HACH TNT822 TNTplus COD Vial Test Installation Guide
- June 12, 2024
- HACH
Table of Contents
- HACH TNT822 TNTplus COD Vial Test
- Product Information
- Product Usage Instructions
- Scope and application
- Instrument-specific information
- Before Starting
- Items to collect
- Sample collection and storage
- Test procedure
- Accuracy check
- Method performance
- Consumables and replacement items
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
HACH TNT822 TNTplus COD Vial Test
Product Information
- The Oxygen Demand, Chemical is a test method used for measuring the chemical oxygen demand (COD) in water and wastewater samples. The method follows the USEPA Reactor Digestion Method and has a low range (3 to 150 mg/L COD) and a high range (20 to 1500 mg/L COD). This test requires the use of specific instruments and TNTplus vials.
- Table 1 provides instrument-specific information for the TNTplus vials, including the required adapters and light shields for each applicable instrument. It is essential to review this table to ensure compatibility and proper setup.
- Before starting the test, it is important to install the light shield in Cell Compartment #2 for DR3900, DR3800, and DR2800 instruments. Additionally, users must review the safety information and check the expiration date on the package.
- The reagent used in this test is corrosive and toxic, so protective measures should be taken to safeguard eyes and skin. In case of spills, it is necessary to immediately flush them with running water. Spilled reagent can affect test accuracy and pose hazards to skin and other materials. The reacted samples containing mercury need to be collected for safe disposal.
- To ensure accurate results, it is recommended to analyze reagent blanks as a quality check, especially at low concentrations or when reagents have been stored for extended periods. One blank per lot of reagents should be analyzed. All tests, including samples and blanks, should be run using the same lot of vials.
- Unused vials should be stored in a closed box to protect them from light exposure.
- It is important to consult the Safety Data Sheets (MSDS/SDS) for the chemicals used in this test and follow the recommended personal protective equipment guidelines. Reacted solutions should be disposed of according to local, state, and federal regulations. For disposal information on unused reagents, refer to the Safety Data Sheets. Consult the environmental, health, and safety staff at your facility or local regulatory agencies for further disposal guidance.
Product Usage Instructions
- Review the instrument-specific information in Table 1 to determine the applicable adapters and light shields for your instrument.
- For DR3900, DR3800, and DR2800 instruments, install the light shield in Cell Compartment #2 before starting the test.
- Check the safety information and expiration date on the package.
- Take necessary precautions to protect eyes and skin when handling the corrosive and toxic reagent. Be prepared to flush spills with running water.
- Collect reacted samples containing mercury for safe disposal.
- Analyze reagent blanks as a quality check, preferably one per lot of reagents, to ensure accurate results. Use the same lot of vials for all tests.
- Store unused vials in a closed box to protect them from light exposure.
- Consult the Safety Data Sheets for the chemicals used and follow recommended personal protective equipment guidelines.
- Dispose of reacted solutions according to local, state, and federal regulations. Refer to Safety Data Sheets for disposal information on unused reagents. Seek guidance from environmental, health, and safety staff at your facility or local regulatory agencies for further disposal guidance.
- Refer to the Consumables and Replacements items section on page 5 for order information.
- Collect the required quantity of samples and follow the test procedure outlined below.
Sample Collection and Storage
- Measure 100 mL of the sample in a blender and blend for 30 seconds or until homogenized. Ignore this step if the sample does not have suspended solids.
- Pour the homogenized sample into a 250mL beaker and stir slowly with a magnetic stir plate. Ignore this step if the sample does not have suspended solids.
Test Procedure
Invert a test vial several times to mix.
Note : For DR1900 instruments, follow the specific instructions mentioned
in the manual regarding selecting the TNTplus number for the test.
USEPA1 Reactor Digestion Method2 Method 8000 3 to 150 mg/L COD (LR) 20 to 1500
mg/L COD (HR) TNTplus® 821/822
Scope and application
For water, wastewater; digestion is required.
- COD ranges 3–150 mg/L and 20– 1500 mg/L COD are USEPA approved (5220 D) for wastewater analyses, Federal Register, April 21,1980, 45(78), 26811-26812
- Jirka, A.M.; Carter, M.J., Analytical Chemistry, 1975, 47(8), 1397
Instrument-specific information
Table 1 shows all of the instruments that have the program for this test. The
table also shows the adapter and light shield requirements for the applicable
instruments that can use TNTplus vials.
To use the table, select an instrument, then read across to find the
applicable information for this test.
Table 1 Instrument-specific information for TNTplus vials
Instrument | Adapters | Light shield |
---|---|---|
DR6000, DR5000 | — | — |
DR3900 | — | LZV849 |
DR3800, DR2800 | — | LZV646 |
DR1900 | 9609900 or 9609800 (A) | — |
Before Starting
- DR3900, DR3800, DR2800: Install the light shield in Cell Compartment #2 before this test is started.
- Review the safety information and the expiration date on the package.
- The recommended temperature for samples and reagents is 15–25 °C (59–77 °F).
- The recommended temperature for reagent storage is 15–25 °C (59–77 °F).
- The reagent that is used in this test is corrosive and toxic. Use protection for eyes and skin and be prepared to flush any spills with running water.
- Spilled reagent will affect test accuracy and is hazardous to skin and other materials. Be prepared to wash spills with running water.
- The reagents that are used in this test contain mercury. Collect the reacted samples for safe disposal.
- Analyze reagent blanks as a quality check for accurate results at low concentrations or when the reagents were in storage for long periods of time. For the best results, analyze one blank (at minimum) per lot of reagents. Refer to Blanks for colorimetric determination on page 3. Run all tests (the samples and the blank) with the same lot of vials. The lot number is on the container label.
- Keep unused (light sensitive) vials in a closed box.
- Use the DRB reactor with 13-mm wells for the digestion. If the reactor has 16-mm wells, put adapter sleeves into the wells.
- DR1900: Go to All Programs>LCK or TNTplus Methods>Options to select the TNTplus number for the test. Other instruments automatically select the method from the barcode on the vial.
- Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective equipment.
- Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory agencies for further disposal information.
Items to collect
Description/Quantity
- COD TNTplus® Reagent Set, LR or HR 1
- DRB200 reactor with 13-mm wells 1
- Blender, 2-speed 1
- Pipet, adjustable volume, 1.0–5.0 mL 1
- Pipet tips, for 1.0–5.0 mL pipet 1
- Test tube rack 1
Refer to Consumables and replacement items on page 5 for order information.
Sample collection and storage
- Collect samples in clean glass bottles. Use plastic bottles only if they are known to be free of organic contamination.
- Test biologically active samples as soon as possible.
- Homogenize samples that contain solids to get a representative sample.
- To preserve samples for later analysis, adjust the sample pH to less than 2 with concentrated sulfuric acid (approximately 2 mL per liter). No acid addition is necessary if the sample is tested immediately.
- Keep the preserved samples at 2–6 °C (36–43 °F) for a maximum of 28 days.
- Correct the test result for the dilution caused by the volume additions.
Test procedure
-
Set the DRB200 reactor power to on. Set the temperature to 150 °C.
-
Measure 100 mL of sample in a blender. Blend for 30 seconds or until homogenized.
If the sample does not have suspended solids, ignore this step. -
Pour the homogenized sample into a 250‑mL beaker and stir slowly with a magnetic stir plate.
If the sample does not have suspended solids, ignore this step. -
Invert a test vial several times to mix.
-
Use a pipet to add 2.0 mL of sample to the test vial
-
Hold the vial by the cap, over a sink. Invert gently several times to mix. The vial gets very hot during mixing.
-
Insert the vial in the preheated DRB200 reactor. Close the lid.
-
Keep the vial in the reactor for 2 hours.
-
When the timer expires, set the reactor power to off. Let the temperature decrease for about 20 minutes to 120 °C or less.
-
Hold the vial by the cap and invert gently several times while the vial is still hot.
-
Put the vial in a test tube rack. Let the temperature of the vial decrease to room temperature.
-
Clean the vial.
-
DR1900 only: Select program 821 (LR) or 822 (HR). Refer to Before starting on page 1
-
Insert the vial into the cell holder. DR1900 only: Push READ. Results show in mg/L COD.
Blanks for colorimetric determination
Analyze reagent blanks as a quality check for accurate results at low concentrations or when the reagents were in storage for long periods of time. For the best results, analyze one blank (at minimum) per lot of reagents. Replace the sample with deionized water in the test procedure to determine the reagent blank value. Subtract the reagent blank value from the sample results automatically with the reagent blank adjust option. Use the blank again for other measurements with the same lot of vials. For storage, keep the blanks in a dark location. Monitor the decomposition of the blanks by periodically measuring its concentration. Measure the reagent blank value when a new lot of reagent is used. To subtract the value of the blanks from a series of measurements:
- Replace the sample with deionized water in the test procedure to determine the reagent blank value. Clean the vial, then put it in the cell holder. Close the lid.
- Set the reagent blank function to on. The measured value of the reagent blank is shown.
Note : As an alternative, record or enter the reagent blank value at a different time. Push the highlighted reagent blank box and use the keypad to enter the value
Interferences
Chloride is the primary interference in this test method and results in a
positive interference. Each COD vial contains mercuric sulfate that will
eliminate chloride interference to a maximum of 2000 mg/L Cl–.
Accuracy check
Standard solution method—LR
Use the standard solution method to validate the test procedure, the reagents and the instrument.
Items to collect:
- COD Standard Solution, 1000-mg/L COD
- 100-mL volumetric flask, Class A
- 10.0-mL volumetric pipet, Class A and pipet filler safety bulb
- Deionized water
- Prepare a 100-mg/L COD standard solution as follows:
- Use a pipet to add 10.0 mL of a 1000-mg/L COD standard solution into the volumetric flask.
- Dilute to the mark with deionized water. Mix well. Prepare this solution daily.
- Use the test procedure to measure the concentration of the prepared standard solution.
- Compare the expected result to the actual result.
Note : The factory calibration can be adjusted slightly with the standard adjust option so that the instrument shows the expected value of the standard solution. The adjusted calibration is then used for all test results. This adjustment can increase the test accuracy when there are small variations in the reagents or instruments.
Standard solution method—HR
Use the standard solution method to validate the test procedure, the reagents and the instrument.
Items to collect:
COD Standard Solution, 300-mg/L, 800-mg/L or 1000-mg/L COD or Oxygen Demand
Standard (contains 617-mg/L COD) or Wastewater Influent Standard Solution,
Mixed cParameter (contains 500-mg/L COD)
- Use the test procedure to measure the concentration of the standard solution.
- Compare the expected result to the actual result.
Note : The factory calibration can be adjusted slightly with the standard adjust option so that the instrument shows the expected value of the standard solution. The adjusted calibration is then used for all test results. This adjustment can increase the test accuracy when there are small variations in the reagents or instruments
Method performance
The method performance data that follows was derived from laboratory tests that were measured on a spectrophotometer during ideal test conditions. Users can get different results under different test conditions.
Program| Standard| Precision (95% Confidence Interval)| Sensitivity Concentration change per 0.010 Abs
change
---|---|---|---
barcode (TNTplus 821, LR)| 75 mg/L COD| 72–78 mg/L COD| —
barcode (TNTplus 822, HR)| 750 mg/L COD| 736–764 mg/L COD| —
Summary of Method
The results in mg/L COD are defined as the milligrams of O2 consumed per liter of sample under the conditions of this procedure. The sample is heated for 2 hours with sulfuric acid and a strong oxidizing agent, potassium dichromate. Oxidizable organic compounds react, reducing the dichromate ion (Cr2O7 2- ) to green chromic ion (Cr3+). When the 3–150 mg/L colorimetric method is used, the amount of Cr6+ that remains is measured. When the 20–1500 mg/L colorimetric method is used, the amount of Cr3+ that is produced is measured. The COD reagent also contains silver and mercury ions. Silver is a catalyst, and mercury is used to complex chloride interferences. The measurement wavelength is 420 nm for the LR or 620 nm for the HR.
Consumables and replacement items
Required reagents
Description | Quantity/Test | Unit | Item no. |
---|---|---|---|
COD TNTplus® Reagent Set, LR, 3 to 150 mg/L COD | 1–2 vials | 25/pkg | TNT821 |
COD TNTplus® Reagent Set, HR, 20 to 1500 mg/L COD | 1–2 vials | 25/pkg | TNT822 |
Required apparatus
Description | Quantity/test | Unit | Item no. |
---|---|---|---|
Blender, 2-speed, 120 VAC option | 1 | each | 2616100 |
Blender, 2-speed, 240 VAC option | 1 | each | 2616102 |
DRB200 Reactor, 115 VAC option, 9 x 13 mm + 2 x 20 mm, 1 block | 1 | each |
DRB200-01
DRB200 Reactor, 230 VAC option, 9 x 13 mm + 2 x 20 mm, 1 block| 1| each|
DRB200-05
Pipet, adjustable volume, 1.0–5.0 mL| 1| each| BBP065
Pipet tips, for 1.0–5.0 mL pipet| 1| 75/pkg| BBP068
Light shield, DR3800, DR2800, DR2700| 1| each| LZV646
Light shield, DR3900| 1| each| LZV849
Recommended standards
Description | Unit | Item no. |
---|---|---|
COD Standard Solution, 300-mg/L | 200 mL | 1218629 |
COD Standard Solution, 800-mg/L | 200 mL | 2672629 |
COD Standard Solution, 1000-mg/L | 200 mL | 2253929 |
Oxygen Demand Standard (BOD, COD, TOC), 10-mL ampules | 16/pkg | 2833510 |
Description | Unit | Item no. |
--- | --- | --- |
Wastewater Effluent Standard Solution, Mixed Parameter, for NH3-N, NO3-N, PO 3–,
4
COD, SO 2–, TOC
4
| 500 mL| 2833249
Wastewater Influent Standard Solution, Mixed Parameter, for NH3-N, NO3-N, PO
3–,
4
COD, SO 2–, TOC
4
| 500 mL| 2833149
Optional reagents and apparatus
Description | Unit | Item no. |
---|---|---|
Flask, volumetric, Class A, 100 mL, glass | each | 1457442 |
Reactor adapter sleeves, 16 mm to 13 mm diameter, for TNTplus vials | 5/pkg |
2895805
Sampling bottle with cap, low density polyethylene, 500-mL| 12/pkg| 2087079
Sulfuric Acid, concentrated, ACS| 500 mL| 97949
Test tube rack, polyethylene, for 13-mm OD vials, 90 holes| each| 2497900
Water, deionized| 4 L| 27256
HACH COMPANY WORLD HEADQUARTERS
Telephone : 970-669-3050
FAX : 970-669-2932
© Hach Company/Hach Lange GmbH, 2007–2014, 2019. All rights reserved.
FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. – Call
toll-free 800-227-4224 Outside the
U.S.A. – Contact the HACH office or distributor serving you. On the Worldwide
Web – www.hach.com;
E-mail – techhelp@hach.com
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>