BRESSER 8855000 40x-1024x Microscope Set Instruction Manual

June 4, 2024
BRESSER

BRESSER 8855000 40x-1024x Microscope Set

TIPS on cleaning

Remove the device from it’s energy source before cleaning (remove plug from socket / remove batteries). Clean the exterior of device with a dry cloth. Do not use cleaning fluids so as to avoid causing damage to electronic components. Toys liable to be cleaned with liquid are to be disconnected from the power supply before cleaning
Clean the lens (objective and eyepiece) only with the cloth supplied or some other soft lint-free cloth (e.g. microfiber). Do not use excessive pressure – this may scratch the lens.

DISPOSAL
Dispose of the packaging material/s as legally required. Consult the local authority on the matter if necessary.  Do not dispose of electrical equipment in your ordinary refuse. The European guideline 2002/96/EU on Electronic and Electrical Equipment Waste and relevant laws applying to it require such used equipment to be separately collected and recycled in an environment-friendly manner.  Empty batteries and accumulators must be dis-posed of separately. Information on disposing of all such equipment made after 01 June 2006 can be obtained from your local authority.

EC Declaration of Conformity
Bresser GmbH has issued a „Declaration of Conformity“ in accordance with applicable guidelines and corresponding standards. The full text of the EU declaration of conformity is available at the following internet address:
www.bresser.de/download/8855000/CE/ 8855000_CE.pdf

UKCA Declaration of Conformity
Bresser GmbH has issued a „Declarati-on of Conformity“ in accordance with applicable guidelines and corresponding standards. The full text of the UKCA declaration of conformity is available at the following internet address: www.bresser.de/download/8855000/UKCA/8855000_UKCA.pdf
Bresser UK Ltd. • Suite 3G, Eden House, Enterpri-se Way, Edenbridge, Kent TN8 6Hf, Great Britain

Warranty & Service
The regular guarantee period is 5 years and be-gins on the day of purchase. You can consult the full guarantee terms and details of our ser-vices at www.bresser.de/warranty_terms.

Here are the parts of your microscope (Fig. 1-6)

  1. 10x WF Eyepiece
  2. 16x WF Eyepiece
  3. Barlow Lens
  4. MicrOcular
  5. Eyepiece supports
  6. Microscope Head
  7. Set Screw
  8. Objective Nosepiece
  9. Objective
  10. Clips
  11. Microscope Stage
  12. LED Illumination (transmitted light)
  13. Microscope Base
  14. Selection Knob for Illumination
  15. Power supply
  16. Focus knob
  17. Color Filter
  18. LED Illumination (reflected light)
  19. 5 Slides, 10 Cover Sips and 5 Prepared Specimens plastic box
  20. Specimens:
  21. Yeast
  22. Shrimp Eggs
  23. Sea Salt
  24. Glue
  25. Transformer with Plug and Power Cable
  26. Specimen Slicer
  27. Shrimp Hatchery
  28. Tweezers
  29. Pipette
  30. Carrying Case
  31. Illumination filter
  32. USB cable

How do I use my microscope?
Before you assemble your microscope, make sure that the table, desk or whatever surface that you want to place it on is stable, and does not wobble.
Apart from that, you will need a power source for your microscope’s light (220V-330V).

How do I operate the electric LED illumination?
For the illuminator, you need the transformer with the plug and power cable (22). First, place the small plug on the power cable into the power connection on the back of the microscope (Fig. 1, 15). Then, insert the large plug into a power outlet.
There are two lights on the microscope. They do not use light bulbs, but rather light-emitting diodes (LED). The first lamp shines onto the specimen from below and the second from above. (The thing that you want to observe with the microscope is called the object or specimen, by the way.) You can use each lamp on its own, or both of them together. There is a selection knob for this (Fig. 1, 14). It has three numbers: I, II and III.
If you select the …

  • I, the light only comes from below (transmitted light).
  • II, the light only comes from above (reflected light).
  • III, both lamps shine light on the specimen.

For transparent objects (transmitted-light objects), number I is best. In order to observe firm, non-transparent objects (direct-light objects), select number II. For semi-transparent objects, it is best to select number III.
It is not recommended to use number III for transmitted-light objects on slides, since the light may cause reflections on the surface of the slide, which will disturb your observation.

When do I use the color filters?
The color filters (Fig. 1, 17) are located below the microscope stage (Fig. 1, 11). They help you when you are observing very bright or clear specimens. Here, you can choose from various colors. This helps you better recognize the components of colorless or transparent objects (e.g. grains of starch, protozoa).

How do I adjust my microscope correctly?
First, loosen the screw (Fig. 1, 7) and turn the microscope head (Fig. 1, 6) into a comfortable viewing position. Each observation starts with the lowest magnification.
Adjust the microscope stage (Fig. 1, 11) so that it goes all the way down to the lowest position. Then, turn the objective nosepiece (Fig. 1, 8) until it clicks into place at the lowest magnification (objective 4x).
Note:
Before you change the objective setting, always move the microscope stage (Fig. 1, 11) to its lowest position. This way, you can avoid causing any damage!
Now, insert the 10x eyepiece (Fig. 1, 1) into the Barlow lens (Fig. 1, 3). Make sure that the Barlow lens is placed all the way into the eye-piece supports (Fig. 1, 5) and is not pulled out (Fig. 2).

How do I observe the specimen?
After you have assembled the microscope with the adequate illumination and adjusted it correctly, the following basic rules are to be ob-served:
Start with a simple observation at the lowest magnification. This way, it is easier to position the object in the middle (centering) and make the image sharp (focusing).
The higher the magnification, the more light you will require for a good image quality. Now place the prepared specimen (20) directly under the objective on the microscope stage (Fig. 3). The object should be located directly over the illumination (Fig. 1, 12). In the next step, take a look through the eye-piece (Fig. 1, 1) and carefully turn the focus knob (Fig. 1, 16) until the image appears clear and sharp.
Now you can select a higher magnification by slowly removing the Barlow lens (Fig. 1, 3) from the eyepiece support (Fig. 1, 5). When the Barlow lens is almost completely pulled out, the magnification can be increased to almost double. If you would like an even higher level of magnification, insert the 16x eyepiece (Fig. 1, 2) and turn the objective nosepiece (Fig. 1, 8) to a higher setting (10x or 40x).

Important tip:
The highest magnification is not always the best for every specimen!

Note:
Each time the magnification changes (eyepiece or objective change, pulling out the Barlow lens), the image sharpness must be readjusted with the focus knob (Fig. 1, 16). When doing this, make sure to be careful. If you move the microscope stage too quickly, the objective and the slide could come into contact and become damaged!

Which light for which specimen?
With this unit, a reflected light and transmitted light microscope, you can observe transparent, semi-transparent as well as non-transparent objects.
The image of the given object of observation is “transported” through the light. As a result, only the correct light will allow you to see some-thing!
If you are observing non-transparent (opaque) objects (e.g. small animals, plant components, stones, coins, etc.) with this microscope, the  light falls on the object that is being observed. From there, the light is reflected back and pass-es through the objective and eyepiece (where it gets magnified) into the eye. This is reflected light microscopy.
For transparent objections (e.g. protozoa), on the other hand, the light shines from below, through the opening in the microscope stage and then through the object.
The light travels further through the objective and eyepiece, where it is also magnified, and finally goes into the eye. This is transmitted-light microscopy.
Many microorganisms in water, many plan components and the smallest animal parts are already transparent in nature. Others have to be prepared. We may make them transpar-ent through a treatment or penetration with the right materials (media), or by taking the thinnest slices from them (using our hand or a specimen slicer), and then examine them. You can read more about this in the following sec-tions.

How do I make thin specimen slices?
Only do this with the supervision of your parents or another adult.
As I already pointed out, the thinnest slices possible are taken from an object. In order to get the best results, we need some wax or paraffin. It is best if you get a candle. Place the wax in a pot and heat it carefully over a low burner. Now, dip the object in the liquid wax a few times. Then, let the wax get hard. Using the specimen slicer (23) or a knife/scalpel, cut the smallest slices from the object that is covered with wax. These slices are to be laid on a slide and covered with a cover slip.

How do I make my own specimens?
Take the object that you want to observe and place it on a glass slide (20). Then, add a few drops of distilled water on the object (Fig. 7) using a pipette (Fig. 7, 26). Now, place a cover slip vertically at the edge of the drop of water, so that the water runs along the edge of the cover slip. Then, slowly lower the cover slip over the water drops (Fig. 8).
Note: The included glue “gum media” (Fig. 5, 21b) is used to make permanent prepared specimens. Use this in place of the distilled water. If you want to keep the object in place on the slide permanently, use the gum media.

Experiments
Use the following web link to find interesting experiments you can try out.
http://www.bresser.de/downloads
Microscope accessories can be found here: http://www.bresser.de/P8859480

How do I install the MicrOcular?
Note: The MicrOcular only works without the includ-ed Barlow lens! The magnification setting is changed when the MicrOcular is used, and it must be readjusted with the focus knob.
First, remove the Barlow lens (Fig. 4, 3) with the current eyepiece from the eyepiece support (Fig. 4) and place the MicrOcular (Fig. 5, 4) in the eyepiece (Fig. 5).

How do I install the software?

  1. You can download the setup file for installa-tion here: www.bresser.de/download/CamLabLite_Junior
  2. Start the installation with a double click on the .exe file and follow the instructions of the installation manager.
  3.  It is not necessary to install a special driver. The camera is automatically recognized by your Windows operating system.

Using the MicrOcular

Preparation
  1. Slide a specimen under your microscope and focus on it.
  2.  Remove the eyepiece and Barlow lens from the eyepiece support and the dust cap from your MikrOkular and then install the MikrOku-lar into the eyepiece support instead of the Barlow lens.
  3. Start your PC if you haven‘t yet and connect your MikrOkular to the USB port of your computer.
  4. Start the „CamLabLite_Junior“ software and activate the camera by clicking on „USB Vi-deo Device.
  5. Click on „Snapshot“ to take a picture or on „Capture“ to record a video.
  6. Click on the floppy disk icon to save the image or capture.

Troubleshooting

Error : Solution

  • No recognizable image
    • Turn on light
    • Readjust focus
    • Check Windows privacy settings
  • Image flickers (while observing with MicrOcular) = Monitor refresh rate not adequate
    • If necessary, increase resolution of graphics card
    • Adjust brightness at the dimmer of the microscope
    • Disable automatic exposure

Make sure your microscope has a long service life.
Clean the lens (objective and eyepiece) only with the cloth supplied or some other soft lint-free cloth (e.g.microfibre). Do not press hard as this might scratch the lens. Ask your parents to help if your microscope is really very dirty. The cleaning cloth should be moistened with cleaning fluid and the lens wi-ped clean using little pressure. Make sure your microscope is always protected against dust and dirt. After use leave it in a warm room to dry off. Then install the dust caps and keep it in the case provided.

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

BRESSER User Manuals

Related Manuals