ASSEMBLING BATTERY
REMOVING BATTERY
Battery Hot-Swap
The detector is equipped with a hot-swappable battery. Hot-Swap is activated for one minute after the battery is removed from the detector.
NOTE
- The detector will turn off if charged battery is not inserted on time.
- Data collection and calibration functions are not available during a Hot-Swap
HOW TO USE MAIN CABLE HOLDER
- Clean the wall before attaching the holder to it.
- Attach the main cable holder using adhesive tape on the back.
- Store the main cable of the detector in the holder.
NOTE
- The strength of the adhesive may be weakened depending on the environment. Avoid attaching the holder too high on the wall.
- Ensure that the holder is firmly fixed on the wall.
- Avoid storing anything other than the main cable.
SPECIFICATION AND DIMENSION OF EACH PART
Specifications are subject to change without notice.
The symbol ~ means alternating current, and the symbol means direct current.
Specifications
Detector
NOTE
- Maximum wireless signal rate derived from IEEE standard specifications. Actual data throughput will vary. Network conditions and environmental factors, including volume of network traffic, building materials and construction, and network overhead, lower actual data throughput rate.
- Recommended Maximum operable distance: 2 m (From the Access Point)
- Wireless antennas: The module adopts the latest 802.11ac technology. The transmitter of the module is powered by host equipment (Detector). The antennas are 2 printed-dipole antennas.
- Wireless module: 802.11 a/b/g/n/ac USB2.0 module is implemented. It supports 2T2R (2 transmit 2 receive) MIMO technology, which delivers throughput up to 300 Mbps.
- Images can be saved by the X-ray generator while the power of the detector is turned on without connecting to a PC. To produce images, X-ray is irradiated at intervals of more than 10 seconds. Check and load the saved images from LG Acquisition Workstation Software.
Detector has been tested with below table’s X-ray condition. This table is only for reference. The legally certified radiologist expert should control X-ray dose.
- Sensor Type: a-Si TFT, X-ray conditions
- Sensor Type: Oxide TFT, X-ray conditions
NOTE
- In the case of the Oxide TFT X-ray condition table, it is only applicable to 14HQ901G-B and 17HQ901G-B models. If the condition table is applied to other models, the desired image may not be obtained.
- Regarding paediatric dosage, it should be much less than for an adult. The certified radiologist should pay special attention to paediatric X-ray dosage levels.
CAUTION
- Always use the detector in places that meet the following environmental requirements.
GRID
Battery
NOTE
- The capacity of the battery pack decreases as the usage period increases.
– The longer the period of use, the higher the possibility that the detector’s operating time has become shorter.
– The battery pack can be replaced at the end of its life. To obtain a replacement, please contact the manufacturer.
Battery Charger
Battery Charger Adapter
Control Box
Cables
Wireless module (LGSBWAC93) Specifications
ENVIRONMENTAL REQUIREMENT
PC System Requirements
INSTALLING CALIBRATION SOFTWARE
How to Install
Launch the installation file of Calibration Software. After the installation
file is launched, follow the instructions displayed on the screen to complete
the installation.
How to Uninstall
To remove Calibration Software from your PC, read and follow the steps below:
Uninstall in the Control Panel
- Click and launch Control Panel in your Start menu.
- Select Programs and Features in Control Panel.
- Select [LG DXD Calibration] from the list of the installed programs.
- Select the [Delete] button when the screen for installation and uninstallation of the program appears.
- Follow the on-screen prompts for proceeding uninstallation and click the [Next] button to complete uninstallation.
Uninstall using the Installation File
Launch the installation file of Calibration Software and follow the instructions on the screen for uninstallation to complete the operation.
NOTE
- To uninstall the program using the installation file, make sure to run the installation fileof the same version as the software installed in the PC.
CONNECTION TYPE
Connection of X-ray Generator – Detector
Select the trigger mode according to the image acquisition method.
– Auto mode: The detector automatically detects and acquires X-ray images.
– Manual mode: The detector recognizes the X-ray generator signals and
acquires images.
Connection of Detector – PC
The status of the connection between the detector and a PC
– Wired Mode: The wired connection between the detector and a PC through the
Control Box
– Wireless (STA) mode: The wireless connection between the detector and a PC
through a wireless AP
– Wireless (AP) mode: Wireless connection between the detector and a PC
through the Soft AP function
Network Connection Mode
Select the network connection mode using (Wired/Wireless Connection button).
– Press (Wired/Wireless Connection button) for at least one second to switch
between the following connection modes, in respective order:
Ethernet/Station/AP mode.
Trigger Cable
The trigger cable is connected to the control box and X-ray Generator. It can only be used in manual mode, and not in automatic mode.
NOTE
- Connection of trigger cable and X-ray generator is to be performed by qualified personnel. The description of each pin is in the common language for this industry.
Block diagram of Trigger Cable connection
Connecting Detector to PC (Wired Mode)
A: Detector
B: Main Cable
C: Control Box
D: LAN Cable
E: PC
F: Object
G: X-ray Switch
H: Power Cord (AC 100-240 V~)
I: X-ray Generator
J: X-ray Generator Interface
K: Trigger Cable
Connecting the Cables
Connect – Wired Connection
Use the LAN cable to connect a PC to the Control Box and connect the detector
to the Control Box with the main cable.
After connecting the cables, proceed to set up the PC as shown below.
-
Launch the [Network and Sharing Center] and click [Change adapter settings].
– Control Panel > Network and Internet > Network and Sharing Center > Change adapter settings -
Go to [Properties] in Local Area Connection.
-
Select [Internet Protocol Version 4 (TCP/IPv4)] and then click [Properties] to set the IP address as follows:
– IP address: 10.10.10.2 ~ 10.10.10.254 (set an IP address other than 10.10.10.99, 10.10.10.100)
– [Subnet Mask]: 255.255.255.0
– [Default Gateway]: 10.10.10.1
– Do not set up DNS. -
Launch Calibration Software, go to (Settings) > [DXD] > [Connection & File Save] and enter the detector’s
IP (10.10.10.100). Click the [Ping] button to check the connection.
- Press the [Ping] button. If connected successfully, the following pop-up will appear:
Connecting Detector to PC (Wireless Mode)
A: Detector
B: AP
C: LAN Cable
D: PC
E: Object
F: X-ray Switch
G: X-ray Generator
H: X-ray Generator Interface
I: Main Cable
J: Control Box
K: Power Cord (AC 100-240 V~)
L: Trigger Cable
Auto Mode
1. Station mode (for the use of external AP)
2. AP mode (for the use of detector internal AP)
NOTE
- Please install the AP and Detector as near as possible without obstacles in between them.
Manual Mode
Making connections-Wireless connection
-
The wireless factory default settings are as follows:
• Station mode (connection via external AP)
– SSID: LGEDXD
• AP mode (connection via detector AP)
– SSID: LGEDXD_SOFTAP -
It is possible to change the Wireless Settings using LG DXD Calibration Software.
– Please see “Wireless AP Settings” for details. -
Press (Wired/Wireless Connection button) for at least one second to switch between the following connection modes, in respective order: Ethernet/Station/AP mode.
-
The Connection method as below.
• Station mode
– PC settings and connection with Detector are same with wired Connection.
• AP mode
– Enter [Wi-Fi] under PC Settings, and enter [Show available networks].
– Attempts are made to connect after checking the DXD wireless AP SSID, which is shown as the research result (the initial value is LGEDXD_SOFTAP). Enter the password to connect.
– If the ping test fails despite the wireless connection being complete, please enter the IP address as follows:
1. Launch the [Network and Sharing Center] and click [Change adapter settings].
– (Control Panel > Network and Internet > Network and Sharing Center > Change adapter settings)
– Select a network adapter.
2. Go to [Properties] in Local Area Connection.
3. Select [Internet Protocol Version 4 (TCP/IPv4)] and then click [Properties] to set the IP address as follows:
– IP address: 10.10.10.2 ~ 10.10.10.254 (set an IP address other than 10.10.10.99, 10.10.10.100)
– [Subnet Mask]: 255.255.255.0
– [Default Gateway]: 10.10.10.1
– Do not set up DNS.
The model and serial number of the product are located on the back and on one side of the product. Record them below in case you ever need service.