IDION iTempShield Battery-Powered Wearable Thermometer User Guide
- June 12, 2024
- IDION
Table of Contents
IDION iTempShield Battery-Powered Wearable Thermometer
Directions For Use
Read these instructions fully before using the iTempShield™.
INDICATIONS
iTempShield is a battery-powered wearable thermometer intended for continuous
measurement of human body temperature on the upper chest via wireless
communication to a smart device application.
iTempShield is intended for single-use and for persons older than 5 years in
healthcare facilities and home environments. The patient is the intended user
of the iTempshield.
Note : iTempShield™ is not FDA approved or cleared.
CONTRAINDICATIONS
iTempShields are contraindicated for patients who exhibit allergic reactions to Silicon adhesives (tape), Silicone rubber, or Polyester.
DESCRIPTION
The iTempShield™ is a wearable device that senses body skin temperature and
communicates with an app on your smartphone or tablet to calculate your
internal body temperature. When the iTempShield™ is connected to your device,
it sends all temperature measurements automatically.
You can wear an iTempShield™ for up to 60 days and check your current
temperature and temperature measurements history at any time. You have the
ability to share the data with your clinician if your clinician has
established an IDION account.
WARNINGS, CAUTIONS, AND NOTES
- The iTempShield™ should not be used as the sole basis for diagnosis or therapy decisions. It must be used in conjunction with your physician.
- The iTempShield™ should be free of any visible defects, discoloration, or damage. Do not use and ask your physician for another iTempShield™ if any visible defects, discoloration, or damage is present.
- No heavy exercise or contact sports when using the iTempShield™ . Exercise may interfere with temperature measurement accuracy.
- Avoid extreme heat or cold for more than a few minutes. This could affect performance or accuracy.
- When taking a shower, Pat dry the iTempShield™ do not rub aggressively. If adhesive lifts up replace it.
- After shower the temperature reading may be lower for several minutes as the chest area is cooler.
- You may also remove the iTempShield™ before taking a shower dry off and re-apply.
- The iTempShield™ may take up to 20 minutes to reach stable body temperature.
- If you feel discomfort or pain from the patch or iTempShield™ , contact your doctor.
- If you see the following, replace adhesive patch right away:
- Adhesive patch’s edges lift easily or curl up more than 3 mm
- Sensor or adhesive patch have shifted from their original location
- When changing clothes, avoid snagging the iTempShield™ . Wear loose-fitting clothes when possible.
- Do not use the iTempShield™ during MRIs, X-Rays, or CT Scans as it may result in physical harm.
- Avoid contact with iTempShield™ during defibrillation.
- Do not use the iTempShield™ in the presence of flammable anesthetics or other flammable substances in combination with air, oxygen-enriched environments, or nitrous oxide to avoid risk of exposure.
- Avoid using the iTempShield™ during electrocautery.
- Avoid placing the iTempShield over compromised skin, excessive hair, implants, ports, subcutaneous or dermal fillers or scar tissue, as it may result in incorrect readings.
- Do not apply over or near pacemakers to avoid any potential interference from the Bluetooth communication.
- The iTempShield™ should not be used near electrical equipment that may affect the shield’s performance.
- Check the iTempShield™ site to ensure skin integrity and to avoid damage or irritation to the skin.
- Incorrect readings may be caused an iTempShield™ that is not placed on an appropriate application site.
- The iTempShield™ placement site should be hair-free, cleaned of debris, and dry before placement.
- The iTempShield™ may not reflect the actual body temperature when used on patients undergoing treatments that may alter their normal temperature regulation (e.g., therapeutic hypothermia, antipyretics).
- Avoid direct heating or cooling of the iTempShield™ . Localized temperature exposure of the shield may result in no or incorrect readings.
- An iTempShield™ that becomes partially dislodged may cause no or incorrect readings.
- Rapid or large changes in ambient temperature may cause no or incorrect readings.
- Periodically check the iTempShield™site for proper adhesion to minimize the risk of incorrect or no readings.
- Change or modifications that are not expressly approved by the manufacturer could void the user’s authority to operate the equipment.
- Do not modify or alter the sensor in any way. Alteration or modification may affect performance and/or accuracy.
- To prevent damage, do not soak or immerse the ™ in any liquid solution.
- Do not attempt to sterilize by irradiation, steam, autoclave, or ethylene oxide as it will damage the iTempShield™.
- Do not attempt to reprocess, recondition, or recycle the iTempShield™ .
- Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the iTempShield™ .
- Keep the iTempShield™away from electrical equipment that emits radio frequencies to minimize radio interference. Radio interference may result in no or inaccurate readings.
- The iTempShield™ does not include an adhesive on the device. Do not attempt to peel away any part of the iTempShieldTM and use the specified adhesive patches for application to the body.
- Keep the iTempShield™ protected from excessive dust.
- If in an environment with a temperature below the operational temperature range specified in the environmental section, ensure the iTempShield™ is covered by warm clothing.
- Do not use adhesive other than the one provided with the iTempShield™ unless prescribed by the physician.
- Do not swallow the iTempShield™.
- Children should be supervised while using the iTempshield.
- Keep the iTempShield™ away from pets.
Note : Loss of essential performance caused by the conditions highlighted
above can result in loss of or inaccurate readings.
Note : Changes or modifications not expressly approved by Idion could
void the user’s authority to operate the equipment.
INSTRUCTIONS
Using the IDION Shield mobile app
Install the IDION Shield app on your mobile device by scanning this QR code
with the device’s camera app.
- Enable “Allow notifications,” and select the box to agree to the Terms and Conditions.
- After viewing the brief iTempShield™ overview on the phone application, select “Enable” and then select “OK” to allow the device to connect with your phone.
- If you are an iPhone user, tap the top of your phone (camera area) to the device. If you are an
Android user, tap the middle of your phone to the device. If the connection is not made, move the device across the back of the phone and try again.
Before using the IDION Shield mobile app, make sure your mobile device:
- Has a near-field communication (NFC) feature
- Has the NFC and Bluetooth features enabled
The Shield sends temperature readings using Bluetooth. For more accurate temperature history:
- Keep Bluetooth enabled on your device at all times
- Keep your Shield and mobile device close to each other, with a maximum range of 32 feet (10 meters) and no physical obstacles (such as a wall) between them.
History Screen
To view your temperature history, tap the History tab. On the history screen
you can review the temperature measurements as a List or as a Chart.
Temperature threshold
When your body temperature exceeds the high temperature threshold value, the
reading is marked with a red dot in the temperature history, and you have
enabled communication with your clinician the data is sent to your clinician
immediately.
Temperatures below the high temperature threshold will be marked with blue
dots.
Temperature Screen (Main Screen)
The Temperature screen shows the latest temperature measurement and Shield
battery level. When the Shield first connects to the app, it sends temperature
readings every 30 seconds. The Shield takes approximately 20min to warm up to
body temperature and sends temperature readings approximately every 5 minutes.
Adding an oral temperature
To add an oral temperature reading, tap the Add button at the bottom of the
Temperature screen. Enter your oral temperature:
- In Fahrenheit,
- With two or three digits and one decimal place (for example 98.6 or 101.2)
- You can also add an activity, such as taking a shower, etc.
To save the temperature, tap the Save button . The Oral temperature will be added to the Temperature History and can be available to your Clinician.
Battery level
The Shield sensor’s battery level is shown on the Temperature screen. Each
sensor has enough battery power to last 60 days. When the sensor battery level
drops below 10%, the app will notify you.
When the low battery notification occurs, follow clinician’s instructions.
Disposal of Product: Comply with local laws in the disposal of the sensor,
battery, and its accessories.
First time Users
It is recommended that prior to applying the ™ sensor to the adhesive patch,
the user download the app and go through the screens to connect to the shield.
With Android phones the sensor is in the back of the phone near the middle.
With the iPhone it is along the top edge.
Applying the iTempShield™ to the body
Prepare chest
a) Select an area on the left or right side of your chest where you will place
Shield.
b) If needed, trim hair on the area to make sure the adhesive tape or patch
will contact the skin.
c) Clean and dry the area.
Adhere sensor to patch or tape
a) Tear a three-inch strip from the tape roll or select a patch.
b) Peel off the inside liner from the adhesive patch or if required the tape.
c) Place the sensor on the adhesive patch or tape, with the sensor’s logo
facing the patch’s sticky side.
Adhere patch or tape to chest
a) If using a patch, peel off the outside liners from the adhesive patch.
b) Adhere the patch to the chest, pressing the patch firmly to secure it to
the skin.
c) Avoid pressing against the sensor. The IDION logo is facing away from the
body.
d) Avoid stretching adhesive patch or tape when applying it to skin.
Avoid touching the adhesive surface.
How to Remove the Adhesive
a) Begin slowly peeling off the adhesive at a low angle.
b) With your other hand, press down skin under adhesive tape to stabilize
skin.
c) Continue slowly pulling tape away from skin and pressing skin down until
you fully remove adhesive tape.
d) After removing adhesive tape from skin, lift iTempShieldT™ off of adhesive
tape.
e) If adhesive residue remains on device, use alcohol wipe to remove residue.
f) Throw away used adhesive tape but keep the iTempShieldT™.
TROUBLESHOOTING GUIDE
App Message | Solution |
---|
Shield association was unsuccessful and there are characters aabbccdd11
displayed.| The iTempShield™ was improperly registered by your clinician. You
must call with that number available so the problem can be rectified.
Shield association was unsuccessful and there is just a phone number| Call the
number your shield is not working properly.
Shield association was unsuccessful due to technical error|
- Restart your mobile device.
- Make sure internet connection and Bluetooth are on and functioning and try to connect your Shield again.
- If it still fails, contact the support number, and replace the shield.
App does not connect via NFC and NFC pop-up disappears|
- Remove your phone from the case and try again to connect your Shield.
Check your connection| Make sure the mobile device and Shield are within 32 feet (10 meters) of each other and without significant obstacles between them
For additional help, contact IDION at 1-203-208-8570
Environmental
Storage/Transport Temperature | 0 degrees C to 30 degrees C |
---|---|
Operating Temperature | 10 degrees C to 43 degrees C |
Storage/Transport Humidity | 40% – 75% RH (non-condensing) |
Operating Humidity | 10% – 95% RH (non-condensing) |
Atmospheric Pressure | 700 hPa to 1060 hPa |
Caution : Using/storing iTempshield outside of the environmental ranges listed above might lead to inaccurate results or a damaged device.
Specifications
Temperature Measurement Accuracy | +/- 0.1 C in the range of 34 C to 42 C |
---|---|
Application Site | Upper Chest, below the left or right collarbone |
Product Use/Battery Life | Minimum of 60 days of continuous use |
Maximum Storage Life | 1 year |
Note: The iTempShield complies with IEC 60601-1 and IEC 80601-2-56.
WIRELESS TECHNOLOGY INFORMATION
Type | Bluetooth Low Energy | NFC |
---|---|---|
Data Transmission Rate | 1 MB/sec | 424 KB/sec |
Max. Output Power | 2.3 dBm | N/A |
Modulation | Spread spectrum | Single frequency |
Frequency Range | 2.4 to 2.483 Ghz | 13.56 MHz |
Antenna Peak Gain | 5.3 dB | N/A |
FCC ID: 2A8QA-ITEMPS, FCC Code: 2A8QA
CAUTION : In order to maintain Bluetooth connectivity with the host
device ensure that iTempShield™ is within specified distance and line of sight
of the host device and avoid stacking units or keeping the close together when
operating. Failure to do so might lead to connection issues and loss of data.
RF Radiation Exposure Statement: This equipment has been exempted from FCC RF
radiation exposure testing and IC RSS 102 RF radiation exposure limits set
forth for an uncontrolled environment.
Note : This device complies with part 15 of FCC Rules and Industry
Canada’s license-exempt RSS. Operation is subject to the following two
conditions: (1) This device may not cause harmful interference, and (2) this
device must accept any interference received, including interference that may
cause undesired operation.
Note : This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses,
and can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Consult the dealer or an experienced radio/TV technician for help.
Note : This device complies with Part 15 of the FCC Rules. Operation is
subject to the following two conditions: (1) the device may not cause harmful
interference, and (2) the device must accept any interference received,
including interference that may cause undesired operation.
Note : When using iTempShield™ consideration should be taken to local
government frequency allocations and technical parameters to minimize the
possibility of interference to/from other wireless devices.
RECOMMENDED SEPARATION DISTANCES
Recommended separation distance between portable and mobile RF communication equipment and the ME equipment
The ME Equipment is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the
ME Equipment can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment
(transmitters) and the ME Equipment as recommended below, according to the
maximum output power of the communication equipment.
Rated Maximum Output of Transmitter (W)| Separation Distance According
to Frequency of Transmitter (M)
80 MHz to 800 MHz
d = 1.17√P| 800 MHz a 2.5 GHz
d = 2.33√??
0.01| 0.12| 0.23
0.1| 0.37| 0.74
1| 0.17| 2.33
10| 3.7| 7.37
100| 11.7| 23.3
GUIDANCE AND MANUFACTURER’S DECLARATION -ELECTROMAGNETIC EMISSIONS
The ME Equipment is intended for use in the electromagnetic environment
specified below. The customer or the user of the ME Equipment should assure
that it is used in such an environment.
EMISSION TEST| COMPLIANCE| ELECTROMAGNETIC ENVIRONMENT GUIDANCE
RF Emissions
CISPR 11| Group 1| The ME Equipment must emit electromagnetic energy in order
to perform its intended function. Nearby electronic equipment may be affected.
RF Emissions
CISPR 11| Class B| Suitable for use in all establishments, including domestic
environments.
Note : The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not offer adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or reorienting the equipment.
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY
The ME Equipment is intended for use in the electromagnetic environment
specified below. The customer or the user of the ME Equipment should assure
that it is used in such an environment.
IMMUNITY TEST| IEC 60601 TEST LEVEL| COMPLIANCE LEVEL|
ELECTROMAGNETIC ENVIRONMENT- GUIDANCE
Electrostatic discharge (ESD) IEC 61000-4-2| +/- 8 kV contact
+/- 15 kV air| +/- 8 kV contact
+/- 15 kV air| Floors should be wood, concrete, or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be at least 30%.
Power frequency (50 / 60 Hz) magnetic field.
IEC 61000-4-8| 30 A/m| 30 A/m| | Guidance – Power frequency magnetic fields
should be at levels characteristic of typical location in a typical hospital
environment.|
Portable and mobile RF communications equipment should be used no closer to
any part of the ME Equipment, including cables, than the recommended
separation distance calculated from the equation applicable to the frequency
of the transmitter.
IMMUNITY TEST| IEC 60601 TEST LEVEL| COMPLIANCE LEVEL|
RECOMMENDED SEPERATION DISTANCE
| | | d = [3,5/(??1)]\√??
| | | 80 MHz to 800 MHz
| | | d = [7/(??1)]\√??
| | | 800 MHz to 2.5 GHz
Radiated RF IEC 61000-4-3| 10 V/m
80 MHz to 2.5 GHz| 10 V/m| where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey a, should be less than the compliance level in
each frequency range b.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2 : These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people.
a) Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the ME Equipment is used
exceeds the applicable RF compliance level above, the ME Equipment should be
observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the
ME Equipment.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than [V1] V/m.
TEST SPECIFICATIONS FOR ENCLOSURE PORT IMMUNITY TO RF WIRELESS
COMMUNICATION EQUIPMENT
TEST
FREQUENCY
| BAND (A)
(MHZ)
| SERVICE (A)| MOCULATION
(B)
| MAXIMUM
POWER (W)
| DISTANCE
(M)
| IMMUNITY
LEVEL (V/M)
385| 380-395| TETRA 400| Pulse modulation
(b) 18 Hz
| 1,8| 0,3| 27
450| 430-470| GMRS 460,
FRS 460
| FM (c) +/-5
kHz
| 2| 0,3| 28
| | | deviation 1
kHz sine
| | |
710| 704-787| LTE Band 13,
17
| Pulse modulation
(b) 217 Hz
| 0,2| 0,3| 9
745
780
810| 800-960| GSM 800/900, TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
| Pulse modulation (b)
18 Hz
| 2| 0,3| 28
870
930
1720| 1700-1990| GSM 1800;
CDMA 1900;
GSM 1900; DECT; LTE
Band 1, 3.4
35: UMTS
| Pulse modulation (b)
217 Hz
| 2| 0,3| 28
1845
1970
2450| 2400-2570| Bluetooth, WLAN, 802.11
b/g/n, RFID 2450, LTE
Band 7
| Pulse modulation (b)
217 Hz
| 2| 0,3| 28
5240| 5100-5800| WLAN
802.11 a/n
| Pulse modulation (b) 217 Hz| 0,2| 0,3| 9
5500
5785
Note: If necessary to achieve the IMMUNITY TEST LEVEL, the distance
between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be
reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4- 3.
(a) For some services, only the uplink frequencies are included.
(b) The carrier shall be modulated use a 50% duty cycle square wave signal.
(c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be
used because while it does not represent actual modulation, it would be worst
case.
WARRANTY
IDION warrants each new iTempShield™ patch against defects in materials or workmanship until the expiration date of the product and agrees to replace any defective product without charge. The life of the product is 60 days after activation if activated prior to expiration date.
WARRANTY EXCLUSIONS
The warranty does not cover damage resulting from accident, misuse or abuse,
or lack of reasonable care. No responsibility is assumed for any special,
incidental, or consequential damages. To obtain warranty replacement, call
IDION Customer Support at 1
203-208-8570
NO IMPLIED LICENSE
Possession of the Equipment does not convey any express or implied license. Sensors designated for single patient use only are licensed to Customer to be used on a single patient only, and Customer. After use of Sensors designated for single patient use only, there is no further license granted by IDION to use the Sensors and they must be returned to IDION.
CAUTION: FEDERAL LAW (U.S.A.) RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN.
For professional use. See instructions for use for full prescribing
information, including indications, contraindications, warnings, precautions,
and adverse events.
If you encounter any serious incident with product, please notify the
competent authority in your country and the manufacturer.
The following symbols may appear on the product or product labeling:
SYMBOL| DEFINITION| SYMBOL| DEFINITION| SYMBOL|
DEFINITION
---|---|---|---|---|---
(blue background)| Follow instructions for use| | Federal Communications
Commission (FCC) Licensing| | Bluetooth
| Consult instructions for use| | Lot code| | Identifies unit has been
registered as a radio device
| Manufacturer| | Catalogue number (model number)| | Medical device
| Date of Manufacture YYYY- MM-DD| | Storage temperature range|
IP57
| IP57 Protected from limited dust ingress. Protected from immersion between
15 centimeters and 1 meter in depth.
| Use By YYYY- MM-DD| | Storage humidity Limitation| | Unique device
identifier
| Caution| | Do not use if package is damaged and consult instructions for
use| | Non-sterile
| Type BF medical device| | Device is single use.| | Comply with local laws in
the disposal of the sensor, battery and its accessories
| U.S. Food and Drug Association| | | |
Product Information
Model: iTempshield
Model #: DWF-1824
Version #: 1
FCC ID: 2A8QA-ITEMPS
Manufacturer: Idion
Manufacturer Location: 67 North Road Chelmsford, MA 01824
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>