JRI Nano SPY Wireless Mini Data loggers User Guide
- June 11, 2024
- jri
Table of Contents
Nano SPY Wireless Mini Data loggers
User Guide
INTRODUCTION
The Nano SPY is a measuring device for 1 or 2 physical parameters (T or TH
depending on the model). Data is transmitted through NANO Link or Relay/Alarm
modules using wireless 2.4GHz radio signal to the JRI-MySirius monitoring
software hosted on the JRI could or the customer’s server.
The Nano SPY complies with EN 12830 with temperature probes only, and is
compatiblewith EN 13486 which defines procedures for periodic verification.
a) Product contents
- 1 Nano SPY
- 1 User guide
b) Symbols
| RECYCLING: do not dispose of in a refuse dump or waste disposal bin.
Comply with existing legislation for disposal.
---|---
| Power source: this device is powered by a 3.6VDC type AA lithium battery (§
ch. V).
| CE LABELING: this device is certified to conform to European
regulations for electrical safety, flammability, disruptive electromagnetic
emissions, and immunity to environmental electrical disturbances.
**| FCC ID : W45 12525
This device complies with part 15 of the FCC rules. Operation is subject to
the following two condition:
1. This device may not harmful interference,
2. This device must accept any interference received, including
interference that may cause undesired operation.
The grantee is not responsible for any changes or modification not expressly
approved by the party responsible for compliance.
Such modifications could void the user’s authority to operate the equipment.
NOTE:** This equipment has been tested and found to comply with the limits
for a Class A digital device, pursuant to Part 15 of the
FCC Rules. These limits are designed to provide reasonable protection against
harmful interference when the equipment is
operated in a commercial environment. This equipment generates, uses and can
radiate radio frequency energy and, if not installed
and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in
which case the user will be required to correct the
interference at his own expense.
Do not use the device under conditions other than those described in the
technical specifications (Risk of fire or explosion).
For any other use than the one mentioned, please get in touch with JRI.
INSTALLATION RECOMMANDATIONS
To ensure optimal radio transmission, a certain number of recommendations must
be respected, as any wireless transmission is subject to disturbances.
a) Sources of disturbances or attenuation
- The presence of obstacles in the wave path between the Nano SPY and the Nano SPY Link (wall, furniture, people…) or near the antenna.
- The thickness of an obstacle in the wave path. The attenuation is greater diagonally than perpendicularly.
- A solid metal wall will not allow transmission by radio. A perforated metal wall will allow waves to pass while attenuating them.
b) Positioning
- The Nano SPY units can be placed either inside or outside the enclosures.
- For installations outside the enclosure, mount the units sufficiently high on the walls to avoid interference with obstacles and foot traffic.
- Insofar as is possible, place the Nano SPY LINK in a central position relative to the measurement points.
- Try to place them in locations where they are visible
- Never place the Nano SPY unit horizontally.
- If difficulties persist it is possible to use Nano SPY ALARMs (repeaters) or connect to another Nano SPY LINK on the Ethernet network.
**** To ensure your safety during installation or an intervention on a device placed in a high position, use proper equipment which is in good condition and provides adequate stability, wear appropriate, non-slip shoes and install warning signs around the work area if the intervention takes place in an area of foot traffic.
PRODUCT DESCRIPTION
a) Control unit b) Mounting
The Nano SPY can be mounted in 2 different ways
- Using a tie wrap to attach it to the monitored product
- Magnetically
The Nano SPY has 2 internal magnets for easy mounting on metallic surfaces. A protective shell is available as an option (Ref: 12715)
OPERATION
The Nano SPY can only be used with the My Sirius software hosted on a Web
platform and with a Nano SPY
Link. See MySirius online help for Nano Spy configuration.
a) Off state
As delivered, the Nano SPY is turned off. It can neither emit nor receive
signals.
The Nano SPY can only be used with the My Sirius software hosted on a Web
platform and with a Nano SPY Link. See MySirius online help for Nano Spy
configuration.
a) Off state
As delivered, the Nano SPY is turned off. It can neither emit nor receive
signals.
b) Activation Once activated, the le Nano SPY automatically declares
itself in MySirius if it is contact with a Link. It starts to measure and
transmit its measurements to My Sirius, at the frequency defined in MySirius,
then flashes regularly as a function of its status.
c) Turning Off d) Actions on the touch button
Press sensitive bouton Mode | < 3″ | > 3’’ | > 8’’ |
---|---|---|---|
Activation | – | during 3’’ | during 3’’ Thee Nano SPY remains |
Mesurement | 1’’ = OK |
1’’ = Technical alarm
3×1’’ = OK but paused
1’’ = In alarm state| Off
Off (If authorized by program)| –|
If the authorization of turning off the device is not programmed via MySirius
then it will be not possible to turn off the Nano Spy.
Using active, corrosive or flammable products or solutions (e.g. acid or
petroleum) on JRI equipment is prohibited.
The JRI equipment is designed for mapping and monitoring the temperature and
humidity of thermal or climatic enclosures within limits described in their
technical data sheet.
For the maintenance of these devices, please refer to the dedicated section.
For any other use than the one mentioned, please get in touch with JRI.
NANO SPY U CONNECTION
The Nano SPY Universal is equipped with a 4-wires cable facilitating
connection to terminal blocks of sensors with analog outputs. These sensors
can, if necessary, be disconnected from the logger for replacement or for the
exchange of the logger itself.
An IP67 Binder connector (optional) can be used to facilitate the connection
of the sensors.
Connect only sensors compatible with the technical characteristics of the
devices
For sensors with 0-1V and 0-20mA outputs, there’s no probe failure detection →
the NANO SPY U doesn’t trigger sensor default alerts.
BATTERY REPLACEMENT
Removing the battery
Open the battery cover 1 with a suitable object (coin) to align the marks
(/!\0 = Open ; 1= Closed)
Remove the battery 2 from its lodging
Replacing the battery
Put the new battery ** in place respecting the polarity 3.
Battery detection is confirmed by the activation of the red LED 4* for few
seconds. The device can be activated after the extinction of the red LED.
KEEP THE BATTERY AWAY FROM FIRE, DO NOT ATTEMPT TO RECHARGE IT OR SHORT-
CIRCUIT IT THE BATTERY MUST BE A LITHIUM 3.6V TYPE AA BATTERY.
USE PREFERABLY THE BATTERIES SUPPLIED BY JRI (PART NBR: 11596)
*Recommended batteries: Saft LS14500 type AA 3.6V 2250mAh
MAINTENANCE
Clean the device with a soft cloth, either dry or slightly moistened with water. To remove stubborn dust, use a cloth soaked in a diluted, non-abrasive detergent. Then wipe carefully with a soft dry cloth. Never use benzene, thinner, alcohol or any type of solvent, which can cause discoloration or deformation of the surfaces.
TECHNICAL FEATURES
a) Common features:
HMI: 1 RGB LED + 1 sensitive bouton
Frequency band: 2.4GHz (from 2400 to 2483.5 MHz)
Maximum Radio Power: 10 dBm
Memory: 10 000 timestamped measures
Resolution: 0.01
Dimensions: 63 mm x 42 mm x 25 mm
Case: Polycarbonate – Food Contact
Power source: 3,6v Lithium battery up to 6 years life time battery,
depending on usage
Weight: ~ 60g
b) Specific features:
Nano SPY T1 Temperature (internal probe) Sensor: PT100 sensitive element
inside unit
Operating range: from -40°C to +85°C
Measurement range: from -40°C to +85°C
Accuracy, standard version: ±0.4°C from -20°C to +40°C / ± 0.5°C outside
this range
IP rating: IP 68
Frequency of measurement and transmission: adjustable from 1 min to 12h
Frequency of recording: adjustable from 1 min to 24h
Response time: ~ 10 min. to 90% of the variation
Standard calibration points: -25°C/+5°C/+25°C
Nano SPY TH (Thermo-Hygro) Sensor: inside unit
Operating range: from -30°C to +70°C
Measurement range: from -30°C to +70°C and 0-100%HR
Accuracy, standard version
Temperature: ± 0.4°C from -20 C° to +40°C / ± 0.5°C outside this range
Humidity at T° from 15 to 25°C ±4% HR from 20% to 80%
From 20% to 80% HR ±5% HR outside of the range
IP rating: IP 40
Frequency of measurement and transmission: adjustable from 1 min to 12h
Frequency of recording: adjustable from 1 min to 24h
Response time: ~ 5 min. to 90% of the variation
Standard calibration points: +2°C/+25°C/+38°C and 20%/50%/80% at 23°C
Nano SPY T2 Remote temperature Sensor: PT100 external probe, non-
withdrawable Ø5x20mm
Operating range: from -30°C to +70°C
Measurement range, standard version: from -50°C to +105°C
Measurement range, incubator version: from +30°C to +50°C
Accuracy, standard version: ±0.3°C from -20°C to +30°C / ±0.5°C outside
this range
Accuracy, incubator version: ±0.2°C from +30°C to +50°C/±0.5°C outside
this range
IP rating: IP 65
Cable length: Flat cable Sentroprene 30 cm, 3 m and 8 m
Frequency of measurement and transmission: adjustable from 1 min to 12h
Frequency of recording: adjustable from 1 min to 24h
Response time: ~ 2 min. to 90% of the variation
Standard calibration points, standard version: -30°C/+5°C/+40°C
Standard calibration points, incubator version: +36°C/+38°C/+45°C
Nano SPY Reference Sensor: External Class A PT 100 – stainless steel
Ø2,9x25mm
Operating range: from -20°C to +50°C
Measurement range: from -196°C to +150°C
Accuracy: ±0.15°C from 0°C to +40°C, ±0.2°C from -30°C to 0°C and from
+40°C to +150°C, ±0.5°C out of this range, ±0.6°C to -196°C
IP rating: IP 65
Cable length: 3 m
Frequency of measurement and transmission: 1 min
Frequency of recording: Adjustable from 1 min to 24h
Response time: ~ 2 min
Standard calibration points: +15°C/+45°C
Nano SPY T3 Extreme temperature data logger Sensor: PT100 external
probe, non-withdrawable
Operating range: from -20°C to 50°C
Measurement range: from -200°C to +200°C
Accuracy, Low temperature version: ±0.2°C from -20°C to 0°C and ±0.5°C
outside this range
Accuracy, High temperature version: ±0.3°C from 0°C to +100°C and ±0.5°C
outside this range
IP rating: IP 65
Cable length: 50 cm and 3 m
Frequency of measurement and transmission: adjustable from 1 min to 12h
Frequency of recording: adjustable from 1 min to 24h
Response time: ~ 2 min. to 90% of the variation
Standard calibration points, Low temperature version: -80°C/-10°C
Standard calibration points, High temperature version: +20°C/+100°C
Nano SPY TH High Accuracy Sensor: Sensitive element PT 1 000, 1/3 B
class, HYGROMER HT-1 cable length: 2m
Operating range: from -40°C to +85°C
Measurement range: from -40°C to +85°C and 0-100 %HR
Accuracy: Refer to the technical data sheet
IP rating: IP 65
Cable length: 2 m
Frequency of measurement and transmission: adjustable from 1 min to
12h
Frequency of recording: adjustable from 1 min to 24h
Response time: ~ 15 seconds without filter to 90% of the variation
Nano SPY Door contact Sensor: Magnetic door contact
Operating range: from -40°C to +85°C
Accuracy: Refer to the technical data sheet
IP rating: IP 65
Cable length: 20 cm
Frequency of measurement and transmission: 1 min
Frequency of recording: adjustable from 1 min to 24h
Response time: ~ 1,7 milliseconds
Nano SPY Digital Sensor: External to control unit
Operating range: from -25°C to +70°C
Measurement range: According to the type of JRI digital probe
Accuracy: Accuracy of the JRI digital probes
IP rating: IP 65
Cable length: N/A
Frequency of measurement and transmission: adjustable from 1 min to 12h
Frequency of recording: adjustable from 1 min to 24h
Response time: ~ 2 min. to 90% of the variation
Standard calibration points: +20°C/+100°C
Nano SPY TWIN, 2 Channels: Internal PT100 probe + External PT100 probe – non-withdrawable Ø5x20mm ****
Sensor: Internal PT100 probe
External PT100 probe – non-withdrawable Ø5x20mm
Operating range: from -30°C to +70°C
Measurement range, Ambient probe – Internal: from -30°C to +70°C
Measurement range, PT100 Probe – External: from -50°C to +105°C
Accuracy, Ambient probe – Internal: ±0.4°C from -20°C to +40°C and ±0.5°C
out of this range
Accuracy, PT100 Probe – External: ±0.3°C from -20°C to +30°C and ±0.5°C
out of this range
IP rating: IP 65
Cable length: 3m
Frequency of measurement and transmission: adjustable from 1 min to 12h
Frequency of recording: adjustable from 1 min to 24h
Response time: ~ 2 min. to 90% of the variation
Standard calibration points: -30°C / 0°C / +40°C
Nano SPY Universal Input type:
PT100 input
measurement range:
resolution:
accuracy (not including probe):| An adjustment of the measurement chain is
mandatory
From -200°C to +300°C
0,01°C
± 0,2°C from -20°C to +50°C
±0,3°C from -80°C to -20°C and from +50°C to +140°C
± 0,5°C beyond these ranges
---|---
Current input
measurement range:
resolution:
accuracy (control unit only):| (No detection of probe failure for 0-20mA
input)
From 0 to 20 mA or 4-20mA
0,001 mA
± 0,01 mA
Voltage input
measurement range:
resolution:
accuracy (control unit only)| (No detection of probe failure for 0-1 V input)
0 à 1V
0,1 mV
±0.5 mV
On/Off or Counting input
Type of input measurement range:
resolution:
accuracy| Dry contact or 0-3,3V max
0 à 65535 – signal from 0 to 200Hz
1
±1
Operating range: from -20°C to 50°C
IP rating: IP 65
Cable length: 2 m
Frequency of measurement and transmission: adjustable from 5 sec to 12h
Frequency of recording: adjustable from 1min to 24h
Response time: according to sensor
Connection: Using a Binder IP67 connector (ref : 12617) or directly on
the analog output
c) Compliance
JRI declares that all our radioelectric devices from the NANO SPY range
(T1,T2,T3,TH) are in compliance with the following standards:
EN 12830| These devices must be verified regularly according to EN13486
(recommended once per year)
---|---
FCC| FCC ID : W45 12525
This device complies with part 15 of the FCC rules. Operation is subject to
the following two condition:
1.This device may not harmful interference,
2.This device must accept any interference received, including interference
that may cause undesired operation.
The grantee is not responsible for any changes or modification not expressly
approved by the party responsible for compliance. Such modifications could
void the user’s authority to operate the equipment.
NOTE: This equipment has been tested and found to comply with the limits
for a Class A digital device, pursuant to Part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial environment. This
equipment generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instruction manual, may cause
harmful interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own
expense.
CE ERM| EN 301489-1 & -17/EN 61010-1/EN 62479/EN 300328(T1, T2, T3, TH, U, TH
High Accuracy) / ETS 300-328 (TH, T3, T2, T1, Reference, Door contact)
2014/53/UE| A copy of the full EU compliance statement can be requested by
email: info@group-mms.com
IC CANADA| This device complies with Industry Canada license-exempt RSS
standard(s). Operation is subject to the following two conditions: (1) This
device may not cause interference, and (2) This device must accept any
interference, including interference that may cause undesired operation of the
device.
This equipment should be installed and operated such that a minimum separation
distance of 20 cm is maintained between the radiator (antenna) and
user’s/nearby person’s body at all times.
OPERATION CAPACITY
FITNESS FOR USE SHEET ACCORDING TO STANDARD 12830
Capacity of operation compliant to EN 12830
model: Nano SPY T1
equipment type: temperature recorder
application: storage
environment: C
accuran class: 1
Test table
Trials| δ
Norms| Requirements| Characteristics| Documents or test
report
---|---|---|---|---
Determination of the error of the temperature range.| 5.3| ±1°C| ±0.5°C|
Metrological qualification report RQCC16001
Determination of response time.| 5.4| <60′| 10′| JRI RECC16003 test report
Determination of the relative time registration error.| 5.5| 1%| 0.002%| JRI
RECC 16005 test report
Variation of the supply voltage. (Earepstreur subjected to assagatic
temperatures)| 5.6.2| 3,2V 13.6V
-30°C 1 +30°C| 2,7 V i 3,6V
-40°C 1 +85°C| JRI RECC16005 test report
Influence of ambient temperature. (Recorder under = at hmite temperatures)|
5_6.3.3| -40°C I +50°C| -40°C i +85°C| JRI RECC 16003 test report
Temperature test with recording in storage and transport condition.| 5.6.4|
-40°C 1 +60°C| -40°C 1 +85°C| JRI RECC16003 test report
Impact resistance.| 5.6.5| EN 60068-2- 27| N/A| Not recognized for storage
Mechanical vibrations| 5.6.6| EN 60068-2-27| N/A| Not required for storage
Degrees of protection provided by the enclosure.| 5.6.7| IP55 selon EN 60529|
Test report /RI RECC 16005
Electrical safety| 5.6.8| EN61010-1: 2010| EMITECH RS-031-PNC-16-103276-2-A
test report. pdf
Ripdite chelectrique.| 5.6.9| NA| INS|
Electromagnetic compatilatine.| –| EN 301489-17 V2.2.1: 2012
EN 301489-1 V1.9.2: 2011
EN 55024: 2010 / Al 2015| EMITECH TC-032-PTC-I5-16147-1-1 test report. pdf
Pour JRI
The Technical and Quality Director: Technical and quality manager
Date : 23101/2017
WARANTY
Our material is guaranteed for one year, parts and labor, against any
manufacturing defect, functional failure or abnormal wear. This guarantee
covers only the replacement of parts recognized to be defective as well as the
repair of the material in question returned shipping paid to our workshops,
and excludes all damages and interest or incidental expenses.
The starting point of the guarantee is the date of invoice of the concerned
product. The invoice must be provided for any request for application of the
guarantee. Repairs under guarantee in no way extend the guarantee period
accorded to the product at the time of sale. Deterioration due to any abnormal
usage or to storage under adverse environmental conditions is excluded from
our guarantee.
MAINTENANCE CONTRACT
How best to optimize your radiofrequency installation?
Radiofrequency measurement systems communicate through Hertzian waves. Many
factors (change in installation, moving, supplemental wall, interference with
another radio system…) can nonetheless modify the radio pathway previously
defined. The use of radiofrequency thus requires periodic monitoring by
recognized specialists.
It is for this reason that JRI has developed for you the maintenance contract.
We simplify your procedures by offering you a fully-integrated solution. This
global service offer includes both maintenance and a metrological service,
ensuring the optimum functioning of your devices or of your installation.
You’ll no longer have to worry about the maintenance of your devices!
This maintenance contract allows you to benefit, for a minimum period of 2
years, from a variety of services such as:
- annual or biannual verification of the material
- an extension of the guarantee
- tele maintenance
- technical support +33 (0) 892 680 933 (0,282 € HT/min)
- replacement of the material onsite or by a return to the factory
- verification of measurement accuracy (metrological certificate)
- battery replacement
- access to new software versions
- intervention within 48 working hours following identification of the fault by our experts
ENVIRONMENTAL PROTECTION
JRI recommends to its customers the disposal of their unusable and/or
irreparable measurement and recording materials in a manner compatible with
the protection of the environment. As the production of waste materials cannot
be avoided, these should be reused through the recycling process best adapted
to the considered materials and to the protection of the environment.
RoHS Directive
The RoHS European directive regulates and limits the presence of dangerous
substances in electronic and electric equipment (EEE).
All new electronic equipment designed, developed and manufactured by JRI are
compliant with the aforementioned Directive 2002/95/CE.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>