic inspections for the oxygen sensor. Check the instruction manual for the oxygen sensor (transmitter) you are using for the actual details of daily and periodic inspections.
Table 8-1 Standard Daily / Periodic Inspection Items for Sensors (Reference) Recommended cycle 1month

Air-point Calibration
Replacement of Oxygen Sensor

Method

See “5 Calibration”.

Others

Periodically check that the calibration gas is flowing at the set flow rate during calibration.

Recommended period 2 years

Method

Refer to the instruction manual of the oxygen sensor (transmitter) used.

8.2 Consumable Parts and Spare Parts

There are no consumable parts for this product (receiver). The spare parts for this product (receiver) are as shown in the table below.

Table 8-2 Spare Part Table

Name

No./Spec.

Q’ty/1 unit

Replacement

Fuse

MF-51NR2A

1

When fuse blowout occurs

54

9 Troubleshooting

9 Troubleshooting

9.1 Phenomena and Countermeasures

Phenomena

Table 9-1 Phenomena and Countermeasures

Cause

Countermeasures

Unable to change data.

CH No.310 is entered with something other than “102”
Receiver problem

During output hold state (Display value is changing)

Analyzer output, display value does not change.

Gas pipe leak During An analyzer error is occurring

Receiver problem

Set “102” in CH No.310.

Replace the receiver or repair by Maker.

Reconfirm receiver status and settings. The output value is held according to the setting during “Temp. rising”, “Calibration”, “Purging”, “Analyzer error”, “Sleep mode”, and “Manual output hold”.

Check for leaks, tighten pipe fittings.

– Turn OFF the power once and turn it ON again after 10 seconds.

– Deal with the error code Refer to “9.2 Error Code Table”

– Press the

key to clear

Replace the receiver or repair by Maker.

Setting error

Analyzer output, display value error

Gas calibration error Gas flow rate is outside range. Poor wiring contact

Check the setting data related to analog output Refer to “7.1 Analog Output”.
– Check the setting data related to calibration – Perform Air-point calibration
Readjust gas flow rate.
Check for poor contact in the analog output wiring and sensor signal wiring.

Gas pipe leak

Check for leaks, tighten pipe fittings.

Sensor deterioration

Replace the sensor.

Setting error

Analyzer output, display value is zero

Gas calibration error Gas flow rate is outside range.

Check the setting data related to analog output Refer to “7.1 Analog Output”.
– Check the setting data related to calibration – Perform Air-point calibration
Readjust gas flow rate.

Flammable gas included in measuring gas

Eliminate flammable gas from measuring gas.

Sensor deterioration

Replace the sensor.

Analyzer output and display value do not match

Output adjustment is inaccurate Output range is different Receiver problem

Perform output adjustment in “CH No.027”.
Check output range setting (CH No.020, 030,031,034,035)
Replace the receiver or repair by Maker.

Filter clogged

Replace the filter (Only transmitters using filters)

Gas pipe leak

Check for leaks, tighten pipe fittings.

Calibration gas pipe clogged

Purging, cleaning and replacing in piping

Slow response

Condensation in the calibration gas pipe Sampling gas pipe and probe clogged

Purging, cleaning in piping and add installation of keep warming structure
Purging, cleaning and replacing in piping/probe

Insufficient sampling flow rate

Adjust sampling gas flow rate. (Only sampling type)

Primary delay time setting value is too large Check the data in CH No.016 and set “0”[sec].

Moving average time setting value is too large Check the data in CH No.017 and set “0”[sec].

Sensor deterioration

Replace the sensor.

55

9 Troubleshooting

9.2 Error Code Table

Error Code

Contents

Table 9-2 Error Code Cause

Countermeasures

E-01 E-02 E-03 E-04 E-05 E-06 E-07 E-20 E-08 E-09 E-10 E-11 E-12 E-13 E-14
E-21
E-35
E-36
E-38
E-45

Abnormality in ROM, RAM, and/or EEPROM

– Receiver abnormality
– Temporary malfunction due to external noise

Turn OFF power supply and turn ON after 10 seconds.

Heater current too over Heater current too under Heater voltage too over Heater voltage too under Heat up does not complete

– Sensor heater abnormality – Setting value abnormality – Receiver abnormality

– Press

key, or turn OFF power

supply and turn ON after 10 seconds.

– To clear the error display,

press

key after the cause of the

error has been eliminated.

Vs voltage too over

Vs voltage too under Ip current too over Ip current too under Vp voltage too over

– Sensor abnormality – Sensor wiring abnormality – Setting value abnormality – Receiver abnormality

– Press

key, or turn OFF power

supply and turn ON after 10 seconds.

– To clear the error display,

press

key after the cause of the

error has been eliminated.

Vp voltage too under

Heater terminals (S+, S-) not closed

Defective contact in wiring between S+ and S-

– Check the wiring.

– Press

key to clear the error display.

Data out of the specified range Data setting error

Calibration value out of range in Zero-point calibration

– Zero-point calibration abnormality
– Sensor abnormality

– Setting data within the specified range.

– Press

key to clear the error display.

Calibration value out of range in Span-point calibration
Calibration value out of range in Air-point calibration

– Span-point calibration abnormality
– Sensor abnormality
– Air-point calibration abnormality
– Sensor abnormality

Ip data out of range in N2 calibration

– N2 calibration abnormality – Sensor abnormality

– Press

key to clear the error display.

– Check the calibration gas concentration and setting value.

– Check the calibration gas flow rate.

– Perform calibration again

Replace the sensor if it does not recover.

E-46

Calibration-value error (Air point)

– Air-point calibration abnormality
– Sensor abnormality

E-60

“Not used” range is selected.

Data setting error about analog output range

Set “CH No.34,35” to other than “0”.

The analyzer needs to repair by Maker when all the actions above should fail.

56

9 Troubleshooting

9.2.1 Oxygen Sensor Failed
If “E-05” or “E-07” occurs during using (after a normal measurement period), the oxygen sensor may have failed due to its life. Follow the steps below to check the cause and restore the analyzer.
“E-05” or “E-07” occur When the error occurred ?

During use

No abnormalities No recovery

Immediately after installation, after sensor replacement, after wiring work.
[Check the wiring connection] Check if there is a connection failure or continuity failure between the oxygen sensor and the receiver. e.g.) – Wrong connection, disconnection,
looseness of wiring – Deterioration and damage of cables

[Check heater resistance] Remove the signal line of the oxygen sensor from the receiver, cool the sensor to room temperature, and check the resistance between “H+” and “H-“.
10 or more
CauseOxygen sensor life failure Replacement the oxygen sensor

2 to 5

[Check the wiring connection] Check if there is a connection failure or continuity failure between the oxygen sensor and the receiver. e.g.) – Wrong connection, disconnection,
looseness of wiring
– Deterioration and damage of cables

No abnormalities No recovery

Please contact us or our agency.

Fig. 9-1 Confirmation Procedure When Occurred “E-05” and “E-07”

Note

If receiver detecting multiple errors at the same time, pressing

key once will display

another error code. Press

key multiple times until the Temp. rising state is temporarily

entered, and check for recurrence of the error.

Timing of occurrence

– During use and sensor usage time is long (1 year or more) High possibility of oxygen sensor life failure.

– During use and sensor usage time is short Check the setting data in CH No.190,192,193.

– After installation, replacement the sensor, maintenance High possibility of poor contacts or incorrect connection.

If the resistance value between “H+” and “H-” is 5 < R <10, cool the sensor sufficiently and then check again.

57

Technical Data

10 Technical Data
10.1 Specifications

1. Power supply

AC 100 to 240 V±10% 50/60 Hz max.50 VA

2. External dimensions 300×250×156 [mm]

3. Weight

Approx. 6.5 kg

4. Measurement item

O2

5. Measurement range 0 to 25vol.%

6. Linearity *

±1.0F.S.F.S.0 to 25vol.% ±2.0F.S.F.S.0 to 10vol.%, 0 to 5vol.%

7. Repeatability *

±0.5F.S.F.S.0 to 25vol.%, 0 to 10vol.%, 0 to 5vol.%

8. Response time *

5 seconds or lessT90

9. Analog output

Item

: O2 concentration

Output

: DC 4 to 20 mA Isolated output Load resistance 550 or less

Output range : HIGH range 0 to 25vol.%Default

LOW range

0 to 10 vol.%Default

Range switching : Key operation/Contact input

Output hold

: No hold/Setting value/ Value 5 seconds before event occurrence

10. Contact output

Number Output
Output item

: 4 points : No-voltage relay contact
capacity AC 250 V 1 A Resistance load DC 30 V 1 A Resistance load
: Refer to “7.2.2 Contact Outputs Items and Setting”

11. Contact input

Number Input
Input item

: 3 points : Input no-voltage contact
Operating current 5 mA contacts close for 0.1 seconds or more : Refer to “7.3.2 Contact Inputs Items”

12. Installation Conditions

Installation location

: Indoor

Ambient temperature : -10 to +50

Ambient humidity

: 90RH or less(Not condensation)

Environmental protection : No waterproof structure

13. Other functions

“Manual/Semi-Auto/Auto calibration”, “Semi-Auto/Auto Purging”, “Abnormality selfdiagnosis”, “Manual output hold”, “Concentration upper/lower alarm”, “Analog output adjustment”, “Simulated output”, “Sleep mode”, “Primary delay calculation”, “Moving average”.

  • “Linearity”, “Repeatability” and “Response time” are reference values when using a TF-10 type probe transmitter (L=500). It depends on the oxygen sensor (transmitter) used.

58

MEMO

10 Technical Data

59

The contents of this manual are subject to change without notice for improvement.
For inquiries regarding product handling, please contact us or our distributors. Inquiry form URL https://www.energys.co.jp/english/inq/all.php ENERGY SUPPORT CORPORATION 1, Aza Kamikobarii, Inuyama, Aichi 484-8505 Japan

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals