micronor sensors FOTEMP1-OEM-MNT-4 OEM 1-Channel Fiber Optic Signal Conditioner PCB Module Instruction Manual
- June 13, 2024
- micronor sensors
Table of Contents
micronor sensors FOTEMP1-OEM-MNT-4 OEM 1-Channel Fiber Optic Signal
Conditioner PCB Module
Product Information
Product Description
The FOTEMP1-OEM-MNT is a PCB-Based Signal Conditioner/Controller.
Fields of Applications
The FOTEMP1-OEM-MNT can be used in various applications.
Features
- 1-Channel Controller
- USB/RS232/SPI interface
- No Analog Output (FOTEMP1-OEM-MNT)
- 4-20mA Output (FOTEMP1-OEM-MNT-4)
Ordering Codes
Configuration | Description |
---|---|
FOTEMP1-OEM-MNT | 1-Channel Controller, USB/RS232/SPI, No Analog Output |
FOTEMP1-OEM-MNT-4 | 1-Channel Controller, USB/RS232/SPI, 4-20mA Output |
Specifications
Parameter | Specification |
---|---|
Optical Interface | ST plug, Single input compatible with all TS Series fiber |
optic
temperature probes
Fiber| 200/220, 0.22NA Step Index Multimode Fiber
Measuring Range| Internal Accuracy
Internal Sampling Rate| 2 Hz
Internal Resolution| SPI Communication
SPI Communication| SPI, Slave, Output Only, 16-bit packet (TI TMP123 format)
Protocol| SPI Clock
Power Required| 12 VDC, 350 mA
Electrical Interface| Power Connector (2-pin Molex): +V, GND DB9 (Female): SPI
(Ground, Clock, Serial Out, Chip Select); RS232 (RxD and TxD) USB
via Type B receptacle 4-20mA (only available on FOTEMP1-OEM-MNT-4,
2pin Molex))
Humidity| 0-95%, Non-condensing
Ingress Protection| IP00
Dimensions| 100mm L x 80mm W x 34mm H
OPTOCON SPI Protocol
The FOTEMP1-OEM-MNT offers various interface options. The SPI interface is
specifically designed to be compatible with the Texas Instruments TMP123
temperature sensor chip.
Temperature Conversion
Temperature (Degree C) | Digital Code |
---|---|
150 C | 0x4B00 |
125 C | 0x3E80 |
25 C | 0x0C80 |
0.0625 C | 0x0008 |
0 C | 0x0000 |
-0.625 C | 0xFFF8 |
-25 C | 0xF380 |
-55 C | 0xE480 |
No Sensor Connected | 0x9600 |
Notice of Proprietary Rights
The design concepts and engineering details embodied in this manual, which are
the property of MICRONOR SENSORS INC. and WEIDMANN TECHNOLOGIES DEUTSCHLAND
GMBH, are to be maintained in strict confidence; no element or detail of this
manual is to be spuriously used, nor disclosed, without the express written
permission of MICRONOR SENORS INC and WEIDMANN TECHNOLOGIES DEUTSCHLAND GMBH.
All rights are reserved. No part of this publication may be reproduced, stored
in a retrieval system, or transmitted in any form or by any means, electronic,
mechanical, photocopying, recording, or otherwise, without prior written
permission from MICRONOR SENSORS INC and WEIDMANN TECHNOLOGIES DEUTSCHLAND
GMBH.
© COPYRIGHT 2016-2023, MICRONOR SENSORS INC.
VENTURA, CALIFORNIA
UNITED STATES OF AMERICA
Revision History
Revision | Date | Notes |
---|---|---|
A | 12/20/2016 | Original Release |
A1 | 10/17/2017 | Added FOTEMP1-OEM-MNT-4 (with 4-20mA output option) |
A2 | 3/21/2018 | Corrected polarity of 4-20mA output connections |
A3 | 7/11/2023 | Updated for Micronor Sensors Inc. |
Product Description
FOTEMP1-OEM-MNT
Figure 1. FOTEMP1-OEM-MNT PCB-Based Signal Conditioner/Controller.
The FOTEMP-OEM-MNT is a small form factor PCB-sandwich module designed for low
cost insertion of high precision GaAs-based fiber optic temperature sensor
technology into OEM applications where common electrical temperature sensors
and thermocouples cannot be used.
This OEM product is compatible with all TS series fiber optic temperature
probes. The sensors are entirely non-metallic and electrically passive,
providing immunity to EMI, RFI, NMR, high voltage, microwaves and radiation
The FOTEMP1-OEM-MNT PCB signal conditioner incorporates three digital
interface; USB, RS232, and SPI. USB interface is Type B jack while RS232 and
SPI share electrical interface connections via a DB9 connector.
Fields of Applications
- EMI, RFI, and Microwave Environments
- High Voltage Environments
- Harsh and Hazardous Environments
- Nuclear Environments
- Aerospace Applications
- Process Monitoring
- Medical Applications (MR)
Features
- Designed for OEM Integration
- Very Small Size, Small Footprint
- Modular Electronics Support Various Device Configurations
- High Accuracy
- Built in USB, RS232, and SPI Interface
- Attractive OEM Volume Discounts
Ordering Codes
Standard configurations typically available from stock:
- FOTEMP1-OEM-MNT 1-Channel Controller, USB/RS232/SPI, No Analog Output
- FOTEMP1-OEM-MNT -4 1-Channel Controller, USB/RS232/SPI, 4-20mA Output
Contact Micronor sales concerning other configurations.
Specifications
Table 1: FOTEMP-OEM-MNT Specifications
Parameter | Specification |
---|---|
Optical Interface | ST plug, Single input compatible with all TS Series fiber |
optic
temperature probes
Fiber| 200/220, 0.22NA Step Index Multimode Fiber
Measuring Range| Standard Calibration: -40°C to +300°C
Extended Calibration: -200°C to +300°C
Internal Accuracy| +/- 0.2°C
Internal Sampling Rate| 2 Hz
Internal Resolution| 0.1°C
SPI Communication
Protocol
| SPI, Slave, Output Only, 16-bit packet (TI TMP123 format)
SPI Clock| 1 MHz max OR 10 MHz max with 1µs delay between bytes
Power Required| 12 VDC, 350 mA
Electrical Interface| Power Connector (2-pin Molex): +V, GND
DB9 (Female): SPI (Ground, Clock, Serial Out, Chip Select); RS232 (RxD and
TxD)
USB via Type B receptacle
4-20mA (only available on FOTEMP1-OEM-MNT-4, 2pin Molex))
Operating Temperature| Operating 0°C to +50°C, Storage -20°C to +70°C
Humidity| 0-95%, Non-condensing
Ingress Protection| IP00
Dimensions| 100mm L x 80mm W x 34mm H
Figure 2: Various Views of Signal Conditioner
OPTOCON SPI Protocol
The OPTOCON FOTEMP-OEM-MNT offers various interface options.
The SPI interface is specifically designed to be compatible with the Texas
Instruments TMP123 temperature sensor chip.
The temperature read is started by pulling /SS low and clocking out a 16-bit
word with the high byte read first. Only the high order 13 bits are relevant
to the temperature reading. The 13 bits are standard two’s complement for a
signed integer. The most significant bit is the sign while the following 12
bits represents temperature in 1/16°C increments. The lowest three bits are
not used. See Table 2 for binary output format of the temperature reading.
The following limitations apply:
- Maximum read out clock speed is 1MHz, or 10MHz with a 1µs “inter byte spacing” clock delay.
- The value of the three least significant bits are always 0.
Table 2: Temperature Conversion
Temperature (Degree C) | Digital Code |
---|---|
150 C | 0x4B00 |
125 C | 0x3E80 |
25 C | 0x0C80 |
0.0625 C | 0x0008 |
0 C | 0x0000 |
-0.625 C | 0xFFF8 |
-25 C | 0xF380 |
-55 C | 0xE480 |
No Sensor Connected | 0x9600 |
Figures 3 and 4 illustrate the “data gap” or “inter byte spacing”. The time between the first and second data byte must be 1µs and be guaranteed > 750ns.
Timing Diagrams
Digital Input/Output Signal Requirements
Table 3: Digital Input/Output Electrical Characteristics
Signal | Condition | Min | Max |
---|---|---|---|
Input Logic Levels |
VCC = 2.7 – 3.6V
|
VIH| 2| —
VIL| —| 0.8V
Output Logic Levels|
VCC = 2.7 – 3.6V
| |
VOL (IOL=2mA)| —| 0.4V
VOH (IOH=-2mA)| 2.4V| —
SPI Clock Frequency| | —| 1 MHz or 10 MHz with 1µs Spacing
Connections
Application Specific Connections
Figure 6: Location of DB9 RS232/SPI and J16 Power Connector
Function: SPI and RS232
Connector: DB9, Female with 9x Socket Contacts
Mating Connector: DB9 Male Plug with 9x Pin Contacts
IMPORTANT NOTE:
The RS232 signals uses 12V signals. A special cable must be used that does not
connect pins 6-9 as the normal RS232 12V handshake signals appearing on these
pins may damage the SPI 5V signals.
Table 4: D89 Signal Connections
Pin | Signal | Notes |
---|---|---|
1 | Not connected | |
2 | RxD (RS232) | Data Out, RS232, DCE, 12V |
3 | TxD (RS232) | Data In, RS232, DCE, 12V |
4 | Not connected | |
5 | GND | GND |
6 | MOSI (SPI) | L2C, SPI, 5V |
7 | /SS (SPI) | L1C, SPI, 5V |
8 | SCK (SPI) | L4C, SPI, 5V |
9 | MISO (SPI) | L3C, SPI, 5V |
Function: Power, J15
Connector: 2C Pin Header,
Mating Connector: 2C, Socket Plug
Function: 4-20mA Analog Output, J_MA
Connector: 2C Pin Header,
Mating Connector: 2C, Socket Plug
Operation
Contents
Figure 10: Key Contents of the FOTEMP Temperature System
The FOTEMP1-OEM-MNT system contains the following:
- FOTEMP1-OEM-MNT Signal Conditioner
- AC Power Supply (Upon request for engineering evaluation applications only)
- CDROM with FOTEMP-ASSISTANT Software, USB Drivers and other support documentation (not shown)
Available Separately:
- TS Series Temperature Probes
SPI Test Connections
The system was connected and tested as shown in Figure 10. The Micronor SPI
Master Emulator displays the both the TMP123-type binary reading and
temperature in °C. From the FOTEMP signal conditioner.
Steps
- Connect TS2 temperature probe to ST optical interface
- Connect 12 V Power to J15 pin header
- Connect SPI interface using DB9 Male Connector With Pin Contacts
Temperature Calibration Using FOTEMP-ASSISTANT Software
The FOTEMP1 will need to be calibrated with the actual temperature sensor
used. This is achieved using a PC running the supplied FOTEMP-ASSISTANT
graphical display/data logging software and connecting to the FOTEMP1 signal
conditioner either via RS232 or USB interface.
For more information regarding calibration using FOTEMP-Assistant please refer
to the FOTEMP-ASSITANT2 Manual.
FOTEMP-Assistant Software Installation
The FOTEMP-ASSISTANT V2 software is provided on the supplied CDROM. It will
work with either USB or RS232 Interface
For more information regarding the installation of FOTEMP-Assistant please
refer to the FOTEMP-ASSITANT2 Manual.
For RS232 operation
- Connect the FOTEMP1 signal conditioner to the PC using Specially Wired RS232 cable. This special cable has Pins 6-9 disconnected so that the 12V RS232 signals do not damage the 5V SPI signals wired to these DB9 pins.
- Click on the Program File and then [PortSettings]. Be sure that COM1 is selected.
- Select [OK] and then [Connect] to start the program. After a few seconds, the program should start and display temperature.
For USB operation
- Connect the FOTEMP1 signal conditioner to the PC using supplied USB cable. If Windows tells you that the device has been properly installed, then continue to Step 3.
- Disconnect the USB cable. Windows was not able to load the correct Virtual Comm Port (VCP) drivers for the FTDI serial interface chip. You will need to install VCP drivers (either 32-bit or 64-bit depending on your system) supplied on the CD. Then repeat Step 1.
- Click on the Program File and then [PortSettings]. Select the COMx port that corresponds to VCP assigned to the FOTEMP unit.
- Select [OK] and then [Connect] to start the program. After a few seconds, the program should start and display temperature.
- If the program will not start, then contact Micronor for assistance.
Note:
If developing custom communications software, contact Micronor for ASCII
Protocol. Sample National Instruments LabVIEW™ Driver is also available.
Warranty Information
Warranty
MICRONOR SENSORS INC. warrants this product to be free from defects in
material and workmanship for a period of 1 (one) year from date of shipment.
During the warranty period we will, at our option, either repair or replace
any product that proves to be defective.
To exercise this warranty, write or call MICRONOR SENORS INC. You will be
given prompt assistance and return instructions. Send the instrument,
transportation prepaid, to the indicated service facility. Repairs will be
made and the instrument returned transportation prepaid. Repaired products are
warranted for the balance of the original warranty period, or at least 90
days.
Limitations of Warranty
This warranty does not apply to defects resulting from unauthorized
modification or misuse of any product or part. This warranty also does not
apply to Fiber Optic Connector interfaces, fuses or AC line cords. This
warranty is in lieu of all other warranties, expressed or implied, including
any implied warranty of merchantability of fitness for a particular use.
MICRONOR SENSORS INC. shall not be liable for any indirect, special or
consequent damages.
Contact Information:
Micronor Sensors Inc.
Phone +1-805-389-6600
2085 Sperry Ave, Suite A-1
Ventura, CA 93003 Email sales@micronor.com
USA URL www.micronor.com
For Europe:
Weidman Technologies Deutschland GmbH
Phone +49-351-843-5990
Washingtonstrasse 16/16a
Fax +49-351-843-5991111
D-01139 Dresden Email info@optocon.de
GERMANY URL
www.optocon.de
Mechanical Reference Drawing
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>