Delta OHM LPUVA02 UVA Radiometer Instruction Manual
- June 13, 2024
- Delta OHM
Table of Contents
- Delta OHM LPUVA02 UVA Radiometer
- Product Information
- INTRODUCTION
- WORKING PRINCIPLE
- INSTALLATION
- ELECTRICAL CONNECTIONS
- MEASUREMENT
- MAINTENANCE
- TECHNICAL SPECIFICATIONS
- SAFETY INSTRUCTIONS
- ACCESSORIES ORDERING CODES
- DECLARATION OF CONFORMITY
- WARRANTY
- TECHNICAL INFORMATION
- DISPOSAL INFORMATION
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Delta OHM LPUVA02 UVA Radiometer
Product Information
Product Name | UVA Radiometer |
---|---|
Model | LPUVA02 |
Manufacturer | Delta OHM |
Website | www.deltaohm.com |
Working Principle
The LPUVA02 radiometer measures the global irradiance in the UVA spectral range on a flat surface (W/m2). It calculates the measured irradiance by summing up the direct solar irradiance and diffuse irradiance. The radiometer uses a solid state sensor with a special filter to achieve the desired spectral match. It has a 50 mm external diameter dome to protect the sensor from weather agents. The response of the radiometer follows the cosine law, ensuring accurate measurements even when the sun is at a low raising angle.
Installation
To install the LPUVA02 radiometer, follow these steps:
- Remove the perforated cap from the silica-gel cartridge.
- Fill the cartridge with the silica-gel crystals from the sealed sachet.
- Close the cartridge with the cap.
- Attach the cartridge to the radiometer body using a coin.
- Ensure the cartridge is tightly screwed to maintain the life of the silica gel.
- Position the shade disk and secure it with the screws.
- The radiometer is now ready for use.
INTRODUCTION
The LPUVA02 radiometer measures the global irradiance in the UVA spectral
range on a flat surface (W/m2). Measured irradiance is the result of the sum
of direct solar irradiance and of diffuse irradiance.
The radiometer can also be used for monitoring indoor UVA emissions.
Available in the following versions:
- LPUVA02: PASSIVE
- LPUVA02AC: ACTIVE with 4..20 mA CURRENT output (0…200 W/m2)
- LPUVA02AV: ACTIVE with 0..1 or 0..5 or 0..10 V VOLTAGE output to be de-fined when ordering (0…200 W/m2).
Each radiometer is individually calibrated at the factory and is distinguished
by its own calibration factor. The calibration is performed by comparison with
the primary stand-ard of the Delta OHM metrology laboratory, using the
emission line at 365 nm of a suitably filtered Xe-Hg lamp.
Note: there is no international standard for the calibration of UVA
radiometers; there-fore, when considering the value of the calibration
coefficient, the method by which it was obtained must be taken into account;
the same radiometer calibrated with differ-ent procedures can have different
calibration factors.
WORKING PRINCIPLE
LPUVA02 radiometer is based on a solid state sensor, the spectral match with
the de-sired curve is obtained using a special filter. The relative spectral
response is reported on figure 2.1. The radiometer is provided with a 50 mm external diameter
dome in order to supply a suitable protection of the sensor to the weather
agents.
The response in accordance with the cosine law has been obtained thanks to the
par-ticular shape of the diffuser and of the housing. The deviation between
the theoretical response and the measured one is shown in the Fig. 2.2
.The
excellent relation between the response of the radiometer and the cosine law
al-lows using the instrument also when the sun has a very low raising (the UVA
diffuse radiation increases as the sun moves away from the zenith, therefore
the error on the direct radiation, due to the imperfect response according to
the cosine law, becomes negligible on the measurement of the global
radiation).
INSTALLATION
Before installing the radiometer, refill the cartridge containing silica-gel
crystals. Silica gel absorbs humidity in the dome chamber and prevents, in
particular climatic condi-tions, condensation on the internal walls of the
domes and measurement alteration.
Do not touch the silica gel crystals with your hands while refilling the
cartridge. Carry out the following instructions in an environment as drier as
possible:
- Loosen the three screws that fix the white shade disk.
- Unscrew the silica gel cartridge using a coin.
- Remove the cartridge perforated cap.
- Open the sachet containing silica gel (supplied with the radiometer).
- Fill the cartridge with the silica gel crystals.
- Close the cartridge with its own cap, paying attention that the sealing O-ring be properly positioned.
- Screw the cartridge to the radiometer body using a coin.
- Check that the cartridge is screwed tightly (if not, silica gel life will be reduced).
- Position the shade disk and screw it with the screws.
- The radiometer is ready for use.
The figure below shows the operations necessary to fill the cartridge with the silica gel crystals.
- The radiometer must be mounted in an easy-to-reach location in order to clean the dome regularly and carry out maintenance. At the same time, make sure that no buildings, constructions, trees or obstructions exceed the horizontal plane where the radiometer lies. If this is not possible, select a site where obstructions in the path of the sun from sunrise to sunset do not exceed 5 degrees of elevation.
- The radiometer must be located far from any kind of obstruction, which might re-flect sunlight (or sun shadow) onto the radiometer itself.
- For fixing, use the holes on the radiometer body (remove the shade disk to access the holes and reposition it after mounting) or the suitable accessories (see the fig-ures below). In order to allow an accurate horizontal positioning, the radiometer is equipped with a levelling device: the adjustment is made by means of the two lev-elling screws that allow adjusting the radiometer inclination. The mast height does not exceed the radiometer plane to avoid measurement errors caused by any re-flection or shadow of the mast itself.
- It is preferably to thermally insulate the radiometer from its mounting bracket en-suring, at the same time, a good electrical contact to ground.
ELECTRICAL CONNECTIONS
LPUVA02… has a 4-pole connector and uses the CPM12AA4… optional cables.
The metallic housing of the radiometer should preferably be grounded ( )
locally. In this case, do not connect the wire of the cable corresponding to
the housing to prevent ground loops.
Only if it is not possible to ground locally the metallic case of the
radiometer, connect the wire of the cable corresponding to the housing to
ground. Note: in LPUVA02AV the housing is not connected to the connector.
LPUVA02 CONNECTIONS
The radiometer LPUVA02 is passive and does not require power supply. It is to be connected either to a millivoltmeter or to a data acquisition system. Typically, the ra-diometer output signal does not exceed 50 mV. In order to better exploit the radiome-ter features, the readout instrument should have 1 µV resolution.
Connector | Function | Color |
---|---|---|
1 | +Vout | Red |
2 | -Vout | Blue |
3 | Housing (C) | White |
4 | Cable shield (SH) | Black |
LPUVA02AC CONNECTIONS
The radiometer LPUVA02AC has 4…20 mA output and requires 10…30 Vdc external power supply. It is to be connected to a power supply and an instrument with 4…20 mA input as shown in fig. 4.2. The load resistance of the instrument reading the signal must be ≤ 500 Ω.
Connector | Function | Color |
---|---|---|
1 | Positive (Iin) | Red |
2 | Negative (Iout) | Blue |
3 | Housing (C) | White |
4 | Cable shield (SH) | Black |
LPUVA02AV CONNECTIONS
The radiometer LPUVA02AV has 0…1 V, 0…5 V or 0…10 V output (depending on the ordered output) and requires external power supply: 10…30 Vdc for 0…1 V and 0…5 V outputs, 15…30 Vdc for 0…10 V output. It is to be connected to a power sup-ply and an instrument with voltage input as shown in fig. 4.3. The load resistance of the instrument reading the signal must be ≥ 100 kΩ.
Connector | Function | Color |
---|---|---|
1 | Output positive (+Vout) | Red |
2 | Output negative Power supply negative (GND) | Blue |
3 | Power supply positive (+Vdc) | White |
4 | Cable shield (SH) | Black |
MEASUREMENT
Below are the ways to calculate the irradiance.
LPUVA02
Each radiometer is distinguished by its own sensitivity (or calibration
factor) S ex-pressed in µV/(Wm-2) and shown in the label on the radiometer
(and in the calibration report).
The irradiance Ee is obtained by measuring with a multimeter the difference of
potential DDP at the ends of the sensor and applying the following formula:
Ee = DDP / S
where:
Ee is the irradiance expressed in W/m2;
DDP is the difference of potential expressed in µV measured by the multimeter;
S is the sensitivity of the radiometer expressed in µV/(Wm-2).
LPUVA02AC
The 4…20 mA output signal corresponds to 0…200 W/m2 irradiance range.
The irradiance Ee is obtained by measuring with a multimeter the current Iout
absorbed by the sensor and applying the following formula:
Ee = 12.5 • (Iout – 4mA)
where:
Ee is the irradiance expressed in W/m2;
Iout is the current expressed in mA absorbed by the radiometer.
LPUVA02AV
The output signal (0…1 V, 0…5 V or 0…10 V depending on the version)
corresponds to 0…200 W/m2 irradiance range.
The irradiance Ee is obtained by measuring with a multimeter the output
voltage Vout of the sensor and applying the following formula:
- Ee = 200 • Vout for the version 0…1 V
- Ee = 40 • Vout for the version 0…5 V
- Ee = 20 • Vout for the version 0…10 V
where:
Ee is the irradiance expressed in W/m2;
Vout is the output voltage expressed in V measured by the multimeter.
MAINTENANCE
In order to grant measurements high accuracy, it is important to keep the glass dome clean. Consequently, the more the dome will be kept clean, the more measurements will be accurate.
You can wash it using water and standard papers for lens. If necessary, use pure ETHYL alcohol. After using alcohol, clean again the dome with water only.
Because of the high temperature changes between day and night, some condensation might appear on the radiometer dome. In this case the performed reading is highly over-estimated. To minimize the condensation, the radiometer is provided with a car-tridge containing dessicant material (silica-gel). The efficiency of the silica-gel crystals decreases over time while absorbing humidity. Silica-gel crystals are efficient when their colour is yellow, while they turn white/translucent as soon as they lose their efficiency. Read instructions at chapter 3 about how to replace the silica-gel crystals. Silica-gel typical lifetime goes from 2 to 6 months depending on the environment where the radiometer works.
To exploit all the radiometer features, it is highly recommended that the calibration be checked annually.
TECHNICAL SPECIFICATIONS
Measuring range | 0…200 W/m2 |
---|---|
Typical sensitivity | 70…200 µV/Wm-2 |
Viewing range | 2p sr |
Impedance | 3 kΩ |
Spectral range | 342 nm…384 nm (1/2) |
330 nm…393 nm (1/10)
320 nm…400 nm (1/100)
Peak: 365 nm
Response time| <0.5 s (95%)
Output| LPUVA02 = µV/Wm-2 LPUVA02 AC = 4…20 mA
LPUVA02 AV = 0…1, 0…5, 0…10 V (depending on the model)
Power supply| 10…30 Vdc (15…30 Vdc only for 0…10 V output)
Operating temperature| -40…+80 °C
Response according to the cosine law| < 8 % (0…80°)
Long term instability (1 year)| <|±3| %
Non linearity| <½±1½ %
Temperature response| < 0.1 %/°C
Weight| 900 g approx.
SAFETY INSTRUCTIONS
General safety instructions
The instrument has been manufactured and tested in accordance with the safety
standard EN61010-1:2010 “Safety requirements for electrical equipment for
meas-urement, control and laboratory use” and has left the factory in perfect
safety tech-nical conditions.
The instrument proper operation and operating safety can be ensured only if
all standard safety measures as well as the specific measures described in
this manual are followed.
Do not use the instruments in places where there are:
- Corrosive or flammable gases.
- Direct vibrations or shocks to the instrument.
- High-intensity electromagnetic fields, static electricity.
User obligations
The instrument operator shall follow the directives and regulations below that
refer to the treatment of dangerous materials:
- EEC directives on workplace safety.
- National law regulations on workplace safety.
- Accident prevention regulations.
ACCESSORIES ORDERING CODES
The radiometer is supplied with shade disk, sylica-gel cartridge, 2 spare sachets, lev-elling device, M12 connector and Calibration Report.
Accessories
- LPSP1 UV-resistant shade disk (spare part).
- LPS1 Only attachment bracket, suitable for mast with diameter 40…50 mm. Installation on horizontal or vertical mast.
- LPRING02 Base with levelling device and adjustable holder for mounting the pyranometer in an inclined position (specify upon ordering on which radiometer model has to be mounted).
- HD2003.79K Kit to mount the radiometer on Æ 40 mm clamping. To install the radiometer on a transverse mast.
- LPS6 Kit for the installation of the radiometer. The kit includes: 750 mm mast, base fitting, graduated support plate, bracket for radiometer.
- CPM12AA4… 4-pole cable. 4-pole M12 connector on one end, open wires on the other end. Available length 2 m (CPM12AA4.2), 5 m (CPM12AA4.5) or 10 m (CPM12AA4.10).
- LPSG Cartridge to contain desiccant silica-gel crystals, complete with Oring and cap (spare part).
- LPG Pack of 5 sachets of silica-gel crystals.
DELTA OHM metrology laboratories LAT N° 124 are ISO/IEC 17025 accredited by ACCREDIA for Temperature, Humidity, Pressure, Photometry / Radiometry, Acous- tics and Air Velocity. They can supply calibration certificates for the accredited quantities.
DECLARATION OF CONFORMITY
WARRANTY
Delta OHM is required to respond to the “factory warranty” only in those cases
provided by Legislative Decree 6 September 2005 – n. 206. Each instrument is
sold after rigorous inspections; if any manufacturing defect is found, it is
necessary to contact the distributor where the instrument was purchased from.
During the warranty period (24 months from the date of invoice) any
manufacturing defects found will be repaired free of charge. Misuse, wear,
neglect, lack or inefficient maintenance as well as theft and damage during
transport are excluded. Warranty does not apply if changes, tampering or
unauthorized repairs are made on the product. Solutions, probes, electrodes
and microphones are not guaranteed as the improper use, even for a few
minutes, may cause irreparable damages.
Delta OHM repairs the products that show defects of construction in accordance
with the terms and conditions of warranty included in the manual of the
product. For any dispute, the competent court is the Court of Padua. The
Italian law and the “Convention on Contracts for the International Sales of
Goods” apply.
TECHNICAL INFORMATION
The quality level of our instruments is the result of the continuous product development. This may lead to differences between the information reported in the manual and the instrument you have purchased. In case of discrepancies and/or inconsistencies, please write to sales@deltaohm.com. Delta OHM reserves the right to change technical specifications and dimensions to fit the product requirements without prior notice.
DISPOSAL INFORMATION
Electrical and electronic equipment marked with specific symbol in compliance with 2012/19/EU Directive must be disposed of separately from household waste. European users can hand them over to the dealer or to the manufacturer when purchasing a new electrical and electronic equipment, or to a WEEE collection point designated by local authorities. Illegal disposal is punished by law.
Disposing of electrical and electronic equipment separately from normal waste helps to preserve natural resources and allows materials to be recycled in an environmentally friendly way without risks to human health.