MPM Products XP-4000 AccuPRO Photometer Radiometer Instruction Manual
- June 15, 2024
- MPM Products
Table of Contents
- XP-4000 AccuPRO Photometer Radiometer
- INTRODUCTION
- AccuPRO COMPONENTS
- BUTTONS AND CONTROLS
- OPERATION––QUICK GUIDE
- USING AccuPRO
- CUSTOMIZING SETTINGS
- INFORMATION
- CALIBRATION
- THEORY OF OPERATION
- WARRANTY, MAINTENANCE AND BATTERY REPLACEMENT
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
OPERATOR’S MANUAL
AccuPRO™ Photometer/Radiometer
XP-4000 & XP-2000
XP-4000 AccuPRO Photometer Radiometer
INTRODUCTION
1.1 GENERAL
The AccuPRO™ (XP-2000) and AccuPRO™ Plus (XP4000) Meter Series features an
advanced digital microprocessor-controlled readout unit calibrated to
accurately detect and display light intensity readings.
The XP-2000 and XP-4000 (with attached sensor) feature UV, blue (XP-4000
only), and white light modes,and are used for fluorescent inspection (non-
destructive testing).
The XP-4000 readout unit is calibrated for HID UV, visible, and blue light
sources. It provides UV-A and blue light readings from 0-100 mW/cm 2 , visible
light readings from 0-5,300 lux (0-500 fc), and is useful for NDT applications
with simple, one-touch PEAK functionality to easily identify the strongest
point of a light source.
1.2 FEATURES
- The meters are compact, lightweight and batteryoperated for convenient use in the factory, field,orany other location wheremeasurements need to betaken.
- The XP-2000 and XP-4000 readout units have amaximum 4-digit autoranging, color LCD screen.
- The sensor is loaded with low electrical impedance,making their linearity vastly superior to that of anyother comparably priced radiometers.
- The sensors are provided with a special diffusersensor window that helps prevent filter degradationand ensures accurate lambertian or cosine response.
INTERFACE
- Simple, three-button interface
- Accurate to within a hundredth of a unit measurement (e.g., 28.96 mW/cm2)
- Intuitive user interface design
- Navigable screen prompts
- One touch PEAK reset
- Toggle between UV/Visible and Blue Light modes
HARDWARE
- Sensor with 3-foot (0.9 m) electrical cord
- Superior band-pass interference filter provides excellent cosine response
- Press and hold the MENU button to turn the meter ON or OFF
- Powered by four AAA rechargeable nickel-metal hydride batteries with AC battery charger included
1.3 PRECAUTIONS
-
The AccuPRO is carefully designed to prevent accidental shock to the operator. However, no engineering design can render safe an instrument that is used carelessly. Therefore, the directions presented in this manual must be read carefully before any measurements are made. Failure to follow directions could result in injury.
-
The UV sensor is designed for use in regions of the spectrum, notably the ultraviolet range, which may be hazardous to the eyes and/or skin. Ultraviolet protective eyewear and facewear are available from Spectro-UV (for instances when UV exposure is
unavoidable). -
Do not leave the exposed sensor head under the light source any longer than necessary to take measurements. Prolonged exposure can result in premature aging of the sensor, necessitating more frequent recalibration to maintain accurate readings.
-
See Section 10, Warranty, Maintenance and Battery information.
-
DO NOT CHARGE ALKALINE BATTERIES (doing so may result in an explosion, potentially causing damage to the unit and/or operators). Only use the supplied batteries.
AccuPRO COMPONENTS
- Unpack and inspect the component(s) for possible damage in shipment. Save the shipping carton and packing materials for future storing or shipping of the AccuPRO components.
- Assure the sensor cable is undamaged and securely attached to the readout unit.
- Conduct a performance test as soon as possible. If any damage is noted, immediately notify the carrier and supplier (do not use the instrument).
- While charging, it is recommended to have thedevice turned OFF.
2.1 AccuPRO™ PARTS
AccuPRO™ Meter with Multi-Wavelength Sensor XP-2000 (AccuPRO) and XP-4000
(AccuPRO Plus) Soft Carrying Case XCC-200 AC Battery Charger (North American
plug shown, see replacement parts for complete list) Rechargeable AAA
Nickel-Metal Hydride batteries (4)
COMPONENTS AND REPLACEMENT PARTS
Part No.| Description
XP-2000| AccuPRO Dual Sensor (UV-A/VIS), Multilingual Display Meter
XP-4000| AccuPRO Plus 3-in-1 Sensor (UV-A/VIS/Blue),
Multilingual Display Meter, Complete with (4)
Rechargeable “AAA” Nickel-Metal Hydride
Batteries, AC Charger and Carrying Case (100-120V/50-50Hz)
XCC-200| Soft Carrying Case
XP-2000/F| XP-2000 (with 230V/50Hz European Plug)
XP-2000/FB| XP-2000 with UK Plug (230V/50Hz)
XP-2000/FA| XP-2000 with Australia/China Plug (220-240V/50Hz)
XP-4000/F| XP-4000 with European Plug (230V/50Hz)
XP-400/FB| XP-4000 with UK Plug (230V/50Hz)
XP-4000/FA| XP-4000 with Australia/China Plug (220-240V/50Hz)
129450| AC Charger with North American Plug (120V/60Hz)
129451| AC Charger with European Plug (230V/50Hz)
129453| AC Charger with UK Plug (230V/50Hz)
129452| AC Charger with Australia/China Plug (220-240V/50Hz)
2.2 TECHNICAL SPECIFICATIONS
DIMENSIONS
Part| Length| Width| Depth| Weight
Readout
Unit| 6 in
(15.2 cm)| 3.0 in
(7.6 cm)| 1.0 in
(2.5cm)| 8 oz
(227 gm)
Sensor| 3.0 in
(7.6 cm)| 2.0 in
(5.1 cm)| 0.5 in
(1.25 cm)| 5.6 oz
(158.7gm)
2.3 ELECTRICAL SPECIFICATIONS
A full charge of the four nickel-metal hydride batteries (using the included
AC battery charger) will take 8 hours and provide approximately 6 hours of
active scanning.
2.4 OPTICAL ACCURACY AND CALIBRATION
The AccuPRO XP-2000 and XP-4000 readout units measure both ultraviolet and
visible light. The AccuPRO Plus (XP-4000) also measures blue light. The LCD
readout features ±0.2% linearity with the sensor sending the linearity
correction data to the readout unit during initial power-up. The sensor is
designed with superior band-pass filters and the optical stacks are assembled
in Class 100 clean workstations to ensure consistent results. For precise
spectral coverage, these high-quality interference filters will effectively
resist degradation. See Section 8 for additional calibrationrelated
information.
If recalibration is required, Contact Spectro-UV at
1-866-230-7305 for assistance.
2.5 ENVIRONMENTAL SPECIFICATIONS
The AccuPRO meter series components are designed to be safe under the
following conditions:
- Indoor use;
- Altitude up to 2,000 m (6,562 ft.);
- Temperature 5°C to 40°C (41°F to 104°F);
- Maximum relative humidity 80% for temperatures up to 31°C (88°F) decreasing linearly to 50% relative humidity at 40°C (104°F);
- Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage;
- Installation Category II;
- Pollution Degree 2.
BUTTONS AND CONTROLS
- The keypad on the readout meter is equipped with three pressure-sensitive buttons that provide easy access to all functions and features.
- Press and hold down the MENU (ON/OFF) button to turn the meter ON.
- The UP and DOWN arrows to the left and right of the MENU button provide access to all displays and features. Selected menu items appear highlighted in blue. Selected Menu items can be confirmed by pressing the MENU button once.
- Pressing the DOWN arrow (while in OPERATION MODE) will reset the PEAK value (highest recorded luminescence since last reset) for visible, UV, and if applicable, blue light sources. If using the AccuPRO Plus (XP-4000), the UP arrow toggles between Visible/UV and blue light readouts.
- Press and hold the MENU button in OPERATION mode to turn the unit OFF.
- Highlighting and selecting the BACK button will return the user to the previous screen without making any changes.
OPERATION––QUICK GUIDE
- To turn the meter on, press and hold the MENU button for two seconds. It is recommended to fully charge the unit before use (see section 10.3).
- The Spectroline® loading screen will appear as the unit powers ON. Information pertaining to your device will briefly appear as well. Select INFORMATION from the MAIN MENU screen to thoroughly examine information pertaining to your device (see Section 6).
- After loading, the unit will boot directly to Operation mode, displaying visible and UV readings by default. To view levels of blue light with the AccuPro Plus (XP-4000), press the UP button.
- To change meter display settings, press MENU. Then select the SETTINGS option. When highlighted, press MENU to enter SETTINGS menu (see SETTINGS section for details).
- To take a measurement, enter Operation Mode. Shine a light source directly over the center of the sensor. The measurement will display on the interface screen.
- The meter instantly displays light level data for each wavelength (intensity/unit area).
- To turn off the meter, press and hold the MENU button for 2 seconds while in Operation mode. The display will then power off. Wait at least 30 seconds after the meter powers down before performing a RESTART (by pressing and holding MENU for 2 seconds). The AccuPRO will only power off from Operation mode, and not from the menus.
- To both conserve battery life and preserve the sensitivity of the sensor head, utilize the “Set Auto-Off” function under the “SETTINGS” menu. (See section 6).
USING AccuPRO
5.1 ON/OFF
Press and hold the MENU button to turn the meter on. Boot screen below
followed by Information screen (section 7) will display. Fully charge before
use.
Press and hold the MENU button to turn the meter ON. The Boot Screen (below)
will load, followed briefly by the Information Screen. (See Section 7 for
details about the Information Screen).
5.2 OPERATION MODE
When turned on, the AccuPRO automatically enters Operation Mode and begins
displaying light readings for white light (555nm) and UV (365nm) wavelengths.
If using the AccuPRO Plus (XP-4000), press the UP arrow to toggle between this
display and blue light (450nm) readout.
5.3 SENSOR READOUTS AND PEAK VALUE
Moving the sensor head will cause the readout values to increase or decrease
depending on the levels of illumination, with the highest recorded value
stored automatically as the PEAK value in the lower left of the readout
screen. Images on pages 5 and 6 detail each readout screen.
Shine the light source being measured directly over the center of the sensor
for the most accurate readings.
The PEAK value stores the maximum luminosity recorded since the last reset.
Simply press the DOWN arrow in operational mode to reset the PEAK value to
zero across visible, UV, and if applicable, blue light readouts.
5.4 SENSOR WAVELENGTHS & RANGES
The single sensor attached to the AccuPRO will display UV-A (365nm) readings
as well as visible light (555nm). In addition, the AccuPRO Plus will also
display blue light at the wavelength of 450nm.
Meter will display “XXXX” should readings in any measured portion of the
spectrum exceed the figures below:
Model No. | Wavelengths Recorded | Range |
---|---|---|
XP-2000 | UV-A/VIS |
UV-A (365nm) Irradiance
Visible (555nm) Illuminance| 0_100 mW/cm2
0_5,300 Lux
0_500 fc
XP-4000| UV-A/VIS/Blue
UV-A (365nm) Irradiance
Visible (555nm) Illuminance
Blue (450nm)| 0_100 mW/cm2
05,300 Lux
0 500 fc
0_100 mW/cm 2
CUSTOMIZING SETTINGS
Press MENU, then press the DOWN arrow. Select the SETTINGS option. Press MENU
again to enter the
SETTINGS menu as shown below: The SETTINGS menu contains the options displayed
below: Press UP or DOWN to highlight the feature to be modified and MENU to
access the next screen. Highlight BACK and press MENU to return to OPERATION
mode.
6.1 CHANGING UNITS OF MEASUREMENT
The AccuPRO provides users the capability of setting the default visible light
unit of measure to either foot-candles (fc) or Lux.
UV and blue light source intensity units willautorange from µW/cm2 to mW/cm2.
To adjust units, highlight and select SET UNITS from the SETTINGS menu. Press
MENU to move to the next screen, as shown below. Vis (Visible), UV, and Blue
(on XP-4000 only) will be listed, along with the BACK button to return to the
previous screen. Select the desired portion of the spectrum using MENU, then
use the arrow keys to select the desired units. The following example displays
the options for visible light: The example below shows the available units for
ultraviolet and blue light: Upon pressing MENU, a brief confirmation screen
will display (e.g., “Units set to Lux”) before returning to the main menu.
Simply press MENU again to return to Operation Mode which will display the new
units.
6.2 ADJUSTING BRIGHTNESS
To change the brightness of the screen, enter the SETTINGS menu. Select
BRIGHTNESS. Press MENU to access brightness settings and choose either
BRIGHTER or DARKER (see example below). When selecting the level of
brightness, consider the ambient
light conditions, user preferences, and battery conservation. Press the MENU
button until desired brightness level is attained. Press BACK to confirm and
exit.
6.3 LANGUAGE OPTIONS
All interactive features and displays of the AccuPRO and AccuPRO Plus are
available in English, Spanish, German, French, and Chinese. Enter the SETTINGS
menu and select the LANGUAGE option to enter the selection menu displayed
below. Upon selection of the desired language, the language menu will refresh
and appear in the newly selected language format.
NOTE: Should the language accidentally be changed to one which you cannot
understand, the first (highest) option on the menu will revert to English as
seen below. Below is the same screen viewed in Chinese. The numbers remain the
same. In this event, press UP until reaching option 1 then MENU to return to
English. The same is applicable to the other languages, with the first option
reverting to English. The languages and their corresponding numbers match
across all language settings. If you accidentally happen to select a language
difficult to understand (for instance, Chinese), select Option Number 1 to
revert back to the default English language setting. The same process is
applicable to the other languages in the selection.
INFORMATION
The INFORMATION screen, accessed via the third selection on the main menu,
displays briefly when the device is powered on, and can be viewed for as long
as desired through the menu option. The INFORMATION Screen can be accessed in
the Main Menu. By default, the INFORMATION Screen briefly displays during
startup. To closely examine the data on the INFORMATION Screen, select
INFORMATION from the Main Menu.
The serial number of your device, the date of your device’s most recent
calibration, and the software version installed on your device is available on
the INFORMATION screen. Important data including the serial number of your
unit, date of last calibration, and software version is avaiable on the
INFORMATION screen. To return to OPERATION Mode, select the BACK option.
CALIBRATION
The fourth and final option on the Main Menu is Calibration. Calibration
requires a precise configuration, and cannot be performed with the device
alone. Contact customer service at
1-866-230-7305 if you believe your
AccuPRO may require calibration.
There is no need to enter this menu under normal circumstances. However,
should you arrive on the calibration screen shown below, simply press MENU
repeatedly to exit. Zeroes will appear in the Response field each time the
MENU button is pressed. The text “Verify Bad” will display (as shown below)
and the device will then automatically exit the calibration screen, returning
to the Main Menu by default.
THEORY OF OPERATION
THE INPUT
OPTICS
The lambertian (cosine) response of the sensor head is desirable for many
measurement applications, especially those where the angle from the source to
the detector is variable or those situations where theangle from the source is
”extended,” such as in the
measurement of a fluorescent lamp at distances comparable to or shorter than
its length. In the latter case, the extended source provides radiation from
many angles, all of which must be properly ”weighted” as to their
effectiveness on the plane represented by
the sensor surface.
In actual practice, it is difficult to make a sensor conform to the ideal
response over the entire solid angle of 2TT steradians. The sensor units of
the AccuPRO meter minimize this problem by being outfitted with optimal
transmission diffusing materials for various spectral regions. These diffusion
materials are mounted close to the surface of the sensors so that the oblique
rays are not obstructed. The spectral range is selected by adding an
appropriate UV interference filter within the optical stack before the sensor
cell assembly.
THE SENSOR CELL
Photovoltaic Operation
When a p-n junction is operated with no externally applied voltage, it is
considered to be operating in the photovoltaic mode. Under this zero applied
voltage condition and low levels of incident light, the p-n junction will
generate a current proportional to the
light power incident on the active surface.
This photon-induced current, or photocurrent, will divide between the diode
parallel dynamic resistance and the parallel load resistance. The dynamic
resistance is normally a high value and is an inverse exponential function of
forward voltage. The direction of current flow will produce a voltage across
the diode that opposes the band-gap potential of the photodiode junction, thus
forward biasing it. As a result, the value of the diode dynamic resistance
(Rd) drops exponentially as the irradiance increases and the photo generated
voltage is a quasilogarithmic function of diode irradiance when the external
load resistance is considered. Another major disadvantage is that Rd typically
has a wide spread of values over different production batches.
One way of achieving sufficiently low load resistance and eliminating the
effect of the diode parallel resistance is to feed the photocurrent into the
virtual ground of an operational amplifier. The output voltage is the result
of the photocurrent being driven by the
amplifier through the feedback resistor and the input impedance Rin = Rf/A
where A is the open loop gain and Rf the feedback resistor. This circuit has a
linearresponse and is low noise due to the almost complete elimination of
leakage current with the zero bias. This results in a proportional voltage
being presented to the signal conditioning section of the electronics.
WARRANTY, MAINTENANCE AND BATTERY REPLACEMENT
10.1 WARRANTY
The warranty policy for the AccuPRO is provided on the Certificate of Limited
Warranty enclosed separately with each unit.
NOTE: For assistance of any kind, including help with a meter under
warranty, contact the Customer Service Department at Spectro-UV. In the United
States and Canada, call toll-free
1-866-230-7305. Include the model
number, serial number and date of purchase. If return of the unit is deemed
necessary, shipping instructions will be provided. If an estimate of charges
for nonwarranty work or other service work is required, a quote will be
furnished upon evaluation of the unit. Out-of-warranty service work will not
be performed without customer approval.
10.2 PREVENTIVE MAINTENANCE
- Immediately clean all spilled materials from the unit and wipe dry. If necessary, moisten a cloth with soap and water to clean plastic surfaces and the sensor head. The sensor surface should be rinsed with ethanol to remove any residual soap and/or organic contaminants.
- Whenever possible, avoid exposure or use in areas that are subject to temperature and humidity extremes, vibration or mechanicalshock, dust or corrosive fumes, or strong electrical or electromagnetic interference.
- It is recommended that the unit be returned tothe factory or a recognized SpectrolineAccuPRO calibration service center for a complete overall check and recalibration atleast every 6 or 12 months, depending on yourfacility’s Standard Operating Procedures. Before returning the units to our factory, contact the Customer Service Department for shipping instructions.
- When the AccuPRO is not in use, store it in a location free from temperature extremes, dust or corrosive fumes, and vibration or mechanical shock.
- If storing for an extended period, place the AccuPRO in its carrying case.
10.3 BATTERY SAFETY AND CHARGING
- The AccuPRO radiometer must be charged overnight or for at least 8 hours when the battery is replaced or after a battery reset (removal and re-installation of the same battery).
- CAUTION: Do not use alkaline batteries in the AccuPRO as they may explode causing damage to the circuitry and possibly the operators. Only use the 700mAh,1.2V rechargeable batteries provided with the AccuPRO and available from Spectroline® (P/N: 129227).
- The AccuPRO should be charged with the power OFF. Charging the unit for long periods while operating is not advised because it will cause the charging/display circuit to malfunction. Should this take place, the unit will need to charge for 8+ hours while turned off for the circuit to reset.
- The battery indicator will turn red at 30%. It is recommended that the AccuPRO be powered down and charged when it reaches this level. At 10% battery life, the unit will automatically power down.
- Disconnect the charger from the meter when not in use.
- While charging, it is recommended to have the device turned OFF.
- Off-State Discharge Rate: The AccuPRO™ battery will naturally discharge approximately 3-4% per day – even when the unit is turned off. Be sure to keep an accurate and up-to-date charging schedule.
CHARGING
- Plug battery charger into AC power outlet.
- Firmly connect charger to port located on the right side of the unit, just below the screen.
- Battery percentage, located in the top right corner of the screen, will begin to increase as the unit charges. It might be necessary to change screens for the battery readout to update.
- A full charge will take 8 hours and is good for approximately 6 hours of operation.
- While charging, it is recommended to have the device turned OFF.
10.4 TROUBLESHOOTING
If unit experiences intermittent screen flickering, this is likely due to an
incomplete charging cycle. To fix, power on the unit and allow for batteries
to fully discharge than plug unit in and power the unit off. Allow unit to
fully charge for 8 hours.
www.Spectro-UV.com
4 Dubon Ct., Farmingdale, NY 11735
866-230-7305