VELLEMAN PCSU1000 PC Oscilloscope Bit Digital Storage DSO Spectrum Analyzer User Manual
- June 13, 2024
- Velleman
Table of Contents
DSO USERMANUAL PCSU1000
Contents
Operation Instructions for Velleman PC Oscilloscope PCSU1000
- DIGITAL STORAGE PC OSCILLOSCOPE
- FFT SPECTRUM ANALYZER
Safety Instructions
1GS/s sampling mode
Controls
Connection
Waveform Parameters Display
Troubleshooting
Adding comment text in the signal screen
Menu Options
Spectral Density marker
Data acquisition to other applications
Data acquisition to Microsoft Excel
1.1 Safety Instructions
SAFETY and WARNINGS
Before making measurements and for safety reasons, it is important to know
some information about the measured unit.
Safe devices are:
- Battery operated equipment
- Equipment supplied via a transformer or adapter.
Do not measure:
- Equipment directly connected to mains.
- Equipment that contains components that are directly connected to mains (dimmers…).
- If necessary to measure above mentioned equipment, use an isolation transformer.
Please remember that the grounds of both channels are interconnected and connected to the PC ground !
1.2 1GS/s oversampling mode
- The 1GS/s sampling rate is in use on blue 0.2us/div, 0.1us/div, 0.05us/div and 0.02us/div ranges.
- Triggering must be ON to get stable waveform images.
- This mode is useable only for repetitive signals.
- This operation mode is called “Random Interleaved Sampling” (RIS) method, sometimes also called “Equivalwent Time Sampling” (ET) mode or “Random Repetitive Sampling”. In this sampling mode the oscilloscope uses successive triggering occurrences to gather the data to construct a picture of a repetitive signal.
1.3 Controls
Spectrum analyzer mode
FREQ. RANGE
Sets the frequency range of the display. It is necessary to slide the screen
using X-POSITION in order to see the full range.
LOG/LIN
To display the frequency on a linear or logarithmic scale.
ZOOM x1, x2, x4, x8
In order to expand the screen X1, X2, X4 or X8
Oscilloscope mode
VOLTS/DIV
Selected value indicates the peak-to-peak voltage required to produce a peak-
to-peak deflection of one major division on the screen
Big Screen
Displays large waveform screen with separate button bar. Use Normal Screen
button to return normal mode.
Note: Big screen is available only in Oscilloscope and Spectrum Analyzer
modes.
Coupling
AC: the input signal is capacitive coupled to the input amplifier/attenuator.
Only the AC components are measured.
GND: the input signal is broken and the input amplifier/attenuator is
connected to earth. Use this position for selecting a reference point on the
display.
DC: the input signal is directly connected to the input amplifier/attenuator.
Both AC and DC voltage are measured.
Probe x1/x10
Use these buttons to adapt the readouts according to the x1/x10-probe setting.
CH1 On, CH2 On
Buttons turn the display of the trace ON or OFF. To get the cursor
measurements of CH2 voltage values switch CH1 off.
Autoset
Automatic setup for the Volts/div, Time/div, and Trigger level to produce a
stable waveform of usable size. The trigger will be set on if the wave
aplitude on the screen is more then 0.5 divisions.
The signal should be repetitive for proper autoset function: Amplitude 5mV to
100V; frequency above 50Hz; duty cycle greater than 10%.
Persist
When this button is down the scope captures many acquisitions of a signal to
the screen. Record points accumulate until you release the button.
When this button is down the scope captures many acquisitions of a signal to
the screen. Record points accumulate until you release the button.
Using Persistence option you can easily analyze worst-case signal variations,
such as jitter or noise.
The Persist option can also be used to locate errors in digital signals. Using
this option you can capture erroneous events even if they only occur once.
Persist option makes it easy to compare known and unknown circuits. Click
“Single” button to capture multiple waveforms on the same screen.
TIME/DIV
Selects the time setting for the beam to sweep one major division on the
screen By selecting different TIME/DIV settings it is possible to zoom the
frozen waveform on the screen.
TRIGGER On/Off
Selects free run mode or trigged mode.
TRIGGER Level
Selects the signal level at which the sweep is triggered.
Triggering reference mark is displayed with the horizontal line on the left
side of the screen.
TRIGGER Channel
Selects the trigger source signal (Ch1, Ch2 or Ext)
TRIGGER Edge
Selects the triggering slope:
Arrow up: Trigger occurs when triggering signal crosses trigger level in a
positive going direction.
Arrow down: Trigger occurs when triggering signal crosses trigger level in a
negative going direction.
> |<
Resets the triggering x-position reference point. Triggering reference mark is
displayed with the vertical line on the bottom of the screen.
RUN
Selects recurrent display update mode (RUN). Pressing the button again freezes
the display.
SINGLE
When button is depressed and the trigger level is reached, refreshment of the
display takes place only once.
X-POSITION SCROLLBAR (Below the waveform display)
Positions the trace horizontally on the screen. Triggering reference point is
displayed with the vertical line on the bottom of the screen.
S/L
Button selects linear (L) or sine(x)/x (S) interpolation. Linear interpolation
connects the data points with straight lines. Sine(x)/x interpolation uses
curves to connect the data points. This smoothed interpolation gives better
waveform display at highest sine frequencies. Linear interpolation is better
for step signals. The S/L selection effects only at TIME/DIV settings 0.2 and
0.1us.
Note: Sine interpolation can result in incorrect top-top signal displaying, at
frequencies above 5MHz.
1GS/s
This 1GS/s sampling rate is in use on 0.2us/div, 0.1us/div, 0.05us/div and
0.02us/div ranges.
CH1 + CH2
CH1 – CH2
XY Plot
INV. CH2
This button appears only in math mode. Toggles between math mode and normal
mode.
Tip: Use Wheel Mouse’s scroll wheel to fine adjust the triggering level and
traces’ y-position.
1.4 Connection
Connect the oscilloscope unit to the USB port.
For Circuit Analyzer (Bode Plotter) option connect function generator PCG10 or
K8016 to LPT1, LPT2 or LPT3.
To select the LPT port address for the function generator click Hardware Setup
on Options menu or select it from Pc-Lab2000 startup screen.
1.5 Troubleshooting
Spectrum analyzer doesn’t work.
-
No arithmetic coprocessor in the computer.
No signal on the oscilloscope display -
No communication with the computer (check that the cable is connected to the USB port)
-
If USB cable is connected, close the program. Disconnect and reconnect the USB cable and run the Pc-Lab2000 program again.
-
RUN button is not ON.
-
The channel concerned is OFF.
-
TIME/DIV switch is in the wrong setting.
-
TRIGGER is ON, set TRIGGER OFF
-
The unit input selection is at GND.
-
Y position is wrongly adjusted.
-
Input amplitude is too large, adjust VOLTS/DIV setting.
If the above tips have no effect, then test on a different computer or different USB port.
Note. Close the program before disconnecting the USB cable.
1.6 Waveform Parameters Display
Waveform Parameters Display
When the menu option “Waveform Parameters” of the View menu is selected the
software automatically calculates various voltage and time parameters of a
signal, such as DC mean, amplitude, rise time etc.
These parameters are displayed in a separate window. Use the check boxes in
the window to select the parameters you which to be displayed.
You may select the parameters to display for frozen waveform as well.
You may even re-open saved waveform data file to perform these measurements.
Please Note: Do not change the scope settings when the re-opened waveform
parameters are to be read.
The green labeled parameters (High, Low, Amplitude, Rise time and Fall time)
are mainly intended for pulse shaped waveform measurements only.
For proper waveform measurement the signal levels must be appropriate. Signal
levels too small will result in noisy and inaccurate measurements. Signal
levels too large will result in clipping and yield incorrect results.
Error indication:
?? Indicates that clipping has occurred.
??? Indicates that there are too few or too many wave cycles in the waveform
data or the signal amplitude is too low. Also too noisy signal and variable
frequency signal causes this indication.
Voltage parameters
DC Mean
The arithmetic mean of the entire waveform data.
Max
The signal’s positive peak voltage.
(Difference between zero and highest value.)
Min
The signal’s negative peak voltage.
(Difference between zero and lowest value.)
Peak-to-Peak
The signal’s peak-to-peak voltage.
(Difference between highest and lowest value.)
High
The statistical maximum level recorded for all the cycles in the signal.
Low
The statistical minimum level recorded for all the cycles in the signal.
Amplitude
The voltage difference between the High and Low of the signal.
AC RMS
The true RMS value of the AC component of the signal is calculated and
converted to voltage.
AC dBV
The measured signal (AC only) is converted to dBV (0dB= 1V).
AC dBm
The measured signal (AC only) is converted to dBm (0dB= 0.775V).
AC+DC RMS
The true RMS value of the wave (AC+DC) is calculated and converted to voltage.
AC+DC dBV
The measured signal (AC+DC) is converted to dBV (0dB= 1V).
AC+DC dBm
The measured signal (AC+DC) is converted to dBm (0dB= 0.775V).
Time parameters
Duty Cycle
The ratio (expressed as a percentage) of the average positive pulse width to
the average period of the signal. The time intervals are determined at middle
reference level.
Duty Cycle = (Positive Pulse Width)/Period x 100%
Positive Width
Average positive pulse width in the waveform.
The time intervals are determined at middle reference level.
The middle reference is the middle point between the high and low levels.
Negative Width
Average negative pulse width in the waveform.
The time intervals are determined at middle reference level.
Rise Time
Time for a signal’s rising edge to go from the low reference level to the high
reference level.
The low reference level is 10% and high reference level is 90% of the pulse
amplitude.
Fall Time
Time for a signal’s falling edge to go from the high reference level to the
low reference level.
The low reference level is 10% and high reference level is 90% of the pulse
amplitude.
Period
The time interval between two consecutive crossings on the same slope of the
signal at the middle reference level.
Frequency
The inverse of the Period of the signal.
Phase
Phase angle in degrees between CH1 and CH2.
For phase measurement the frequency of CH1 must be equal to the frequency of
CH2.
Phase measurement is time consuming process. In slow PCs this reduces the
display update rate.
1.7 Add comment text in the signal screen
For explanation and documentation, each measurement can be supplied with a
comment text.
This text will be saved together with the waveform data to the disk file.
To enter the text:
- Right mouse click into the screen.
- Text box will open, to write your comment.
- Click Add Text on Screen or Remove to remove previously inserted text.
- Right click on the screen to position your text.
- Click Close.
To make the text transparent with the background, check Transparent text.
The text will have the same color as the vertical time/frequency markers.
Menu Options
File Menu
Edit Menu
Options Menu
View Menu
Math Menu
Help Menu
2.1 File Menu
Note: Default subdirectory (folder) \DATA for image and data files is created
when the program is run the first time.
Open Image
Opens an image file and displays it on the screen.
Open DSO Data
Opens and displays the waveform data saved in text format using the Save DSO
Data option.
Save Image
Saves the image to a file in Windows Bitmap (*.BMP) format.
Save DSO Data
Saves the waveform data in text format. All captured data (4096
points/channel) is saved.
Save FFT Data
Saves the FFT data in text format. Only the portion of the data displayed on
the screen is saved (250 points).
Save Settings
Saves the Oscilloscope, Spectrum Analyzer and Transient Recorder settings to a
file. Also
Function Generator settings (frequency, amplitude, offset and duty cycle) are
stored to the file.
Recall Settings
Loads a previously-stored settings file to the oscilloscope.
Print
Prints the image in color.
You can edit the image caption.
Print Setup
Selects a printer and sets printer options before printing. The available
options depend on the printer you select.
Exit
Terminates the program.
Calibrate & Exit
Makes the oscilloscope calibration, saves the calibration values to the
WinDSO.INI file and terminates the program. This option should be used when
the new oscilloscope has been on for about 1h.
This option performs following operations:
- The fine adjustment of the trace Y-position (offset) on different Volt/Div and Time/Div scales.
- Sets the trace labels (on the left side of the screen) to correspond the trace GND level.
- Sets the triggering level mark to correspond the triggering level.
2.2 Edit Menu
Copy
Copies the image to the Windows’ clipboard.
Paste
Pastes the image residing in Windows’ clipboard to the screen.
2.3 Options Menu
FFT Window
Set the window function used to taper the original signal before calculating
the FFT..
Spectrum analyzer supports six different FFT windows
- Rectangular
- Bartlett
- Hamming
- Hanning
- Blackman
- Flat top
The Hamming window is set as default at the startup
Background information
It is common practice to taper the original signal before calculating the FFT
(Fast Fourier Transformation). This reduces any discontinuities at the edges
of the signal. This is done by multiplying the signal with a suitable window
function. By choosing a tapered window it is possible to achieve a good
compromise between main lobe width and sidelobe levels of a spectral line. The
undesirable spectral leakage can be reduced using a tapered window, at the
expense of some broadening around individual spectral lines. Many different
windows have been designed for this purpose.
The choice of a suitable one depends on the nature of the signal or data, and
on the type of information to be extracted from its spectrum. In general, a
good FFT window has a narrow main spectral lobe to prevent local spreading of
the spectrum, and low sidelobe levels to reduce ‘distant’ spectral leakage.
In some cases it may be best to leave the data alone — in effect, to use a
rectangular window. For example, if a signal has close spaced components of
roughly the same amplitude, the rectangular window will probably offer the
best chance of resolving them. Conversely, if the amplitudes are very
different, a window with low sidelobes will reduce leakage around the large
component, and should make the small one easier to detect.
Comparison of the window functions
Window type | Window characteristics | Applications |
---|---|---|
Rectangular | Narrow mainlobe Slow rolloff rate | Broadband random (white noise) |
Closely spaced sine-wave signals
Bartlett| Narrow mainlobe Fast rolloff rate|
Hamming| Good spectral resolution Narrow mainlobe| Closely spaced sine-wave
signals
Nanning| High maximum sidelobe level Good frequency resolution Reduced leakage
Fast rolloff rate| Narrowband random signals Sine wave or combination of sine-
wave signals
Blackman| Wide mainlobe
Fast rolloff rate|
Flat top| Good amplitude accuracy Wide mainlobe
Poor frequency resolution More spectral leakage| Sine wave with need for
amplitude accuracy
FFT Options
Maximum
Maximum value of each frequency is displayed in Run mode.
This option can be used for recording signal levels as a function of frequency
(Bode plot). You can use spreadsheet to display the frequency response curve
including the frequency labels. On File menu, click Save FFT Data to export
the data to the spreadsheet.
RMS Average
Use this averaging mode to reduce signal fluctuations.
RMS averaging provides an excellent estimate of the true signal and noise
levels of the input signal.
Vector Average
Use this averaging mode to reduce random or uncorrelated noise in the
synchronous signal you want to display.
Vector averaging requires a trigger – set Trigger ON.
The signal of interest must be both periodic and phase synchronous with the
trigger.
Vector averaging reduces the noise floor for random signals since they are not
phase coherent from time record to time record.
If not trigged, the signal will not add in phase and instead will cancel
randomly.
Hardware setup
Select the LPT port where the hardware is connected.
Operation mode
- Oscilloscope connected to USB port.
- Demo mode. (No hardware needed).
Select the LPT port address where the Function Generator PCG10 or K8016 is
connected 378, 278 or 3BC
You’ll find the address from Windows’ Device Manager:
- Click “System” icon in Control Panel and then the Device Manager tab.
- Click the plus sign next to the “Ports”.
- Double-click “Printer Port (LPTx)”.
- Click the Resources tab to see the Input/Output address.
Select the LPT port communication speed for the Function Generator
Normal
This can be used in most cases.
Slow
Select this option if the waveform of the function generator is corrupte
Colors
Select the display colors.
Select the color for various items on the waveform display.
To change the color of an item, click the corresponding button. This will open
a dialog in which you can select the new color.
Full color selection is possible only if True Color (24 bit) palette is used.
There are restrictions in the color combinations with other palettes.
Click Bright Screen or Black Screen button to resets all colors to the Default
settings.
Trigger Options
Noise Reject
Select this option to get stable triggering on noisy signals.
This option works only in the Run mode and only in the Real-time Sampling
mode.
2.4 View Menu
RMS value
Displays AC RMS value of the signal
When this option is selected the true RMS AC value of the signal is displayed
on the screen.
- If CH1 is on the RMS value of CH1 is shown
- If CH1 is off the RMS value of CH2 is shown dBm Value
Displays AC dBm value of the signal
Sample Rate
Displays the sampling rate on the top of the screen.
When this option is selected the dBm value of the signal’s AC component is
displayed on the screen.
- If CH1 is on the dBm value of CH1 is shown
- If CH1 is off the dBm value of CH2 is shown
The dBm value displayed on the screen: 0 dBm = 1 milliwatt at 600 ohms ( 0.775 Vrms)
Waveform Parameters
Software automatically calculates various voltage and time parameters of a
signal, such as DC mean, amplitude, rise time etc.
These parameters are displayed in a separate window. Use the check boxes in
the window to select the parameters you which to be displayed.
Markers
Displays Markers on the screen.
Bright Grid
Brightens the signal grid on the screen.
Dot Join
On: The dots of the acquired waveform data are connected by lines.
Off: Only the dots of the acquired waveform data are displayed.
Markers in ocilloscope mode
- Two horizontal markers for measuring voltage. The voltage difference and the absolute voltage values (in parentheses) are displayed.
- Two vertical markers for measuring time and frequency
Note: The voltage markers give preference to channel CH1 if both channels are being used.
Markers in spectrum analyzer mode
- Marker function for absolute and relative voltage measurement is provided.
- The absolute voltage level in dBV or the voltage difference in decibels (dB) can be measured.
- Noise level can be measured using the Spectral Density marker.
- One vertical marker is provided for the frequency measurement.
Moving the markers
- Place the mouse pointer over a dashed marker line.
- Press and hold the left mouse button. The markerline turns solid.
- Drag the marker to the appropriate position
dB
The term dB or decibel is a relative unit of measurement used to describe
power or voltage difference.
Equation to calculate a dB value based on the ratio of two voltages V2 and V1
is:
dBV
dBV = The dB value is obtained with respect to 1Volt. The dBV is an absolute
unit of voltage. It expresses voltages as a ratio relative to 1 volt.
Equation to calculate voltage V from dBV value is:
Equation to calculate voltage V from dBV value is:
dBm
A unit of measurement of signal level in an electrical circuit, expressed in
decibels referenced to 1 milliwatt.
In a circuit with an impedance of 600 ohms, 0 dBm gives an equivalent voltage
level of 0.775 Vrms.
The dBm value displayed on the screen: 0 dBm = 1 milliwatt at 600 ohms ( 0.775
Vrms)
2.5 Spectral Density
The Spectral Density marker may be used when measuring the density of random
or noise signals since it properly takes into account the frequency bin width
and the FFT window function used by the spectrum analyzer when measuring
noise-like signals.
The Spectral Density marker readout is automatically normalized to 1 Hz.
The displayed unit is:
Note: The Spectral Density marker should not be used to measure discrete
frequency components as it will provide misleading level readings.
The Spectral Density is simply the magnitude of the spectrum normalized to a 1
Hz bandwidth. This measurement approximates what the spectrum would look like
if each frequency component were really a 1 Hz wide piece of the spectrum at
each frequency bin.
When measuring broadband signals such as noise with a spectrum analyzer, the
amplitude of the spectrum changes with the frequency span. This is because the
FFT bin width changes and the frequency bins have a different noise
bandwidth.
The Spectral Density marker normalizes all measurements to a 1 Hz bandwidth
and the noise spectrum becomes independent of the frequency span. This allows
measurements with different spans to be compared.
If the noise is Gaussian in nature, then the amount of noise amplitude in
other bandwidths may be approximated by scaling the Spectral Density
measurement by the square root of the noise bandwidth.
Example:
Spectrum analyzer can be used to measure the spectral density of this noise
signal.
In the spectrum analyzer select the following menu options:
- Options / FFT Options / RMS Average
- View / Markers (FFT) f & Spectral Density dBV/sqrt(Hz)
This image shows the spectrum of the band limited noise signal.
The analysis of noise in the frequency domain shows the distribution of the
noise amplitude as a function of frequency.
Using the Spectral Density marker and the Frequency marker, the voltage
spectral density (VSdBV) and the noise bandwidth (BN) can be read from the
spectrum analyzer display.
First convert the voltage spectral density to V/Hz
This can be accomplished using the following calculation:
This is the magnitude of the spectrum normalized to a 1 Hz bandwidth.
You may calculate the noise voltage over any bandwidth by multiplying this
value by the square root of the bandwidth.
Assuming a 6 kHz bandwidth, the total output noise voltage is:
(See the Vrms value of the oscilloscope waveform image of this noise signal.)
2.6 Math Menu
The result of mathematical operation of channel 1 and 2 is displayed.
One of the following functions can be selected:
- Ch1 + Ch2
- Ch1 – Ch2
- XY Plot
- Invert Ch2
XY Plot:
Ch1 data is displayed on Y axis
Ch2 data is displayed on X axis
A button is provided to toggle between Math mode and Normal mode.
2.7 Help Menu
Contents
Displays this help file
Installing Windows NT4 driver
Gives instructions for the Windows NT4, Windows 2000 and Windows XP users
About
Displays information of the program version
Data Transfer
Data acquisition to other applications
Data acquisition to Microsoft Excel
3.1 Data acquisition to other applications
Pc-Lab2000 software package includes a DLL (Dynamic Link Library) DSOLink.DLL,
installed to Windows’ SYSTEM32 folder.
This DLL allows you to write custom applications in Excel, Visual Basic,
Delphi or any other 32-bit Windows application development tool that supports
calls to a DLL.
The DLL gives you direct access to real-time data and settings information
from the oscilloscope.
The complete example programs are located on the VELSOFT CD. Those may be used
as a starting point how to construct your customized application programs.
Note: Before running the following example programs: The oscilloscope software
must be running and Run or Single button pressed and trace displayed on the
oscilloscope screen.
Description of the procedures of the DSOLink.DLL
ReadCh1
ReadCh2
Syntax
PROCEDURE ReadCh1(Buffer: Pointer);
PROCEDURE ReadCh2(Buffer: Pointer);
Parameter
Buffer: A pointer to the data array of 5000 long integers where the data
will be read.
Description
Read all the data and the settings of channel 1 or channel 2 of the PCSU1000.
As a return the following data is put to the buffer:
[0] : Sample rate in Hz
[1] : Full scale voltage in mV
[2] : Ground level in A/D converter counts. The value may be beyond the 0…255
range if GND level
is adjusted beyond the waveform display area.
[3…4098] : The acquired data in A/D converter counts (0…255), from PCSU1000.
The triggering point of the PCSU1000 is at the data location [1027].
Running the DSOLink in Delphi
Check the \PC-Lab2000 tools\PCSU1000 – PCS500 – PCS100 – K8031\Data transfer
DSOLink_DLL\DSOLink_Demo_VB\ folder on the Velleman CD to locate the demo
files.
This folder contains a ready to run DSOLink_Demo.EXE program and its source
code.
You may copy the files to any folder and use Delphi to examine, edit and
compile the files.
Example (in Delphi)
var
data: array[0..5000] of longint;
procedure ReadCh1(Buffer: Pointer); stdcall; external ‘DSOLink.dll’;
procedure TForm1.Button1Click(Sender: TObject);
var i: longint;
p:pointer;
begin
p:= @data[0];
ReadCh1(p);
memo1.clear;
memo1.lines.add(‘Sample rate [Hz]’+chr(9)+inttostr(data[0]));
memo1.lines.add(‘Full scale [mV]’+chr(9)+inttostr(data[1]));
memo1.lines.add(‘GND level [counts]’+chr(9)+inttostr(data[2]));
memo1.lines.add(”);
begin
for i:=0 to 20 do
memo1.lines.add(‘Data (‘+inttostr(i)+’)’+chr(9)+chr(9)+inttostr(data[i+3]));
end; end;
Running the DSOLink in Visual Basic
Make sure that the file DSOLink.DLL is copied to Windows’ SYSTEM32 folder.
Check the \PC-Lab2000 tools\PCSU1000 – PCS500 – PCS100 – K8031\Data transfer
DSOLink_DLL\DSOLink_Demo_VB\ folder on the VELSOFT CD to locate the demo
files.
This folder contains a ready to run DSOLink_Demo.EXE program and its source
code.
You may copy the files to any folder and use Visual Basic to examine, edit and
compile the files.
Example (in Visual Basic)
Option Explicit
Dim DataBuffer(0 To 5000) As Long
Private Declare Sub ReadCh1 Lib “DSOLink.dll ” (Buffer As Long)
‘This reads the settingsd and 4096 bytes of data from CH1 to the data buffer.
‘The first 21 values are displayed.
Private Sub Read_CH1_Click(Index As Integer)
Dim i As Long
List1.Clear
ReadCh1 DataBuffer(0)
List1.AddItem “Sample rate [Hz]” + Chr(9) + Str(DataBuffer(0))
List1.AddItem “Full scale [mV]” + Chr(9) + Str(DataBuffer(1))
List1.AddItem “GND level [counts]” + Chr(9) + Str(DataBuffer(2))
List1.AddItem “”
For i = 0 To 20
List1.AddItem “Data(” + Str(i) + “)” + Chr(9) + Chr(9) + Str(DataBuffer(i +
3))
Next
End Sub
Running the DSOLink in Borland C++ Builder
The following files are available in the \PC-Lab2000 tools\PCSU1000 – PCS500 –
PCS100 K8031\Data transfer DSOLink_DLL\DSOLink_Demo_BCB\ folder on the VELSOFT
CD for development with Borland C++Builder:
DSOLink.dll the dynamically linked library
DSOLink.h the C/C++ header file for function prototypes
DSOLink.lib the import library
DSOLink_demo.cpp demo source
- Create a new project in Borland C++ Builder.
- Add the import library to your project using Project | Add to Project menu option.
- Add a #include statement in the main unit that includes DSOLink.H.
- Finally, add code that calls the DLL functions.
DSOLink.h
//————————————————————————–//
DSOLink.h #ifdef
__cplusplus
extern
“C”
{
/ Assume C declarations for C++ /
endif
define FUNCTION __declspec(dllimport)
FUNCTION stdcall ReadCh1(int* ptr);
FUNCTION stdcall ReadCh2(int* ptr);
ifdef __cplusplus}
endif
//—————————————————————————
Example (in Borland C++Builder)
//————————————————————————–//
DSOLink_demo.cpp
include <vcl.h>
pragma hdrstop
include “DSOLink.h”
include “DSOLink_demo.h”
//————————————————————————–#pragma
package(smart_init)
pragma resource
“.dfm”
TForm1
Form1;
//————————————————————————–fastcall TForm1::TForm1(TComponent* Owner)
: TForm(Owner)
{} //————————————————————————–void
fastcall TForm1::Button1Click(TObject *Sender)
{
int data[5000];
ReadCh1(data);
Memo1->Clear();
Memo1->Lines->Add(“Sample rate [Hz]: “+IntToStr(data[0]));
Memo1->Lines->Add(“Full scale [mV]: “+IntToStr(data[1]));
Memo1->Lines->Add(“GND level [counts]: “+IntToStr(data[2]));
Memo1->Lines->Add(“”);
for (int i = 0; i < 20; i++)
{ Memo1->Lines->Add(“Data “+IntToStr(i)+char(9)+IntToStr(data[i+3]));
}
}
//—————————————————————————
Note: If the import library is not compatible to your Borland C++ version, you
can create an import library by running IMPLIB on the DLL.
IMPLIB works like this:
IMPLIB (destination lib name) (source dll)
For example,
IMPLIB DSOLink.lib DSOLink.dll
3.2 Data acquisition to Microsoft Excel
Pc-Lab2000 software package includes a DLL (Dynamic Link Library) DSOLink.DLL,
installed to Windows’ SYSTEM32 folder.
This DLL allows you to write custom applications in Excel, Visual Basic,
Delphi or any other 32-bit
Windows application development tool that supports calls to a DLL.
Transferring Waveform Data to an Excel Spreadsheet
The example Excel macro shows you how to collect data directly into the
spreadsheet from the Velleman PC oscilloscopes without the need for other
software.
-
Open Microsoft Excel and start a new workbook document.
-
Select the View / Toolbars Menu and Select Forms.
• The Forms toolbar appears. -
Create a Button
• In the Forms toolbar click on the “Button” button: the mouse-pointer will become a small cross.
• On Excel worksheet, use the mouse to draw a rectangle to mark where you want your button to appear.
• When you release the mouse after drawing the rectangle the “Assign Macro” dialog-box will appear. -
Type Macro name: ReadAll and click New button.
• Microsoft Visual Basic edit window will open. A subroutine called ReadAll has been created. -
Substitute the default text: Sub ReadAll()
End Sub with the following text in the edit window:
(Use Copy and Paste.)
Option Explicit
Dim DataBuffer1(0 To 5000) As Long
Dim DataBuffer2(0 To 5000) As Long
Private Declare Sub ReadCh1 Lib “DSOLink.dll ” (Buffer As Long)
Private Declare Sub ReadCh2 Lib “DSOLink.dll ” (Buffer As Long)
Sub ReadAll()
Dim i As Long
ReadCh1 DataBuffer1(0)
ReadCh2 DataBuffer2(0)
With ActiveSheet
For i = 0 To 99
.Cells(i + 1, 2) = DataBuffer1(i)
.Cells(i + 1, 3) = DataBuffer2(i)
Next i
End With
End Sub
-
Press Alt+F11 to return to Excel.
-
Type following texts to column A:
Sample rate [Hz] Full scale [mV] GND level [counts] Data 0
Data 1
Data 2
… -
Start oscilloscope program for PCSU1000, PCS500, PCS100 or K8031 and click Run or Single button.
-
Click the button on the Excel worksheet. The created macro will execute and the data described in column A will appear to the worksheet columns B and C.
• Rows 4…4099 contain the acquired data in A/D converter counts (0…255) for PCSU1000 and PCS500.
• Rows 4…4083 contain the acquired data in A/D converter counts (0…255) for PCS100 and K8031.
• The triggering point of PCSU1000 and PCS500 is on row 1030 and of PCS100 and K8031 on row 4.
The three first rows contain the scope settings and the rest of the rows
contain the raw oscilloscope data in A/D converter counts (0…255).
Using the Sample rate, Full scale and GND level values it is possible to
reconstruct the waveform data into engineering units (e.g. Volts and seconds)
for further analysis.
© 2005 … Velleman
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>