LEISTER G1200 Thermocouple Quick Response Thermometer User Manual
- June 13, 2024
- LEISTER
Table of Contents
- LEISTER G1200 Thermocouple Quick Response Thermometer
- Product Information
- Product Usage Instructions
- About this documentation
- Safety
- The device at a glance
- Operation
- Measurement Basics
- Operation and maintenance
- Error and system messages
- Technical data
- Service
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
LEISTER G1200 Thermocouple Quick Response Thermometer
Product Information
The G 1200 is a compact manual Thermocouple Quick Response Thermometer designed for use with exchangeable Type K sensors. It is produced by an undisclosed manufacturer.
Table of Contents
- About this documentation
- Safety
- The device at a glance
- Operation
- Operation and maintenance
- Error and system messages
- Technical data
- Disposal
- Service
About this documentation
The documentation provides important information on the operation and usage of the G 1200 Thermocouple Quick ResponseThermometer. It is recommended to read and familiarize yourself with this document before using the device.
Foreword
It is essential to carefully read and understand this document before operating the G 1200 Thermocouple Quick Response Thermometer. Keep this document easily accessible near the device for quick reference.
Product Usage Instructions
- Ensure you have read and understood the documentation provided.
- Make sure the device is powered off before proceeding.
- Connect the Type K sensor to the appropriate connection port on the device.
- Power on the G 1200 Thermocouple Quick Response Thermometer.
- To access the configuration menu, press the corresponding button (refer to the operating elements section in the documentation).
- Adjust the measuring input according to your requirements using the configuration menu.
- Follow the operating and maintenance notices provided in the documentation for proper usage and care of the device.
- Monitor the battery indicator to ensure sufficient power for accurate measurements.
- If the battery needs to be replaced, refer to the documentation for instructions on changing the battery.
- In case of any error or system messages, consult the documentation for troubleshooting steps.
- Refer to the technical data section for detailed specifications of the G 1200 Thermocouple Quick Response Thermometer.
- Dispose of the device according to the disposal guidelines provided in the documentation.
- For any service-related issues, refer to the manufacturer’s information in the documentation.
- Calibration and adjustment services, as well as compatible accessories, are available (refer to the respective sections in the documentation).
About this documentation
Foreword
Read this document carefully and familiarise yourself with the operation of
the device before you use it. Keep this document ready to hand and in the
immediate vicinity of the device so that it is available to the personnel/user
for reference at all times in case of doubt.
The user must have carefully read and understood the operating manual before
begin-ning any work.
Legal notices
The liability and warranty of the manufacturer for damages and consequential
damag-es are voided with misuse, disregarding this document, disregarding
safety notices, assignment of inadequately qualified technical personnel and
arbitrary modifications of the device.
This document is entrusted to the recipient for personal use only. Any
transmission, duplication, translation into other languages or excerpts from
this operating manual require the consent of the manufacturer.
The manufacturer assumes no liability for print errors.
Further information
Software version of the device:
V1.0 or later
For the exact product name, refer to the type plate on the rear side of the
device.
NOTE
For information about the software version, press and hold the ON button to
switch on for longer than 5 seconds. The series is shown in the main display
and the software version of the device is shown in the secondary display.
Safety
Explanation of safety symbols
DANGER
This symbol warns of imminent danger, which can result in death, severe bodily
injury, or severe property damage in case of non-observance.
CAUTION
This symbol warns of potential dangers or harmful situations, which can cause
damage to the device or to the environment in case of non-observance.
NOTE
This symbol indicates processes, which can have a direct influence on
operation or can trigger an unforeseen reaction in case of non-observance.
Foreseeable misuse
The fault-free function and operational safety of the device can only be
guaranteed if applicable safety precautions and the device-specific safety
instructions for this docu-ment are observed. If these notices are
disregarded, personal injury or death, as well as property damage can occur.
DANGER
Incorrect area of application!
In order to prevent erratic behaviour of the device, personal injury and
property dam-age, the device must be used exclusively as described in the
chapter Description in the operating manual.
- The device is not suitable for use in explosion-prone areas!
- The device must not be used for diagnostic or other medical purposes on pa-tients!
- Not suitable for use with requirements on functional safety, e.g. SIL!
- Do not use in safety / emergency stop devices!
Safety instructions
CAUTION
Risk of injury at the temperature sensor!
Devices with insertion probes entail the risk of stab injuries due to the
pointed probe design.
There is a risk of burns when measuring in high (and very low) temperatures.
Use gloves if necessary.
CAUTION
Empty batteries and batteries of inferior quality can leak more easily, which
can destroy the device. Please also observe the instructions in the chapter
“Operation and maintenance”.
NOTE
This device does not belong in children’s hands!
NOTE
The device housing is not designed for continuous contact with foods. For
measurements in food in accordance with EC Regulation 1935 / 2004, suitable
temperature probes must be used
Intended use
The device is thermometer for type K interchangeable probes.
It is designed, with the appropriate probe, for precise and instantaneous
temperature
measurements in the following media:
- Liquids, gases, soft plastic materials, bulk material
- Food
Qualified personnel
For commissioning, operation and maintenance, the relevant personnel must have
adequate knowledge of the measuring process and the significance of the
measurements. The instructions in this document must be understood, observed
and followed. In order to avoid any risks arising from interpretation of the
measurements in the concrete application, the user must have additional
expertise. The user is solely liable for damages/danger resulting from
misinterpretation due to inadequate expertise.
The device at a glance
Display elements
Display
- Battery indicator: Evaluation of the battery status
- Unit display: Display of units or min/max/hold
- Main display: Measurement of the current temperature or value for min/max/hold
- Auxiliary display: Measurement of the current temperature in min/max/hold mode with unit
Connections
Thermos couple plug Connection for temperature sensor
CAUTION
The device ensures extensive protection against splashing water and rain.
However, immersion in water is not possible due to the socket opening.
- Protect the connector opening/contacts from soiling and moisture!
- Dry damp device and plug connectors as quickly as possible!
Operating elements
- On / Off button
- Press the brief: Switch on the device
- Activate/deactivate lighting
- Long press: Switch off the device
- Reject changes in a menu
- Up / Down button
- Press briefly
- Display of the min/max value
- Change value of the selected parameter
- Long press: Reset the min/max value of the current measure-ment
- Both simultaneously: Rotate display, overhead display
- Function button
- Press briefly
- Freeze measurement (Hold)
- Call up the next parameter
- Operating status
- The device is in the measured value display
- The device is in a menu
Operation
Opening the configuration menu
- Press the Function button for 2 seconds to open the Configuration menu.
- (ONF appears in the display. Release the Function button
Parameter | Values | Meaning |
---|---|---|
|
POFF
| Shut-off time
|
OFF
| No automatic shut-off
| 0:15, 0:30, 1:00,
4:00, 12:00
| Automatic shut-off after a selected time in hours:minutes, during which no buttons have been pressed
L,TE
| Backlight
|
OFF
| Backlight deactivated
| 0:15, 0:30, 1:00,
2:00, 4:00
| Automatic shut-off of the backlight after a selected time in
minutes:seconds, during which no buttons have been pressed
|
ON
| No automatic shut off of the backlight
UN,T
| Display unit
| °(| Temperature display in °C
| °F| Temperature display in °F
IN,T
| Factory settings
|
NO
| Use current configuration
|
YES
| Reset device to factory settings.
After confirming with the function-button , the display
shows: IN,T DONE
Adjustment of the measuring input
The temperature input can be adjusted with the zero point correction and the
gradient correction. If an adjustment is made, you change the pre-adjusted
factory settings. This is signalled with the display text T.OF or T.SL when
switching on.
- Switch the device off.
- Hold the down button and press the On/Off button briefly to switch on the device and open the Adjustment menu.
- The display shows the first parameter. Release the down button.
Parameter | Values | Meaning |
---|---|---|
|
T.OF| Zero point correction
| 0.0| No zero point correction
| -5.0 … 5.0| Zero point correction in °C. (at °F -9.0 .. 9.0)
T.SL| Gradient correction of the temperature
| 0.00| No gradient correction of the temperature
| -5.00 … 5.00| Gradient correction in %
Formula used by device:
- Temperature = °C: Display = (measured value – T.OF) * (1 + T.SL / 100)
- Temperature = °F: Display = (meas. value – 32 °F – T.OF) * (1 + T.SL / 100) + 32 °F
Adjustment example:
Temperature unit= °C, the adjustment is carried out in 2 separate steps at 0
°C (e.g. ice water) and at a reference temperature (e.g. clinical thermometer
with water bath 37 °C).- First set the values of t.oF and t.SL in the
adjustment menu to 0
Zero point: Set the temperature probe to 0 °C and let it adjust.
- Start the adjustment menu and enter the display value at 0 °C for t.oF
- After leaving the menu, the device should now display 0.0 °C.
- Slope: Set the temperature sensor to the reference temperature and let it adjust.
- Calculate slope correction:
- Start adjustment menu and enter the calculated value at t.SL
- Exit menu, the device should now display the reference temperature
Measurement Basics
Sensor/device accuracy
The device can be equipped with different interchangeable sensors. The
temperature sensors are divided into the following classes according to EN
60584-2.
class | deviation | temperature range |
---|---|---|
1 | ±1.5 °C or ±0.4 % of measured value * | -40 … 1000 °C |
2 | ±2.5 °C or ±0.75 % of measured value * | -40 … 1200 °C |
3 | ±2.5 °C or ±1.5 % of measured value * | -200 … 40 °C |
The system accuracy of the measuring chain (device + probe) is the sum of the accu-racy of the probe and the accuracy of the characteristic curve and cold junction of the device.
Possible measuring errors
- Immersion depth
- Liquid: Immerse to a depth of at least 20 mm and then stir. Otherwise, measuring errors can occur due to the heat transmission of the sensor tube if the immer-sion depth is too shallow.
- Gas: Immerse as far as possible into the gas to be measured so that the measur-ing sensor is subjected to a heavy flow.
Surface effects and poor heat transfer
Special measuring sensors are required for this purpose. Surface
characteristics, de-sign of the measuring sensor, heat transfer and
environmental temperature influence the measurement result.
NOTE: Thermally conductive paste between the measuring sensor and surface
can also in-crease measurement accuracy in some cases
Cooling / evaporation
When measuring the air temperature, the probe should be dry, otherwise the tempera-ture measured will be too low.
Response time
An adequate wait time must be observed for the measuring process before
reading the measured value. The response time T90 describes the time in which
the displayed measured value reached 90% of the end value.
Operation and maintenance
- Operating and maintenance notices
NOTE
The device and temperature probe must be handled with care and used in
accordance with the technical data. Do not throw or strike
NOTE
Plugs and sockets must be protected from soiling.
NOTE
If the not used for an extended period of time, the batteries must be removed.
Leaks from the batteries are avoided as a result.
Battery
Battery indicator
- If the empty frame in the battery display blinks, the batteries are depleted and must be replaced. However, the device will still operate for a certain length of time.
- If the BAT display text appears in the main display, the battery voltage is no longer adequate for operation of the device. The battery is fully depleted.
Changing battery
DANGER
Danger of explosion!
Using damaged or unsuitable batteries can generate heat, which can cause the
batter-ies to crack and possibly explode!
- Only use high-quality and suitable alkaline batteries
CAUTION
Damage!
If the batteries have different charge levels, leaks and thus damage to the
device can occur.
- Only use high-quality and suitable alkaline batteries!
- Do not use different types of batteries!
- Remove depleted batteries immediately and dispose of them at a suitable collection point.
NOTE
Unnecessary unscrewing endangers the protection against moisture and should
therefore be avoided.
NOTE
Read the following handling instructions before replacing batteries and follow
them step by step.
If disregarded, the device could be damaged or the protection from moisture
could be diminished.
- Unscrews the Phillips screws (A)and remove the cover.
- Carefully replace the two Mignon AA batteries (B). Ensure that the polarity is correct! It must be possible to insert the batteries in the correct position without using force.
- The O-ring (C) must be undamaged, clean and positioned at the intended depth.
- Fit the cover (D) on evenly. The O-ring must remain at the intended depth!
- Tighten the Phillips screws (A).
Error and system messages
Display | Meaning | Possible causes | Remedy |
---|---|---|---|
—- | No suitable meas- uring probe con- nected |
Measurement far outside of the measuring range
| Incorrect measuring probe
Measuring probe or device defect
| Connect a suitable meas- uring probe
Measurement leaves the permissible range
Send in for repair
No display, unclear char- acters or no response when buttons are pressed|
Battery depleted System error Device is defective| Battery depleted Error in
the device| Replace battery Send in for repair
BAT| Battery depleted| Battery depleted| Replace battery
ERR.1| Measuring range exceeded| Measurement too high
Incorrect measuring probe connected
Measuring probe or device defect
| Stay within allowable measurement range
Check measuring probe
Send in for repair
ERR.2| Measuring range is undercut| Measurement too low
Incorrect measuring probe connected
Measuring probe or device defect
| Stay within allowable measurement range
Check measuring probe
Send in for repair
ERR.7| System error| Cold junction out- side operating range or defect| Stay within allowable operating temperature
Send in for repair
SYS ERR| System error| Error in the device| Switch device on/off Replace batteries Send in for repair
Technical data
Sensor connection | Connection socket for miniature flat plug |
---|---|
Supported sensor type | Thermocouple type K (NiCr-Ni) |
Measuring range | -65.0 … +1200.0 °C (-85.0 … +2192 °F) |
Accuracy
(at nominal temperature)
| Note: The system accuracy depends on the accuracy of the used temperature
sensor!
| Device| ±0.1 % of m.v. ±1 °C
Cold junction| ±0.3 °C
Temperature drift| ±0.01 % of m.v. / K ±0.025 °C / K
Measuring cycle| approx. 3 measurements per second
Display| 3-line segment LCD, additional symbols, illuminated (white, luminous
duration adjustable),
Display orientation 180° rotatable (overhead display)
Additional functions| Min/max/hold
Adjustment| Offset and gradient correction
Housing| Break-proof ABS housing
| Protection rating| IP40 (sensor socket)
IP65 / IP67 (rest of the housing)
Dimensions LWH| 108 54 28 mm
Weight| approx. 125 g, incl. batteries
Nominal temperature| 25 °C
Oberating conditions| -20 to 50 °C; 0 to 85 %RH (non-condensing)
Current supply| 2 * AA batteries (mignon)
| Current require- ment| approx. 1 mA, approx. 3 mA with backlight
battery life| Service life > 2500 hours with alkaline batteries (without
backlighting)
| Battery indicator| 4-stage battery status indicator,
Replacement indicator for depleted batteries: “BAT”
---|---|---
Auto-power-OFF function| The device switches off automatically if this is
activated
Directives and stand- ards| The devices conform to the following Directives of
the Coun- cil for the harmonisation of legal regulations of the Member States:
2014/30/EU EMC Directive 2011/65/EU RoHS
Applied harmonised standards:
EN 61326-1:2013 Emission limits: Class B
Immunity according to Table 1 Additional errors: < 1 % FS
EN IEC 63000:2018
The device is intended for mobile use and/or stationary operation in the scope of the specified operating conditions without further limitations.
Disposal
Separation by material and recycling of device components and packaging must take place at the time of disposal. The valid regional statutory regulations and directives applicable at the time must be observed.
NOTE
The device must not be disposed of with household waste. Return it to us,
freight prepaid. We will then arrange for the proper and
environmentallyfriendly disposal. Private end users in Germany have the
possibility of dropping off the device at the municipal collection centre.
Batteries must be removed beforehand! Please dispose of empty batteries at the
collection points intended for this purpose
Service
Manufacturer
If you have any questions, please do not hesitate to contact us:
- GHM Messtechnik GmbH
- GHM GROUP – Greisinger
- Hans-Sachs-Str. 26
- 93128 Regenstauf | GERMANY
- Email: info@greisinger.de
- www.greisinger.de
- WEEE reg. no. DE 93889386
Calibration and adjustment service
The purpose of the calibration is to verify the precision of the measuring
device by comparing it with a traceable reference. Both ISO calibration
certificates and DAkkS calibration certificates are available from Greisinger.
Explanation
The ISO standard 9001 is applied for the iso-calibration certificates. These
certificates area affordable alternative to the DAkkS calibration certificates
andprovide information of the traceable reference, a list of individual values
and documentation. The DAkkS calibration is based on DIN EN ISO/17025, the
accreditation basis is recognised worldwide. These certificates offer high-
quality calibration and consistently high quality. The DAkkS calibration
includes any necessary adjustment with the purpose of minimising a deviation
of the measuring device.
NOTE
The device is delivered with a test report. This confirms that the measuring
device has been adjusted and tested, without making any statement about the
accuracy of a temperature sensor.
NOTE
Only the manufacturer can check the basic settings and make corrections if
necessary
Accessories
Spare parts:
GB-AA-2 : art. no. 479249 Spare batteries AA (2 pcs.)
Suitable sensors:
GTE 130 OK | art. no. 601483 | Insertion probe without cable, Ø1,5 mm |
---|---|---|
GF 1TK-T3 | art. no. 609695 | Compact immersion probe with silicone |
handle, Ø3 mm
GF 1TK-E3| art. no. 609697| Compact insertion probe with silicone
handle, Ø3 mm
GF 1TK-E1.5| art. no. 609699| Compact insertion probe with silicone
handle, Ø1,5 mm
GF 1TK-L3| art. no. 611299| Compact air probe with silicone handle, Ø3
mm
GTF 300| art. no. 600072| Wire probe for measurements in fractions of a
second
GTF 40K-620| art. no. 610829| T-handle insertion probe for bulk
material, soil, compost
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>