Gikaida H2S Voice Type Multifunctional Gas Detector User Manual
- June 15, 2024
- Gikaida
Table of Contents
Gikaida H2S Voice Type Multifunctional Gas Detector
Product Information
Specifications
- Product Name: Voice Type Multifunctional Gas Detector
- Usage : Safety monitoring for toxicity, harmfulness, and explosion prevention
- Application: Petroleum, chemical industry, environmental protection, metallurgy, refining, gas, biochemical medicine, agriculture, fire protection, archaeology, and other industries
- Working Principle: Electrochemical and catalytic combustion
- Dimensions : 7035140mm
- Weight : 210g
Product Usage Instructions
Startup Process
Keep the air around the detector clean before turning it on.During the
startup process, the detector will perform vibration and light self-tests.
Please observe whether the self-tests pass:
- Sound
- Display
- Light
- Vibration
Gas Concentration Display
After the boot is completed, the normal gas main surface will be displayed,
and the real-time gas concentration can be seen. Press the left button to view
the system status information.
Gas Concentration Levels:
- EX Alarm:
- L Alarm: 20%LEL
- H Alarm : 50%LEL
- R Alarm: 100%LEL
- O2 Alarm:
- L Alarm: 19.5%
- H Alarm : 23.0%
- R Alarm: 30.0%
- H2S Alarm:
- L Alarm: 10PPM
- H Alarm: 35PPM
- R Alarm : 100PPM
- CO Alarm:
- L Alarm: 50PPM
- H Alarm : 150PPM
- R Alarm: 1000PPM
Function Settings
The menu of the detector contains the following options:
- Alarm Record
- Set Low Alarm
- Set High Alarm
- Zero Adjust
- Calibrate Gas
- Set Time
- Set Backlight
- Set Password
- Channel Info
- Set Language
- Exit
Setting Low Alarm Concentration
- In the menu list, choose “Set L Alarm” and press the power key to enter the setting interface.
- Use the up and down keys to select the channels of combustible gas, oxygen, hydrogen sulfide, and carbon monoxide.
- After selecting the channels you need to set, press the power key to enter the page of setting low alarm concentration.
- Adjust the alarm concentration using the up and down keys.
- Press the “Yes” key to save the alarm settings.
Setting High Alarm Concentration
- In the menu list, choose “Set H Alarm” and press the power key to enter the setting interface.
- Use the up and down keys to select the channels of combustible gas, oxygen, hydrogen sulfide, and carbon monoxide.
- After selecting the channels you need to set, press the power key to enter the page of setting high alarm concentration.
- Adjust the alarm concentration using the up and down keys.
- Press the power key to save the alarm settings.
Thank you for using our products, please be sure to read this manual first when you are ready to use th s product, read th s manual and follow the nstructions prov ded so that you can be able to fully enjoy the products and serv ces prov ded by our company wh le avo d ng unnecessary behav oral damage or other accidents.
Product introduction
portable voice type multi gas detector is a safety device that can continuously detect the concentration of leaking gas. It adopts advanced integrated circuit technology, embedded computer control, and high-quality imported gas sensors, with excellent sensitivity and repeatability. It uses dot matrix LCD display to support Chinese-English interface and Chinese- English voice prompt, so that users can quickly understand the product and use and maintain it simply. High-strength engineering plastics, good seismic resistance, high strength, high-grade atmospheric appearance and dust-proof, waterproof and explosion-proof function, etc.
The detector is widely used in petroleum, chemical industry, environmental protection, metallurgy, refining, gas, biochemical medicine, agriculture, fire protection, Archaeology and other industries and places that need safety monitoring for toxicity and harmfulness, explosion prevention. The detector can effectively predict the concentration of dangerous gases and alarm, so as to ensure the safety of workers and production equipment.
Structure introduction
Structure
Working principle
Electrochemical and catalytic combustion
Technical parameter
Gas type | Range | Low alarm | High | Resolutio |
---|---|---|---|---|
EX | (0-100)¾LEL | 20%LEL | 50%LEL | l¾LEL |
H2S | (0-l00)ppm | l0ppm | 35ppm | lppm |
co | (0-1000)ppm | 50ppm | 150ppm | lppm |
02 | (0-30)%vol | 19.5%vol | 23.5%vol | 0.l¾vol |
Other gas type please contact us.
Precision:| :S ±5%FS
Response time| T<30s
Display:| LCD display real-time data and station of system
Alarm:| Audible-visual-vibration alarm withvoice alarm
Work| -20’C -so·c; <95%RH non-condensing
Power supply:| DC3.7V (Rechargeable lithium batteryl800mAh)
Explosion-Proof grade:| ExibIIBT3Gb
Battery charge
time:
| 4h-6h
Working time:| Over 10 hours
Sensor life:| 2 years
Protection grad| Ip65
Dimensions:| 7035140mm
Weight:| 210g
Function and operation
Tum on, Tum off, Charging
Tum on
- When the detector is off, press and hold the middle power button for about 3 seconds, release the button when the LCD screen displays the software version interface, and the indicator light flashes twice, at this time, the LCD screen displays the boot welcome page, and you can hear the voice prompt “Welcome to use multi-function voice gas detector
- Keep the air around the detector clean on the page waiting to be turned on. During the startup process, the detector will perform vibration and light self-test, please pay attention to observe whether it is normal. The above status is normal, indicating that the sound, display, light, and vibration self-test pass
- After the boot is completed, the normal gas main surface can be displayed, and the real-time gas concentration can be seen. Press the left button to view the system status information
Turn off
- Keep the air around the detector clean on the page waiting perior vibration and light scar-test, pocase pay attention to observe whether it is normal. The above status is normal, indicating that the sound, display, light, and vibration self- test pass
Charging
- The USB charging cable is plugged in when the detector is in shutdown mode, the current power will be displayed on the screen, when there is only one grid left, the LCD will pop up a low battery page every 60 seconds, and there has a voice indicating“Low battery, please charge”.
- When the battery is less than 1 grid, the detector will automatically shut down to prevent data loss inside the detector and unpredictable damage caused by insufficient voltage on the internal sensitive components of the instrument
Function setting
- The menu of detector contain as follow: Alarm Record, Set Low Alarm, Set High Alarm, Zero Adjust, Calibrate Gas, Set Time, Set Backlight, Set Password, Channel Info, Set Language, Exit.
- At the detecting interface, press Right key, then input the password, default password is 1111
Then you will get the menu.
Set Low Alarm
Under the menu list choose the Set L Alarm, press power key then you can
enter the setting interface;you can select the channels of combustible gas,
oxygen, hydrogen sulfide and carbon monoxide by up and down keys; after
selecting the channels you need to set, press the power key to enter the page
of setting low alarm concentration; The alarm concentration can be
adjusted according to the Up and Down, and after the Yes key is pressed, the
alarm is saved.
Set High Alarm
Under the menu list choose the Set H Alarm, press power key then you can
enter the setting interface;you can select the channels of combustible gas,
oxygen, hydrogen sulfide and carbon monoxide by up and down keys; after
selecting the channels you need to set, press the power key to enter the page
of setting high alarm concentration; The alarm concentration can be adjusted
according to the Up and Down, and after the power key is pressed, the alarm is
saved.
Zero Adjust
Under the menu list choose the Zero Adjust, press power key then you can
enter the setting interface;you can select the channels of combustible gas,
oxygen, hydrogen sulfide and carbon monoxide by up and down keys; after
selecting the channels you need to set, press the power key to enter the page
of zero adjust interface. The data is your gas concentration, after the Yes
key is pressed, the alarm is saved.
Warning : You must do this in the clean air (The 02 must adjust in the N2 gas environment) , ensure there is no detected gas in the air. Or thedetect value will be wrong !
Cali Gas
Prepare gas cylinders, pressure limiting valves, flow meters, gas pipes, and
calibration hoods before calibration. When the arrow in menu selection bar
points to the calibration setting, press the middle confirmation key to switch
to the interface for selecting the calibration setting channel; you can
scootie channst be ine and dress, Are ho sir on and” ” ” keys; After
choosing the setting channel, press the OK key to enter the calibration
setting page,
Open the pressure limiting valve of the cylinder with the standard concentration, and adjust the flow to 200-400ml/min. connected to our gas detector through a gas pipe and a calibration gas hood. observe the AD value displayed on the instrument, at this time, the AD value should be rising, wait for about 1 minute for the AD value to rise to the peak value and be stable and not floating. Use ” ” and ” ” to adjust the concentration value of the calibration gas. For example, if the carbon monoxide concentration of the gas cylinder is 250 ppm, you need to adjust the concentration value displayed on the instrument to 250 ppm. Click OK to save.
Calibration successful
Warning : Non professional personnel are strictly prohibited from carrying out this operation, otherwise all consequences will be borne by themselves. The tester has been calibrated uniformly when it is out of the factory. If the user wants to re-calibrate, please follow the steps strictly, first set the zero point, then re-calibrate the settings. If the user fails to operate this setting, please contact the manufacturer back in time for calibration
Set Backlight
Under the menu list choose the Set Backlight, press power key then you can
enter the setting interface;Modify the type by press Up or Down key. After
modified press the save is ok. Auto: the light will keep for 30 second;
Normally on: the light will always be working.
Set Password
Under the menu list choose the Set Password, press power key then you can
enter the setting interface; Before modify the password, you need to validate
the password first, then you can change it.
The default password is 1111
Attention : please save the password properly. Once lost, it can only be returned to the manufacturer to resume the factory settings.
Channel Info
Under the menu list choose the Channel info, press power key then you can
enter the info interface; By pressing the Up or Down to choose the gas type,
will display the gas name, low alarm, high alarm, detecting range
Language
Under the menu list choose the Set Language, press power key then you can
enter the setting interface;
Use
- When the detector is turned on, the LCD screen is in the standby state of displaying four concentrations of EX gas, O2, H2S and CO in real time.
- When the concentration of EX gas is detected to be higher than the low alarm value set by the system, the detector will broadcast the voice “Please note that the concentration of EX gas exceeds the standard” and continue to simulate the alarm sound; the alarm lamp above the detector and the internal vibration motor will be turned on simultaneously; When the alarm value is low, the voice, light and vibration alarm states disappear.
O2 alarm
- When the concentration of O2 is detected to be lower than the low alarm value set by the system, the detector will broadcast the alarm signal “Please note that the concentration of oxygen is too low”, which is higher than the high alarm threshold set by the system. The detector will broadcast the alarm signal “Please note that the concentration of oxygen exceeds the standard” and always simulate the alarm sound; the alarm light on the detector will voice the alarm signal. When the concentration of O2 gas is monitored by the detector and restored to the normal value, the alarm state of voice, light and vibration disappears.
- Note : O2 concentration in normal air is 20.9%. Users can adjust the value of low and high alarm according to actual needs.
- When the concentration of H2S gas is detected to be higher than the low alarm value set by the system, the detector will broadcast the voice “Please note that the concentration of H2S gas exceeds the standard” and continue to simulate the alarm sound; the alarm lamp above the detector and the internal vibration motor will be turned on simultaneously; When the alarm value is low, the voice, light and vibration alarm states disappear.
When the concentration of CO gas is detected to be higher than the low alarm value set by the system, the detector will broadcast the voice “Please note that the concentration of CO gas exceeds the standard” and continue to simulate the alarm sound; the alarm lamp above the detector and the internal vibration motor will be turned on simultaneously; When the alarm value is low, the voice, light and vibration alarm states disappear.
Note :The alarm voice support Chinese or English, as same as the language you choose.
Warning:
- The detector only alarm when it is detecting;
- Do not charging detectors at the gas monitoring site to avoid fire or explosion caused by sparks during the plugging
- Try not to charge the detector on startup, so as not to affect charging speed.
- The detector has the function of intelligent tracking zero point. Please turn on under the condition of clean air.
Notes
- Prevent the detector from falling down or being subjected to severe vibration.
- In the presence of high concentration gas, the detector may not be used normally.
- Please strictly operate and use according to the instruction manual, otherwise it may cause the test result to be inaccurate or damage the detector.
- This product should not be stored or used in environments containing corrosive substances (such as chlorine of relatively high concentration) or in other harsh environments, including too high or too low temperature, high humidity, electromagnetic field and strong sunlight exposure
- If there is dirt on the surface of the instrument after long-term use, please gently wipe it with clean soft cloth dipped in water. Do not use corrosive solvents and hard materials to wipe the surface of the machine. Otherwise, it may lead to scratches or damage on the surface of the instrument.
- Please unload the lithium batteries and sensors of the discarded portable gas detector and send them to the designated location or return to our company. Do not throw them into the garbage can at will
- Any failure or failure to be described in this manual should be contacted with our company for settlement..
- In the explosive gas environment, the battery pack can not be dismantled or replaced, nor can the battery pack be charged. In explosive gas environment, peripheral plug-in devices without explosion-proof certification can not be used, nor can sensors be replaced.
- This product has obtained explosion-proof certification. Users are not allowed to change or change the components or structures affecting explosion-proof performance without permission..
Common faults and Solutions
Fault | Failure cause | Operation |
---|
Can not turn on
| Low battery| Please get charging
Detector’s down| Please contact the factory
Circuit fault| Please contact the factory
No detect value| Circuit fault| Please contact the factory
Detecting not true| Sensor over life| Please contact the factory to change
Long time no cali| Get to calibration
Time display wrong
| Battery power is exhausted| Please get charging and re- setting time
Strong electromagnetic| Re-setting time
Set zero fault| The sensor zero drift too high| Calibrate it or change sensor
---|---|---
After detection the value not
back to 0(exce
pt O2 )
|
Zero drift
|
Zero calibrate
Detector
display whole range
|
Sensor bad
|
Change new sensor
Storage
The detector should be stored in the ventilation room with ambient
temperature -10 to 55℃ and humidity <85%. Avoid direct sunlight, and air can
not contain harmful gases or impurities that are corrosive to the detector.
Package
Packing box | 1 |
---|---|
Portable gas detector | 1 |
Charger | 1 |
USB line | 1 |
Manual | 1 |
mask | 1 |
Appendix – Sensor Selection Table
Gas to be measured | Measuring range | **** | **** | **** |
---|---|---|---|---|
**** | 0-100%lel | 0-100%vol | 1%lel/1%vol | **** |
O2 | 0-30%vol | 0-30%vol | 0.1%vol | **** |
H2S | 0-100ppm | 0-50/200/1000ppm | 0.1ppm | **** |
CO
| 0-1000ppm| 500/2000/5000
ppm
| 1ppm| 50
l50ppm
****| 0-5000ppm| 0-1%/5%/10%vol| 1ppm/0.1%vol| 1 000
2000ppm
NO| 0-250ppm| 0-500/1000ppm| 1ppm| 50
l50ppm
NO2| 0-20ppm| 0-50/1000ppm| 0.1ppm| 5
l0ppm
SO2| 0-20ppm| 0-50/1000ppm| 0.1/1ppm| 5
l0ppm
CL2| 0-20ppm| 0-100/1000ppm| 0.1ppm| l0ppm
H2| 0-1000ppm| 0-5000ppm| 1ppm| 50
l50ppm
NH3| 0-100ppm| 0-50/500/1000ppm| 0.1/1ppm| 20
50ppm
PH3| 0-20ppm| 0-20/1000ppm| 0.1ppm| 5
l0ppm
HCl| 0-20ppm| 0-20/500/1000ppm| 0.001/0.1ppm| 5
l0ppm
CLO2| 0-50ppm| 0-10/100ppm| 0.1ppm| 5
l0ppm
HCN| 0-50ppm| 0-100ppm| 0.1/0.01ppm| 10
20ppm
C2H4O| 0-100ppm| 0-100ppm| 1/0.1ppm| 20
50ppm
O3| 0-10ppm| 0-20/100ppm| 0.1ppm| ****
5 ppm
CH2O| 0-20ppm| 0-50/100ppm| 1/0.1ppm| 5
l0ppm
HF| 0-100ppm| 0-1/10/50/100ppm| 0.01/0.1ppm| 2
5 ppm
C7H8/C8H10| 0-20ppm| 0-1/10/50/100ppm| 0.01/0.1ppm| 5
10 ppm
FAQ’s
What is the default password for accessing the menu?
The default password is 1111.