Introduction/overview

The PSDS-HSBK allows simple display of strain bridge based measurements such as load cells and pressure gauges with sensitivity up to +/-480 mV/V.
Up to six calibration ranges are available allowing for different loading modes (tension and compression) or different sensors. Each calibration range will remember settings that contribute to the user experience such as selected units and tare values.
TEDS devices using templates 33, 40 and 41 can be connected and will update the viewed calibration. The last twenty TEDS devices connected will be remembered and recognized when connected again reverting to the last user experience settings for that device. TEDS can be disabled and the internal calibration ranges become available again.
There are also up to six display modes available. These determine what is visible on the LCD display and what actions are available from the set of three soft keys.
Full configuration is available with a free PC based toolkit. Some simple configuration such as two point calibration is available from within the handheld using the menu system.

Getting started

This document is designed to give a very quick overview of the PSDS-HSBK and its general, standard functionality. For more details, see the full manual.

Hardware overview

mantracourt PSDS HSBK Portable Sensor Display - Frontmantracourt PSDS HSBK Portable Sensor Display -
Backmantracourt PSDS HSBK Portable Sensor
Display - Top

The Load Cell connector fitted to PSD Standard Product is BINDER 770-8. The cable fitted mating connector is a BINDER 771 8-way male connector. There are three versions available with different field cable diameters.

Cable pair Suggested Colour Connector Pin Function
1 White 1 Loadcell Reference +ve
Black 2 Loadcell Reference -ve
2 Green 3 Loadcell Signal +ve
Black 4 Loadcell Signal -ve
3 Red 5 Loadcell Excitation + ve
Black 6 Loadcell Excitation – ve
4 Blue 7 TEDS
Black 8 Ground
Screen Grey Cable screen should only be connected to chassis of the

sensor.
If this cannot be achieved, then it should be connected to Excitation -ve.

Setting up without the Toolkit

Quick, limited setup is possible using just the handheld and no toolkit. To access all the configuration features, you will need to use the toolkit.

Calibration

Please note that this must be done in the correct order for the calibration to work correctly. Changing the Sensitivity after inputting mV/V readings will cause the calibration to be void. Always make sure you finish with Apply mantracourt PSDS HSBK Portable Sensor Display -
Calibrationmantracourt PSDS HSBK Portable
Sensor Display - Calibration1mantracourt
PSDS HSBK Portable Sensor Display - Calibration5

Change units

In all standard display modes, scrolls through the available units for the selected calibration.

Change decimal places

In all standard display modes, a 2 second press shifts the decimal place position for the selected units.

Overload/underload alarm

Please note, the values saved for the thresholds are applied in the calibrated units of the currently selected range. This means that different ranges will trigger at different loads if the calibrated unit is different.

mantracourt PSDS HSBK Portable Sensor Display -
Calibration2mantracourt PSDS HSBK
Portable Sensor Display - Calibration3

Using TEDS
Plug in a TEDS enabled load cell.

mantracourt PSDS HSBK Portable Sensor Display - Using
TEDSmantracourt PSDS HSBK Portable Sensor
Display - Using TEDS 1

Toolkit

Toolkit overview

More complex setup is possible using the supplied toolkit software.
Install the toolkit onto your windows PC and connect the USB to the PSDS-HSBK.
Then open the toolkit: mantracourt PSDS HSBK Portable Sensor Display -
Toolkit

Navigation is achieved using the icons along the top of the toolkit. mantracourt PSDS HSBK Portable Sensor Display - Toolkit
1

For further information, please see the PSDS-HSBK User Manual.

Document Title: PSDS-HSBK Quick Start
Applies To: PSD (Portable Sensor Display)
Part Number: 517-952
Issue Number: 01.00
Dated: 23rd May 2022

In the interests of continued product development, Mantra court Electronics Limited reserves the right to alter product specifications without prior notice.

mantracourt PSDS HSBK Portable Sensor Display - iconwww.mantracourt.com

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

mantracourt User Manuals

Related Manuals