TTBIO EZclick Couplings Dental Unit Instruction Manual

June 13, 2024
TTBIO

TTBIO-logo

TTBIO EZclick Couplings Dental Unit

Product Description

Separate the EZclick couplings from the hose.

Connect the handpiece exactly onto the EZclick coupling until it clicks audibly.

Removal from the handpiece (for EZclick-N series): Pull back the EZclick coupling ring and remove from the handpiece.

EZclick Couplings:

  • EZclick-K: Standard 4-hole connection and anti-suctionvalve.
  • EZclick-N: Standard 4-hole connection and anti-suction valve.

Fixed Tool:

  • EZclick-L / EZclick-LED: Standard 4-hole connection, anti-suction valve, two electrical contacts, and Xenon Bulb.
  • EZclick-KL / EZclick-KLED / EZclick-K(5-Hole): Standard 4-hole connection, anti-suction valve, two electrical contacts, and Xenon Bulb.
  • EZclick-NL / EZclick-NLED: Standard 4-hole connection, anti-suction valve, two electrical contacts, and Xenon Bulb.

Indications for Use:

The EZclick series coupling attaches the handpiece to the dental unit tubing to deliver water and air to the handpiece.

Use environment/Intended user:

This device is to be connected to a dental high-speed handpiece and operated by a qualified professional (dentist) in the clinic.

Patient population:

Any patient (except babies). The user’s discretion is required prior to use.

  • Indications: N/A
  • Contraindications: N/A

The clinical benefits of our coupling enable the transformation of the air supplied by a dental unit to drive the high-speed handpiece.

Specifications

EZclick Couplings Specifications
EZclick & EZclick-K & EZclick-N coupling Standard 4-hole connection and anti-

suction valve.
EZclick-K (5 Hole) coupling| Standard 5-hole connection, anti-suction valve, and optical
rod.
EZclick-L & EZclick-KL & EZclick-NL coupling| Standard 4-hole connection, anti-suction valve, two electrical
contacts, and Xenon Bulb.
EZclick-LED & EZclick-KLED & EZclick-NLED coupling| Standard 4-hole connection, anti-suction valve, two electrical
contacts, and LED Bulb.

Table of EZclick couplings and suitable handpieces

Handpiece EZclick couplings
KAVO LUX handpiece EZclick-KL/EZclick-KLED/EZclick-K(5-Hole)
KAVO handpiece EZclick-K
TTBIO LUX handpiece EZclick-L/EZclick-LED
TTBIO handpiece EZclick
NSK Mach LUX handpiece EZclick-NL/EZclick-NLED
NSK Mach handpiece EZclick-N

Safety Notes

  1. Once EZclick couplings are damaged (i.e. deformed caused by a hit) or improperly used (i.e. sterilized above suggested temperature), be sure not to continue using them.
  2. Before operation, always make sure the product and its accessories meet the specified requirements.
  3. It is the duty of the user to use only faultless tools, for the correct purpose, and protect patients and third parties from all dangers.
  4. Please do not look directly into the lighting bulb, or it may hurt your eyes when you use the EZclick coupling with light.

Installation

  1. Connection to the hose:
    • Attach the EZclick couplings to the hose.
    • Use a fixed tool to fix the EZclick couplings.
    • Rotate the thread ferrule of the hose in a clockwise direction.
    • Spray oil onto the O-rings (for EZclick & EZclick-K series).

EZclick Couplings Instruction Manual

Product Description

TTBIO-EZclick-Couplings-Dental-Unit-
fig-1

Indications for Use: The EZclick series coupling attach the handpiece to the dental unit tubing to deliver water and air to the handpiece. Use environment/Intended user: This device is to be connected to dental high speed handpiece and operated by qualified professional (dentist) in the clinic.
Patient population: Any patient (except babies). The user’s discretion is required prior use.

  • Indications: N/A
  • Contraindications: N/A

The clinical benefits of our coupling enable to transform the air supplied by a dental unit to drive the high speed handpiece.

Specifications

  • EZclick & EZclick-K & EZclick-N coupling
  • Standard 4-hole connection and anti-suction valve.
  • EZclick-K (5 Hole) coupling
  • Standard 5-hole connection, anti-suction valve, and optical rod.
  • EZclick-L & EZclick-KL & EZclick-NL coupling
  • Standard 4-hole connection, anti-suction valve, two electrical contacts and Xenon Bulb.
  • Xenon Bulb ideal voltage range: 3.4~3.6 Vd.c. (for Ezclick-L & EZclick-KL)
  • Xenon Bulb recommended voltage: 3.5 Vd.c. (for EZclick-L & EZclick-KL)
  • Xenon Bulb ideal voltage range: 3.0-3.5 Vd.c. (for EZclick-NL)
  • Xenon Bulb recommended voltage: 3.3 Vd.c. (for EZclick-NL)
  • EZclick-LED & EZclick-KLED & EZclick-NLED coupling
  • Standard 4-hole connection, anti-suction valve, two electrical contacts and LED Bulb
  • LED Bulb ideal voltage range: 3.0-3.4 Vd.c. (for EZclick-LED & EZclick-KLED)
  • LED Bulb recommended voltage: 3.2 Vd.c. (for EZclick-LED & EZclick-KLED)
  • LED Bulb ideal voltage range: 3.0-3.4 Vd.c. (for EZclick-NLED)
  • LED Bulb recommended voltage: 3.2 Vd.c. (for EZclick-NLED)

Table of EZclick couplings and suitable handpieces

Handpiece EZclick couplings
KAVOⓇ LUX handpiece EZclick-KL/EZclick-KLED/

EZclick-K(5-Hole)

KAVOⓇ handpiece| EZclick-K
TTBIO LUX handpiece| EZclick-L/

EZclick-LED

TTBIO handpiece| EZclick
NSK MachⓇ LUX handpiece| EZclick-NL/

EZclick-NLED

NSK MachⓇ handpiece| EZclick-N

Safety Notes

  1. Once EZclick couplings are damaged (i.e. deformed cause by hit) or improperly used (i.e. sterilized above suggested temperature), be sure not to continue using them.
  2. Before operation, always make sure the product and its accessories meet the specified requirements.
  3. It is the duty of the user to use only faultless tools, for the correct purpose, and protect patients and third parties from all dangers
  4. Please do not look directly into the lighting bulb, or it may hurt eyes when you use the EZclick coupling with light.

Installation

Connection to the hose

  • Attach the EZclick couplings to the hose.
  • Use fixed tool to fix the EZclick couplings
  • Rotate the thread ferrule of the hose in a clockwise direction.
  • Spray oil onto the 0-rings. (for EZclick & EZclick-K series)

Removal from the hose

  • Use fixed tool to remove the EZclick couplings
  • Rotate the thread ferrule of the hose in an anti-clockwise direction.
  • Separate the EZclick couplings from the hose

Connect the handpiece
Connect the handpiece exactly onto the EZclick coupling until it clicks audibly.

Removal from the handpiece (for EZclick-N series)
Pull back the EZclick coupling ring and remove from the handpiece.

Fixed Tool

  1. Do not activate the coupling system during installation.
  2. Please choose the correct connected interface of the fixed tool according to the different types.

EZclick-N series

  1. Connection to the coupling
    Connect the fixed tool to the coupling. Ensure that the pin of the fixed tool is firmly inserted to the coupling hole until you hear the “Click” sound.

  2. Approach in the use of the coupling

    • as a coupling wrench
    • as a protection-against-dust tool
    • as a bulb replacement wrenchTTBIO-EZclick-Couplings-Dental-Unit-fig-9

Trouble Shooting

Replacing the bulb (for EZzclick, EZclick-K series)
Unscrew the cap by turning it anti-clockwise. Pull a bulb out of the coupling and insert a new one. Screw the cap by turning it clockwise.

Replacing the bulb (for EZclick-N series)
Refer to Fixed Tool section.

  1. Note the position of two contact lugs in the couplings.
  2. Do not touch the bulb immediately after operation.
  3. The LED bulb must be inserted into the polar position in the correct orientation.
  4. Inserted in the incorrect orientation may result in failure of lighting. If so, please rotate it 180° and re-insert the LED bulb.

Replacing the O-ring (for EZclick, EZclick-K series)
Press the sides of a defective O-ring between two fingers until a loop is formed and then slide the O-ring out of couplings and remove it.
Insert these new O-rings in the grooves and spray them with lubrication oil.

When sliding the O-ring, do not use any tools, such as tweezers, etc.

Cleaning, Disinfection and Sterilization

Cleaning exterior

  • The outside surface of the EZclick couplings can be cleaned with an alcohol solution.
  • Cleaning the EZclick couplings in an ultrasonic cleaner may result in malfunction.

Disinfection

  • The outside surface of the EZclick couplings can be disinfected with chemical disinfectants. Do not immerse EZclick couplings in disinfectants.

Sterilization

  • The EZclick couplings can be sterilized at 135°C.

Disposal

Follow your local and national laws, directives, standards and guidelines for disposal.

  • Medical device
  • Packaging

Disclaimer

  1. Information in this operation manual was carefully reviewed. We cannot assume any liability for damage arising from improper use of the hand piece or from use of this manual
  2. Products are subject to change without notice.
  3. This product is only for dental hand piece treatment and use for other purpose may damage the handpiece and cause severe consequence.
  4. Please use the product correctly according to the instruction.

KaVoⓇ & MULTlflex® are the registered trademarks of Kaltenbach & Voigt GmbH Co., Germany. NSK MachⓇ is the registered trademark of NAKANISHI INC. Japan.

Reporting

Any serious incident and adverse event (e.g. bur detached accidentally and poor clamping) caused by device deficiency must be reported to the manufacturer, the authorized representative and the local competent authorities.

Quality and Service

  1. All of our products have passed strict QC procedures before delivery.
  2. Please note that improper use or not using the recommended parts and accessories or modifying our products without our approval may shorten the lifetime of the product or cause damage.
  3. Please contact us or dealers with proof of purchase for consultation whenever you have any questions.
  4. eIFU use the ref on https://www.ttbio.com/download.php.

Symbol

| Manufacturer| | Date of manufacture
---|---|---|---
| CE mark

(European Conformity)

| | Serial number
| Authorized representative in the European Community| | Sterilizable up to the temperature specified
| Data Matrix Code for product information including UDI

(Unique Device Identification)

| |

Caution

| Instructions for use|

Manufactured by

TIBIO CORP.

  • 2F.,NO.7, 6TH ROAD INDUSTRY PARI
  • TAICHUNG. TAIWAN 40755
  • TEL: 886-4-23595958