AIR BOX P02008 Personal Air Sampling Pump User Manual
- October 30, 2023
- AIR BOX
Table of Contents
AIR BOX P02008 Personal Air Sampling Pump
Product Information
- Product Name: XL
- Flow Range: Adjustable 5 -17 l/min
- Battery Pack: 14.4 Volt 26.Ah Lithium Li-on
- Charge Time: 6-8 hours
- Duration: 10+ Hours @ 16 l/min, 22+ Hours @ 8 l/min
- Display: OLED, elapsed run time and battery indicator
- Material: HF ABS
- Waterproof: IP65
- Dimensions: H38cm x W24cm x D12cm
- Mast extends: (Up to) 2 meters
- Weight: 4.2kg
- Certification: CE Marked
Product Usage Instructions
- Before using the XL pump, ensure it is fully charged. A battery level indicator will show the available power when the unit is first switched on.
- To charge the battery, plug the charger into the charging socket. A red LED will be illuminated. The LED will change to green when the battery is fully charged. It takes approximately 6-8 hours to completely charge the battery.
- Before switching on the pump, ensure any yellow waterproof covers are removed from the air outlet and from the bottom of the air sampling head.
- If the pump is shut down between setting the flow and starting the test, the flow will have to be re-calibrated by repeating the steps above.
- To set the test mode, press start/pause. The display will show ‘Set test mode’. Press the arrows to select ‘Free Running’ or ‘Timed test’.
- If ‘Free Running’ is selected:
- Set the flow rate using the arrows.
- Press start/pause to begin testing.
- If ‘Timed Test’ is selected:
- Set the flow rate using the arrows.
- Set the desired test time using the arrows.
- Press start/pause to begin testing.
- If the pump switches itself off during either Free Running or Timed Test (i.e. if the battery dies or the sampling head gets blocked) on restart it will either show:
- ‘Test Incomplete, Restart Test?’. Press start/pause to resume test and see where the previous test got to.
- ‘Test Completed Check Flow’, in which case continue as per instructions.
- ‘Low battery’, in which case replace the battery within 10 minutes of disconnecting the old one.
Specifications
Specifications:- (tested with 8.0μm MCE filter)
- Flow Range: Adjustable 5 -17 l/min
- Battery Pack: 14.4 Volt 26.Ah Lithium Li-on
- Charge Time: 6-8 hours
- Duration: 10+ Hours @ 16 l/min, 22+ Hours @ 8 l/min Display: OLED, elapsed run time and battery indicator Material: HF ABS
- Waterproof: IP65
- Dimensions: H38cm x W24cm x D12cm
- Mast extends: (Up to) 2 meters
- Weight: 4.2kg CE Marked
Warranty Registration
Don’t forget to register your new Airbox product by filling in the
registration form provided and returning it to us at the address below.
Alternatively, you can do this online at:
www.airboxsp.com
Technical support
For all technical support or customer service inquiries
please contact us:
- telephone: +44(0)1903 524 600
- email: sales@airboxsp.com
Airbox Sampling Products, Technology Centre, Easting Close, Worthing, West Sussex BN14 8HQ Thank you for choosing this Airbox product. To ensure that you get the best possible use from it, please read this manual carefully. Airbox Sampling Products is part of the FermionX Group
Setting up and using the XL pump
Before you use the XL pump make sure it is fully charged. A battery level indicator shows the available power when the unit is first switched on.
Charging the battery
When the charger is plugged into the charging socket a red LED will be
illuminated. The LED will change to green when the battery is fully charged.
It takes approximately 6-8 hours to completely charge the battery.
Removing the waterproof covers
IMPORTANT: Before switching the unit on, ensure any yellow waterproof covers are removed from the air outlet and from the bottom of the air sampling head.
Switching on and setting the flow rate
-
Before switching on the XL, connect a flow meter to the air sampling head.
-
Press start/pause to switch on the unit. The display will show, in this order, the serial number, the current firmware version and the battery status.
-
If start/pause is pressed the display shows ‘Flow not set’.
-
Press either arrow to open and start Calibration Mode and to Set Flow.
-
The pump is set at aprox 8 l/min as a default, to fine-tune this:-
-
Press start/pause to start calibration.
NOTE: Run for 45 seconds to allow to stabilize. Press up/down arrows to adjust flow rate. (A single press advances the display by one increment, holding while pressing advances the display by 10 increments). Check the flow rate against the flow meter and adjust it according to flow requirements. (Any increment changes will be saved for the post-test flow check, see 2.3). -
To confirm and set the flow rate press stop.
IMPORTANT: If the pump is shut down between setting the flow and starting the test, the flow will have to be re-calibrated by repeating the steps above.
Setting test mode
- Press start/pause the display will show ‘Set test mode’.
- Press the arrows to select ‘Free Running’ or ‘Timed test’.
If Free Running is selected
- Press start/pause the pump runs and the timer starts counting, (you can pause during the test by pressing start/pause)
- Press stop during the test and the display shows ‘Test Stop? Are you sure?’ If you are not sure, press start/pause to continue test, if you are sure:-
- Press stop display shows ‘Test completed – Check Flow’.
- Press stop display shows ‘Connect flow meter’ (now reconnect your flow meter to check the flow rate).
- Press start/pause, the display shows ‘Flow Check’ (any increment changes made when setting the flow rate (see 2) will be shown).
NOTE: Wait 45 seconds for the flow rate to stabilize.
- Press stop display shows ‘Flow Check Complete’.
- Press stop to shut down.
If Timed Test is selected
- Press start/pause the display shows ‘Enter Run Time’
- Press arrows to adjust time:-
- Press start/pause to move between (Hrs) 00 (Mins) 00 (Sec) 00.
- Press start/pause the pump runs and the timer starts. When the time has elapsed the display shows ‘Test Completed – Check Flow’.
- Press stop display shows ‘Test completed – Check Flow’.
- Press stop display shows ‘Connect flow meter’ (now reconnect your flow meter to check the flow rate).
- Press start/pause display shows ‘Flow Check’ (any increment changes made when setting the flow rate (see 2) will be shown.
NOTE: Wait 45 seconds for the flow rate to stabilize.
- Press stop display shows ‘Flow Check Complete’
- Press stop to shut down.
Note: If the pump switches itself off during either Free Running or Timed Test (i.e if the battery dies or the sampling head gets blocked) on restart it will either show:-
- ‘Test Incomplete, Restart Test?’, Press start/pause to resume test and see where the previous test got to.
- ‘Test Completed Check Flow’, in which case continue as instructed above.
- ‘Low battery’, in which case either replace the battery (within 10 minutes of disconnecting the old one).
Maintenance
Replacing the battery
The XL pump has been specifically designed to make replacing the battery a
quick and easy task. Simply undo the two screws on the front of the battery
housing, pull off the terminal connectors, remove the battery, and replace.
- NOTE: This process is made easier if the pump is lying on its back.
- IMPORTANT: The terminals on the battery are color-coded red and black to match the red and black wires of the unit. When fitting a new battery make sure that these match up correctly.
Replacing the mast
The mast is secured on the unit in a threaded housing, so to replace just
unscrew and remove it. We would also recommend removing the mast when cleaning
the unit (3.3).
Cleaning the unit
The casing of the XL pump is made from a hard-wearing ABS structural foam,
making it incredibly rugged and also very easy to clean. Tested to IP65
standards (‘Protected against low-pressure water jets from any direction.’)
the unit can be cleaned under a shower hose.
NOTE: Do NOT Submerge the product fully in water.
Servicing & Warranty
The XL pump is covered by our exclusive warranty and in the event that the
unit fails we will undertake the repair free of charge for up to 2 years from
the date of purchase. The warranty covers the costs of materials and labour
(with the shipping costs to be covered by the customer) and is valid under the
following conditions:-
- Please complete and return the warranty registration card.
- The defect must not have been caused by improper handling or misuse.
- Return the unit to us in the decontamination bag provided by Airbox. We will not assume responsibility for transport damage under any circumstances.
- A short description of the defect must be included with the returned unit. A full warranty statement is available on our website or from our customer service team.
Troubleshooting
The pump will not start, the red LED illuminated
Battery voltage may be too low, check the battery level and recharge the
battery. If the pump still doesn’t start, contact the Airbox service
department.
The pump will not start, no display
Flat battery, recharge or replace the battery.
The flow rate doesn’t reach the required level
- Ensure any yellow waterproof safety caps have been removed.
- Check the tube for kinks and that the sampling head is not blocked.
- If the flow rate is still insufficient, contact the Airbox service department.
The pump stops while running, red LED illuminated
- Check the tube for kinks and that the sampling head is not blocked.
- Check the battery level and recharge if required. Replace the battery if necessary.
- If the flow rate is still insufficient, contact the Airbox service department.
The pump fails to run for a specified duration
Check the battery level and recharge if required. Replace the battery if
necessary.
Parts
Product Code | Name | Description |
---|---|---|
P02008 | Variflow XL | VARIFLOW XL Pump |
P02008/KIT | Variflow XL Kit | Sampling pump + Kit (incl head, spare mast) |
P01839 | Sampling head | AIRBOX SAMPLING HEAD + 2x O-rings + |
2x Filter support pads
P029347LF| Charger for Variflow 1Ah| Charger for P029347LF – 12V 1Ah Lead acid
SLA – (Choose lead)
PT27259| Battery for Variflow/ V8/Lowflow| Battery for Variflow/V8/Lowflow –
Lead acid 12V 9Ah Recharge
PT29482LF| Variflow XL mast| Telescopic mast for Variflow – 2m
PT28297LF KIT| Cap + plug Kit (x10)| Sampling head cap + plug Kit (pack of 10)