GIMA LW05202SL Standard Electric Hospital Bed User Manual
- June 12, 2024
- GIMA
Table of Contents
LW05202SL Standard Electric Hospital Bed
Product Information
Product Name: Professional Medical Products
Model Numbers: LW05202SL (GIMA 27683), LW05202S
(GIMA 27684)
Manufacturers:
- Mingguang Longway Technology Co., Ltd., China
- Lotus NL B.V., Netherlands
- Gima S.p.A., Italy
Contact Information:
-
Mingguang Longway Technology Co., Ltd.: No. 59 Lingji Rd.,
Industrial Park, Mingguang, Anhui, China -
Lotus NL B.V.: Koningin Julianaplein 10, 1 e Verd, 2595AA, The
Hague, Netherlands -
Gima S.p.A.: Via Marconi, 1 – 20060 Gessate (MI) Italy
-
Email: gima@gimaitaly.com, export@gimaitaly.com
-
Website: www.gimaitaly.com
Product Usage Instructions
1. General Information
1.1 Structure of the manual and procedure for consulting
The instruction manual is an integral part of the provision of
this bed. It is intended for customers, owners, safety personnel,
users, and maintainers.
1.1.1 Symbols Used
The following pictograms can be found in the manual:
-
Operation must be carried out by qualified personnel
only or an assistance centre. -
Read the instruction manual.
-
Operation that must be done by two
persons. -
Warning! Danger.
-
Warning! Important information.
1.1.2 Aim and content
The aim of the manual is to provide guidance on the usage and
maintenance of the product.
1.1.3 Maintenance
The manual includes information on maintenance procedures for
the product.
1.2 Introduction
The introduction section provides an overview of the
product.
1.2.1 Intended use
The product is intended for medical use.
3. Product Description
3.4.1 Technical specification
The manual provides detailed technical specifications of the
product.
3.4.2 Electric components
The manual includes information about the electric components
used in the product.
4. Assembly
4.1 Unpacking
The manual provides instructions on how to unpack the
product.
4.2 Assembly
The manual includes step-by-step instructions for assembling the
product.
5. Use
5.1 Use of the medical device
The manual provides instructions on how to use the medical
device.
6. Troubleshoot
The manual includes troubleshooting information to help resolve
any issues that may arise.
8. Warranty
8.1 Warranty
The manual provides information about the warranty coverage for
the product.
8.2 Expected lifespan
The manual includes information about the expected lifespan of
the product.
PROFESSIONAL MEDICAL PRODUCTS
LETTI PER DEGENZA HOMECARE BEDS LITS MÉDICALISÉS À DOMICILE CAMAS DE CUIDADO
EN EL HOGAR
È necessario segnalare qualsiasi incidente grave verificatosi in relazione al
dispositivo medico da noi fornito al fabbricante e all’autorità competente
dello Stato membro in cui si ha sede. All serious accidents concerning the
medical device supplied by us must be reported to the manufacturer and
competent authority of the member state where your registered office is
located. Il est nécessaire de signaler tout accident grave survenu et lié au
dispositif médical que nous avons livré au fabricant et à l’autorité
compétente de l’état membre où on a le siège social. Es necesario informar al
fabricante y a la autoridad competente del Estado miembro en el que se
encuentra la sede sobre cualquier incidente grave que haya ocurrido en
relación con el producto sanitario que le hemos suministrado
LW05202SL (GIMA 27683) LW05202S (GIMA 27684)
Mingguang Longway Technology Co., Ltd. No. 59 Lingji Rd., Industrial Park,
Mingguang, Anhui, China Made in China Lotus NL B.V. Koningin Julianaplein 10,
1 e Verd, 2595AA, The Hague, Netherlands
Gima S.p.A. Via Marconi, 1 – 20060 Gessate (MI) Italy gima@gimaitaly.com –
export@gimaitaly.com www.gimaitaly.com
M27683-M-Rev.0-04.23
2
ENGLISH
Summary
1. GENERAL INFORMATION
3
1.1 Structure of the manual and procedure for consulting ……………………………………………………
3
1.1.1. Symbols Used …………………………………………………………………………………………………………………………………………….. 3
1.1.2 Aim and content ………………………………………………………………………………………………………………………………………….. 3
1.1.3 Maintenance……………………………………………………………………………………………………………………………………………….. 3
1.2 Introduction……………………………………………………………………………………………………………. 3
1.2.1 Intended use……………………………………………………………………………………………………………………………………………….. 3
2 AVVERTENZE DI SICUREZZA
3
2.1 General provision ……………………………………………………………………………………………………. 3 2.2 Provisions relating to electrical hazard ………………………………………………………………………… 3 2.3 Provisions of use……………………………………………………………………………………………………… 3 2.4 Use ……………………………………………………………………………………………………………………….. 3 2.5 Contraindication to use…………………………………………………………………………………………….. 4 2.6 Residual risk …………………………………………………………………………………………………………… 4
3. PRODUCT DESCRIPTION
4
3.1 Reference to illustrations ………………………………………………………………………………………….. 4 3.2 Description …………………………………………………………………………………………………………….. 5 3.3 Optional extras ……………………………………………………………………………………………………….. 5 3.4 Technical specification ……………………………………………………………………………………………… 6
3.4.1 Technical specification ………………………………………………………………………………………………………………………………….. 6
3.4.2 Electric components……………………………………………………………………………………………………………………………………… 7
4. ASSEMBLY
7
4.1 Unpacking………………………………………………………………………………………………………………. 7 4.2 Assembly ……………………………………………………………………………………………………………….. 7
5. USE
10
5.1 Use of the medical device …………………………………………………………………………………………. 10
6. TROUBLESHOOT
11
7.ROUTINE MAINTENANCE ……………………………………………………………………………………………. 12 7.1 Routine maintenance table ……………………………………………………………………………………….. 12 7.2 Cleaning metal partse ………………………………………………………………………………………………. 12 7.3 Cleaning plastic parts ……………………………………………………………………………………………….. 12 7.4 Cleaning wooden parts …………………………………………………………………………………………….. 12 7.5 Replacement of worn parts……………………………………………………………………………………….. 13 7.6 Storage ………………………………………………………………………………………………………………….. 13 7.7 Disposal …………………………………………………………………………………………………………………. 13
8. WARRANTY
13
8.1 warranty ………………………………………………………………………………………………………………. 13 8.2 Expected lifespan…………………………………………………………………………………………………….. 13
ENGLISH
1. GENERAL INFORMATION 1.1 Structure of the manual and procedure for
consulting
The instruction manual is an integral part of the provision of this bed. The
manual is for customers, owners, safety personnel, users and maintainers.
1.1.1. Symbols Used The following pictograms can be found in the manual.
Operation must be carried out by qualified personnel only or an assistance
centre.
Read the intruction manual
Operation that must be done by two persons
Warning! Danger
Warning! Important information
Use protective gloves
1.1.2 Aim and content This manual provides with information on proper use,
technical characteristics and safety warnings; it cannot, however, replace the
user appropriate experience. 1.1.3 Maintenance This manual must be kept in an
accessible and well-known place by all the operators, protected against
humidity, heat and light.
1.2 Introduction
1.2.1 Intended use The device intended to be used as a bed only. This bed is
intended to be used in private home, care home, hospital, or community
environment, by users over 12 years old. Different uses are not contemplated.
This product can only be paired with the products featured in the list of
compatible accessories. Use with other products must be evaluated in advance
and may invalidate the warranty.
2 AVVERTENZE DI SICUREZZA
2.1 General provision Do not proceed to assembling this product before
consulting
this instruction manual. Every alteration of the product unbeknown to the
manufacturer is the full responsibility of whoever makes the changes.
Alterations made without the authorization of Longway Medical Technology Co.,
Ltd. will invalidate any warranty and invalidate the declaration of conformity
to the applicable directives.
3
It is forbidden to use the product in a different way from the one described
in this user manual.
Longway Medical does not any responsibility for malfunctions or damages to
objects and persons due to the incorrect
use of the product. Operators must carefully read this manual, follow the
given instructions and familiarize themselves with the procedures
for the use and maintenance of this product. When the patient is not under
surveillance, the medical bed should be set in the lower position with the
siderails up in
order to reduce the risk of falls and injuries. Avoid any inappropriate use of
the power cord. The dangers arising could be – for example – kinks, cuts and
other me-
chanical damages. In the event that cables from other equipment are located
inside the medical bed, it is necessary to take precautions in order to avoid
them to be crushed between other compo-
nents of the medical bed. Due to the limited space under the medical bed, it
is neces-
sary to use it with suitable lifting devices. Use and perform any maintenance
on the product following the instruction provided by this manual and use only
spare
parts and assistance approved by Longway Medical. Do not use the product for
purposes other than those for
which it was built and designed for. Retain this manual for consulting or
personnel training. It is intended to be kept together with the product if
sold or used by new users.
2.2 Provisions relating to electrical hazard
Power supply must be provided to the machine using a protected line in order
to avoid overcurrent through automatic magneto-thermal switch and diff; these
devices must provide enough
electrical power to protect the size of the power wires. If extension cords
are necessary, the power wires must end with a plug which is indissolubly
plugged or with other devices able to avoid water entering during the cleaning
process
of the bed. The medical bed must be provided with a suitable equipment in
order to maintain the power wire free from any mobile part or mechanism of the
medical bed when this is functioning, or while it is moved, in order to avoid
any dam-
ages to the power wire. Power cords must be adequately protected against
damages that may be caused if they come in contact with one or more moving
parts or they rub against sharp angles and surfaces inside the medical bed.
2.3 Provisions of use
Warnings on the use of the medical device are shown below. Before operating
the moving parts of the bed, make sure they don’t damage any pipe, electric
cable, catheters and other devices else and ensure they will not make the user
uncomfortable or in any sort of problematic situation; see
this manual for information on the procedure. When provided, wheels have the
sole purpose of moving the bed when sanitation, therapy etc. are necessary and
they should be kept braked at any other time.
2.4 Use
The daily use of the product is allowed only when each and every safety
security feature and functional part is well-functioning
4
ENGLISH
and correctly maintained. The bed is intended to be used in private or care
home’s, hospitals and community environment by users over 12 years old. The
only possible use is the one described in this manual; we recommend to read
the instructions carefully in order to be fully aware of the functionality of
the product.
2.5 Contraindication to use
It cannot be used when moving parts are not connected or welded, or show
abnormalities. · It cannot be used in environment subject to water splashing
or critical and adverse weather conditions. · It cannot be used if
defectiveness is shown: this can cause unexpected situations of danger.
2.6 Residual risk
If the requirements displayed in this manual are not observed · Risk of limbs
crushing if limbs are placed under the moving parts during downward movements.
· Risk of injury or fire if the electrical components come in contact with
fluids. · Risk of damages to the product if this is subject to critical or
adverse weather conditions.
3. PRODUCT DESCRIPTION
CODES: LW05202(B) Standard Electric Homecare Bed(with wooden slat) LW05203(B)
Economy Standard Electric Homecare Bed(with wooden slat) LW05207(B) Electric
Homecare Bed(with wooden slat) LW05208(B) Economy Electric Homecare Bed(with
wooden slat) LW05202S Standard Electric Homecare Bed New LW05202SL Extra Low
Standard Electric Homecare Bed New LW05206 Heavy Duty Homecare Bed LW05211 Two
Function Homecare Bed LW05212 Three Function Homecare Bed LW05204 Electric Low
Bed
3.1 Reference to illustrations
1. Monkey pole 2. Handle 3. Back section 4. Button/remote with cable 5.
Wheels with brake 6. Footboard 7. Foot section 8. Side panels 9. Thigh section
10 Pelvis section 11 Mattress stopper 12 Electric actuator 13 Power Led 14
Trendelenburg/Antitrendelenburg activation
(not for LW05211 and LW05212) 15 Bed base height adjustment (not for LW05211)
16 Legs-back simultaneous adjustment 17. Legs section adjustment 18. Back
adjustment 19. Wheel with brake
The manufacturer reserves the right to bring modifications to the product and
its instruction manual without advance notice or modifying earlier
productions. The manufacturer is relieved of any responsibility for damages
resulting from misuse by untrained personnel, unauthorized alterations or
operations, use of non-original spare parts, exceptional events, total or
partial disregard of the instructions displayed in this manual.
Bed control unit
ENGLISH
5
3.2 Description
The bed was designed and built to provide high-quality standards and
sturdiness. The medical device is composed of a bed base, named applied part,
a footboard and a support headboard. This three joint, four section bed with
adjustable height has the following specifics: · Steel square-meshed net frame
200×89 cm which is height adjustable using electronic controls provided in 43
or 86 cm · Back section adjustable from 0° to 70° using the electronic pusher
(24V dc). · Fixed pelvis section · Legs section adjustable from 0° to 30°
using the electronic pusher (24V dc). · Back section continuously adjustable
together with the legs section · Removable wooden siderails · Headboard and
footboard height 44 cm · This bed is intended to be used in private or care
home, hospital or community environment, by users over 12 years old. However,
it can also be used as room furniture in the same environment. Use with other
products must be evaluated in advance. · Supplied wheels with brakes for
headboard and footboard of 12,5 cm Ø
3.3 Optional extras
ITEM CODE
LW10205
DESCRIPTION Deluxe Bedrail
PHOTO
OPTIONAL EXTRA MAY NOT BE AVAILABLE
LW10206 N-shaped Bedrail
OPTIONAL EXTRA MAY NOT BE AVAILABLE LW10207 Lift Round Pole
The use of the different types of mattresses from those shown requires the
observance of the minimum height for the later parapets and the requirements
set by the UNI 60601-2-52 standard.
WARNING! Taking into account the safety devices for the patient, the
manufacturer declines all responsability in the case of using a different type
of mattress which fails to comply with the minimum height of the parapets set
by the UNI 60601-2-52 standard.
6
ENGLISH
3.4 Technical specification
3.4.1 Technical specification
LW05202(B) LW05207(B) LW05202S LW05202SL
Bed mass
120 kg
Max capacity (mattress + user + accessory)
180 kg
Maximum dimension 2150 X 1020 mm
Bed base height from the floor (min/max)
Backboard/footboard height
380 / 790 mm 270 / 680 mm (LW05202SL)
560 mm
Bed base size
2000 X 900 mm
Head section
790 X 715 mm
LW05203(B) LW05208(B)
LW05211 LW05212
LW05204
LW05206 (120cm / 140cm)
108kg
70kgs(LW05211) 110kgs(LW05212)
200kg
165kgs/178kgs
180kgs
170kgs
180kgs
280kgs
2120 x 1030mm
2080 x 1020mm
2200 x 1015mm
2150 x 1320mm / 1520mm
380/790 mm
420mm(LW05211) 330/650mm(LW05212)
315/685mm
380/790 mm
560 mm 2000x900mm 790x 715 mm
560 mm 2000x900mm 790x 715 mm
560 mm
560 mm
2000x900mm
2000×1200/ 1400mm
790x 715 mm 820/840x715mm
Pelvis section
860 X 205 mm
860x 205 mm
860x 205 mm
860x 205 mm 890/910x205mm
Legs section
790 X 295 mm
790x 295 mm
790x 295 mm
790x 295 mm 820/840x295mm
Foot section
790 X 440 mm
790x 440 mm
790x 440 mm
790x 440 mm 820/840x440mm
Angle moving back section (max)
Angle moving legs section (max)
Moving legs section height from the bed base (max)
Storage and packaging temperature (in a dry environment)
Temperature range of usage environment
70° 25° 210 mm -10 ÷ 50°C 10 ÷ 50°C
70° 25° 210 mm -10 ÷ 50°C 10 ÷ 50°C
70° 25° 210 mm -10 ÷ 50°C 10 ÷ 50°C
70° 25° 210 mm
70° 25° 210 mm
-10 ÷ 50°C 10 ÷ 50°C
-10 ÷ 50°C 10 ÷ 50°C
Mattress weight (max)
30 kg
30 Kg
30 Kg
30 Kg
30 Kg
User weight (max)
Packaging sizes HEADBOARD/ FOOTBOARD + BED SPRING
Packaging sizes SIDE PANELS
Packaging sizes MONKEY POLE
135 kg
1240 X 410 / 450(B) X 960 mm
(LW05202(S); 1240 X 390 / 410(B)
X 930 (LW05207)
2050 x 150 x 110 mm
2074 x 400 x 400 mm
1240 x 410/450(B) x 960 mm(LW05203); 1240×390/410(B) x930(LW05208)
2050 x 150 x 110 mm
2074 x 400 x 400 mm
960x290x1040mm(LW05211); 960x215x-
1040mm(bed body) 1550x315x940mm (bed
base)-LW05212
2050 x 150 x 110 mm
2074 x 400 x 400 mm
2150x285x950mm
2050 x 150 x 110 mm
2074 x 400 x 400 mm
1115x270x1300/1500mm
2050 x 150 x 110 mm
2074 x 400 x 400 mm
ENGLISH
7
3.4.2 Electric components Supply voltage 230V AC Operating voltage 24V DC MAX
power consumption 220 W MAX actuators thrust force 6000 N MAX hand control
voltage 24V DC
4. ASSEMBLY
4.1 Unpacking
Il manuale di istruzioni è parte integrante della fornitura. Il manuale è
destinato a clienti, proprietari, personale addetto alla sicurezza, utenti e
manutentori.
Used symbols The following pictograms can be found in the manual.
The bed comes disassembled into its parts in nr, 2 boxes. Make sure all the components are in good condition before proceeding.
When the packed product is transported eith a trolley be careful to avoid injuries from accidental roll.
Open the package and extract content with caution, making sure not to demage it.
After unpacking the product, make sure it is not damaged; otherwise, please contact the authorized dealer. Damages caused during incorrect transport, unpacking or assembly are not covered under warranty.
Bed components:
BOX 1
BOX 2
Bed base and motor with inbuilt control unit
Headboard and footboard
Wooden siderails
Operations must be carried out by qualified personnel only
4.2 Assembly
The product must be transported following the directions on the package; if
manual handling is necessary, keep in mind that one operator can carry a
maximum load of 25 kg.
Lift and move the bed slowly in order to avoid injuries from biomechanical
overload of the spine. If moving the product is necessary, we recommend the
use of protective gloves and the presence of at least two operators to lift
the product.
8
ENGLISH
The assembly operation of the medical device must be carried out by at least two operators in order to avoid damages to objects or persons.
Use protective gloves
The device is supplied dismantled
Bed base – Insert the two inserts at the end of the bed base, head side. The inserts must be assembled so that one of them protrudes more than the other. – Slightly tighten the two bolts, head side – Push the bed base, foot side, against the two inserts. – Tighten the two bolts at the foot side. – Complete tightening all the bolts.
Motor and in-built control unit assembly – When the bed base is assembled,
before proceeding and fixing it to the headboard and the footboard, the bed
base must be overturned so that its hooks are pointing upwards.
For LW05207(B)/LW05208(B) – Starting from this position of the bed base,
proceed on assembling the motor unit by pushing and make sure it is fixed to
the two pins – Slide out the two lids from the slots, where the hooks for the
bed shall be put – Flip the motor and match the slots with the two protruding
hooks and push on. – Reinsert the two lids into the motor over the hooks to
secure
Before locating the hooks, please pay attention to the the icons affixed on
the motor to ensure correct orientation
HEAD SIDE
LEG SIDE
For LW05202(B)/LW05203(B)/LW05202S/LW05202SL/LW05206/LW05204/LW05211/LW05212 The control unit is fixed on the main bed plate Labels are applied on the control unit with symbols illustrating the connection point of the motor and control sockets. Implement the links according to the following matches:
ENGLISH
9
M1 M2 M3 M4 Handset
Head Motor Foot Motor Legrest Elevating Motor Backrest Elevating Motor Hand control
HEADBOARD /FOOTBOARD assembly · Hook the bed base onto the headboard
/footboard and push tightly in order to block it in position
· Tighten the bolts
WIRING assembly For LW05207(B)/LW05208(B) 23. Headboard motor connection 24. Footboard motor connection 25. Remote connection
Connect the motor of headboard
For LW05202(B)/LW05203(B)/LW05202S/LW05202SL/LW05206/LW05204 /LW05211/LW05212
To prevent cables from being damaged when the engines are activated, follow
the instructions below: · Connect the cables of the test/foot motors directly
on the control unit. · Connect the height adjustment motor cables contained in
the heads directly on the control unit · Connect the remote control cable to
the control unit · Connect the mains cable to the general 230V power supply. ·
Bring the head/foot motors to a high position. · Carry out all section
movements, mattress platform, and trendelenburg to check for interference
between cables and moving parts. · Make sure all cables are free and without
risk of entangling in moving parts of the bed
Insert the monkey pole (loops from parts box)
10
ENGLISH
In order to avoid the damage of the cables when the engine is starting, strictly follow the following instruction: – Connect the headboard/footboard cables straight to the central motor unit – Connect the remote cable to the central motor unit – Connect the mains cable to a 230V general electric supply – Put the headboard/footboard motors in a high position – Perform movements of the sections, mattress platform and trendelenburg in order to verify possible interference between the cables and the moving parts – Make sure all the cables are free from the risk of entanglement with the moving parts of the bed
Take care when handling the footboard the risk of crushing and damage to the product is real as the adjustable foot-lifter is not securely locked into the head or foot board.
Remove the plug from the socket before moving the bed
SIDE PANELS assembly – Lift the bed base approximately 1/4 to make the insert of the panels easier – Insert the panels from below until the stops click in position
Conservare il dispositivo elettromedicale in un luogo in cui la spina possa
essere facilmente staccata dalla pre-sa di corrente
5. USE 5.1 Use of the medical device
Keep the electro-medical device in a place where the plug can be easily
disconnected from socket
The remote is provided with safety lock thanks to a magnet which turns on
automatically when the device has not been used for more than 30/40 seconds.
The device is not provided with a main switch, just plug into the mains socket
to access all the features. Pulling the plug off the mains socket works as
emergency response on the power supply.
The bed must be seton its lowest height and the siderails raised while not
supervised in order to reduce the risk of falling.
ENGLISH
11
13. Power Led 14. Trendelenburg/Anti -trendelenburg activation 15. Bed base height adjustment 16. Legs-back simultaneous adjustment 17. Legs section adjustment 18. Back adjustment 19. Wheel with brake
Power led (Green means on) Command reception led (Red means on) Handset Safety Lock Magnet
The WARNING! symbol is on when one of the components of the device is being
moved by the electric motors.
In order to move the wooden siderails, it is necessary to push the brown
button which activates the locking pin on the headboard and footboard, then
proceed as following: – LIFTING: grab an end of the panel and pull up the
higher wooden siderail in order to activate the locking pin; if this clicks,
the attachment has been successful. Repeat this operation for the headboard
and then the footboard or vice versa. – LOWERING: grab an end of the siderail,
push the brown button, then lower the rail all the way down. Repeat this
operation with the other end. To activate the monkey pole: Insert the monkey
pole on the provided hook on the bed base, making sure to insert the locking
stops into the right holes on the hook.
6. TROUBLESHOOT
When operations are represented by this symbol they must be carried out by
qualified personnel only.
INCONVENIENCE
SOLUTION
The bed doesn’t answer to the UP/DOWN command
– Check the plug is correctly plugged in – Repeat the activation process of the remote – Verify and clean the contacts – Reduce the load
The LED command of the remote is not working
– Replace the LED
12
ENGLISH
The battery does not charge properly The device moves intermittently
The panels are not properly inserted Unusual sounds when the device is
operating
– Contact the technical support service
– Remove the movement connectors from the remote and clean the contacts with a
dry alcohol pad – Lubricate the machine
– Contact technical support
The company responsible for supplying the device and for technical support service is: Mingguang Longway Technology Co., Ltd. No. 59 Lingji Rd., Industrial Park, Mingguang, Anhui, China Tel:+86-550-8138-128 Fax.:+86-550-8021-399
7. ROUTINE MAINTENANCE
These operations can be carried out by the user. The device needs ordinary
maintenance just like any other product, so that it is maintaned in safe
perfect functioning condition. Food, hair and any other accumulating debris –
like dust – must be removed daily; a domestic vacuum cleaner can be used for
this operation.
The following table shows the recommended frequency for maintenance operations. Constant monitoring and attention by the user shall encourage carrying out the actual necessary operations.
7.1 Routine maintenance table
Frequency
6 months
Cleaning metal parts Proceed with cleaning these parts according to use and
necessity
Cleaning plastic parts Proceed with cleaning these parts according to use and
necessity
Cleaning wooden parts Proceed with cleaning these parts according to use and
necessity
1 year
General check · Check the wheels and tighten the bolts · Check paint for rust or chipping · Check plastic parts are intact and the state of wear, consider replacement if reqd · Check if the electric motors are working properly and lubricate the screws
If transferred to third parties or restored
Disinfection Warning! Disinfection substances must be used by specifically
qualified personnel only The disinfection of this device must be carried out
by specialized centers according to the dispositions
made by the competent health authority.
For more information contact the authorized technical support service.
7.2 Cleaning metal partse
Painted metal parts can be cleaned using sprays for domestic use only if they
are alcohol-free and solvent-free, alcohol and solvents can discolour or
damage painting. Chromed parts can be cleaned using ordinary domestic products
and before being carefully dried.
7.3 Cleaning plastic parts
Plastic parts can be cleaned using an ordinary plastic cleaner following the
instructions and
contraindications of the product. Avoid solvent-based and alcohol-based
products. Do not use jets and avoid splashing water.
7.4 Cleaning wooden parts
Wooden parts can be cleaned using an ordinary non bleach cleaners, following
the instructions and contraindications of the product. Avoid solvent-based and
alcohol-based products. Do not use jets and avoid splashing water.
ENGLISH
13
7.5 Replacement of worn parts
Operations of replacement of worn parts must be carried out by qualified personnel only
If, during the at least yearly routine inspection, damaged or worn parts are
revealed, these must be immediately replaced as recommended by the inspecting
personnel. We recommend an independent inspection is completed at leat every
12 months.
Original spare parts must be used when replacing damaged parts, we cannot
accept responsibility for damages to objects or persons if this is not adhered
to.
List of basic common spare parts: · platform electric motor kit · handset ·
castor with brake · Headboard/footboard motor
WARNING! Any maintenance and/or replacement operation must be carried out
exclusively when the bed is not being used by the user and free from any
accessories (Linen, mattress, panels, etc.) and is UNPLUGGED.
7.6 Storage
If the medical device remains unused for a long time, it must be kept in a dry
environment away from direct sunlight and humidity in order to avoid damages
to the device. Make sure the medical device is kept in perfect clean
condition.
7.7 Disposal
Before proceeding with dismantling, the device must be unplugged.
This symbol on the device means that the product at its end of life must be
collected separately from household waste in accordance with European
Directive 2012/19/UE on electric and electronic waste (RAEE).
The packaging material is 100% recyclable. For disposal follow local
regulations. The packaging material (plastic bags, polystyrene, etc.) must be
kept out of reach of children as potential source of danger.
For further information on treatment, recovery and recycling of this product,
please contact the local office according to the applicable law.
8. WARRANTY 8.1 warranty
The product is covered under a legal 12-month warranty starting from the
delivery date (d. legs 24/2002) to the customer of Longway Medical and Dir. CE
1999/4 against any kind of manufacturing defect.
Throughout the duration of warranty, the maintenance of the device is
guaranteed entirely at the expense of the manufacturer when it is established
that the problem is a manufacturing defect. Warranty will be invalid should
routine maintenance be neglected.
Warranty does not cover parts that tend to get worn during normal use (wheels,
etc.)
Warranty does not cover: 1. Damages from misuse or different use than the one
specified in this manual. 2. Damages from incorrect assembly by the customer.
3. Damages from alteration, repair and mishandling by unauthorized personnel.
4. Shipping and packaging of the product under warranty are at the expense of
the dealer or the manufacturer.
Warranty does not entitle the customer to demand compensation in any way.
Spare parts are available for 5 years from the date the product is
discontinued.
If customers detect problems, they must immediately inform the authorized
dealer, providing information on the detected problem using data from the
specific data decal on the device, filling the given defect and failure
reporting form.
The customer, after making arrangements with the authorized dealer, will take
care of the shipment of the product for warranty repair.
Throughout the duration of warranty, technical intervention will be made by
the manufacturer who will make sure that the problem is a manufacturing defect
and not a fault of a different nature. This manual must be kept by the owner
and must stay with the device for the exercise of guarantee.
Warranty starts from the delivery date.
8.2 Expected lifespan
The average lifespan of your homecare bed is 5 years. Depending on the
frequency and way of use and maintenance, the lifespan of your homecare bed
will increase of decrease.
14
SYMBOL LEGEND
Manufacturer
Date of manufacture
Product code
Lot number Authorized representative in the European community
Caution: read instructions (warnings) carefully
Consult instructions for use
Medical Device compliant with Regulation (EU) 2017/745 Medical Device
Unique device identifier
WEEE disposal
Imported by
GIMA WARRANTY TERMS The Gima 12-month standard B2B warranty applies.
Disposal: The product must not be disposed of along with other domestic waste.
The users must dispose of this equipment by bringing it to a specific
recycling point for electric and electronic equipment.
ENGLISH
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>