SUTO S601 Stationary Compressed Air Purity Monitor Instruction Manual
- June 15, 2024
- SUTO
Table of Contents
- S601 Stationary Compressed Air Purity Monitor
- Safety Instructions
- Registered Trademarks
- Application
- Features
- Technical Data
- Dimensions
- Installation on Site
- Installation on Site
- Setup
- Operations and Configurations
- Software Installation
- Optional Accessories
- Maintenance
- Disposal or Waste
- Calibration
- Warranty
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Instruction and Operation Manual
S601 Stationary Compressed Air Purity Monitor
Dear Customer,
Thank you for choosing our product.
Please read the manual in full and carefully observe operating instructions
stated before you start up the device. The manufacturer cannot be held liable
for any damage that occurs as a result of nonobservance or non-compliance with
this manual.
Should the device be tampered with in any manner other than a procedure which
is described and specified in the manual, the warranty is void and the
manufacturer is exempt from liability.
The device is destined exclusively for the described application.
SUTO offers no guarantee for the suitability for any other purpose.
SUTO is also not liable for consequential damage resulting from the delivery,
capability or use of this device.
Safety Instructions
Please check if this instruction manual matches the product type.
Please observe all notes and instructions indicated in this manual. It
contains essential information which must be observed before and during
installation, operation and maintenance. Therefore this instruction manual
must be read carefully by the technician as well as by the responsible user /
qualified personnel.
This instruction manual must be available at the operation site of the flow
sensor at any time. In case of any obscurities or questions, regarding this
manual or the product, please contact the manufacturer.
WARNING!
Compressed air!
Any contact with quickly escaping air or bursting parts of the compressed air
system can lead to serious injuries or even death!
-
Do not exceed the maximum permitted pressure range. (See sensors label.)
-
Only use pressure tight installation material.
-
Avoid that persons get hit by escaping air or bursting parts of the instrument.
-
The system must be pressure less during maintenance work.
WARNING!
Voltage used for supply!
Any contact with energized parts of the product, may lead to an electrical shock which can lead to serious injuries or even death! -
Consider all regulations for electrical installations.
-
The system must be disconnected from any power supply during maintenance work.
Laser Safety Information
This product contains a laser-based counter that is Class 1 product (as
defined by 21 CFR, Subchapter J of the Health and Safety Act of 1968) when
used under normal operation and maintence. Service procedures on the counter
can result in exposure to invisible radiation. Service should be performed
only by factory-authorized personnel.
ATTENTION!
Permitted operating parameters!
Observe the permitted operating parameters. Any operation exceeding this
parameters can lead to malfunctions and may lead to damage on the instrument
or the system.
- Do not exceed the permitted operating parameters.
- Make sure the product is operated in its permitted limitations.
- Do not exceed or undercut the permitted storage and operating temperature and pressure.
The product must be maintained and calibrated frequently, at least annually.
General safety instructions
- It is not allowed to use the product in explosive areas.
- Please observe the national regulations before/during installation and operation.
Remarks
- It is not allowed to disassemble the product.
- Always check the compressed air connectors in terms of stability and tightness.
ATTENTION!
Measurement values can be affected by malfunction!
The product must be installed properly and frequently maintained. Otherwise it
may lead to wrong measurement values, which can lead to wrong results.
- Always observe the direction of the flow when installing the sensor. The direction is indicated on the housing.
- Do not exceed the maximum operating temperature at the sensors tip.
- Avoid condensation on the sensor element as this will affect the accuracy enormously.
Storage and transportation
- Make sure that the transportation temperature is between -10 … +70°C.
- For transportation it is recommended to use the packaging which comes with the sensor.
- Please make sure that the storage temperature of the sensor is between -10 … +50°C.
- Avoid direct UV and solar radiation during storage.
- For the storage the humidity must be <90%, no condensation.
ATTENTION!
Equipment may get damaged!
Please make sure, that your measuring point is free of excessive
contamination/dirt. This should be maintained before every measurement.
- Observe the measuring point always before measurement if it is free of contamination like water drops, oil drops or other rough contamination.
- Should water hit the inner electronics, the senors could be seriously damaged.
- Check your measurement point with the enclosed test kit.
Registered Trademarks
SUTO® | Registered trademark of SUTO iTEC |
---|---|
MODBUS® | Registered trademark of the Modbus Organization,Hopkinton, USA |
HART® | Registered trademark of the HART Communication Foundation, Austin, USA |
Android™,Google Play | Trademark s of Android™, Google Play |
Application
The S601 Stationary Compressed Air Purity Monitor is a compact wallmounted
device that measures and records three major quality parameters of compressed
air (particle quantity, dew point and temperature).
The S601 is mainly used in industrial environments, and not developed to be
used in explosive areas.
Features
- All-in-one device measures particle concentration, dew point and oil vapor
- Measures additionally the temperature and pressure
- Open protocol outputs enable integration with your management system
- Modbus/RTU (RS-485) and Modbus/TCP (Ethernet) included
- Compact design and easy setup
- Compressed air connection via 6 mm tube
- Integrated data logger saves data for later analysis
- Dew point measurement from -100 … +20°C Td
- Oil vapor measurement from 0.001… 5.000 mg/m³
- The detection limit of oil vapor down to 0.001 mg/m3
- Particle measurement in two versions:
◦ A1260: 0.3 < d ≤ 5.0 µm
◦ A1263: 0.1 < d ≤ 5.0 µm
Technical Data
5.1 General Data
Data logger| Internal, 100-million values
Parameter| Measuring parameter| Range| Reference
Particle| Option A1260:
0.3 < d ≤ 0.5 µm
0.5 < d ≤ 1.0 µm
1.0 < d ≤ 5.0 µm| Annex 1 /DIN 14644
Option A1263:
0.1 < d ≤ 0.5 µm
0.5 < d ≤ 1.0 µm
1.0 < d ≤ 5.0 µm
Dew point| -100 … +20°C Td| DIN 8573
Oil vapor| 0.001 … 5.000 mg/m3| ZLG/AIM 07120604
Pressure| 0.3 … 1.5 MPa| DIN 1301
Temperature| 0 … +40°C| DIN 60751
Reference settings| ISO 1217, 20°C 1000 mbar
Measurement principle| Parameter| Principle
Particle quantity| Laser optical detection
Dew point| Polymer humidity sensor, oscillating crystal
Oil vapor| PID
Resolution of oil vapor sensor| 0.001 mg/m3
Medium| Compressed air, non corrosive components
Humidity of the medium| < 40% rH, non condensation
Temp. of the medium| 0 … +40°C
Operation pressure| 0.3 … 1.5 MPa / 43.5 … 217.6 psig
UV lamp lifetime (oil vapor)| 6,000 working hours or 1 year, whichever comes
first
Housing material| Sheet steel, powder-coated on the outside
Protection class| IP54
Dimension| Please observe the drawings on the next page
Display| 5” graphics color display, 800 x 480 Pixel with touch screen
interface
Weight| 15 kg
5.2 Electrical Data
Power supply | Input: 100 … 240 VAC, 50/60 Hz, 50 VA |
---|
5.3 Accuracy
Particle|
---|---
(Per ISO 21501-4)|
Option A1260| 30…70% of d > 0.3 µm
| 90… 110% of d ≥ 0.45 µm
Option A1263| 30…70% of d > 0.1 µm
| 90… 110% of d ≥ 0.3 µm
Dew point| ±1 °C Td (0 … 20 °C Td)
±2 °C Td (-70 … 0 °C Td)
±3 °C Td (-100 … -70 °C Td)
Oil vapor| 5% of value ± 0.003 mg/m3
Temperature| ± 0.2 K
Pressure| ± 0.08 bar
Dimensions
Dimensions in mm:
Installation on Site
Please make sure that all components listed below are included in your package.
Qty.
| Description|
Item No.
---|---|---
1| S601 Stationary Compressed Air Purity Monitor in a wall mountable cabinet|
D500 0601
1| USB OTG memory stick| A554 0087
1| Operation and instruction manual| No P/N
1| Purge filter for pre-measurement (test kit)| A554 0604
1| 6 mm Teflon hose with quick connector on one end and compressed air
coupling on the other end, 1.5 m| A554 3316
1| Depending on orders:
M12 connector or M12 cable (5 or 10 m)
| Connector: C219 0059 Cable: A553 0104 or A553 0105
1| M12 connector| C219 0059
1| Certificate of calibration| No P/N
If you need replacement materials from the list above or need further useful
accessories, please contact the manufacturer or your local distributor.
7.1 Select Installation Location
The device needs to be set up next to the measuring point.
The air supply tube should not be bended too strongly and should be installed
with a big curve radius.
ATTENTION!
The device is designed to be operated indoors only
7.2 Mount the Unit at the Wall
7.3 Electrical Connections
The S601 provides the following electrical connectors:
- A power supply connector (The cable is inserted through a PG plug and wires are connected to the internal screw terminals.)
- An M12 RS-485 connector: To connect S601 to a RS-485 network
- A RJ-45 Ethernet connector: To connect S601 to a TCP/IP network
In addition, two alarm output relays are provided inside the S601 cabinet.
7.3.1 Power Supply
The power cable for the 230 VAC is connected to the terminals inside of the cabinet, as shown in the picture on the left.
ATTENTION!
Please observe voltage ranges and power levels!
7.3.2 RS-485 Networking (Modbus/RTU)
Through the M12 connector at the bottom, the S601 can be connected to the
RS-485 network over the Modbus/RTU protocol.
The RS-485 network requires a termination resistor at both far ends of the
RS-485 bus. Please set them up correctly.
The RS-485 networking uses the following 3 pins in the M12 connector.
The pin functions are described below.
Pin No.| Assignment|
Function
---|---|---
4| Data +| Data + signal
5| Data –| Data – signal
1| GND RS-485| Ground of RS-485
RS-485 cable
Only cables according to the recommendations of EIA 485 standard should be
used. A maximum of 30 devices may be connected to one segment. The bus cable
must be laid at a distance of at least 20 cm from other cables. It should be
laid in a separate, conductive, and earthed cable trunk. It must be ensured
that no potential differences occur between the individual devices on the bus.
RS-485 cable specification
- Impedance: 135 – 165 Ohm @ 3 to 20 Mhz
- Cable capacity: < 30 pF/m
- Cable diameter: > 0.64 mm
- Cross section: > 0.34 mm², conforms to AWG 22
- Loop resistance: < 110 Ohm/km
- Screening: Cu shielding braid or shielding braid and shielding foil
- Outer diameter for power and sensor cable: 4.5 … 8 mm
7.3.3 TCP/IP Networking (Modbus/TCP)
Through the RJ-45 connector at the bottom, the S601 can be connected to the
TCP/IP network over the Modbus/TCP protocol.
Remove the protection cap and plug in the network cable (RJ-45).
7.3.4 Alarm Outputs
S601 provides two relays for alarm outputs (230 VAC, 3 A) in the S601 cabinet.
The following figure shows the back of the S601 display where the two alarm
output relays (terminals H and I) are housed.
7.4 Compressed Air Connection
The S601 offers a compressed air inlet and an outlet at the bottom of the
cabinet, next to the electrical connectors.
Installation precautions
ATTENTION!
Permissible pressure!
Please observe the maximum permissible incoming pressure. It must be in
between 0.3 and 1.5 MPa. If the pressure exceeds, it will damage the device.
if the pressure is too low, the volume flow will not be high enough, which
will lead to wrong results.
ATTENTION!
Before you connect the device to your point of measurement, make sure that
there is no rough contamination such as water / oil drops or heavy dust. This
could damage the sensor units. Please use the purge filter test kit to check
the contamination.
To check the measuring point, do the following:
- Connect the purge filter test kit onto your measuring point.
- Open the purge valve on the test kit and purge some air for a short period.
- Check the filter in the test kit for high contamination of water, oil or dust.
- If the filter is contaminated severely, stop using the S601 to measure because this may lead to serious damage. In case you are not sure, please contact the manufacturer.
7.5 Connect Modbus/RTU Sensors/Devices to a Master
Sensors and devices with a Modbus/RTU output can be connected to a Modbus
master device. This master can be either SUTO displays and gateways or any
third-party Modbus/RTU master. See below specifications of the Modbus/RTU
connections.
7.5.1 Modbus/RTU Cable Length
A Modbus/RTU bus line has limitations which shall not be exceeded, otherwise
the communication might not be stable.
- The maximum total bus length must not exceed 1200 m.
- The stub length to each node must not exceed 30 cm.
See the picture below for details.
7.5.2 Modbus/RTU Wiring and Cable Type
To ensure a stable communication a twisted-pair bus cable must be chosen to
set up the Modbus/RTU communication.
- Bus-cable specifications must follow the EIA485 Standard, must be twisted pair and shielded, for example 2 x 2 x 0.22 mm2 , Li-2YCY (A553 0123).
- The shield must be connected at one end to the master GND connection.
- At the end of the bus, a 120 Ohm resistor should be placed a termination resistor.
See the below picture for details.
7.5.3 Daisy-Chain using RS-485 Splitter
SUTO devices with M12 connectors can be easily connected in a Modbus/RTU
daisy-chain using a M12 RS-485 splitter (A554 3310). Furthermore this allows
to easily place the M12 termination resistor (C219 0055) at the last splitter
in the bus-chain.
See below example for reference.
Installation on Site
Connect Modbus/RTU slaves with M12 connectors to a daisy-chain using M12
RS-485 splitters.
The M12 RS-485 splitter (A554 3310) comes with two M12 connectors to easily
wire the chain.
7.5.4 Topology of Modbus/RTU Daisy-Chain
The recommended bus topology is to connect devices as a daisy-chain, as shown
in the figure in section 7.5.1 Modbus/RTU Cable Length. Other connection
topologies are not recommended and must be avoided.
Avoid a connection of slaves to the master in ring or star topology.
Setup
The S601 is configured ex-work and ready to measure out of the box.
The setup settings are saved in the device and will not be lost even aftera
power failure.
To ensure an accurate measurement, before measurement, configure the altitude
where the device is placed. The device also enables you to change the ex-work
settings. For more information, see section 9.4.
ATTENTION!
If you have problems setting up your device, feel free to contact the
manufacturer or your local dealer for assistance.
If a power failure occurs during the measurement, the measurement data will
not be saved.
Operations and Configurations
When the S601 is powered on, the initialization screen is displayed with an
active progress bar.
During the initialization, the device configures the sensors and runs
initialization routines. The value screen appears, as shown in the left.
9.1 Main Screen
The following figure shows the main screen in the value view.
9.1.1 Quick Buttons and Icons
Graphic | To switch to the graphic view of measurement data |
---|---|
Value | To switch to the value view of measurement data |
Menu | To access the operation menus |
To scroll down and up the screen | |
To take a screenshot |
9.1.2 Icons in the Status Bar
USB stick connected. Press the icon to disconnect the stick.
Type of the connected ensor does not match with the sensor type configured in
the device.
Logger status
Calibration is expired.
Please contact the manufacturer or your local dealer.
System errors occur. Press this icon to get further information.
Sensor unit is not matching with configuration.
RTC backup battery status
Alarm triggered
9.2 Value Screen
Shows the measured values in real time.
Remark: During the first five minutes, the S601 automatically performs a purge process to ensure any remaining particles in the system are blown out. During this period, the counting numbers on the Particle pane appear green and blink.
9.3 Graphic Screen
Shows the dynamic graphs of all measurements. To view the graphic screen,
press Graphic in the bottom bar.
The graphic view is pre-configured in the factory. You can view the S601
measurement graph without configuring anything. In case that you want to
manipulate the graph, follow the instructions indicated in the following
figure.
9.4 Menu
Enables you to change the S601 settings. You can enter the main menu by
pressing Menu in the quick buttons bar at the bottom.
9.4.1 Sensor Settings
Enables you to configure the measurement units and altitude for the S601. To
change the measurement units. To ensure an accurate oil vapor measurement,
enter the altitude where the S601 is placed. The valid values can be positive
only.
If the S601 is placed at a negative altitude, please enter 0 instead of the
actual negative value.
9.4.2 Alarm Setting
Enables you to define alarms in the S601. The S601 offers two alarm relay
outputs and an optical alarm indication (flashing value).
Channel | Select a channel to be configured. |
---|---|
Unit | The unit is shown automatically after the channel is selected. |
Hysteresis | The alarm is activated when the channel reading reaches the |
threshold and deactivated when the reading is the
threshold minus hysteresis (high alarm) or the threshold plus hysteresis (low
alarm). For example: The threshold is 30 and hysteresis is 5. High alarm is
activated on channel reading ≥30 and deactivated on readings <25. Low alarm
is activated on channel reading ≤30 and deactivated on readings >35.
Direction| UP indicates a high alarm. The alarm is activated when the reading
is higher than the threshold.
Down indicates a low alarm. The alarm is activated when the reading is lower
than the threshold.
Relay| Select a relay. Relay I and Relay H are equipped by default. In
addition, the optional relay module offers 8 additional
relays (Ext. Relay 0 to Ext.Relay 7). These 8 relays can only be displayed and
used when the optional relay module
is installed.
Add button| Click Add to add a channel to set alarm parameters.
Delete button| Click Delete to delete the selected channel.
After setting the alarms, you can view the activated alarms or dismiss some alarms through Menu > Alarm > Activated alarm.
ATTENTION!
Before installing the relay module, turn off the S601.
After the relay module is installed, power on the S601 again. Otherwise, the
normal use of the relay module may be affected.
9.4.3 Logger Setting
Enables you to start and stop the logger and view logger status.
Start time | Shows Logger start time. |
---|---|
Sample /Channel | Show the number of recorded samples per logging channel. |
Logger channel | Shows the total number of the recorded channel. |
Sample rate | Shows the sampling rate. |
To change the sampling rate, click More settings.
Status| Shows the logger status.
Key start| Click to start logging immediately.
Time start| Click to configure a scheduled start for logging.
Select channel| Click to select the channel to log for.
More setting| Click to configure the logging interval, logging mode, and
memory mode.
• Average: Turning Average on means to log the average value during an
interval; turn the average off means the logged value is an instantaneous one.
• Memory mode: Choose a mode as needed.
◦ Full stop: Logging stops if the memory is full.
◦ Wrap around: If the memory is full, new logging data overwrites the oldest
data.
9.4.4 Files
Shows all recorded log files and screenshots as well as memory status.You can
select a log file to view log details or to delete the file.
9.4.5 Service Info
Enables you to view the contacts of the service provider, which can be input
using the S4C-Display software.
For more information about how to use S4C-Display, see section 9.5 Configuring
S601 using S4C-Display。 9.4.6 System Setting
Enables you to configure the S601 general settings.
Password| To configure a password to protect the S601 from unauthorized
operations.
---|---
Back light| To adjust the back light brightness and the dimming timeout.
Calibrate touch screen| To calibrate the touch accuracy.
Language| To select the user interface language.
Date time| To select the user interface language.
Date time| To configure the date and time.
Device info| • To view the device software and hardware information, such as
S/N and firmware versions of the system and the sensors.
• To update the new sensor S/N after an exchange service.
Reset| To reboot the display.
9.4.7 Communication
The menu enables to configure the communication settings for S601.
Click the desired button and follow the instruction shown on the display.
Modbus master| S601 works as the Modbus master. Enter the Modbus communication
parameters for S601 to communicate with Modbus slaves.
---|---
Field-bus RS485| S601 works as the slave. Enter the port (1247) that S601 uses
to communicate with the master.
Field-bus Ethernet| Enter an IP address for the S601 or select DHCP for the
S601 to be assigned with a dynamic IP address.
Web transmitter| Enter the IP address or domain name of the S4M server.
9.5 Configuring S601 using S4C-Display
To change the S601 settings, in addition to the S601 display, you can use the
configuration software S4C-Display. You can download S4CDisplay for free from
the SUTO web page at www.suto-itec.com.
The following sections describe how to change the settings using the S4C-
Display software, but focus on the Field Bus Interface settings.
9.5.1 RS-485 Settings
Return error value| Is the value the salve device will return as a measurement
value in case of any error condition.
---|---
Protocol selection| • Choose SUTO Protocol if the peer device is using SUTO
software.
• Choose Modbus if the S601 is connected to a Modbus network.
Address| Each device on the RS-485 network needs to have a unique device
address. Please ensure that address 0 is not used and that there are no
duplicated addresses.
9.5.2 Ethernet Settings
Return error value| Is the value the salve device will return as a measurement
value in case of any error condition.
---|---
Protocol selection| Please choose SUTO Protocol if using SUTO software, chose
Modbus if device is connected to a Modbus network.
Get IP config automatically| After you select this option, the network router
will dynamically assign the S601 an IP address. This is convenient but not
recommended in industrial networks. It is recommended to assign a static IP
address to the S601.
Use the following IP….| Enter the static IP address for the S601.
9.5.3 Modbus Interface
The default settings of the Modbus interface are as follows:
Communication parameters (Modbus/RTU)
Baud rate | : 19200 |
---|---|
Device address | : Last two digits of serial number |
Framing / parity /stop bit | : 8, N, 1 |
Response time | : 1 second |
Response delay | : 0 ms |
Inter-frame spacing | : 7 char |
Communication parameters (Modbus/TCP)
DHCP | : Yes |
---|---|
MAC | : Set ex-factory |
IP address | : Dynamic or Static |
Subnet | : Dynamic or Static |
Gateway | : Dynamic or Static |
Timeout | : >= 200 ms |
Response message that the device returns to the master:
- Function code: 03
The information for the byte order is shown in the table below:
Byte Order
| Sequence| Data Type
---|---|---
1st| 2nd| 3rd|
4th
1-0-3-2| Byte 1
(MMMMMMMM*)
| Byte 0
(MMMMMMMM *)
| Byte 3 (SEEEEEEE)| Byte 2
(EMMMMMMM *)
| FLOAT
1-0-3-2| Byte 1| Byte 0 LSB| Byte 3 MSB| Byte 2| UINT32 INT32
1-0| Byte 1
MSB
| Byte 0
LSB
| —| —| UINT16
INT16
1-0| Byte 1 XXX *| Byte 0 DATA| —| —| UINT8 INT8
*S: Sign, E: Exponent, M: Mantissa, XXX: no value
Explanations of MSB and LSB
MSB| MSB refers to Most Significant Byte first, which follows the BigEndian
byte order.
For example, if the main system follows the MSB first order:
When the 4-byte floating number, in the order of Byte1-Byte0Byte3-Byte2, is
received from the slave (sensor), the master must change the byte order to
Byte3-Byte2-Byte1-Byte0 for the correct display of the value.
---|---
LSB| LSB refers to Least Significant Byte first, which follows the
LittleEndian byte order.
For example, if the main system follows the LSB first order:
When the 4-byte floating number, in the order of Byte1-Byte0Byte3-Byte2, is
received from the slave (sensor), the master must change the byte order to
Byte0-Byte1-Byte2-Byte3 for the correct display of the value.
Modbus holding register table
Sensor Description| Channel Description| Holding register|
Data type| No. of byte| Unit| Res.| R/ W| Func
Code
---|---|---|---|---|---|---|---|---
Dew point sensor| Temperature| 2| FLOAT_L| 4| °C| 0.1| R| 3
Dew point| 4| FLOAT_L| 4| °C Td| 0.1| R| 3
Serial number sensor 1| 10| UINT32_L| 4| | 1| R| 3
Serial number sensor 2| 12| UINT32_L| 4| | 1| R| 3
Device status Dew point| 14| UINT32_L| 4| | 1| R| 3
Particle counter| 0.1 < d ≤ 0.5 um| 20| FLOAT_L| 4| cn/m³| 1| R| 3
0.5 < d ≤ 1.0 um| 22| FLOAT_L| 4| cn/m³| 1| R| 3
1.0 < d ≤ 5.0 um| 24| FLOAT_L| 4| cn/m³| 1| R| 3
0.3 < d ≤ 0.5 um| 26| FLOAT_L| 4| cn/m³| 1| R| 3
Serial number| 30| UINT32_L| 4| | 1| R| 3
Device status particle| 32| UINT32_L| 4| | 1| R| 3
Oil vapor sensor| Oil vapor| 40| FLOAT_L| 4| mg/m³| 0.001| R| 3
Pressure| 42| FLOAT_L| 4| bar| 0.01| R| 3
Serial number sensor| 48| UINT32_L| 4| | 1| R| 3
Device status oil vapor| 50| UINT32_L| 4| | 1| R| 3
Display unit| Serial number display| 60| UINT32_L| 4| | 1| R| 3
Serial number sensor rack| 62| UINT32_L| 4| | 1| R| 3
Device status display| 64| UINT32_L| 4| | 1| R| 3
Remarks:
- Device status information will be defined in a later state.
- Display unit serial number is the same number as the serial number of the complete S601 unit.
Software Installation
Following software is available for download from www.suto- itec.com:
S4C-Display | Free configuration software |
---|---|
S4A | Free logger readout and data analysis software |
S4M | Multiple-device data acquisition and analysis software |
After download, please follow the onscreen instructions to install the software.
Optional Accessories
To buy the following optional accessories, please contact the manufacturer or your local dealer.
- Teflon hoses
- Sorts of adapters
- Relay module (P/N: A1510)
Maintenance
Use a moist fabric to clean the device. For the use in GMP-area, it ispossible to disinfect through wipe disinfection. For this please contacthe manufacturer or your local dealer.
ATTENTION!
Please dry the device after cleaning using a clean and dry fabric. Always take
care, that the fabric for cleaning is not too wet because water could get into
the device and cause damage.
Disposal or Waste
Electronic devices are recyclable material and do not belong in the household
waste.
The sensor, the accessories and its packing must be disposed according to your
local statutory requirements.
The dispose can also be carried by the manufacturer of the product, for this
please contact the manufacturer.
Calibration
For calibration we recommend you use our exchange service, where the complete sensor unit is exchanged with a newly-calibrated unit. This saves time and costs.
ATTENTION!
Please save all your measurement data on an external device before returning
the instrument tocalibration and service. It might be necessary to reset the
displays storage during calibration and service.
After you exchange a new sensor unit, remember to input its serial number to
the S601 through Menu > System Setting > Device info.
Warranty
Please find the warranty as a separated warranty card included with the
instrument delivery.
The warranty does not cover any wear parts or consumables, therefore the UV
lamp with limited lifetime as well as the internal filter are not covered by
the warranty.
SUTO iTEC GmbH
Grißheimer Weg 21 D-79423 Heitersheim Germany
Tel: +49 (0) 7634 50488 00
Email: sales@suto-itec.com
Website: www.suto-itec.com
SUTO iTEC (ASIA) Co., Ltd.
Room 10, 6/F, Block B, Cambridge Plaza
188 San Wan Road, Sheung Shui, N.T. Hong Kong
Tel: +852 2328 9782
Email: sales.asia@suto-itec.com
Website: www.suto-itec.com
SUTO iTEC Inc.
5460 33rd St SE Grand Rapids, MI 49512 USA
Tel: +1 (616) 800 -7886
Email: sales.us@suto-itec.com
Website: www.suto-itec.com
All rights reserved ©
Modifications and errors reserved
S601_im_en_2023-2
References
- ITEC Entertainment | A TAIT Company
- SUTO iTEC - Messtechnik für Druckluft und Gase
- SUTO iTEC - Messtechnik für Druckluft und Gase
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>