mintti health Breeze Infrared Thermometer User Manual

June 10, 2024
mintti health

mintti health Breeze Infrared Thermometer User Manual
mintti health Breeze Infrared Thermometer

Introduction

Breeze measures the body temperature through receiving the infrared energy radiation from the human body. There are two modes available for the device, forehead temperature mode and ear temperature mode.
The measurement result will be directly shown on the LED screen. The main components of the product include infrared temperature sensors, signal receiving processor, buttons, buzzer, LED display, battery, etc.

Product intended use/Indications for Use

The infrared thermometer is intended for the measurement and monitoring of human body temperature by physician in healthcare environment or lay person at home. The ear mode of the device is not intended for infants less than three months.

Contradictions: There are no known cortrnindications
Side-effects: Not applicable.

Safety Information

Warnings:

  • The device is not intended for Professional Use.
  • The measurement results cannot replace the physician diagnosis.
  • Read the instruction manual before using this device.
  • The thermometer should be placed where children can’t get access.
  • Do not measure against the eyes.
  • Do not try to change the product default settings.
  • Use the thermometer in a stable temperature environment. If the environment temperature changes too often, for example from outdoor to indoor, please put the thermometer indoor for about half an hour before measuring.
  • Do not measure body temperature immediately when the current two measured temperatures differ significantly, but 10 minutes later to perform the measurement.
  • Keep the thermometer away from any liquid to avoid any damage to the device.
  • Do not use in high or low temperature environment for a long time.
  • Do not collide, and avoid any falling and disassembly by the user
  • Do not use in strong electromagnetic interference environment.
  • Do not operate the thermometer when the ear is not clean, or the skin of the forehead is sweating.
  • The dispose of the thermometer should be accorded to local regulations
  • Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally
  • Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this device could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation
  • Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the device, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.

Principle

Any object with the temperature of higher than absolute zero degree will transmit some infrared radiation energy according to its own temperature.
The radiation energy and its distribution per wavelength are closely associated with its surface temperature. Based on the principle, it is possible to measure the forehead surface temperature and then adjust the offset between forehead temperature and actual body temperature, which will realize the correct display of body temperature.

Product features

  • Dual modes of measurement: Forehead temperature, ear temperature. ;
  • High definition green color LED backlight, clearly and softly display;
  • High temperature indication function (default beeps when temperature is 38.0°C and above);
  • Store 32 sets of measurement data;
  • Dual temperature units: °Cand °F;
  • Automatic shut-down and power-saving;
  • Sound ON/OFF function, which acts as an indication if the measurement is finished;
  • Sound can be muted by a continuous pressing for 4s after the device is on.

Technical Parameters

Measurement Mode Ear Mode 34.0°C – 42.4°C ( 93.2°F-108.3°F)
ForeheadMode 34.0°C – 42.4°C ( 93.2°F-108.3°F)
Measurement accuracy* Ear Mode ±0.2°C/0.4°F
ForeheadMode ±0.2°C/0.4°F
Display resolution 0.1°C/°F
Operating environment 16° C – 35 ° C (60.8° F to 95 ° F)5 85% moisture

noncondensing
Storage condition| X20 °C- 55 ° C (-4° F to 131 ° F)= 90% moisture noncondensing
Power supply| DC1.5Vx2 AAA
---|---
Power consumption| When off 5 10 uw
When measuring< 200 mw
Auto power-off| 30 seconds
Weight| About 62 g (without battery)
Dimensions| 161.5mm 35.8mm 55.9mm(length x width x height)
Product accessories| 2x AAA Batteries, lx User Manual,lx Maintenance & Service card and lx QC Certificate

  1. Clinical accuracy characteristics and procedures are available from the manufacturer on request.
  2. “ASTM laboratory accuracy requirements in the display range of 37 to 39 °C (98 to 102 °F) for IR thermometers is ±0.2 °C (±0.4 °F), whereas for mercury-in-glass and electronic thermometers, the requirement per ASTM Standards E667-86 and E1112-86 is ±0.1 °C (±0.2 °F).”

Shape Structure

Shape Structure

  1. Forehead cover;
  2. Infrared probe;
  3. Measuring key and Memory key;
  4. LED display;
  5. Conversion key and ON/OFF
  6. Handle
  7. Battery cover
Function Key Functional Description
Measuring key Press to measure the temperature
Memory key Press to get the current 32 sets of measuring data
Conversion key Long press to adjust °C / °F
ON/OFF key Long press the key, device will be power off

Note: Forehead cover is a reusable component of the thermometer.

Definition of display Icon

Icon definition Icon Definition
Measuring model Forehead temperature measurement mode
Measuring model Ear temperature measurement mode
Memory symbol Record the current 32 sets of

measurement data
Temperature unit| | Celsius degree; Fahrenheit degree
°F and °C is convertable
Battery capacity| | Indicates low battery status
Memory symbol| | Record the current 32 sets of
measurement data
| Caution
| The electronic waste should professionals after| be disposed by use
| Manufacturer
| BF type application device
| Keep dry
| Upward
| Fragile, handle with care
IP22| 2 Protected against solid foreign objects of
012.5 mm and greater ;
2 If keepin the thermometer in 15 degree angle, the water drops can still be prevented.

Settings

A: Memory mode:

Take off the bottom cover and install the battery, then all symbols will be displayed on the screen. Then press the Memory key to check the current 32 sets of measurement data.
Note:a. The record order is 1,2,3…20, 21,…,32.
b.The record will be deleted when uninstalling the battery.

B: Temperature unit switch setting:
When the device is off, press the memory key for Ss until the LED display “– – °C or “- – – – °F “, then press the memory key one more time to switch to the right unit. Then the device will be power off automatically after 3s, which means the unit setting is done.

Note: The default temperature unit is “°C. The setting will come back to the default when uninstalling the battery.

C: Voice switch setting:
When the device is on, long press the Memory key for 4s, then the device will beep once, and the voice can be set ON or OFF.

Note: The sound key can be set on/off depending on the situation. e.g. it can be off when silence is appreciated.

D: Shutdown key:
When the device is on, long press the Memory key for about 2s, it will be powered off automatically

Measuring method

Forehead temperature measurement.

  • Install the batteries and press the ON/OFF key, and make sure the icon shows up before the measurement. Be sure to keep the forehead cover not removed before the measurement.
  • Hold the thermometer and near the temporal artery of forehead, then press the measurement key. When buzzer beeps, the temperature will be displayed on the LED screen.
    The thermometer must be kept near the forehead before the end of the measurement.
    Measuring Method

Ear temperature measurement

  • Install the batteries and press the ON/OFF key

  • Take off the forehead cover, and make sure the “ ” icon shows up before the measurement.
    Be sure that the forehead cover is removed and converted to the ear measuring mode.

  • Press the measuring key after inserting the probe into the ear. When the buzzer beeps, the temperature will be displayed on the LED screen.
    Ear Temperature Measurement

Warning

  • Do remember to put on the disposable probe cover before the measurement.

  • Do remember to change the disposable probe cover after each measurement or for different person.

  • Do not push hard into the ear to avoid hurting the eardrum.

  • Only use probe covers that are Melodicare-approved for use with this device, Otherwise, the results will be inaccurate.

  • • When the measured temperature value is 38°C or above, the thermometer will send out an indication and the backlight will turn red.
    Note:

    1. If no indication can be observed when using, please check whether the sound function is turned off, or contact the service in time.
    2. If thermometer produces a false indication, especially when used for infants or children, please contact the service in time.
  • Keep the battery beyond the child’s reach and prevent it from being swallowed. If the battery is swallowed, please contact the doctor immediately.

  • The measurement results of the thermometer is only for reference and cannot replace the clinical diagnosis result. If feeling unwell, please consult the doctor in time.

Caution

  1. All memory records will be lost when uninstalling the battery.
  2. All settings will come to default when uninstalling the battery. If the settings need to be adjusted please install the battery and renew the settings.
  3. The thermometer will be power off automatically after 30s without any operation.
  4. When the measurement value is between 37.5-37.9°C, it is defined as low fever, please consult the doctor for further diagnostic.
  5. When there is any wound in the ear measurement area, please do not use the thermometer to avoid aggravating the severity of the wound.

Replacement of batteries

  • • The battery life is approx. 3000 readings per year and 200 minutes when in constant use.
  • • Take off the bottom cover, and remove the old battery.
  • • Pay attention to the positive and negative when installing a new battery.

Caution

  1. It it’s not to ue used for one month, please remove the battery to prevent any leakage.
  2. Do not put waste battery in the fire to avoid any exploding.
  3. The dispose of the waste battery should be accorded to local regulations.

Cleaning and maintenance

Use only the Melodicare-approved substances and methods listed in this chapter to clean or disinfect your equipment. Warranty does not cover damage caused by using unapproved substances or methods.
Melodicare has validated the cleaning and disinfection instructions included in this User Manual. It is the responsibility of the healthcare professional or lay person to ensure that the instructions are followed so as to ensure adequate cleaning.

General Points

Keep your thermometer free of dust and dirt. To prevent the device from damage,please follow the procedure:

  • Use only recommended cleaning substances and disinfectants listed in this manual. Others may cause damage (not covered by warranty), reduce product lifetime or cause safety hazards.
  • Unless otherwise specified, do not immerse any part of the device in liquid.
  • Do not pour liquid onto the system.
  • Do not allow liquid to enter the case.
  • Never use abrasive material (such as steel wool or silver polish).
  • Inspect the device after they are cleaned and disinfected.

Warning

  1. If you spill liquid on the probe sensor or LED, or they are accidentally immersed in liquid, contact your service personnel or service engineer
  2. Before cleaning, make sure that the device is switched off and disconnected from the power line
  3. After cleaning with Isopropanol, allow the thermometer to air dry. Use a soft, dry cloth to clean the thermometer body and LED display.
  4. Do not use abrasive cleaners. Never submerge this thermometer in water or any other liquid.
  5. Please keep the inner cavity of the sensors and probes clean, otherwise it will affect the measuring accuracy

If the point(s) has been in contact with the patient without the replaceable ear coverco-packaged, e.g. surface of infrared probe and protective cover, then cleaning is required after each use to avoid contamination.
If there has been no patient contact and there is no visible contamination, including the handle, battery cover and compartment LED, keys and inside and outside surface of the protective cover, daily cleaning may be not necessary, and the suggestion is to do it weekly. But cleaning after each patient is recommended.
Water should be applied and removed using a clean, soft, non-abrasive cloth or paper towel.

To do initial cleaning to the thermometer, please follow these steps:

  1. Switch off the thermometer and uninstall the battery
  2. Wipe the entire exterior surface of the thermometer for two minutes by using a soft cloth which has been dampened with the cleaning solution thoroughly for 30s until no visible contaminants remain under room temperature.

Point-of-Use Processing

If the point(s) has been in contact with the patient without the replaceable ear cover co-packaged, e.g. surface of infrared probe and protective cover, then cleaning is required after each use to avoid contamination.
If there has been no patient contact and there is no visible contamination, including the handle, battery cover and compartment LED, keys and inside and outside surface of the protective cover, or when the probe is used with the replaceable ear cover, daily cleaning may be not necessary, and the suggestion is to do it weekly. But cleaning after each patient is recommend-ed, and only the legally market protective cover suggested by the manufacturer can be used with the device to make sure the measurement accuracy of the device.
Water should be applied and removed using a clean, soft, non-abrasive cloth or paper towel.

To do surface-cleaning to the thermometer, please follow these steps:

  1. Switch off the thermometer and uninstall the battery
  2. Wipe the entire exterior surface of the thermometer for two minutes by using a soft cloth which has been  dampened with the cleaning solution thoroughly for 30s until no visible contaminants remain under room temperature.

Thorough Cleaning

The aevice shoula oe tnoroughly cleanea atter the point-of-use processing after each use.
The validated cleaning agents for cleaning the monitor and reusable accessories are:
Isopropanol (70%)
Cleaning substances should be applied and removed using a clean, soft, non- abrasive cloth or paper towel.

To do surface-cleaning to the thermometer, please follow these steps:

  1. Switch off the thermometer and uninstall the battery
  2. Wipe the entire exterior surface, including the handle, battery cover and compartment, LED, keys and inside surface of the protective cover, of the equipment for five minutes using a soft cloth which has been dampened with the cleaning solution thoroughly for 30s until no visible contaminants remain under room temperature.
  3. Wipe off the cleaning substances for three minutes with a fresh cloth or towel, dampened with tap water for 30s after cleaning until no visible cleaning substances remain under room temperature.
  4. Dry the thermometer in a ventilated and cool place for 10 minutes under room temperature.

Troubleshooting

Diagnosis Solution

The screen shows
“Lo” or “Hi”| 1.Check the measurement site. It is unable to measureaccurately when the forehead is with water, sweating, hair-covered or applying cosmetics.2.Check the operation environment.Environment changes will greatly influence the measurement result.Temperature change is significant. It’s better to use after 10 minutes to achieve a new heat balance.4.    The measured body temperature is higher than 42.4° C or lower then 34.0 ° C5. The measured object is not correct.
Screen flashes
after
measurement| The environment temperature is not among the normal working arrange which is 16° C – 35 ° C ( 60.8° F to 95 ° F)
Buttons have
no response| 1.Install and uninstall the battery 2.Check if the setting is in the progress
No display or
display abnormal| Uninstall the battery and re-install
Power off when open| Check the battery, uninstall and re-install the battery

EMC Information

Guidance and manufacturer ‘s declaration-electromagnetic emissions

BREEZE is intended to be used in the electromagnetic environmentspecified below. The customer or the user of BREEZE should assurethat it is used in such an environment.
Emissions test| Complianc| Electromagnetic environment-guidance
RF emissionsCISPR 11| Group r| BREEZE uses RF energy only for its internalfunction. Therefore, its RF emissions are verylow and are not likely to cause anyinterference in nearby electronic equipment.
RF emissionsCISPR 11| Class B| BREEZE is suitable for use in all establish- ments other than domestic and those directlyconnected to the public low- voltage powersupply network that supplies buildings usedfor domestic purposes.
Harmonicemissions IEC 61000-3-2| N/A
Voltagefluctuations/flickeremissions IEC 61000-3-3| N/A
Guidance and manufacturer ‘s declaration-electromagnetic immunity

BREEZE is intended to be used in the electromagnetic environmentspecified below. The customer or the user of BREEZE should assurethat it is used in such an environment.
Immunity test| IEC60601test level| Compliancelevel| Electromagneticenvironment -guidance
Electrostaticdischarge(ESD)IEC 61000-4-2| ±2,±4,±6kVfor Contactdischarge+2 +4 +8kVi–        /-air discharge| ±2,±4,±6kVfor Contactdischarge±2,±4,±8kVair discharge| Floors should be wood,concrete or ceramic tile. Iffloors are covered withsynthetic material, the relativehumidity should be at least30%
Electrical fasttransient/burstIEC 61000-4-4| ±2 kV for a.c.power lines±1 kV for d.c.power lines| N/A| Mains power quality should bethat of a typical commercial orhospital environment.
SurgeIEC 61000-4-5| ±1 kV line(s)to line(s)+2 kV line(s)to earth| N/A| Mains power quality should bethat of a typical commercial orhospital environment.
Voltage dips,shortinterruptionsand voltagevariations inpower supplyinput linesIEC 61000-4-11| <5%UT(>95     infor 0.5dip     UT) cycle40% UT(60% dip in UT)for S cycles(30% dip in UT)for 25 cycles<5% UT (>95%dip in UT for 5s| N/A| 70%  UTMains power quality should bethat of a typical commercial orhospital environment. If theuser of BREEZE requirescontinued operation duringipower mains interruptions, it isrecommended that BREEZEbe powered from anuninterruptible power supplyor a battery
Powerfrequency(50/60Hz)magnetic fieldIEC 61000-4-8| 3A/m| 3A/m| Power frequency magneticfields should be at levelscharacteristic of a typicallocation in a typicalcommercial or hospitalenvironment
NOTE UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer ‘s declaration-electromagnetic immunity

BREEZE is intended for use in the electromagnetic environmentspecified below.
Immunity test| IEC 60601 test level| Compliance level| Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6| 3Vrms 150kHz to 80MHz| N/A| Portable and mobile RF communications equipment should be used no closer to any part of BREEZE including caoles, than the recommend- ed separation distance calculated from the equation applicable to the frequency of the transmitter.Recommended separation distance
Radiated RF
IEC 61000-4-3| 3V/m
80kHz to 2.5GHz| 3V/m| d=1.2 P 80MHz to 800MHz d=2.3 P 800MHz to 2.5MHz Here P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths form fixed RF transmitters, as determined
by an electromagnetic site
survey, should be less than the compliance b level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:

NOTE 1 At 90MHz and 800MHz, he higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnet- ic propagation is affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, anelectromagnetic site survey should be considered. If the measured field strength in the location in which BREEZE is used exceeds the applicable RF compliance level above, BREEZE should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating BREEZE
b
Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
Recommended separation distances between portable and mobile RF communications equipment and BREEZE

BREEZE is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of BREEZE can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and BREEZE as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter| Separation distance according to frequency of transmitterm
Ratedmaximumoutput power| 150kHz to 80MHz
d -1.2 P| 80MHz to 800MHz
d =1.2 P| 800MHz to 2.5GHz
d =2.3 P
0.01| 0.01| 0.12| 0.23
0.1| 0.1| 0.38| 0.73
1| 1| 1.2| 2.3
10| 10| 3.8| 7.3
100| 100| 12| 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.NOTE 1 At 80MHz and 800MHz, the separation distance for the higher frequency range applies.NOTE 2 These guidelines may not apply in all situations. Electromag­netic propagation is affected by absorption and reflection from structures, objects and people.

Commitment to quality and after-sales service

The warranty period of the product is one year. Note: The free service during warranty period does not include the failure and damage due to user’s personal reasons or unauthorized disassem-bling.

Prompt
Please keep the purchase vouchers to facilitate future maintenance.

Manufacturer information

Company Name: Suzhou Melodicare Medical Technology Co., Ltd.
Address: Room 301, Building F2, No. 32 Hongxi Road, Suzhou
High-Tech Zone, Jiangsu Province, China.

Zip Code: 215000
Telephone: +86(0)791-85230171
Website: www.minttihealth.com