ANRITSU CORPORATION G0430A PCIe5 Re Driver Set User Manual

June 13, 2024
ANRITSU CORPORATION

ANRITSU CORPORATION G0430A PCIe5 Re Driver Set

ANRITSU CORPORATION G0430A PCIe5 Re Driver Set

Safety Symbols

To prevent the risk of personal injury or loss related to equipment malfunction, Anritsu Corporation uses the following safety symbols to indicate safety-related information. Ensure that you clearly understand the meanings of the symbols BEFORE using the equipment. Some or all of the following symbols may be used on all Anritsu equipment. In addition, there may be other labels attached to products that are not shown in the diagrams in this manual.

Symbols used in manual

DANGER  This indicates a very dangerous procedure that could result in serious injury or death if not performed properly.
WARNING This indicates a hazardous procedure that could result in serious injury or death if not performed properly
CAUTION This indicates a hazardous procedure or danger that could result in light-to-severe injury, or loss related to equipment malfunction, if proper precautions are not taken.

Safety Symbols Used on Equipment and in Manual

The following safety symbols are used inside or on the equipment near operation locations to provide information about safety items and operation precautions. Ensure that you clearly understand the meanings of the symbols and take the necessary precautions BEFORE using the equipment.

This indicates a prohibited operation. The prohibited operation is indicated symbolically in or near the barred circle.
This indicates an obligatory safety precaution. The obligatory operation is indicated symbolically in or near the circle.
This indicates a warning or caution. The contents are indicated symbolically in or near the triangle.
This indicates a note. The contents are described in the box.
SymbolsThese indicate that the marked part should be recycled.

G0430A
PCIe5 Re-Driver Set
Operation Manual
15 April 2022 (First Edition)
28 November 2022 (Seventh Edition)

Copyright © 2022, ANRITSU CORPORATION.
All rights reserved. No part of this manual may be reproduced without the prior written permission of the publisher.
The operational instructions of this manual may be changed without prior notice. Printed in Japan

For Safety

CAUTION

Static Sensitive

Always take the following anti-static measures to prevent the internal circuit from being damaged when using this equipment.

  • Wear a wrist strap connected to the ground terminal of this equipment.
  • Connect the ground wires of this equipment, external measuring instruments and DUT before connecting a coaxial cable.
  • Eliminate static electricity charged between the cores and outer conductors of an external device and the coaxial cable, before connecting this equipment and the external device

Use in a Residential Environment

This equipment is designed for an industrial environment.

In a residential environment, this equipment may cause radio interference in which case the user may be required to take adequate measures.

Use in Corrosive Atmospheres

Exposure to corrosive gases such as hydrogen sulfide, sulfurous acid, and hydrogen chloride will cause faults and failures.

Note that some organic solvents release corrosive gases.

Anritsu Warranty

Anritsu Corporation provides the following warranty against stoppages arising due to manufacturing error, and against problems with operation occurring even though the procedures outlined in the operation manual were followed.

The equipment is warranted for one year from delivery and will be replaced free of charge only once under warranty.

After the expiration of one-year warranty period, replacement will not be made on the failed equipment. Please consider purchasing a new one.

The hardware and software warranties are not valid under any of the following conditions:

  • The fault is due to mishandling, misuse, or unauthorized modification or repair of the equipment by the customer.
  • The fault is due to severe usage clearly exceeding normal usage.
  • The fault is due to improper or insufficient maintenance by the customer.
  • The fault is due to natural disaster, including fire, wind or flood, earthquake, lightning strike, or volcanic ash, etc.
  • The fault is due to damage caused by acts of destruction, including civil disturbance, riot, or war, etc.
  • The fault is due to explosion, accident, or breakdown of any other machinery, facility, or plant, etc.
  • The fault is due to use of non-specified peripheral or applied equipment or parts, or consumables, etc.
  • The fault is due to use of a non-specified power supply or in a nonspecified installation location.
  • The fault is due to use in unusual environments(Note).
  • The fault is due to activities or ingress of living organisms, such as insects, spiders, fungus, pollen, or seeds.

In addition, this warranty is valid only for the original equipment purchaser. It is not transferable if the equipment is resold.

Anritsu Corporation shall assume no liability for damage or financial loss of the customer due to the use of or a failure to use this equipment, unless the damage or loss is caused due to Anritsu Corporation’s intentional or gross negligence.

Note:

For the purpose of this Warranty, “unusual environments” means use:

  • In places of direct sunlight
  • In dusty places
  • In liquids, such as water, oil, or organic solvents, and medical fluids, or places where these liquids may adhere
  • In salty air or in place chemically active gases (sulfur dioxide, hydrogen sulfide, chlorine, ammonia, nitrogen dioxide, or hydrogen chloride etc.) are present
  • In places where high-intensity static electric charges or electromagnetic fields are present
  • In places where abnormal power voltages (high or low) or instantaneous power failures occur
  • In places where condensation occurs
  • In the presence of lubricating oil mists
  • In places at an altitude of more than 2,000 m
  • In the presence of frequent vibration or mechanical shock, such as in  cars, ships, or airplanes

Anritsu Corporation Contact

In the event of this equipment malfunctions, please contact a sales office listed in Anritsu Service and Sales office. Contact information can be found on the last page of the printed version of this manual, and is available in a separate file on the PDF version.

Notes On Export Management

  • We assume no responsibility if this product is re-exported and used without our approval since it may not comply with the safety standards, etc., of specific countries.
  • This product and its manuals may require an Export License/Approval by the Government of the product’s country of origin for re-export from your country.
    Before re-exporting the product or manuals, please contact us to confirm whether they are export-controlled items or not.
    For more details, visit https://www.anritsu.com/support/export-procedures.
    When you dispose of the products or manuals subject to export control, they must be physically destroyed or shredded so as not to be unlawfully used for military purpose.

About This Manual

This operation manual explains the configuration, connection, and operation of the G0430A PCIe5 Re-Driver Set.

Chapter 1 Configuration of PCIe5 Re-Driver Set

The G0430A PCIe5 Re-Driver Set (hereafter G0430A) is an equalizer for 32 Gbaud NRZ signal. With G0430A for channel loss compensation, the MP1900A Signal Quality Analyzer-R can perform tests according to the PCIe 5.0 standard.

CAUTION

  • When signals are input to this modules, avoid excessive voltage beyond the rating. Otherwise, the circuit may be damaged.
  • When output is used at the 50 Ω GND terminator, never feed any current or input signals to the output.
  • As a countermeasure against static electricity, ground other devices to be connected (including experimental circuits) with ground wires before connecting the I/O connector.
  • The outer conductor and core of the coaxial cable may become charged as a capacitor. Use any metal to discharge the outer conductor and core before use.
  • To protect this modules from electrostatic discharge failure, a conductive sheet should be placed onto the workbench, and the operator should wear an electrostatic discharge wrist strap. Always ground the wrist strap to the workbench antistatic mat or the frame ground of this modules.

The G0430A consists of a printed circuit board, cables, software, adapters, screws, and posts. Refer to the following table for the G0430A configuration.

Table 1-1 G0430A Configuration

No. Name Q ty Remar k s
1 MAEQ-39904B EVM board 1 Linear Equalizer EVM board
2 Mini USB cable 1 Cable
3 USB flash drive 1 The following files are stored.
  • GUI software

| | |

  • Operation manual

| | |

  • Preset files

| | |

  • Test report

4| 2.92mm Carlisle cable| 2| Cable
5| K220B| 2| Precision Adapter DC to 40 GHz, 50 n
6| DC cable| 1| A pair of black and red cords
7| GND Connection Cable| 1| Cable
8| ASB-316.5E| 7| Posts
9| 3NPS6Ni+SW+WBS| 7| Screws
10| +3.3V label| 1| For MAEQ-39904B  EVM board Red
| | | Terminal

Table 1-2 Required Components

Model Name Q ty Remarks
J1728A Electrical Length Specified Coaxial Cable (0.4m, K Connector) 2 Jl

728A or equivalent
NIA| DC Power Supply| 1| Single power supply*

(+3.3 V, 1.5 A)

Chapter 2 Connection Procedure

Installation

Support MAEQ-39904B EVM board with screws and posts.

Installation

Place the MAEQ-39904B EVM board on a horizontal and level surface. Stick the +3.3V label near the red terminal.

Installation

Setup

CAUTION

Never feed a voltage of more than +3.3 V for MAEQ-39904B EVM board. Doing so may cause damage to it.

  1. Turn on the MP1900A. Check that the DUT is turned OFF.

  2. Set the DC power supply as below.
    DC voltage: +3.3 V
    Current Limit: 1.5 A
    Output: OFF

  3. Connect the cables (refer to Figure 2.2-1).
    GND Connection Cable (7):
    Connect MP1900A and the MAEQ-39904B EVM board.
    2.92mm Carlisle cable (4):
    Attach to the MAEQ-39904B EVM board.
    Mini USB cable (2):
    Connect MP1900A or PC and the MAEQ-39904B EVM board.
    K220B (5):
    Connect MAEQ-39904B Ch1 output to Ch2 input with two K220Bs.
    Note: Unused channels (Ch0 and Ch3) do not need to be terminated.

  4. Connect the DC power supply to MAEQ-39904B EVM board with the DC cable (6).

  5. Connect the DUT, MAEQ-39904B EVM board, and the ED of MP1900A referring Figure 2.2-1 “Cable Connection”.

  6. Turn on the output of DC power supply. Typical current is 260 mA

Refer to the following figure for connecting each part.

Connection Procedure

CAUTION

To prevent a short circuit, be careful not to allow the 2.92mm Carlisle cables and their end connectors to contact the surface of the MAEQ-39904B EVM board or the DUT.

Connection Procedure

Chapter 3 Installing Software and Operation

Preparation

Prepare the G0430A for operation according to the following steps.

  1. Install the GUI software on the host PC or MP1900A.
    Note:
    The GUI Software works only on Windows 10.
    If the OS on the MP1900A is Windows 7, install the GUI
    Software on an external PC installed with Windows 10.
    The GUI software is stored on the USB flash drive.
    Folder name: \EVM Kit Zip File\EVM Setup vx.x.x
    (x.x.x represents the software version)
    Executable file: setup.exe

  2. Verify jumpers are present in the positions shown in Figure 3.1-1 and Table 3.1-1.
    Installing Software and Operation Board| Jumper| Setting| Description
    ---|---|---|---
    Main Board (Green)| JP1| ON| 3.3 V to AVCC
    Controller (Blue)| JP1, JP2, JP8, JP3, JP4| 1-2, ON| Micro-controller programming setup
    JP6| OFF| Do not use.

  3. Connect the device under the test (hereafter, DUT) outputs to the MAEQ-39904B Ch1 inputs with the Carlisle connector cables.

  4. Connect between the MAEQ-39904B Ch2 outputs and the MP1900A ED inputs by using 2.92mm Carlisle cables.

  5. Make sure the necessary hardware connections for the G0430A are in place, and launch the User Control software by going to Start > All Programs > MACOM Technology Solutions.

  6. Make sure the software is active, click Connect.
    The device will be recognized and displayed in the upper-right corner of the window.
    When the Bad COM Port Settings dialog box is displayed, open Device Manager, expand Ports (COM & LPT), and check the number displayed after USB Serial Device.Installing Software and Operation
    Figure 3.1-2 Port Number Indication of EVM board
    Change the Port setting for software from Auto to the USB Serial
    Device number (for example, COM5).
    Also, check that I2C Display is set to 7-Bit.

Installing Software and Operation

  1. On the right side of the Channel Control tab, click Reset with No OTP Download to reset the device.
    The Channel Control tab helps you control the settings for each of the four channels individually.
    The high frequency boost can be controlled by using the slide bar or by inputting a decimal number between 1 to 127 (1in Figure 3.1-4). Channel control also allows control over the settings like, low frequency boost, bias control, and disabling the channel (2 and 3 in Figure 3.1-4).
    On the individual channel tab, each of the four channels can be configured differently.Installing Software and Operation
How to Optimize Return Path

For Receiver LEQ Test with PCIe Gen5 (32GT/s), it is important to optimize Return Path (DUT Tx to BERT ED). The procedure is shown below.

Requirements for using the G0430A When testing the System, connect the G0430A between DUT Tx and ED. Refer to Chapter 2 “Connection Procedure” for how to set up the G0430A. Though the G0430A is not required for AIC Test, it is the same as the method of optimizing Return Path for System / AIC test.

CAUTION

The maximum recommended input amplitude for the IC mounted on the MAEQ-39940B EVM board is 1.0 Vp-p (differential).

The G0430A is used with Ch1 and Ch2 of the IC on the MAEQ-39904B EVM board connected in series. Therefore, if the input amplitude for Ch1 exceeds 800 mVp-p (differential), there is a risk that the input amplitude for Ch2 in the subsequent stage could exceed 1 Vp-p (differential).

By using an oscilloscope with a bandwidth of at least 50 GHz, make sure that the input waveform amplitude for the MAEQ-39904B EVM board is less than or equal to 800 mVp-p (differential).

Preparation of preset files
To optimize the Return Path, you are required to prepare preset files for MAEQ-39904B.

  1. Copy the zip file from the USB flash drive to the PC or MP1900A.
    \Macom_MAEQ-39904_EQ_Preset.zip

  2. Extract the zip file, and then make sure that the following 21 files
    are in the directory you specify.
    MAEQ-39904_EQ_P0.txt
    MAEQ-39904_EQ_P1.txt:
    MAEQ-39904_EQ_P19.txt
    MAEQ-39904_EQ_P20.txt

  3. Click the Channel Control tab, click Load File shown in Figure 3.2-1, and select a preset file.
    Installing Software and Operation

Optimization procedure

  1. Refer to the “Receiver Link Equalization Test” section in the MX183000A High-Speed Serial Data Test Software Operation Manual and set up the test configuration according to the figure below. Turn on the DUT.

  2. Start MX183000A, set Specification to 5.0(32.0 GT/s), and then turn off all the stresses below.
    ● SJ
    ● RJ
    ● DM
    ● CM
    Installing Software and Operation

  3. In the MX183000A main screen, select the Setting check box for LEQ Test, and follow the steps below:
    3a. Set PPG Starting Preset to P5.
    3b. Set DUT Initial Preset (Preset Hint Tx) to P9.
    3c. Set DUT Target Preset (Change Preset) to P9.
    3d. Click Apply.
    3e. Click Option, and then on the PPG/ED tab of the Option dialog box, set CTLE Gain to 0.0 dB.
    Installing Software and Operation

  4. Turn the G0430A on, load channel control parameters shown in Table A-1 from a preset file named MAEQ-39904_EQ_P0.txt first as shown in Figure 3.2-1.

  5. Click Link Start.
    Installing Software and Operation

  6. Establish Loopback.
    Installing Software and Operation 6a. If Linkup Speed is not 32.0 Gbps and the LTSSM State is not
    Loopback. Active. Lead, increase the preset file number of the redriver , for example at intervals of four, EQ_P0 -> EQ_P4  > … – > EQ_P20. Each time the preset file number set to the re-driver changes, make sure that the LTSSM is in the Loopback state displayed.
    6b. If Loopback cannot be established despite the preset file number is set to EQ_P20, set the Preset values in steps 3b and 3c to P8, set the preset file number of the re-driver to EQ_P0, and then retry to check the Loopback state according to step 6a.
    6c. If Loopback cannot be established even with Preset P8, redo step 6b with the Preset values set to P7, P6, … P0.
    6d. If Loopback cannot be established even with Preset P0 in step 6c, set CTLE Gain to OFF (step 3e.) and redo step 3 to 6c.
    Installing Software and Operation
    6e. If Loopback is established according to steps 6a to 6d, check the Preset values of MX183000A and the preset file number of the re-driver.
    6f. If Loopback cannot be established up to step 6e, remove ISI from the measuring system and retry steps 6a to 6e.
    If Loopback cannot be established by the steps above, the measurement environment may be incorrect (cable connection error, etc.) or there may be a problem with the DUT, so review the measurement environment again.

  7. BER optimization
    Perform Link Training with the Preset values, CTLE Gain setting,
    and the preset file number of the re-driver specified in step 6 to
    measure BER.
    7a. Optimize EQ to find the setting that gives the optimal BER.
    7b. Set the EQ confirmed in step 6.
    7c. Run Auto Search.
    Click Operate MP1900A to switch over to MP1900A GUI.
    Installing Software and Operation
    Click the Auto Search button.

Set as below and click Start:
Mode: Fine(NRZ)
Item: Threshold & Phase
CTLE Auto Adjust: OFF (if CTLE Gain setting is ON)
Installing Software and Operation
7d. Wait until Auto Search is completed, and then check BER.
7e. Increase the preset file number of the re-driver by 1 or 2 at a time, and then check the BER again. If the optimal preset BER numbers are detected in succession, repeat until you find the preset number at which the BER deteriorates. If an error-free condition is achieved with multiple preset file numbers, use the preset file with the middle number of a series of preset file numbers as the optimum setting.
7f. Click the Auto Search button in step 7c and check the BER again.
When an error-free condition is achieved with the steps so far, optimization of Return Path is complete successfully. Then, go to step 8. Otherwise, follow steps 7g to 7j.
7g. CTLE adjustment Adjust the CTLE Gain value or set CTLE Gain to OFF to minimize Total BER.
Installing Software and Operation
7i. Run Auto Search
Click Operate MP1900A to switch over to MP1900A GUI
Installing Software and Operation
Click the Auto Search button.

Set as below and click Start:
Mode: Fine(NRZ)
Item: Threshold & Phase
CTLE Auto Adjust: ON (if CTLE Gain setting is ON)
Installing Software and Operation
7j. Wait until Auto Search is completed, and then check BER.
With the steps so far, if an error is still detected, it is determined that there is an error in the Rx of the DUT.

  1. Confirm the setting parameters for optimization.
    These parameters for optimization are determined according to the procedure above (steps 1-7), as shown below:
    ● Preset setting confirmed in step 6.
    DUT Initial Preset (Preset Hint Tx)
    DUT Target Preset (Change Preset)
    Installing Software and Operation
    ● Threshold, Phase, and G0430A settings confirmed in step 7. CTLE setting is effective only when step 7i is followed.
    Installing Software and Operation
    Now, the Return Path is successfully optimized.
    Installing Software and Operation
    The following table shows recommend combinations of values for Figure 3.1-4 “Channel Control tab”.

Appendix A Recommended Channel Control Values

Table A-1 Recommended Channel Control Values

Set No.| ch1_eq_hf| CH1 EQ Bias Control| ch1_eq_lf (dB)| ch2_eq_hf| CH2 EQ Bias Control| ch2_eq_lf (dB)
---|---|---|---|---|---|---
0| 28| 2| 3.0| 7F| 3| 3.0
1| 20| 2| 3.0| 7F| 3| 3.0
2| 18| 2| 3.0| 7F| 3| 3.0
3| 10| 1| 3.0| 7F| 3| 3.0
4| 8| 1| 3.0| 7F| 3| 3.0
5| 1| 1| 3.0| 7F| 3| 3.0
6| 1| 1| 3.0| 78| 3| 3.0
7| 1| 1| 3.0| 70| 3| 3.0
8| 1| 1| 3.0| 68| 3| 3.0
9| 1| 1| 3.0| 60| 3| 3.0
10| 1| 1| 3.0| 58| 3| 3.0
11| 1| 1| 3.0| 50| 2| 3.0
12| 1| 1| 3.0| 48| 2| 3.0
13| 1| 1| 3.0| 40| 2| 3.0
14| 1| 1| 3.0| 38| 2| 3.0
15| 1| 1| 3.0| 30| 2| 3.0
16| 1| 1| 3.0| 28| 2| 3.0
17| 1| 1| 3.0| 20| 2| 3.0
18| 1| 1| 3.0| 18| 2| 3.0
19| 1| 1| 3.0| 10| 1| 3.0
20| 1| 1| 3.0| 1| 1| 3.0