HACH AT1000 Series Titration Bundle User Manual

June 14, 2024
HACH

AT1000 Series Titration Bundle

Product Specifications

Dimensions

Width: 22 cm (8.7 in)

Depth: 40 cm (15.7 in)

Height: 36 cm (14.2 in)

Weight

4 kg (8.8 lb)

Power Requirements

Input: 24 VDC, 2.5 A

Main Supply Voltage Fluctuation

Altitude: 2,000 m (6,562 ft) maximum

Operating Temperature

20 to 80% relative humidity, non-condensing

Storage Temperature

Overvoltage Category: II

Pollution Degree

2

Protection Class

II

Environmental Conditions

Indoor use only

Certifications

Safety: IEC/EN 61010-1, UL/CSA C22.2 61010-1

EMC: IEC/EN 61326-1

Warranty

1 year (EU: 2 years)

Product Usage Instructions

Safety Information

DANGER: Indicates a potentially or imminently
hazardous situation which, if not avoided, will result in death or
serious injury.

WARNING: Indicates a potentially or imminently
hazardous situation which, if not avoided, could result in death or
serious injury.

CAUTION: Indicates a potentially hazardous
situation that may result in minor or moderate injury.

NOTICE: Indicates a situation which, if not
avoided, may cause damage to the instrument. Information that
requires special emphasis.

Precautionary Labels

Read all labels and tags attached to the instrument. Personal
injury or damage to the instrument could occur if not observed. A
symbol on the instrument is referenced in the manual with a
precautionary statement. This symbol, if noted on the instrument,
references the instruction manual for operation and/or safety
information.

This symbol indicates that a risk of electrical shock and/or
electrocution exists.

This symbol indicates the presence of devices sensitive to
Electro-static Discharge (ESD) and indicates that care must be
taken to prevent damage with the equipment.

Electrical equipment marked with this symbol may not be disposed
of in European domestic or public disposal systems. Return old or
end-of-life equipment to the manufacturer for disposal at no charge
to the user.

Compliance and Certification

CAUTION: This equipment is not intended for use
in residential environments and may not provide adequate protection
to radio reception in such environments.

1. The equipment may not cause harmful interference.

2. The equipment must accept any interference received,
including interference that

FAQ

Q: Can this product be used outdoors?

A: No, this product is designed for indoor use only.

Q: What is the warranty period for this product?

A: The warranty period is 1 year, with an extended warranty of 2
years for customers in the European Union.

DOC022.52.93074
TitraLab® AT1000 series workstations
User Manual 12/2022, Edition 10

Table of Contents
Section 1 Specifications…………………………………………………………………………………………………….. 3
Section 2 General information…………………………………………………………………………………………… 5 2.1 Safety information………………………………………………………………………………………………………… 5 2.1.1 Use of hazard information……………………………………………………………………………………… 5 2.1.2 Precautionary labels……………………………………………………………………………………………… 5 2.1.3 Compliance and certification………………………………………………………………………………….. 6 2.1.4 Restriction of hazardous substances (RoHS)…………………………………………………………… 6 2.2 Product overview………………………………………………………………………………………………………….. 7 2.3 Instrument connections…………………………………………………………………………………………………. 8 2.4 Product components…………………………………………………………………………………………………….. 9
Section 3 Installation………………………………………………………………………………………………………… 11 3.1 Installation guidelines………………………………………………………………………………………………….. 11 3.2 Connect to AC power………………………………………………………………………………………………….. 11 3.3 Install the syringe……………………………………………………………………………………………………….. 12 3.4 Install the sensor storage tubes……………………………………………………………………………………. 13 3.5 Install the stir bar and the beaker………………………………………………………………………………….. 13 3.6 Prepare the tubes……………………………………………………………………………………………………….. 13 3.7 Connect the tubes………………………………………………………………………………………………………. 14 3.8 Install the sensor………………………………………………………………………………………………………… 14 3.8.1 Install the legacy adapter…………………………………………………………………………………….. 14 3.8.1.1 Set the legacy adapter settings…………………………………………………………………….. 15 3.8.2 Connect the sensor…………………………………………………………………………………………….. 15 3.9 Install the titrant and the reagent…………………………………………………………………………………… 16 3.10 Tidy the work area…………………………………………………………………………………………………….. 18
Section 4 Install accessories……………………………………………………………………………………………. 19 4.1 Install an external pump………………………………………………………………………………………………. 19 4.2 Install an external propeller………………………………………………………………………………………….. 21 4.3 Install a balance…………………………………………………………………………………………………………. 21 4.4 Install PC software……………………………………………………………………………………………………… 22 4.5 Install the syringe verification kit…………………………………………………………………………………… 22
Section 5 User interface and navigation…………………………………………………………………………..25 5.1 Keypad……………………………………………………………………………………………………………………… 25 5.2 Home screen……………………………………………………………………………………………………………… 25 5.3 Menu structure…………………………………………………………………………………………………………… 26
Section 6 Startup………………………………………………………………………………………………………………. 29 6.1 Configure the instrument……………………………………………………………………………………………… 29 6.2 Install the applications…………………………………………………………………………………………………. 30 6.3 Prepare the instrument for measurement………………………………………………………………………. 30
Section 7 Standard operations………………………………………………………………………………………… 33 7.1 Calibration…………………………………………………………………………………………………………………. 33 7.1.1 Calibrate the sensor……………………………………………………………………………………………. 33 7.1.2 Calibrate the titrant……………………………………………………………………………………………… 34 7.1.3 Auto leveling calibration………………………………………………………………………………………. 35 7.2 Sample preparation…………………………………………………………………………………………………….. 36 7.3 Get a sample measurement…………………………………………………………………………………………. 36 7.4 Change the application settings……………………………………………………………………………………. 38 7.4.1 Sample name…………………………………………………………………………………………………….. 38 7.5 Manage the data log…………………………………………………………………………………………………… 39 7.6 Purge………………………………………………………………………………………………………………………… 39
1

Table of Contents Section 8 Maintenance……………………………………………………………………………………………………… 41
8.1 Maintenance schedule………………………………………………………………………………………………… 41 8.2 Clean the instrument…………………………………………………………………………………………………… 41 8.3 Clean the sensor………………………………………………………………………………………………………… 41 8.4 Replace the tubes………………………………………………………………………………………………………. 42 8.5 Replace desiccant cartridge contents……………………………………………………………………………. 42 8.6 Replace the syringe electro-valve block…………………………………………………………………………. 42 8.7 Maintenance menu……………………………………………………………………………………………………… 42
8.7.1 Syringe activation……………………………………………………………………………………………….. 43 8.7.2 Pump activation………………………………………………………………………………………………….. 43 8.7.3 Syringe management………………………………………………………………………………………….. 43 8.7.4 Syringe verification……………………………………………………………………………………………… 43 8.7.5 Pump cassette replacement…………………………………………………………………………………. 44 8.7.6 Other maintenance options………………………………………………………………………………….. 45 8.8 Storage and transportation…………………………………………………………………………………………… 46 8.8.1 Prepare the instrument for storage…………………………………………………………………………46 8.8.2 Prepare the instrument for short-term storage………………………………………………………… 47 8.8.3 Prepare the instrument for shipment……………………………………………………………………… 47 Section 9 Troubleshooting………………………………………………………………………………………………..49 Section 10 Replacement parts and accessories………………………………………………………………57 Appendix A Change application parameters…………………………………………………………………… 63 Appendix B Frequently asked questions………………………………………………………………………… 73 Section 13 Configure the instrument………………………………………………………………………………. 75
2

Section 1 Specifications

Specifications are subject to change without notice.

Specification Dimensions (W x D x H)

Details 22 x 40 x 36 cm (8.7 x 15.7 x 14.2 in.)

Weight

4 kg (8.8 lb)

Power requirements

Instrument: Input 24 VDC, 2.5 A
External power supply: Input 100­240 VAC, 50­60 Hz, 1.5 A, Class I; output 24 VDC, 2.5 A 60 VA

Main supply voltage fluctuation

±10% of nominal voltage

Altitude

2,000 m (6,562 ft) maximum

Operating temperature

15 to 35 °C (59 to 95 °F)

Relative humidity

20 to 80%, non-condensing

Storage temperature

­5 to 40 °C (23 to 104 °F)

Overvoltage category

II

Pollution degree

2

Protection class

II

Environmental conditions

Indoor use

Certifications EMC requirements

Safety: IEC/EN 61010-1, UL/CSA C22.2 61010-1 EMC: IEC/EN 61326-1
This product is intended to be used in a domestic or basic electromagnetic environment.

Warranty

1 year (EU: 2 years)

3

Specifications
4

Section 2 General information
In no event will the manufacturer be liable for damages resulting from any improper use of product or failure to comply with the instructions in the manual. The manufacturer reserves the right to make changes in this manual and the products it describes at any time, without notice or obligation. Revised editions are found on the manufacturer’s website.
2.1 Safety information
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including, without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent permitted under applicable law. The user is soley responsible to identify critical application risks and install appropriate mechanisms to protect processes during a possible equipment malfunction. Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment. Make sure that the protection provided by this equipment is not impaired. Do not use or install this equipment in any manner other than that specified in this manual.
2.1.1 Use of hazard information
DANGER
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in minor or moderate injury.
NOTICE
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special emphasis.
2.1.2 Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement. This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
This symbol indicates that a risk of electrical shock and/or electrocution exists.
5

General information
This symbol indicates the presence of devices sensitive to Electro-static Discharge (ESD) and indicates that care must be taken to prevent damage with the equipment.
Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.
2.1.3 Compliance and certification
CAUTION
This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such environments.
Canadian Radio Interference-Causing Equipment Regulation, ICES-003, Class A: Supporting test records reside with the manufacturer. This Class A digital apparatus meets all requirements of the Canadian InterferenceCausing Equipment Regulations. Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne sur les équipements provoquant des interférences. FCC Part 15, Class “A” Limits Supporting test records reside with the manufacturer. The device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions:
1. The equipment may not cause harmful interference. 2. The equipment must accept any interference received, including interference that
may cause undesired operation.
Changes or modifications to this equipment not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at their expense. The following techniques can be used to reduce interference problems:
1. Disconnect the equipment from its power source to verify that it is or is not the source of the interference.
2. If the equipment is connected to the same outlet as the device experiencing interference, connect the equipment to a different outlet.
3. Move the equipment away from the device receiving the interference. 4. Reposition the receiving antenna for the device receiving the interference. 5. Try combinations of the above.
2.1.4 Restriction of hazardous substances (RoHS)
Dieses Informationsblatt enthält Angaben, die ausschließlich für den Export dieses Gerätes in die Volksrepublik China erforderlich sind. This document contains information which is only required for the export of this instrument into the People’s Republic of China. Type: AT1xxx
6

(Parts) (Resistor) ( Diode)

General information

Name: TitraLab AT1000 Table 1 (Hazardous Substance)

(Pb) (Hg) (Cd) (Cr6+)

(PBB)

X

O

O

O

O

X

O

O

O

O

(PBDE) O O

SJ/T11363-2006 O: Indicates that the concentration of the hazardous substance in all homogeneous materials in the parts is below the
relevant threshold of the SJ/T11363-2006 standard.
SJ/T11363-2006 X: Indicates that the concentration of the hazardous substance of at least one of all homogeneous materials in the parts
is above the threshold of the SJ/T11363-2006 standard.
, , : This table shows where these substances may be found in the supply chain of “Company” electronic information products, as of the date of sale of the enclosed product. Note that some of the component types listed above may or may not be a part of the enclosed product.

, . ( , ) . . The Environmentally Friendly Use Period (EFUP) for all enclosed products and their parts are per the symbol shown here, unless otherwise marked. Certain field-replaceable parts have a different EFUP (for example, battery modules) and so are marked to reflect such. The Environmentally Friendly Use Period is valid only when the product is operated under the conditions defined in the product manual.

2.2 Product overview
The instrument operates with digital and analog sensors to do potentiometric (AT1000 series) and volumetric (KF1000 series) titration. Measurement applications are installed on the instrument to automate the measurement process. Instructions show on the display when user intervention is required.

7

General information

1 Keypad

6 Beaker

11 Sensor holder

2 Display

7 Syringe protection cover

12 Pump 2 input/output

3 Sensor storage tubes 4 USB port1

8 Syringe input/output 9 Tube clips

13 Pump 1 input/output 14 Pump access cover

5 Tube holder

10 Syringe Note: Depending on the model, there will be 1 or 2 syringes and syringe input/output ports, and 0, 1 or 2 pumps. Refer to Table 2.

Table 2 Instrument configurations

Model

Syringes

Pumps

AT1102

1

0

AT1112

1

1

AT1122

1

2

AT1222

2

2

2.3 Instrument connections
DANGER
ElectricaI shock hazard. Externally connected equipment must comply with an applicable country safety standard (e.g., IEC 60950-1 or IEC 62368-1 for IT equipment) and the circuit intended to be connected to the equipment shall not exceed the SELV (safety extra low voltage) safe level.

1 A second USB port is found on the rear of the instrument, but the instrument recognizes only one USB storage device connection at a time.
8

General information
Use the USB port on the side of the instrument for the USB applications key supplied with the instrument. Use the USB port on the rear of the instrument to connect to a printer, mouse, keyboard or a USB hub.

1 24 V external power supply port 2 Sensor 1 port 3 Sensor 2 port

4 External pump port 5 External propeller port 6 Serial port

7 USB port 8 Ethernet port

2.4 Product components
Make sure that all components have been received. Refer to the packing list in the box. If any items are missing or damaged, contact the manufacturer or a sales representative immediately.
Figure 1 Contents of the instrument box

Note: For identification, the outlet tube from the syringe is fitted with a blue marker.

1 Instrument
2 Tube holder (1 for each syringe position on the instrument)

3 Sensor storage tubes (3x) 4 Power supply

5 Power cord

9

General information
Figure 2 Contents of the application box

1 Beakers (5 x 50 mL and 5 x 150 mL) 2 Tube with anti-diffusion tip (if necessary for the
application) 3 Conical adapters (quantity depends on application) 4 Magnetic stir bars 5 Legacy sensor adapter (not in all application kits) 6 Sensor (type and quantity depends on application)

7 Syringe holding ring (1 for each syringe) 8 Syringe (refer to Table 2 on page 8 for quantity)
9 USB applications key 10 Glass bottles (not in all application kits) 11 Bottle caps (type and quantity depends on application)

10

Section 3 Installation
CAUTION
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.
NOTICE
This is a class A product. There may be potential difficulties in ensuring electromagnetic compatibility in other environments, due to conducted as well as radiated disturbances. In a domestic environment this product may cause radio interference in which case the user may be required to take adequate measures.
NOTICE
Network and access point security is the responsibility of the customer that uses the wireless instrument. The manufacturer will not be liable for any damages, inclusive however not limited to indirect, special, consequential or incidental damages, that have been caused by a gap in, or breach of network security.
The instrument is available in different configurations (refer to Table 2 on page 8). This manual supplies instructions for the installation of an instrument with one syringe and one pump. Adjust the installation procedure as applicable to accommodate the number of syringes and pumps in the instrument.
3.1 Installation guidelines
· This instrument is for indoor use only. · The mains plug must be easily accessible so the power can be disconnected quickly
in case of emergency. · Protective earth ground (PE) connection is required. · Keep the instrument away from temperature extremes, including heaters, direct
sunlight and other heat sources. · Put the instrument on a stable and level surface in a well ventilated place. · Make sure that there is at least 15 cm (6 in.) of space on all sides of the instrument to
prevent electrical parts from overheating. · Do not operate or keep the instrument in dusty, damp or wet locations. · Always keep the surface of the instrument and all accessories dry and clean.
3.2 Connect to AC power
CAUTION
Electrical shock and fire hazards. Make sure that the supplied cord and non- locking plug meet the applicable country code requirements.
WARNING
Fire hazard. Use only the external power supply that is specified for this instrument.
WARNING
Electrocution hazard. Protective Earth Ground (PE) connection is required.
11

Installation
1. Connect the power cord to the power supply. 2. Connect the power supply to the instrument (refer to Instrument connections
on page 8). 3. Connect the power cord to an electrical outlet.
3.3 Install the syringe
Before syringe installation, set the instrument power to on. Push the power button on the front of the instrument. Make sure that the startup sequence shows on the display. The syringe holder lowers to its operating position. Note: Ignore any warning messages related to missing applications that show on the display. The sensor holder has two positions: one over the magnetic stirrer and the second at 180° to the right. Move the sensor holder away from the instrument to the second position. Refer to the illustrated steps that follow.
CAUTION
In step 6, tighten the syringe using the metallic part at the top. Do not hold the glass section of the syringe. Do not tighten too much. To install a second syringe, do steps 5 through 7 again.
12

Installation
3.4 Install the sensor storage tubes
Put the three sensor storage tubes into the holder that is on the side of the instrument (refer to Product overview on page 7). Keep the sensor in a storage tube when not in use.
3.5 Install the stir bar and the beaker
Add the stir bar to the beaker, and then attach the beaker to the sensor holder.
3.6 Prepare the tubes
Remove any bends in the end of the tubes. 13

Installation
3.7 Connect the tubes
Arrow symbols identify the inlet and outlet ports for the syringe and the pump connections. The “up” arrow is the outlet port. The “down” arrow is the inlet port. Turn the tube connectors on the inlet and outlet ports of the syringe and pump until they click. The syringe outlet tube has a blue ring on it. If anti-diffusion tips are necessary, remove the pre-installed outlet tube from the syringe and install the tube from the application kit with the pre- installed anti-diffusion tip. Push the outlet tubes into the tube holder slots so that they are correctly attached.
3.8 Install the sensor
3.8.1 Install the legacy adapter
Go to Connect the sensor on page 15 if no legacy adapter is included in the application kit. 1. Connect the measuring, reference and temperature sensors to the legacy adapter. 2. Make sure that the instrument display shows the Home screen. Connect the legacy
adapter cable to a sensor socket on the rear panel of the instrument.
14

Installation

1 Temperature sensor

2 Reference sensor

3 Measuring sensor

3. The legacy adapter commissioning wizard automatically starts. Follow the instructions that show on the display. Select the parameter based on the connected sensor.

Option

Description

pH

Select this parameter if the connected sensor is a pH analogic sensor.

Metal/RedOX/Color Select this parameter if the connected sensor is a Pt-Pt (metallic) analogic sensor or PTM450/OPT300 sensor.

ISE

Select this parameter if the connected sensor is an ion selective

sensor.

Refer to the application documentation to enter the related information for the selected parameter.

3.8.1.1 Set the legacy adapter settings Do the steps that follow to set a sensor name in the legacy adapter settings or set the
legacy adapter settings.

1. Make sure that the legacy adapter is connected to the instrument.
2. From the home screen, select Settings>Legacy settings and then push Select.
3. If more than one legacy adapter is connected, select the legacy adapter to change the settings.
4. Select an option2. Refer to the application notes for additional information.

Option Input sensor name
Input manual temperature Input pHIso Reset settings

Description
Enters the name of the sensor used. If specified, follow the recommendations on the application notes for the sensor name.
Sets the temperature used in the titration if no temperature sensor is connected.
(pH parameter only) Sets the pH ISO value (default=7).
Sets the legacy adapter settings to the default configuration. Starts the legacy adapter commissioning wizard.

3.8.2 Connect the sensor
Use a conical adapter to hold the sensor tightly in the sensor holder.

2 The options shown are based on the parameter selected on the legacy adapter commisioning wizard. 15

Installation

Connect the sensor to an available sensor port on the rear of the instrument. After the sensor is connected, make sure that the sensor icon shows in the banner at the top of the display.
NOTICE
Make sure that the sensor tip is 5 to 10 mm above the top of the magnetic stir bar to prevent any contact with the bar during operation.
Refer to the illustrated steps that follow.

3.9 Install the titrant and the reagent
CAUTION
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols.
16

Installation
CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and national regulations.
NOTICE
This is only applicable to instruments with two syringes installed. It is recommended to install the applications (refer to Install the applications on page 30) before installing the titrant. Applications using titrant from syringe 1 are loaded on line 1 of the home screen (refer to Home screen on page 25) and applications using titrant from syringe 2 are loaded on line 2. After the applications are installed, the correct titrant can be connected to the applicable syringe. Optional: Fill a desiccant cartridge with an applicable desiccant. Put the desiccant cartridge into the adapter on the titrant bottle cap. Refer to the illustrated steps that follow, step 1. Loosen the tube connector on the bottle cap. Push the inlet tube through the connector. Make sure that the end of the tube is at the bottom of the bottle. Tighten the connector on the bottle cap. Use the same procedure to connect the second titrant bottle if a second syringe is installed on the instrument. Refer to the “Application Note” on the USB applications key to identify the correct pump to connect to the reagent bottle. Refer to the illustrated steps that follow.
17

Installation
3.10 Tidy the work area
Attach the tubes to the instrument with the clips on the electrovalve and the sensor holder. Refer to the illustrated steps that follow.
18

Section 4 Install accessories

4.1 Install an external pump

Table 3 External pump specifications (subject to change without notice)

Specification

Details

Dimensions (W x D x H)

11.4 x 11.4 x 10.0 cm (4.49 x 4.49 x 3.94 in.)

Weight

280 g (0.62 lb)

Power requirements

12 VDC, 0.4­0.54 A

Altitude

2,000 m (6,562 ft) maximum

Operating temperature

15 to 35 °C (59 to 95 °F)

Relative humidity

20 to 80%, non-condensing

Storage temperature

­5 to 40 °C (23 to 104 °F)

Installation category

II

Pollution degree

2

Certifications

EMC IEC/EN 61326-1

Warranty

1 year (EU: 2 years)

NOTICE
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including, without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent permitted under applicable law. The user is solely responsible to identify critical application risks and install appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire section before unpacking, setting up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this equipment in any manner other than that specified in this manual.
Read all labels and tags attached to the equipment. Personal injury or damage to the equipment could occur if not observed. A symbol on the equipment is referenced in the manual with a precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.

Products marked with this symbol indicates that the product contains toxic or hazardous substances or elements. The number inside the symbol indicates the environmental protection use period in years.

This accessory (item no. LZE142) is used to remove any excess sample from the beaker to control the volume of sample before titration. The inlet and outlet tubes are preinstalled on the pump. Make sure that the tube from the outlet port of the pump is installed in an applicable liquid waste container. A blue tag is attached to the inlet tubes that extract the liquid from the beaker.
To install the pump, refer to the illustrated steps that follow. After installation, check that the pump and tubes are correctly installed by selecting Maintenance > Pump activation from the instrument home screen (refer to Pump activation on page 43).

19

Install accessories
To adjust the height of the tube in the sample, unscrew the top of the fitting and slide the metal tube up or down to the required height before retightening the fitting. Refer to Figure 3. 20

Figure 3 Tube adjustment

Install accessories

4.2 Install an external propeller
This accessory (Item no. LZE143) is used as an alternative to the standard magnetic stir bar, for more viscous samples.
1. Install the propeller into the slot in the middle of the sensor holder. 2. Connect the propeller to the external propeller port on the rear of the instrument (refer
to Instrument connections on page 8). 3. From the home screen, select Settings>Options. Set the option Propeller Stirrer.
Note: Push the up and down arrow keys to highlight the option. Push the left arrow to set the option ON/OFF then push Done to confirm.
4.3 Install a balance
This accessory is used to automatically send accurate weights to the instrument. The balance is a laboratory type balance and uses the RS232 communication protocol with no flow control. When a balance is connected, the window shown at the end of a titration that lets users enter a weight, shows the value from the balance. The balance automatically sends the weight and unit to the instrument in textual format:
· Messages end with , or <LF+CR> · The numeric value of the weight must be transmitted before the unit of weight · Allowed units of weight are g, mg and kg · The unit of weight specifies the end of the data string used by the instrument · Space characters are ignored
1. Set the RS232 parameters on the balance to:
· 9600 baud · no parity · no handshake · 8 data bits · 1 stop bit 2. Connect the balance to the serial port on the rear of the instrument (refer to Instrument connections on page 8) using an applicable cable. 3. From the home screen, select Settings > Options and make sure that the Balance option is checked.
21

Install accessories
4.4 Install PC software
The optional PC software connects with the TitraLab® workstation through an Ethernet network. The software can be used to: · Control a TitraLab® workstation to start and stop analyses · Control a sample changer to start and stop analyses · Control a balance for accurate sample weighing · Show run-time data directly from the workstation · Manage data stored locally or on a server (search, compare, delete, print, etc.) · Export data to files for use in other software applications
Full product documentation and on-line help is available with the software.
4.5 Install the syringe verification kit
The optional syringe verification kit enables the verification of the dispensing system based on the ISO8655 standard. The test adds different volumes of liquid on a vessel put on a balance. The weights are used to calculate the real volume added. The real volume is compared with the nominal volume. The results are used to calculate the error in the volume added and the syringe precision. There are three verification modes available:
· Fully automated–The balance is connected to the instrument. The system makes the additions and calculates the results.
· Partially automated–The balance is not connected to the instrument. The user does the balance readings and the system calculates the results.
· Dispense only–The system only makes the additions. The user does the balance readings and calculates the results.
The syringe verification kit includes all of the necessary items: dispensing tips, tubes, syringe holder, pipette holder, support stand, software key and label with the license number. Make sure to keep the label with the license number. A license key is necessary to enable the syringe verification option on the instrument. The first time the option is selected, the instrument shows the MAC address and the Firmware version. Send the MAC address, firmware version and the license number to the manufacturer or a sales representative to get the license key. Note: The license key is only necessary when the option is selected for the first time. Items to collect:
· Analytical balance with resolution 0.1 mg, repeatability 0.2 mg and uncertainty 0.2 mg · Temperature sensor · Deionized water “quality 3” (ISO 3696)
Prepare the syringe verification kit as follows:
1. Put an empty beaker on the balance. Refer to the illustrated steps that follow. 2. For the fully automated mode, connect the balance to the instrument. 3. Install the support stand near the balance. Put the pipette tip adapter above the
beaker on the balance. 4. Connect the tube from the pipette holder to the syringe outlet port. 5. Connect the tube from the water bottle to the syringe inlet port. 6. On the instrument, manually enter the temperature. As an alternative, connect and
put a temperature sensor on the water bottle to get the temperature automatically. Note: The temperature setting is necessary for the fully automated and partially automated modes. The dispense only mode does not adjust the temperature.
22

Install accessories
23

Install accessories
24

Section 5 User interface and navigation
5.1 Keypad

1 Power 2 Printer Key Power Printer
Selection keys (contextual) Navigation keys
Home

3 Selection keys 4 Navigation keys

5 Home key

Description
Sets the instrument power to on or off. Push the key for 2 seconds to set the power to off.
The printer key only operates if a printer is connected to the instrument. Sends the data currently shown on the display to a connected printer when pushed. A sound is heard if the current display cannot be printed. A graph is automatically printed at the end of the measurement, if the option is selected (Settings > Options).
Used to select options shown above them in the footer bar. The options available are applicable to the current operation (e.g. calibration, measurement, etc.).
Scrolls through menus and data, enter numbers and letters, enter checkbox settings and set options for the syringe and the pump.
Push this key at any time to go directly to the home screen. A sound is heard if the key is disabled (e.g. during a calibration or measurement).

5.2 Home screen

1 Available options from this screen
2 If two syringes are installed, shows applications for syringe 2

5 Screen name 6 Information icons (refer to Table 4)

3 Applications for syringe 1 4 Screen icon

7 Option available by pushing the selection key below 8 Arrow keys available for use in the screen

Table 4 shows the information icons that can be shown in the header bar.

25

User interface and navigation

Icon Description

Table 4 Information icons

A printer is connected to the instrument

A USB key is connected to the instrument

A sensor is connected to sensor port 1

A sensor is connected to sensor port 2

The data log file is full. Refer to Manage the data log on page 39 for the options available to manage the data log file.

A measurement is in operation using the PC software. The keyboard is locked.

5.3 Menu structure

Option

The three primary menus are shown in Table 5, Table 6 and Table 7. Table 5 Home menu
Description

Data log

Manage the data log file

Calibration

Electrode and titrant calibration procedures

Settings

See the table entry below

Maintenance

See the table entry below

Purge Option
Applications

Remove air from the syringe, and from the syringe and pump tubes
Table 6 Settings menu Description Manage the installed applications

Operators Date & Time

Set up and manage the instrument operators Set the date and time

26

Option Brightness

User interface and navigation
Table 6 Settings menu (continued) Description
Adjust the display brightness

Sounds

Set the sound on or off for specific operations

Language

Set the operating language

Network

Give a name to identify the instrument

Legacy settings

Specify the legacy sensor data

Show instrument and attached hardware information Info

Restore defaults

Set the instrument to the default configuration

Options

Set the operating parameters

Security Option
Syringe activation

Set a password and specify which options are password protected
Table 7 Maintenance menu Description Make sure that the syringe is working correctly

Pump activation

Make sure that the pump is working correctly

Stirring activation

Make sure that the stirrer is working correctly

Live measure

Quickly check measurement data for connected sensors. Not available for all sensors

Syringe management

Replace the syringe or change the syringe volume

27

User interface and navigation

Option

Table 7 Maintenance menu (continued) Description

Replace the pump cassette for installed pumps Pump cassette replacement

Reagent replacement

Use this option when replacing reagents

Maintenance schedule Maintenance summary

Manage the maintenance tasks
See the number of days remaining on specified devices before maintenance is required

Syringe verification

When enabled, starts the syringe verification procedure.

28

Section 6 Startup

CAUTION
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols.
CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and national regulations.

CAUTION
Personal injury hazard. Never use the instrument without the syringe cover installed.

6.1 Configure the instrument

1. From the home screen, select Settings. Refer also to Menu structure on page 26. 2. Select an option, then push Select.

Option Applications
Operators Date + Time Brightness Sounds Language Network

Description
Changes (refer to Change the application settings on page 38), makes copies, exports and removes application data. Make sure that the duplication function does not make more than five applications for each syringe installed.
Adds, changes and removes operators.
Sets the instrument date and time.
Sets the brightness of the display.
Sets the sound options.
Sets the language.
Give a name to the instrument. This name is used to connect the instrument to a PC. If a printer is connected, this name is printed on the hard-copy output. Restart the instrument if the name is changed.

Legacy settings Specify the sensor data when the legacy adapter is used.

Info

Shows information about the instrument and the attached hardware.

Restore Defaults

Sets the instrument to the default configuration. The restored settings are:
· The language is set to English · All password security is removed · The security password is set to 0000 · No sound on: Key press, Information and Analysis results OK · Sound on: Instrument ready, Instrument switch off, Errors and
Analysis results not OK · The application parameters view is set to basic mode · Sets the syringe to empty when the instrument is set to off · The temperature display unit is set to °C · No print of the measurement curve · No print of the derivative curves · No balance is connected · No propeller stirrer is connected · The default stirring is set to 10% · The user is set to “Default user”

29

Startup

Option Options
Security
3. Push Back.

Description
Sets the following options:
· Set the application parameters view to expert or basic mode · When the instrument is set to off, sets the syringe to empty into the
titrant bottle · Changes the temperature display unit · Prints the measurement and derivative curves if a printer is connected · Specify if a balance is connected · Specify if an external propeller stirrer is connected.
Note: When the instrument is set to on, the application parameters view is always set to basic mode. Changes made to application data with the instrument in expert mode are kept.
Change the password. Specify which of these options are password protected:
· Settings (all items except Info) · Maintenance schedule and summary · Delete data log · Export data log · Electrode calibration · Titrant calibration · Sample volume calibration · Syringe verification test

6.2 Install the applications
Use the supplied USB key to install the applications. The instrument can install a maximum of five applications for each syringe installed. For two syringes, the installed applications shown on the top line of the display refer to syringe one and the installed applications shown on the second line refer to syringe two.
If any errors occur during installation, refer to Troubleshooting on page 49.
1. Push Home to go to the main menu. 2. Connect the USB key to the USB port on the side of the instrument. The applications
on the USB key show on the display.
3. Push the arrow keys to highlight and select an application to install. Push the left or right arrow key to select it. Do this step again to select additional applications to install.
4. Push Import to install the selected applications. 5. Push OK to complete the installation. The installed applications show on the home
screen. Note: To install more applications, push Home to go to the home screen, then remove the USB key and reconnect it.

6.3 Prepare the instrument for measurement
1. From the home screen, select Purge. All attached devices are listed. 2. Select All elements to purge all the attached devices, or select one device to purge.
Push Select. Air is removed from the device and filled with liquid from the bottle. 3. Push OK when the operation has completed. 4. Make sure that there are no air bubbles in the device. Do step 2 again if there are any
air bubbles.
30

Startup
5. Select the next device to purge if individual devices are being selected. 6. Push Exit when all the tubes are filled with reagent and the device has no air
bubbles. Note: If a few small air bubbles can be seen on the inner wall and/or piston of the syringe, they can be left without effecting system performance.
31

Startup
32

Section 7 Standard operations

CAUTION
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.

CAUTION
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols.
CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and national regulations.

CAUTION
Personal injury hazard. Never use the instrument without the syringe protection cover in place.
CAUTION
Chemical exposure hazard. Never remove the stir bar from the beaker before the end of a titration.

7.1 Calibration

7.1.1 Calibrate the sensor

1. From the home screen, select Calibration, then push Electrode calibration.
2. If more than one sensor is installed, push the up and down arrow keys to highlight the sensor to use, then push Select.
3. If more than one application includes calibration parameters for the sensor, push the up and down arrow keys to highlight the application to use, then push Select. Calibration information shows on the display.
4. If necessary, select an icon for more information or to change some data.

Option

Description

Electrode

Shows more information about the sensor.

Operator

Changes the operator ID. Select from a list of applicable operators.

Buffer or Standard Shows more information about the buffer set or standard.

5. Do the instructions that show on the display, then push Start to start the calibration. Calibration data shows on the display.
6. If the default stirring speed needs to be adjusted, push the up and down arrow keys to increase or decrease the speed. Note: This adjustment only applies to the current operation. The standard default stirring speed for the calibration is not changed.
7. Select Stop at any time to stop the calibration. Results are then calculated from the calibration data available before Stop is selected.

33

Standard operations

8. For pH sensors only:

Option Description

Yes Continue with the next calibration buffer solution in the sequence.

No

Stop the calibration. The calibration can still be validated if at least one buffer

calibration was successful.

9. When the calibration is complete, push the left and right arrow keys to see the different measurement views.
10. Push Reject or Validate.

Option Reject

Description
Select Cancel to go back to the result display or Confirm to reject the calibration and use the default or previous calibration value.

Validate The calibration is accepted and the new values stored.

7.1.2 Calibrate the titrant

1. From the home screen, select Calibration, then push Titrant calibration.
2. If more than one titrant is installed, push the up and down arrow keys to highlight the titrant to calibrate, then push Select.
3. If more than one application contains a titrant calibration method, push the up and down arrow keys to highlight the application to use, then push Select.
4. Calibration information shows on the display. If necessary, select an icon for more information or to change some data.

Option

Description

Information Shows more information about the calibration.

Operator

Changes the operator ID. Select from a list of applicable operators.

5. Fill a beaker with the recommended standard amount that shows on the display. If necessary, add more of the solvent specified in the application note until the sensor is correctly installed in the sample.
6. Carefully put a magnetic stir bar into the beaker. Make sure there is no liquid spill.
7. Attach the beaker to the sensor holder.
8. Make sure that the icon at the bottom of the display is highlighted. Do the instructions that show on the display adjacent to this icon. Refer to Connect the sensor on page 15 to make sure that the tubes and sensor are correctly aligned.
9. Push Start to start the calibration. Calibration data shows on the display.
10. If the default stirring speed needs to be adjusted, push the up and down arrow keys to increase or decrease the speed. Note: This adjustment only applies to the current operation. The standard default stirring speed for the application is not changed.

34

Standard operations

11. Two options are available during the procedure:

Option Description

Stop Aborts the calibration and no results are calculated. If selected during the Replicate Sample option, then all data in the series is lost.

Skip

Stops the current operation and goes directly to the next step in the procedure. Results are then calculated from calibration data available before Skip is selected. Results can be less accurate using this option.

12. Push Reject or Continue.

Option Description

Reject

Rejects the calibration. Select Cancel to go back to the result display or Confirm to reject the calibration. If this is the first calibration, select Confirm to reject the calibration and use the default or previous calibration values. If this is a Replicate Standard calibration, select Confirm to reject only the current calibration in the series.

Continue Select from one of these options:

· Replicate Standard: Do the calibration again using the same standard
· Save & Exit: Keep the calibration results and exit the calibration procedure
· Reject & Exit: Reject the calibration results and use the default or previous calibration values, and exit the calibration procedure

7.1.3 Auto leveling calibration
This option is only available when at least one installed application contains an auto leveling calibration method and the method is set to active. The calibration makes sure that the sample volume in the measurement cell is the same before each titration.
An external pump must be installed for this procedure.

1. From the home screen, select Calibration, then push Autoleveling calibration.
2. If more than one application contains an auto leveling calibration method, push the up and down arrow keys to highlight the application to use, then push Select.
3. Calibration information shows on the display. If necessary, select an icon for more information or to change some data.

Option

Description

Information Shows more information about the calibration.

Operator

Changes the operator ID. Select from a list of applicable operators.

4. Make sure that the external pump and tubes are correctly installed (refer to Install an external pump on page 19).
5. Make sure that the sensor, tubes and stir bar are correctly installed in an empty beaker, as if preparing for a standard measurement.
6. Accurately measure the necessary volume of sample and pour it into the empty beaker.
7. Adjust the position of the tube from the external pump (refer to Figure 3 on page 21) so that it touches the surface of the liquid. Then add more sample until the end of the tube is immersed in 5 to 10 mm of the liquid.
8. Push Start to start the calibration.
9. The leveling procedure starts and sample is removed for the preset time. Make sure that this time is large enough for all the necessary sample to be removed and for air to be removed for some seconds at the end.

35

Standard operations

10. The titration measurement starts after the leveling procedure completes. If the default stirring speed needs to be adjusted, push the up and down arrow keys to increase or decrease the speed. Note: This adjustment only applies to the current operation. The standard default stirring speed for the calibration is not changed.
11. Two options are available during the calibration procedure:

Option Description
Stop Aborts the calibration and no results are calculated. If selected during a Replicate Standard calibration, then all data in the series is lost.
Skip Stops the leveling operation and goes directly to the next step in the procedure. Results are then calculated from the data available before Skip is selected.
12. When the calibration is complete, push the left and right arrow keys to see the different measurement views. The result of the titration is the calculated volume of sample.
13. Push Reject or Continue.

Option Description

Reject

Rejects the calibration. Select Cancel to go back to the result display or Confirm to reject the calibration. If this is the first calibration, select Confirm to reject the calibration and use the default or previous calibration values. If this is a Replicate Standard calibration, select Confirm to reject only the current calibration in the series.

Continue Select from one of these options:

· Replicate Standard: Do the calibration again using the same sample
· Save & Exit: Keep the calibration results and exit the calibration procedure
· Reject & Exit: Reject the calibration results and use the default or previous calibration values, and exit the calibration procedure

14. If the calibration is successful, the calculated volume of sample is kept in the application parameters. This value is shown in the measurement start view for that
application and is used in the next titration calculations of the application.

7.2 Sample preparation
When this application is installed (refer to Install the applications on page 30), use it to prepare one or more measurement samples. Note: If the PC software is installed (refer to Install PC software on page 22) and the AS1000 sample changer is installed, sample preparation can be done with or without leveling. But if the sample preparation is done without the software and sample changer, then leveling must be deactivated (Active = No). It is recommended to read the related “Application Note” from the USB applications key for more instructions.
1. From the home screen, select the sample preparation application, then push Select. 2. Make sure that the reagent used to prepare the sample is installed with the correct
pump. Note: The application parameters can be changed in basic or expert mode. Refer to Change the application settings on page 38. 3. Obey the instructions on the display. 4. Push Next to prepare more samples or Exit to go back to the home screen.

7.3 Get a sample measurement
Use this option to get sample measurements with one of the installed applications.

36

Standard operations

1. From the home screen, select the measurement application, then push Select. Application information shows on the display.
2. Read the related “Application Note” from the USB applications key for more instructions.
3. If necessary, select an icon for more information or to change some data.

Option

Description

Information Shows more information about the application.

Operator Changes the operator ID. Select from a list of applicable operators.

Sample

Sample Name: Change the specified name of the sample.
Type: Push the left and right arrow keys and select the sample type (Sample, QC or Define blank) to be used for the measurement. If Define blank has been selected before, two more sample types are available (QC with blank and Sample with blank).

4. Fill a beaker with the recommended sample amount that shows on the display. If necessary, add more of the solvent specified in the application note until the sensor is correctly installed in the sample.
5. Carefully put a magnetic stir bar into the beaker. Make sure there is no liquid spill.
6. Attach the beaker to the sensor holder.
7. Make sure that the icon at the bottom of the display is highlighted. Do the instructions that show on the display adjacent to this icon. Refer to Connect the sensor on page 15 to make sure that the tubes and sensor are correctly aligned.
8. Push Start to start the measurement. Measurement data shows on the display.
9. If the default stirring speed needs to be adjusted, push the up and down arrow keys to increase or decrease the speed. Note: This adjustment only applies to the current operation. The standard default stirring speed for the application is not changed.
10. Two options are available during the procedure:

Option Description

Stop Aborts the measurement and no results are calculated. If selected during the Replicate Sample option, then all data in the series is lost.

Skip

Stops the current operation and goes directly to the next step in the procedure. Results are then calculated from measurement data available before Skip is selected. Results can be less accurate using this option.

11. When the measurement is complete, push the arrow keys to see the different measurement views.
12. Push Next for these options:

Option
Replicate Sample

Description
Use this option to start the same titration on the same sample. This is used to study the repeatability by successively analyzing several parts of the same sample. At the end of each measurement, a window shows the average value, the standard deviation and the relative standard deviation.

New Sample Use this option to start the same titration on a new sample. No standard deviation or relative standard deviation measurements will be done.

13. Push Exit to go back to the home screen.

37

Standard operations
7.4 Change the application settings
NOTICE
The applications installed on the instrument have been pre-defined to optimize the measurement process. Changing these default application parameters will have an effect on the measurement process and measurement results. Only qualified personnel should change these parameters or use the instrument in Expert mode. Refer to the application note before changing any parameters.
NOTICE
In Basic mode changes to parameters are secure. In Expert mode changes to parameters are not secure and can cause the application to run incorrectly.
Refer to Change application parameters on page 63 for a list of parameters that can be changed in basic and expert mode. Refer to Sample name on page 38 for special features on the sample name.
1. From the home screen, select Settings. 2. Select Applications followed by Edit. 3. If more than one application is installed, choose an application to change from the
installed list and then push Edit. 4. Push the left and right arrow keys to jump to the previous or next parameter group.
Push the up and down arrow keys to scroll through the application parameters in the parameter group. 5. The Edit key is only available when a parameter can be changed. Push Edit to change the parameter. Enter the new details or select from a list. Note: A description of the selected parameter is shown on the bottom of the display.
7.4.1 Sample name
The two wildcard characters ? (question mark) and (asterisk) can be used to automatically increment the sample name. Use the question mark character to increment the name using an integer. For example, where Water is the name of the sample:
· Select Settings > Applications > Edit and change the sample name to Water ? · The sample name for the first measurement is Water 1 · The sample name for the next measurement is Water 2, etc. · The integer value is always set back to 1 when the instrument is set to off
Use the asterisk character in the name to increment the name using the instrument date and time in YYMMDDHHMMSS format. For example, where Water is the name of the sample:
· Select Settings > Applications > Edit and change the sample name to Water
· The sample name for the first measurement made on 27th February 2015 at
15:30:25 is Water 150227153025 · The sample name for the next measurement made on 28th February 2015 at
16:32:47 is Water 150228163247, etc. · The value is not reset when the instrument is set to off
Use both characters in the name to increment the name using an integer and the instrument date and time. For example, where Water is the name of the sample:
· Select Settings > Applications > Edit and change the sample name to Water ?

  • · The sample name for the first measurement made on 27th February 2015 at
    15:30:25 is Water 1 150227153025
    38

Standard operations

· The sample name for the next measurement made on 27th February 2015 at 17:01:27 is Water 2 150227170127, etc.
· The integer value is always set back to 1 when the instrument is set to off

7.5 Manage the data log

To select data to view, delete or export, specify data filters

1. From the home screen, select Data log. 2. Select an option, then push Select.

Option

Description

View data log Views measurement data. Select individual lines of data to view more content.

Export data log Exports measurement data from the system to an external device. Preview data selection before it is exported. Make sure that an external device is connected to the instrument (e.g., a USB key, external hard drive, etc.).

Delete data log Removes measurement data from the system. Previews data selection before it is removed.

3. Specify the data selection parameters. Push the left and right arrow keys to make a selection. Push the up and down arrow keys to select an option.

Option

Description

Result type

Sets the type of result available.

Application

Sets the available applications.

Date

Sets the date range.

Operator

Sets the available operators.

4. If View data log was the selected option, push View to see the selected data.

· Push the up and down arrow keys to select a line of data and push Detail to see more data
· If Electrode calibration is selected as the Result type, push the left and right arrow keys to select the sensor if more than one sensor is installed. Push the up and down arrow keys to select a line of data, then push the left and right arrow keys to see related graphs
· If only one Application is selected push the up and down arrow keys to select a line of data and push Detail to see more data, or push the left and right arrow keys to see related graphs
5. If Export data log or Delete data log was the selected option, push Preview to see the selected data, then push Export or Delete to start the procedure.

7.6 Purge

Use this procedure to remove air bubbles from the system. Refer to Prepare the instrument for measurement on page 30 for instructions.

39

Standard operations
40

Section 8 Maintenance

CAUTION
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.
NOTICE
Do not disassemble the instrument for maintenance. If the internal components must be cleaned or repaired, contact the manufacturer.

8.1 Maintenance schedule

Table 8 shows the recommended schedule of maintenance tasks. Facility requirements and operating conditions may increase the frequency of some tasks.

Table 8 Maintenance schedule

Task

1 day 7 days 12 months

Clean the instrument on page 41. Also do this procedure immediately after any liquid spills on the instrument.

x

Clean the sensor on page 41.

x

If applicable, Replace desiccant cartridge contents on page 42.

x

Replace the syringe. Refer to Syringe management on page 43.

x

Replace syringe electro-valve block. Refer to Replace the syringe electro- valve block on page 42.

x

Examine the syringe connections and inlet and outlet tubes for leaks and damage. Replace as

necessary. Refer to Replacement parts and accessories on page 57 for part numbers and to

x

Replace the tubes on page 42.

Replace the pump cassettes. Refer to Pump cassette replacement on page 44.

x

Examine the pump connections and inlet and outlet tubes for leaks and damage. Replace as

necessary. Refer to Replacement parts and accessories on page 57 for part numbers and to

x

Replace the tubes on page 42.

Examine the bottle caps and connections for leaks and damage. Replace as necessary. Refer to Replacement parts and accessories on page 57 for part numbers.

x

8.2 Clean the instrument
NOTICE
Never use flammable or corrosive solvents to clean any part of the instrument. Use of these solvents can degrade the environmental protection of the instrument and may void the warranty.
Clean the exterior surface with a moist cloth or with a mixture of water and mild detergent. Dry with a soft cloth.
8.3 Clean the sensor
Obey the cleaning instructions given in the documentation supplied with the sensor, and any other instructions given in the application note.

41

Maintenance
8.4 Replace the tubes
After installing a new tube, make sure that it is correctly installed by turning the tube connectors on the inlet and outlet ports of the syringe or pump until they click. Refer also to Connect the tubes on page 14.
8.5 Replace desiccant cartridge contents
When used, it is recommended to replace the desiccant cartridge contents weekly, but this will vary depending on the humidity of the laboratory. A humidity indicator, such as color changing silica gel crystal, can be used to show when replacement is necessary.
8.6 Replace the syringe electro-valve block
DANGER
Electrocution hazard. Remove power from the instrument before this procedure is started.
NOTICE
It is recommended that this procedure is only done by a qualified service engineer. Refer to the illustrated steps that follow.
8.7 Maintenance menu
Select Maintenance from the home screen. 42

Maintenance

Refer also to Table 7 on page 27.
8.7.1 Syringe activation
Do a check of the syringe. Make sure the syringe fills and empties correctly.

1. From the maintenance menu, push Syringe activation. Push the arrow keys to select an option.

Option

Description

Fill

Fills the syringe with titrant solution. The process stops automatically when

the syringe is full.

Empty to bottle Discards the contents of the syringe into the titrant bottle.

Empty to beaker Discards the contents of the syringe into the beaker. Make sure that the tube from the outlet port of the syringe is inside the beaker.

Stop

Stops the operation.

If a second syringe is installed, the Toggle option is shown on the display. Push Toggle to change to the second syringe, or to operate both syringes.
8.7.2 Pump activation
This option only applies to instruments with pumps installed. Do a check of the pump. Make sure the pump fills and empties correctly.
1. From the maintenance menu, push Pump activation. Push the up and down arrow keys to select an option.
Option Description
Start Starts the pump. The reagent is pumped through the tubes into the beaker. Make sure that the tube from the outlet port of the pump is in the beaker.
Stop Stops the operation.

If more than one pump is installed, the Toggle option is shown on the display. Push Toggle to change to the second pump, to an external pump or to operate all pumps.

8.7.3 Syringe management
Use this option to replace the syringe, or change the syringe volume.

1. From the maintenance menu, push Syringe management.

Option Syringe replacement
Syringe volume change

Description
If two syringes are installed, select the syringe to replace. Select to discard the contents of the syringe into the titrant bottle or the beaker, and then obey the instructions on the display. Refer also to Install the syringe on page 12 for the installation.
If two syringes are installed, select the syringe to change. Select the new syringe volume.

8.7.4 Syringe verification
Use the syringe verification option to examine the syringe accuracy based on the ISO 8655 standard. For the best results, follow the procedure in the application note supplied with the syringe verification kit. Note: A license key is necessary to enable the syringe verification option on the instrument. Contact the manufacturer or a sales representative to get a license key.
1. From the home screen, select Settings, then push Select. 2. From the settings screen, select Options, then push Select.

43

Maintenance

3. Push the left or right arrow key to select the expert mode, then push Done.
4. From the home screen, select Maintenance, then push Select.
5. From the maintenance screen, select Syringe verification, then push Select. Note: When the option is selected for the first time, enter the license key.
6. Push the up and down button to select and change the syringe test options.

Option

Description

Syringe

Selects the syringe. Used when there are two syringes. Shows the syringe information: syringe number and volume.

Operator

Selects the operator ID

Configuration Sets the test settings: test volume and number of replicates. The total dispensed liquid for the test shows when the two parameters are set. Note: Make sure to use the applicable beaker and balance for the test based on the amount of total dispensed liquid.

A field to write the test conditions is available in the test conditions settings.

Start

Starts the syringe verification procedure.

7. Push Start to start the syringe verification procedure. Follow the procedure supplied with the syringe verification kit.
8. To cancel the syringe verification test and not save data to the data log, push Stop.
9. To cancel the syringe verification test and save the syringe verification results for fully tested volumes to the data log, push Skip.
10. When the test is complete Pass or Fail shows. Push the left or right arrow to show the results for the different volumes. Push the up and down arrow to show more results data for the selected volume.

8.7.5 Pump cassette replacement

DANGER

Electrocution hazard. Remove power from the instrument before this procedure is started.

This option only applies to instruments with pumps installed.

44

Maintenance
To replace the pump cassette, refer to the illustrated steps that follow and obey the instructions on the display.

8.7.6 Other maintenance options

1. Select an option, then push Select.

Option

Description

Stirring activation Do a check of the magnetic stirrer. Push the up and down arrow keys to increase or decrease the stirring speed.

45

Maintenance

Option Live measure
Reagent replacement Maintenance schedule Maintenance summary

Description
This option is not available for all sensors. The option shows continuous measurement data with connected sensors to quickly check measurements. The installed applications and the automatic additions to the sample are set to off. Continuous measurements are not temperature compensated, so measurement differences may occur in the same sample between continuous measurements and measurements that use installed applications with temperature compensation.
Use this option to replace the reagents. Obey the instructions on the display.
See the list of maintenance tasks. Push Edit to change the default value.
See the number of days remaining for maintenance tasks. After doing a task, push Reset to set the number of days remaining to the default value.

8.8 Storage and transportation
8.8.1 Prepare the instrument for storage
CAUTION
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols.
CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and national regulations.

1. Remove the sensor from the sensor holder and disconnect it from the sensor port on the rear of the instrument.
2. Clean the sensor. Refer to Clean the sensor on page 41.
3. Put the sensor in a storage tube that contains a storage solution. Refer to the sensor documentation for storage instructions.
4. From the home screen, select Maintenance > Syringe activation and empty the contents of the syringe.
5. If a second syringe is installed, push Toggle to select it and empty the contents.
6. Set the instrument power to off.
7. Carefully remove the tubes from the titrant bottle. If pumps are installed, remove the tubes from the reagent bottles. Seal the bottles with the original bottle cap.
8. Flush the tubes with water or an applicable cleaning solution (refer to the application note for more information).
9. Drain the used beakers and clean with deionized water.
10. If storing the instrument for more than a week, remove the syringe from the instrument (refer to Install the syringe on page 12). Flush the syringe with water or an applicable cleaning solution and store in its box.

46

Maintenance 8.8.2 Prepare the instrument for short-term storage
CAUTION
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols.
CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and national regulations. Use this procedure if storing the instrument overnight or over a weekend. 1. Carefully clean the measurement cell using an applicable organic solvent to dissolve any residue and oil. 2. Fill the measurement cell with a new working medium. 3. Add titrant to the working medium until it becomes brown in color. 4. Check that the measurement cell is tightly installed. 5. Check the position of all the elements in the sensor holder. 6. Set the instrument power to off.
8.8.3 Prepare the instrument for shipment
1. Refer to Prepare the instrument for storage on page 46. 2. Put the instrument and the necessary accessories in a package with all the packaging
materials. 3. Ship the instrument at a temperature ­5 to 40 °C (23 to 104 °F) and relative humidity
up to 80%. 4. Make sure that the storage location for the instrument does not have dust,
condensation or chemical evaporation.
47

Maintenance
48

Section 9 Troubleshooting

Refer to the following tables for problem messages and possible causes and solutions.

Table 9 Measurement

Error/Warning

Description

Possible solution

Measurement unstable

The measurement from the sensor is not stable

Make sure that there are no bubbles around the sensor element or junction

Measurement out of range

The sample concentration is outside the sensor measurement limits

Adjust the sample preparation to be inside the sensor measurement range
Use a different sensor with a more applicable measurement range

Temperature out of range

The sample temperature is outside the sensor measurement limits

Adjust the sample temperature to be inside the sensor measurement range
Use a different sensor with a more applicable measurement range

Make sure that the sensor is operating correctly

Max. ordinate reached

The titration is stopped because the measured signal is above or below the ordinate stop point set for this application, depending on the curve direction

Refer to the application note for best practices and do the titration again
Adjust the maximum ordinate for the application (refer to Change application parameters on page 63)

Below minimum range

This message is specific to amperometric applications; the equivalence point has not been found

Select an application range more adapted to the sample

Above maximum range

This message is specific to amperometric applications; the equivalence point has not been found

Select an application range more adapted to the sample

This message is specific to amperometric Insufficient curve resolution applications; the equivalence point has not
been found

Refer to the application note for best practices and do the titration again
Clean the sensor using the recommended procedures in the sensor documentation

Not enough excess titrant

This message is specific to amperometric applications with back titration; the equivalence point has not been found

Refer to the application note for best practices and do the titration again
Check the excess titrant defined in the application file and adjust if necessary or select an application range more adapted to the sample

Too much excess titrant

This message is specific to amperometric applications with back titration; the equivalence point has not been found

Refer to the application note for best practices and do the titration again
Check the excess titrant defined in the application file and adjust if necessary or select an application range more adapted to the sample

Flat signal

The titration is stopped because the amount of sample is below the detectable limit of the application

Refer to the application note for best practices and do the titration again
Select an application range more adapted to the sample

Analysis skipped

The titration has been stopped manually by the user; the equivalence point has not been found

Preferably use an automatic end titration detection instead of a manual skip

49

Troubleshooting

Table 9 Measurement (continued)

Error/Warning

Description

Possible solution

Refer to the application note for best practices and do the titration again

Equiv. point not found

The titration fails to determine the equivalence point

Adjust the equivalence point detection value for the application (refer to Change application parameters on page 63)

Contact technical support

Out of range

The computed result is outside the accepted limits

Refer to the application note for best practices and do the titration again
Adjust the result minimum and maximum values for the application (refer to Change application parameters on page 63)

Equation calc. error

An equation variable is unknown (e.g. the ordinate of the equivalence point is unknown or the volume of the equivalence point is unknown)

Make sure that all the equation variables are available (refer to the application note for the equation description and values)

A required electrode is not connected!
Application {application name}
Electrode: Type: {type}:
Name: {name}:

The applicable IntelliCAL electrode type for Connect the IntelliCAL electrode specified in the

the application is not connected.

application note.

The “Sensor name” set in the legacy adapter Change the “Sensor name” setting in the legacy

settings does not agree with application

adapter setting to agree with the sensor name

settings.

specified in the application note.

Electrode calibration has expired

A titration or titrant calibration has been done with an expired sensor calibration date

Calibrate the sensor and if necessary adjust the calibration frequency (refer to the application note for the recommended frequency)

Titrant calibration has expired

A titration has been done with an expired titrant calibration date

Calibrate the titrant and if necessary adjust the calibration frequency (refer to the application note for the recommended frequency)

The titrant used by an application has been modified and this application became incompatible with applications already installed. Please go to application management to solve the issue

The application data has been changed and the application cannot be used

Remove the application or change the titrant to a titrant that is already installed

System error

A system error causes the titration to stop (e.g. a sensor has been disconnected or the volume of reagent is too low)

Make sure that all the system elements are connected and working correctly (refer to Maintenance menu on page 42)

Max. vol. reached

Refer to the application note for best practices

The titration is stopped because the supplied and do the titration again

volume of reagent is more than the maximum Adjust the maximum volume stop point for the

quantity

application (refer to Change application

parameters on page 63)

No titrant information

There is no description of the titrant

Install the application again

50

Troubleshooting

Error/Warning
No measure received
The stirrer/propeller stirrer does not operate
Printer: communication error Burette fails to deliver Burette fails to fill Burette failed to read delivered volume System failure

Table 9 Measurement (continued)

Description

Possible solution

Make sure that the sensor is correctly connected to the instrument

No measurement can be done

Disconnect the sensor and connect it again
Change the sensor if no measurement is received using the Live Measure option (refer to Other maintenance options on page 45)

The stirrer (or the propeller stirrer) does not automatically start when the measurement starts.

Make sure that the stirrer configuration is set correctly. Identify if the applicable stirrer is selected in the Settings>Options menu.

The red LED on the printer flashes to show that the printer is in error mode. There is no paper in the printer.

Replace the paper roll in the printer. Refer to the printer documentation.

Syringe cannot deliver any reagent

Contact technical support

Syringe cannot be filled with reagent

Contact technical support

Syringe cannot calculate the delivered volume of reagent

Contact technical support

Operational failure

Contact technical support

Table 10 Calibration

Error/Warning

Description

Possible solution

Temperature variation too high

The variation in temperature is too high across all the calibration solutions when more than one is used

Make sure that the maximum temperature variation across all the calibration solutions is no more than 5 °C before starting a calibration

Temperature out of range

The temperature of the calibration solution is Make sure that the temperature of the calibration outside the specified temperature range for solution is inside the specified range before the buffer or standard used in the calibration starting a calibration

Calibration solution already The calibration solution has already been

used

used for a calibration

Use another calibration solution

Slope out of range

The pH sensor calibration slope is outside the accepted limits

Make sure that there are no bubbles around the sensor element or junction
Clean the sensor using the recommended procedures in the sensor documentation

Offset out of range

The pH sensor calibration offset is outside the accepted limits

Make sure that there are no bubbles around the sensor element or junction
Clean the sensor using the recommended procedures in the sensor documentation

Buffer value under electrode detection limit

The value of the calibration solution is less than the low limit for sensor detection

Use a calibration solution that is inside the sensor measurement range

Cell constant out of range

The calibration value of the cell constant is outside the accepted limits

Make sure that the cell constant minimum and maximum values are correct for the application and adjust them if incorrect (refer to Change application parameters on page 63)

The connected electrode cannot be calibrated or there is no application that contain calibration parameters

There is no installed application that contains calibration parameters, or the sensor is not a sensor that can be calibrated (e.g. PtPt or ORP sensors)

Make sure that at least one application that contains calibration parameters for the sensor has been installed, and that the sensor is a sensor that can be calibrated (e.g. pH or conductivity sensors)

51

Troubleshooting

Error/Warning No titrant to calibrate

Table 10 Calibration (continued)

Description

Possible solution

There is no method to calibrate the titrant into the application. If more than one application is installed, there is no method to calibrate the titrant into all the applications.

The titrant cannot be calibrated with the current application. Refer to the related application note to find if the titrant calibration is available.

The titrant calibration has been deactivated (Basic or Expert mode: Active = No)

Activate the titrant calibration (Active = Yes)

Table 11 USB storage device

Error/Warning

Description

Possible solution

Exporting {application name} access forbidden to the applications folder on USB flash drive

Error when exporting data to the USB storage device

Make sure that the USB storage device is correctly installed and the USB icon can be seen in the header bar (refer to Table 4 on page 26)
Remove the USB storage device and make sure the USB icon is no longer seen in the header bar, then install the device again and make sure that the icon can be seen (refer to Table 4 on page 26)

Exporting {application name} read only access to the applications folder on USB flash drive

The USB storage device is read only

Unlock the storage device or use another one

Exporting {application name} error during writing. Verify USB flash drive is not full

Error when exporting data to the USB storage device

Make sure that the USB storage device has enough available space or use another one

USB flash drive is not installed. Install and export applications again

The instrument cannot find a USB storage device

Make sure that the USB storage device is correctly installed and the USB icon can be seen in the header bar (refer to Table 4 on page 26)
Remove the USB storage device and make sure the USB icon is no longer seen in the header bar, then install the device again and make sure that the icon can be seen (refer to Table 4 on page 26)

Application {application name}: maximum number of applications is reached

The maximum number of applications for each line has been reached when copying applications to the instrument from the USB storage device

Remove any unwanted applications, then push the Home key to return to the home screen. Remove the USB storage device and make sure the USB icon is no longer seen in the header bar, then install the device again and import the new application again

52

Troubleshooting

Table 11 USB storage device (continued)

Error/Warning

Description

Possible solution

Application {application name}: requires at least one titrant that cannot be installed

The titrant needed by the application cannot be installed on the instrument because the applications currently installed use another titrant

Single syringe instrument:
Remove all installed applications, then push the Home key to return to the home screen. Remove the USB storage device and make sure the USB icon is no longer seen in the header bar, then install the device again and import the new application again
Two syringe instrument where the application uses only one titrant:
Remove all installed applications from the line of applications using that titrant and then push the Home key to return to the home screen. Remove the USB storage device and make sure the USB icon is no longer seen in the header bar, then install the device again and import the new application again
Two syringe instrument where the application uses two titrants:
· If the two titrants are not already used by the installed applications, remove all installed applications, then push the Home key to return to the home screen. Remove the USB storage device and make sure the USB icon is no longer seen in the header bar, then install the device again and import the new application again
· If one of the two titrants is already installed, remove all installed applications from the line of applications using the other titrant and then push the Home key to return to the home screen. Remove the USB storage device and make sure the USB icon is no longer seen in the header bar, then install the device again and import the new application again

Application {application name}: requires too many titrants for this instrument model configuration

The instrument has a single syringe and the application needs a two syringe instrument

Change to a two syringe instrument

Error while saving application {application name}

Error when copying from the USB storage device to the instrument memory

Contact technical support

Unable to read applications from device. Applications have been removed, please load them from USB flash drive

While updating the application tiles on the display, an installed application has been detected as corrupt

Remove the corrupt application and try to load it again

Application {application name} The selected application file is

corrupted

invalid

Try to load the original application file again, or contact technical support for a replacement file

Application {application name} The hardware does not support

incompatible with hardware

the application

Change to a compatible instrument

Application {application name} (too new model)

This selection is not allowed because the instrument software only supports older application files

Update the instrument software

53

Troubleshooting

Error/Warning Application {application name} (untrusted)
Application {application name} (obsolete)
Application {application name} (syringe to replace)
Error while serializing data
Cannot write to USB mass storage

Table 11 USB storage device (continued)

Description

Possible solution

This selection is not allowed because the application has not been authenticated

Contact technical support to get an authenticated application file

The selection is allowed. When the The automatic update will set the new application

application is exported, it will

parameters to the default values. It is not

automatically update to the latest recommended to use an obsolete application, but to

version

install the latest version

The selection is allowed but the syringe recommended in the application file is not the one installed on the instrument

Change the installed syringe to the one recommended by the application (refer to Syringe management on page 43)

Error when creating a file on the USB storage device

Change or format the USB storage device and try the export again

Remove the USB storage device and make sure the USB storage device is not installed USB icon is no longer seen in the header bar, then
install the device again try the export again

USB storage device is corrupt

Change or format the USB storage device and try the export again

Error/Warning
Edit user failed. Be sure that name {user name} is not already used
Fail to save the modification of the application
Impossible to remove application {application name}

Table 12 Data updates Description The update to the operator name cannot be done The update to the application data cannot be done
The application cannot be erased

Error while deleting data

The data cannot be erased

Possible solution
Check that the name is not already used
Try again, if the problem continues, contact technical support Try again, if the problem continues, contact technical support Try again, if the problem continues, contact technical support

Table 13 Battery

Error/Warning

Description

Possible solution

Battery voltage low

Battery voltage is below 2.2 V. The date and time are no longer available

Contact technical support

Set the date and time again. If the date and time Realtime clock has been lost The date and time are no longer available are not available the next time the instrument is
set to on, contact technical support

Table 14 Balance

Error/Warning

Description

Possible solution

Format of message received from the balance is not supported

The data received from the balance is not compatible with the format of messages accepted by the instrument

Configure the balance to be compatible with the format of messages accepted by the instrument (refer to Install a balance on page 21)

No connection with balance

The instrument is not receiving data from the balance

Check the balance cable is installed correctly

The measurement received is out The weight sent by the balance is not in the Adjust the sample. Adjust the acceptance

of range

acceptance limits

limit for the application

54

Troubleshooting

Table 15 Syringe verification

Error/Warning

Description

Possible solution

Format of message received from the balance is not supported

The data received from the balance is not compatible with the format of messages accepted by the instrument (fully automated mode only).

Configure the balance to be compatible with the format of messages accepted by the instrument (refer to Install a balance on page 21)
Make sure that the weight measurement is stable and did not get to the maximum weight.

No connection with balance

The instrument is not receiving data from the balance (fully automated mode only).

Make sure that the balance cable is installed correctly.

Temperature sensor not detected

The instrument is not receiving data form the temperature sensor (fully and partially automated modes only)

Make sure that the temperature sensor is installed and connected correctly.

Incorrect input value

The input value is out of range (fully and partially automated modes only).

Make sure that the syringe verification configuration values are correct.

No operator recorded

The operator value is not defined on the syringe verification screen.

Set an operator on the syringe verification configuration.

Serialization error

There was an error during the results export.

Remove the USB storage device and make sure the USB icon is no longer seen in the header bar, then install the device again try the export again. Try again, if the problem continues, contact technical support.

USB mass storage writting Error when exporting data to the USB

error

storage device

Make sure that the USB storage device is correctly installed and the USB icon can be seen in the header bar (refer to Table 4 on page 26).

55

Troubleshooting
56

Section 10 Replacement parts and accessories

WARNING

Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the instrument or equipment malfunction. The replacement parts in this section are approved by the manufacturer.

Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to the company website for contact information.

Table 16 Replacement parts and accessories

Description

Item no.

Syringe 2.5 mL, glass Syringe 5 mL, glass Syringe 10 mL, glass Syringe 25 mL, glass

LZE100 LZE101 LZE102 LZE103

O-ring for 2.5 mL syringe O-ring for 5 mL syringe O-ring for 10 mL syringe O-ring for 25 mL syringe

LZE104 LZE105 LZE106 LZE107

Set of beakers 50 mL (5x) Set of beakers 150 mL (5x)

LZE108 LZE109

Delivery tubing with anti-diffusion tip

LZE112

Set of tubes (2x) for syringe, suction & delivery
Set of tubes (2x) for pump, suction & delivery

LZE114 LZE115

57

Replacement parts and accessories

Table 16 Replacement parts and accessories (continued)

Description

Item no.

Bottle stopper GL25 Bottle stopper GL45 Bottle stopper S40 Bottle stopper GL32

LZE116 LZE117 LZE118 LZE186

Glass bottle, 1L

LZE119

Desiccant tube empty with cotton
Desiccant tube filled with molecular sieve

LZE120 LZE121

Conical adapter NS14.5/23
Conical adapter, large cut (for 7.5 mm diameter)

LZE122 LZE182

Tube fitting (2x)

LZE124

Spare pump cassette

LZE125

58

Replacement parts and accessories

Table 16 Replacement parts and accessories (continued)

Description

Item no.

Magnetic stir bar, PTFE, 20 x 6 mm (5x) LZE136

Tube holder, 4 positions

LZE141

Sample leveling pump

LZE142

Propeller stirrer, 70 mm shaft

LZE143

Power supply (without power cable)

LZE144

Power cable (EU)

YAA080

59

Replacement parts and accessories

Table 16 Replacement parts and accessories (continued)

Description

Item no.

Power cable (US)

XLH055

Power cable (UK)

XLH057

RS-232 adapter cable

LZE145

Syringe protection cover

LZE156

Electro-valve block and connector

LZE165

Sensor storage tubes

LZE185

60

Replacement parts and accessories

Table 16 Replacement parts and accessories (continued)

Description

Item no.

RS-232 cable for balance connection to LZE196 TitraLab

Thermo-Drucker USB printer
AT1000/KF1000 syringe volume verification kit

LQV161.99.10000 LZE228

Table 17 Electrodes and electrode accessories
Description
MTC301, platinum combined electrode, 1 m cable MTC306, silver combined electrode, 1 m cable MTC695, Pt-Pt electrode, 1 m cable PHC705, pH probe, red rod, glass, general purpose, standard PHC725, pH probe, glass PHC805, pH probe, glass, general purpose CDC401 conductivity probe, standard, 1 m cable ISECa calcium combined selective Legacy adapter OPT300 photocolorimetric probe PTM450, NP photocolorimetric titration module, without power pHG311-9 pH electrode, pH 0-14, FX pHG301-9 pH electrode, glass REF251 reference electrode, red rod, double junction, banana CL114 cable FX/COAX/1m/BNC REF361 reference electrode, Ag/AgCl, reverse sleeve, FX Adaptor BNC male to banana REF451 reference electrode, calomel, double junction, FX ISE25Cu-9 ion selective, copper, FX CL111 cable FX/COAX/1m/banana

Item no.
MTC30101 MTC30601 MTC69501 PHC70501 PHC72501 PHC80501 CDC40101 LZW9660C.97.002
LZV662 E71T001 X61T005 E11M004 E11M001 E21M001 A94L114 E21M003 LZE184 E21M005 E41M006 A94L111

61

Replacement parts and accessories
62

Appendix A Change application parameters

Parameter Name Advisable syringe

NOTICE
The applications installed on the instrument have been pre-defined to optimize the measurement process. Changing these default application parameters will have an effect on the measurement process and measurement results. Only qualified personnel should change these parameters or use the instrument in Expert mode. Refer to the application note before changing any parameters.

NOTICE
In Basic mode changes to parameters are secure. In Expert mode changes to parameters are not secure and can cause the application to run incorrectly.

NOTICE
The Active parameter is available to change in some parameter groups. It specifies if the method should be used during the analysis. Changing this from Yes to No means that the method will not be used which will have an effect on the measurement process and measurement results.

NOTICE
The Hide result parameter (e.g. R2 hide, R3 hide, etc.) is available to change in some parameter groups. It specifies if the result is shown on the display. Changing this from Yes to No means that the calculated measurement result will not be shown. This will not have an effect on the measurement process and measurement results.

The parameters available for change can be different for each instrument configuration and the application being edited. The parameters marked with an “x” are available for change.

Table 18 Application

Description

Basic mode Expert mode

Name of the application

x

x

The application has been designed with this type of syringe

Parameter Name Amount Unit Min. amount Max. amount Resolution

Table 19 Sample Description Name of the sample Amount of the sample Unit of the sample Minimum permitted sample amount Maximum permitted sample amount Number of decimals to show for the sample amount

Basic mode x x

Expert mode x x x x x x

Parameter Name

Description Name of the QC

Table 20 QC Basic mode x

Expert mode x

Parameter Description

Description Description of the blank

Table 21 Blank

Basic mode x

Expert mode x

63

Change application parameters

Table 22 Electrode

Parameter

Description

Basic mode Expert mode

Type

Type of measurement electrode

Part of name

Part of the name used to identify matching electrodes if the preferred electrode is not connected

Recommended electrode Recommended electrode to be used for this analysis

x

Calibration frequency

Recommended frequency of calibration (0 = no reminder)

x

x

Stability criterion

Refer to the working procedure for the recommended value

x

Max. stability time

Maximum time limit for the stability check

x

Stirring speed

Speed of the stirrer during electrode calibration

x

x

pH electrode parameters

Stirring duration

Stirring duration before buffer measurement (0 = buffer

x

measurement with stirring)

Max. temp. variation

Maximum permitted temperature variation during calibration

Min. slope

Minimum permitted slope limit for the calibration result

x

Max. slope

Maximum permitted slope limit for the calibration result

x

Min. offset

Minimum permitted offset limit for the calibration result

x

Max. offset

Maximum permitted offset limit for the calibration result

x

Calibration mode

Set the electrode calibration mode

x

Buffer set

List of available buffers for electrode calibration

x

x

Number fixed buffers

The number of fixed buffers to use

Fixed buffer #n

Buffer to be used during calibration

x

Conductivity electrode parameters

Calibration mode

Set the electrode calibration mode

x

Standard

Standard used for electrode calibration

x

x

Min. cell constant

Minimum permitted cell constant for the calibration result

x

Max. cell constant

Maximum permitted cell constant for the calibration result

x

Temp. compensation

Follow Natural Water Standards ISO/DIS_7888

x

Linear temp. coeff.

Coefficient for linear temperature compensation mode

x

Reference temp.

Reference temperature for linear temperature compensation mode

x

PtPt electrode parameters

Working mode

Specify if the probe is used in imposed voltage/current (AC or DC)

x

Current (sample analysis) Set the value of the imposed current during sample analysis

x

Current (titrant calibration) Set the value of the imposed current during titrant calibration

x

Voltage (sample analysis) Set the value of the imposed voltage during sample analysis

x

Voltage (titrant calibration) Set the value of the imposed voltage during titrant calibration

x

64

Change application parameters

Parameter

Description

Table 23 Titrant

Name

Name of the titrant

Titrant concentration Nominal concentration of the titrant

Resolution

Number of decimals of the nominal concentration of the titrant

Unit

Unit of the titrant

Real concentration Real concentration of the titrant used for result calculation

Resolution

Number of decimals of the real concentration of the titrant

Location

The syringe that contains the titrant

Titrant calibration

Active

Select No to skip this method during titrant calibration

Calibration frequency Recommended frequency of titrant calibration (0 = no reminder)

Stirring speed

Speed of the stirrer during titrant calibration

Incremental addition

Predose type

Type of the predose (in volume or in ordinate)

Predose volume

Titrant volume added at the beginning of the titration

Predose ordinate

Ordinate measurement to reach to stop the predose

Predose speed

Speed in mL/min during the predose in ordinate

Delay

Stirring time before the titration starts

Max. vol. stop point Stops the titration when this volume is reached

Ordinate stop point Stops the titration when this ordinate is reached

Stop on last EQP

Stops the titration on the last equivalent point detection

Min. ordinate

Contact technical support before changing. The minimum ordinate value is used in the regulation algorithm during the dynamic addition

Max. ordinate

Contact technical support before changing. The maximum ordinate value is used in the regulation algorithm during the dynamic addition

Min. angle

Contact technical support before changing. The minimum tangent angle value is used in the regulation algorithm during the dynamic addition

Max. angle

Contact technical support before changing. The maximum tangent angle value is used in the regulation algorithm during the dynamic addition

Min. stability time

Minimum time after an increment addition

Max. stability time Maximum time after an increment addition

Stability criterion

Stability criterion after an increment addition

Increment size

Increment size when the addition is monotonic

Min. increment size Minimum increment volume when the addition is dynamic

Max. increment size Maximum increment volume when the addition is dynamic

EP detection

EP1 ordinate

Ordinate of this end point

Basic mode
x

Expert mode
x x x x x x

x

x

x

x

x

x

x x x x x x x x x
x
x
x
x x x x x x

x

65

Change application parameters

Parameter

Description

Table 23 Titrant (continued)

IP detection

Derivative filter

Contact technical support before changing. Number of points used for building the derivative

Detection threshold Contact technical support before changing. The minimum value of the derivative for inflexion point detection

IP1 min. ordinate

Minimum permitted value for this inflexion point

IP1 max. ordinate

Maximum permitted value for this inflexion point

Angle detection

Maximum abscissa

Contact technical support before changing. The maximum abscissa is used in the algorithm for equivalence calculation

Curve filter

Contact technical support before changing. Number of points used for building the titration curve

Curvature filter

Contact technical support before changing. Number of points used for the curvature calculation

Detection threshold Contact technical support before changing. The minimum value of the curvature for equivalence detection

Linearity threshold

Contact technical support before changing. The minimum value of the curvature used in the algorithm for equivalence calculation

EQP min. ordinate Minimum permitted value for this equivalence point

EQP max. ordinate Maximum permitted value for this equivalence point

Stop on flat signal Stops titration when a flat signal is detected

Derivative filter

Contact technical support before changing. Number of points used for building the derivative used in flat signal detection

Min. derivative

Contact technical support before changing. Minimum threshold on derivative for flat signal detection

Max. derivative

Contact technical support before changing. Maximum threshold on derivative for flat signal detection

Derivative number

Contact technical support before changing. Number of derivative values between the minimum and maximum thresholds to stop the titration on a flat signal

Titration calculation

Resolution

Maximum number of decimals to show for the real concentration of the titrant

Titrant cal. limit check Activate the titrant calibration result boundary check

Min. titrant conc.

Minimum value of the real concentration of the titrant

Max. titrant conc.

Maximum value of the real concentration of the titrant

Titrant cal. EQP index Index of the equivalence point used for the titrant calibration calculation

Standard name

Name of the standard

Standard amount

Amount of standard to be used for titer determination

Unit

Unit of the standard amount

Min. amount

Minimum amount of the standard

Basic mode

Expert mode

x x x x

x x x x x x x x x x x x

x
x x x x x x x x

66

Change application parameters

Parameter

Description

Table 23 Titrant (continued)

Max. amount

Maximum amount of the standard

Concentration

Concentration value of the standard

Unit

Unit of the concentration value of the standard

Resolution

Number of decimals of the concentration value of the standard

Standard equivalents Number of exchanged equivalents of the standard (H+; e-; …)

Titrant equivalents Number of exchanged equivalents of the titrant (H+; e-; …)

Molar weight

Molar weight of the standard

Basic mode
x

Expert mode
x x x x x x x

Parameter
Active Time Stirring speed Message
Active Reagent Pump ID Time Stirring speed
Active Time
Solution name Solution concentration Unit Resolution
Active Message Stirring speed
Active Message Stirring speed

Table 24 Methods Description
Agitation method Select No to skip this method during analysis Duration of the stirring Speed of the stirrer during this method Information shown during the analysis
Auxiliary addition method Select No to skip this method during analysis Name of the reagent to add Pump identifier Duration of the pump activation Speed of the stirrer during the addition
Leveling method Select No to skip this method during analysis Duration of the pump activation
Autoleveling calibration Name of the solution used for autoleveling calibration Concentration value of the standard Unit of the concentration value of the standard Number of decimals for the concentration value of the standard
Message method Select No to skip this method during analysis Information shown during the analysis Speed of the stirrer during this method
Input value method Select No to skip this method during analysis Information shown during the analysis Speed of the stirrer during this method

Basic mode Expert mode

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

67

Change application parameters

Parameter Resolution Hide Min. value Max. value
Active Max. stability time Stability criterion Stirring speed Result 1 (R1) name R1 max. resolution R1 hide R1 limit check R1 min. R1 max. R1 QC min. R1 QC max.
Stirring duration Pump ID Addition duration Preparation duration
Addition
Active Stirring speed Measured parameter Predose type Predose volume Predose ordinate Predose speed Delay Addition mode Min. ordinate

Table 24 Methods (continued)

Description

Basic mode Expert mode

Maximum number of decimals for the input value

The result is not shown if Yes is selected

x

x

Minimum permitted input value

x

Maximum permitted input value

x

Measurement method

Select No to skip this method during analysis

x

x

Time limit for the stability check

x

Refer to the working procedure for the recommended value

x

Speed of the stirrer during the measurement

x

x

Name of the result

x

x

Maximum number of decimals to show for the result

x

The result is not shown if Yes is selected

x

x

Activate the sample result boundary check

x

Minimum permitted result value

x

Maximum permitted result value

x

Minimum permitted QC result value

x

Maximum permitted QC result value

x

Sample batch preparation method

Duration of the stirring during the addition

x

x

Pump identifier

x

x

Duration of the addition of reagent by the pump

x

x

Duration of the sample batch preparation

x

x

Excess titrant addition method

Excess addition done immediately at application start or just before titration

x

x

Analysis titration method

Select No to skip this method during analysis

x

x

Speed of the stirrer during titration

x

x

Parameter measured by the probe

Type of the predose (in volume or in ordinate)

x

Titrant volume added at the beginning of the titration

x

x

Measurement ordinate to reach to stop the predose

x

x

Speed in mL/min during the predose ordinate

x

Stirring time before the titration starts

x

x

Titrant addition mode (monotonic or dynamic)

x

Contact technical support before changing. The minimum ordinate

x

value is used in the regulation algorithm during the dynamic

addition

68

Change application parameters

Table 24 Methods (continued)

Parameter

Description

Basic mode Expert mode

Max. ordinate

Contact technical support before changing. The maximum ordinate

x

value is used in the regulation algorithm during the dynamic

addition

Max. vol. stop point

Stops the titration when this volume is reached

x

Ordinate stop point

Stops the titration when this ordinate is reached

x

Stop on last EQP

Stops the titration on last equivalent point detection

x

Min. angle

Contact technical support before changing. The minimum tangent

x

angle value is used in the regulation algorithm during the dynamic

addition

Max. angle

Contact technical support before changing. The maximum tangent

x

angle value is used in the regulation algorithm during the dynamic

addition

Min. stability time

Minimum time after an increment addition

x

Max. stability time

Maximum time after an increment addition

x

Stability criterion

Stability criterion after an increment addition

x

Increment size

Increment size when the addition is monotonic

x

Min. increment size

Minimum increment volume when the addition is dynamic

x

Max. increment size

Maximum increment volume when the addition is dynamic

x

Back titration mode

Excess addition mode

x

Excess vol.

Volume of the excess titrant

x

Excess titrant name

Name of the excess titrant

Excess titrant conc.

Nominal concentration of the excess titrant

x

Resolution

Number of decimals of the concentration value of the excess titrant

x

Excess titrant conc. unit Unit of the excess titrant concentration

x

Real conc. of excess titrant Real concentration of the excess titrant

x

x

Excess equivalents

Number of exchanged equivalents of the excess (H+; e-; …)

x

Sample equivalents

Number of exchanged equivalents of the sample (H+; e-; …)

x

Titrant equivalents

Number of exchanged equivalents of the titrant (H+; e-; …)

x

EP detection

EP1 ordinate

Ordinate of this end point

x

IP detection

Derivative filter

Contact technical support before changing. Number of points used

x

for building the derivative

Detection threshold

Contact technical support before changing. The minimum value of

x

the derivative for inflexion point detection

IP1 min. ordinate

Minimum permitted value for this inflexion point

x

IP1 max. ordinate

Maximum permitted value for this inflexion point

x

Angle detection

Maximum abscissa

Contact technical support before changing. The maximum

x

abscissa is used in the algorithm for equivalence calculation

69

Change application parameters

Parameter Curve filter
Curvature filter
Detection threshold
Linearity threshold
EQP min. ordinate EQP max. ordinate Stop on flat signal Derivative filter
Min. derivative
Max. derivative
Derivative number
Result x (Rx) name Rx max. resolution Rx hide Rx limit check Rx min. Rx max. Rx QC min. Rx QC max. Rx unit Rx EQP index Rx molar weight Rx sample equivalents Rx titrant equivalents
Result x (Rx) name Rx max. resolution Rx hide Rx limit check Rx min.

Table 24 Methods (continued)

Description

Basic mode Expert mode

Contact technical support before changing. Number of points used

x

for building the titration curve

Contact technical support before changing. Number of points used

x

for the curvature calculation

Contact technical support before changing. The minimum value of

x

the curvature for equivalence detection

Contact technical support before changing. The minimum value of

x

the curvature used in the algorithm for equivalence calculation

Minimum permitted value for this equivalence point

x

Maximum permitted value for this equivalence point

x

Stops titration when a flat signal is detected

x

Contact technical support before changing. Number of points used

x

for building the derivative used in flat signal detection

Contact technical support before changing. Minimum threshold on

x

derivative for flat signal detection

Contact technical support before changing. Maximum threshold on

x

derivative for flat signal detection

Contact technical support before changing. Number of derivative

x

values between the minimum and maximum thresholds to stop the

titration on a flat signal

Titration calculation

Name of the result

x

x

Maximum number of decimals to show for the result

x

The result is not shown if Yes is selected

x

x

Activate the sample result boundary check

x

Minimum permitted result value

x

Maximum permitted result value

x

Minimum permitted QC result value

x

Maximum permitted QC result value

x

Unit of the result

x

Index of the equivalence point used for the result calculation

x

Molar weight of analyte used for the result calculation

x

Number of exchanged equivalents of the sample (H+; e-; …)

x

Number of exchanged equivalents of the titrant (H+; e-; …)

x

Titration calculation using an equation

Name of the result calculated with the equation

x

x

Maximum number of decimals to show for the result

x

The result is not shown if Yes is selected

x

x

Activate the sample result boundary check

x

Minimum permitted result value

x

70

Parameter Rx max. Rx QC min. Rx QC max. Rx equation Rx unit Rx user value

Change application parameters

Table 24 Methods (continued) Description Maximum permitted result value Minimum permitted QC result value Maximum permitted QC result value Equation used for the result shown Select the result unit Value that may be used in the equation computation as Fx

Basic mode Expert mode x x x

x

x

x

x

71

Change application parameters
72

Appendix B Frequently asked questions

Table 25 Application files

Question

Cause

Procedure

Why can’t I load any applications

Wrong file placement on the USB storage device
The user is not in the home screen

Make sure that the file path on the USB storage device is correct
Push the Home key to make sure that you are in the home screen before connecting the USB storage device

Why is the instrument not recognizing the applications USB storage device, even though I have removed and reconnected the device and I see the USB icon in the header bar

Reconnecting the device has been done too quickly

Remove the device and wait until the icon disappears from the header bar, then push the Home key to make sure that you are in the home screen before reconnecting the device

Question
Why does the volume delivered not correspond to the volume displayed

Cause

Table 26 Titration

The volume of the installed syringe is different from the syringe configuration

Procedure
Use the recommended syringe for the application and make sure the volume is correct

Table 27 Mechanical

Question

Cause

Procedure

Why is the equivalent point not being detected at the expected volume

Air bubbles are trapped in the tubes

Select the Purge option on the instrument to purge the tubes of air bubbles (refer to Prepare the instrument for measurement on page 30)

How can I remove an air bubble that is trapped on top of the syringe

Air bubbles are trapped in the syringe

Select the Purge option on the instrument to purge the syringe of air bubbles (refer to Prepare the instrument for measurement on page 30)

Why is titrant leaking along the sealing above the electrovalve

The tubes have not been correctly installed

Turn the tube connector until a “click” is heard

Why is titrant leaking along the sealing below the electrovalve

The syringe has not been installed tight enough
The electrovalve is leaking

Make sure that the syringe is installed correctly
Change the electrovalve as described in Replace the syringe electro-valve block on page 42

Why is the titrant not being delivered into the tubes

The titrant bottle is empty or the tube is not correctly installed in the bottle

Make sure that the titrant bottle is installed correctly

The electrovalve is blocked Change the electrovalve as described in Replace

or defective

the syringe electro-valve block on page 42

On my two syringe instrument, why is the titrant not being delivered into the beaker but back into the titrant bottle

The procedure is not being done on the correct syringe

Select the Syringe activation option on the instrument, and Toggle the syringes to make sure that the correct syringe is being used in the
application

The active syringe is not the correct syringe

Erase all the installed applications and install them again. The applications on line 1 will be allocated to syringe 1 and the applications on line 2 will be allocated to syringe 2. Make sure the titrant and installation is correct for the syringe (refer to Install the titrant and the reagent on page 16)

73

Frequently asked questions

Question
On my two pump instrument, using the Pump activation option, why is the reagent not being delivered into the beaker

Table 27 Mechanical (continued)

Cause

Procedure

The procedure is not being done on the correct pump

Select the Pump activation option on the instrument, and Toggle the pumps to make sure that the correct pump is being used

The active pump is not the Change to expert mode and change the name of

correct pump

the pump

The pump is defective

Change the pump as described in Pump cassette replacement on page 44

Table 28 Password Question How can I retrieve my password to access expert mode

Procedure Contact technical support

74

Section 13 Configure the instrument

1. From the home screen, select Settings. Refer also to Menu structure on page 26. 2. Select an option, then push Select.

Option

Description

Applications

Changes (refer to Change the application settings on page 38), makes copies, exports and removes application data. Make sure that the duplication function does not make more than five applications for each syringe installed.

Operators Date + Time Brightness Sounds Language

Adds, changes and removes operators. Sets the instrument date and time. Sets the brightness of the display. Sets the sound options. Sets the language.

Network

Give a name to the instrument. This name is used to connect the instrument to a PC. If a printer is connected, this name is printed on the hard-copy output. Restart the instrument if the name is changed.

Legacy settings Specify the sensor data when the legacy adapter is used.

Info

Shows information about the instrument and the attached hardware.

Restore Defaults

Sets the instrument to the default configuration. The restored settings are:
· The language is set to English · All password security is removed · The security password is set to 0000 · No sound on: Key press, Information and Analysis results OK · Sound on: Instrument ready, Instrument switch off, Errors and
Analysis results not OK · The application parameters view is set to basic mode · Sets the syringe to empty when the instrument is set to off · The temperature display unit is set to °C · No print of the measurement curve · No print of the derivative curves · No balance is connected · No propeller stirrer is connected · The default stirring is set to 10% · The user is set to “Default user”

75

Configure the instrument
Option Options
Security
3. Push Back.

Description
Sets the following options:
· Set the application parameters view to expert or basic mode · When the instrument is set to off, sets the syringe to empty into the
titrant bottle · Changes the temperature display unit · Prints the measurement and derivative curves if a printer is connected · Specify if a balance is connected · Specify if an external propeller stirrer is connected.
Note: When the instrument is set to on, the application parameters view is always set to basic mode. Changes made to application data with the instrument in expert mode are kept.
Change the password. Specify which of these options are password protected:
· Settings (all items except Info) · Maintenance schedule and summary · Delete data log · Export data log · Electrode calibration · Titrant calibration · Sample volume calibration · Syringe verification test

76

HACH COMPANY World Headquarters P.O. Box 389, Loveland, CO 80539-0389 U.S.A. Tel. 970-669-3050 800-227-4224 (U.S.A. only) Fax 970-669-2932 [email protected] www.hach.com

HACH LANGE GMBH Willstätterstraße 11 D-40549 Düsseldorf, Germany Tel. +49 (0) 2 11 52 88-320 Fax +49 (0) 2 11 52 88-210 [email protected] www.de.hach.com

HACH LANGE Sàrl 6, route de Compois 1222 Vésenaz SWITZERLAND Tel. +41 22 594 6400 Fax +41 22 594 6499

© Hach Company/Hach Lange GmbH, 2014­2016, 2018, 2019­2022. All rights reserved.

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

HACH User Manuals

Related Manuals