OHAUS 3000 Series Defender Bench Scale Instruction Manual
- October 30, 2023
- Ohaus
Table of Contents
3000 Series Defender Bench Scale
Instruction Manual
INTRODUCTION
This manual covers installation, and maintenance instructions for the Ohaus
Defender™ 3000 Series Base. Please read this manual completely before
installation and operation.
SAFETY PRECAUTIONS
For safe and dependable operation of this product, please comply with the following precautions:
- Operate the base only under ambient conditions specified in these instructions
- Ensure that the load cell cable does not pose an obstruction or tripping hazard
- Do not operate the base in hazardous environments or unstable locations
- Do not drop loads on the base
- Do not lift the base by the top frame; always lift from the bottom frame when moving the base
- Service should only be performed by authorized personnel
Intended Use
Use the instrument exclusively for weighing as described in the operating
instructions. Any other type of use and operation beyond the limits of
technical specifications without written consent from OHAUS, is considered as
not intended. This instrument complies with current industry standards and the
recognized safety regulations; however, it can constitute a hazard in use. If
the instrument is not used according to these operating instructions, the
intended protection of the instrument may be compromised and OHAUS assumes no
liability.
INSTALLATION
Unpacking
Unpack and inspect the product to make sure that all components have been
included. The package includes the following:
- Base
- Weighing Pan
- Warranty Card
- Instruction Manual
When purchased as a complete Defender 3000 Series Scale, will also include:
- 3000 Series Indicator
- Column Assembly
Assembly
Weighing Pan
Place the weighing pan securely over the rubber load pads on the top frame of
the base.
Wiring Connections
Connect the load cell cable to an indicator using the wiring codes in Table 1
below. Refer to the indicator manual for information about load cell
connections, setup and calibration.
Note: When purchased as a Defender Series Scale, the base is already pre-
wired to the 3000 Series Indicator.
TABLE 1. SIX WIRE LOAD CELL CONNECTION.
FUNCTION | WIRE COLOR |
---|---|
+ Excitation | Green |
– Excitation | Black |
+ Signal | Red |
– Signal | White |
+ Sense | Blue |
– Sense | Brown |
Shield | Yellow |
Selecting the Location
To ensure accuracy, proper performance and safety, locate and operate the base
on a stable, level surface. Avoid locations with rapid temperature changes or
excessive dust, air currents, vibrations, electromagnetic fields or heat.
Level the base by adjusting the four leveling feet until the bubble in the
level indicator (located at the rear of the base) is centered. A wrench may be
needed to loosen the locking nut above each leveling foot. When the base is
level, retighten the locking nuts up against the base to lock each foot into
place.
Note: Ensure that the base is level each time its location is changed.
MAINTENANCE
Cleaning
The base components should be kept clean and free of excessive material build
up.
- Damp cloth with water and a mild detergent may be used to wipe clean the external surfaces—do not use acids, alkalis, strong solvents or abrasive materials and chemicals
Troubleshooting
Aside from installing components and leveling adjustments, the Defender Series
Base does not require any other adjustments as shipped from the factory.
Operational difficulties that may be encountered can often be traced to simple
causes such as:
- Loose or incorrect wiring connections
- Obstructions to the base frame
- Unstable environments
- Incorrect calibration or setup of the indicator
If the troubleshooting section does not resolve or describe your problem, contact your authorized Ohaus service agent. For service assistance or technical support in the United States call toll-free 1-800-526-0659 between 8:00 AM and 5:00 PM EST. An Ohaus product service specialist will be available to provide assistance. Outside the USA, please visit our web site, www.ohaus.com to locate the Ohaus office nearest you.
TECHNICAL DATA
Technical Specifications
The technical data is valid under the following ambient conditions:
Ambient temperature: –10ºC to +40 ºC
Relative humidity: 10% to 90% relative humidity, non-condensing
Height above sea level: Up to 4,000m
TABLE 2a. SPECIFICATIONS
MODEL| D15BR| D30BR| D60BR| D60BL| D150BL|
D150BX| D300BX
---|---|---|---|---|---|---|---
Capacity| 15 kg| 30 kg| 60 kg| 60 kg| 150 kg| 150 kg| 300 kg
Maximum Resolution| 6000d| 6000d| 6000d| 6000d| 7500d| 7500d| 6000d
Pan Dimensions| 305 x 355 x 75 mm| 420 x 550 x 80 mm| 500 x 650 x 90 mm
Base Construction| Stainless Steel platform with painted steel frame and
rubber leveling feet
Safe Overload Capacity| 125% of capacity
Repeatability (std. deviation)| 1d
Linearity| ±1d
Load Cell Cable| 2 m L x 6-wire| 2.3 m L x 6-wire
Load Cell Type| 350 Ohm, aluminum, single point
Load Cell Excitation| 5-15V DC/AC
Load Cell Rated Output| 2mV/V
Load Cell Protection| IP67
Net Weight| 9 kg| 16 kg| 28 kg
Shipping Weight| 11 kg| 19 kg| 32 kg
MODEL| D15VR| D30VR| D60VR| D60VL| D150VL|
D150VX| D300VX
---|---|---|---|---|---|---|---
Capacity| 15 kg| 30 kg| 60 kg| 60 kg| 150 kg| 150 kg| 300 kg
Maximum Resolution| 6000d| 6000d| 6000d| 6000d| 7500d| 7500d| 6000d
Load Cell Capacity| 30 kg| 50 kg| 100 kg| 150 kg| 300 kg| 300 kg| 500 kg
Pan Dimensions| 305 x 355 x 75 mm| 420 x 550 x 80 mm| 500 x 650 x 90 mm
Base Construction| Stainless Steel platform with stainless steel frame and
rubber leveling feet
Repeatability (std. deviation)| 1d
Linearity| ±1d
Load Cell Cable| 2 m L x 6-wire| 2.3 m L x 6-wire
Load Cell Type| 350 Ohm, stainless steel, single point
Load Cell Excitation| 5-15V DC/AC
Load Cell Rated Output| 2mV/V
Load Cell Protection| IP67
Safe Overload Capacity| 125% of capacity
Net Weight| 9 kg| 21 kg| 33 kg
Shipping Weight| 11 kg| 24 kg| 37 kg
Drawings
TABLE 3. DIMENSIONS.
Base | A | B | C | D | E |
---|---|---|---|---|---|
Pan Depth | Pan Width | **Height of Pan to Surface of Table*** | |||
Height of Pan | Depth Including Level Indicator | ||||
D15BR, D30BR, D60BR | 355 mm | 305 mm | 123 mm | 75 mm | 380 mm |
D60BL, D150BL | 550 mm | 420 mm | 136 mm | 80 mm | 575 mm |
D150BX, D300BX | 650 mm | 500 mm | 143 mm | 90 mm | 675 mm |
D15VR, D30VR, D60VR | 355 mm | 305 mm | 119 mm | 75 mm | 380 mm |
D60VL, D150VL | 550 mm | 420 mm | 145 mm | 80 mm | 575 mm |
D150VX, D300VX | 650 mm | 500 mm | 154 mm | 90 mm | 675 mm |
*For leveling purposes, the feet may be extended up to an additional 11mm.
TABLE 4. ACCESSORIES.
Description | Part Number |
---|---|
Column Kit, 35 cm painted steel | 80251743 |
Column Kit, 70 cm painted steel | 80251744 |
Column Kit, 35 cm stainless steel | 80251745 |
Column Kit, 70 cm stainless steel | 80251746 |
Compliance
This product conforms to the EU Directive 2011/65/EU (RoHS).
The complete declaration of Conformity is available online at
www.ohaus.com/ce.
Disposal
In conformance with the European Directive 2002/96/EC on Waste Electrical and
Electronic Equipment (WEEE) this device may not be disposed of in domestic
waste. This also applies to countries outside the EU, per their specific
requirements.
Please dispose of this product in accordance with local regulations at the
collecting point specified for electrical and electronic equipment.
If you have any questions, please contact the responsible authority or the
distributor from which you purchased this device.
Should this device be passed on to other parties (for private or professional
use), the content of this regulation must also be related.
For disposal instructions in Europe, refer to
www.ohaus.com/weee.
Thank you for your contribution to environmental protection.
ISO 9001 Registration
In 1994, Ohaus Corporation, USA, was awarded a certificate of registration to
ISO 9001 by Bureau Veritus Quality International (BVQI), confirming that the
Ohaus quality management system is compliant with the ISO 9001 standard’s
requirements. On May 21, 2009, Ohaus Corporation, USA, was re-registered to
the ISO 9001:2008 standard.
LIMITED WARRANTY
Ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period. During the warranty period Ohaus will repair, or, at its option, replace any component(s) that proves to be defective at no charge, provided that the product is returned, freight prepaid, to Ohaus. This warranty does not apply if the product has been damaged by accident or misuse, exposed to radioactive or corrosive materials, has foreign material penetrating to the inside of the product, or as a result of service or modification by other than Ohaus. In lieu of a properly returned warranty registration card, the warranty period shall begin on the date of shipment to the authorized dealer. No other express or implied warranty is given by Ohaus Corporation. Ohaus Corporation shall not be liable for any consequential damages. As warranty legislation differs from state to state and country to country, please contact Ohaus or your local Ohaus dealer for further details.
Ohaus Corporation
7 Campus Drive
Suite 310
Parsippany, NJ 07054, USA
Tel: (973) 377-9000
Fax: (973) 944-7177
With offices worldwide
www.ohaus.com P/N 80251065 E © 2016 Ohaus Corporation,
all rights reserved
Printed in China
References
- Lab Balance | Industrial Scales | Lab Equipment | Lab Instruments | OHAUS
- Lab Balance | Industrial Scales | Lab Equipment | Lab Instruments | OHAUS
- CE Declarations | OHAUS
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>