HI98120
Waterproof ORP
& Temperature Tester
INSTRUCTION MANUAL
HI98120 Waterproof ORP and Temperature Tester
Dear Customer,
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before using the tester. For
more information about Hanna Instruments and our products, visit
www.hannainst.com or e-mail us at
sales@hannainst.com.
For technical support, contact your local Hanna Instruments Office or e-mail
us at tech@hannainst.com.
Preliminary Examination
Remove the tester and accessories from the packing material and examine it
carefully. If you require any further information, please contact Hanna
Instruments technical support team at
tech@hannainst.com.
Each HI98120 is delivered in a cardboard box and is supplied with:
- HI70022 ORP test solution, 470 mV, 20 mL sachet (6 pcs.)
- HI73120 ORP electrode
- HI73128 Electrode removal tool
- 1.5V batteries (4 pcs.)
- Instrument quality certificate
- Instruction manual
Note: Save all packing material until you are sure that the tester works correctly. Any damaged or defective item must be returned in its original packing material with the supplied accessories.
General Description & Intended Use
HI98120 is a compact ORP and temperature tester. It features a two-buttons
operation system and is easy to use. The compact and waterproof casing is
designed to float if accidentally dropped in the water. The tester is factory
calibrated and measures temperature in both °C and °F. It displays a stability
tag that will disappear once the reading has stabilized.
Battery level and low-battery indicator are clearly displayed on the LCD to
alert the user in the event that low battery power could adversely affect the
readings.
BEPS (Battery Error Prevention System)
The tester will automatically shut off if there is not enough power to get an
accurate reading.
ORP Probe
HI98120 is supplied together with HI73120 ORP replaceable electrode.
Temperature Sensor
The stainless steel temperature sensor facilitates faster and more accurate
temperature measurement.
Specifications
Range| ±1000 mV
-5.0 to 60 °C/ 23.0 to 140.0 ° F
---|---
Resolution| 1 mV
0.1 °C/ 0.1 °F
Accuracy (@25 °C/77 °F)| ±2 mV
±0.5 °C/ ±1.0 °F
Calibration| Factory calibrated
Electrode| HI73120 ORP electrode (included)
Battery type| 1.5 V (4 pcs.)
Battery life| Approx. 300 hours of use
Auto-off| After 8 minutes
Environment| -5 to 50 °C (23 to 122 °F); RH 100%
Dimensions| 171 x 41 x 26 mm (6.7 x 1.6 x 1.0″)
Weight| 83 g (2.9 oz.)
Functional Description
LCD Display
- Stability tag
- Battery life percentage indicator
- Low battery indicator
- Measurement unit
- First LCD line
- Second LCD line
Operational Guide
Turning the Tester ON / OFF
Press and hold /MODE. Al LCD segments will be displayed for a few second
followed by battery life percentage indicator (e.g. % 100 BATT). To turn the
tester OFF, from measurement mode, press MODE. “OFF” will be displayed on the
second LCD line and the tester will turn off.
HOLD Mode
From measurement mode, press and hold SET/HOLD until HOLD is displayed. mV
reading will be frozen on the LCD. Press either buttons to resume measurement
mode.
Setup
Setup mode allows the selection of temperature unit.
To enter the Setup mode, press /MODE until “TEMP“ and the current temperature
unit are displayed on the second LCD line (e.g. “TEMP °C”).
Use SET/HOLD to change the temperature unit, and press /MODE to return to
measurement mode.
Measurement
Place the electrode in the solution to be tested while stirring it gently.
The measurement should be taken after the stability tag ( ) has disappeared.
The ORP (mV) value is displayed on the first LCD line and the temperature on
the second LCD line.
Note: If measurements are taken successively, rinse the probe thoroughly
in distilled or deionized water to eliminate crosscontamination. After
cleaning, rinse the probe with some of the sample to be measured.
Maintenance
Please read the information below to ensure the highest possible accuracy.
- If measurements are taken successively, rinse the probe thoroughly in distilled or deionized water to eliminate crosscontamination.
- Clean the electrode monthly using by keeping the electrode in HI7061 General purpose cleaning solution for 30 minutes and rinsing the probe thoroughly with water.
- When not in use, add a few drops of HI70300 Storage solution to the protective cap. Never store the probe in distilled or deionized water.
Electrode Replacement
Use supplied removal tool (HI73128) to replace the ORP electrode.
Insert the tool into the probe cavity. Rotate the tool counterclockwise. Pull
the electrode out by using the other side of the tool. Insert a new ORP
electrode following the above instructions in reverse order.
Battery Replacement
Battery life percentage indicator is displayed at power on. If the battery
level drops below 5%, the symbol lights up indicating that the batteries
should be replaced soon.
If the battery level is not adequate to guarantee an accurate reading, the
tester automatically turns off.
To replace the batteries, follow the next steps:
- Turn OFF the tester.
- Remove the four screws on the top of the tester to open the battery compartment (figure 1).
- Remove the old batteries.
- Insert four new 1.5V AA batteries in the battery compartment while paying attention to the correct polarity (figure 2).
- Close the battery compartment using the four screws.
Note: Only use the battery type specified in the manual. Old batteries should be disposed in accordance with local regulations.
Accessories
Electrodes
HI73120| Replaceable ORP electrode
HI73128| Electrode removal tool
ORP Test Solution
HI7021M| 240 mV @25 °C, ORP test solution, 230 mL
HI7022M| 470 mV @25 °C, ORP test solution, 230 mL
Electrode Cleaning Solution
HI7061M| General purpose cleaning solution, 230 mL
Electrode Storage Solution
HI70300M| Electrode storage solution, 230 mL
Pretreatment Solution
HI7091L| Pretreatment reducing solution
HI7092M| Pretreatment oxidizing solution, 230 mL
Other Accessories
HI740026P| Replacement 1.5 V batteries (12 pcs.)
Certification
All Hanna Instruments conform to the CE European Directives. Disposal of
Electrical & Electronic Equipment. The product should not be treated as
household waste. Instead hand it over to the appropriate collection point for
the recycling of electrical and electronic equipment which will conserve
natural resources.
Disposal of waste batteries. This product contains batteries, do not dispose
of them with other household waste. Hand them over to the appropriate
collection point for recycling. Ensuring proper product and battery disposal
prevents potential negative consequences for the environment and human health.
For more information, contact your city, your local household waste disposal
service, the place of purchase or go to
www.hannainst.com.
Recommendations for Users
Before using this tester, make sure that it is entirely suitable for your specific application and for the environment in which it is used. Avoid touching the electrode at all times. Any variation introduced by the user to the supplied equipment may degrade the tester’s performance. For yours and the tester’s safety do not use or store the tester in hazardous environments.
Warranty
HI98120 is warranted for a period of one year against defects in workmanship
and materials when used for its intended purpose and maintained according to
instructions. The electrode is warranted for a period of six months. This
warranty is limited to repair or replacement free of charge. Damage due to
accidents, misuse, tampering or lack of prescribed maintenance is not covered.
If service is required, contact your local Hanna Instruments Office. If under
warranty, report the model number, date of purchase, serial number and the
nature of the problem. If the repair is not covered by the warranty, you will
be notified of the charges incurred.
If the instrument is to be returned to Hanna Instruments Office, first obtain
a Returned Goods Authorization (RGA) number from the Technical Service
department and then send it with shipping costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete protection.
Hanna Instruments reserves the right to modify the design, construction, or
appearance of its products without advance notice.