4AILE 111 Blood Pressure Monitor User Manual
- June 14, 2024
- 4AILE
Table of Contents
- 4AILE 111 Blood Pressure Monitor
- INSTRUCTION AND INTENDED USE
- IMPORTANT INFORMATION ON BLOOD PRESSURE AND ITS MEASUREMENT
- Using your Monitor for the First Time
- Measurement Procedure
- SYMBOL DESCRIPTIONS
- Care and Maintenance
- Warranty
- Certifications
- Technical Specifications
- EMC Declaration
- Warranty Card
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
4AILE 111 Blood Pressure Monitor
A Special Thank You …
Thank you for choosing a 111 blood pressure instrument. We’re proud of the
care and quality that goes into the manufacture of each and every item that
bears our name. Only the finest materials are used to assure you of a timeless
instrument designed for optimum performance. You’ll quickly appreciate the
results, for you now own one of the finest digital blood pressure instruments
that money can buy. With proper care and maintenance, your 111 automatic blood
pressure monitor is sure to provide you with many years of dependable service.
Please read the following instructions and general information which will
prove helpful in allowing you to enjoy your 111 of Digital In hospitals and
physician’s offices throughout the world, where accuracy and dependability are
critical, 111 professional diagnostic products are the instruments of choice.
Now you too can enjoy the benefits of 111 engineering and quality in the home. This feature rich instrument was designed to simplify the measurement of blood pressure and pulse rate at home and deliver consistent, dependable results.Your 111 digital blood pressure monitor is a fully automatic digital blood pressure measuring device for use on the upper arm. It enables very fast and reliable measurement of the systolic and diastolic blood pressure as well as the pulse by way of the oscillometric method. This device offers accuracy and has been designed to be user friendly. Read this booklet thoroughly before attempting to use your new 111 Digital Blood Pressure Monitor. Thank you for your patronage. It is indeed our pleasure to serve you. Sincerely, Shenzhen Yuezhongxing Technology Co., Ltd.
- E-mail: support@4aile.com
- Phone:: +1888-875-0846
- Website: 4aile.com
Shenzhen Yuezhongxing Technology Co., Ltd. No.2 Zhenye Road, Pingshan District, Shenzhen, China
- Contact: support@4aile.com
- Web: 4aile.com
EC IREPI
EUROPECERT UG (haftungsbeschrankt),
INSTRUCTION AND INTENDED USE
This manual is for the 111 models. It is a fully automatic digital blood pressure measuring device for use by adults on the upper arm at home or in your doctor’s/nurse’s office. The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with arm circumference ranging from ( D22-42cm D 22-52cm). The devices detect the appearance of irregular heartbeats during measurement It enables very fast and reliable measurement of systolic and diastolic blood pressure as well as pulse through the oscillometric method. This device offers clinically proven accuracy and has been designed to be user friendly. Before using, please read this instruction manual carefully and then keep it in a safe place. Please contact your doctor for further questions on the subject of blood pressure and its measurement. The user is an intended operator, the functions of monitoring blood pressure and pulse rate can be safely used by user. The routine clean and changing batteries can be performed by the user. Contraindication(s) Use of this instrument on patients under dialysis therapy or on anticoagulant, antiplatelets, or steroids could cause internal bleeding. intended patient population This monitor is intended for use by adults only. Consult with a physician before using this instrument on a child. Intended placement of cuff Place the cuff on your bare upper arm one inch above the bend of your elbow. Make sure the tubing falls over the front center of your arm so that the sensor is correctly placed. Pull the end of the cuff so that it’s evenly tight around your arm.
Warning:
-
When patient use this device by themselves, the PATIENT is an intended OPERATOR.
-
Do not service or maintain while this device is in use.
-
All functions can be safely used by patient.
-
Patient can perform the maintenance: changing batteries and cleaning.
-
DO NOT use this monitor on infants, toddlers, children or persons who cannot express themselves.
-
DO NOT use this monitor on an injured arm or a arm under medical treatment.
-
DO NOT apply the arm cuff on your arm while on an intravenous drip or blood transfusion.
-
DO NOT use this monitor in areas containing high frequency (HF) surgical equipment, magnetic resonance imaging (MRI) equipment, computerized tomography (CT) scanners. This may result in incorrect operation of the monitor and/or cause an inaccurate reading.
-
Consult with your physician before using this monitor if you have common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation; arterial sclerosis; poor perfusion; diabetes; pregnancy; preeclampsia or renal disease. NOTE that any of these conditions in addition to patient motion, trembling, or shivering may affect the measurement reading.
-
NEVER diagnose or treat yourself based on your readings. ALWAYS consult with your physician.
-
This product contains small parts that may cause a choking hazard if swallowed by infants, toddlers or children.
-
Stop using this monitor and consult with your physician if you experience skin irritation or discomfort.
-
Consult with your physician before using this monitor on a wrist where intravascular access or therapy, or an arterio-venous (A-V) shunt, is present because of temporary interference to blood flow which could result in injury.
-
Consult with your physician before using this monitor if you have had a mastectomy.
-
Consult with your physician before using this monitor if you have severe blood flow problems or blood disorders as cuff inflation can cause bruising.
-
DO NOT take measurements more often than necessary because bruising, due to blood flow interference, may occur.
-
ONLY inflate the arm cuff when it is applied on your arm.
-
Remove the arm cuff if it does not start deflating during a measurement.
-
DO NOT use this monitor for any purpose other than measuring blood pressure.
-
During measurement, make sure that no mobile device or any other electrical device that emits electromagnetic fields is within 12 inches (30cm) of this monitor. This may result in incorrect operation of the monitor and/or cause an inaccurate reading.
-
DO NOT disassemble or attempt to repair this monitor or other components. This may cause an inaccurate reading.
-
DO NOT use in a location where there is moisture or a risk of water splashing this monitor. This may damage this monitor.
-
DO NOT use this monitor in a moving vehicle such as in a car.
-
DO NOT drop or subject this monitor to strong shocks or vibrations.
-
DO NOT use this monitor in places with high or low humidity or high or low temperatures.
-
During measurement, observe the arm to ensure that the monitor is not causing prolonged impairment to blood circulation.
-
DO NOT use this monitor in high-use environments such as medical clinics or physician offices.
-
DO NOT use this monitor with other medical electrical (ME) equipment simultaneously. This may result in incorrect operation of the monitor and/or cause an inaccurate reading.
-
Avoid bathing, drinking alcohol or caffeine, smoking, exercising and eating for at least 30 minutes before taking a measurement.
-
Rest for at least 5 minutes before taking a measurement.
-
Remove tight-fitting, thick clothing and any accessories from your wrist while taking a measurement.
-
Remain still and DO NOT talk while taking a measurement.
-
ONLY use this monitor on persons whose wrist circumference is within the specified range of the cuff.
-
Ensure that this monitor has acclimated to room temperature before taking a measurement. Taking a measurement after an extreme temperature change could lead to an inaccurate reading. We recommend waiting for approximately 2 hours for the monitor to warm up or cool downwhen the monitor is used in an environment within the temperature specified as operating conditions after it is stored either at the maximum or at the minimum storage temperature. For additional information of operating and storage/transport temperature.
-
DO NOT crease the arm cuff excessively.
-
DO NOT use this monitor after the durable period has ended. Never apply the arm cuff over a wound, as this can cause further injury;
No modification of this equipment is allowed. -
Do not modify this equipment without authorization of the manufacturer.
-
If this equipment is modified, appropriate inspection and testing must be conducted to ensure continued safe use of the equipment.
-
The time required for the device to warm from the minimum storage temperature between uses until the device is ready for its intended use when the ambient temperature is 20 C.
-
The time required for the device to cool from the maximum storage temperature between uses until the device is ready for its intended use when the ambient temperature is 20 C.
Warning:
- Use of power adapters
- Adapter: input 100-240v, 50/60hz output DC SV 1A
- Do not to position the device to make it difficult to operate the disconnection device while using adaptor.
- Do not use cuffs, AC adapters or batteries other than those included with this product or replacement parts supplied by the manufacturer.
- Prone to damage from water leakage, high temperature, moisture, direct sunlight and more or more corrosive gas environment. Avoid using this product in the aforementioned environments.
- Remove the battery if the blood pressure monitor is not likely to be used for some time.
- Do not use the batteries and the AC adapter to provide power at the same time.
- The separate ac adapter which is intended to connect USB interface of Blood Pressure Monitor has not been evaluated according to !EC 60601-1. The safety of the product shall be reappraised when it power supply by a separate ac adapter.
- DO NOT fold or kink the air tube while taking a measurement. This may cause an injury by interrupting blood flow
- Avoid compression and restriction of the connection tubing.
- Please stop using this device and contact distribution or manufacture in the event of changes in the performance of this device.
- Do not use accessories and other parts except for manufacture specified.
- When display” ¢0 “, please replace batteries immediately.
- The battery can support to conduct 10 times measurement.
- waiting for approximately 2 hours for the monitor to warm up or cool down when the monitor is used in an environment within the temperature specified as operating conditions after it is stored either at the maximum or at the minimum storage temperature.
- This device contains sensitive electronic components. Avoid strong electrical or electromagnetic fields in the direct vicinity of the device (e.g. mobile telephones, microwave ovens) during use. These can lead to erratic results.
- Do not attempt to service or repair this device yourself. Should a malfunction occur, refer to local distributor or the manufacturer.
- Significant risks of reciprocal interference posed by me equipment during specific investigations or treatments.
Note: To obtain the greatest accuracy from your blood pressure instrument, it is recommended that the instrument be used within the specified temperature and the relative humidity, please see the Technical Specifications.
Note: The cuff is treated as the applied part.The user should contact the manufacturer for assistance, if needed, in setting up, using or maintaining the device. The standard material used for the bladder and tubing is latex- free.
Attention: Self-measurement means control, not diagnosis or treatment. Unusual values must always be discussed with your doctor. Under no circumstances should you alter the dosages of any drugs prescribed by your doctor the pluse monitoring function is not suitable for checking the frequency of heart pacemakers!
Note: To obtain the greatest accuracy from your blood pressure instrument, it is recommended that the instrument be used within the specified temperature and the relative humidity, please see the Technical Specifications.
Note: The cuff is treated as the applied part.The user should contact the manufacturer for assistance, if needed, in setting up, using or maintaining the device.
IMPORTANT INFORMATION ON BLOOD PRESSURE AND ITS MEASUREMENT
How does high or low blood pressure arise?
Your level of blood pressure is determined in the circulatory center of the
brain and adjusts to a variety of situations through feedback from the nervous
system. To adjust blood pressure, the strength and speed of the heart (Pulse),
as well as the width of circulatory blood vessels isaltered.Blood vessel width
is controlled by fine muscles in the blood vessel walls.Your level of arterial
blood pressure changes periodically during heart activity: During the “blood
ejection” (Systole) the value is highest (systolic blood pressure value). At
the end of the heart’s “rest period” (Diastole) pressure is lowest (diastolic
blood pressure value). Blood pressure values must lie within certain normal
ranges in order to prevent particular diseases.
WHICH VALUES ARE NORMAL
Please refer to the diagram below(Picture-01)
There are six grids in the display of device. Please refer to the picture-01-01. Different grids represent different interval scales of WHD.
Blood pressure is very high if your diastolic pressure is above 90 mm Hg and/or your systolic blood pressure is over 160 mmHg, while at rest. In this case, please consult your physician immediately. Long-term values at this level endanger your health due to continual damage to the blood vessels in your body. If your systolic blood pressure values are between 140 mm Hg and 159 mmHg and/or the diastolic blood pressure values between 90 mmHg and 99 mmHg, consult your physician. Regular self-checks are necessary. If you have blood pressure values that are too low, (i.e., systolic values under 90 mmHg and/or diastolic values under 60 mm Hg), consult your physician. Even with normal blood pressure values, a regular self-check with your blood pressure monitor is recommended. You can detect possible changes in your values early and react appropriately. If you are undergoing medical treatment to control your blood pressure, keep a record of values along with time of day and date. Show these values to your physician. Never use the results of your measurements to independently alter the drug doses prescribed by your physician. The method of verifying the calibration of blood pressure monitor in pressure gauge test mode: Step 1: blood pressure monitor in time mode, long press the setting key and then press the memory key to enter the static pressure mode. Step 2: connect the special plug with the windpipe and insert the blood pressure monitor armband interface with the equipment integrated with the gas source and the manometer. The static pressure of the instrument is displayed in the high pressure area. Comparing the pressure displayed by the device with the pressure shown by the manometer, the difference is less than 3mmHg as qualified.
Further information
- If your values are mostly normal under resting conditions but exceptionally high under conditions of physical or psychological stress, it is possible that you are suffering from so-called “labile hypertension.” Consult your doctor.
- Correctly measured diastolic blood pressure values above 120mmHg require immediate medical treatment.
What can be done if regular high or low values are obtained?
- Consult your doctor.
- Increased blood pressure values (various forms of hypertension) are associated with considerable health risks over time. Arterial blood vessels in vessel walls (Arteriosclerosis). A deficient supply of blood to important organs (heart, brain, muscles) can result from arteriosclerosis. Furthermore, the heart will become structurally damaged with increased blood pressure values.
- Components of your blood pressure monitor
The symbols on the Picture-OS
- Blood pressure monitor indicator
- Battery symbol
- Mute symbol
- Date/Time display
- Memory symbol
- Average value symbol
- USER symbol
- Heartbeat symbol
- Systolic blood pressure
- Diastolic blood pressure
- Pulse display
- Number of group symbol
- Irregular heartbeat symbol
- Systolic
- Diastolic
Features of Model 111
- Talking function
- Double users: 2*99 sets memory
- Date/time display
- Irregular heartbeat checking
- Average value function
- Low battery display
- WHO function
- Auto power-off
- External power adapter support
- Cuff self-checking function
Note: Arm circumference should be measured with a measuring tape in the middle of the relaxed upper arm. Do not force cuff connection into the opening. Make sure the cuff connection is not pushed into the AC adapter port.
Using your Monitor for the First Time
Activating the pre-installed batteries
Battery Installation
Use only 1.SV “AAA” alkaline batteries with this device.
- Press the hook on the bottom of the battery cover and lift the cover off in the direction of the arrow.
- Install 4 “AAA” size batteries so the + (positive) and – (negative) polarities match the polarities of the battery compartment, replace the battery cover. Make sure that the battery cover is securely in position. Battery replacement
Low Battery Indicator
- When the Low Battery Indicator appears on the display, turn the monitor off and remove all the batteries. Replace with 4 new batteries at the same time. Long-life alkaline batteries are recommended.
- To prevent the damage of monitor from leaked battery fluid, please take out of battery if the monitor unused in a long time(generally more than 3 months). If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Contact a physician immediately.
- Attached battery is only for testing the function of the monitor, Long-life alkaline batteries are recommended.
- Dispose of the device, components and optional accessories according to applicable local regulations. Unlawful disposal may cause environmental pollution
- Battery is dangerous stuff, do not mix it with other rubbish.
System Settings
After you load the battery or connect power for the monitor, long press the
SET button for more than 3 seconds, and then you can start to set when
flashing numbers appear on the screen.
A. Setting the User
Press the SET button, then press the USER button. When display A(/B) on the
screen, press the USER button to switch to user B(/A).
B. Setting the Year
Press SET button for more than 3 seconds, when the YEAR display is flashing,
press MEM button continuously and it will increase continuously 1 by 1 until
2099, once the year set is ok, press SET button to confirm.
C. Setting the Month/Date
After completing the YEAR setting, the MONTH/DATE will flash on the screen,
press MEM button, the month will increase by 1, press SET button to confirm.
DO the same way to set the Date, press SET button to confirm.
D. Setting the Time
After completing the MONTH/DATE setting, the TIME will flash on the screen,
press MEM button, the hour will increase by 1, press SET button to confirm. DO
the same way to set the minute, press SET button to confirm.
E. Setting the Speaker
After completing the TIME setting, the SP will flash on the screen, press MEM
button to switch ON or OFF, press SET button to confirm.
F. Setting the Unit
The original unit display on the screen is mm Hg, after you completing SP
setting, the PA will flash on the scree, press MEM button to switch ON, the
unit will change from mmHg to kPa.
G. Record Delete
Turn off the machine, press MEM button for 3 seconds, don’t let your finger go, press SET button for 3 seconds at the same time, when EEEE appear on the screen, it means the memories have been deleted.
Note
When you choose to delete the records, you will delete all measurement record
at one time. If you decide to delete the all record, please keep the record in
another way, in case you need it some days later. Take the battery out won’t
lead to a record missing.
Measurement Procedure
Note: You should always be seated and calm before and during measurement.
Before measurement
- Avoid eating and smoking as well as all forms of exertion directly before measurement. These factors influence the measurement result. Find time to relax by sitting in an armchair in a quiet atmosphere for about ten minutes before taking a measurement.
- Remove any garment that fits closely to your upper arm.
- Always measure on the same arm (normally left).
Always compare measurements taken at the same time of day, since blood pressure changes during the course of the day, as much as 20-40 mmHg.
S.2Common sources of error:
Note: Comparable blood pressure measurements always require the same
conditions!
- Conditions should always be quiet.
- All efforts by the user to support the arm can increase blood pressure. Make sure you are in a comfortable, relaxed position and do not flex any of the muscles in the measurement arm during the measurement. Use a cushion for support if necessary.
- If the arm artery lies considerably lower or higher than the heart, an erroneously high or low blood pressure will be measured! Each 25-30cm difference in height between your heart and the cuff results in a measurement error of 10 mm Hg!
- Cuffs that are too narrow or too short result in false measurement values. Selecting the correct cuff is extremely important. Cuff size is dependent upon the circumference of the arm (measured in the center). The permissible range is printed on the cuff. Cuff works Under the pressure range 0-300MMHG
The wide range rigid cuff is:B.7″-16.5″ (22-42 cm) B.7″-20.4″ (22-52 cm) Note: Only use approved cuffs!
- A loose cuff or a sideways protruding air pocket causes false measurement values.
- With repeated measurements, blood accumulates in the arm, which can lead to false results. Consecutive blood pressure measurements should berepeated after a 1 minute pause or after your arm has been held up In order to allow the accumulated blood to flow away. If you decide to take your Averaging Mode measurement again, be sure to wait at least one minute beforehand.
Fitting the Cuff
Please refer to picture-06
- The cuff is preformed for easier use. Remove tight or bulky clothing from your upper arm.
- Wrap the cuff around your upper left arm. The rubber tube should be on the inside of your arm extending downward to your hand. Make certain the cuff lies approximately 1/2″ to 3/4″ (1 to 2 cm) above the elbow. Important! The on the edge of the cuff (Artery Mark) must lie over the artery which runs down the inner side of the arm.
- To secure the cuff, wrap it around your arm and press the hook and loop closure together.
- There should be little free space between your arm and the cuff. You should be able to fit 2 fingers between your arm and the cuff. Cuffs that don’t fit properly result in false measurement values. Measure your arm circumference if you are not sure of proper fit.
- Lay your arm on a table (palm upward) so the cuff is at the same height as your heart. Make sure the tube is not kinked.
- Remain seated quietly for at least two minutes before you begin the measurement.
Measure Procedure
Refer to picture-07 The monitor is designed to take measurements and store the
measurement values in memory for two people using User ID A and User ID B.
- Sit comfortably in a chair with your feet flat on the floor.
- Select your User ID (A or B).
Stretch your arm forward on the desk and keep relaxing, make sure the palm of hand is upturned. Make sure arm is in correct position, to avoid body movement. Sit still and do not talk or move during the measurement. After the cuff has been appropriately positioned on the arm and connected to the blood pressure monitor, the measurement can begin:
- Press the START/STOP button. The pump begins to inflate the cuff. In the display, the increasing cuff pressure is continually displayed.
- After automatically reaching an individual pressure, the pump stops and the pressure slowly falls. The cuff pressure is displayed during the measurement.
- When the device has detected your pulse, the heart symbol in the display begins to blink.
- When the measurement has been concluded, the measured systolic and diastolic blood pressure values, as well as the pulse will be displayed.
- The appearance of this symbol signifies that an irregular heartbeat was detected.This indicator is only a caution. It is important that you be relaxed, remain still and do not talk during measurements.
- The measurement results are displayed until you switch the device off. If no button is pressed for 60 seconds, the device switches off automatically
NOTE:
Patient Position:
- Comfortably seated
- Legs uncrossed
- Feet flat on the floor
- Back and arm supported
- Middle of the CUFF at the level of the right atrium of the heart
Recommended Use Methods
- Recommendation that the PATIENT relax as much as possible and not talk during the measurement PROCEDURE
- Recommendation that 5 min should elapse before the first reading is taken
- Any reading can be affected by the measurement site, the position of the PATIENT, exercise, or the PATIENT’S physiologic condition
- Performance of the AUTOMATED SPHYGMOMANOMETER can be affected by extremes of temperature, humidity and altitude
- To stop the inflation or measurement, push the START/STOP button. Themonitor will stop inflating, start deflating, and will turn off.
- After the monitor has detected your blood pressure and pulse rate, the cuff automatically deflates. Your blood pressure and pulse rate are displayed.
- The monitor will automatically turn off after one minute.
Sirregular Heartbeat Detector
This symbol – indicates that certain pulse irregularities were detected during
the measurement. In this case, the result may deviate from your normal basal
blood pressure – repeat the measurement. Information for the doctor on
frequent appearance of the Irregular Heartbeat Symbol. This instrument is an
oscillometric blood pressure monitor device that also analyzes pulse frequency
during measurement. The instrument is clinically tested. If pulse
irregularities occur during measurement, the irregular heartbeat symbol is
displayed after the measurement. If the symbol appears more frequently (e.g.
several times per week on measurements performed daily) or if it suddenly
appears more often than usual, we recommend the patient to seek medical
advice. The instrument does not replace a cardiac examination, but serves to
detect pulse irregularities at an early stage.
Error Indicates
The following symbol will appear on the display when measuring abnormal
SYMBOL | CAUSE | CORRECTION |
---|---|---|
Er 1 | Sensor abnormal | 1. Don’t talk or shake your arm before |
measuring;
2. Check if the pump is working or not.
Er 2
| Blood pressure can’t be
measured
| Check the wearing position of the cuff and remeasure according to the require, posture.
Er 3
| Blood pressure is abnormal| Check the wearing position of the cuff and
remeasure according to the required posture
Er 4| Blood pressure arm cuff cannot be pressurized| 1. Check whether the air
path is open;
2. Check whether the arm cuff is too loose.
Er 5
|
Arm cuff is pressurized too fast
| 1. Check whether any blockage in the air path;
2. Check whether the arm cuff is too tight.
Er 6
| Excessive pressure interference| Measure the posture according to the requirements. DO NOT move your body in the measurement.
Er 7
| The pressure exceeds the limit| Measure the posture according to the requirements. DO NOT move your body in the measurement.
Trouble Removal
PROBLEM | CHECK | CAUSE AND SOLUTION |
---|---|---|
Check the battery | Replace new one | |
No | power | |
power | Check the polarity position | Installation for proper placement of the |
batteries polarities
No inflation
| Whether the plug insert| Inset into the air socket tightly
---|---|---
Whether the plug broken or leak| Change a new cuff
Err and stop working| Whether move the arm when inflate| Keep the body
peaceful
Check if chatting when measured| Keep quite when measure
Cuff leak| Whether the cuff wrap too loose| Wrap the cuff tightly
Whether the cuff is broken| Change a new cuff
Please contact the distributor if you can’t solve the problem, do not
disassemble the unit by your self
SYMBOL DESCRIPTIONS
The following symbols may appear in this manual, on the Digital Blood Pressure Monitor 111, or on it’s accessories. Some of the symbols represent standards and compliances associated with the Digital Blood Pressure Monitor 111 and its use.
Memory
At the end of a measurement, this monitor automatically stores each result
with date and time. Each unit stores 99 sets measurements for 2 users, totally
198 sets (User A and B) Viewing the stored values With the unit off, press the
Memory button. The display first shows “A”, then shows an average of all
measurements stored in the unit. Please note: Measurements for each user are
averaged and stored separately. Be certain that you are viewing the
measurements for the correct user. Pressing the Memory button again displays
the previous value. To view a particular stored memory, press and hold the
Memory button to scroll to that stored reading.
Discontinuing a Measurement
If it is necessary to interrupt a blood pressure measurement for any reason
(e.g the patient feels unwell), the Start/Stop button can be pressed at any
time. The device then immediately lowers the cuff pressure automatically.
Battery Change Indicator Batteries discharged- replacements required
When the batteries are discharged, the battery symbol will flash as soon as
the instrument is switched on. You cannot take any further measurements and
must replace the batteries .
The battery compartment is located on the back side of the unit.
- Remove cover from the bottom plate, as illustrated below picture-OB
- Insert the batteries ( 4 x size AAA ) . Always use AAA long life batteries or alkaline 1.Sv batteries.
- The memory retains all values although date and time must be reset – the year number therefore flashes automatically after the batteries are replaced.
- To set date and time, follow the procedure described in Section 4.2.
Which batteries and which procedure?
Use four new, longlife 1.SV AAA batteries. Do not use batteries beyond their
expiration date. If the monitor is not going to be used for a prolonged period
the batteries should be removed
Using rechargeable batteries
You can also operate this instrument using rechargeable batteries.
- Only use “NiMH” reusable batteries!
- If the battery symbol the batteries must be removed and recharged! They must not remain inside the instrument, as they may become damaged through total discharge even when switched off. The batteries must NOT be discharged in the blood pressure monitor! If you do not intend to use the instrument for a week or more, always remove the rechargeable batteries!
- Recharge these batteries using an external charger and follow manufacturer’s instructions carefully.
Using the AC Adapter
You may also operate this monitor using the AC adapter (output SV DC/lA with
Micro USB plug). Use only the approved AC adapter to avoid damaging the unit.
- Ensure that the AC adapter and cable are not damaged.
- Plug the adapter cable into the AC adapter port on the right side of the blood pressure monitor.
- Plug the adapter into your electrical outlet. When the AC adapter is connected, no battery current is consumed.
Note: No power is taken from the batteries while the AC adapter is connected to the monitor. If electrical power is interrupted,(e.g., by accidental removal of the AC adapter from the outlet) the monitor must be reset by removing the plug from the socket and reinserting the AC adapter connection.
Care and Maintenance
Wash hands after each time measurement. If one device is used by different patients, wash hands before and after each use.
- Do not expose the device to either extreme temperatures, humidity, dust or direct sunlight.
- The cuff contains a sensitive air-tight bubble. Handle this cuff carefully and avoid all types of stress through twisting or buckling.
- Clean the device with a soft, dry cloth. Do not use gas, thinners or similar solvents. Spots on the cuff can be removed carefully with a damp cloth and soapsuds. The cuff with bladder must not be washed in a dishwasher, clothes washer, or submerged in water.
- Handle the tube carefully. Do not pull on it. Do not allow the tubing to kink and keep it away from sharp edges.
- Do not drop the monitor or treat it roughly in any way. Avoid strong vibrations.
- Never open the monitor! This invalidates the manufacturer’s warranty.
- Batteries and electronic instruments must be disposed of in accordance with the locally applicable regulations, not with domestic waste.
Accuracy Test
Sensitive measuring devices must be checked for accuracy from time to time. We
recommend a periodical inspection of your unit by an authorized dealer every 1
year. Please turn to local distributor or the manufacturer.
Warranty
Your blood pressure monitor is guaranteed for 1 year against manufacturers’ defects for the original purchaser only, from date of purchase. The warranty does not apply to damage caused by improper handling, accidents, professional use, not following the operating instructions or alterations made to the instrument by third parties. Warranty only applies to the instrument. All accessories including the cuff are guaranteed for one year, USB charging cable is not included. There are no user serviceable parts inside. Batteries or damage from old batteries is not covered by the warranty.
Note: According to international standards, your monitor should be checked for accuracy every year.
Certifications
Device standard:
This device is manufactured to meet the European blood pressure monitors: EN
ISO 81060-1 :2007 • !EC 80601-2-30 • IEC60601-1-11 • IEC60601-1
Electromagnetic compatibility: Device fulfills the stipulations of the
International standard IEC60601-1-2
Technical Specifications
- Model: 111
- Weight: 270g (Batteries and AC adapter are not included)
- LCD Digital Display Size: 76 (W) x 62 (L) x 2.8 (H) mm [ 3″(W)x2.44″(L)x0.11 “(H))
- Accessories: lxMain Device, lxCuff, lxUsers manual, lxWarranty card Operating Conditions: Temperature: SC to 40C;Humidity: 15% to 93% RH; Storage And Shipping Conditions:Temperature: -25C to 70C; Humidity:S 93% RH;
- Atmospheric pressure range: 70kPa~106kPa
- Measuring method: Oscillometric
- Pressure sensor: Resistive
- Measuring range: 0-280mmHg;
- Blood pressure detection range: systolic blood pressure 55-25SmmHg, diastolic blood pressure 25-200mmHg
- Pulse: 40 to 199 per minute
- Cuff pressure display range: <300mmHg
- M emory: Automatically stores the last 99 measurements for 2 users (total 198) Measuring resolution: 1 mm Hg
- Accuracy: Pressure within ± 3 mmHg / pulse ± 5 % of the reading
- Power source: a) 4*AAA batteries, 1.5 V b) AC adapter INPUT: 100-240VAC S0/60HZ OUTPUT: SV DC 1A
- Accessories: Wide range rigid cuff D8.7″-16.5″ (22-42 cm) D8.7″-20.4″ (22-52 cm) Automatically power off : 60 seconds
- Users: Adult
Expected service life of the device and accessories: 5 years Technical alterations reserved! The SPHYGMOMANOMETER was clinically investigated according to the requirements of ISO 81060-2:2013
EMC Declaration
- This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided, and this unit can be affected by portable and mobile RF communications equipment.
- Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in incorrect operation of the unit.
- Caution: This unit has been thoroughly tested and inspected to assure proper performance and operation!
- Caution: this machine should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, this machine should be observed to verify normal operation in the configuration in which it will be used
Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used
in such an environment.
Immunity test| !EC 60601
test level
| Compliance
level
| Electromagnetic environment –
guidance
Conducted RF
!EC
61000-4-6
| 3 Vrms
150 kHz to 80 MHz
3 V RMS
outside the ISM band, 6 V RMS
in the ISM and amateur bands 80%
AM at lkHz
| | Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance d; 0.35 v’p
d; l. 2v’p
| 10 V/ m| 10 V/ m| 80MHz to 800MHz : d; l. 2v’p 800MHzto 2.7GHz: d; 2 . 3v’p Where, P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance.
Field strengths from fixed RF transmitters , as determined
| 80 MHz to| 80 MHz to
| 2.7 GHz| 2.7 GHz
| 80% AM at| 80% AM at
Radiated RF| lkHz| lkHz
!EC
| |
61000-4-3| |
Test
frequenc y (MHz)
| Band (MHz)| Service (MHz)| Modulation| Maximum power(w)| Distance (m)|
IMMUNITY TEST
---|---|---|---|---|---|---
385| 380-390| TETRA 400| Pulse Modulation 18 Hz| 1.8| 0.3| 27
450| 430-470| GMRS 460,
FRS 460
| FM
±5kHz deviation lkHz sine
| 2| 0.3| 28
710| 704-787| LTE Band 13, 17| Pulse Modulation
217 Hz
| 0.2| 0.3| 9
745
780
810| 800-960| GSM 800/ 900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
| Pulse Modulation
18 Hz
| 2| 0.3| 28
870
930
1720
| ****
1700-19
| GSM 1800;
CDMA 1900;
GSM 1900; DECT;
LTE Band 1,3
4,25;UMTS
| ****
Pulse Modulation
217 Hz
| ****
2
| ****
0.3
| ****
28
1845
1970
2450
| ****
2400-25
7090
| Bluetooth, WLAN 802.11
b/ g/ n, RFID 2450,
LTE Band 7
| ****
Pulse Modulation
217 Hz
| ****
2
| ****
0.3
| ****
28
| | | by an electromagnetic site survey, should be less than the compliance
level in each frequency range. Interference may occur in the vicinity of
equipment marked wit h the following symbol:
---|---|---|---
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies . NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from struct ures,
objects and people.
A. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment
to fixed RF transmitters, an electro magnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnorma l performance is observed, additiona l measures may be necessary, such as re-orienting or relocating the device.
B. Over the frequency range 150 kHz to 80 MHz, field stre ngth s should be less than 3 V/m.
Guidance and manufacture ‘ s declaration – electromagnetic emission
The device is intended for use in the electromagnetic envir onment specified below. The customer of the user of the device should assure that
it is used in such an env i ronment.
Rated maximum output power of transmitter (W)| Separation distance
according to frequency of transmitter (m)
---|---
150 KHz
to 80 MHz
| 80 MHz to 800 MHz
| 800 MHz to
2.7 GHz
0,01| 0.12| 0.12| 0.23
0,1| 0.38| 0.38| 0.73
| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, the recommended separation distanced in metres (m) can be estim ated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufa cturer.
NOTE 1 At 80 MHz and 800 MHz, the separati on distance for the higher frequency range applies .
NOTE 2 These guide lin es may not apply in all situations. Elec tr omagnet ic propagation is affected by absorption and reflection from structures, objects and people.
Guidance and manufacture ‘ s declaration – electromagnetic immunity
The device is inte nded for use in the electromag netic environment specified below. The customer or the user of the device, should assur e that it is used in such an environment.
Emission test
|
Compliance
| Electromagnetic
environment – guidance
---|---|---
RF emissions CJSPR 11
|
Group 1
| The device use RF energy only for its internal function. Therefore, its RF
emissions are very low and are not lik ely to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11|
Class B
| The device is suitable for use in all establishments, including domestic
establishments other than domestic and those directly connected to the public
low-voltage power supply network that supplies buildings used for domestic
purposes.
Harmonic
emissions
!EC 61000-3-2
| Not
app licable
Voltage fluctuations/fli cker emissions
!EC
61000-3-3
| Not applicable
Recommended separation distances between portable and mobile RF communications equipment and the device. The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the communications equipment.
5240| 5100-5800| WLAN 802.11
a/ n
| Pulse Modulatio n
217 Hz
| 0.2| 0.3| 9
---|---|---|---|---|---|---
5500
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the
transmitting
antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m.
The 1 m test distance is permitted by !EC 61000-4-3.
a ) For some services, only the uplink frequencies are included.
b ) The carrier shall be modulated using a 50% duty cycle square wave signal.
c ) As an alternative to FM modulation. 50% pulse modulation at 18 Hz may be used because while it does not represent actual modulation, it would be worst case.
The MANUFACTURER should consider reducing the minimum separa tion distance, based on RISK
MANAGEMENT, and using higher IMMUNITY TEST LEVELS that are appropriate for the reduced
minimum separation distance. Minimum separation distances for higher IMMUNITY TEST LEVELS
shall be
calculated using the following equation:
Where P is the maximum power in W, d is the minimum separation distance in m, and E is the
IMMUNITY
TEST LEVEL in V/ m.
Warranty Card
Blood pressure is very high if your diastolic pressure is above 90 mmHg and/or your systolic blood pressure is over 160 mm Hg, while at rest. In this case, please consult your physician immediately. Long-term values at this level endanger your health due to continual damage to the blood vessels in your body. If your systolic blood pressure values are between 140 mmHg and 159 mmHg and/or the diastolic blood pressure values between 90 mmHg and 99 mmHg, consult your physician. Regular self-checks are necessary. If you have blood pressure values that are too low, (i.e., systolic values under 90 mmHg and/or diastolic values under 60 mmHg), consult your physician. Even with normal blood pressure values, a regular self-check with your blood pressure monitor is recommended. You can detect possible changes in your values early and react appropriately. If you are undergoing medical treatment to control your blood pressure, keep a record of values along with time of day and date. Show these values to your physician. Never use the results of your measurements to independently alter the drug doses prescribed by your physician.
- E-mail: support@4aile.com
- Phone:: + 1888-875-0846