MARTINDALE PD230SRD DC Proving Unit Instruction Manual
- June 13, 2024
- MARTINDALE
Table of Contents
MARTINDALE PD230SRD DC Proving Unit
Product Information
The product is a proving unit manufactured by Martindale Electric. It is designed to provide a safe and reliable method for testing the functionality of test lamps, voltage testers/indicators, and digital multimeters (DMMs).
The proving unit comes in different models, including PD230SRD, PD440S, PD440SRD, PD440SRDX, PD690S, and PD690SRD.
The product manual provides comprehensive instructions and safety information for the proper use and maintenance of the proving unit.
For technical support, customers can contact Martindale Electric at +44 (0)1923 441717 or support@martindale-electric.co.uk.
Product Usage Instructions
Safety Information
- Before using the equipment, it is essential to read and understand all the safety information provided in the manual. Failure to comply with the warnings and instructions can result in serious injury or damage.
- Pay particular attention to the warnings, precautions, and technical specifications.
- Save the manual for future reference and check the manufacturer’s website for any updated instructions or product information.
Precautions
Follow these precautions to ensure safe operation:
- Always check the unit for cracks or damage before use. Ensure it is free from dust, grease, and moisture. Do not use if damaged.
- Ensure the battery cover is fitted before use.
- Avoid severe mechanical shock or vibration and extreme temperatures.
- Remove batteries when the unit is not in use for an extended period to prevent corrosion.
- Do not short the output terminals.
- The equipment should only be used by a suitably trained and qualified person.
- Avoid using the equipment in wet conditions or with wet hands.
- Using the equipment in a manner not specified by the manufacturer may impair its protection capabilities.
Meaning of Symbols and Markings
The manual explains the meaning of various symbols and markings found on the equipment. These symbols indicate caution, risk of danger, electric shock, double or reinforced insulation (Class II), compliance with relevant EU Directives, and UK Conformity Assessed marking. It also provides information on the proper disposal of the equipment according to relevant EU Directives.
Introduction
Inspection
Prior to use, inspect the shipping carton, unit, and accessories for any
signs of damage. If any damage is detected, contact your distributor
immediately.
Description
The manual provides a detailed description of the proving unit, including
its features and specifications. Refer to this section for a better
understanding of the product.
SAFETY INFORMATION
Always read before proceeding.
REMEMBER : SAFETY IS NO ACCIDENT
- Before using this equipment read all instructions and cautionary markings in this manual and and on the equipment. These instructions contain both information and warnings that are necessary for the safe operation and maintenance of this product. Failure to understand and to comply with the warnings and instructions can result in serious injury, damage or even death.
- Particular attention should be paid to the Warnings, Precautions and Technical Specifications.
- Read and save these instructions for future reference. Updated instructions and product information are available at: www.martindale-electric.co.uk
Precautions Warnings
These warnings apply to the proving units.
-
To reduce the risk of electric shock, fire or injury:
It is important that proper safety measures are taken when working with voltages exceeding the extra low voltage (ELV) threshold of 50V AC RMS or 120V DC. -
Before use check the unit for cracks or any other damage. Make sure the unit is free from dust, grease and moisture. Also check any associated accessories for damage. Do not use if damaged.
-
Do not use if the battery cover is not fitted
Cautions:
- Avoid severe mechanical shock or vibration and extreme temperature.
- To avoid corrosion from leaking batteries, remove the batteries when the unit is not in use for an extended period.
- Do not short the output terminals.
- The equipment must only be used by a suitably trained and qualified person.
- Do not use it in wet conditions or with wet hands.
- If this equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
Meaning of Symbols and Markings
- Caution: risk of danger & refer to instructions
- Caution: risk of electric shock
- Equipment protected by double or reinforced insulation (Class II)
- Equipment complies with relevant EU Directives
- The equipment complies with relevant UK Conformity Assessed marking
- End-of-life disposal of this equipment should be in accordance with relevant EU Directives
INTRODUCTION
Inspection
Examine the shipping carton for any sign of damage. Inspect the unit and any
accessories for damage. If there is any damage then consult your distributor
immediately.
Description
- Voltage Indicating LEDs
- Calibration card (PD440SRDX only)
- Output voltage terminals
- AC/DC selection and output indication switch
- DC output indicates LED
- AC output indicates LED
- Battery low LED
The PD230SRD / PD440S / PD440SRD / PD440SRDX / PD690S /
PD690SRD are portable battery-powered proving units for testing of contact- type voltage detectors or other units with voltage detection capability, e.g. multimeters or multifunction testers. The table below shows the differences in the output voltage for each model .
After the voltage detector probes are pushed down onto the Proving Device (PD) unit output voltage terminals, the PD unit outputs voltage level(s) for period(s) indicated in the table before the voltage drops to 0V and the terminal’s output goes into low impedance mode
PD440S / PD690S
- The PD440S and PD690S output a fixed AC voltage for 5s before the output voltage is reduced to 0V and the terminal output goes into low impedance mode.
- The PD440S fixed output voltage is 440V AC 50Hz nominal. The PD690S fixed output voltage is 690V AC 50Hz nominal.
PD230SRD / PD440SRD / PD440SRDX / PD690SRD
The output voltage can be set to be to AC or DC mode using the AC/DC output
selection push button. The output voltage reduces in steps over a period of
10s.
For example, the PD230SRD outputs voltage in the following steps: 230V for 4s, 120V for 2s, and 50V for 4s before the output voltage is reduced to 0V and the terminal output goes into low impedance mode.
Product
| Output voltage step down|
NO LOAD output voltage step (50Hz nominal AC V / DC V)
---|---|---
PD690SRD| | 690| 440| 230| 120| 50
PD440SRD / PD440SRDX| | 440| 230| 120| 50|
PD230SRD| | 230| 120| 50| |
PD690S| | 690| | | |
PD440S| | 440| | | |
Accessories
All units come with the following accessories:
- 6 x 1.5V AA alkaline batteries
- Instructions
- MG3 magnetic hanger *
An excluding kits where a PD carry case is included
Battery Installation
Refer to Section 4.1 (Battery Replacement) for the battery installation
instructions for all units.
OPERATION
Caution Always make sure the precautions and limitations of the unit being tested are observed. If necessary refer to the specification of the unit being tested. Also refer to the graph of typical output voltage under various load conditions at the rear of these instructions.
AC / DC output voltage mode selection and mode indication
AC / DC output voltage mode selection and mode indication can be carried
out using the AC / DC output selection push The push button functions are
described below.
Function | Operation step/s | Description |
---|
Output mode indication
|
1. Momentarily press the push button while the proving device is not being used
| Output mode indications are as follow, AC output mode – “AC” LED ON,
“DC” LED OFF
DC output mode – “DC” LED ON, “AC” LED OFF
The LED will time out after approximately 10s. It is not necessary to wait for the LED to time out before proving a voltage indicator. If the LED is left to time out, then a minimum of 3s is required before pressing the button again or proving a voltage indicator.
|
1. Carry out “Output
| 1. Each time the push button is momentarily pressed the Output Mode changes to the opposite mode
2. The default startup output mode does not change
Change output| mode indication” function.
mode| 2. Momentarily press
| push button again
| 1. Carry out “Output|
If the LED fails to flash then this indicates that default output mode is already set
to the mode indicated by the “AC” or “DC” LED
| mode indication” function.
| 2. Press push button
| (and keep it pressed).
Change default| The LEDs will change to
startup output| indicate the new output
mode| mode
| 3. Release push button
| when the LED stops
| flashing (after about 3
| seconds)
Proving Unit LED indications
| LED indicators
---|---
| | Output mode| Output terminals voltages indicators
Product| Battery low| AC (50Hz)| DC| 690V| 400V|
230V| 120V| 50V| High voltage
PD690SRD| | | | | | | | |
PD440SRD| | | | | | | | |
PD440SRDX| | | | | | | | |
PD230SRD| | | | | | | | |
PD690S| | | | | | | | |
PD440S| | | | | | | | |
The previous table shows the LED indicators for each PD product type.
For example, the PD690SRD includes the following LED indicators- 690V, 440V, 230V, 120V, 50V.
Battery low indication
All PD products support the battery low LED, which is illuminated when the
battery voltage is low.
PD440S / PD690S
On these PD products the “High voltage” LED is ON when the live voltage is
present at the voltage output terminals. If the loading impedance of the
voltage indicator (at the output terminals) is less than 1kΩ then the “High
voltage” LED will start flashing.
PD230SRD / PD440SRD / PD440SRDX / PD690SRD
On these PD products there are multiple LED to indicate the voltage at the
PD output terminals when a voltage indicator is under test at the terminals.
Refer to the graph of typical output voltage under various voltage indicator
load conditions on the specification page. Also refer to the “PD Output
voltages table” under section 2.2. The LED indication will depend on the
loading impedance of the voltage indicator under test as well as the PD output
voltage steps in the “PD Output voltage table”. If the loading impedance of
the voltage indicator (at the output terminals) is less than 1kΩ then the
“50V” LED will start flashing.
Voltage Output Mode (AC / DC)
AC LED and DC LED indicate the PD output voltage mode.
- Proving a Test Lamp, Voltage Tester / Indicator, DMM etc
- Place one probe of the unit under test into the right hand socket of the proving unit until it makes contact with the terminal. Place the other probe into the left hand terminal of the proving unit and gently press It is important to follow this procedure in order to ensure PD output voltage is applied to the equipment under test from the start of the PD output voltage test cycle.
- The PD output voltage test cycle starts from when the voltage indicator is applied to the left hand terminal of the PD.
- On the PD440S and PD690S, the output voltage test cycle is 5s long.
- On the PD230SRD / PD440SRD / PD440SRDX / PD690SRD, the output voltage test cycle is up to 10s long.
- At the end of these PD units’ test cycles, the output voltage drops to 0V and the output goes into low impedance. For SRD and SRDX versions, a minimum of 3s is required between tests.
- Observe that the required indicators on the unit under test illuminate, then withdraw the probe from the right hand terminal first, and then the left.
- If none of the proving unit LED’s illuminate when performing the above tests, check the condition of the proving unit batteries and replace them if required (see section 4.1).
Test Lamp, Voltage Tester / Indicator Cables
During the above tests, emphasis should also be placed upon the flexing of
the UUT cable along its length, and particularly at the entry points to the
hand held elements, to confirm that the cable has not fractured. It may be
necessary to perform this test a number of times as the proving unit times out
after 5s on the PD440S /PD690S and after 10s on the PD230SRD / PD440SRD /
PD440SRDX / PD690SRD.
CALCHECK (PD440SRDX only)
The CALCHECK feature may be used to check that a multifunction tester,
insulation tester or continuity tester is maintaining ongoing accuracy between
calibrations. There are two insulation test resistance values, 0.5MΩ and 1MΩ.
There are also two continuity test resistance values, 0.5Ω and 1Ω.
One probe of the tester being checked should be placed on the desired resistance value and the other probe should be placed on the COM terminal, following the instruction manual for the tester being used. The CALCHECK feature is not designed to be used to check loop testers, where the high test current will damage the proving device. The CALCHECK (X models only) feature can be used as a regular check that instruments are not drifting between calibrations.
Records can be kept using the CALCHECK Verification Measurement sheet at the back of this manual. An initial measurement can be taken soon after calibration and the results recorded. Further checks can be made periodically, e.g. every month, and the results also recorded on the sheet.
Further copies of the sheet can be downloaded from www.martindale- electric.co.uk/manuals/.
MAINTENANCE
Battery Replacement
Remove the rear battery cover by turning the quarter turn screw 90° anti-
clockwise and then lifting the cover. Observing correct polarity fit 6 new
1.5V, AA alkaline batteries (IEC LR6, NEDA 15A). Replace the battery cover by
positioning the top onto the rear casing slots, lowering the cover and then
locking by turning the quarter turn screw 90° clockwise.
Note: Do not mix old and new batteries.
Cleaning
If contamination is found, clean with a damp soft cloth and, if necessary,
a mild detergent or alcohol. Do not use abrasives, abrasive solvents, or
detergents which can cause damage to the unit. If a mild detergent is used,
the unit should subsequently be thoroughly cleaned with a water dampened soft
cloth. After cleaning, dry and allow to remain in a dry environment for 2
hours before use.
Repair and Service
There are no user serviceable parts in this unit other than those that may
be described in section 4. Return to Martindale Electric if faulty. Our
service department will quote promptly to repair any fault that occurs outside
the guarantee period. Before the unit is returned, please ensure that you have
checked the unit and batteries.
Storage Conditions
The instrument should be kept in cool, dry conditions and not subjected to
shock, scratching or other damage, prolonged direct harsh sunlight, extremes
of temperature and in such a manner as to preserve the working life of the
unit. It is strongly advised that the unit is not kept in a tool box where
other tools may damage it.
WARRANTY AND LIMITATION OF LIABILITY
This Martindale product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is 2 years and begins on the date of receipt by the end user. This warranty extends only to the original buyer or end- user customer, and does not apply to fuses, disposable batteries, test leads or to any product which, in Martindale’s opinion, has been misused, altered, neglected, contaminated, or damaged by accident or abnormal conditions of operation, handling or storage.
Martindale’s authorised resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Martindale.
Martindale’s warranty obligation is limited, at Martindale’s option, to refund of the purchase price, free of charge repair, or replacement of a defective product which is returned to Martindale within the warranty period.
This warranty is the buyer’s sole and exclusive remedy and is in lieu of all other warranties, expressed or implied, including but not limited to any implied warranty of merchantability or fitness for a particular purpose. Martindale shall not be liable for any special, indirect, incidental or consequential damages or losses, including loss of data, arising from any cause or theory. Since some jurisdictions do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any part of any provision of this warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the validity orenforceability of any other provision or other part of that provision. Nothing in this statement reduces your statutory rights
CALCHECK verification measurements
(PD440SRDX only)
PD model
PD serial no.
Test instrument model
Test instrument serial no.
Test instrument date of calibration
Initial measurement
Date | Insulation | Continuity | Initials |
---|---|---|---|
0.5MΩ | 1MΩ | 0.5Ω | 1Ω |
Periodic verification measurements
Date | Insulation | Continuity | Initials |
---|---|---|---|
0.5MΩ | 1MΩ | 0.5Ω | 1Ω |
Periodic verification measurements
(PD440SRDX only)
Date | Insulation | Continuity | Initials |
---|---|---|---|
0.5MΩ | 1MΩ | 0.5Ω | 1Ω |
Specification
Proving Units
- PDS
- PDSRD
- PDSRDX
Electrical
Product
| Output voltage step down| ****
NO LOAD output voltage step (50Hz nominal AC V / DC V)
---|---|---
PD690SRD| | 690| 440| 230| 120| 50
PD440SRD / PD440SRDX| | 440| 230| 120| 50|
PD230SRD| | 230| 120| 50| |
PD690S| | 690| | | |
PD440S| | 440| | | |
All Units
Output loading: See typical output voltage vs. loading graph
CALCHECK – PD440SRDX only
-
Insulation test resistance nominal values:
0.5MΩ (max test voltage 1kV) 1.0MΩ (max test voltage 1kV) -
Continuity test resistance nominal values:
0.5Ω (max test current 200mA) 1.0Ω (max test current 200mA)
General
- Power : Internal batteries Internal batteries: 6 x 1.5V, AA alkaline batteries (IEC LR6, NEDA 15A)
- Dimensions : 150 x 80 x 35mm. Weight packed: 435g approx. (with batteries)
- Includes : 6 x 1.5V AA alkaline batteries, instructions
Environmental
- Location : Suitable for indoor use, and outdor use in dry weather conditions only.
- Operating temperature : -10°C to 40°C at max. 70% R.H.
- Altitude : up to 2000m Pollution degree: 2
Typical output voltage AC rms vs loading (kΩ)
Check out what else you can get from Martindale:
- 18th Edition Testers
- Accessories
- Calibration Equipment
- Continuity Testers
- Electricians’ Kits
- Environmental Products
- Full Calibration & Repair Service
- Fuse Finders
- Digital Clamp Meters
- Digital Multimeters
- Labels
- Microwave Leakage Detectors
- Safe Isolation Kits
- Multifunction Testers
- Non-trip Loop Testers
- Pat Testers & Accessories
- Phase Rotation Testers
- Proving Units
- Socket Testers
- Thermometers & Probes
- Test Leads
- Voltage Indicators
- Specialist Drummond Testers
Martindale Electric Co. Ltd. Metrohm House, 12 Imperial Park, Imperial Way, Watford WD24 4PP. T : +44 (0)1923 441717 F : +44 (0)1923 446900 www .martindale-electric.co.uk sales @martindale-electric.co.uk
Ver. D1.0
Due to policy of continuous development, Martindale Electric reserves the
right to alter equipment specification and description outlined in this
document without prior notice. No part of this document shall be deemed to be
part of any contract for the equipment unless specifically referred to as an
inclusion within such contract.
© 2023 Martindale Electric Co. Ltd. Registered in England No. 3387451. LITPD
References
- Martindale Electric - Manufacturer of Voltage Indicators, PAT Testers & Socket Testers
- Manuals - Martindale Electric
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>