TOOTOO MEDITECH CO LTD DTS-280 VET Dental System User Manual
- June 13, 2024
- TOOTOO MEDITECH CO LTD
Table of Contents
TOOTOO MEDITECH CO., LTD. All rights reserved.
DTS-280 VET Dental
System Operation Manual
DTS-280 VET Dental System
WE ARE COMMITTED TO PROVIDING VETERINARIANS WITH THE WONDERFUL PRODUCTS AND AFTER SALES SUPPORT
Intellectual Property Right
The intellectual property rights of this product and its instruction manual
belong to TOOTOO MEDITECH CO., LTD.(hereinafter referred to as the TOOTOO),
including , but not limited to patent, trademark, copyright etc TOOTOO
reserves final interpretation right of this instruction manual.
TOOTOO have the right to use the instruction as confidential information. Any
individual and/or organization shall not disclose the instruction of all or
part of the information by any means without TOOTOO’s written permission. Nor
shall any other person or organization be allowed to obtain all or part of the
information of this instruction manual by any means.
No individual and/or organization shall publish, modify, reproduce, issue,
rental, adapted, and translated into other languages without TOOTOO’s written
permission. Statement
TOOTOO reserves the right to modify the content of this manual without prior
notice.
TOOTOO reserves the right to change the technology without prior notice.
TOOTOO reserves the right to modify the product specifications without prior
notice.
TOOTOO does not guarantee the information in any forms, including (but not
limited to) the responsibility of proposing the implied merchantability and
suitability for a specific purpose.
TOOTOO in only the following conditions is considered to be responsible for
the safety, reliability and performance of the instruments, i.e.:
Assembly operation, expansion, adjustment, improvements and repairs carried
out by TOOTOO authorized personnel.
Relevant electrical equipment in line with national standards; The instrument
is operated according to the instruction manual.
TOOTOO is not responsible for the products’ safety, reliability or operation
status in the following conditions:
Components are disassembled, stretched or debugged; Non TOOTOO authorized
personnel repairs or alterations to the instruments; Product may not in
accordance with the manual.
Introduction
Sincerely thanks for selecting the Veterinary Dental System manufactured by
TOOTOO.
TOOTOO Dental Systems are designed with unique and practical features to make
your work even more satisfying and productive.
Video lnstructions
As for how to assemble and how to maintain TOOTOO dental system. All
instructions are conveniently delivered to you by video instruction step by
step.
Control Top
The top is a non-porous work surface which is easy to sanitize. lt provides s
a detachable stainless plate with magnetism around the edge to supply a safe
area to store the tools and prevent burs falling on the floor during
procedures.
The main body is made from powder coated metal and aluminum that won’t warp or
deteriorate with age.
Silent Oil-Free Air Compressor
TOOTOO takes an Oil free compressor – low maintenance, oil free air. Mounts
underneath the unit into the body, with coolant system and noise reduction
system, supplying air wherever the unit is to be used. lt is more convenient
than a noisy compressor which is usually situated in another room and requires
a length of air line.
Color Coded System
All supply lines on TOOTOO dental systems are color coded to simplify service
and maintenance requirements.
Reliable Air Lines and Pipeline
All inter air lines are bundled with brass air tip in case loosen after lang
time using. Outer air lines are locked with stainless connectors. And using
the rubber pipeline instead of PCV pipeline to against the aging and crack.
Disinfectant System
This system is designed for occupational health and patient safety. lt is
designed to help reduce the risk of atmospheric pollution and the transfer of
viruses, such as FIV from patient to patient and reduce unpleasant mouth
odors.
Air Particle Filters
Air particle filters are installed on TOOTOO dental systems to help reduce
particles entering the air supply and reduce the potential damage to
Handpiece. also reduce the unpleasant environmental odors.
LED Advantage
The self-generated LED provides daylight-quality, neutrally-coloured white
light at 22,000 Lux. LED’s offer a robust, lang lasting Light source that
barely heat up, are shock resistant and that do not emit any harmful IR and
UVA rays. SUPER dental system is designed LED advantage into high speed
handpiece and ultrasonic scaler.
Please read this instruction manual and all other auxiliary materials (video)
carefully before installing and using the product, it will be helpful to work
with it better.
1.1 Application
Veterinary dental system is a dental platform that integrates polishing,
repairing, cutting and auxiliary root canal treatment. lt is used for dental
beauty, tooth removal, oral tumor cutting, etc., providing a convenient and
favorable operating environment for animal dentists.
1.2 Note
TOOTOO dental system should be operated by a qualified person – for Veterinary
use only
1.4 Performance parameter
1.5 Environment requirements for equipment
Getting Started
-
When you take the machine away from the bottom plate, do not take the Control Top, but take the Body, in case damage the Control Top.
-
Loosen the four castor wheels under the body to move machine or lock them to fix it unmovable.
-
Unscrew the coolant distilled water supply clear bottle, then push out the accessory box, which includes all handpieces and pedal inside. Take out of all accessories, assembly and put them in holder correspondingly, and place the pedal on the floor in front of machine.
-
Fill the clear plastic water bottle with distilled water, not too full, and screw the bottle into the hole with pipe under the Control Top, tight the bottle. Fill the red plastic water bottle with disinfectant liquid, dilute with distilled water, and screw it on another side. (Refer to the video: How to Assemble DTS)
-
Place the 3 way air/water syringe in its handpiece holder.
-
Assemble the handpieces according to the manufacturers instructions. Screw the gray handpiece air/water line onto the handpiece and place in correctly labeled handpiece holder on the dental system.
-
Connect the Hast to power supply with the supplied power cord.
-
Turn on the power and the compressor start to pressurize. lt needs 1 minutes to get ready on working pressure at the first time by using.
-
TOOTOO factory sets the working pressure. You can forget it. But you still can adjust individual handpiece pressure by corresponding knob easily.
(Refer to the video: The Control Panel lnstruction).
lf adjustment is required, make sure your compressor is set to deliver the correct working pressure. When you depress the pedal, the air pressure regulator gauge should be in Green area (Yellow area means too low pressure and Red area means too high). -
Check handpiece and ultrasonic scaler operation:
a) Remove High Speed Handpiece from holder and depress foot control pedal, then handpiece should now operate.
b) Replace the High Speed handpiece and remove the Low Speed Handpiece from holder and depress foot control pedal, then handpiece should now operate.
c) Replace the Low Speed handpiece and remove the Ultrasonic Scaler Handpiece from holder and depress foot control pedal, then handpiece should now operate.
NOTE: The DTS is fitted with automatic handpiece holders. Only one handpiece (HS or LS) can be used at any one time. Make sure that put the HS (LS) handpiece back to the holder before picking up LS (HS) handpiece. -
Check water and air supply to the 3 way air water syringe. Depress the right hand button for air and left hand button for water or both together for mist.
-
Pressurize water system by moving the coolant water system switch to ON (water drop icon), located on the control panel.
NOTE: When filling distilled water/ disinfectant bottles de-pressurize the water system, Moving the system switch to OFF (water drop/cup icon with slash) -
Water volume. Remove HS/LS/ Ultrasonic Scaler Handpiece from holder and depress foot control pedal. Adjust water spray volume by flow control knob, and then, if required, to get a fine mist by rotating vaporization control knobs.
-
Adjust Ultrasonic Scaler Handpiece by the white color knob, located on the last of control panel.
-
HS and LS handpieces should be lubricated regularly to get optimal performance and life. Recommend lubricated before using each time. The foot pedal should be depressed for 15 seconds after oiling to ensure the lubricant will reach the turbines and working portions of the handpiece.
NOTE: The lubricant is placed in the smaller of the 2 large holes. “Air entry hole” at the base of the handpiece. Using before and after using for 30 minutes, please drop with lubricant.
Product Structure and Guide
Front
-
3way air/water syringe
-
High speed handpiece
-
High speed handpiece (Straight type or Contra-angle type)
-
Ultrasonic scaler handpiece
-
Handpiece pressure gauge
-
Water ON/OFF toggle switch
-
Disinfectant ON/OFF toggle switch
-
Water volume adjustment knob
-
Mist volume adjustment knob
-
HS pressure/speed adjustment knob
-
LS pressure/speed adjustment knob
-
Ultrasonic scaler power adjustment knob
-
Control Top – Detachable stainless plate with magnetism
-
Clear bettle for distilled water
-
Red bettle for disinfectant
-
Control pedal
-
Body – Air compressor group
-
Caster
Back -
Air filter drain valve
-
Compressor tank drain valve
-
Host power ON/OFF button
-
Host power input port
3.1 Coolant Disinfectant system and Filtration
TOOTOO dental system is supplied with a clear and a red water bettle. The
clear bottle should be filled only with distilled water to assist in
preventing mineral or bacteria build up in the handpieces. This water bettle
supplies the handpieces, 3 way air water syringe and ultrasonic scaler
coolant.
The coolant system for your dental system, has been preset. You can turn on
and off it by toggle switch to get the coolant water supply.
TOOTOO water system allows you to choose between distilled water and
disinfectant liquid by toggle switch, depending on the dental procedure you
wish to perform. The chlorhexidine is light sensitive and should be kept in
the red bettle provided.
Refilling Coolant Bottles
To refill the distilled water or disinfectant liquid, toggle the pressure
switch on your dental system to OFF – depressurizing the system. After the
system has been depressurized, unscrew the bottle and refill with either
distilled water or disinfectant liquid. After filling the coolant bottle screw
it back into the holder and toggle the pressure switch to ON, repressurizing
the system.
DO NOT OVERTIGHTEN THE BOTTLES.
Please Note: When the day’s dental procedures are complete, toggle the
pressure ON/OFF switch to the OFF position, depressurizing the coolant bottles
NOTE: Failure to use distilled water will avoid the machine warranty, which
will result in crystat deposits forming, damage handpieces gear/ motor and
block valves.
Filtration System
The use of distilled water is mandatory in the water bottles, as local water
may clog the system and can corrode the brass valves in the system, apart from
the risk of a bacterial build up.
Chlorine in tap water can cause flocculation, clogging filters and valves. lf
you have no choice to use tap water. You need buy lnline filter, and fix it on
the pipeline in the supply bottles. This kind of filtration help prevent
particles from entering the system and blocking handpieces. This kind of
filter is located in the distilled water and disinfectant bottles.
Attention: This kind of inline filter need to be checked and replaced if they
appear dirty.
Cleaning
Bottles may be sterilized with ethylene oxide or by chemical sterilization
3.2 3 Way Air/Water Syringe
Your choice of air, water, or fine mist spray. Depress the right-hand button
for air (with air injection icon) and the left-hand button for water (with
water drop icon) and both for a fine mist.
User could select water or disinfectant by the water/disinfectant toggle
switch.
The 3 way syringe is ideal for flushing away debris du ring prophy or drying
areas during endodontic procedures.
- Out ring housing
- Air
- Water
Tip Removal
The tip of the 3 way syringe may be removed for cleaning and autoclaving.
Simply depress the outer ring housing and remove the tip. (See diagram as
above).
3.3 High Speed Handpiece
The High Speed Handpiece is designed to operate well at pressure gauge in
Green area (Yellow means not enough, Red means too high pressure).
The handpiece is designed at speeds of up to 400,000 rpm. A very light touch
is necessary as handpieces operates with very low torque and high speed.
Pressure on the handpiece of 30gms (1ounce) or more, will cause the bur to
stall.
Distilled water can be ejected by HS handpiece. Fine mist is easily adjusted
by knob on control panel.
Note: All handpieces should be lubricated prior to use.
Please refer to ( 2-Getting Start No.15)
Bur Replacement
- Depress the head back of handpiece with your thumb and pull out the bur.
- Insert the bur into the hole all the way until it stops, and then depress the head back of handpiece with your thumb, at the same time, depress the bur again until it stops.
- To change the Bur repeat step 1. and step 2.
To increase the life of the Handpiece, Oil between each use and use a new Bur between each patient. See Handpiece manual.
3.4 Low Speed Handpiece
The low speed handpiece kinds include Straight type without coolant and
Contra angel type with water pipeline.
Both of them can adjust the air pressure and can adjust the speed and rotation
direction by the knob located on handpiece itself. The straight type reducing
head is designed for polishing. The speed of the polishing cup should be
maintained at less than 5,000 r.p.m. The reducing attachment therefore allows
the motor speed and torque tobe maintained while reducing the polishing speed,
not generate excessive heat. But the Contra-angel type can get more position
and application than Straight type, also it can coolant by distilled water
from pipeline. Please refer to the video to understand how to install the
polishing head (lock and unlock), adjust speed and direction.
3.5 Ultrasonic Scaler Handpiece
Find the red dot to match and assemble the ultrasonic scaler handpiece.
Using the plastic wrench to lock and unlock the tips (refer to the handpiece
manual).
it allows users to set power output at their desired level of operation by
white knob which is located on control panel.
Please refer to the video to understand how to install the tip and adjust
power.
The ultrasonic scaler with LED light technology has been designed for use in
the areas of scaling, periodontology, endodontics, and micro-preparation.
Lightweight handpiece with pedal on/ off control.
3.7 Silent Oil-Free Air Compressor
The compressor is perhaps the most crucial part of any dental system as it
supplies the valuable air needed to run handpieces and pressurize the water
and cooling system. factory has set ready, you can forget any setting, very
simple to use. Just attention: Connect The Host to The Appropriate Voltage.
High voltage used on this equipment. lsolate from mains before removing any
covers, which should only be carried out by a qualified electrician. Ensure
power outlet is adequately rated for this unit. lsolate power supply and
relieve air pressure prior to service tasks. This appliance is not intended
for use by persons (including children) with reduced physical, sensory or
mental capabilities, or lack of experience and knowledge, unless they have
been given supervision or instruction concerning use of the appliance by a
person responsible for their safety.
System saf ety
Please read safety instructions carefully. For the sake of safety, please be
sure to be aware of the followings:
4.1 Attentions
-
Correct cable connection
Ensure the correct voltage and securely grounding. -
Avoid exposure of all electronic cables
Do not touch any electronic devices and circuits inside the device -
Please stop the device in case of suspected failure lf there is potential danger in the device or it cannot be operated properly, contact authorized technical personnel.
-
Correct device (all handpiece) connection Connect device correctly to avoid difficulties in operation due to wrong connection, resulting in disconnection.
4.2 Precautions
- Fasten power cord to prevent poor power contact.
- Pay attention to shock-, water-, humidity-, pressure-, and fire proof.
- Pay attention to handling strength and position during moving and handling, to prevent damage or dump the instrument.
- Untrained personnel are prohibited from operating the equipment. Please use a reliable grounding device.
- Users shall undertake corresponding duties due to any improper cleaning, maintenance and operation.
- In case that the equipment host is disassembled without permission of TOOTOO, it shall no langer be guaranteed by the Company with technical maintenance service. Please contact authorized personnel or TOOTOO for support in case of any technical roblems.
Maintenance
5.1 Handpiece Maintenance & Sterilization
All handpieces have a working life but it can be extended by maintaining with
the correct operating pressures and speeds as well as adequate lubrication.
The working life of most handpieces turbines is 12-24 months, however, it will
depend upon the degree of care and the frequency of use.
Note: Make sure that the turbine cartridge and head cavity are cleaned
thoroughly before autoclaving.
lf autoclaved with bloody particles in the turbine housing, autoclaving will
cause particles to be scorched and damage to the turbine. Prior to
autoclaving, lubricate items.
Note: Do not autoclave the handpiece with a bur.
Do not exceed temperatures of 275 ° F or 135 °
Lubricant after cooling down, do not allow hand- piece to sit in sealing bag
overnight or for a lang period.
5.2 Air Compressor Maintenance
Turn off the power before Daily Maintenance
- Drain moisture for the air filter (open the valve No.19).
- Drain the moisture from the compressor tank (open the valve No.20) .
Drain all water and empty the gas every time.
Yearly Maintenance:
lt is recommended to replace the inlet filters and pressure regulator filters
every year, please contact the after-sale service personnel for dealing with
it.
Troubleshooting
The chapter describes problems (faults) and suggested measures that are common
in using the product.
6.1 High Speed Handpiece
6.2 Low Speed Handpiece
6.3 3Way Air/W ater Syringe
6.4 Air Compressor
Check the supply voltage ang grounding.
Check whether the power cord in socket is well plugged.
Contact the after-sales personnel.
Warranty
The warranty of this equipment starts from the date of leaving the factory.
During the warranty period, the equipment cannot be used normally due to
problems such as materials and process defects. SUPER undertakes after sales
service such as equipment maintenance and parts replacement.
Any damage caused by improper use or over-range use is not covered by the
warranty. lf repair or replacement of parts is required, the cost will be
borne by the user.
lf the reworked equipment was found to have been unauthorized disassembly,
TOOTOO will not provide after-sales service such as warranty, free maintenance
and parts replacement.
The warranty statement (including its restrictions) is exclusively issued by
TOOTOO and covers all other warranties.
lntellectual Property Right
The intellectual property rights of this product and its instruction manual
belong to TOOTOO MEDITECH CO., LTD.(hereinafter referred to as the TOOTOO),
including , but not limited to patent , trademark , copyright etc TOOTOO
reserves final interpretation right of this instruction manual.
TOOTOO have the right to use the instruction as confidential information. Any
individual and/or organization shall not disclose the instruction of all or
part of the information by any means without TOOTOO’s written permission. Nor
shall any other person or organization be allowed to obtain all or part of the
information of this instruction manual by any means.
No individual and/or organization shall publish, modify, reproduce, issue,
rental, adapted, and translated into other languages without TOOTOO’s written
permission.
Statement
TOOTOO reserves the right to modify the content of this manual without prior
notice.
TOOTOO reserves the right to change the technology without prior notice.
TOOTOO reserves the right to modify the product specifications without prior
notice.
TOOTOO does not guarantee the information in any forms, including (but not
limited to) the responsibility of proposing the implied merchantability and
suitability for a specific purpose.
TOOTOO in only the following conditions is considered to be responsible for
the safety, reliability and performance of the instruments, i.e.:
Assembly operation, expansion, adjustment, improvements and repairs carried
out by TOOTOO authorized personnel.
Relevant electrical equipment in line with national standards; The instrument
is operated according to the instruction manual.
TOOTOO is not responsible for the products’ safety, reliability or operation
status in the following conditions:
Components are disassembled, stretched or debugged;
Non TOOTOO authorized personnel repairs or alterations to the instruments;
Product may not in accordance with the manual.
TOOTOO MEDITECH CO., LTD.
Add: RM607, HuaTe BLDG, LongGang District, LongGang, Shenzhen, China
Web: www.newtootoo.com
Tel:0086-755-28306385
Email: info@newtootoo.com
Service Email: service@newtootoo.com