Spengler 513 700 Series Mobi Tensiometer User Manual
- June 14, 2024
- Spengler
Table of Contents
- Spengler 513 700 Series Mobi Tensiometer
- Product Information
- Product Usage Instructions
- FAQ (Frequently Asked Questions)
- SYMBOLOGY
- INTRODUCTION
- INSTRUCTIONS FOR USE
- INTENDED MEDICAL USE
- WHAT DOES “BLOOD PRESSURE” MEAN?
- WHAT IS “NORMAL” BLOOD PRESSURE?
- CHOOSE AN CUFF THAT SUITS THE PATIENT’S MORPHOLOGY
- USING THE MOBI ANEROID TENSIOMETER
- MAINTENANCE AND MAINTENANCE VERIFICATION
- CLEANING AND DISINFECTION
- TECHNICAL SPECIFICATIONS
- PRECAUTIONS FOR USE
- CONTRAINDICATION
- WARRANTY
- STANDARD
- QUALITY ASSURANCE
- CE MARKING
- ABOUT COMPANY
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Spengler 513 700 Series Mobi Tensiometer
Product Information
Specifications
- Product Name: Mobi Tensiometer
- Manufacturer: SPENGLER
- Model Numbers: 513 702, 513 700, 513 701, 513 703
- Measurement Method: Auscultatory method (requires a stethoscope)
- Measurement Range: Systolic Blood Pressure (SBP) and Diastolic Blood Pressure (DBP)
Product Usage Instructions
Step 1: Reading the Leaflet
Please read the provided leaflet before using the Mobi tensiometer. Keep
the leaflet for future reference.
Step 2: Intended Medical Use
The Mobi tensiometer is a professional medical diagnosis device used by
trained health professionals. It measures the pressure exerted by blood on
arteries using the auscultatory method. This method requires the use of a
stethoscope. The Mobi tensiometer accurately measures maximum systolic blood
pressure (SBP) and minimum diastolic blood pressure (DBP).
Step 3: Understanding Blood Pressure
Blood pressure is the pressure exerted by blood flow on the walls of
arteries. The highest pressure, known as systolic pressure, occurs when the
left ventricle contracts. The lowest pressure, known as diastolic pressure,
occurs when the ventricles relax. Blood pressure can fluctuate throughout the
day and can be influenced by factors such as stress, nutrition, physical
activity, medications, and diseases.
Step 4: Normal Blood Pressure
According to the World Health Organization (WHO), normal blood pressure is
defined as systolic pressure below 130mmHg and diastolic pressure below
90mmHg. However, individual blood pressure can vary. Please refer to the WHO
classifications table below for more information:
Blood Pressure Classification in Adults (mmHg)| Systolic Blood Pressure
(mmHg)| Diastolic Blood Pressure (mmHg)
---|---|---
Optimum| < 120| < 80
Normal| 120-129| 80-84
Upper Normal| 130-139| 85-89
Grade 1 HBP| 140-159| 90-99
Grade 2 HBP| 160-179| 100-109
Grade 3 HBP| >= 180| >= 110
Step 5: Warnings
The following warnings are important to ensure accurate measurements and
patient safety:
- Possible measurement error: Only use the Mobi tensiometer cuffs and accessories from SPENGLER to avoid measurement errors.
FAQ (Frequently Asked Questions)
-
Can I use any stethoscope with the Mobi tensiometer?
Yes, you can use any stethoscope with the Mobi tensiometer as long as it provides clear and accurate sound. -
Can I use the Mobi tensiometer for children?
The Mobi tensiometer is designed for use with adults. For pediatric patients, please consult a healthcare professional for appropriate blood pressure monitoring devices. -
How often should I calibrate the Mobi tensiometer?
It is recommended to follow the calibration instructions provided by SPENGLER. Calibration frequency may vary based on usage and specific requirements.
SYMBOLOGY
-
Manufacturer
-
See the directions for use
-
Product reference
-
Batch number
-
Protect from humidity
-
Protect from light and sun
-
Storage temperature range
-
Dispose of used product at a collection point for recycling in accordance with local regulations.
-
This product complies with directive 93/42/CEE regarding medical devices.
INTRODUCTION
- Thank you for purchasing a Mobi tensiometer.
- When you choose a SPENGLER instrument, you are choosing experience and excellence that has been recognized since 1907.
- Mobi is the result of several years of research and testing focused on providing health professionals with an instrument with unrivalled accuracy, design and practicality.
INSTRUCTIONS FOR USE
Please read the leaflet before use and keep it for future reference.
INTENDED MEDICAL USE
- The Mobi tensiometer is a medical diagnosis professional device (used by health professionals trained in its use) that measures the pressure exercised by blood on arteries by the auscultatory method (a stethoscope is therefore required).
- It accurately measures maximum systolic blood pressure or SBP (resulting from the left ventricle contracting) and minimum diastolic blood pressure or DBP (resulting from ventricles relaxing).
WHAT DOES “BLOOD PRESSURE” MEAN?
- Blood pressure is the pressure exercised by blood flow on the walls of arteries. The highest pressure, produced in the arteries by each heart beat, is called “systolic pressure” whereas the lowest is “diastolic pressure”.
- Blood pressure fluctuates several times a day and can also vary with stress, nutritional factors, physical activity, medicines or diseases.
WHAT IS “NORMAL” BLOOD PRESSURE?
According to the WHO (World Health Organization), normal blood pressure is systolic pressure less than 130mmHg (millimetres of mercury) and diastolic pressure below 90mmHg. However, blood pressure can vary depending on the individual. The WHO classifications table can be found below.
WARNINGS
The warnings in this document highlight conditions or practices which could
cause injury, disease or potentially death of a patient.
- Warning: Possible measurement error. Only use the tensiometer cuffs and Mobi accessories from SPENGLER to avoid measurement errors.
- Warning: Risk of inaccurate measurement. Before use, check that all the connection points, the cuff and tube are completely airtight. Excessive leakage can affected measurement.
- Warning: No changes can be made to this equipment as this could affect measurements.
CHOOSE AN CUFF THAT SUITS THE PATIENT’S MORPHOLOGY
SPENGLER is taking on the fight by informing health professionals in using an cuff that suits a patient’s morphology, as the size of the cuff impacts the quality of measurements:
- A too narrow cuff over-estimates blood pressure.
- A too large cuff could under-estimate blood pressure.
Note:
- The index marker showing the arterial index on the cuff must be within the range shown on Mobi cuffs.
- When the cuff is positioned on the patient’s arm, the “index” marker shown on the cuff must be within the range shown on Mobi cuffs. If the “index” marker or the edge of the marker does not reach this range, we recommend that you see the table below in order to adapt the cuff to a patient’s morphology.
SPENGLER innovation: With its magnetic connectors, the “Life Link®“ tube of the Mobi tensiometer means that the cuff can be changed effortlessly in one second.
AMBIDEXTROUS: CHANGE THE CONFIGURATION
The Mobi is especially designed to be ambidextrous, change the position of the
following parts:
- Decompression valve,
- Spoon
- LifeLink® connection
Instructions: changing the Life Link® connection
-
Unscrew the bulb and remove it from the device. With the decompression valve closed, unscrew the magnetic “Life Link®“ connector with a screwdriver. (Schema 1)
-
Remove the cartridge by pulling on the decompression valve. To make it easier, you may use a tumbling tool through the hole of the removed magnetic connector.
Warning: The cartridge includes an O-ring. Make sure that you do not lose or damage it during the process -
Before replacing the cartridge in the right position, keep pushing the bulb upwards in order to line up both openings enabling cartridge positioning. Replace the cartridge, making sure that holes are placed up and down. They will allow air flow between the bulb and the manometer.
-
When the cartridge is in place, you may screw the “Life Link®“ connector back.
Instructions: changing the bulb
- Unscrew the bulb and then remove it
- Unscrew the 2 screws and invert the bulb
- Tighten the 2 screws
- Tighten the bulb
USING THE MOBI ANEROID TENSIOMETER
Steps
- Leave the patient to rest for 5 minutes
- Place the patient in a comfortable position, with legs uncrossed, and the back and arm elevated, ideally without moving or speaking
- Check that the marker index for arterial index is within the range shown on the Mobi cuffs
- Keep the cuff at heart height during measurement
- Keep the pneumatic circuit closed (check the decompression valve)
- Inflate the cuff for an initial SBP estimation (when the blood circulates again, it causes a noticeable beating sound on the stethoscope) until radial pulse disappears (so-called Korotkoff sounds)
- Inflate again to 30mmHg above the estimated SBP
- Deflate by opening the valve at a speed of 2mmHg per second and simultaneously read the blood pressure on the manometer
- Read the SBP for phase I of Korotkoff sounds
- Read the DBP for phase V of Korotkoff sounds (complete disappearance of sounds noted at the stethoscope)
- Measure the heart rate
- Continue decompression until complete purge of the cuff
Warning: A blood flow stoppage carries risks.
Recommandation:
- Use K5* for adult auscultation
- Use K5 for auscultation of pregnant patients, so long as the sounds are noticeable when the cuff is deflated, in which case the K4 can be used
- Use K4* for auscultation of children between 3 and 12 years.
- See the table describing the phases of Korotkoff sounds.
Environmental specifications:
- Operating temperature: +10°C to +40°C
- Operating relative humidity: 20% to 85%
- Storage temperature: -20°C to +70°C
- Relative humidity at storage: 20% to 85%
MAINTENANCE AND MAINTENANCE VERIFICATION
Comply with current regional regulations for the disposal of manual tensiometers or accessories.
Metrological inspection by an approved body
Check accuracy and watertightness:
- At least once every 2 years (manometer)
- At least once per year (cuffs, tube, bulb)
- After each maintenance operation or repair or replacement of removable parts.
- Note: In accordance with international and national metrological standards, the operator or the body responsible for maintenance must ensure traceability of benchmark manometer used for calibration.
- Note: The manufacturer undertakes to deliver the spare parts replacement following the instructions from the end user.
Removable parts & accessories:
Please contact your approved retailer to purchase the parts and accessories
below:
- Replacement bulb,
- LifeLink® tube,
- Cuffs (all sizes).
CLEANING AND DISINFECTION
- Warning: Never immerse the manometer in liquid or in a decontamination device. Use a neutral detergent to clean the surface of product. Should disinfection be necessary, use a cloth soaked in 75% alcohol.
- Warning: Avoid placing your Mobi too close to heat, excessive cold, solvents or greases. Avoid long-term contact with sweat, as the PVC could easily stiffen (bulb and tube). Keep hands dry.
- Warning: To clean the cuff, do not use alcohol but use warm, soapy water to avoid damaging the printing.
TECHNICAL SPECIFICATIONS
-
Mobi Manometer
- Accuracy: +/-3mmHg
- Operating range: 0 to 300mmHg
- Displayed range: 0 to 300mmHg
- Indication type: radial scale
- Scale grading: increments of 2 mmHg
- Tube adapter: magnetic adapter for Life Link® connection
- Pressure generation: bulb
- Pressure reduction: air release valve
- Case: ABS
- Mechanism: copper-beryllium alloy
- Bulb: PVC
- Spoon: ABS
- Valve: Zamack Z3
-
Life Link® tube
PVC tube with magnetic nickel steel connection. -
Mobi cuff
- Cuff: TPU.
- Tube adapter: magnetic adapter for Life Link® connection.
PRECAUTIONS FOR USE
- Avoid impacts, particularly on the dial glass.
- Do not expose to heat, sun or humidity, dust.
- Avoid twisting the tubing while using Life Link® and be careful to hold the connector (not the tube) while unplugging to change cuff
CONTRAINDICATION
The magnetic Life Link® connectors must be held over 5 cm from any ferromagnetic or electronic implant with sensitivity to magnetic fields.
WARRANTY
- The Mobi tensiometer is guaranteed against any part and labour or operating fault, in accordance with the manufacturer’s specifications, under normal maintenance and usage conditions. The 3-year guarantee for the tensiometer (excluding cuff and tube) starts from the date of purchase from an approved retailer.
- SPENGLER’s obligations are limited during the guarantee period to the repair or replacement of components deemed to be defective by SPENGLER
- Guarantee for Mobi accessories:
- Life Link® tube: 1 year guarantee
- Mobi inflation bulb: 1 year guarantee
- Mobi cuff: 1 year guarantee
STANDARD
Mobi is compliant with the European reference standard NF EN ISO 81060-1.
QUALITY ASSURANCE
NF EN ISO 13485
CE MARKING
CE certification aneroid tensiometer provided by GMED, notified body no. 0459. Class Im device (measurement of blood pressure by auscultatory method) in accordance with European directive 93/42/CEE regarding medical devices.
ABOUT COMPANY
- Spengler S.A.S.
- 30 rue Jean de Guiramand
- 13290 Aix en Provence
- FRANCE
- Tél. : + 33 (0) 4 42 52 89 89
- Fax : + 33 (0) 4 42 52 89 09
- Mail : contact@spengler.fr
- www.spengler.fr
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>