3S ANALYZERS 3S-UVO UV254 Absorption Sensor for Organic Load User Manual
- October 30, 2023
- 3S ANALYZERS
Table of Contents
USER MANUAL
3S‐UVO
UV254 absorption sensor for organic load
** Electrical equipment marked with this symbol can not be disposed of
through home or public waste disposal systems after 12 August 2005. In
accordance with local and national European regulations (EU Directive 2002/96
/ EC), users must return the equipment which is unsuccessful or can no longer
be used to the manufacturer, which have to provide free of charge disposal.
Note:** To return devices at the end of their useful life, accessories
supplied by the manufacturer and all auxiliary items for recycling, contact
the manufacturer or the vendor of the device to arrange proper disposal.
SAFETY INFORMATION
1.1 Warnings and safety information
Before installing and operating the analyzer, read this manual thoroughly.
Please pay particular attention to all the labels applied to the analyzer and
to all the hazard information indicators in this manual.
This symbol indicates that you must refer to this manual for proper use of the
equipment. Only qualified operators, properly trained on the use and
maintenance of the analyzer can carry out service activities on the equipment.
This symbol indicates the existence of a risk of electric shock and/or
electrocution. Only operators qualified for these activities can perform
maintenance and control operations on the equipment bearing this label, always
after unplugging it.
Parts involved:
- input terminal block in the upper box
The instrument operates with low power UV radiation. Do not look directly at
the light source and do not disassemble the light source enclosure.
Parts involved:
- UV source
The manufacturer shall not be held responsible under any circumstances for
improper use of the equipment.
The head of department and the machine operator must comply with the following
rules and with the provisions of current legislation on the safety and health
of workers.
The use, maintenance, and repair of the instrument are permitted only to
persons authorized for such operations. These operators must be physically and
mentally capable to perform such activities, which can not be performed under
the influence of alcohol and drugs.
When the instrument is not being used it must be protected from voluntary or
involuntary activation, after disconnecting the power supply.
Failure to follow the instructions given and/or failure to pay attention to
the hazard indicators may cause serious risks of physical damage to operators
and breaks or malfunctioning of the analyzer.
All the components of the instrument are placed within a panel closed by a
door with a special key, supplied only to maintenance operators.
The instrument must then be used under operating conditions with the door
closed.
GENERAL INFORMATION
2.1 Technical specifications
Measured parameters | CODeq, TOCeq, BOD eq, SAC254, UV254Abs |
---|---|
Measuring principle | UV254 absorption photometry |
Wavelength | 254 nm |
Measuring range | Low range COD 0 ‐ 370 mg/l eq. KHP / TOC 0 ‐ 150 mg/l |
High range COD 0 ‐1000 mg/l eq. KHP / TOC 0 ‐ 400 mg/l
Reproducibility| ± 2.5 % of the full scale
Limit of detection| 0.75 ppm (KHP equivalents)
Analysis Frequency| 1 s
Sample| Pressure: pressure‐free vessel (depth up to 60 m) Temperature: 5 ‐ 50
°C (41 ‐ 122 °F)
Flow Rate: 80 to 500 mL/min Connection: 6 mm (¼‐in.)
Drain: pressure‐free, atmospheric drain
Body material| Stainless steel 316L
Dimensions| Æ 50 mm, L 175.8 mm
Weight| Approx. 1 Kg (2.2 lbs)
Power Supply| Voltage: 12 VDC (powered by 3S‐PC1000 controller)
Power consumption: max. 0.5 VA
Outputs| ModBUS RTU RS485
Installation| With optional fast‐loop reservoir (not included), pipe‐mounted
or wall mounted with appropriate brackets
Protection Grade| IP68
2.2 Instrument description
The 3S‐UVO is a sensor for water monitoring. The sensor works with the
principle of UV fluorescence and can be used to detect many common pollutants
such as hydrocarbons/oil‐in‐ water, BTEX, PAH/PAC. In many cases the
measurements can be correlated with sum parameters like Code, Toes, Bode. The
design is compact and robust, the stainless steel body offer great protection
up to a depth of 60 m.
2.3 Applications
The instrument can be used to monitor the concentration of various parameters
in aqueous samples and it finds application in civilian and industrial
wastewater control, oil industry, hydrocarbons transport and storage and every
other application that requires a fast and sensitive determination of possible
pollutants.
2.4 Method description
The probe is sensitive to dissolved organic matter, through absorption
measuring in the UV region.
Most organic molecules in aqueous solution have spectral characteristics
capable of absorbing a fraction of energy associated with a light beam to
which it is exposed. This characteristic is due to the presence of chromophore
groups (aromatic bonds, double
covalent bonds and triple bonds) which is typical of organic substances.
Therefore, the greater the amount of organic molecules dissolved in the volume
of the measuring cell, the less the light intensity, at this specific
wavelength, that will reach the detector.
The intensity of the absorbance is measured by using as a reference the
intensity detected for another wavelength (in the visible region) which is not
influenced by the presence of organic substances.
This photometric investigation method allows, using the Lambert‐Beer law, to
calculate the absorbance and after calculating a calibration curve, to
determine the concentration of the organic substances expressed as TOC, DOC,
COD, BOD or as SAC254 Absorption coefficient.
The related results are expressed in concentration units (mg/l, ppm, ppb) for
cumulative parameters obtained by correlation or extinction per meter m (1/m)
as required by DIN 38404‐ 3 standard for SAC254.
The ultraviolet light source of the probe is a highly‐stable LED.
INSTALLATION
3.1 Opening the package
For safety reasons, when removing the packaging of the equipment, please check
for any visible defects and, if necessary, inform the supplier.
Parts inside the package apart from the user manual:
A 3S‐UVO UV254 absorption sensor for organic load
B Probe cable (6 m)
3.2 Product code
The product code is an alphanumeric code that identify your 3S Analyzers
product and its configuration. For the 3S fluorescence sensor the code is the
following:
3S‐UVO‐X
X = | L | Low range COD 0 ‐ 370 mg/l eq. KHP / TOC 0 ‐ 150 mg/l |
---|---|---|
H | High range COD 0 ‐1000 mg/l eq.KHP / TOC 0 ‐ 400 mg/l |
3.3 Wall mounting dimensions
Recirculating sample reservoir cod. A46U10020 is included in the scheme as a
reference, has to be purchased separately.
3.4 Mounting the instrument (example with 3S‐PC1000)
The 3S‐PC1000 controller and the sample reservoir must be mounted
vertically on a wall or support suitable for their weight and not subject to
vibrations. Use suitable screws (not included in the supply) and fasten them
only on the side brackets (ear clips) of the instrument and in the holes of
the tank metal plate. Mount them so as to get the display at eye height (160
cm, 63 in).
Since the probe connections and flow sensor connectors are on the right side of
the analyzer, install sample reservoir underneath the analyzer, in a way that
is reachable from the right side. Please, also consider that the surrounding
space must allow easy opening of the analyzer door and easy access to the
sample reservoir for cleaning or mantainance.
A minimum distance of 10 cm is required between the sides of the instrument
and any other obstacle.
The sample reservoir (cod. A46U10020) can be mounted preferably under the
instrument. The sample line must be connected to the inlet below the
container, optionally a flow sensor can be installed on the same line to detect
the presence of the sample.
The reservoir has a side arm to drain the excess liquid and to maintain a
constant sample flow.
The side arm must be connected to the drain.
When the container is installed in a proper position the probe can be inserted
into its slot and secured with the clamp.
Finally, attach the probe connector to the analyzer.
CALIBRATION
4.1 About the method
The probes are calibrated using standard solutions which are analyzed in
the same way as the sample.
In order to ensure correct measurement performance, the probes should be
calibrated periodically, best results are obtained if they have been recently
cleaned and serviced.
Due to the nature of some analtycal methods the concentration/signal plot is
not linear in the whole range of our interest. Therefore the analyzer uses a
multi‐point calibration curve. The first point is the blank (zero), which is
usually done by analyzing demineralized water. A part from the blank, other
points are needed for the calibration curve, covering the whole range of
interest. In case the calibration curve is linear only a blank and a span
points are needed.
The probe can be calibrated internally (via the ModBUS interface) or
externally (via the controller).
The probe is commonly calibrated in COD(KHP), to correlate the measurement to
the actual sample value the probe can be further calibrate with a process
calibration.
4.2 Internal calibration
The probe has the possibility to store blank and slope values. If a linear
calibration is enough, the probe can be calibrated internally via the ModBUS
interface (see section 3.7).
After collecting the values for blank and span, trace a calibration curve and
take note of offset and slope values. Overwrite the corresponding registers to
calibrate the probe.
The probe will now use the new calibration curve.
4.3 External calibration
If a non‐linear calibration curve or deeper data analysis are needed, the
probe can be calibrated using an external controller.
In this case the logic must be implemented in the controller. The 3S‐PC1000
controller is able to manage two probes with independent calibration curves,
up to five points each.
Please refer to the 3S‐PC1000 or other controller user manual for the
instructions to perform a multi‐point calibration with the 3S‐UVO probe.
4.4 Blank calibration
The blank calibration is simply performed by analyzing demineralized water.
The blank calibration is particularly sensitive to impurities so is advisable
to thoroughly clean the probe before starting the calibration.
To perform a blank calibration, submerge the probe in pure water. If the
result is stable, take note of it.
4.5 Span calibration
The instructions shows a multi‐point calibration. Single‐point calibration is
performed in the same way.
Some probes require a non‐linear calibration. In that case we need to make
measurements at different concentrations to draw the calibration curve. If the
probe response is linear only one point is necessary.
As an example the next paragraph describes a 5‐point calibration with a
COD(KHP) standard solution.
COD(KHP) is the measure of the chemical oxygen demand when the solute is
potassium hydrogen phthalate (KHP). After the calibration with the standard
solution the measurement can be converted in the final measurement unit with a
process calibration.
To perform a multi‐point calibration, proceed as follow:
Chemicals
- Potassium hydrogen phthalate (KHP)
- Demineralized water
Instrumentation
- 2 L glass beaker
- 1 L volumetric flask
- 100 mL measuring cylinder
Stock solution
Prepare a stock solution (10,000 mg/l COD) of KHP by weighting 8.5 g and
inserting it into the
- L volumetric flask, fill the flask with demineralized water upto the flask mark.
Blank
The blank is measured using pure water. If the application is not particularly
demanding, tap water can be used. The blank is the first point of our
calibration curve.
Standard
We need other four points in addition to the blank. It is better if the four
points cover the full scale.
Procedure
Fill the 2 L beaker with 2000 ml (precisely measured) of the blank water.
Clean the probe with a clean cloth and a drop of isopropanol. Place the probe
into the beaker, dipped 5 cm into the solution, at least 5 cm from the
container walls. You can use a stand to help the probe stay in position. Make
sure no air bubbles are trapped below the sensor.
Turn on the stirrer at 500 rpm and wait at least 90 seconds for the sensor to
give a stable reading. Take note of the sensor reading, this is your blank.
Now, prepare the other standard solutions. In this example we will calibrate
the 0 ‐ 300 mg/l COD probe, we’ll need to prepare four standard solutions that
will cover the whole range.
From the 10,000 mg/l standard solution take the amount listed in the table
below and add to the beaker.
Concentration (mg/l) | Volume addition (ml) |
---|---|
0 | 0 |
75 | 15 |
150 | 15 |
225 | 15 |
300 | 15 |
For every addition, wait for the stabilization of the measurement then take
note of it.
When you have collected all the points, insert them in the controller (follow
the controller instruction on how to do it).
Note: the increase in the total volume bears a small error in the final
concentration which is usually negligible. If a more accurate calibration is
needed the operator must prepare fresh standard solution for every calibration
point.
Eventually, go to CONFIGURATION > CALIBRATION DATA and fill the calibration
table with the collected data.
The calibration is now complete.
4.6 Process cal
Once the probe is calibrated with the standard solution we can proceed
further and align the measurement to process parameters such as TOC, COD, BOD.
These parameters depends on the nature of the specific substances contained in
the sample, therefore it is not always possible to prepare standard solutions
for them.
What we can do is compare the measurement to a laboratory analysis (cuvette
tests or other analytical methods), calculate a conversion factor and apply to
the probe measurement.
The process calibration can be performed through the following steps:
- Take a sample representative of the water stream to be analyzed, at least 1 liter. Follow good sampling techniques to have reliable results.
- Determine the concentration of the parameter of interest (COD, TOC, BOD) using a reference instrument or a laboratory analysis of the sample .
- Submerge the probe in the sample and wait for a stable result.
- Calculate a multiplication factor then follow the instructions of the controller to apply it to the measurement.
- The calibration is now over, the following measurements will be calculated with the new process factor.
3S Analyzers S.r.l. Italy
www.3s‐analyzers.eu
November 2021
3S‐UVO User Manual v1.0
Documents / Resources
| 3S
ANALYZERS 3S-UVO UV254 Absorption Sensor for Organic
Load
[pdf] User Manual
3S UVO UV254 Absorption Sensor for Organic Load, 3S UVO, UV254 Absorption
Sensor for Organic Load, Absorption Sensor, Sensor
---|---
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>