hetaida HTD8813 Non-Contact Infrared Body Thermometer User Manual

June 12, 2024
HeTaiDa

hetaida HTD8813 Non-Contact Infrared Body Thermometer

Safety Information

  • This device may only be used for the purposes described in these instructions. The manufacturer cannot be held liable for damage caused by incorrect application.
  • The noncontactinfrared body thermometeris designed to minimize the possibility of hazards from errors in the software program by following sound and lightengineering design processes, Risk Analysis and Software Validation.

Warning: Warnings are identified by the WARNING symbol shown above.

  • The NonContact Infrared Body Thermometer is to be operated by consumers in the home setting and primary care setting as screening tool. This manual, accessories, Directions for Use, all precautionary information, and specifications should be read before use.
  • This product is designed to measure human body temperature on the forehead. Do not use it for any other purpose.
  • This product is intended in the home setting and primary care setting as screening tool.
  • Do not use the thermometer if it malfunctions or has been damaged in any matter.
  • When the ambient temperature of the thermometer changes too much, such as moving the Thermometer from one place of lower temperature to another place of higher temperature, Allow the thermometer to remain in a room for 30 minutes where the temperature is between15℃to 40℃(59℉- 104℉).
  • Remove primary batteries if equipment is not likely to be used for long time.
  • This product is not waterproof, do not be immersed in water or other liquid; If cleaning and disinfection, please follow the “Care and Storage” section requirements.
  • Do not touch the sensor of infrared detection with your fingers.
  • If a cold compress on the forehead fever patients, or take other measures to cool down the temperature data will low, should be avoided in this case to measure body temperature.
  • If measure human forehead temperature , please select “body” mode; for measure other objects, liquids, food and other temperature please select “surface” mode.
  • This product must be operated in a stable environment, if the ambient environment was mutations, please should be note whether there is fog on the sensor, if any, before using accordance with the “Care and Storage” section to removing the fog.
  • Do not near strong electrostatic field or strong magnetic fields, thus avoiding the impact on the accuracy of the measurement data.
  • Do not mix the old and new batteries to avoid damage to the product.
  • It may affect the accuracy of measurements when the forehead is covered by hair, perspiration, cap or scarf.
  • The measuring result of this product is only for your reference. If you have any doubt, please measure the temperature in other methods.
  • The device should be kept out of the reach of children/pets. When not in use, store the device in a dry room and protect it against extreme moisture, heat, lint, dust and direct sunlight. Never place any heavy objects on the storage case.
  • Do not throw batteries into fire.
  • Only use recommended batteries. Do not use rechargeable batteries.
  • This thermometer will irreplaceable the diagnostic in hospitals.
  • Do not fall, disassemble or modify the device.
  • Do not use this device if you think it is damaged or notice anything unusual.
  • This device comprises sensitive components and must be treated with caution. Observe the storage and operating conditions described in the ‘Technical Specifications’ section.
  • Not servicing/maintenance while the thermometer is in use.
  • When using, shall not touch battery and the patient simultaneously.
  • Do not use the device if it is damaged/ degraded/loosened in any way. The continuous use of a damaged unit may cause injury, improper results, or serious danger.
  • Based on the current science and technology, other potential allergic reactions are unknown.
  • This equipment needs to be installed and put into service in accordance with the information provided in the ACCOMPANYING DOCUMENTS.

Overview

The HeTaiDa Non Contact Infrared Body Thermometers are designed to be used for intermittent measurement and monitoring of human body temperature by consumers in the home setting and primary care setting as screening tool.

Description of Non Contact Infrared Body Thermometer

Device principle and introduction

The HeTaiDa Noncontact infrared body thermometer are hand-held, reusable, battery-operated devices, which can measure human body temperature on the forehead, the skin temperature on one’s forehead. The operation principle is based on Infrared Sensor technology. The IR sensor can output different signal when measuring different object temperatures or in different ambient temperature, and the ASIC can turn the signal from IR Sensor to a digital value and display it on the LCD.

Description on Controls, Indicators, and Symbols

hetaida-HTD8813-Non-Contact-Infrared-Body-Thermometer-
FIG-1

  1. IR sensor
  2. Liquid crystal display(LCD)
  3. MODE button
  4. On/Scan button
  5. Battery Cover
  6. Plastic Shell
  7. Surface mode
  8. Body mode
  9. Data indicator
  10. Indicator of the measurement result
  11. Low Battery indicator
  12. Voice on/off indicator
  13. Fahrenheit/.Celsius
  14. Memory number
  15. Memory indicator

Thermometer Applications

Thermometer Model Number|

Thermometer Style

| Adult| Pediatric
---|---|---|---
Ear| Forehead| Ear| Forehead
HTD8813| Non Contact Infrared Body Thermometers| | | |

Equipment Symbols

hetaida-HTD8813-Non-Contact-Infrared-Body-Thermometer-
FIG-2

Technical Specifications

hetaida-HTD8813-Non-Contact-Infrared-Body-Thermometer-
FIG-7

Calculated values of the indicators according to ISO 80601-2-56

Forehead mode

hetaida-HTD8813-Non-Contact-Infrared-Body-Thermometer-
FIG-9

Note: the above value is calculated from the clinical data of HTD8818A.

Safety Classification of ME EQUIPMENT

Note: Not intended to be sterilized. Not for use in an OXYGEN RICH ENVIRONMENT

Operation

Battery installation
Caution: The Non-Contact Infrared Body Thermometer does not operate with dead batteries and does not input outer power. Install new batteries.

  1. Pull the battery downward, toward the bottom of the Non-Contact Infrared Body Thermometer, and remove the battery access door;
  2. Insert two pieces AAA size batteries according to the “+” and “-”;
  3. Close the battery cover.

How to Operate

  • Before Applying the Thermometer

Be sure to read and understand all warnings listed of the instructions before use.

  • The thermometer is aligned with the middle of the forehead to measure body temperature (between the eyebrows above) and keep the vertical distance, press the On/measure button, temperature display immediately, see ?igure 2.
  • When the ambient temperature of the thermometer changes too much, such as moving the Thermometer from one place of lower temperature to another place of higher temperature, Allow the thermometer to remain in a room for 30 minutes where the temperature is between15℃ to 40℃.
  • The ambient temperature around the test person should be stable, should keep away from the larger flow fan, air-conditioning vents and so on.
  • When people moving from one place of lower temperature to another place of higher temperature, should at least remain in the test environment more than 5 minutes, to be consistent with the ambient temperature after the re-measure-ment.
  • Wait at least 1 second for the next measurement. If the continuous measurement of five times, it is recommended to wait at least 30 seconds and then continue measurement.
  • You cannot use the thermometer in place where the sun is strong.
  • If for some reason the low forehead temperature measurement can try to align behind the ears. See figure 3.
General Setup and Use

Start measuring

  1. Turn on the thermometer by pressing the On/measure button. The thermometer will perform self-test with all segments displayed for 2 seconds.
  2. Align staff forehead to keep the distance, and then press the On/measure button to start the measurement, read the data.

Note:

  1. After full display over, you will hear a rattle or “beep beep beep” three times, which means that the measurements have been completed, while the target value of the measured temperature is displayed on the LCD, while backlit display according to the appropriate setting among the three colors red, green, yellow one of. And the Green means ready for next measurement. When 37.4℃-38.0℃, it’s yellow ,means slight fever warning. Please pay attention to body temperature. When the body temperature is above 38.1℃, it’s red , means high fever. Please take action to cool down or go for a doctor.
  2. To ensure the accuracy of the measurement, wait at least 30 seconds after 5 consecutive measurements.

Mode conversion

  • When the device is running, pressing the MODE button to cycle conversion between “body” mode and “surface” mode. “body” mode is used for measuring human body temperature, the “surface ” mode is used to measure the surface temperature. (The factory default is “body” mode).

Recalling and Erasing Memory Data

  • The last temperature taken before the thermometer powers off is stored in memory, up to 50.
    • In the boot or shutdown state, short press the MEMO button to view the history of measured values.
    • An empty memory cell shows “—℃” or “—℉”.
    • Temperature readings can be stored in memory. Up to 50 temperature readings can be stored into the memory cells and automatically overwrite historical data.
    • In boot mode, press the MEMObutton until the LCD display “CLR”, which means that all stored data is cleared completely.

Parameter settings
This product can be set according to the subjects of different colors and different environments data to meet the different characteristics of populations or individuals.
Long press the SET button to modify the measurement parameters

HTD8813 as below step

  1. Unit Set-F1
    • Under the boot mode .Long press SET button to enter F1, press the “MODE” or “MEMO”button to switch Celsius and Fahrenheit temperature units, press the SET button to confirm the unit settings (factory default is Celsius).
  2. Fever alert set-F2
    • Under F1 state, press SET button to enter the F2, press the “MODE” button to decrease 0.1 ℃, press the “MEMO” button plus 0.1 ℃, long press to accelerate the speed of temperature regulation, and finally press the SET button to save. (The factory default is 38.1 ℃)
  3. Prompt sound settings-F3
    • Under F2 state , short press SET button to enter F3, press MODE button or MEMO button to set voice switch, and press the SET button to confirm the settings. (The factory default is the voice Prompt to open).
    • Prompt sound settings ON/OFF function
    • Under the boot mode .Short press SET button to take on or take off the Prompt sound function.
  4. Restore to factory setting function
    • Under the boot mode, long press MODE button until LCD display “rst”. Two seconds later ,former F1-F3 parameter back to factory setting.

 Troubleshooting

hetaida-HTD8813-Non-Contact-Infrared-Body-Thermometer-
FIG-5

Replacing the Battery

  1. Open and release battery cover following indicator on the surface of battery cover. Before changing the battery be sure the system is already power off.

  2. Remove the battery and replace with 2 new one, type AAA, Make sure align properly as indicated inside the battery cover .

  3. Slide the battery cover back in until it snaps in place.
    Do not dispose of used batteries in household waste. Take them to special local collection sites.

  4. In case, if system is latched up after changing battery. You may not follow up the process of rule one. Just take off battery, waiting 30 sec, then load battery again.

Warning

Do not recharge, disassemble or dispose of in fire.

  1. The typical service life of the new and unused batteries is 2000 measurements for the operation time is 18s.
  2. Only use the recommended batteries, do not recharge non-rechargeable batteries and do not burn them.
  3. Remove the batteries if the thermometer is not to be used for a long period.

Cleaning, Care and Storage

The lens is very delicate.
It is very important to protect the lens from dirt and damage.

  • Use a clean, soft cloth to clean the surface of the device and LCD. Do not use solvents or immerse the device into water or other liquids.
  • Always keep the thermometer within the storage temperature range (- 20℃to 55 ℃or – 4℉to 131℉) and humidity range (≤93% non-condensing).
  • It is recommended to store the thermometer in a dry location free from dust. Do not expose the thermometer to direct sunlight, high temperature/ humidity or any extreme environment, otherwise the function will be reduced.
  • When the ambient temperature of the thermometer changes too much, such as moving the thermometer from one place of lower temperature to another place of higher temperature, allow the thermometer to remain in a room for 30 minutes where the temperature is between15℃ to 40℃.

Disposal

  • Used batteries should not be disposed of in the household rubbish. Used Batteries should be deposited at a collection point.
  • At the end of its life, the appliance should not be disposed of in household rubbish. Enquire about the options for environment-friendly and appropriate disposal. Take local regulations into account.

Warranty

Our company warrants Non Contact Infrared Body Thermometer at the time of its original purchase and for the subsequence time period of one year.

The warranty does not cover the followings:

  • The device series number label is torn off or cannot be recognized.
  • Damage to the device resulting from misconnection with other devices.
  • Damage to the device resulting from accidents.
  • Changes performed by users without the prior written authorization of the company.
  • Batteries and packaging are not covered under warranty

When asked to provide warranty service, you must have a purchase date and purchase stamp dealers (including dealers name and address) of the warranty card. Be sure to ask the dealer to purchase this product signature on the warranty card. When asked to provide warranty service, please put the product to get our distribution points for repair. Products outside the warranty expires, will be charged accordingly.

Note:

  1. If you have any problems with this device, such as setting up, maintaining or using, please contact with SERVICE PERSONNEL of HeTaiDa Technology Co., Ltd. Don’t open or repair the device by yourself.
  2. Please report to HeTaiDa Technology Co., Ltd. if any unexpected operation or events occur.
  3. The thermometer is initially calibrated at the time of manufacture. If this thermometer is used according to the use instructions, periodic re-adjustment is not required. If at any time you question the accuracy of temperature measurements, please contact us timely.
  4. The patient is an intended operator. The patient can measure and change battery. Under normal circumstances and maintain the device and its accessories according to the user manual.

EMC Declaration

  1. This equipment needs to be installed and put into service in accordance with the information provided in the ACCOMPANYING DOCUMENTS;
    This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided, and this unit can be affected by portable and mobile RF communications equipment.

Caution

  • Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in incorrect operation of the unit.
  • This unit has been thoroughly tested and inspected to assure proper performance and operation!
  • This machine should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, this machine should be observed to verify normal operation in the configuration in which it will be used.

Guidance and manufacture’s declaration – electromagnetic emission
The Non Contact Infrared Body Thermometer is intended for use in the electromagnetic environment specified below. The customer of the user of the Non Contact Infrared Body Thermometer should assure that it is used in such an environment.

Emission test/Compliance

  • RF emissions CISPR 11 : Group 1
  • RF emission CISPR 11: Class B
  • Harmonic emissions IEC 61000-3-2: Not applicable
  • Voltage fluctuations/ flicker emissions IEC 61000-3-3: Not applicable

NOTE

  1.  At 80 MHz and 800 MHz, the higher frequency range applies.
  2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
  3. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Non Contact Infrared Body Thermometer is used exceeds the applicable RF compliance level above, the Non Contact Infrared Body Thermometer should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Non Contact Infrared Body.
  4. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
  • a) For some services, only the uplink frequencies are included.
  • b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
  • c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual modulation, it would be worst case

Body temperature

  • Body temperature varies from person to person and ?luctuates during the course of the day. For this reason, it is suggested to know one’s normal, healthy forehead temperature to correctly determine the temperature.
  • Body temperature runs approximately from 35.5℃ to 37.8℃(95.9°F-100°F). To determine if one has a fever, compare the temperature detected with the person’s normal temperature. A rise over the reference body temperature of 1℃(1°F) or more is generally indication of fever.
  • Different measurement sites(rectal, axillary, oral, frontal, auricular) will give different readings. Therefore it is wrong to compare the measurement taken from different sites.
  • Here below are typical temperatures for adults, based on different measurement sites:
    • Rectal: 36.6℃ to 38℃/97.9°F-100.4°F
    • Axillary: 34.7℃ to 37.3℃/94.5°F-99.1°F

ASTM laboratory accuracy requirements in the display range of 37℃(98°F to 102°F) for IR thermometers is ±0.2℃(±0.4°F), whereas for mercury in-glass and electronic thermometers, the requirement per ASTM Standards E667-86 and E1112-86 is ±0.1℃(±0.2°F).
Caution: This infrared thermometer meets requirements established in ASTM Standard (E1965-98) Except of clause 5.2.2. It displays subject’s temperature over a range of 34.0~43.0℃. Full responsibility for the conformance of this product to the standard is assumed by(Hetaida Technology Co., Ltd. Add: Room 801, 802, 803, 804, 901, 2# Building Scienti?ic Research Center, Songhu Intelligent Valley, No.6 Minfu Road, Liaobu Town, Dongguan City, Guangdong Province, P.R.China

Foreword

  • The non contact Infrared body thermometer operating Instructions intend to provide the necessary information for proper operation of HTD8813, thermometer model.
  • Only body mode was reviewed and certified by notified body.
  • General knowledge of Infrared thermometer and an understanding of the features and functions of the HTD8813 thermometer model are prerequisites for proper use.
  • The non contact infrared body thermometer is a medical device, and can be used repeatedly, which using life is 5 years.
  • Please read the manual first before using it, if not fully understand the usages, please stop using the thermometer.
  • Do not operate any of the models HTD8813 thermometer without completely reading and understanding these instructions.

Notice
Purchase or possession of this device does not carry any express or implied license to use with replacement parts which would, alone or in combination with this device, fall within the scope of one of the relating patents.

Contacts

For further information contact:
Hetaida Technology Co., Ltd.

  • Address: Room 801, 802, 803, 804,901, 2# Building Scientific Research Center, Songhu Intelligent Valley, No.6 Minfu Road, Liaobu Town, Dongguan City, Guangdong Province, P.R.China
  • Tel :+860769-83326886
  • Contact: Tom. Chen
  • E-mail: [email protected]
  • Address: Share Info Consultant Service LLC Repräsentanzbüro Heerdter Lohweg 83, 40549 Düsseldorf
  • Tel: 0049 1767 0057022
  • Fax: 0049 1767 0057022
  • Email: [email protected]
  • [email protected]

Software version : V31 Ver. NO.: A.10
Ver. Date: 2022.03.03