aqua viva SKRX Smart Rx 1.5 l-h Peristaltic Dosing Pump Instructions
- June 12, 2024
- aqua viva
Table of Contents
ORP PUMP
SKRX Smart Rx 1.5 l-h Peristaltic Dosing
Instructions
General Features
- Power Supply 220-240 Vac
- Display 4 digits with 7 segments
- Bnc connection for probe Redox/ORP
- Switch 3 position Priming, ON, Stand by
- Peristaltic pump flowrate and pressure look at label on product
- Measure Range Redox/ORP: 400÷999 mV (resolution: 10 mV)
- Dosage method ON-OFF ( Set Point adjustable)
- Calibration method Redox/ORP 1 Point (465 mV)
Switch position:
-
Priming position “the pump running” (Switch position MOM)
-
Normal position “the instrument measure Redox and dosing. (Switch position ON)
Keep pressed key for 5 seconds to run calibration menu:
Redox probe: the pump displays C465 press key and wait 60 seconds
Displays quality probe ex: C100
Press the button to stop the calibration.
Keep pressed key until flashing Setpoint value and press again to increase the value.
After 3 seconds from releasing the key the value will be saved and the pump will display mem. the range of values is included between 400 and +990 mV with an increase of 10 mV.
When the real measure is less than 400 mV and higher than 999 mV the measure displayed flashes. -
Stand by position “The pump is stop” (Switch position OFF)
Keep Press key for 3 seconds to run Setup menu:
Redox setup menu:
Adjust dosage method High (“Hi”) or Low (“Lo”).
ver. 08_2019
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>