OHAUS Valor 1000 Series Compact Scale Instruction Manual
- June 4, 2024
- Ohaus
Table of Contents
OHAUS Valor 1000 Series Compact Scale
INTRODUCTION
This manual contains installation, operation and maintenance instructions for the Ohaus ValorTM 1000 Series Scales. Please read the manual completely before using the scale.
Safety Information
Definition of Signal Warnings and Symbols
Safety notes are marked with signal words and warning symbols. These show
safety issues and warnings. Ignoring the safety notes may lead to personal
injury, damage to the instrument, malfunctions and false results.
WARNING For a hazardous situation with medium risk, possibly resulting in
severe injuries or death if not avoided.
CAUTION For a hazardous situation with low risk, resulting in damage to
the device or the property or in loss of data, or minor or medium injuries if
not avoided.
ATTENTION For important information about the product. May lead to
equipment damage if not avoided.
NOTE For useful information about the product.
Warning Symbols
Safety Precautions
CAUTION: Read all safety warnings before installing, making connections, or servicing this equipment. Failure to comply with these warnings could result in personal injury and/or property damage. Retain all instructions for future reference.
- Before connecting power, verify that the AC adapter’s input voltage range and plug type are Ð compatible with the local AC mains power supply.
- Do not position the equipment such that it is difficult to reach the power connection.
- Make sure that the power cord does not pose a potential obstacle or tripping hazard.
- Operate the equipment only under ambient conditions specified in these instructions.
- The equipment is for indoor use only.
- Do not operate the equipment in wet, hazardous or unstable environments.
- Do not allow liquids to enter the equipment.
- Do not load the equipment above it’s rated capacity.
- Do not drop loads on the platform.
- Do not place the equipment upside down on the platform.
- Use only approved accessories and peripherals.
- Disconnect the equipment from the power supply when cleaning.
- Service should only be performed by authorized personnel.
WARNING: Never work in an environment subject to explosion hazards! The
housing of the instrument is not gas tight. (Explosion hazard due to spark
formation, corrosion caused by the ingress of gases).
CAUTION: Battery is to be replaced only by an authorized Ohaus service
dealer. Risk of explosion can occur if the rechargeable battery is replaced
with the wrong type or if it is not properly connected. Dispose of the
rechargeable battery according to local laws and regulations.
Intended Use
This instrument is intended for use in businesses and light industry. It shall
only be used for measuring the parameters described in these operating
instructions. Any other type of use and operation beyond the limits of
technical specifications, without written consent from OHAUS, is considered as
not intended. This instrument complies with current industry standards and the
recognized safety regulations; however, it can constitute a hazard in use.
If the instrument is not used according to these operating instructions, the
intended protection provided by the instrument may be impaired.
Controls
Button | Functions |
---|
| On/Off
- Press (when off): Turn the scale on
- Long Press (when on): Turn the scale off
Zero
- Short Press (when on): Zero the scale
Yes
- Press (in Menu): Confirm (Enter)
**| Tare
Short Press: Enter / clear a Tare value Menu
Long Press: Enter User Menu No
**Press (in Menu): Move to next menu or selection
Notes:
Short Press: Press less than 3 seconds.
Long Press: Press and hold for more than 3 seconds.
Display
Item | Description |
---|---|
1 | Battery charge symbol |
2 | Center of Zero symbol |
3 | Negative symbol |
4 | Stable weight symbol |
5 | NET symbol |
INSTALLATION
Unpacking
Carefully remove your Valor 1000 scale and each of its components from the
package. Save the packaging to ensure safe storage and transport.
Please read the manual completely before installing and using the Valor 1000
scale to avoid incorrect operation.
Components included:
-
Scale
-
Quick Start Guide or NLF User Guide or User Manual
Scan the QR code to obtain the user manual Power adapter + attaching plug -
Stainless steel pan*
-
In-use cover*
-
4 dry batteries*
-
Rechargeable battery*
-
Sealing cover, screw, wire and lead seal*
*The included components vary depending on the region.
Location
Use the scale on a firm, steady surface. Avoid locations with excessive air current, vibrations, heat sources, or rapid temperature changes. Adjust the leveling feet so the bubble is centered in the circle.
Power
Connect the AC Adaptor to the AC mains supply. Connect the plug to the DC jack
on the bottom of the scale. The scale may be operated on the AC Adapter
connected to mains supply or 4 alkaline batteries.
Note: The AC adapter maybe optional or included with the scale, depending
on the region.
Rechargeable Battery Power
Allow the battery to charge for 12 hours before using the scale on battery power. The instrument can be operated during charging, and the battery is protected against overcharging. For maximum operating time, the battery should be charged at room temperature.
Note: Availability is depending on region as accessory.
WARNING: Only charge the battery when the ambient temperature is between 0 and 140 °F / 0 and 40 °C.
Initial Calibration
When the scale is operated for the first time, a Span Calibration is
recommended to ensure accurate weighing results. Before performing the
calibration, be sure to have the appropriate calibration weight. You can also
contact an OHAUS servicing scale dealer for more information.
When the weighing unit is set to g or kg, the scale is calibrated in kg. When
the weighing unit is set to lb, oz or lb:oz, the scale is calibrated in lb.
Refer to section 4.1 for Span calibration procedures and section 4.3 for unit setting.
Maximum Capacity (kg)| Maximum Capacity (lb)| Span
Calibration Mass (kg)| Span Calibration Mass (lb)
---|---|---|---
2 kg| 5 lb| 2 kg| 5 lb
3 kg| 6 lb| 3 kg| 6 lb
5 kg| 10 lb| 5 kg| 10 lb
6 kg| 15 lb| 6 kg| 15 lb
10 kg| 20 lb| 10 kg| 20 lb
15 kg| 30 lb| 15 kg| 30 lb
20 kg| 50 lb| 20 kg| 50 lb
30 kg| 60 lb| 30 kg| 60 lb
OPERATION
Turning Scale On/Off
Press On/Off to turn the scale on. Then the scale displays the software version. After the scale is turned on, long press On/Off for more than 3 seconds to turn the scale off.
Weighing
Once the scale has a stable zero weight and the tare weight of any packaging or container has been captured, place the product being weighed on the platform — the display will show the weight of the product in the selected unit of measure.
SETTINGS
The Menu function in the scale allows several settings to be changed. Long press Menu for 3 seconds until either C.A.L or L.I.G.H.t is displayed. Then release the Menu key.
Calibration [C.A.L]
For non-approved models, calibration is always allowed. But the Calibration
menu is locked and hidden in user menu in some regions.
To show “Calibration” in user menu, you need to enter the service menu.
- Press and hold and at the same time for more than 10 seconds until you see ramMP. Release both keys to enter the service menu.
- Short press No until you see LOCK. Press Yes to enter the LOCK menu.
- Press No to change to NO.
- Press Yes to save the change.
- Short press No until you see End.
- Press Yes to exit the service menu.
Now the “Calibration” is available in user menu.
Once the scale displays C.A.L, press Yes to enter this menu to perform a calibration.
Press No to exit Calibration and to move on to next menu.
For approved models, calibration is not permitted after LFT switch is pressed (LFT ON).
The calibration menu is only shown when the scale is not in LFT/Approved mode before verification and sealing. Refer to section 5. Legal For Trade for more information.
SPAN [SPAN]
Span calibration uses two calibration points, one at zero load and the other
at specified full load (span). For detailed calibration mass information,
please refer to the table in section 2.4.
With no load on the balance, enter the Span Calibration to begin the process.
- Once the scale displays C.A.L, press Yes to enter the calibration menu and press Yes again to perform a Span calibration.
- After the scale shows –C–, the calibration mass value will be shown in the screen. Place weight(s) of shown value and press Yes.
- Then the scale shows –C–. Once the span calibration is completed successfully, the scale displays Done and exits the user menu automatically.
If incorrect calibration masses are used, the scale displays Cal E and exits the user menu automatically.
To abort calibration at any point, press No.
Lin [LIN]
When performing a linearity calibration, the scale will prompt you for a zero weight the platform without anything on it and weight display at zero — a Midpoint weight equal to 50% of the Span weight, and a Span weight equal to the full capacity of the scale.
Maximum Capacity (kg)| Maximum Capacity (lb)| Linearity
Calibration Points (kg)| Linearity
Calibration Points (lb)
---|---|---|---
2 kg| 5 lb| 0, 1 kg, 2 kg| 0, 3lb, 5 lb
3 kg| 6 lb| 0, 1.5kg, 3 kg| 0, 3lb, 6 lb
5 kg| 10 lb| 0, 3kg, 5 kg| 0, 5lb, 10 lb
6 kg| 15 lb| 0, 3kg, 6 kg| 0, 10lb, 15 lb
10 kg| 20 lb| 0, 5kg, 10 kg| 0, 10lb, 20 lb
15 kg| 30 lb| 0, 10kg, 15 kg| 0, 15lb, 30 lb
20 kg| 50 lb| 0, 10kg, 20 kg| 0, 30lb, 50 lb
30 kg| 60 lb| 0, 15kg, 30 kg| 0, 30lb, 60 lb
With no load on the balance, enter the Linearity Calibration to begin the process.
- After entering the calibration menu, press No to move to linearity calibration menu LIN and press Yes to start a linearity calibration.
- After the scale shows –C–, the calibration mass value (a Midpoint weight) will be shown in the screen. Place weight(s) of shown value and press Yes.
- The scale shows–C–, then the calibration mass value (a span weight) will be shown in the screen. Place weight(s) of shown value and press Yes.
- The scale shows –C–. Once the linearity calibration is completed successfully, the scale displays Done and exits the user menu automatically.
If incorrect calibration masses are used, the scale displays Cal E and exits the user menu automatically.
To abort calibration at any point, press No.
End [End]
Press Yes to move to the next Menu, or press No to return to Span calibration.
Light [L.I.G.H.t]
This sets the operation of the display’s backlight.
ON = always on
OFF = always off
AUTO = Turns on when a button is pressed or the displayed weight changes.
Alternatively, turns off the backlight when the displayed weight does not
change for more than 10 seconds.
Unit [U.N.I.t]
The Valor 1000 allows one weighing unit to be used at a time. Units can be changed in this menu. Available units vary for different regions or models.
Unit | Display |
---|---|
Gram | g |
Kilogram | kg |
Pound | lb |
Ounce | oz |
Pound: Ounce | lb:oz |
Press Yes to enter the Unit menu. The currently used weighing unit will be
shown. To change the unit, press No to move to the next unit.
Press Yes to set the displayed unit to the weighing unit and exit the Unit
menu.
Auto Shut-off [a.-O.F.F]
Set the automatic shut off time for the scale.
A00 = the scale powers off after 5 minutes of no activity when it is battery
powered only; Auto shut-off function is disabled when scale is powered by AC
adapter. A01 = powers off after 1 minute of no activity
A05 = powers off after 5 minutes of no activity
A10 = powers off after 10 minutes of no activity
Off = disabled
Beeper Key [BP.key]
ON = turns on the beeper sound when you press any key.
OFF = turns off the beeper sound when you press any key.
End [e.n.d]
Press the No to return to the C.A.L menu, or press the Yes to exit to normal operation.
LEGAL FOR TRADE (for approvable T models)
When the scale is used in trade or a legally controlled application, it must be set up, verified and sealed in accordance with local weights and measures regulations. It is the responsibility of the purchaser to ensure that all pertinent legal requirements are met. As the requirements vary by jurisdiction, the purchaser is advised to contact their local weights and measures office for instructions about putting the balance into service.
Settings
Before verification and sealing, perform the following steps:
- Confirm that the available units are permitted by the local weights and measures regulations.
- Perform a calibration as explained in Section 4.1.
- Press the LFT switch under the LFT switch cover as shown below. This cover is located at the bottom of the scale.
- Remove the LFT switch cover
- Turn on the scale
- Press the LFT switch. After that you will see “LFt.ON” is displayed on the screen.
Verification
A weights and measures official must perform the verification procedure. Contact the local weights and measures office for more information.
Sealing
After the scale has been verified, it must be sealed by the weights and measures official to prevent undetected access to the legally controlled settings. Refer to the illustrations below for the sealing.
MAINTENANCE
Cleaning
WARNING: Electric Shock Hazard. Disconnect the equipment from the power
supply before cleaning.
Make sure that no liquid enters the interior of the instrument.
Attention: Do not use solvents, harsh chemicals, ammonia or abrasive
cleaning agents.
The housing may be cleaned with a cloth dampened with a mild detergent if necessary.
Troubleshooting
The following table lists common problems and possible causes and remedies. If the problem persists, contact OHAUS or your authorized dealer.
TABLE 5-1. TROUBLESHOOTING.
Display | Possible Cause | Possible Solution |
---|---|---|
CAL E | The environment is not stable. | Move balance to a suitable location and |
calibrate again.
Incorrect calibration masses are used.
| Use calibration masses to calibrate accodrding to the correct calibration
points shown in the screen during calibration procedures.
System error| If error persists, please contact OHAUS service (www.ohaus.com
– More – Contact Us)
Err 8.1| A wrong weighing pan is used when power on.| Use OHAUS original
weighing pan before power on.
The load on the pan is over the initial weight setting.| Remove the load from
the pan before power on.
Err 8.2| A wrong weighing pan is used when power on.| Use OHAUS original
weighing pan before power on.
The pan is not installed when power on.| Install the weighing pan before power
on.
Err 8.3| The weight on the pan is too heavy.| Reduce sample size until the
weight is within the weighing capacity.
Err 8.4| The pan is not properly installed.| Install the weiging pan properly.
Err 8.5| The tare value is a negative value.| Press Zero instead of Tare.
Err 9.5
Err 13
| System error| Disconnect the power and restart.
If error persists, please contact OHAUS service
(www.ohaus.com – More – Contact Us)
Lo.Bat| Battery very low| Replace dry batteries or charge rechargeable battery.
–NO–
| The value is out of zero range. The zero range is dependent on region.|
Perform zero again according to regional requirement.
——| The environment is not stable.| Move balance to a suitable location.
Service Information
If the troubleshooting section does not resolve or describe your problem, contact your authorized OHAUS service agent. For service assistance or technical support in the United States call toll-free 1-800-672-7722 ext. 7852 between 8:00 AM and 5:00 PM EST. An OHAUS product service specialist will be available to provide assistance Monday through Friday. Outside the United States, Canada and Puerto Rico, please visit our web site, www.ohaus.com to locate the OHAUS office nearest you.
TECHNICAL DATA
- Equipment Ratings: Indoor use only
- Altitude: 2000m
- Operating temperature: 0 to 40°C
- Humidity: Maximum relative humidity 80% for temperatures up to 31 °C decreasing linearly to 50% relative humidity at 40°C.
- Electrical supply: 12VDC, 0.5A. For use with certified or approved power supply, which must have a SELV and limited energy circuit output.
- Voltage fluctuations: Mains supply voltage fluctuations up to ±10% of the nominal voltage.
- Overvoltage category (Installation category): II
- Pollution degree: 2
Specifications
TABLE 7-1. Specifications
**Model*** | V12P3 | V12P6 | V12P15 | V12P30 |
---|---|---|---|---|
Capacity × Readability (Max x d non- approved) | 6 lb x 0.001 lb |
3 kg x 0.0005 kg
3,000 g x 0.5 g
96 oz x 0.02 oz
6 lb x 0.02 oz
| 15 lb x 0.002 lb
6 kg x 0.001 kg
6,000 g x 1 g
240 oz x 0.05 oz
15 lb x 0.05 oz
| 30 lb x 0.005 lb
15 kg x 0.002 kg
15,000 g x 2 g
480 oz x 0.1 oz
30 lb x 0.1 oz
| 60 lb x 0.01 lb
30 kg x 0.005 kg
30,000 g x 5 g
960 oz x 0.2 oz
60 lb x 0.2 oz
Maximum Displayed Resolution (gram)| 1:6,000| 1:6,000| 1:7,500| 1:6,000
Maximum Displayed
Resolution (pound)
| 1:6,000| 1:7,500| 1:6,000| 1:6,000
Weighing Units| g, kg, lb, oz, lb:oz
Application Mode| Weighing
Tare Range| To capacity by subtraction
Stabilization Time| ≤2 seconds
Power requirements| 4 x D cell Alkaline battery, AC adapter or rechargeable
battery
Calibration| Digital with external weight
Typical Battery Life| 1500 hours with backlight turned off
Construction| ABS plastic housing, stainless steel pan, plastic in-use cover
Safe Overload| 150% of capacity
Specified Temperature Range| 32 to 104° F / 0 to 40 °C
Display Type| Single display – LCD with white backlight
Display Size| 1.0 in / 25.4 mm
Pan Size| 9.65 x 7.48 in / 245 x 190 mm
Scale Dimensions (W x D x H)| 10.2 x 10.6 x 4.8 in / 260 x 270 x 122 mm
Shipping Dimensions (W x D x H)| 12.6 x 12.6 x 6.9 in / 319 x 319 x 175 mm
Net Weight| 5.5 lb / 2.5 kg
Shipping Weight| 6.8 lb / 3.1 kg
- Certain regions are available with V12PRx models, which are standard with AC adapter and rechargeable battery.
**Availability is dependent on region.
TABLE 7-2. Specifications
Model | V12P2T | V12P5T | V12P10T | V12P20T |
---|---|---|---|---|
Capacity × Readability (Max x d non- approved) | 5 lb x 0.0005 lb |
2 kg x 0.0002 kg
2,000 g x 0.2 g
80 oz x 0.01 oz
| 10 lb x 0.001 lb
5 kg x 0.0005 kg
5,000 g x 0.5 g
160 oz x 0.02 oz
| 20 lb x 0.002 lb
10 kg x 0.001 kg
10,000 g x 1 g
320 oz x 0.05 oz
| 50 lb x 0.005 lb
20 kg x 0.002 kg
20,000 g x 2 g
800 oz x 0.1 oz
Maximum Displayed Resolution (gram)| 1:10,000| 1:10,000| 1:10,000| 1:10,000
Maximum Displayed Resolution (pound)| 1:10,000| 1:10,000| 1:10,000| 1:10,000
Certified Capacity ×
Readability
(Max x e approved)
| 5 lb x 0.002 lb
2 kg x 0.001 kg
2,000 g x 1 g
80 oz x 0.05 oz
| 10 lb x 0.005 lb
5 kg x 0.002 kg
5,000 g x 2 g
160 oz x 0.1 oz
| 20 lb x 0.01 lb
10 kg x 0.005 kg
10,000 g x 5 g
320 oz x 0.2 oz
| 50 lb x 0.02 lb
20 kg x 0.01 kg
20,000 g x 10 g
800 oz x 0.5 oz
Approved Resolution
(gram)
| 1:2,000| 1:2,500| 1:2,000| 1:2,000
Approved Resolution (pound)| 1:2,500| 1:2,000| 1:2,000| 1:2,500
Class| Class III NTEP / Measurement Canada
Weighing Units| g, kg, lb, oz
Application Mode| Weighing
Tare Range| To capacity by subtraction
Stabilization Time| ≤2 seconds
Power requirements| 4 x Alkaline battery (not included) or AC adapter
(included)
Calibration| Digital with external weight
Typical Battery Life| 750 hours with backlight turned off
Construction| ABS plastic housing, stainless steel pan, in-use cover
Safe Overload| 150% of capacity
Communication| Optional RS232
Specified Temperature Range| 32 to 104° F / 0 to 40 °C
Display Type| Dual display – LCD with white backlight
Display Size| 1.0 in / 25.4mm
Pan Size| 9.65 x 7.48 in / 245 x 190 mm
Scale Dimensions (W x D x H)| 10.2 x 10.6 x 4.8 in / 260 x 270 x 122 mm
Shipping Dimensions (W x D x H)| 12.6 x 12.6 x 6.9 in / 319 x 319 x 175 mm
Net Weight| 5.5 lb / 2.5 kg
Shipping Weight| 6.8 lb / 3.1kg
TABLE 7-3. Specification
Model | V12P3T | V12P6T | V12P15T | V12P30T |
---|---|---|---|---|
Capacity × Readability (Max x d non approved) |
3 kg x 0.0002 kg
3000 g x 0.2 g
|
6 kg x 0.0005 kg
6000 g x 0.5 g
|
15 kg x 0.001 kg
15000 g x 1 g
|
30 kg x 0.002 kg
30000 g x 2 g
Maximum
Displayed Resolution (d)| 1:15,000| 1:12,000| 1:15,000| 1:15,000
Model| V12P3T- M| V12P6T- M| V12P15T- M|
V12P30T- M
Capacity ×
Readability
(Max x e approved)| 3 kg x 0.001kg 3000 g x 1 g| 6 kg x 0.002 kg
6000 g x 2 g
| 15 kg x 0.005 kg
15000 g x 5 g
| 30 kg x 0.01 kg
30000 g x 10 g
Approved Resolution| 1:3,000| 1:3,000| 1:3,000| 1:3,000
Weighing Units| g, kg
Application Mode| Weighing
Tare Range| To capacity by subtraction
Stabilization Time| ≤2 seconds
Power requirements| 4 x Alkaline battery (included), AC adapter (optional) or
rechargeable battery (optional)
Calibration| Digital with external weight
Typical Battery Life| 750 hours with backlight turned off
Construction| ABS plastic housing, stainless steel pan
Safe Overload| 150% of capacity
Communication| Optional RS232
Specified Temperature
Range| 0 to 40 °C
Display Type| Dual display – LCD with white backlight
Display Size| 25.4mm
Pan Size| 245 x 190 mm
Scale Dimensions (W x D x H)| 260 x 270 x 122 mm
Shipping Dimensions (W x D x H)| 319 x 319 x 175 mm
Net Weight| 2.5 kg
Shipping Weight| 3.1 kg
Compliance
Compliance to the following standards is indicated by the corresponding mark on the product.
Mark | Standard |
---|---|
CE | This product complies with the applicable harmonized standards of EU |
Directives 2011/65/EU (RoHS), 2014/30/EU (EMC), 2014/35/EU (LVD) and
2014/31/EU (NAWI). The EU Declaration of Conformity is available online at
www.ohaus.com/ce.
| This product complies with the EU Directive 2012/19/EU (WEEE) and 2006/66/EC
(Batteries). Please dispose of this product in accordance with local
regulations at the collecting point specified for electrical and electronic
equipment.
For disposal instructions in Europe, refer to
www.ohaus.com/weee.
| EN 61326-1
Important notice for V12P…-M verified weighing instruments in the EU
When the instrument is used in trade or a legally controlled application, it
must be set up, verified and sealed in accordance with local weights and
measures regulations. It is the responsibility of the purchaser to ensure that
all pertinent legal requirements are met.
Weighing Instruments verified at the place of manufacture bear the following
supplementary metrology marking on the descriptive plate.
Weighing Instruments to be verified in two stages have no supplementary
metrology marking on the descriptive plate. The second stage of conformity
assessment must be carried out by the applicable weights and measures
authorities.
If national regulations limit the validity period of the verification, the
user of the weighing instrument must strictly observe the re-verification
period and inform the weights and measures authorities.
As verification requirements vary by jurisdiction, the purchaser should
contact their local weights and measures office if they are not familiar with
the requirements.
ISED Canada Compliance Statement:
This Class B digital apparatus complies with Canadian ICES-003.
ISO 9001 Registration
The management system governing the production of this product is ISO 9001
certified.
FCC Supplier Declaration of Conformity
Unintentional Radiator per 47CFR Part B
Trade Name: OHAUS CORPORATION
Model or Family identification: V12…
Issuing Party that Assembled the Product:
Ohaus Instruments (Changzhou) Co., Ltd.
2F, 22 Block, 538 West Hehai Road, Xinbei District, Changzhou
Jiangsu 213022
China
Phone: +86 519 85287270
Responsible Party – U.S. Contact Information:
Ohaus Corporation
7 Campus Drive, Suite 310
Parsippany, NJ 07054
United States
Phone: +1 973 377 9000
Web: www.ohaus.com
FCC Compliance Statement:
Note: This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses and
can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
LIMITED WARRANTY
OHAUS products are warranted against defects in materials and workmanship from
the date of delivery through the duration of the warranty period. During the
warranty period OHAUS will repair, or, at its option, replace any component(s)
that proves to be defective at no charge, provided that the product is
returned, freight prepaid, to OHAUS.
This warranty does not apply if the product has been damaged by accident or
misuse, exposed to radioactive or corrosive materials, has foreign material
penetrating to the inside of the product, or as a result of service or
modification by other than OHAUS. In lieu of a properly returned warranty
registration card, the warranty period shall begin on the date of shipment to
the authorized dealer. No other express or implied warranty is given by OHAUS
Corporation. OHAUS Corporation shall not be liable for any consequential
damages.
As warranty legislation differs from state to state and country to country,
please contact OHAUS or your local OHAUS dealer for further details.