OHAUS 83021171 Explorer Semi Micro Balance Instruction Manual

June 10, 2024
Ohaus

Explorer™ Balances
Instruction ManualOHAUS 83021171 Explorer Semi Micro
Balance

INTRODUCTION

1.1 Description
The Explorer balance is a precision weighing instrument that will provide you with years of service if properly cared for. The Ohaus Explorer balances are available in capacities from 120 grams to 35,000 grams.
1.2 Features
Modular Design:
Ohaus Explorer balances are composed of two interconnected modules: a Terminal and a Base. Depending on the user’s needs, the unit can be operated with the Terminal either attached to, or remote from, the Base, with a single interconnect cord 1 meter long.
Touch Controls: Quick, graphical access to all control functions, over a dozen applications and many features.
Touchless Sensors: can be assigned functions when activated: Zero, Print, Ionizer+Tare, Automatic draftshield door, ionizer etc. ”
Statistics: available in Weighing, Counting, Percent Weighing, Check Weighing, Dynamic Weighing, Filling.
Ionizer (EX225D/AD and EX225/AD only) The Ionizer generates bipolar ions continuously by DC corona discharge separately from positive and negative discharge electrodes and directs the ionized air onto the charged body to eliminate static electricity. The generated ions are well-balanced in polarity and can eliminate static electricity regardless of the polarity of the charged body.
Static electricity
Generally, insulators such as plastics are prone to be charged with static electricity when the ambient relative humidity is below 45%RH. This may affect weighing and cause a measurement error of several milligrams. The Ionizer can eliminate static electricity very effectively.
How to activate the Ionizer
The Ionizer can be activated or turned off by either the touchless sensors (see section 5.4.9) or through the function icons by pressing the Ionizer icon on the application home screen.
1.3 Safety Precautions
Definition of Signal Warnings and Symbols
Safety notes are marked with signal words and warning symbols. These show safety issues and warnings. Ignoring the safety notes may lead to personal injury, damage to the instrument, malfunctions and false results.
WARNING For a hazardous situation with medium risk, possibly resulting in severe injuries or death if not avoided.
CAUTION For a hazardous situation with low risk, resulting in damage to the device or the property or in loss of data, or minor or medium injuries if not avoided.
ATTENTION For important information about the product. May lead to equipment damage if not avoided.
NOTE For useful information about the product.
Warning Symbols

**** General hazard
Explosion hazard
Electrical shock hazard

Safety Notes
CAUTION: Read all safety warnings before installing, making connections, or servicing this equipment. Failure to comply with these warnings could result in personal injury and/or property damage. Retain all instructions for future reference.

  • Before connecting power, verify that the AC adapter’s input voltage range and plug type are compatible with the local AC mains power supply.
  • Do not position the equipment such that it is difficult to reach the power connection.
  • Make sure that the power cord does not pose a potential obstacle or tripping hazard.
  • When shipping or transporting this product, follow the applicable regulations for equipment containing lithium-ion batteries.
  • Operate the equipment only under ambient conditions specified in these instructions.
  • This equipment is for indoor use only.
  • Do not operate the equipment in wet, hazardous or unstable environments.
  • Do not allow liquids to enter the equipment.
  • Do not load the equipment above its rated capacity.
  • Do not drop loads on the platform.
  • Do not place the equipment upside down on the platform.
  • Use only approved accessories and peripherals.
  • Disconnect the equipment from the power supply when cleaning.
  • Service should only be performed by authorized personnel.

WARNING: Never work in an environment subject to explosion hazards! The housing of the instrument is not gas tight. (Explosion hazard due to spark formation, corrosion caused by the ingress of gases).
CAUTION: Battery is to be replaced only by an authorized Ohaus service dealer. Risk of explosion can occur if the rechargeable battery is replaced with the wrong type or if it is not properly connected. Dispose of the rechargeable battery according to local laws and regulations.
Safety precautions for the Ionizer:
WARNING: ELECTRICAL SHOCK HAZARD
AVOID TOUCHING THE ELECTRODE PINS WHILE THE IONIZER IS TURNED ON.
1.4 Intended User
This instrument is intended for use in laboratories, pharmacies, schools, businesses and light industry. It must only be used for measuring the parameters described in these operating instructions. Any other type of use and operation beyond the limits of technical specifications, without written consent from OHAUS, is considered as not intended. This instrument complies with current industry standards and the recognized safety regulations; however, it can constitute a hazard in use. If the instrument is not used according to these operating instructions, the intended protection provided by the instrument may be impaired.

INSTALLATION

2.1 Unpacking
Carefully remove your Explorer balance and each of its components from the package. The included components vary depending on the balance model (see table below). Save the packaging to ensure safe storage and transport.

Included Component| Photo| EX125
EX1D
EX22525D¹
EX225¹| EX124¹
EX224¹
EX324¹| EX(2232 EX4232 EX623 EX1103| EX2202² EX4202² EX6201² EX6202² EX10201 EX10202| EX12001
Ex2001
EX35001
---|---|---|---|---|---|---
Terminal| | | X| X| X| X| X
Weighing Base| | | X| X| X| X| X
Weighing Platform| 80mm Diameter| | X| | | |
90 mm Diameter| | | X| | |
130 mm Diameter| | | | X| |
190 x 200 mm| | | | | X|
311 x377 mm| | | | | | X
Sub-Platform| 80mm Diameter| | X| | | |
130 mm Diameter| | | | X| |
Spider| | | | | | | X
Support Caps (4 Pcs)| | | | | | X|
Wind Ring| | | X| | | |
| | | X| | |
Wind Shield| | | | | | X|
Draft Shield Assembly| | | X| X| X| |
Glass Shelf Assembly| | | X| | | |
Power Adapter| | | X| X| X| X|
Power Cord| (Country specific)| | X| X| X| X| X
Use Cover| Terminal| | X| X| X| X| X
Warranty Card| | | X| X| X| X| X
USB Extension Cable| | | X| | | |
Dust Cover| | | X| | | |

Notes:

  1. Includes Automatic draftshield door Model
  2.  Includes ExCal Model

2.2 Installing Components
Refer to the illustrations and instructions below to identify and assemble your Explorer balance with its components.
All components must be assembled before using the balance.
2.2.1 Terminal Setup
When the Explorer is delivered, the Terminal is already attached (docked) to the Base. No additional setup is necessary. However, for Draft Shield models, the Terminal needs to be detached temporarily in order to assemble parts of the Draft Shield. Refer to the illustrations and instructions below to identify and assemble your Explorer Balance.
Note: The Terminal is identical for all Explorer Balance models.
2.2.2 Installing the Wind Ring, Weighing Platform on Semi-Micro models
Note: Assembly instructions also apply to models with the suffix M or N.

  1. Place the Wind Ring in position.

  2. Insert the Platform Assembly onto the Mounting Cone located in the center of the Balance.OHAUS 83021171 Explorer Semi Micro Balance - icon 13 2.2.3 Installing the Wind Ring, Weighing Platform on Models EX124¹ , EX224¹ , and EX324¹
    Note: Assembly instructions also apply to models with the suffix M or N.

  3. Place the Wind Ring in position.

  4. Insert the Platform Assembly onto the Mounting Cone located in the center of the Balance.

2.2.4 Installing the Sub-Platform and Weighing Pan on Models EX223² , EX423² , EX623 and EX1103

  1. Place the Sub-platform in position.
  2. Position the Platform on the Sub-platform.

EXPLORER BALANCES
2.2.5 Installing the Weighing Pan on Models EX12001, EX24001 and EX35001

  1. Position the Platform on the base.

Notes:

  1. Includes Automatic draftshield door Model
  2. Includes ExCal Model

2.2.6 Installing the Draft Shield Glass and Doors on Draft Shield Models

  1. Carefully remove each of the four glass Draft Shield sections from the packaging.
    NOTE: Before installing side doors, be sure to set the lock tabs in the proper position, as shown here.OHAUS 83021171 Explorer Semi Micro Balance -
shown here

  2. Follow the steps below to install side, front and top doors of the Draft Shield.

Side Door Installation
OHAUS 83021171 Explorer Semi Micro Balance - shown here
1 Position glass door to slide. Open lock tab (top); insert in top groove and slide inside bottom groove. Close lock tab.
Front Panel Installation
OHAUS 83021171 Explorer Semi Micro Balance - shown here
2Press button to separate Terminal from Base. Seat panel’s hooked ends in base. Swing panel to vertical position.

Top Door Installation

OHAUS 83021171 Explorer Semi Micro Balance - shown here
3

Insert Top Door’s hooked ends in grooved pegs. Press gently to seat. Check for smooth movement. Close door.
2.2.7 Installing the Glass Shelf (on Semi-Micro models)

OHAUS 83021171 Explorer Semi Micro Balance - shown here
4

Align and press down the shelf on the two supporting pegs.
Note: If the sliding doors and top lid have been installed, first open the sliding glass doors on both sides and open the top lid to make space for the glass shelf.
2.2.8 Installing the Weighing Pan & Wind Shield – Non-Draft Shield Models
Note: Assembly instructions also apply to models with the suffix M or N.

  1. Place the four Support Caps onto the Balance as shown.
  2. Place the Weighing Pan on the Support Caps
  3. Place the Wind Shield onto the Balance as shown.

OHAUS 83021171 Explorer Semi Micro Balance - shown here
5

Note: The balance can be operated without the Wind Shield. However, depending on ambient conditions the displayed results may be slightly less stable.
2.3 Selecting the Location
Avoid excessive vibrations, heat sources, air current, or rapid temperature changes. Allow sufficient space.

OHAUS 83021171 Explorer Semi Micro Balance - shown here
6

Note: Interface cables connect to the terminal. The terminal can be detached and mounted on a wall or positioned on a table separate from the balance.
2.4 Leveling the Equipment
The Explorer has an illuminated level indicator as a reminder that the balance should be leveled for accurate weighing. There is a level bubble in a small round window on
the front of the Base. To level the balance, adjust the Leveling Thumbwheels at each corner until the bubble is centered in the circle.
Be sure the equipment is level each time its location is changed. Note: The feet cannot rotate. The Thumbwheels must be used for leveling.
See the Level Assist screen in the Information Menu.
To see the Level Assist menu, touch More on the bottom-right corner of the main screen. Then touch Level Assist.
The Level Assist screen shows which direction to turn the thumbwheels to center the level bubble.

OHAUS 83021171 Explorer Semi Micro Balance - shown here
8

2.5 Connecting Power
2.5.1 Models with AC Adapter
For models supplied with an AC adapter, connect the DC output connector to the power receptacle on the rear of the base. Then connect the AC power cord to a suitable electrical outlet.
Note: EX12001, EX24001 and EX35001 do not have AC Adapter
Caution: For use with CSA certified (or equivalent approved) power source, which must have a limited current output.
Note: Allow equipment to warm up for 60 minutes for optimal weighing performance.
Standby: When unit is plugged in, it starts in Standby mode. Press the Standby icon to start.

2.5.2 Models with AC Power Cord
For models with an AC power cord, connect the power cord to a suitable electrical grounded outlet

2.5.3 Power ON Button on EX12001, EX24001 and EX35001
The models EX12001, EX24001 and EX35001 have no AC Adapter. Just connect the power plug to a suitable electrical outlet and press the ON button on the side of the base (see figure below).OHAUS 83021171 Explorer Semi Micro Balance -
shown here 11

Power ON button on the side of the base (EX12001, EX24001 and EX35001)
2.6 Connecting the Interface
Use the built-in RS-232 (10101) Port to connect either to a computer or a printer with a standard (straight-through) serial cable. Or connect using the scale’s USB ports.
Note: For configuration and interface commands, see the Communication Menu Settings section.
For Connecting, Configuring and Testing the Printer/Computer Interface, and for sample Print Output Formats, see the Printing section.

Interface connections on back of Terminal.

Thread printer cable along cable coils on bottom of balance. Or pass cable through groove near release button.
2.7 Adjusting the View Angle of the Terminal
To adjust the Terminal display’s viewing angle, press in the Terminal Adjust buttons on both sides of the Terminal.

OHAUS 83021171 Explorer Semi Micro Balance - shown here
15

EXPLORER BALANCES

2.8 Remote Terminal Operation
The Terminal communicates with the weighing base via the Terminal cable. This cable must be plugged into the Terminal for the Explorer to display properly. If desired, the Explorer balance may be operated either with the Terminal attached (as-shipped), or remotely (up to 1 meter away).
2.8.1 Separating the Terminal from the Weighing Base

  1. To detach, press both the Release buttons inward (both at the same time) and gently pull the Terminal towards you (outward) until the Terminal is detached. These Release buttons disengage the two hooks holding the Terminal to the Base. A cable is attached to the Terminal. Take care to not damage or disconnect this cable.
  2. To reattach the Terminal, press in the two Release buttons and slide the Terminal into the Base until the Terminal hooks click and engage to hold the Terminal in place.

Release Button

Base and TerminalOHAUS 83021171 Explorer Semi Micro Balance - shown
here 17

Terminal Cable

Note: If a longer operation distance is required, a Terminal Extension cable is available as an accessory.
2.8.2 Terminal Mounting
If desired, the Terminal may be mounted to a wall or table using fasteners (not supplied) that are appropriate for the type of mounting surface The Terminal bracket will accommodate 4mm (#8) screws. Locate the mounting holes as shown in Figure 2-1.   OHAUS 83021171 Explorer Semi Micro Balance - shown
here 19

2.9 Initial Calibration
When the Balance is first installed, and when it is moved to another location, it must be calibrated to ensure accurate weighing results. The Explorer Balances have built in AutoCal which can calibrate the balance automatically and does not require calibration masses. If preferred, the balance can be manually calibrated with external masses. Have the appropriate calibration masses available before beginning calibration. Refer to the Calibration Section for masses and calibration procedure.

OPERATION

3.1 Overview of Display, Home Screen
This equipment utilizes a touch-sensitive display. Touch areas and Buttons to control the equipment’s functions.
CONTROLS

*Battery pack is only for EX12001, EX24001, EX35001 and their related models.
3.2 Principal Functions and Main Menu
Weighing: Press Tare to set the display to zero. Place an item on the pan. Display indicates gross weight. Taring: With no load on the pan, press Tare to set the display to zero. Place an empty container on the pan and press Tare. Add material to the container and its net weight is displayed. Remove container and container’s weight appears as a negative number. Press Tare to clear.
MENU & SCREEN NAVIGATION
Touch Menu to open the menu list.
Touch and drag the Scroll bar to view additional items.
Calibration: Touch to view calibration options.
User Management: Touch to view user management settings.
User Settings: Touch to view user preferences.
Balance Setup: Touch to view balance settings.
Application Modes: Touch to view application modes.
Weighing Units: Touch to view weighing units.
GLP and GMP Data: Insert user data for traceability.
Communication: Touch to view COM Device Settings and Print Settings.

Library: Touch to delete all Library records.
Diagnostics: Touch to view Diagnostics items.
Factory Reset: Touch to do a Factory reset.
Information: Touch to view list of topics.

3.3 Overview of Parts and Features
OHAUS 83021171 Explorer Semi Micro Balance - shown here
22
3.4 Overview of Parts and Features –Draft Shield ModelsOHAUS 83021171
Explorer Semi Micro Balance - shown here 23

3.5 Overview of Parts and Features – Non-Draft Shield ModelsOHAUS
83021171 Explorer Semi Micro Balance - shown here 24 3.6 Overview of Parts and Features – EX12001, EX24001 and EX35001OHAUS 83021171 Explorer Semi Micro Balance - shown here
25

3.7 Overview of Parts and Features – TerminalOHAUS 83021171 Explorer
Semi Micro Balance - shown here 26

3.8 Functions and Their Icons
Seven Icons are initially available to operate the Balance and access the Menu. These Icons can be customized based on user preference. For setup information, see User Settings.
3.8.1 Standby
Touch Standby to shut down the display.
Note: After initial start-up (Section 2), in Standby your balance needs no warm-up time and is immediately available for weighing by touching the Start-up icon.
3.8.2 Printing Data
Touch Print to send the displayed value to a printer or computer via the active COM port.
Note: Ensure proper setup of connections, printing and interface parameters.
Note: Data may also be printed using the Touchless Sensors by configuring these for Printing.
Data may also be printed using the P command from a computer connected to the COM port.
3.8.3 Zero Operation
Remove the load from the pan and touch Zero to set the display to zero. When weighing pan is empty, the >0< indicator turns on when the measurement is within + ¼ division (d) of the zero setting.
Note: The balance also features Auto Zero Tracking (AZT) that automatically maintains a Center of Zero within the tolerances set in the AZT menu (see Balance Settings).
3.8.4 Touch Less Sensor Status
Explorer balances have four Touch Less sensors that can be assigned a unique function when activated (e.g., zero, print, ionizer+tare, Automatic draftshield door etc).
To assign a task for each Touch Less sensor, press Sensors.
Note: To activate a sensor, move an object over it (from a distance of 1-2 inches or 3 cm). The sensor shows a green light and beeps when activated. If the sensor cannot be activated (sensor is disabled during certain instances, like when the Menu is displayed) the sensor shows a red light.
3.8.5 Auto Cal
When Auto Cal is ON, the balance performs a self calibration. Auto Cal will automatically calibrate the Balance (using the internal mass) each time a change in temperature significant enough to affect accuracy, or after every 11 hours and for Semi-Micro models, the Auto Cal system is triggered by 3 hours as default setting.
Touch Auto Cal™ to initiate. (Default status is ON.)
Note: It is not available in ExCal.
For Non-Approved models, when AutoCal Internal Calibration is turned off, it changes to span calibration. OHAUS 83021171 Explorer Semi Micro Balance -
shown here 27For Approved models, when Auto Cal Internal Calibration is turned off, it is hidden.OHAUS 83021171 Explorer
Semi Micro Balance - shown here 28

3.8.6 Menu
Touch Menu to enter the balance’s menu list.
3.8.7 More
Touch More to access Level Assist, Zero, Tare, Units, Pretare, Cal Test, Calculator, Stopwatch, Graduation and Routine Test.
3.8.8 Level Assist
For assistance with Leveling the Balance; the Level Assist screen shows which direction to turn the thumbwheels to center the level bubble.
3.8.9 Graduations
Touch Graduations to set 1d or 10d resolution for display.
3.8.10 Routine Test
Touch Routine Test to perform 10 times repeatability test by using the test weight close to working point.
3.8.11 Applications
The balance can be configured to operate in various Application modes.
Touch Applications to choose – or the top button in the Application area.
3.8.12 Taring a Container
Taring sets the weight of a container as zero, so only the weight of objects held in the container (net weight) is displayed.
Manual Tare – Place the empty container on the pan and touch Tare when stable.
Add material to the container. The net weight of the material is displayed.
To clear the Tare value, remove the container from the pan and touch Tare.
g Kg mg 3.8.13 Weighing Units
The balance can be configured to measure in a variety of weighing units, including three (3) custom units. Touch Units to choose from the displayed list. (Scroll down to see all options.)
Note: The active weighing Unit can also be accessed by touching the unit area of the Main Display line; If a unit is not displayed in the list, it must first be turned on via the User Menu.
3.8.14 Ionizer (only for EX225D/AD and EX225/AD)
The Ionizer is used to eliminate static electricity. To use the Ionizer, touch the Ionizer icon. A pop-up window will appear with the text: ‘The ionizer is neutralizing the static charges.
3.8.15 Preset Tare
To enter a Preset Tare, press the Icon and then enter a value; enter 0 to clear the Preset Tare. The display will show the PT symbol and the tare value as a negative  number.
3.8.16 Calibration Test
is used to compare a known calibration weight against the last stored span calibration data.
3.8.17 Calculator
To use the built-in calculator application, touch Calculator. A four function calculator with addition, subtraction, multiplication and division mathematical operations is available.

  • Perform a numeric operation using the keypad.
  • Press = to view the result.
  • To Clear the display: Press CE to clear the entry, C to clear all
  • Touch x to close and return to the current Application.
    Note: Calculator is not available in software version 1.10.

3.8.18 Stopwatch
To use the built-in stopwatch application, touch Stopwatch. A countdown timer with interval (lap) timer is available.

  • Press Start to begin using the Stopwatch timer.
  • Press Reset to reset the timer to Zero.
  • Touch Close to close and return to the current Application.

3.8.19 Logout
Used for User logout and return to User Login screen. This feature is only active if a user profile has been created and a User has logged in.
3.9 Ionizer (EX225D/AD and EX225/AD only) and Glass Shelf Step 1.
Place the glass tube in front of the ionizer while the IR sensor has activated the ionizer

Step 2.
After about 3 to 5 seconds, removethe sample. The static electricity is now eliminated

Step 3.
Sampling

APPLICATION

The balance can be configured to operate in various Application modes.
Touch Applications to choose – or the top button in the Application area:OHAUS 83021171 Explorer Semi Micro Balance - shown here
34

Note: If an application mode does not appear in the list of available Applications it must be enabled in the User Menu: touch Menu, then select Application Modes. The full list of modes appears. Touch the one you want to enable. Then touch Enabled in the setup menu for that application mode – Percent Weighing in the example below. Each enabled mode appears in the Applications menu.OHAUS 83021171 Explorer Semi Micro Balance - shown here
35

The Explorer Balance incorporates the following Applications.OHAUS
83021171 Explorer Semi Micro Balance - icon 49

4.1 Weighing
Note: Before using any application, be sure the balance has been leveled and calibrated.
Use this application to determine the weight of items in the selected unit of measure.
Note: The Explorer is shipped with all units enabled.
Weighing

  1. In the upper left portion of the home screen, select Weighing (this application is the default).
  2. Press Tare or Zero if necessary to begin.
  3. Place objects on the pan to display the weight. When stable, the * appears.
  4. The resulting value is displayed in the main Weighing Line in the active unit of measure.

The WEIGHING Home screen Main Display Line Second Display Line (not shown)
Reference Fields  Application Buttons Functions

4.1.1 Edit Settings
To view or adjust the current settings
Touch the Edit Settings button. The Settings screen appears.
Minimum Weight: establish a minimum weight value to be used to verify a reading.
Sample Name: establish a name for your sample  Batch ID: establish an ID for your sample batch.
Sample ID: establish an ID for your sample.
Clear All: clear all items in Edit Settings to default settings.
To adjust the Minimum Weight value, touch the Minimum Weight button.
If an actual weight is below the established Minimum Weight value, the screen will show Below Minimum Weight.

A numeric input window appears.
Use the keys to enter the desired Minimum Weight, then press .
The display reverts to the previous screen with the new Minimum Weight value highlighted: orange.
To return to the Weighing home screen, touch Return to Application at bottom- right of that screen.

To input a sample name, touch the Sample Name button. Enter the desired sample name, and then press.
You can use bar code scanner to input sample name, batch ID, Sample ID, or use RFID reader to get the sample name.

Note: Since there are many brands of RFID device in the market, OHAUS tested and confirmed that below one from RFIDeas (www.RFIDeas.com) is compatible:
RDR-6081AKU-C06. Since there are many brands of barcode scanners in the market, OHAUS tested and confirmed that below ones from Datalogic are compatible:
Heron series
Quick Scan series
Gryphon 4100 series
Gryphon 4400 series
To input a batch ID, touch the Batch ID button. Enter the desired ID, and then press√. To input a Sample ID, touch the Sample ID button. Enter the desired ID, and then press√.
To clear the Minimum Weight, Sample Name, Batch ID and Sample ID you set, touch the Clear All button. Touch Yes to confirm.

4.1.2 Application Setup
The Application can be customized for various user preferences.
Touch the Setup Icon (wrench) to enter Application Setup.  The Application Setup Menu is now displayed.
Select the list item to view or change the setting as desired.
To return to the Application home screen, touch Done.

The Weighing Application Setups are below (defaults in Bold)

Enabled On, Off Turns Application on/off
RFID On, Off To read RFID code to get sample name
Secondary Unit On, Off To show the Second Display line
Minimum Weight On, Off To enable Min Weight
Statistics Off, Manual, Auto
Statistics Result Standard, Continuous To select the Statistics print

output:
Standard: print all the Statistics result
Continuous: print sample weights without other Statistics data. Note: this menu will only appear when the Statistics menu is enabled.
Lock Settings| On, Off| Press to lock the current setting(s) so they cannot be changed.
Reference fields| |
Min Weight| On, Off| Min Weight is displayed
Gross| On, Off| Gross weight is displayed
Net| On, Off| Net weight is displayed
Tare| On, Off| Tare weight is displayed
Print Options| | See Printing Section.
Min Weight| On, Off, Below Minimum Weight| Below Minimum Weight: both the weight value and the text “Below Minimum Weight” will be printed when weight is below minimum.

Note: When RFID is turned on, the RS232 printing function will be disabled.
4.1.3 Statistics for Weighing
The Statistics function allows you to compare a number of samples and examine the relative deviation of each sample from the mean, along with other statistical data. A minimum of three samples is required. Statistics can be determined manually (requires key press after each step) or automatically (weights are automatically recorded when stable). To use the Statistics function, enable it in the Application Setup menu. This feature is available in Weighing, Counting, Check Weighing, Dynamic/Animal Weighing and Filling applications.
Application Setup: First, enable Statistics
Touch the Setup Icon (wrench) to enter Setup menu to enable Statistics.
The Statistics options screen appears.OHAUS 83021171 Explorer Semi Micro
Balance - shown here 48

Enable the desired mode (Manual, Auto)
Press to return to the Setup menu. After you click your preference, the window will close automatically.

Accumulating displayed values & viewing data – Manual Once Statistics-Manual is enabled, two buttons appear on the Application screen: Statistics and Accumulate.
Application – Auto Mode
Statistics Auto mode is similar to Manual mode, except the Accumulate button is not shown. The weight of each sample is automatically captured. The current sample must  be removed before capturing the next sample. When stable, the Statistics button lights briefly to show the item is added to the set, and the data set quantity increments by one.  Continue to add items to build the data set.
Press the Statistics button to view the results, as shown here.
Press Print to send results to the enabled COM port.
Press Close to return to the Application home screen without clearing the current statistics set.
Press Clear Data to clear results and return to the home screen.

Note: If you set Statistics Result in weighing application setup to continuous, the print output will only include sample weights without other statistic data. Statistic Result menu will only appear when the Statistics menu in weighing application setup is activated.

4.2 Parts Counting
Use this application to count samples of uniform weight. Three different modes can be selected:
Standard Counting, Check Counting, or Fill Counting.
4.2.1 Standard Parts Counting (default)
In the upper left portion of the home screen, select Counting.
The default (or last) Average Piece Weight (APW) is displayed.
Place objects on the pan to display the number of pieces.

The PARTS COUNTING- Standard Home screen
Main Display Line Second Display Line Reference Fields
Application Buttons
Functions
Auto Optimization improves counting accuracy: it re-calculates the piece weight automatically as parts are added.
By default Auto Optimization is Off.

4.2.1.1 Edit Settings: To view or adjust the current settings, touch the Edit Settings button:
The Settings screen is now displayed.
Settings available: APW, Sample Size, Sample Name, Sample ID, Batch ID, Clear All.
Note: please refer to section 4.1.1 for details about setting Sample Name, Sample ID, Batch ID and Clear All.
Functions available: Recalculate APW, Return to Application
Establish an Average Piece Weight (APW)
Each time a new type of part is counted, the nominal weight of one piece (Average Piece Weight or APW) must be established using a small quantity of pieces. This APW is stored until replaced by another APW.
There are two methods to establish the APW value:

  1. The actual APW is known; or
  2. The APW must be derived by weight. For this case the current sample size will be used To choose a different value, establish it beforehand, using the Sample Size button. (See below.)

Set a known Average Piece Weight (APW)
To adjust the APW value directly, touch the APW button.
A numeric input window appears.
Key in the desired APW Weight, then press .
The display returns to the previous screen with the new APW value highlighted.

Set a new Average Piece Weight (APW) – Derived
To establish a new APW, touch the Recalculate APW button.
The APW Screen is displayed.
Note: The sample size that is displayed will be used.
To use a different sample size, change that first. (See below.)
Follow the screen instructions to establish a new APW.
Actions available: Tare, Cancel, or Accept
Press Accept to capture the value and return to the Settingsscreen.
The Settings screen returns with the new APW value highlighted. To return to the Counting home screen, touch Return to Application.
The home screen shows 10 pieces at the new APW

To return to the Counting home screen, touch Return to Application.

4.2.1.2 Sample Size
The sample size can be 1 to 10 000 pieces. The default sample size is 10. Once a sample size is changed, the balance will immediately open the Recalculate APW screen, expecting to establish a new APW.
The current sample size is shown on the Settings screen. To adjust the sample size, touch the Samples button.

A numeric input window appears.
Key in the desired Sample Size, then press √.
The next screen appears, with the message to place the number of pieces entered in the numeric input window.

Place the indicated number of pieces on the platform.
Then touch either Tare, Cancel, or Accept.
Press Accept to capture the value and return to the Settings screen. The Settings screen appears with the new APW value highlighted.
To begin counting pieces, touch Return to Application.
4.2.2 Application Setup
The Application can be customized for various user preferences.
Touch the Setup Icon (wrench) to enter Application Setup.

The Application Setup Menu appears.
Select the list item to view or change the setting as desired.
To return to the Application home screen, touch Done.

The Counting Application Setups are below (defaults in Bold).

Item Available Settings Comments
Enabled On, Off Turns Application on/off
Auto Optimization On, Off
Counting Mode Standard Counting, Check Counting,
Fill Counting Select the Mode
Secondary Unit On, Off To show the Second Display line
Statistics Off. Manual, Auto
Lock Settings On, Off Press to lock the current setting(s) so they cannot be

changed.
Reference fields| |
APW| On, Off| The APW is displayed
Gross| On, Off| Gross weight is displayed
Net| On, Off| Net weight is displayed
Tare| On, Off| Tare weight is displayed
Target| On, Off| Used for Check-Fill only
Difference| On, Off| Used for Check-Fill only
Over Limit| On, Off| Used for Check-Counting only
Under Limit| On, Off| Used for Check-Counting only
Print Options| | See Printing Section.
APW| On, Off|
Sample Size| On, Off|
Target| On, Off| Used for Check-Fill only
Difference| On, Off| Used for Check-Fill only
Over Limit| On, Off| Used for Check-Counting only
Under Limit| On, Off| Used for Check-Counting only

4.2.3 Parts Counting – Check
This feature permits establishing a set quantity of pieces as criteria for similar items that can be quickly checked against a sample.

The PARTS COUNTING – Check Home screen
Main Display Line
Second Display Line
Reference Fields
Application Buttons
Functions

The default (or last) Average Piece Weight (APW) is displayed. Place objects on the pan to display the Check Status (number of pieces).
4.2.3.1 Edit Settings: To use the Parts Counting-Check mode, the Over Limit and Under Limit values must be set.
The Average Piece Weight and standard sample size can also be set. Follow these steps:
Touch the Edit Settings button. The Settings screen appears.
The buttons for APW, Samples, Sample Name, Sample ID, Batch ID, Clear All.
Note: please refer to section 4.1.1 for details about setting Sample Name, Sample ID, Batch ID and Clear All.
Over Limit and Under Limit show default (or previously set) values. To change any value, touch the button. A keyboard screen appears.

(See keyboard screens, next page.)
To set a new average piece weight, press the Recalculate APW button.
A new screen appears with the message: Place reference weight on the pan. Press Accept to continue.
After following these instructions, press Return to Application on the main Edit Settings screen. OHAUS 83021171 Explorer Semi Micro Balance - shown
here 73

4.2.4 Parts Counting – Fill
This feature permits establishing a set quantity of pieces as a reference, so when pieces are added, their percentage of the reference amount appears. A Progress bar indicates the percentage. In the upper left portion of the home screen, select Parts Counting. (The option will appear if it has been enabled in the Application Modes menu.)

he PARTS COUNTING – Fill Home screen
Main Display Line
Second Display Line
Reference Fields
Application Buttons
Functions

The default (or last) Average Piece Weight (APW) is displayed.
Place objects on the pan to display the Fill Status.
4.2.4.1 Edit Settings: Touch the Edit Settings button.
The Settings screen appears.
The buttons for APW, Samples and Target show default (or previously set) values.
To change values, touch the button.
A keyboard screen appears.
Key in the new value, and press .
Note: please refer to section 4.1.1 for details about setting Sample Name, Sample ID, Batch ID and Clear All.
To set a new average piece weight, press Recalculate APW. A new screen appears saying: Place reference weight on the pan. Press Accept to continue.
This establishes a new reference weight with a sample.
Alternatively, fill in values in the screens for APW, Samples and Target, as shown below.
After following these instructions, and setting all values as desired, press Return to Application.OHAUS 83021171 Explorer Semi Micro Balance - shown
here 79

4.2.5 Statistics for Parts Counting
Use the Statistics function to compare a number of samples and examine the relative deviation of the samples along with other statistical data. A minimum of three samples is required.
Statistics can be determined manually (Manual mode, with key press after each step) or automatically (Auto Mode: weights are automatically recorded when stable).
To use the statistics function, enable it first in the Application Setup menu.
Application Setup
First, Enable StatisticsTouch the Setup
Icon wrench (as above) to enter Setup menu.

OHAUS 83021171 Explorer Semi Micro Balance - shown here
80

Press Statistics and choose either Manual or Automatic, then press Close.
Press Done on the main options menu to return to the home screen.
Accumulating displayed values & Viewing data – Manual
Once Statistics is enabled, two buttons appear on the Application screen:39 Statistics and Accumulate.
To add the sample to the data set, place the item on the pan and press Accumulate.
When stable, the Statistics button lights briefly to show the item is added to the set and the data set quantity increments by one.

Continue to add items and press Accumulate to build the statistical data set At any time, press the Statistics button to view the results.   The Statistics Results screen
Press Print to send to the enabled COM port
Press Close to return to the Application home screen with the current statistics set or
Press Clear Data to clear the results and return to the Application home screen.

Statistics – Automatic
Statistics Automatic mode is similar to manual mode (above) except the first stable weight is automatically captured. Thus the Accumulate button is not used.
4.3 Percent Weighing
Use Percent Weighing to measure the weight of a sample displayed as a percentage of a pre-established Reference Weight.
The default (or last) Reference Weight is displayed.
Place an object on the pan. The difference between the sample and the Reference Weight is displayed as both a weight and a percentage.
To Access Percent Weighing: Press the upper left portion of the home screen, so the available Application Modes are visible.
Select Percent Weighing.

The Percent Weighing Home Screen appears.

The PERCENT WEIGHING Home screen
Main Display Line
Second Display Line
Reference Fields
Application Buttons
Functions

4.3.1 Edit Settings
To view or adjust the current settings
To adjust the Application settings, touch the Edit Settings button.
(See above.)
The Settings screen is now displayed.
Settings available: Reference, Reference Adjust, Sample Name, Sample ID, Batch ID, Clear All.
Note: please refer to section 4.1.1 for details about setting Sample Name, Sample ID, Batch ID and Clear All. Functions Available: Recalculate Reference,
Return to Application

Establishing a Reference Weight
There are two methods to establish the Reference Weight value:

  1. Direct: If the actual Reference Weight is known, it can be entered directly;
  2. Derived: The sample’s weight can be used to establish the reference weight.

Establishing a Known Reference Weight
The current Reference weight value is shown on  the Settings screen.To adjust the
Reference weight directly, touch the Reference button. A numeric input window appears.  Key in the desired Reference Weight, then press .
The display returns to the previous screen with the new value highlighted.
To return to the Percent Weighing home screen, touch Return to Application.
Using a Sample to Establish a Derived Reference Weight To establish a new Reference weight value, touch the Recalculate Reference button.
The New Reference screen appears. Follow  the screen instructions to establish a new Reference Weight.

Actions available: Tare, Cancel, or Accept.
Press Accept to capture the value and return to the Settings screen.
The display returns to the Settings screen with the new Reference weight value highlighted momentarily.
To return to the Percent weighing home screen, touch Return to Application.
Reference Weight Adjustment (Reference Adjust)
The Reference Weight adjustment can be 1 to 1000 percent. Examples: 100% means the sample is displayed as compared to 100% of the Reference Weight.
25% means the sample is displayed as compared to 25% of the Reference Weight.
The Edit Settings screen shows current Reference Adjust value. To change it, touch Reference Adjust button. A numeric input window appears. Key in the desired  Reference Weight Adjustment ,then press √.

The display returns to the previous screen with the new value highlighted momentarily.
To return to the Percent weighing home screen, touch Return to Application.
4.3.2 Application Setup
The Application can be customized for various user preferences. Press Setup (wrench) from the home screen. The Setup Menu appears.
Select the item to view or change.
To return to the Application home screen, press Done.

Percent Weighing Application Setups (defaults in Bold)

Item Available Settings Comments
Enabled On, Off Turns Application on/off
Secondary Unit On, Off To show the Second Display line
Lock Settings On. Off Press to lock the current setting(s) so they cannot be

changed.
Reference fields| |
Reference Weight| On, Off| The Reference is displayed
Difference (weight)| On, Off| Displays the Difference to the reference weight.
Difference (percent)| On, Off| Displays the Difference to the reference weight as percentage.
Gross| On. Off| Gross weight is displayed
Net| On. Off| Net weight is displayed
Tare| On. Off| Tare weight is displayed
Print Options| | See Printing Section.
Reference Weight| On, Off|
Reference Adjust| On, Off|
Difference (weight)| On, Off|
Difference (percent)| On, Off|

4.4 Check Weighing
Check Weighing is used to compare the weight of a sample against target limits. Three different modes can be selected: Over and Under, Nominal Weight Tolerance, or Nominal Percent Tolerance.
4.4.1 Standard Check Weighing (default)
In the upper left portion of the home screen, select Check Weighing.
The default (or last) Check weigh limits are displayed. Place objects on the pan. The Under/Accept/Over status is shown in the progress bar area while the actual weight of the item is shown on the main Display Line.

The CHECK WEIGHING Home screen
Main Display Line
Second Display Line
Reference Fields
Application Buttons
Functions

4.4.2 Edit Settings
To view or adjust the current settings
Touch the Edit Settings button. The Settings screen appears.
The Application Settings Screen.
Settings available: Over Limit, Under Limit, Sample Name, Sample ID, Batch ID, Clear All.
Note: please refer to section 4.1.1 for details about setting Sample Name, Sample ID, Batch ID and Clear All.
Functions Available: Return to Application The check limit values are displayed in the settings screen.
To set the Over Limit value, touch the Ove r Limit button To set the Under Limit value, touch the Under Limit button A numeric input window appears.
Enter the desired Limit Weight, then press .
The display returns to the previous screen with the new value highlighted momentarily.
To return to the CHECK WEIGHING home screen, touch Return to Application.

EXPLORER BALANCES

4.4.3 Application Setup
The Application can be customized for various user preferences.
Touch the Setup Icon (wrench) to enter the Application Setup from the home screen:
The Setup Menu is now displayed.
Select the list item to view or change the setting as desired.
To return to the Application home screen, touch Done.

Check Weighing Application Setups (defaults in Bold)

Item Available Settings Comments
Enabled On, Off Turns Application on/off
Limits Setting Mode Over and Under, Nominal/Weight Tol, Nominal/Percent Tol.

To select the Mode
Secondary Unit| On, Off| To show the Second Display line
Audible Signal| On, Off| Beeps if reading is under or over.
Statistics| Off, Manual, Auto|
Display Setting| Weight, Check Status| To show Weight or Check Status on Main Display Line
Lock Settings| On, Off| Press to lock the current setting(s) so they cannot be changed.
Reference fields| |
Over Limit| On, Off| Over weight limit
Under Limit| On, Off| Over weight limit
Gross| On, Off| Gross weight is displayed
Net| On, Off| Net weight is displayed
Tare| On, Off| Tare weight is displayed
Target| On, Off| Used for Nominal Checkweigh
Positive Tolerance| On, Off| Used for Nominal Checkweigh
Negative Tolerance| On, Off| Used for Nominal Checkweigh
Print Options| | See Printing Section.
Over Limit| On, Off|
Under Limit| On, Off|
Target| On, Off| Used for Nominal Checkweigh
Positive Tolerance| On, Off| Used for Nominal Checkweigh
Negative Tolerance| On, Off| Used for Nominal Checkweigh

4.4.4 Check Weighing – Nominal/Weight Tolerance Mode
With this feature you can establish a Target or Nominal weight value and pre- set tolerance in weight.
To enable this feature, enter the Setup menu.

Change the mode to Nominal/wt. Tolerance.
The Check Weighing options screen returns.
Press Done to begin Check Weighing.

CHECK WEIGHING – Nominal/Wt Tolerance home screen
Main Display Line
Second Display Line (not shown)
Reference Fields
Application Buttons
Functions

To use this mode, touch Edit Settings to set the Nominal, and Upper and Lower Tolerance values (in weight). (See keyboards below.)
Press Return to Application when settings are complete. Note: please refer to section 4.1.1 for details about setting Sample Name, Sample ID, Batch ID and Clear All.OHAUS 83021171 Explorer Semi Micro Balance - Done to begin
4

4.4.5 Check Weighing – Nominal/Percent Tolerance Mode
With this feature you can establish a Target or Nominal weight value and pre- set tolerance in percent.
To enable this feature, enter the Setup menu.

Change the mode to Nominal, % Tolerance. The Check Weighing options screen returns.
Press Done to begin Check Weighing.

The CHECK WEIGHING – Nominal/Percent Tolerance home screen
Main Display Line
Second Display Line (not shown)
Reference Fields
Application Buttons
Functions

To use this mode, touch Edit Settings to set the Target Weight, and Upper and Lower Tolerance values (in percent). (See keyboards below.)
Press Return to Application when settings are complete.
Note: please refer to section 4.1.1 for details about setting Sample Name, Sample ID, Batch ID and Clear All.

OHAUS 83021171 Explorer Semi Micro Balance - Done to begin
7

4.4.6 Display Settings
The results can be displayed with the check status (Under, Accept, or Over) either in the Progress bar (Weighing) or in the main Weighing Line (Check Status).
To display the results in the main Weighing Line, set the Display Setting to Check Status, using the Application Setup screen. (The default value is Weight.)

The CHECK WEIGHING home screen withDisplay set to Check Status
Main Display Line
Second Display Line
Reference Fields
Application Buttons
ApIcon
Functions

To enable audible signals for Under, Accept, and Over, Touch Audible Signal-On in the Application Setup screen.

4.4.7 Statistics for Check Weighing
Use the Statistics function to compare a number of samples and examine the relative deviation of the samples along with other statistical data. A minimum of three samples is required.
Statistics can be determined manually (with key press after each step) or automatically (weights recorded automatically when stable). To use the statistics function, enable it first in the Application Setup menu.
Application Setup First, Enable Statistics.

Touch the Setup Icon wrench to enter Setup menu.  Touch Statistics. Choose either Manual or Automatic, then press Done. Accumulating Displayed Values & Viewing Data – Manual
Once Statistics is enabled, two buttons appear on the Application screen: Statistics and Accumulate.
To add the sample to the data set, place the item on the pan and press Accumulate.
When stable, the Statistics button momentarily lights to show the item is added to the set and the data set quantity increments by one.
Continue adding items and pressing Accumulate to build the statistical data set
At any time, press the Statistics button to view the results.
The Statistics Results screen
Press Print to send to the enabled COM port Press Close to return to the Application home screen with the current statistics set.
Press Clear Data to clear the results and return to the Application home screen.

Statistics – Automatic
Statistics Automatic mode is similar to manual mode (above) except the first stable weight is automatically captured. Thus the Accumulate button is not used.
4.5 Dynamic/Animal Weighing
Use this application to weigh an unstable load, such as a moving animal. Three different start/reset modes can be selected: Manual (start and stop via key press), Semi-Automatic (auto-start with manual reset), and Automatic (start and stop automatically).
Dynamic Weighing – Manual (default) To set the mode, touch the Setup wrench: When you make a selection, the Setup options screen re-appears.
Press Done. On the Dynamic Weighing home screen, the default (or last) Averaging Time is displayed. Follow the screen instructions.
The DYNAMIC WEIGHING Home screen
Main Display Line
Second Display Line
Reference Fields
Application Buttons
Functions

Place objects on the pan.
Press the Begin Dynamic button.
The balance begins a countdown (averaging process).
During the countdown, the information line displays the time remaining.
If necessary, press Cancel to quit and return to the Home screen.
When the countdown ends, the result is displayed and held. Press Reset to clear the held weight and return to the Initial screen.
Note: Clear the pan before beginning a new Dynamic Weighing cycle.

4.5.1 Edit Settings
To view or adjust the current settings:
Press the Edit Settings button.
(See Home Screen, previous page.)
The Settings screen appears.
The Application Settings Screen.
Settings available: Averaging Time, Sample Name,
Sample ID, Batch ID, Clear All.
Note: please refer to section 4.1.1 for details about setting Sample Name, Sample ID, Batch ID and Clear All. Functions Available: Return to Application
The Averaging Time can be set from 1 to 99 seconds. The default value is 10. To change it, press the button. A numeric input window is displayed.
Key in the desired Averaging Time, then press Done. The display returns to the previous screen with the new value highlighted.
To return to the Dynamic Weighing home screen, touch Return to Application.

4.5.2 Application Setup
The Application can be customized for various user preferences.
Touch the Setup wrench to enter the Application Setup from the home screen.

The Setup Menu appears.
Select the list item to view or change the setting as desired.
To return to the Application home screen, press Done.

Dynamic Weighing Application Setups (defaults in Bold)

Item Available Settings Comments
Enabled On, Off Turns Application on/off
Auto Sample Mode Manual, Semi Automatic Automatic To select the Mode
Statistics Off, Manual, Auto
Lock Settings On, Off Press to lock the current setting(s) so they cannot be

changed.
Reference fields| |
Averaging Time| On, Off| Time in Seconds
Gross| On, Off| Gross weight is displayed
Net| On, Off| Net weight is displayed
Tare| On, Off| Tare weight is displayed
Print Options| | See Printing Section.
Averaging Time| On, Off|
Auto Print| On, Off| To print automatically after weighing process is finished.

4.5.3 Dynamic Weighing – Semi-Auto
The cycle begins when an object (animal) is placed on the pan, but the held value must be reset manually.
To enable this feature, touch the Setup Icon (wrench) in the top-center of the main screen to enter the
Dynamic Weighing Setup menu.
Change the mode to Semi Automatic.
The Dynamic – Semi-Automatic screen appears with the default (or last) Averaging Time displayed.

Place specimen on the Pan. The cycle starts automatically.
After the cycle ends, the weight reading is held. Press the Reset button to clear the reading and return to zero.

4.5.4 Dynamic Weighing – Automatic
The cycle begins when an object (animal) is placed on the pan, and the held value is automatically reset when the object is removed from the pan (10-second delay).
To enable this feature, enter the Dynamic Weighing Setup menu.
Press Auto Sample, and change the mode to Automatic. The Dynamic – Automatic screen appears with the default
(or last) Averaging Time displayed.
Place objects on the pan to begin.
The cycle starts automatically.
After the cycle ends, remove the sample and the balance returns to the Home screen automatically.

4.5.5 Statistics for Dynamic Weighing

Use the Statistics function to compare a number of samples and examine the relative deviation of the samples along with other statistical data. A minimum of three samples is required. Statistics can be determined manually (Manual mode, with key press after each step) or automatically (Automatic Mode: weights are automatically recorded when stable). To use the statistics function, enable it first in the Application Setup menu.
Application Setup
First, Enable Statistics

Touch the Setup Icon wrench to enter Setup menu.

Press Statistics. Choose either Manual or Automatic mode, then press Done.

Accumulating Displayed Values & Viewing Data – Manual
Once Statistics is enabled, two buttons appear on the Application screen: Statistics and Accumulate.
To add the sample to the data set, place the item on the pan and press Accumulate.
When stable, the Statistics button momentarily lights to show the item is added to the set and the data set quantity increments by one.
Continue adding items and pressing Accumulate to build the statistical data set
At any time, press the Statistics button to view the results.

The Statistics Results screen
Press Print to send to the enabled COM port
Press Close to return to the Application home screen with the current statistics.
Press Clear Data to clear the results and return to the Application home screen.
Statistics – Automatic
Statistics Automatic mode is similar to manual mode (above) except the first stable weight is automatically captured. Thus the Accumulate button is not used. ****

4.6 Filling
This application allows the user to slowly fill a container to a pre- determined target weight. The progress bar displays the filling status, and within 10 percent of the target value the progress bar converts to fine resolution (+/– 10%) for accurate results.
In the upper left portion of the home screen, select Filling. The default (or last) Target weight is displayed. Place objects on the pan to begin.

The FILLING Home screen
Main Display Line
Second Display Line
Reference Fields
Application Buttons
Functions

The progress bar displays the filling status, and within 10 percent of the target value, the progress bar convertsto fine resolution display.
4.6.1 Edit Settings
To view or adjust the current settings:
Touch the Edit Settings button.
The Settings screen appears:  Settings available: Target Weight, Sample Name, Sample ID, Batch ID, Clear All.
Note: please refer to section 4.1.1 for details about setting Sample Name, Sample ID, Batch ID and Clear All.
Functions Available: Recalculate Target Weight, Return to Application
Establish a Target Weight
Each time a new item is used for filling, the Target Weight (fill-to weight) must be established.
There are two (2) methods to establish the Target Weight:

  1. The actual Target Weight is known; or
  2. The Target Weight can be derived from a sample.
    Establish using a Known Target Weight
    The current Target Weight appears on the Settings screen, on the Target Weight button. (See above.)
    To adjust the Target weight value directly, touch the Target Weight button.
    A numeric input window appears.
    Enter the desired Target Weight, then press .The display returns to the previous screen with the new value highlighted.
    Press Return to Application to go to the Filling home screen.

Using a Sample to establish new Target Weight – Derived
To establish a new Target Weight value, press the Recalculate Target Weight button.  The New Target Screen appears, with a message: Place the target weight on the pan. Press Accept to continue.
Place the weight on the pan (or in a tared container).
Press Accept to return to the Settings screen. The display returns to the Settings screen with the new Target Weight value highlighted momentarily.
To return to the Filling home screen, touch Return to Application.
4.6.2 Application Setup
The Application can be customized for various user preferences.
Touch the Setup Icon (wrench) to enter the Application Setup menu from the home screen.

The Setup Menu appears.
Select the list item to view or change the setting as desired.
To return to the Application home screen, touch Done.

The Filling Application Setups (defaults in Bold)

Item Available Settings Comments
Enabled On, Off Turns Application on/off
Secondary Unit On, Off To show the Second Display line
Statistics Off, Manual, Auto
Lock Settings On, Off Press to lock the current setting(s) so they cannot be

changed.
Reference fields| |
Target Weight| On, Off| The Target weight is displayed
Difference (Weight)| On, Off| Displays the Difference to the Target weight
Difference (Percent)| On, Off| Displays the Difference to the Target weight
Gross| On, Off| Gross weight is displayed
Net| On, Off| Net weight is displayed
Tare| On, Off| Tare weight is displayed
Print Options| | See Printing Section.
Target Weight| On, Off|
Difference (Weight)| On, Off|
Difference (Percent)| On, Off|

4.6.3 Statistics for Filling
Use the Statistics function to compare a number of samples and examine the relative deviation of the samples along with other statistical data. A minimum of three samples is required. Statistics can be determined manually (Manual mode, with key press after each step) or automatically (Automatic Mode: weights are automatically recorded when stable). To use the statistics function, enable it first in the Application Setup menu. Application Setup
First, Enable Statistics

Touch the Setup Icon wrench (see below) to enter Setup menu.
Touch Statistics. Choose either Manual or Automatic mode, then press Done.

Accumulating Displayed Values & Viewing Data – Manual
Once Statistics is enabled, two buttons appear on the Application screen: Statistics and Accumulate.
To add the sample to the data set, place the item on the pan and press Accumulate.
When stable, the Statistics button momentarily lights to show the item is added to the set and the data set quantity increments by one.
Continue adding items and pressing Accumulate to build the statistical data set.
At any time, press the Statistics button to view the results.. The Statistics Results screen
Press Print to send to the enabled COM port
Press Close to return to the Application home screen with the current statistics set.
Press Clear Data to clear the results and return to the Application home screen.

Statistics – Automatic
Statistics Automatic mode is similar to manual mode (above) except the first stable weight is automatically captured. Thus the Accumulate button is not used.
4.7 Totalization
Totalization measures the cumulative weight of a sequence of items. Two different Totalization accumulation modes can be selected: Manual accumulation (requires key press) and Automatic accumulation. The cumulative total may exceed the capacity of the Balance.
4.7.1 Totalization – Manual (Default)
Touch the upper left portion of the home screen.
A drop-down menu appears. Select Totalization if it is present.
(If not, touch the Menu icon in the bottom-left corner, then choose Application Modes, scroll down to find Totalization, press it, then press Enabled to turn it On. It will now appear in the Application Modes menu on the home screen. Place objects on the pan to begin.

The TOTALIZATION Home screen
Main Display Line
Second Display Line
Reference Fields
Application Buttons
Functions

The sample weight is shown on the Main Display Line.
Press Accumulate to add the weight of the item to the total.
When stable, the new total is shown on the secondary Display LineRemove the item from the weighing pan, then add the next item and continue as above.
When finished, press the Clear Total button to reset the Second Display Line to zero.

4.7.2 Application Setup
The Application can be customized for various user preferences. Touch the Setup Icon (wrench) to enter the Application Setup from the home screen.

The Setup Menu appears.
Select the list item to view or change the setting as desired.
To return to the Application home screen, touch Done.

The Totalization Application Setups (defaults in Bold)

Item Available Settings Comments
Enabled On, Off Turns Application on/off
Auto Sample Mode On, Off To select the mode
Reference fields
Samples On, Off To display the number of samples weighed
Average (weight) On, Off Displays the Average
Standard deviation On, Off Displays the standard deviation
Minimum (weight) On, Off Displays the Minimum value
Maximum (weight) On, Off Displays the Maximum value
Range On, Off Displays the Range
Gross On, Off Gross weight is displayed
Net On, Off Net weight is displayed
Tare On, Off Tare weight is displayed
Print Options See Printing Section.
Samples On, Off
Average (weight) On, Off
Standard Deviation On, Off
Minimum (weight) On, Off
Maximum (weight) On, Off
Range On, Off

4.7.3 Totalization – Automatic
The weight of the item is added automatically (when stable).
To enable this feature:

  1. Touch the Setup Icon (wrench) to enter Application Setup.
  2. Press Auto Sample to turn this feature On.

Clear the pan (if necessary). Add a new sample.
The item weight is shown on the Main Display Line.
The weight of the item is automatically added (when stable) to the total. The new total is shown on the Second Display Line.
Remove the item from the weighing pan, then add the next item and continue the same as in manual mode.
When finished, press the Clear Total button to reset the Second Display line to zero. 4.8 Formulation
Use this application for compounding and recipe making. The number of components can be 2 to 99.
Formulation has two available modes of operation: Free Formulation and Recipe Formulation.
4.8.1 Free Formulation (default)
This mode of Formulation allows the user to freely add components. A filler material is asked for at the end.
In the upper left portion of the home screen, select Formulation.

The FREE FORMULATION Home screen
Main Display Line
Second Display Line
Reference Fields
Application Buttons
Functions
To start the compounding process, press Begin Formulation.
A prompt says Place component 1 then press Accept.
The item weight is displayed on the main display Line.
Place the first ingredient on the pan (or in a tared container).
Press Accept to store the first component.
A prompt says Place item 2. Press Accept.
Continue adding components and pressing Accept to store the weight of the individual components until the formula is complete.
The Total line shows the total weight of all the components.
Press Print at any time in the process to print the recipe.
Then press End Formula.
The display may prompt for a Filler material.
If the formula calls for a filler material, press OK.
A popup message window appears, prompting for the Filler Material to be added.
Add the filler if needed, then press Accept.
(If no filler is needed, press Accept without adding filler.)
The results are then displayed in a pop-up window.
Free Formulation Results
Press Print to send data to the enabled COM port
Press Close to close the Results window and return to the Application home screen

EXPLORER BALANCES

4.8.2 Application Setup
The Application can be customized for various user preferences.
Touch the Setup icon (wrench) to enter Application Setup from the home screen.
The Setup Menu appears.
Select the list item to view or change the setting as desired.
To return to the Application home screen, touch Done. The Formulation Application Setups (defaults in Bold)

Item Available Settings Comments
Enabled On, Off Turns Application on/off
Mode Free Formulate,
Recipe Based To select the mode
Filler On, Off To prompt for a filler material
Secondary Unit On, Off To show the Second Display line
Reference fields
Total On, Off Total is displayed
Gross On, Off Gross weight is displayed
Net On, Off Net weight is displayed
Tare On, Off Tare weight is displayed
Target On, Off For Recipe Based mode
Difference (weight) On, Off For Recipe Based mode
Difference (percent) On, Off For Recipe Based mode
Print Options See Printing Section.
Total On, Off
Target On, Off For Recipe Based mode
Difference (weight) On, Off For Recipe Based mode
Difference (percent) On, Off For Recipe Based mode

4.8.3 Recipe Formulation
Recipe Formulation requires the formula to be entered first, followed component by component.
To enable this feature, press the Setup icon (wrench) to enter the Formulation Setup menu.
Change the mode to Recipe Based.
Press Done in the main setup options screen.+OHAUS 83021171 Explorer Semi
Micro Balance - fig 1

Press Begin Formulation.
If no items have been previously set up, the Settings screen appears.
If items already exist in memory, Formulation begins.
Press the Add New button.
A new default component is added to the end of the Formula (default Name, Weight).OHAUS 83021171 Explorer Semi Micro Balance - fig
2

To change the default component Name, Press Edit Name.
The alpha-numeric entry screen appears.
EXPLORER BALANCES
Enter the new name, then press .
The display returns to the previous screen with the new value highlighted momentarily.OHAUS 83021171 Explorer Semi Micro Balance - fig
3

The display returns to the previous screen with the new value highlighted momentarily.
Continue adding components as required for your formula.
To change the order of a component in the formula, touch the item to highlight it, then press Move Up or Move Down. To delete a component, touch the item to highlight it, and press Delete Item.
To clear all components, press Clear All. A confirmation is required.
When finished, press Return to Application.OHAUS 83021171 Explorer Semi
Micro Balance - fig 4

Press Begin Formulation. Follow the screen instructions:
Add the required component in the formula.
The component weight is shown on the Main Display Line.
The progress bar displays the fill status. Within 10 percent of the target value the progress bar converts to fine resolution display.
Press Accept to add the weight of the component to the total.
The new total appears on the Second Display Line.
The Difference to the target is displayed for each component (in weight and percent).
Follow the screen instructions to continue adding components and pressing Accept to complete the Formula.
When finished, the Formulation Results screen is displayed.
Formulation Results
Press Print to send the results to the enabled COM port.
To close the Results window and return to the Application home screen, press Close.

4.9 Differential Weighing
Differential weighing stores weight values so samples can be dried or processed and the difference in weight calculated. Up to 99 samples can be stored.
Differential weighing has two operating modes:

  • Auto-Process On, where the initial and final weights are sequentially gathered for a set of items, with no need to return to Item Settings.
  • Auto-Process Off, where initial and final weights are gathered separately for each item, after each item is selected in Item Settings.

  Touch the Setup icon (wrench) to enter Application Setup from the home screen.

OHAUS 83021171 Explorer Semi Micro Balance - fig 7 4.9.1 Item Settings
Press Item Settings to set up items in memory.
To set up the item, Press Add New.![OHAUS 83021171 Explorer Semi Micro Balance

  • fig 8](https://manuals.plus/wp-content/uploads/2023/04/OHAUS-83021171 -Explorer-Semi-Micro-Balance-fig-8.jpg)New items are added with default name and weight information (no initial or final weights).
    To change the default Item Name, press Edit Name.
    The alpha-numeric entry screen appears.
    Enter the new name, then press .
    The display returns to the previous screen with the new value highlighted momentarily. Continue adding items as required using Add New.
    New items can be added at any time.
    To delete an item, touch the item to highlight it, then press Delete Item.
    To Clear the item initial weight press Clear Initial Weight.
    (This feature is disabled if no initial weight is stored.)
    To Clear the item’s final weight, press Clear Final Weight.
    (This feature is disabled if no final weight is stored.)
    To clear all items, press Clear All, then confirm.
    When finished, press Return to Application.
    On the main Application screen, follow the screen instructions:
    Add the first item, then press Accept to store the value.
    The initial weights of the defined item are recorded first, then the final weights of the items are collected next.
    Once the item’s initial weight and a final weight are stored, the values are displayed along with the Difference in weight and percent.
    The items are stored in memory. To begin a new Differential weighing session, delete the initial and/or final weights.OHAUS 83021171 Explorer Semi Micro
Balance - fig 14

4.9.2 Application Setup
The Application can be customized for various user preferences.
Touch the Setup icon (wrench) to enter Application Setup from the home screen.
The Setup Menu appears.
Select the list item to view or change the setting as desired.
To return to the Application home screen, touch Done.

Reciprocal Proportion
When it is turned on, Diff.(wt.) and Diff.(%) will be counted as the following:
Diff (wt) = Initial weight – Final weight
Diff (%) = (Initial weight – Final weight) / Initial weight
When it is turned off, the result will be counted by Final weight minus Initial weight.
Absolute Value
When it is turned on, the result of Diff.(wt.) will get an absolute value. It will become positive if it is negative in the first place.
The Differential Weighing Application Setups (defaults in Bold)

Item Available Sett Comments
Enabled On, Off Turns Application on/off
Auto-Process Mode On, Off To select the mode
Reference fields
Initial Weight On, Off The initial weight is displayed
Final Weight On, Off The final weight is displayed
Difference (Weight) On, Off Displays the Difference to initial
Difference (Percent) On, Off in weight
Gross On, Off Displays the Difference to initial
Net On, Off in percent
Tare On, Off Gross weight is displayed
Print Options Net weight is displayed
Initial Weight On, Off Tare weight is displayed
Final Weight On, Off See Printing Section.
Difference (Weight) On, Off
Difference (Percent) On, Off
Reciprocal Proportion On, Off
Absolute Value On, Off Absolute value is displayed

4.9.3 Differential Weighing with Auto-Process On
Differential weighing with Auto Process On requires the initial and final weights to be sequentially gathered for a single item.
To enable this feature, touch the Setup icon (wrench) from the home screen to enter Application Setup.

Press Auto Process to turn it On. Follow the screen instructions to begin. OHAUS 83021171 Explorer Semi Micro Balance - fig 12

Follow the screen instructions:
Press Item Settings to set up items in memory if necessary.
Use the buttons as necessary.
Follow the same steps as shown on page 57.
Press Return to Application when ready.
A message prompts for Item 1 to be placed on the pan again, then to press Accept to establish Final Weight.![OHAUS 83021171 Explorer Semi Micro Balance

4.10 Density Determination
The Explorer can be used to determine an object’s density. Four types of density determination can be made:

  1. Solids more dense than water
  2. Solids less dense than water
  3. Liquid density
  4. Porous material (impregnated with oil)

A density determination kit is designed to be used with Ohaus Explorer balances. Illustrations in this procedure refer to the density kit. However, you may use whatever lab apparatus that suits the requirements for density measurements. The balance software includes a built-in reference density table for water at temperatures between 10°C and 30°C. Review this entire section before attempting density measurements. When making density measurements the material should weigh at least 10 mg on an analytical balance and 100 mg on a precision balance.
Density can be determined manually (with a key press after each step) or automatically (weights are automatically recorded). Before making density measurements, establish the Application Settings.OHAUS 83021171 Explorer
Semi Micro Balance - fig16

4.10.1 Measuring the Density of a Solid Using Water (default)
Press the Setup button (the wrench) to open the Density
Determination Settings screen.
Confirm the following Setups are selected:

  •  Density Type: Solid
  • Liquid Type: Water
  • Porous Material: Off
    Press Done to return to the Density Determination home screen.
    The Edit Settings Screen.
    Settings available: Water Temp, Sample Name, Sample ID,
    Batch ID, Clear All.
    Note: please refer to section 4.1.1 for details about setting
    Sample Name, Sample ID, Batch ID and Clear All.

Functions Available: Return to Application To adjust the water temperature value, touch the Water Temp button.
The balance calculates water density based on the water OHAUS 83021171
Explorer Semi Micro Balance - fig 17temperature value entered (internal lookup table). Measure the actual water temperature using a precision thermometer.
A numeric input window appears.
Enter the actual water temperature, then press Done.
The display returns to the previous screen with the new value highlighted.
To return to the Density Determination home screen, touch Return to Application.
To begin the Density Determination process, press Begin Density Determination. (See below.)
OHAUS 83021171 Explorer Semi Micro Balance - fig 20Once the necessary weights have been determined, the density of the sample is displayed in g/cm3(along with the weight in air, weight in water) on the Application screen.
Press Begin Density Determination to reset all the weight values and restart the process.
4.10.2 Application Setup
The Application can be customized for various user preferences.

OHAUS 83021171 Explorer Semi Micro Balance - fig21

The Setup Menu appears.
Select the list item to view or change the setting as desired.
To return to the Application home screen, touch Done.
The Density Application Setups (defaults in Bold)

Item Available Settings Comments
Enabled On, Off Turns Application on/off
g/cm3  0.01mg balance model: 0.1g/cm 3 , 0.01g/cm 3, 0.001g/cm 3 ,

0.0001g/cm 3 , 0.00001g/cm 3
 0.1mg balance model: 0.1g/cm 3 , 0.01g/cm 3 , 0.001g/cm 3 , 0.0001g/cm 3
 1mg balance model: 0.1g/cm 3 , 0.01g/cm 3 , 0.001g/cm 3
 0.01g balance model: 0.1g/cm 3 , 0.01g/cm 3
 0.1g balance model: 0.1g/cm 3| To select the display resolution of Density’s weighing result.
Density Type| Solid, Liquid| To select either solid or liquid
Liquid Type| Water, Other| Liquid used during measurements
Porous Material| On, Off| To select Porous material
Auto Print Result| On, Off| To automatically print results
Auto Sample Mode| On, Off| To select the mode
Reference fields| |
Liquid Density| On, Off| Test liquid’s density is displayed
Water Temperature| On, Off| The water temp is displayed
Oil density| On, Off| The test oil’s density is displayed
Weight in Air| On, Off| The sample weight is displayed
Oiled Weight| On, Off| The sample weight is displayed
Weight in Liquid| On, Off| The sample weight is displayed
Sinker volume| On, Off| Displays volume of the Sinker
Print Options| | See Printing Section.
Porous Material| On, Off|
Liquid Type| On, Off|
Liquid Density| On, Off|
Water Temperature| On, Off|
Weight in Air| On, Off|
Weight in Liquid| On, Off|

 4.10.3 Measuring the Density of a Buoyant Material Using Water
Press the Setup button (the wrench) to open Settings screen.
Confirm the following Setups are selected:
 Density Type: Solid
 Liquid Type: Water
 Porous Material: Off
Press Done to return to the Density Determination home screen.
Follow the same procedure as Solid Material above, except in Density Determination step 2, push the sample down into the liquid until it is fully submerged.
4.10.4 Measuring the Density of a Solid Using an Auxiliary Liquid
To enable this feature, enter the Density Setup menu and select the following: Density Type: Solid; Liquid Type: Other; Porous Material: Off. (See Application Setup.) OHAUS 83021171 Explorer Semi Micro Balance - fig
23

Confirm the default values displayed (Liquid Density, etc) are correct.
To edit the default values, touch Edit Settings and the application Settings screen appears.
Settings available: Liquid Density, Sample Name, Sample ID, Batch ID, Clear All.
Note: please refer to section 4.1.1 for details about setting Sample Name, Sample ID, Batch ID and Clear All.
Functions Available: Return to Application
To adjust the Liquid Density value, touch the Liquid Density button.
Measure the actual liquid density of the Density setup and enter it.
A numeric input window appears.
Key in the density in g/cm 3 , then press .
The display returns to the previous screen with the new value highlighted. OHAUS 83021171 Explorer Semi Micro Balance - fig24To return to the Density Determination home screen, touch Return to Application.
Begin the Density determination process as per above. 4.10.5 Measuring the Density of a liquid using a Calibrated Sinker (not supplied)
Enable this feature, enter the Density Setup menu and select the following; Density Type: Liquid. (See Application Setup.)
Note: when the Density Type is set to Liquid, the Liquid type and Porous material selections are disabled.![OHAUS 83021171 Explorer Semi Micro Balance

Confirm that the default value displayed (sinker volume) is correct. To edit the default values, touch Edit Settings and the Settings screen is now displayed:
Settings available: Sinker Volume, Sample Name, Sample ID, Batch ID, Clear All.
Note: please refer to section 4.1.1 for details about setting Sample Name, Sample ID, Batch ID and Clear All.
Functions Available: Return to Application To adjust the Sinker volume value, touch the Sinker Volume button.
A numeric input window is displayed.
Key in the desired value, then press .
The display returns to the previous screen with the new value highlighted.
To return to the Density Determination home screen, touch Return to Application.
Press Begin Density Determination to start the process.
OHAUS 83021171 Explorer Semi Micro Balance - fig 28Step 1 of 2 – Weigh the Sinker in Air.Follow the screen instructions, then press Accept to store the sinker weight (“in air”).

Step 2 of 2 – Weigh the Sinker Submerged in the Test Liquid.
Follow the screen instructions, then press Accept to store the sinker weight (submerged in liquid). Once the necessary weights have been determined, the density of the Liquid sample is displayed in g/cm 3  (along with the weight in air, weight in water) on the Application screen.
Press Begin Density Determination to reset all the weight values and restart the process.OHAUS 83021171 Explorer Semi Micro Balance - fig
29

4.10.6 Measuring the Density of Porous Material Using Oil
To enable this feature, enter the Density Determination
Setup menu, and set the following:

  • Density Type: Solid
  • Liquid Type: Water
  • Porous Material: On

OHAUS 83021171 Explorer Semi Micro Balance - fig32

Confirm the default values displayed (Water Temp) are correct.
To edit the default values, touch Edit settings.
The Settings screen appears.
Settings available: Water Temp, Oil Density, Sample Name, Sample ID, Batch ID, Clear All.
Note: please refer to section 4.1.1 for details about setting Sample Name, Sample ID, Batch ID and Clear All.
Functions Available: Return to Application The balance calculates water density based on the water temperature value entered (look-up table).
Measure the actual water temperature using a precision thermometer.
To adjust the Water Temperature or Oil Density values, touch the Water Temp or Oil Density button. Numeric input windows appear.
Key in the desired value, then press .
The display returns to the previous screen with the new value highlighted.
To return to the Density Determination home screen, touch Return to Application.OHAUS 83021171 Explorer Semi Micro Balance -
fig33

Press Begin Density Calculation.
Step 1 of 3 – Weigh the Un-Oiled Sample in Air.
Follow the screen instructions then press Accept to store the dry sample weight (in air).OHAUS 83021171 Explorer Semi Micro Balance -
fig34

Step 2 of 3 – Weigh the Oiled Sample in Air.
Follow the screen instructions then press Accept to store the sample weight (oiled).
Step 3 of 3 – Weigh the Oiled Sample Submerged in Liquid.Follow the screen instructions then press Accept to store the oiled sample weight (submerged in liquid).
Once the necessary weights have been determined, the density of the sample is displayed in g/cm 3 (along with the weight in air, un-oiled and oiled, and weight in water) on the Application screen
The value stays on the display until Begin Density Determination is touched.
Press Begin Density Determination to reset all the weight values and restart the process.
OHAUS 83021171 Explorer Semi Micro Balance - fig 35

4.11 Peak Hold
The Peak Hold application allows the user to capture and store the highest stable or unstable weight in a series of weight measurements.
Three different start/reset modes can be selected:

  • Manual (start and stop the capture via key press),
  • Semi-Automatic (start automatically/reset manually); and
  • Automatic (start and stop the capture automatically).

This application allows both stable and unstable weights to be captured.
4.11.1 Peak Hold – Semi-Automatic (default)
In the upper left portion of the home screen, select Peak Hold.
Follow the screen instructions to begin.OHAUS 83021171 Explorer Semi Micro
Balance - fig 36Place item on the pan. The capture process starts automatically.The highest reading is held on the display.
To release the reading, press End Peak Hold.
The following screen appears.
Clear the pan or press Tare to reset the display and re-start the process.OHAUS 83021171 Explorer Semi Micro Balance - fig
37

4.11.2 Application Setup
The Application can be customized for various user preferences.
Touch the Setup icon (wrench) to enter Application Setup from the home screen.
The Setup Menu appears.
Select the list item to view or change the setting as desired.
To return to the Application home screen, touch Done. OHAUS 83021171
Explorer Semi Micro Balance - fig40The Peak Hold Application Setups (defaults in Bold).

Item Available Settings Comments
Enabled On, Off Turns Application on/off
Auto Sample Mode Manual,

Semi Automatic
Automatic| To select the mode of operation
Stable Weight Only| On, Off| To capture Stable values only
Secondary Unit| On, Off| To show the Second Display line
Reference fields| |
Gross| On, Off| Gross weight is displayed
Net| On, Off| Net weight is displayed
Tare| On, Off| Tare weight is displayed
Print Options| | See Printing Section.
Stable Weight Only| On, Off|

4.11.3 Peak Hold – Manual
For manual start and reset, choose Manual mode.
To enable this feature, enter the Peak Hold Setup menu:
  Change the mode to Manual.OHAUS 83021171 Explorer Semi Micro Balance - fig
42

Follow the screen instructions: Place item on the pan.
Press Begin Peak Hold to begin the capture process.
The following screen appears:OHAUS 83021171 Explorer Semi Micro Balance -
fig43The highest stable weight reading is held indefinitely.
Press End Peak Hold then clear the pan to reset the display and re-start the process.
4.11.4 Peak Hold – Automatic
For automatic start and reset, choose Automatic mode.
To enable this feature, enter the Peak Hold Setup menu.
Change the mode to Automatic.OHAUS 83021171 Explorer Semi Micro Balance -
fig82

Place item on the pan. The capture process starts automatically.OHAUS
83021171 Explorer Semi Micro Balance - fig 83The highest stable weight reading is held until the weight is removed.
The process starts automatically once the weight is removed from the pan (after 10 second delay).
4.12 Ingredient Costing (not available in semi-micro models)
This application allows the user to determine the cost of a formula or recipe based on the known cost and quantity of the components or ingredients. Up to 99 components are allowed.
In the upper left portion of the home screen, select Ingredient Costing. Follow the screen instructions to begin.OHAUS 83021171 Explorer Semi Micro
Balance - fig 1

4.12.1 Item Settings
Press Item Settings to set up items in memory.
The Edit Settings Screen.
Settings available: Add New, Edit name, Edit Cost, Edit Unit, Delete Item, Clear All Functions Available: Return to Application To set up the item(s) in memory, Press Add New.
A new item is added (default Name) with default info (cost per weight and weight unit). To change the default item name, Press Edit Name.
The alpha-numeric entry screen appears.
Enter the new name, then press .
The display returns to the previous screen with the new value highlighted momentarily.
To change the default component cost per unit,
Press Edit Cost.
The numeric entry screen appears.
Enter the new weight, then press .
The display returns to the previous screen with the new value highlighted momentarily. To change the default component Unit, press Edit Unit.
The Weighing Units screen appears.
Continue adding items as required, using Add New.
New items can be added at any time.
To delete an item, touch the item to highlight it, then press Delete Item.
To clear all items, press Clear All. A confirmation is required.
When finished, press Return to Application. Once setup is complete and the application screen is displayed, follow the screen instructions:
Add the first item then press Accept to store the value in memory.
The weight of the item is recorded, then the screen prompts for the next component. Unit Cost, Total Weight and Total Cost are also displayed.

Continue follow the screen instructions to add the items to the pan and touch the Accept button until each item has been weighed.
Press the View Data button at any time to show the current results: (Item, Item Cost and Item Weight). After the last item has been weighed, press View Data to see
the results. (Cancel the current data set by pressing Clear Data.)
Ingredient Costing Results.
Press Print to send the data to the enabled COM port.
Press Close to close the results window and return to the Application home screen.

4.12.2 Application Setup
The Application can be customized for various user preferences.Touch the Setup Icon (wrench) to enter Application Setup from the home screen.OHAUS 83021171
Explorer Semi Micro Balance - fig 8

The Setup Menu is now displayed.
Select the list item to view or change the setting as desired.
To return to the Application home screen, touch Done.
The Ingredient Costing Application Setups (defaults in Bold)

Item Available Settings Comments
Enabled On, Off Turns Application on/off
Currency label None, various To select the Currency label
Reference fields
Unit Cost On, Off To display the Item cost
Total Weight On, Off To display the Total weight
Total Cost On, Off To display the Total cost
Gross On, Off Gross weight is displayed
Net On, Off Net weight is displayed
Tare On, Off Tare weight is displayed
Print Options See Printing Section.
Total Weight On, Off
Total Cost On, Off

 4.13 Pipette Adjustment (not available in EX12001, EX24001, EX35001)
Pipette adjustment checks the accuracy and precision values of pipettes by weight analysis. The balance has a built-in density reference table for water at temperatures between 10°C and 30°C. If other liquids are used for pipette calibration, enter the liquid’s density at current room temperature in g/cm 3 . Since all calculations are made within the balance, the atmospheric pressure is also required.
Pipette adjustment can be accomplished manually (with a key press after each step) or automatically (weights are automatically recorded when stable). Calculations made within the balance provide results of inaccuracy and imprecision. Up to 30 samples can be used. OHAUS 83021171 Explorer Semi
Micro Balance - fig 9

4.13.1 Pipette Adjustment – Manual (default)
Confirm the default values displayed are correct:
Nominal, Inaccuracy, Imprecision, Density, Atmospheric Pressure, Water Temperature, etc.
To edit the default values, touch the Edit Settings button.
The Edit Settings Screen.
Settings available: Nominal Capacity, Water Temp, Inaccuracy, Atmospheric Press,
Imprecision, Pipette Name, Pipette Number
Functions Available: Return to Application
To adjust the pipette Nominal Capacity value, touch the Nominal button.
Enter the Nominal Capacity in the keyboard screen that appears, and press .
The display returns to the previous screen with the new value highlighted.
Note: Capacity can be measured in mL or uL, depending on the Application Setup. To adjust the water temperature, touch the Water Temp button.
The balance calculates water density based on the water temperature value entered. Measure the actual water temperature using a precision thermometer.
A numeric input window is displayed.
Key in the desired temperature, then press .
The display returns to the previous screen with the new value highlighted.
Continue to enter the required information following the above procedure:OHAUS 83021171 Explorer Semi Micro Balance - fig
11

Pipette Inaccuracy, via the Inaccuracy button
Barometric Pressure, via the Barometric Pressure button
Pipette Imprecision, via the Imprecision button
Pipette Name, via the Pipette Name button
Pipette Number, via the Pipette number button
When finished, press Return to Application.
The Application home screen appears, and you can begin the
Pipette Adjustment process.
4.13.2 Begin Pipette Adjustment ProcessOHAUS 83021171 Explorer Semi
Micro Balance - fig 183 Step 1 – Dispense the First Sample
Follow the screen instructions, then press Accept to store the first sample weight (liquid).
Continue to follow the screen instructions by dispensing samples and pressing Accept to store each sample weight (liquid).
The default number of samples is 10.
It can be changed in Setup, from 2 to 30 samples.OHAUS 83021171 Explorer
Semi Micro Balance - fig 20 Note: To view either the results or graph screens again from the home screen, press the View Result button.
To start a new Pipette Adjustment process, press Begin Pipette Adjustment.
4.13.3 Application Setup
The Application can be customized for various user preferences.
Touch the Setup Icon (wrench) to enter the Application Setup from the home screen.
The Setup Menu appears.
  Select the list item to view or change the setting as desired.
To return to the Application home screen, touch Done.

The Pipette Adjustment Application Setups (defaults in Bold)

Item Available Settings Comments
Enabled On, Off Turns Application on/off
Auto Sample Mode On, Off To select the mode of operation
Number of Samples 2 – 30 (default 10) To select the number of samples
Liquid Type Water, Other Liquid used during measurements
Unit of Pressure ATM, inHg, KPa, To select Atmospheric Pressure unit
Volume Unit (Pipette mbar, mmHg, PSIA Nominal capacity unit
capacity unit) ml, μl To show the Second Display line
Secondary Unit On, Off Press to lock the current setting(s) so
Lock Settings On, Off they cannot be changed.
Reference fields
Nominal Volume On, Off Nominal capacity displayed
Inaccuracy On, Off Inaccuracy value displayed
Imprecision On, Off Imprecision value displayed
Liquid Density On, Off Liquid density value displayed
Pressure On, Off Barometric Pressure value displayed
Water Temperature On, Off Water temperature is displayed
Print Options See Printing Section.
Nominal Volume On, Off
Inaccuracy On, Off
Imprecision On, Off
Liquid Density On, Off
Pressure On, Off
Water Temperature On, Off

4.14 Statistical Quality Control (SQC)
The Statistical Quality control (SQC) feature is useful during various types of process filling operations, to monitor and/or control the process to eliminate under- and over-filling.
As samples are weighed and stored in the balance, a Trend Analysis is developed and can be displayed.
Up to 10 batches with up to 80 samples in a batch are visible for quality control purposes. Each batch of samples is shown, with the maximum, minimum, standard deviation and mean values for each batch. An on-going examination of the relative deviation of the samples along with other statistical data can be viewed.
The filling process operation can be effectively monitored by observing the results on the View Trace screen.
SQC collection can be done manually (Manual mode, requires key press after each step) or automatically (Auto Mode, the weights are automatically recorded when stable).
In the upper left portion of the home screen, select SQC. OHAUS 83021171
Explorer Semi Micro Balance - fig 23 4.14.1 Edit Settings
Confirm the default values (Nominal, Inaccuracy, ImprecisionDensity, Atmospheric Pressure, Water Temperature, etc.) are correct.
To view/edit the default values, touch the Edit Settings button.
The Settings screen appears.
Settings available: Nominal Target weight,
+Tolerance 1, –Tolerance 1,
+Tolerance 2, –Tolerance 2
Sample Name, Batch ID
Functions Available: Return to Application
Note: please refer to section 4.1.1 for details about setting
Sample Name and Batch ID.
To adjust the Nominal Target weight value, touch the Nominal button.
A numeric input window appears.
Key in the desired Nominal Weight, then press  The display returns to the previous screen with the new value highlighted.

Continue to enter the required information following the above procedure:
Positive Tolerance 1, via the +Tolerance 1 button.
Negative Tolerance 1, via the –Tolerance 1 button.
Positive Tolerance 2, via the +Tolerance 2 button.
Negative Tolerance 2, via the –Tolerance 2 button.
Tolerances can be in either weight (default) or percent – See Application Setup, Section 4.14.3.
When finished, press Return to Application and the SQC home screen appears.

4.14.2 Begin the SQC ProcessOHAUS 83021171 Explorer Semi Micro Balance -
fig 27 The Process screen is then displayed.
Follow the screen instructions, then press Accept to store the sample weight.
The current number of Samples, along with the Average, Sum, Minimum value, Maximum value, and Range are displayed. Continue to follow the screen instructions by weighing samples and pressing Accept to store each sample’s weight.
The sample size is 10 (default).
It can be changed in Setup from 2 to 130 pieces.
At any time, press End Batch to see the Individual results screen, showing results for the current Batch.
Once all the samples have been weighed, the Results screen appears automatically.
If the COM interface is enabled, press Print to send the results either to a computer or a printer.
Press Close to return to the current Batch process.
A message offers the chance to save the batch, if desired. To view the process results of this batch (and any other saved batch), press the View Batch History button.
The Batch History screen shows the basic information – Maximum, Minimum, Average (mean), and Standard Deviation – for all currently saved batches.
Press Select to see (and print) the Batch Information screen for any saved batch.
Press Clear Batches to reset SQC results and return to the SQC home screen
Press Close to return to the SQC home screen to begin another batch or to view process results.OHAUS 83021171 Explorer Semi Micro Balance - fig
31

4.14.3 Application Setup
The Application can be customized for various user preferences.
Touch the Setup Icon (wrench) to enter Application Setup from the home screen.
The Setup Menu is now displayed.
Select the list item to view or change the setting as desired.
To return to the Application home screen, touch Done. The SQC Application Setups (defaults in Bold)

Item Available Settings Comments
Enabled On, Off Turns Application on/off
Number of Samples 2 – 130 (default 10) To select the number of samples in a

Batch
Tolerance Mode| Weight,
Percent| Tolerance used during measurements
Auto Sample Mode| Manual,
Semi Automatic,
Automatic| To select the mode of operation
Lock Settings| On, Off| Press to lock the current setting(s) so they
Reference fields| | cannot be changed.
Number of Batches| On, Off| The total batches
Number of Samples| On, Off| The total samples
Average| On, Off| The Average value is displayed
Total| On, Off| The Sum value is displayed
Maximum| On, Off| The Maximum value is displayed
Minimum| On, Off| The Minimum value is displayed
Range| On, Off| The Range is displayed
Print Options| | See Printing Section.
Number of Batches| On, Off|
Number of Samples| On, Off|
Average| On, Off|
Total| On, Off|
Maximum| On, Off|
Minimum| On, Off|
Range| On, Off|

4.15 Fill Weight Variation
Fill weight variation application is the workstation to check whether the individual pill weight is in the tolerance of the average of total pill weight. To ensure the consistency of dosage units, each unit in batch should have a drug substance content within a narrow range around the label claim.
As samples are weighed and stored in the balance, you can check the result by press View Result on the screen. A maximum number of 99 samples can be tested. Each sample is shown with weight, difference compared to the average weight, and if it is accepted or failed.
In the upper left portion of the home screen, select Fill Weight Variation.OHAUS 83021171 Explorer Semi Micro Balance - fig
33 4.15.1 Edit Settings
To edit values, touch the Edit Settings button. The Settings screen appears.
Touch the Sample Name button. Enter the desired sample name, and then press.
You can also use bar code scanner or RFID reader to input sample name.
Note: Since there are many brands of RFID device in the market, OHAUS tested and confirmed that below one from RFIDeas (www.RFIDeas.com) is compatible:
RDR-6081AKU-C06. Touch the Batch ID button. Enter the desired ID, and then press.OHAUS 83021171 Explorer Semi Micro Balance - fig
34Touch the Sample ID button. Enter the desired sample name, and then press.
Then touch the Return to Application button to return to the main screen.
Touch the Samples Average button.
A numeric input window appears.
Key in the number of samples, then press.
The default value is 20 pieces.
A new screen appears with the message:
Place samples to average. Press Accept to continue.
Place all samples together on the pan to calculate the sample average weight. Then touch the Accept button to continue.
Touch the +/- Limit % button to set the variation percent of the sample average weight that can be accepted. Sample weight within this value will be counted as accepted.
The default value is 7.50%.
4.15.2 Begin the Fill Weight Variation ProcessOHAUS 83021171 Explorer
Semi Micro Balance - fig 40 After that follow the onscreen instruction to place sample 2 on the pan, and touch the Accept button to continue.
When a sample is placed, the screen will show if the result is Under, Accept or Over the +/- Limit % value you set.
Continue to follow the screen instructions to weigh samples and touch the Accept button to store each sample’s weight.
The maximum sample number is 99.
It can be changed from 1 to 99.
Once all the samples have been weighed, the Results screen appears automatically.
Setup the proper connectivity and press Print to send the results to PC or printer.
See Communication Setup, Section 5.10 for details.
Insert a USB flash drive to the balance if you need to save the result to it. Once the drive is inserted, you will see a Save to USB button. Touch the button to save the result.
Touch Close to return to the current process.
To view the result again later, touch the View Result button.OHAUS 83021171
Explorer Semi Micro Balance - fig 41

4.15.3 Application Setup
The Application can be customized for various user preferences.
Touch the Setup Icon (wrench) to enter Application Setup from the home screen.
The Setup Menu is now displayed.
Select the list item to view or change the setting as desired.
To return to the Application home screen, touch Done.
When Auto Tare is enabled, the sample will be tared automatically after Accept is pressed and the sample weight is stored.
The Fill Weight Variation Application Setups (defaults in Bold) ****

Item Available Settings Comments
Enabled On, Off Turns Application on/off
RFID On, Off Enable to scan RFID code
Lock Settings On, Off Press to lock the current setting(s) so they
Reference cannot be changed.
Difference Weight On, Off
Difference Percent On, Off Weight difference is displayed
Average On, Off Weight difference percentage is displayed
+/- Limit % On, Off Weight average is displayed
Gross On, Off +/- Limit value is displayed
Net On, Off Gross weight is displayed
Tare On, Off Net weight is displayed
Print Options Tare weight is displayed
Average On, Off
+/- Limit % On, Off

4.16 Statistics
Statistics are used to compare a number of samples and examine the relative deviation of the samples along with other statistical data. A minimum of three samples is required. Statistics can operate manually (key press after each step) or automatically (weights automatically recorded when stable). To use the statistics feature, enable it in the Application Setup menu. This feature is available in the application modes marked with √ in the table below.

Application Mode Statistics Function
Basic Weighing
Parts Counting
Percent Weighing
Check Weighing

Dynamic / Animal
Weighing|
Filling|
Totalization| N/A
Formulation| N/A
Differential Weighing| N/A
Density Determination| N/A
Peak Hold| N/A
Ingredient Costing| N/A
Pipette Adjustment| N/A
SQC| N/A
Fill Weight Variation| N/A

4.17 Secondary Unit
Secondary unit allows the display of the weighing result in a second unit. After enable it in each application’s setup menu, touch the displayed second unit on the screen while weighing to change it. This feature is available in the application modes marked with √ in the table below.

Application Mode Secondary Unit
Basic Weighing
Parts Counting
Percent Weighing
Check Weighing

Dynamic / Animal
Weighing| N/A
Filling| √
Totalization| N/A
Formulation| √
Differential Weighing| N/A
Density Determination| N/A
Peak Hold| √
Ingredient Costing| N/A
Pipette Adjustment| N/A
SQC| N/A
Fill Weight Variation| √

4.18 Library
When an item is processed on a regular basis, the item’s data may be stored in memory for future use. This memory is referred to as the Balance’s Library.
The following data is stored for the Application used:

Application
mode| Library
function| Characters
of record
name| Type of data to be stored in
library| Comments| Number of
library records
---|---|---|---|---|---
Basic Weighing| | 25| Minimum Weight. Sample Name. Sample ID. Batch ID| Only Admin can edit min-weight value.| 1000 records without limit to specific approbation
PadsCounting| .| 25| APW, Sample Size. Over Limit. Under Limit. Target Sample PCS. Sample Name. Sample ID. Batch ID| Sample Size can be set to 1-10000 pieces.
Percent Weighing| .| 25| Reference Weight, Reference Adjust. Sample Name. Sample ID. Batch ID| N/A
CheckWeighing| | 25| Over Limit. Under Limit. Nominal Weight. +Weight Tolerance. – Weight Tolerance. +Percent Tolerance. -Percent Tolerance. Sample Name. Sample ID.
Batch ID| N/A
Dynamic/Ani mat
Weighing| | 25| Averaging Time. Sample Name. Sample ID. Batch ID| Averaging Time can be set to 1-99 seconds.
Filling| -,| 25| Target Weight Sample Name, Sample ID. Batch ID| N/A
Totalization| N/A| N/A| N/A| EX can keep totalization records with up to 999 ‘Accumulate’ operations
Formulation| ..1
(only in “Recipe Based’ node)| NIA| Item #, Item Name. Item Weight| Number of items in one record can be 2 to 99.
Differential Weighing| N/A| N/A| N/A| Up to 99 items are
allowed to store in one record.
Density Determinatio n| | 25| Water Temperature. Oil Density. Sinker Volume. Liquid Density. Sample Name. Sample ID.
Batch ID| N/A
Peak Hold| NIA| NA| N/A| N/A
Ingredient Costing| NIA| tsliA| WA| EX can keep last item settings. Up to 99 items are allowed in one formula/recipe.
Pipette Adjustment| | 25| Nominal Volume, Inaccuracy. Imprecision. Water
Temperature. Pressure. Liquid Density. Pipette Name. Pipette Number| Number of samples can be set to 2-30 for one test.
SOC| v| 25| Nominal Weight +Tolerance 1. – Tolerance 1. +Tolerance 2. – Tolerance 2| Up to 10 batches can be stored. Number of samples can be set to 2 – 130 for one batch test.
FIN Weight Variation| N/A| N/A| N/A| EX can keep last item settings. Number of samples can be set to 299 for one test.

4.18.1 Creating a Library Record
To create a Library record, the Settings screen is used. From the home screen, touch the Edit Settings button.
The Settings screen is now displayed.

OHAUS 83021171 Explorer Semi Micro Balance - fig 46

4.18.2 Retrieving a Library Record
Load a Library record from the home screen.
Press the Library bar in the upper right area of the screen.
(It shows the name of the most recently used library file.)
The Library manager popup is now displayed.
Touch the Library Record to highlight it, then touch Recall to load the record’s settings into the Application.
Then press Return to Application to return to the home screen with the Library name displayed and the stored parameters active.
4.18.3 Deleting a Stored Library Record
To delete a stored record, follow “Retrieving a Library Record” above, but press Delete instead of Recall.
4.19 Additional Features
4.19.1 Weigh Below
The Explorer Balance is equipped with a weigh below hook for weighing below the balance.
Note: before turning the balance over, remove the Pan, Pan Support, Wind Ring/Shield and Draft Shield elements (if present) to prevent damage.
CAUTION: Do not place the balance on the pan support cone or Load cell Pins To use this feature, remove power from the balance, then remove the protective cover for the weigh below opening (2 screws). The protective cover is reversible for easy storage. OHAUS 83021171 Explorer Semi Micro Balance -
fig 49

The balance can be supported using lab jacks or any other convenient method. Ensure the balance is level and secure. Power on the Balance, then use a string or wire to attach items to be weighed.

5.1 Menu Navigation
All menu navigation is performed by touching the display. To enter the Menu, touch Menu from any Application Home screen. The Main menu appears, with buttons for Main, Back and Done. Continue touching the appropriate list item to navigate to the Menu items or touch a button to change location.

OHAUS 83021171 Explorer Semi Micro Balance - Fig1

5.1.1 Changing Settings
To change a menu setting, navigate to that setting using the following steps:
Enter the Menu
From any Application screen, Touch Menu.
The Main Menu List appears on the display.
Select the Sub-Menu
Scroll down to the selected item of the Main Menu List
and touch it. The Sub-Menu appears.
Select the Menu Item
Continue until the desired setting is chosen in the
Menu list. Touch the setting to change it.
The changed setting will be displayed as highlighted for 2 seconds to confirm the changed value.

OHAUS 83021171 Explorer Semi Micro Balance - Fig2

Exit the Menu and Return to the Current Application
After the setting is confirmed, touch Done to return to the Application.
Note: at any time the Main, Back & Done buttons can be touched to navigate to the desired area of the menu or return to the current Application.
5.2 Main Menu
The Explorer balance Main menu structure is illustrated below.
OHAUS 83021171 Explorer Semi Micro Balance - Fig3

5.3 Calibration
Explorer Balances offer a choice of six calibration methods: Internal Calibration, Automatic Calibration
(Auto Cal), Calibration Adjust, Span Calibration, User Calibration and Calibration Test.
ExCal models have three calibration methods: Span Calibration, User Calibration and Calibration test.
Note: Do not disturb the balance during any calibration.
5.3.1 Calibration sub-menu

OHAUS 83021171 Explorer Semi Micro Balance - Fig4

5.3.2 Internal Calibration (AutoCal™)
Calibration is accomplished with the internal calibration mass. Internal calibration can be performed at any time, provided the balance has warmed up to operating temperature and is level.
With the Balance turned ON and no load on the pan, touch Auto Cal™ Internal Calibration, and then touch Calibration.
The Balance begins to calibrate.

OHAUS 83021171 Explorer Semi Micro Balance - Fig5

Alternately, touch the shortcut Calibration Icon from any Application home screen to initiate an Internal Calibrations
Note:
For Non-Approved models, when Auto Cal Internal Calibration is turned off, it changes to span calibration.
For Approved models, when Auto Cal Internal Calibration is turned off, it is chidden
The display shows the status, then returns to the current application.
To cancel at any time, press Cancel.
5.3.3 Automatic Calibration
When Automatic Calibration is set ON, the balance performs a self calibration:

  • when it senses a temperature change of 1.5°C
  • or every 11 hours

Auto Cal will automatically calibrate the Balance (using the internal mass) each time there is a change in temperature significant enough to affect accuracy.
5.3.4 Auto™ Interval (Hours)
Enter this menu to set hours for the balance to start
Internal Calibration (Auto Cal™).
The value can be set from 3-11 hours.

5.3.5 AutoCal™ Adjustment
Use this calibration method to adjust the zero calibration point, without affecting the span or linearity calibration.
Calibration Adjust may be used to adjust the result of the Internal calibration by +100 divisions.
Note: Before making a calibration adjustment, perform an Internal Calibration. To verify whether an adjustment is needed, place a test mass equal to the span calibration value on the pan and note the difference (in divisions) between the nominal mass value and the actual Balance reading. If the difference is within +1 division, calibration adjustment is not required. If the difference exceeds +1 division, calibration adjustment is recommended. (See Table 5-1, below.)
Example:

Actual weight reading:
Expected weight reading:
Difference Weight (d):
Difference weight in digits:| 200.014
200.000 (Test mass value)
0.014
–14 (Adjust value)
---|---

To perform a Calibration Adjustment, touch Auto Cal Adjustment from the Calibration Menu; Enter the value (positive or negative divisions) to match the difference noted earlier in the procedure and press√ .
Recalibrate using Internal Calibration. After calibration, place the test mass on the pan and verify that the mass value now matches the displayed value. If not, repeat the procedure until Internal Calibration reading agrees with the test mass.
Once completed, the balance stores the Adjustment value and the display returns to the current application.
5.3.6 Span Calibration
Span calibration uses two calibration points, one at zero load and the other at specified full load (span).
Refer to Table 5.1 for Span values. For detailed calibration mass information please see the specification tables in section 9.
Note: Factory default settings are shown in bold.
TABLE 5-1 Calibration Masses

Model Span Calibration Points Weight Class
EX125D; EX125DM 50g, 100g ASTM Class 1
EX125; EX125M 50g, 100g ASTM Class 1
EX225D; EX225DM 100g, 150g, 200g ASTM Class 1
EX225D/AD; EX225DM/AD 100g, 150g, 200g ASTM Class 1
EX225/AD; EX225M/AD 100g, 150g, 200g ASTM Class 1
EX124 50g, 100g ASTM Class 1
EX224 100g, 150g, 200g ASTM Class 1
EX324, M, N 150g, 200g, 300g ASTM Class 1
EX223 100g, 150g, 200g ASTM Class 1
EX423 200g, 300g, 400g ASTM Class 1
EX623 300g, 400g, 500g, 600g ASTM Class 1
EX1103, M, N 500g, 1000g ASTM Class 1
EX2202 1000g, 2000g ASTM Class 1
EX4202 2000g, 3000g, 4000g ASTM Class 1
EX6202 3000g, 4000g, 5000g. 6000g ASTM Class 1
EX10202, M, N 5000g, 10,000g ASTM Class 1
EX6201 3000g, 4000g, 5000g. 6000g ASTM Class 2
EX10201 5000g, 10,000g ASTM Class 1
EX12001 3000g,5000g,10000g, 12000g ASTM Class 2
EX24001 10000g,15000g,20000g, 24000g ASTM Class 1
EX35001 10000g,20000g.30000g, 35000g ASTM Class 1

With the balance turned ON and no load on the pan, touch Span Calibration to initiate the procedure. Additional calibration values to be used are shown on the display. The best accuracy is achieved using the mass closest to the full span value.
Note: To change the span calibration point, touch the alternate weight shown on the display. Follow the screen instructions and place the specified calibration weight on the scale when prompted to do so. When complete, the display shows the Span calibration status and returns to the current application.
5.3.7 User Calibration
User Calibration is used to calibrate the balance using a user defined mass.
With no load on the scale, press User Calibration from the calibration sub- menu.
The display shows the current full load (span) calibration point and calibration unit of measure.
To change the value to a user-defined weight, touch ENTER VALUE.
Enter the weight via the popup screen. The balance then prompts for the User- defined weight.
Place the specified calibration weight on the balance and the calibration process completes.
The display shows status, then returns to the current application.
5.3.8 Calibration Test
Use Calibration Test to compare a known calibration weight against the stored span calibration data. With no load on the scale, press Cal Test to begin the process.
The balance captures the zero point, then prompts for the span value.
The display shows status, followed by the difference between the current calibration weight and the stored calibration data.
After 3 seconds, the display returns to the current application.
5.4 User Management
Enter this menu to set user profiles and password policy.
5.4.1 User Profiles
To create new users or edit or delete the current users.

To add new user, press the New button and insert a user name, and then press√.

OHAUS 83021171 Explorer Semi Micro Balance - Fig8

To edit a user, select the user you want to edit and press the Edit button.
You can change User Name, Password and Group for each user.
Note: to change the password of the currently logged-in user, please go to the User
Settings menu. Pleaser refer to section 5.5.2 Change Password for details.

OHAUS 83021171 Explorer Semi Micro Balance - Fig9

To edit the password for the user, press the Password button and input a new password. After that press√.
Password must contain 8~10 characters with combination of letters and numbers.

To delete a user, select the user you want to delete and press Delete. Then press Yes to confirm.

For Group selection, you can set up to 100 Operator, 10 Supervisor, 2 Log Viewer and 1 Admin who is by default in total. Please check the following table for the permission of each role.
Admin can create, edit or delete Supervisor, Operator and Log Viewer, while Supervisor can create, edit or delete Operator. Operator and Log Viewer cannot access User management menu.

Settings and Application| Admin| Supervisor| Operator| Log Viewer
---|---|---|---|---
User management| √| √| x| x
Password policy| √| x| x| x
User settings| √| √| √| x
Balance setup| √| √| x| x
Calibration setting| √| √| x| x
External calibration| √| √| √| x
Internal calibration| √| √| √| x
Date & Time| √| x| x| x
Application modes settings| √| √| x| x
Turn a unit ON/OFF| √| √| x| x
Device & Print settings| √| √| x| x
GLP| √| √| x| x
Balance name| √| x| x| x
Edit library| √| √| √| x
System log| √| √| x| √
Diagnostic| √| √| x| x
Routine test| √| √| √| x
Factory reset| √| √| x| x
Edit Settings (when library record is activated)| √| √| x| x
Edit Settings| √| √| √| x
Minimum Weight| √| √| x| x

5.4.2 Password Policy
To set when password will expire. There are three selections, 30 days, 60 days and 90 days.
This function can also be turned off.
5.4.3 Auto Standby (minutes)
To set the time period for the balance to enter standby mode if no screen activity for x minutes.
Default value = OFF (disabled). Range is 1 to 240 minutes.
Note: When the balance enters standby mode, the current user will be automatically logged out. You need to log in the next time you use the balance.
5.5 User Settings
When using the Balance for the first time, enter this sub-menu to set User- defined values like Language, Display Settings, Volume, Functions Configuration and the Touchless Sensor setup.
5.5.1 User Settings Sub-menu

OHAUS 83021171 Explorer Semi Micro Balance - Fig12

Factory default settings are shown below in bold.
5.5.2 Change Password
Change the password for the user who is logged in at present.
Password must contain 8~10 characters with combination of letters and numbers.
5.5.3 Language
Set the language displayed for menus and displayed messages.

English Japanese
German Russian
French Polish
Spanish Portuguese
Italian Czech
Chinese Hungarian
Korean Turkish

OHAUS 83021171 Explorer Semi Micro Balance - Fig13

5.5.4 Volume
Controls the built-in Speaker loudness.
Default value= 5 (mid-level)
0= disabled (Off)
Range is 0 to 10
5.5.5 Display Settings
Use this sub-menu to change the setting for:
Screen Brightness: Default value= 100
Range is 20 to 100
Touch Calibrate
“Touch the screen at the center of the ring. O Press here.” (First top-left, then bottom-right.)
Auto Dim (Dims the display if no Screen Activity for x minutes)
Default value = OFF (disabled). Range is 1 to 240 minutes.

OHAUS 83021171 Explorer Semi Micro Balance - Fig14

5.5.6 Functions Configuration
Touch Functions Configuration to see a table of Selected and
Available functions. Press > to move a function from Available to Selected, and press < to move them back.
To move all functions to the right, Press >> and press << to move them back.

OHAUS 83021171 Explorer Semi Micro Balance - Fig15

5.5.7 Touch Less Sensor Setup
If desired, assign Print, Zero, Tare, Calibrate, Automatic draftshield door or
Draft Shield Light to the Touchless Sensors Terminal Right, Terminal Left, Base Right, and Base Left.

OHAUS 83021171 Explorer Semi Micro Balance - Fig16

Default setting for Automatic draftshield door models:
Terminal left: left door
Terminal right: right door
Note: EX12001, EX24001 and EX35001 do not have Base sensors.
5.5.8 Draft Shield light
Set the status of the lighted Draft Shield feature.
OFF = disabled.
ON = enabled.
5.5.9 Ionizer (seconds)
Set the working time of the ionizer, 3 or 5 seconds.

OHAUS 83021171 Explorer Semi Micro Balance - Fig17

Normally, 3 seconds is enough to remove the static electricity from a sample.
However, if the air is very dry, or if the sample volume is large, a longer working time might be used: 5 seconds.
Note: In order to get the best effect, it’s recommended to position the samples at the center of the glass shelf.
5.6 Balance Setup
Enter this sub-menu to customize Balance display functionality.
5.6.1 Balance Set-up sub-menu

OHAUS 83021171 Explorer Semi Micro Balance - Fig18 Factory default settings are shown below in bold.
5.6.2 Stable Range

Set the amount the reading can vary while the stability symbol remains on.

0.5 D = 0.5 graduations
1 D = 1 graduation
2 D = 2 graduations
3 D = 3 graduations
5 D = 5 graduations

Note: When LEGAL FOR TRADE (below) is set to ON, the setting is forced to 1 D.
The setting is locked when the Security Switch is set to the ON position.
5.6.3 Filter level
Set the amount of signal filtering.
LOW = faster stabilization time with less stability.
MEDIUM = normal stabilization time with normal stability.
HIGH = slower stabilization time with more stability.
The setting is locked at the current setting when the security switch is set to the ON position.
5.6.4 Auto Zero Tracking
Set the automatic zero tracking functionality.
OFF = disabled.
0.5 D = display maintains zero up to a drift of 0.5 graduation per second
1 D = display maintains zero up to a drift of 1 graduation per second.
3 D = display maintains zero up to a drift of 3 graduations per second.
Note: When Legal for Trade is set to ON, the AZT setting is forced to 0.5 D.
The OFF setting is still available. The setting is locked at the current setting when the security switch is set to the ON position.
5.6.5 Audible Signal
Set to turn on a beep sound when weighing is stable.
OFF = disabled.
Stable Weight Only = enabled the beep sound when weight is stable
5.6.6 Auto Tare
Set the automatic Tare.
OFF = disabled.
ON = enabled.
5.6.7 Gross Indicator
Set the symbol displayed for gross weights.
OFF = no symbol is displayed.
GROSS = the G symbol is displayed.
BRUTTO = the B symbol is displayed.
5.6.8 Legal For Trade
Use this menu to set the Legal for Trade status.
OFF = standard operation.
ON = operation complies with Weights and Measures regulations.
Note: When Legal for Trade is set to ON, the menu settings are affected as follows:
Calibration Menu:
Automatic Calibration is forced to ON and hidden. Auto Cal ™ internal Calibration, and Calibration Test functions are available. All other functions are hidden.
For EX…N… models:
– Automatic Calibration will be locked at its current setting.
– If you set Auto Cal™ internal Calibration to be ON before you turn on Approved
Mode, Auto Cal™ internal Calibration menu will still be available.
– If you set Autoclave™ internal Calibration to be OFF before you turn on Approved
Mode, Auto Cal™ internal Calibration menu will be locked.
Balance Setup Menu:
Filter Level is locked at the current setting.
Stability Indicator Range is locked at 1 Division.
Auto Zero Tracking is limited to 0.5 Division and OFF.
Auto Tare and Gross Indicator are locked at their current setting.
Graduations is forced to 1 Division and the menu item is hidden.
For EX…N…models, graduations will be locked at its current setting.
Weighing Units menu: all units are locked at their current settings.
Communication Menu:
Stable Weight Only is locked ON.
For EX…N…models, auto print mode selections are limited to OFF, On Stability, and Interval. Continuous is not available.
Numeric Value Only is locked OFF.
Note: The security switch must be in the locked position to set Legal for Trade to ON. The security switch must be in the unlocked position to set Legal for Trade to OFF. See Section 6.
5.6.9 Graduations
Set the displayed readability of the balance.
1 Division = standard readability.
10 Divisions = readability is increased by a factor of 10.
For example, if the standard readability is 0.01g, selecting 10 Divisions will result in a displayed reading of 0.1g.
5.6.10 Date & Time
Set the current Date and Time.
Change the format (if desired), then enter the current value.
Select √ to return to the sub-menu.

5.7 Applications
Enter this sub-menu to activate the desired Balance Applications. Only one application can be running at a time.
(The use of each Application is described in detail in Section 4.)
5.7.1 Turning an Application ON/OFF

Touch line of the desired application.
In the Item option screen, touch the Enabled line to toggle it on or off.
Once an Application is enabled (turned on) it may be chosen from the
Applications shortcut (top-left of all home screens).
OHAUS 83021171 Explorer Semi Micro Balance - Fig22

The current menu item status is shown: OFF = disabled, ON = enabled
5.8 Weighing Units
Enter this sub-menu to activate the desired units of measure.
Note: Due to national laws, the indicator may not include some of the units of measure listed.
5.8.1 Units Sub-menu

OHAUS 83021171 Explorer Semi Micro Balance - Fig23

Notes: If LEGAL FOR TRADE is set to ON, the Unit menu is locked.
The Custom Units names can be customized up to 3 characters.
5.8.2 Turning a Unit ON/OFF
Select the desired unit, then touch the item to toggle it either on or off.
The current menu item status is shown.
OFF = disabled
ON = enabled
OHAUS 83021171 Explorer Semi Micro Balance - Fig24

5.8.3 Custom Units
Use each Custom Unit to display weight in an alternative unit of measure. The custom unit is defined using a conversion factor, where the conversion factor is the number of custom units per kilogram expressed in scientific notation (Factor x 10^Exponent).
For example: To display weight in troy ounces (32.15075 troy ounces per kilogram) enter a Factor of 0.321508 and an Exponent of 2.
Set the status.
OFF = disabled
ON = enabled
When Custom Unit is set to ON, the Factor, Exponent and Least Significant Digit must be set.
Factor
Set the conversion factor using the numeric keypad.
Settings of 0.00001 to 1.9999999 are available. The default setting is 1.000000
Exponent
Set the factor multiplier.
0 = multiply the Factor by 1 (1×10°)
1 = multiply the Factor by 10 (1×10¹)
2 = multiply the Factor by 100 (1×10²)
3 = multiply the Factor by 1000 (1×10³)
–3 = divide the Factor by 1000 (1×10 -3)
–2 = divide the Factor by 100 (1×10 -2 )
–1 = divide the Factor by 10 (1×10 -1 )
Least Significant Digit
Set the graduation.
Settings of 0.00001, 0.00002, 0.00005, 0.0001, 0.0002, 0.0005, 0.001, 0.002, 0.005, 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50, 100, 200, 500 and 1000 are available.
Note: Least Significant Digit setting selections are dependent on the Factor and Exponent settings.
Selections are limited.
5.9 GLP and GMP Data
Enter this menu to set the Good laboratory Practices (GLP) data.
5.9.2 GLP Data Sub-menu
5.9.3 Header
Enables the printing of GLP headings. There are up to 5 headings available.
Alphanumeric settings up to 25 characters are available for each Header setting. .
5.9.4 Balance Name
Set the balance identification.
Alphanumeric settings up to 25 characters are available. The default setting is blank.
5.9.5 Project Name
Enter this menu to set the Project identification.
Alphanumeric settings up to 25 characters are available. The default setting is blank.
5.10 Communication
Enter this menu to define external communication methods and to set printing parameters.
Data may be output to either a printer or PC (See Section 5.9.8 for output samples.)
Factory default settings are shown in bold.
Notes: The COM3 menu is only available if the optional Interface is installed.
5.10.1 Communication Sub-menu

OHAUS 83021171 Explorer Semi Micro Balance - icon12

There are two types of USB interfaces on the Explorer:
Use this interface to transfer weighing data directly to a USB flash drive. See the following 5.10.13 Save to USB flash drive for detailed instructions.
Use this interface to connect the balance to a computer or printer.
Choosing an item brings up another menu level (RS232 shown):
OHAUS 83021171 Explorer Semi Micro Balance - icon15 Choosing a Settings item brings up yet another menu level; the device settings are dependent on the COM chosen (RS232 shown)
Device Settings Menu: (RS232 shown)
5.10.2 Baud Rate
Set the baud rate (bits per second).
1200  =1200 bps
2400= 2400 bps
4800 = 4800 bps
9600 = 9600 bps
19200 = 19200 bps
38400 = 38400 bps
5.10.3 Transmission
Set the data bits, stop bit, and parity.
7 EVEN 1 = 7 data bits, even parity
7 ODD 1 = 7 data bits, odd parity
7 NONE 1 = 7 data bits, no parity
8 NONE 1 = 8 data bits, no parity
7 EVEN 2 = 7 data bits, even parity
7 ODD 2 = 7 data bits, odd parity
7 NONE 2 = 7 data bits, no parity
8 NONE 2 = 8 data bits, no parity
5.10.4 Handshake
Set the flow control method.
NONE = no handshaking
XON-XOFF = XON/XOFF handshaking
HARDWARE = hardware handshaking (COM1 menu only)
5.10.5 Alternate Command
Enter this sub-menu to set a different command character for the P (Print), T (Tare) or Z (Zero) commands.
Note: The selected character can only be used for one command.
Alternate Print Command
Set the alternate command character for Print.
Settings of A to Z are available. The default setting is P, except T&Z.
Alternate Tare Command
Set the alternate command character for Tare.
Settings of A to Z are available. The default setting is T, except P&Z.
Alternate Zero Command
Set the alternate command character for Zero.
Settings of A to Z are available. The default setting is Z, except P&T.
Print Settings Menu:
Note: If LEGAL FOR TRADE is set to ON, the print Output-Stable setting is not reset.
5.10.6 Print Output
Print Stable Weight Only
Set the printing criteria.
OFF = values are printed immediately, regardless of stability.
ON  = values are printed only when the stability criteria are met.
Print Numeric Value Only
Set the printing criteria.
OFF = All data values are printed
ON = Only numeric data values are printed
Print Options
Set the way to print.
Printer = Print the result through a printer.
PC = Transfer the result to a computer.
5.10.7 Auto Print
Set the automatic printing functionality.
OFF = disabled
ON STABLE = printing occurs each time the stability criteria are met.
INTERVAL = printing occurs at the defined time interval.
CONTINUOUS = printing occurs continuously.
When ON STABLE is selected, set the conditions for printing.
LOAD = Prints when the displayed load is stable.
LOAD ZERO = Prints when the displayed load or zero reading is stable.
When INTERVAL is selected, set the time interval using the numeric keypad.
Settings of 1 to 3600 seconds are available.
5.10.8 Print Calibration Data
Set the automatic Calibration Data printing functionality.
OFF = disabled
ON = enabled
5.10.9 Print Content
Enter this sub-menu to define the content of the printed data.
See details below:
Header
Set the status.
OFF = the user defined header is not printed.
ON = the user defined header is printed.
Note: The header information must be defined using the H x “text” interface command.
(See Section 5.1.)
Date and Time
Set the status.
OFF = the time and date are not printed.
ON = the time and date are printed.
Application Name
Set the status.
OFF = the Application Name is not printed.
ON = the Application Name is printed.

Sample Name
Set the sample name.
OFF = the Sample Name is not printed.
ON = the Sample Name is printed.

Batch ID
Set the batch ID
OFF = the Batch ID is not printed.
ON = the Batch ID is printed.
Balance ID
Set the status.
OFF = the Balance ID value is not printed.
ON = the Balance ID value is printed.

Balance Name
Set the status.
OFF = the Balance Name is not printed.
ON = the Balance Name is printed.
User Name
Set the status.
OFF = the User Name is not printed.
ON = the User Name is printed.
Project Name
Set the status.
OFF = the Project Name is not printed.
ON = the Project Name is printed.

Sample ID
Set the sample ID
OFF = the Sample ID is not printed.
ON = the Sample ID is printed.

Result
Set the status.
OFF = the displayed reading is not printed.
ON = the displayed reading is printed.
NUMERIC ONLY = only the numeric portion of the displayed reading is printed.
Gross
Set the status.
OFF = the gross weight is not printed.
ON = the gross weight is printed.

Net
Set the status.
OFF = the net weight is not printed.
ON = the net weight is printed.
Tare
Set the status.
OFF = the tare weight is not printed.
ON = the tare weight is printed.
Information
Set the status.
OFF = the reference info is not printed.
ON = the reference info is printed.
Note: See below for more clarification
Signature Line
Set the status.
OFF = the Signature Line is not printed.
ON = the Signature Line is printed.

Note: signature line will show the user name which is logged in at present. If admin account is logged in, Admin will appear.
Verified Line
Set the status.
OFF = the Verified by Line is not printed.
ON = the Verified by Line is printed.

Note: “Verified by” is to be signed by an officer from local metrological regulation office to assert that the weighing and/or calibration results are in accordance with regulations.
Note: The Information is dependent on the application and the ACCUMULATE setting. Examples are shown below.

Mode Accumulate set OFF Accumulate set ON
Weighing None N: 10

Total: 10.000 kg
Avg: 1.000 kg
Std: 0.001 kg
Min: 0.999 kg
Max: 1.001 kg
Diff: 0.002 kg
Counting| APW: 0.100 kg| N: 10
Total: 1000 Pcs
Avg: 100 Pcs
Std: 1 Pcs
Min: 99 Pcs
Max: 101 Pcs
Diff: 2 Pcs
Check Weighing| Under: 0.995 kg
Over: 1.005 kg| Under: 0.995 kg
Over: 1.005 kg
N: 10
Total: 10.000 kg
Avg: 1.000 kg
Std: 0.001 kg
Min: 0.999 kg
Max: 1.001 kg
Diff: 0.002 kg
Dynamic| Level: 0| Level: 0
N: 10
Total: 10.000 kg
Avg: 1.000 kg
Std: 0.001 kg
Min: 0.999 kg
Max: 1.001 kg
Diff: 0.002 kg

5.10.10 Print Layout
This sub-menu is used to define the format of the data output to a printer or computer.
Format
Set the printing format.
Multiple Lines = a multiple line printout is generated. A CRLF is added after each data output.
Single Line = a single line printout is generated. A TAB delimiter is added between each data output.
Feed
Set the paper feed.
1 Line = move the paper up one line after printing.
4 Lines = move the paper up four lines after printing.
Form = a form feed is appended to the output.
5.10.11 Data Transfer Function
Output weighing results directly to a PC application. Setup is easy and no additional software is required.
Note: Data Transfer Function is not supported in Win7, Win8 and Win10. OHAUS provides SPDC software for Win7, Win8 and Win10 users.
OFF = do not print.
ON = print the specified settings.
Click the Start Menu in Windows XP system and click “Settings” –>open Control Panel.
Double click Accessibility Options in Control Panel.

Select the General tab in Accessibility Options.

Check Use Serial Keys, and click the Settings button.

Select the Serial Port, and set the Baud rate to 9600.
After selecting, click OK to close setting for serial keys.
Close the Control Panel.
Run Excel to open one blank sheet. Focus the cursor on one item.
At this time, if scale sends data to the PC through the RS232 port, the data will be put into the cell, and the cursor will automatically move to the next vertical cell.
Note: If the weighing value is a negative number, set the target cell in TEXT format. Otherwise, Excel will not distinguish it as a negative number.
Label Printing Menu:
OFF
= deactivate label printing function
ON = activate label printing function
When Label Printing menu is set to ON, Label Template menu appears with default selection, English label template “100mm x 75mm”.
This function is compatible with any label printer using the ZPLII programming language. We recommend Zebra GC420t label printer.

5.10.12 Label Template
There are 2 built-in templates:

  • 1 – 100mm x 75mm, which is designed based on label size “100mm x 75mm”. This template includes all basic information “Data/Time”, “Sample ID”, “Sample Name”,  “Batch ID”, “Result”, “Gross Weight”, “Tare Weight”, “Net Weight”, and “User Name”.

  • 2 – 100mm x 10mm, which is designed for small label size “100mm x 10mm” with weighing result only.

To select another label template, touch the other numbers. Templates #3 – #5 are empty templates in Explorer until they are edited by OHAUS label Designer and written to the balance. Overall 5 label templates can be stored in the balance.
All templates including the default one can be edited via OHAUS Label Designer. Please go to the website below to download the software. For how to use the software, please contact an authorized dealer to obtain the software’s instruction. https://dmx.ohaus.com/WorkArea/DownloadAsset.aspx?id=28916
5.10.13 Save to USB flash drive
OFF
= deactivate save to USB function
ON = activate save to USB function
When this menu is activated, insert the USB flash drive into the balance’s USB port:

  1. Locate the USB port on the back side of the Explorer terminal.
  2. Insert the USB flash drive into the USB port.

After the save to USB function is enabled and a USB flash drive is inserted, Save to USB button appears in main screen.
To save weighing data to USB:

  1. Touch Save to USB. Once saved, the button will shortly change color to orange.
  2. A folder named “EX” will automatically be created on the USB flash drive.
    The weighing data is in the folder named by the date of printing, e.g.
  3. Double click the txt file to open it in notepad.
    The files can easily be changed to ‘.csv’ files to be viewed in Excel.
    To change the file type into ‘.csv’, simply single click on the file name and change ‘.TXT’ into ‘.csv’.OHAUS 83021171 Explorer Semi Micro Balance -
Fig3

Notes:

  • The first time a USB flash drive is connected to the balance there might be some delay before the button Save to USB works. The reason is that the balance is creating the necessary folders on the USB flash drive where the data will be stored.
  •  Save to USB applies to Weighing, Parts Counting, Percent Weighing, Check Weighing and Filling.

To automatically save weighing data to USB:

  1. Enable “Auto Print” in the “Communication – USB – Print Settings – Auto Print” menu.
  2. The On Stable, Interval and Continuous selections are available.
    Auto Save to USB applies to Weighing, Parts Counting, Percent Weighing, Check Weighing and Filling. The weighing data is in the folder named by WGH_AUTO.

5.11 Library
When an item is processed on a regular basis, the item’s data may be stored in memory for future use. This memory is referred to as the Balance’s Library.
Delete All Record: Select to delete all library records.
5.12 Diagnostics
Use this menu to enable the Diagnostics (troubleshooting) functions, and to enter the Service menu;
5.12.1 Routine Test
To set settings for daily repeatability test and perform it.
You can set Number of Repeatability Tests, Test Weight ID, Test Weight Value and Test Weight Class in this menu.
To perform repeatability test, press Repeatability Test and then follow the onscreen instruction.

When the repeatability test finishes, it will pop out the results automatically. Press Print to print repeatability test report. Press Min- weight to show the reference minimum weight. After exiting the routine test, the results will be deleted automatically.

OHAUS 83021171 Explorer Semi Micro Balance - Fig8

5.12.2 Level Bubble Light (not available in EX12001, EX24001, EX35001)
To verify proper operation of the illuminated Level Bubble.
When selected, this light should blink.
5.12.3 Draft Shield Light (Analytical models)
To verify proper operation of the Draft Shield Light.
When selected, this light should blink.
5.12.4 Sensors
To verify proper operation of each Touchless Sensor.
When operated, each sensor should light and make a sound.
5.12.5 Automatic draftshield door
To verify proper operation of each Touchless Sensor.
When operated, each sensor should light and make a sound.
Note: Only for Automatic draftshield door models.
5.12.6 Data Maintenance
To check Software Upgrade, Balance Info and System Log.
All the changes made on balance setting will keep records in system log file. e.g. every printed data, date/time change, balance setting changes, perform calibration action, user
log in/log out, user account create/edit/delete and etc.
These system log files can be reviewed and exported as a PDF to a USB flash drive.
When exporting the system log files, the two touchless sensor’s LEDs will blink.
Maximum 5000 records can be stored in balance memories. When the memories are full, a message will pop up for exporting the log files to a USB flash diver. Then the pop-up message will show the log file has deleted.
In addition, calibration logs can be viewed separately in System Log. To do this:
Activate System Log selection in the Date
Maintenance menu.
The System Log Review menu appears. Touch it.

Touch View Calibration to view calibration logs.

5.12.7 Service Mode
Used to enter the Ohaus Service menu (Service personnel only). A password is required to access this menu;
5.13 Factory Reset
Use this sub-menu to reset the menus to their Factory default settings.
5.13.1 Factory Reset Sub-menu![OHAUS 83021171 Explorer Semi Micro Balance

5.13.2 Reset Calibration
Select YES to return all Calibration menu items to their factory settings.
5.13.3 Reset User Settings
Select YES to return all User Settings menu items to their factory settings.
5.13.4 Reset Balance Setup
Select YES to return all Balance Setup menu items to their factory settings.
5.13.5 Reset Application Modes
Select YES to return all Application Modes menu items to their factory settings.
5.13.6 Reset Weighing Units
Select YES to return all Weighing Units menu items to their factory settings.
5.13.7 Reset GLP and GMP Data
Select YES to return all GLP and GMP Data menu items to their factory settings.
5.13.8 Reset Communication
Select YES to return all Communication menu items to their factory settings.
5.13.9 Reset All (Global Reset)
Select YES to return all the menu items to their factory settings. The two touchless sensor’s LEDs will blink during resetting and go out when the reset is finished.
This selection will clear all the user and library records.
5.14 Securing the Menu and Key Lock menu settings
A slide switch is used to secure the Lock menu settings. When the switch is set to the ON position, the Lock menu settings may be viewed but not changed. This switch is located under the Base.
Set the position of the switch to ON by sliding the external Lock Switch to LOCKED as shown in Section 6.
When the switch is in the ON position, the startup display includes the LOCK ON message.
Note: This switch is also used in conjunction with the Legal for Trade menu item. When the Legal for Trade menu is set to ON, the switch must be set to the ON position to prevent calibration and changes to metrologically significant settings. Refer to Section 6 for more information. EX12001, EX24001 and EX35001 switch 5.15 Information
Information is available from any application as well as from the Main Menu. A brief descrtiption of the
item selected is displayed.
This feature is menu based.
5.15.1 Using Information

  • Touch the Info Icon to bring up the menu (UPPER RIGHT)
  • Information topics, beginning with Level Assist
  • Touch the word or the “+” in front of it, to open a topic.
  • Touch x to close the Information topic.

6. LEGAL FOR TRADE
When the balance is used in trade or a legally controlled application it must be set up, verified and sealed in accordance with local weights and measures regulations. It is the responsibility of the purchaser to ensure that all pertinent legal requirements are met.
*For EX…N…models, please refer to Legal for Trade Setting and Sealing flyer for details.
LEGAL FOR TRADE SETTING AND SEALING
6.1 Settings
Before verification and sealing, perform the following steps in Order:

  1. Verify that the menu settings meet the local weights and measures regulations.
  2. Units menu should be reviewed. Verify the units turned On meet the local weights and measures regulations.
  3. Perform a calibration as explained in Section 5.
  4. Set the position of the Security Switch to the locked position.
  5. Set Legal for Trade to ON in the Balance Setup menu.

6.2 Verification
A weights and measures official or authorized service agent must perform the verification procedure.
6.3 Sealing
After the Balance has been verified, it must be sealed to prevent undetected access to the legally controlled settings. Before sealing the device, ensure that the security switch is in the Locked position and the Legal for Trade setting in the Balance Setup menu has been set to ON.
If using a wire seal, pass the sealing wire through the holes in the security switch and Bottom Housing as shown. If using a paper seal, place the seal over the security switch and Bottom Housing as shown.OHAUS 83021171 Explorer Semi
Micro Balance - Fig16

EX12001, EX24001 and EX35001:
OHAUS 83021171 Explorer Semi Micro Balance - Fig17

PRINTING

7.1 Connecting, Configuring and Testing the Printer/Computer Interface Use the built-in RS-232 Port to connect either to a computer or a printer. If connecting to a computer, use HyperTerminal or similar software. (Find HyperTerminal under Accessories/Communications in Windows XP.) Connect to the computer with a standard (straight-through) serial cable. Choose New Connection, “connect using” COM1 (or available COM port). Select Baud=9600; Parity=8 None; Stop=1; Handshaking=None.  lick OK. Choose Properties/Settings, then ASCII Setup. Check boxes as illustrated: (Send line ends…; Echo typed characters…; Wrap lines…)
Use RS232 Interface Commands (Section 9.6.1) to control the balance from a PC. Note: When the HyperTerminal configuration is complete, it will automatically print the results of a Cal Test operation, and echo print commands sent to the scale. Printer Connection
Two printers compatible with Explorer Balances are available from Ohaus. (See Section 9.) Connect a straight-through serial cable to the scale’s RS232 port. See sample Print Output Formats below. Two printers compatible with explorer balances are available from Ohaus. Connect a straight through serial cable to the scale

7.2 Printout Examples
Examples for each Application are shown with all items turned ON in the Print Content menu. The default values
for Header lines 1-5 are also shown.
Note: If a Library name has been selected, “Library Name:” will appear below “Application” in the printout.
User name and Signature line will show the user name which is logged in at present. If admin account is logged in, Admin will appear.

OHAUS 83021171 Explorer Semi Micro Balance - Fig1 OHAUS 83021171 Explorer Semi Micro Balance - Fig2 OHAUS 83021171 Explorer Semi Micro Balance - Fig3

OHAUS 83021171 Explorer Semi Micro Balance - Fig4Note: In order to print Net weight, the Tare key needs to be pressed.
OHAUS 83021171 Explorer Semi Micro Balance - Fig5

OHAUS 83021171 Explorer Semi Micro Balance - Fig6

OHAUS 83021171 Explorer Semi Micro Balance - Fig7OHAUS 83021171 Explorer Semi Micro Balance -
Fig8OHAUS 83021171 Explorer Semi Micro Balance -
Fig9 Internal Calibration, Span Calibration and User Calibration:

OHAUS 83021171 Explorer Semi Micro Balance - Fig10

Note: User calibration is for InCal models only.
The CAL TEST printout is automatically printed when a Calibration Test is performed.

OHAUS 83021171 Explorer Semi Micro Balance - Fig11

7.3 Output Format
The Result Data, and G/N/T data, is output in the following format.

OHAUS 83021171 Explorer Semi Micro Balance - Fig12

  1.  In certain cases, a Label field of up to 11 characters is included.
  2. Each field is followed by a single delimiting space (ASCII 32).
  3. The Weight field is 9 right justified characters. If the value is negative, the “–“ character is located at the immediate left of the most significant digit.
  4. The Unit field contains the unit of measure abbreviation up to 5 characters.
  5. The Stability field contains the “?” character if the weight reading is not stable. The Stability field and the following Space field are omitted if the weight reading is stable.
  6. The G/N field contains the net or gross indication. For net weights, the field contains “NET”. For gross weights, the field contains nothing, “G” or “B”, depending on the GROSS INDICATOR menu setting.
  7. The Termination Characters field contains CRLF, Four CRLF or Form Feed (ASCII 12), depending on the LINE FEED menu setting.
  8. The PFMT command can be used to control the printout format of OHAUS products to be compatible with the printout format of other brands’ products.
    The command format is as the following:
    PFMT [n]
    n:
    0 default OHAUS mode
    1 MT mode
    2 Sartorius modes

MAINTENANCE

8.1 Calibration
Periodically verify calibration by placing an accurate weight on the balance and viewing the result. If calibration is required, perform a Balance internal calibration.
8.2 Cleaning
WARNING: Electric Shock Hazard. Disconnect the equipment from the power supply before cleaning. Make sure that no liquid enters the interior of the instrument.
Attention: Do not use solvents, harsh chemicals, ammonia or abrasive cleaning agents.
The housing may be cleaned with a cloth dampened with a mild detergent if necessary.
8.3 Battery Power
CAUTION: Battery is to be replaced only by an authorized Ohaus service dealer. Risk of explosion can occur if the rechargeable battery is replaced with the wrong type or if it is not properly connected. Dispose of the rechargeable battery according to local laws and regulations.
8.4 Troubleshooting
TABLE 8-1. TROUBLESHOOTING

Symptom / Display Possible Cause Remedy
Balance will not turn on No power to Balance Verify connection and voltage
Poor accuracy Improper calibration
Unstable environment Perform calibration

Move balance to suitable location
Cannot calibrate| Calibration Menu locked
LFT set to on
Unstable environment
Incorrect calibration masses| Turn Calibration menu lock off
Turn LFT off
Move balance to suitable location
Use correct calibration masses
Cannot change menu settings| Sub-menu locked
LFT set to on| Unlock sub-menu
Turn LFT off
Low Reference weight| Reference weight too small
The weight on the pan is too small to define a valid reference weight.| Increase sample size
Invalid Piece Weight| Average piece weight is too small| Increase average piece weight
Operation Timeout| Weight reading is not stable| Move balance to suitable location
| Busy (tare, zero, printing)| Wait until completion

8.5 Service Information
If the troubleshooting section does not resolve your problem, contact an Authorized Ohaus Service Agent. Please visit our website www.ohaus.com to locate the Ohaus office nearest you. An Ohaus Product Service Specialist will be available to assist you.

TECHNICAL DATA

9.1 Specifications
Ambient conditions

  • Indoor use only
  • Altitude: 2000 m
  • Operating temperature: 10 to 30°C
  • Humidity: Maximum relative humidity 80% for temperatures up to 31 °C decreasing linearly to 50% relative humidity at 30°C.
  • Electrical supply: 100 – 240V~, 0.5A, 50/60Hz for models of EX12001x, EX24001x and EX35001x; 12VDC, 1.5A. (For use with certified or approved power supply, which must have a SELV and limited energy output.)
  • Mains supply voltage fluctuations: up to ±10% of the nominal voltage
  • Installation category II
  • Pollution degree: 2

Materials

  • Base Housing; die-cast Aluminum, Painted
  • Base: IP54 waterproof protected (EX12001, EX24001 and EX35001)
  • Top Housing, Base, Terminal: plastic (ABS)
  • Weighing Platforms: 18/10 stainless steel
  • Weighing Pan: 304 SST (EX12001, EX24001 and EX35001)
  • In-use Cover, Terminal: plastic (PET)
  • Draft Shield; Glass, plastic (ABS)

Battery Specification

  • Battery Type: Lithium-Ion, 2600mAh, 14.8V
  • Battery Operating Time: 10 hours* with 50% brightness
  • Battery Charging Time: 9 hours
  • Operating Temperature Range: -20°C ~ 60°C
  • Charging Temperature Range: 0°C ~ 45°C

Note:

  • The number of hours the balance can run on the battery depends on many factors. Backlight, IR Sensors and Interface Options will all reduce the hours before a recharge is needed. For optimum battery life, turn these features off.
    WARNING:
    Stop charging the battery if charging is not completed within the specified time.
    Stop using the battery if the battery becomes abnormally hot, or if there is an odder, discoloration or deformation.
    Stop using the battery if abnormal conditions are detected during use, charging or storage.
    TABLE 9-1. SPECIFICATIONS

MODEL| EX125D
EX125DM| EX125
EX125M| EX225D
EX225DM| EX225D/AD
EX225DM/AD| EX225/AD
EX225M/AD
---|---|---|---|---|---
Capacity (g)| 52 g /120 g| 120 g| 120g/220g| 120g/220 g| 220 g
Readability d, fine range| 0.01 mg| 0.01 mg| 0.01 mg| 0.01 mg| 0.01 mg
Readability d, full range| 0.1 mg| 0.01 mg| 0.1 mg| 0.1 mg| 0.01 mg
Repeatability (sd.), s5% of kill load| 0.01 mg| 0.01 mg| 0.01 mg| 0.01 mg| 0.01 mg
Repeatability (sd.), 5% of full load to fine range max| 0.02 mg| 0.02 mg| 0.02 mg| 0.02 mg| 0.02 mg
Repeatability (sd. ), fine range max to full range| 0.1 mg| 0.02 mg| 0.1 mg| 0.1 mg| 0.02mg
Linearity deviation, typical| ± 0.06 mg
Linearity deviation| ± 0.1 mg
Verification interval, e
(EX…M.. model only)| 1 mg
Approval Class (EX…M.. model only)|
Span Calibration Points (g)| 50g, 100g| I 50g, 1009| I 100g, 150g, 200g| I 100g, 150g, 200g| I 100g, 150g, 2009
Weighing units| g, mg, ct, oz , ort, dwt, Grain, mom, msg, tical, tola, baht, Custom UM 1, Custom Unit 2, Custom Unit 3
Weighing units
(EX…M.. model)| q, mg, ct
Applications| Weighing, Parts Counting, Percent Weighing, Check Weighing, Dynamic Weighing, Filling, Totalization,
Formulation, Differential Weighing, Peak Hold, Density Determination, Pipette Adjustment, Statistical
Quality Control, Fill Weight Variation
Stabilization time (typical)| 0.1 mg: 52 seconds, 0.01 mg: 58 seconds
Sensitivity Temperature Drift| 0.80 ppm/K
Min-Weight (typical) (USP, K=2, U=0.10%)| 20 mg
Min-Weight (optimal)
(USP, K=2, U=0.10%, SRP50.41d)| 8.2 mg
Display| Full-Color VGA Graphic LCD
Display size| 5.7 in / 14.5 cm (diagonal)
Backlight| White LED
Controls| 4-wire resistive touch screen
Communication| RS-23Z 2 USB ports
Balance power input| 12 VDC, 0.5A
Power supply| AC Adapter Input 100-240 VAC 0.6A 50-60 Hz
AC Adapter Output 12 VDC 1.5A
Platform size (diameter)| 80 mm/3.1 in
Terminal Housing dimensions (WxDxH)| 195 x 155 x 68 mm
7.7 x 6.1 x 2.7 inch
Base Housing dimensions (WxDxH)| 230 x 260 x 350 mm
9.1 x 10.2 x 13.8 inch
Assembled dimensions (WxDxH)| 230 x 393 x 350 mm
9.1 x 15.5 x 13.8 inch
Shipping dimensions (W x D x H)| 632 x 467 x 640 mm
24.9 x 18.4 x 25.2 inch
Net weight| 7 kg / 15.4 lb| 7.5 kg / 16.5 lb
Shipping weight| 12 kg / 26.5 lb| 12.6 kg / 27.8 lb

Note 1: M = EC Type approved
Note 2: Default Calibration weights shown in Bold
Note 3: * Availability is dependent on region.

MODEL| EX124
EX124M
EX124M/AD| EX224
EX224M
EX224M/AD
EX224N
EX224N/AD| EX324
EX324M
EX324M/AD
EX324N
EX324NIAD
---|---|---|---
Capacity (g)| 120 g| 220 g| 320 g
Readabihty d (g)| 0.0001 g ( EX N..model: 0.0001 or 0.001 )
Repeatability (sd.), S5% of full load (g)| 0.00008 g
Repeatability (sd.), 5% of full bad to full range (g)| 0.0001 g
Linearity (g)| + 0.0002 g
Linearity deviation, typical (g)| ±0.00006 g
Verification interval, e (EX…M.. I EX…N..model only)| 0.001 g
Approval Class
(EX…M.. / EX…N..model only)| I
Span Calibration Points (g)| 50g,1009| I 100g, 150g, 200g| I 150g, 200g, 300g
Weighing units*| g, mg, kg, ct, oz , ort, dwt, Grain, lb, mom, msg, tical, tola, baht, Hong Kong Tael,
Singapore Tael, Taiwan Tael, N, Custom Unit 1, Custom Unit 2, Custom Unit 3
Weighing units (EX…M.. model)| g, mg, ct
Applications| Weighing, Parts Counting, Percent Weighing, Check Weighing, Dynamic Weighing, Filling,
Totalization, Formulation, Differential Weighing, Peak Hold Density Determination, Pipette
Adjustment, Ingredient Costing, Statistical Quality Control, Fill Weight Variation
Stabilization time (typical)| Within 2 seconds| Within 3 seconds
Sensitivity Temperature Drift (PPM/t)| 2.| I 1.5| 2.
Min-Weight (typical) (g) (USP, K=2, U=0.10%)| 160 mg
Min-Weight (optimal) (g)
(USP, K=2, U=0.10%, SRP,C0.41d)| 82 mg
Display| Full-Color VGA Graphic LCD
Display size| 5.7 in / 14.5 cm (diagonal)
Backlight| White LED
Controls| 4-wire resistive touch screen
Communication| RS-232, USB
Balance power input| 12 VDC, 0.5A
Power supply| AC Adapter Input: 100-240 VAC 0.6A 50-60 Hz
AC Adapter Output: 12 VDC 1.5A
Platform size (diameter)| 90 mm / 3.5 in
Terminal Housing dimensions (WxDxH)| 195 x 155 x 68 mm
7.7 x 6.1 x 2.7 inch
Base Housing dimensions (WxDxH)| 230 x 260 x 350 mm
9.1 x 10.2 x 13.8 inch
Assembled dimensions (WxDxH)| 230 x 393 x 350 mm
9.1 x 15.5 x 13.8 inch
Net weight| 6.9 kg 115.2 lb
Shipping weight| 9.7 kg 121.3 lb
Shipping dimension| 550 x 385 x 551 mm5
21.7 x 15.2 x 21.7 inch

Note 1: M = EC Type approved
N = NTEP certified and Measurement Canada approved
Note 2: Default Calibration weights shown in Bold
Note 3: * Availability is dependent on region.
TABLE 9-3. SPECIFICATIONS

MODEL| EX223 EX223/E EX223N EX223NIE| X423E
EX423/E EX423M EX423N EX423N/E| EX623| EX1103
EX1103M
EX1103N
---|---|---|---|---
Capacity (g)| 220 g| 420 g| 620 g| 1100 g
Readability d (g)| 0.001 g ( EX…N..model: 0.001 or 0.01)
Repeatability (sd.), 55% of full load (g)| 0.0008 g
Repeatability (sd.), 5% of full load to full range (q)| 0.001 q
Linearity deviation, typical (q)| ±0.0006 q
Linearity deviation (g)| + 0.002 g
Verification interval, e (EX…M.. / EX…N..model only)| 0.019
Approval Class
(EX…M.. / EX…N..model only)| II| I
Span Calibration Points| 100g, 150g, 2009| 200g, 300g, 4009| 3009, 6009 500g 400g, ,| 500g,10009
Weighing units*| g, mg, kg, ct, oz, ort, dwt, Grain, lb, mom, msg, tidal, tola, baht, Hong Kong Teal,
Singapore Tae , Taiwan Teal, N, Custom Unit 1, Custom Unit 2, Custom UM 3
Weighing units (EX…M.. model)| g, mg, ct
Applications| Weighing, Parts Counting, Percent Weighing, Check Weighing, Dynamic Weighing, Filling,
Totalization, Formulation, Differential Weighing, Peak Hold, Density Determination, Pipette
Adjustment, Ingredient Costing, Statistical Quality Control, Fill Weight Variation
Stabilization time (typical)| Within 1.5 seconds
Sensitivity Temperature Drift (PPWC)| 3| I3| I3| I3
Min-Weight (typical) (g) (USP, K=2, U=0.10%)| 2.
Min-Weight (optimal) (g) (USP, K=2, U=0.10%, SRP50.41d)| 0.82 g
Display| Full-Color VGA Graphic LCD
Display size| 5.7 in / 14.5 cm (diagonal)
Backlight| White LED
Controls| 4-wire resistive touch screen
Communication| RS-232, USB
Balance power input| 12 VDC, 0.5A
Power supply| AC Adapter Input : 100-240 VAC 0.6A 50-60 Hz
AC Adapter Output : 12 VDC 1.5A
Platform size (diameter)| 130 mm / 5.1 in
Terminal Housing dimensions (WxDxH)| 195 x 155 x 68 mm
7.7 x 6.1 x2.7 inch
Base Housing dimensions (WxDxH)| 230 x 260 x 350 mm
9.1 x 10.2 x 13.8 inch
Assembled dimensions (WxDxH)| 230 x 393 x 350 mm
9.1 x 15.5 x 13.8 inch
Net weight| 6.9 kg 115.2 lb
Shipping weight| 9.7 kg 121.3 lb
Shipping dimension| 550 x 385 x 551 mm5
21.7 x 15.2 x 21.7 inch

Note 1: M = EC Type approved
N = NTEP certified and Measurement Canada approved
Note 2: Default Calibration weights shown in Bold
Note 3: * Availability is dependent on region.
TABLE 9-4. SPECIFICATIONS (Cont.)

MODEL| EX2202
EX22021E
EX2202N/E| EX4202 EX42021E EX4202M EX4202N EX4202N/E| EX6202 EX62021E EX6202N EX6202NIE| EX1 0202
EX10202M
EX10202N
---|---|---|---|---
Capacity (g)| 2200 g| 4200 g| 6200 g| 10200 g
Readability d (g)| 0.01 g c EX…N..model: 0.01 or 0.1 )
Repeatability (sd.), S.5% of full load (g)| 0.008 g
Repeatability (sd.), 5% of full load to full range (q)| 0.01 g
Linearity deviation, typical (q)| ±0.006 g
Linearity deviation (g)| ±0.02 g
Verification interval, e (EX…M.. / EX…N..model only)| 0.1
Approval Class
(EX…M.. / EX…N..model only)| II| 1
Span Calibration Points| 1000g, 2000g| 2000g, 3000g,
4000g| 3000g, 4000g,
5000g, 6000g| 2500g, 5000g,
7500g, 10,000g
Weighing units*| g, kg, ct, oz, ort, dwt, Grain, lb, mom, msg, tical, tola, baht, Hong Kong Tael, Singapore Tael,
Taiwan Tael, N, Custom Unit 1, Custom Unit 2, Custom Unit 3
Weighing units (EX…M.. model)| g, kg, ct
Applications| Weighing, Parts Counting, Percent Weighing, Check Weighing, Dynamic Weighing, Filling,
Totalization, Formulation, Differential Weighing, Peak Hold, Density Determination, Pipette
Adjustment, Ingredient Costing, Statistical Quality Control, Fill Weight Variation
Stabilization time (typical)| Within 1 second
Sensitivity Temperature Drift (PPM/IC)| 3| I 3| I 3| I 3
Min-Weight (typical) (g) (USP, K=2, U=0.10%)| 16 g
Min-Weight (optimal) (g)
(USP, K=2, U=0.10%, SRP50.41d)| 8.2 g
Display| Full-Color VGA Graphic LCD
Display size| 5.7 in / 14.5 cm (diagonal)
Backlight| White LED
Controls| 4-wire resistive touch screen
Communication| RS-232, USB
Balance power input| 12 VDC, 0.5A
Power supply| AC Adapter Input: 100-240 VAC 0.6A 50-60 Hz
AC Adapter Output 12 VDC 1.5A
Platform size| 190 x 200 mm
7.5 x 8.0 in
Terminal Housing dimensions (WxDxH)| 195 x 155 x 68 mm
7.7 x 6.1 x 2.7 inch
Base Housing dimensions (WxDxH)| 230 x 260 x 98 mm
9.1 x 10.2 x 3.9 inch
Assembled dimensions (WxDxH)| 230 x 393 x 98 mm
9.1 x 15.5 x 3.9 inch
Net weight| 4.3 kg / 9.5 lb| 5.0 kg/11.0 lb
Shipping weight| 6.8 kg / 15.0 lb| 7.5 kg / 16.5 lb
Shipping dimensions| 550 x 385 x 291 mm
21.7 x 15.2 x 11.5 inch

Note 1: M = EC Type approved
N = NTEP certified and Measurement Canada approved
Note 2: Default Calibration weights shown in Bold
Note 3: * Availability is dependent on region.
TABLE 9-5. SPECIFICATIONS (Cont.)

MODEL| EX6201
EX6201M
EX6201N| EX1 0201
EX10201M
EX10201N
---|---|---
Capacity (g)| 6200 g| 10200 g
Readability d (g)| 01 g
Repeatability (sd.), s5% of kill load (g)| 0.08 g
Repeatability (sd.), 5% of full load to full range (q)| 0.1 g
Linearity deviation, typical (q)| ±0.06 q
Linearity deviation (q)| ±0.2 q
Verification interval, e (EX…M.. I EX…N..model only)| 0.1 g
Approval Class (EX…M.. I EX…N..model only)|
Span Calibration Points| 3000g, 4000g, 5000g, 6000g| 2500g, 5000g, 7500g,10,0009
Weighing units*| g, kg, ct, oz, ort, dwt, Grain, lb, mom, msg, tical, tola, baht, Hong Kong Tael, Singapore Tael,
Taiwan Tael, N, Custom Unit , Custom Unit 2, Custom Unit 3
Weighing units (EX…M.. model)| g, kg, ct
Applications| Weighing, Parts Counting, Percent Weighing, Check Weighing, Dynamic Weighing, Filling,
Totalization, Formulation, Differential Weighing, Peak Hold, Density Determination, Pipette
Adjustment, Ingredient Costing, Statistical Quality Control, Fill Weight Variation
Stabilization time (typical)| Within 1 second
Sensitivity Temperature Drift (PPM/C)| 5| I 3
Min-Weight (typical) (g) (USP, K=2, U=0.10%)| 160 g
Min-Weight (optimal) (g)
(USP, K=2, U=0.10%, SRPs0.41d)| 82 g
Display| Full-Color VGA Graphic LCD
Display size| 5.7 in 114.5 cm (diagonal)
Backlight| White LED
Controls| 4-wire resistive touch screen
Communication| RS-232, USB
Balance power input| 12 VDC, 0.5A
Power supply| AC Adapter Input: 100-240 VAC 0.6A 50-60 Hz
AC Adapter Output: 12 VDC 1.5A
Platform size| 190 x 200 mm
7.5 x 8.0 in
Terminal Housing dimensions (WxDxH)| 195 x 155 x 68 mm
7.7 x 6.1 x 2.7 inch
Base Housing dimensions (WxDxH)| 230 x 260 x 98 mm
9.1 x 10.2 x 3.9 inch
Assembled dimensions (WxDxH)| 230 x 393 x 98 mm
9.1 x 15.5 x 3.9 inch
Net weight| 5.0 kg / 11.0 lb
Shipping weight| 7.5 kg /16.5 lb
Shipping dimensions| 550 x 385 x 291 mm
21.7 x 15.2 x 11.5 inch

Note 1: M = EC Type approved
N = NTEP certified and Measurement Canada approved
Note 2: Default Calibration weights shown in Bold
Note 3: * Availability is dependent on region.
TABLE 9-6. SPECIFICATIONS (Cont.)

MODEL| EX12001
EX12001M
EX12001N| EX24001
EX24001M
EX24001N| EX35001
EX35001M
EX35001N
---|---|---|---
Capacity (g)| 12000 g| 24000 g| 35000 g
Readability d (g)| 0.1 g (EX…N..model: O. lg or 1g
Repeatability (sd.), .S5% of full load (g)| 0.08 g
Repeatability (sd.), 5% of kill load to full range (q)| 0.1 q
Linearity deviation, typical (q)| 10.06 q
Linearity deviation (g)| t0.2 g
Verification interval, e (EX…M.. / EX…N..model only)| 1 g
Approval Class
(EX…M.. / EX…N..model only)| II
Span Calibration Points| 3000g, 5000g, 10000g,
120009| 10000g, 15000g, 20000g,
24000g| 10000g, 20000g, 30000g,
35000g
Weighing unite| g, kg, ct, oz, ort, dwt, Grain, lb, mom, msg, tical, tola, baht, Hong Kong Tael, Singapore Tael,
Taiwan Tael, N, Custom Unit 1, Custom Unit 2, Custom Unit 3
Weighing units (EX…M.. model)| g, kg, ct
Applications| Weighing, Parts Counting, Percent Weighing, Check Weighing, Dynamic Weighing, Filling,
Totalizatbn, Formulation, Differential Weighing, Peak Hold, Density Determination,
Ingredient Costing, Statistical Quality Control, Fill Weight Variation
Stabilization time (typical)| Within 1 second
Sensitivity Temperature Drift (PPM/C)| 5
Min-Weight (typical) (g) (USP, K=2, U=0.10%)| 160 g
Min-Weight (optimal) (g)
(USP, K=2, U=0.10%, SRPs0.41d)”| 82 g
Display| Full-Color VGA Graphic LCD
Display size| 5.7 in / 14.5 cm (diagonal)
Backlight| White LED
Controls| 2-wire resistive touch screen
Communication| RS-232, USB
Balance power input| 12 VDC, 0.5A
Power supply| AC Adapter Input 100-240 V— 0.5A 50/60 Ft
AC Adapter Output: 24 VDC 0.625A
Platform size| 377 x 311 mm
14.8 x 12.2 in
Terminal Housing dimensions (WxDxH)| 195 x 155 x 68 mm
7.7 x 6.1 x 2.7 inch
Base Housing dimensions (W x D x H)| 377 x 311 x 120 mm
14.8 x 12.2 x 4.7 inch
Assembled dimensions (W x D x H)| 442 x 311 x 120 mm
17.4 x 12.2 x 4.7 inch
Net weight| 10 kg/ 22 lb
Shipping weight| 12 kg / 26.5 lb
Shipping dimension| 665 x 525 x 330 mm
26.2 x 20.7 x 13.0 inch

Note 1: M = EC Type approved
N = NTEP certified and Measurement Canada approved
Note 2: Default Calibration weights shown in Bold
Note 3: * Availability is dependent on region.
9.2 Drawings and Dimensions
Fully assembled dimensions
OHAUS 83021171 Explorer Semi Micro Balance -
Dimensions

9.3 Parts and Accessories
TABLE 9-7. PARTS

DESCRIPTION PART NUMBER
Terminal in-use Cover 83033633
Universal AC Adapter (without power Cord) 46001884
Power Cord  US 83033672
Power Cord  EU 83033673
Power Cord  UK 83033674
Power Cord  AU 83033675
Power Cord  JP 83033676
90 mm Pan Assembly 83033640
130 mm Pan 83033641
190 x 200 mm Pan 83033643
377 x 311 mm Pan (EX12001/24001/35001) 30076186

TABLE 9-8. ACCESSORIES

DESCRIPTION PART NUMBER
Terminal Extension Cable kit 83021083
Density Determination kit 80253384
Tower Mount for Terminal 83021102
Draft Shield kit 83021084
Security Device 80850043
Grid Pan 30369673
Printer SF-40A  30064202(EU);

30064203(AM)
30045641 (AP)
Interface Cable, Printer| Contact Ohaus
Interface Cable, PC 25 pin| Contact Ohaus
Interface Cable, PC 9 pin| Contact Ohaus
Interface Cable, USB (Type A to B)| 83021085
Ethernet Option Kit| 83021082
External Rechargeable Battery Pack| 30041295
Adjustable Rolling Feet| 30041470
Auxiliary Display(AD7-RS)| 30472064
Static Ionizer, ION-100A|  30130302(AM);
30130303(AP);
30095929(EU)

9.4 Communication
9.4.1 Interface Commands
Commands listed in the following table will be acknowledged by the balance. The balance will return “ES” for invalid commands.
TABLE 9-10. EXPLORER INTERFACE COMMAND LIST

Command Characters’) Function
IP Immediate Print of displayed weight (stable or unstable). Attention: for

EX…N…models, when LFT ON, IP could only print stable displayed weight.
Ph| Print displayed weight according to “Stable only” setting in the communication menu. Attention: when LFT ON, P could only print stable displayed weight.
CP| Continuous Print. For EX…N…models , when LFT ON, CP could not work.
SP’)| Print on Stability.
SLP| Auto Print stable non-zero displayed weight. Attention: the corresponding settings in the communication menu are modified too.
SLZP| Auto Print stable non-zero weight and stable zero reading. Attention: the corresponding settings in the communication menu are modified too.
xP| Interval Print x = Print Interval (1-3600 sec) OP ends interval Print. Attenticn: the corresponding settings in the communication menu are modified too.
OP| See above.
H| Enter or get Print Header Lines. Attention: header text length is up to 25 characters, the setting format is “H x “header line text””; and “H x” could get the corresponding header line in the balance.
Z| Same as pressing Zero Key
T| Same as pressing Tare Key.
xT| Establish a preset Tare value in displayed unit. X = preset tare value. Sending OT clears tare (if allowed).
PT| Prints Tare weight stored in memory.
PM| Print current application mode (weighing mode).
xM| Set current application mode to x. x depends on application, use application list2).
M| Scroll to the next enabled mode.
PU| Print Current weighing unit: g, Kg, lb, oz, etc….
xU| Set balance to unit x: g, Kg etc. x depends on unit list3).
U| Scroll to the next enabled unit.
ON| Brings out of Standby
OFF| Goes to Standby.
C| Begin Span Calibration, same as trigger from calibration menu. Attention: when LFT ON, the operation is not allowed.
IC| Begin internal Calibration, same as trigger from calibration menu.
UC| User Calibration (uses default weight), same as trigger from calibration menu. Attention: when LFT ON, the operation is not allowed.
AC| Abort Calibration. Attention: when LFT ON, the operation is not allowed.
xUC| Set user defined weight and trigger one user calibration. Attention: the user defined weight is only used for this command.
WI 1 0| Left door open, right door closed.
WI 0 1| Left door closed, right door open.
WI 1 1| Both doors open.
WI 0 0| Both doors closed.

TABLE 9-10. EXPLORER INTERFACE COMMAND LIST (continued)

Command Characters’) Function
PSN Print Serial Number.
PV Print terminal software version, base software version and LFT ON (if LFT

is set ON).
xff| Set Counting APW (x) in grams. (must have APW stored)
Pit| Print Counting application APW.
x%| Set Percent application reference weight (x) in grams. (must have reference weight stored)
P%| Print Percent application reference weight.
xAW| Set Dynamic Weigh Level to x. (x = 1 – 99 seconds)
xAW| Set Dynamic Weigh Mode. x = A (Automatic), S (Semi-Automatic), M (Manual)
PAW| Print Dynamic Weigh Level.
BAW| Start Dynamic Weigh cycle. (Manual Mode)
CW| Clear locked weight (weight < threshold) in Dynamic Weigh (same as button “Reset”) & Display Hold (same as button “End Peak Hold”).
xCO| Set Checkweighing Over Limit in grams x.
xCU| Set Checkweighing Under Limit in grams x.
PCO| Print Checkweighing Over Limit.
PCU| Print Checkweighing Under Limit.
xCM| Set Checkweigh mode (1=over/under, 2=Target/weight tolerance, 3=target/% tolerance
xCT%| Set Checkweighing target in grams x for percent tolerance mode.
PCT%| Print Checkweighing Target for percent tolerance mode.
xCTW| Set Checkweighing target in grams x for weight tolerance mode.
PCTW| Print Checkweighing Target for weight tolerance mode.
xC%| Set Checkweighing % tolerance x. Attention: when x is a positive value, it is used to set the +tolerance value; vice versa.
PC%| Print Checkweighing `)/0 tolerance.
xCW| Set Checkweighing weight tolerance x. Attention: when x is a positive value, it is used to set the +tolerance value; vice versa.
PCW| Print Checkweighing weight tolerance.
xDH| Set Display Hold mode (Peak Hold). x = A (Automatic), S (Semi-Automatic), M (Manual)
xD| Set 1 second print delay (set x = 0 for OFF, or x = 1 for ON).
xFL| Set filter level to x (1 = low, 2 = med, 3 = high).
xAL| Set Auto-zero to x (x = 1 for Od, x = 2 for 0.5d, x = 3 for 1d, x = 4 for 3d).
Esc R| Resets all Balance menus to factory defaults. Attention: The binary code of this commands is “1 B 20 52 OD OA” or “1 B 52 OD OA”.
PID| Print current user Name.
xID| Program user Name. Attention: only allowed numeric input.
xTL| Set Totalize Mode. x = A (Automatic), M (Manual).
PTIME| Print current time.
PDATE| Print current date.
xTIME| Set Time, x format: hh:mm:ss.
xDATE| Set Date, x format: mm/dd/yyyy.
CA| Continuous weight, same as CP.
SA| Stable load, same as SLP.
xA| Interval Print x = interval in sec (1-3600) 0 = off, same as xP.
OA| Set AutoPrint off, same as OP.
SC| Begin Span Cal, same as C.
xAM| Set Animal Mode to Auto, Semi-Auto, Manual. Same as xAW(A/S/M)
?| Prints current mode, same as PM.
xS| 0 = print unstable data, same as IP; 1 = print stable only1), same as SP.
xRL| 0 = disable response; 1 = enable response. This command only controls the “OK!” response.

General Notes:
Commands sent to the indicator must be terminated by a carriage return-line feed (CRLF).
Alternate command characters may be defined by the user.
Data output is always terminated with a carriage return-line feed (CRLF).
Note 1) There is 40-second timeout control for print under stable requirement. If the unstable condition continues over 40 seconds, balance will respond “ES”.
Note 2) Application list:

Index Application Index Application
0 Weighing 11 Ingredient Adjustment
1 Parts Counting 12 Pipette Adjustment
2 Percent Weighing 13 SQC
3 Check Weighing 14 Fill Weight Variation
4 Dynamic Weighing
5 Filling
6 Totalization
7 Formulation
8 Differential
9 Density Determination
10 Peak Hold

Note 3) Unit list:

Index Unit Index Unit
0 Milligram 11 Mesghal
1 Gram 12 Hong Kong Tael
2 Kilogram 13 Singapore Tael
3 Carat 14 Taiwan Tael
4 Ounce 15 Tical
5 Ounce Troy 16 Tola
6 Pound 17 Baht
7 Pennyweight 18 Custom Unit 1
8 Grain 19 Custom Unit 2
9 Newton 20 Custom Unit 3
10 Momme

9.4.2 RS232 (DB9) Pin Connections
Pin 2: Balance transmit line (TxD)
Pin 3: Balance receive line (RxD)
Pin 5: Ground signal (GND)
Pin 7: Clear to send (hardware handshake) (CTS)
Pin 8: Request to send (hardware handshake) (RTS)

9.5 The USB Interface
The Ohaus USB Interface is a unique solution to the problem of connecting a balance to a computer using a Universal Serial Bus (USB). USB devices are categorized into classes such as disk drives, digital cameras, printers, etc. Balances do not have a commonly used class so the Ohaus USB interface uses a generic interface based on the RS232 serial standard.
Data sent from the balance to a computer is in USB format. The USB data is directed to a virtual port. This port then appears as an RS232 port to the application program.
When sending a command from a computer to the balance, the application program sends a command to the virtual port as if it were an RS232 port. The computer then directs the command from the virtual port to the computers USB connector where the balance is connected. The port receives the USB signal and reacts to the command.
System Requirements

  • PC running Windows 98® , Windows 98SE® , Windows ME® , Windows 2000® , Windows XP® or Windows 7® , Windows 10® .
  • Available USB port (Type A, 4-pin, female)

USB Connection
The balance’s USB port terminates with a mini USB.
A USB Cable (Mini B/male to type A/male) is required (not supplied).

  1. Ensure that the balance is powered on and working properly.
  2. Power on the computer and verify that its USB port is enabled and working properly.
  3. Plug the cable’s USB connectors into the computer’s USB port and the balance’s USB port. Windows® should detect a USB device and the New Hardware Wizard will be initialized.

Virtual Port Software Installation

  1. Download the USB driver from the following address on the OHAUS website and then double click to install.
    https://us.ohaus.com/en-US/Support/Softwareand-Drivers

  2. After the installation is finished, click Finish, the virtual port should be ready to be used. Windows® typically adds the virtual port in sequence after the highest number COM port. For example, on PC’s equipped with up to 4 COM ports, the virtual port will be COM5.
    When using the USB interface with programs that limit the number of COM port designations (e.g. Ohaus MassTracker allows only COM1, 2, 3, & 4), it may be necessary to assign one of these port numbers to the new virtual port.

OHAUS 83021171 Explorer Semi Micro Balance - Fig14

This can be done in the Port Settings of the Device Manager utility, found in the Windows Control Panel.
Balance Setup for USB Interface
A) The Ohaus Interface is preset to communicate using the following settings: 9600 baud, 8 bit, no parity, no handshake, 1 stop bit. This is a fixed setting and cannot be changed.
B) Configure the balance to the desired USB and printing parameters.
SETUP DEFINITIONS

PRINT / Stable – On
PRINT / Auto Print – Continuous
PRINT / Auto Print – On stability
PRINT / Auto Print – Interval
PRINT / Auto Print – Off
USB / Baud, Parity, Handshake| Balance transmits stable data only.
Balance repeatedly sends data as fast as possible. Balance automatically outputs data upon stability.
Balance sends data every (xx) seconds.
Only sends data when PRINT is pressed.
Set parameters to match printer or computer.
---|---

Operation
USB OUTPUT
With the Adapter installed the balance will operate in one of the following three ways:

  • USB = On, Auto Print = Off, Stable = On or Off
    Press PRINT to send the display data based on the settings in the Menu.
    If Stable is set to On, the balance will wait for a stable reading before sending data.
    If balance has connect USB flash diver, the data will automatic go to USB.

  • USB = On, Auto Print = On, Stable = On or Off
    The balance will automatically send data.
    If Stable is set to On, only stable values will be sent.

  • USB = Off
    The interface is turned off.

Data sent from the Interface is in standard ASCII format terminated with a [CRLF] (carriage return-line feed).
The output format is as follows:

[weight] [space] [unit] [space] [stability indicator] [space] [legend] [CR] [LF]| 10 characters (right justified)
1 character
5 characters max (left justified)
1 character
1 character “?” when unstable or below min-weight, blank when stable
1 character
10 character TOTAL, hh:mm:ss (time interval), etc.
1 character
1 character
---|---

Output examples: (Note: * and _ indicate spaces)

  • Manual, Continuous or On-Stable printing unstable reading
  • Interval Printing (example:15sec interval) unstable reading
  • Accumulate Mode (Manual printing only)

USB INPUT
The balance will respond to various commands sent via the interface adapter.
Terminate the following commands when with a [CR] or [CRLF].
Explorer Commands

?
0A
SA
CA
(n)A
C
T
V
(Esc)R
P
0S
1S
P
SP
IP
CP
SLP
SLZP
xP
0P
PM
M
PU
U
T
Z
PV| print current mode, object below min-weight
turn Auto-print off
Auto-print on, prints on stability
Continuous Auto-print
Auto-print on 1 to 3600 second intervals (n = 1 to 3600)
perform span calibration
tare balance, same as pressing ON-ZERO
print software version
reset balance to factory defaults
same as pressing PRINT
print unstable data
print stable data only
same as pressing Print
print stable weight only
immediate print of displayed weight (stable or unstable)
Continuous print of weights
Auto-print stable non-zero weight only
Auto-print stable non-zero weight and zero reading
Auto-print on 1 to 3600 second intervals (x = 1 to 3600)
turns auto-print off
print current mode
advance to the next enabled mode
print current unit
advance to the next enabled unit
same as pressing Tare
same as pressing Zero
print software version
---|---

Auto-Print Operation
Once Auto-Print is activated in the menu, the balance will send data as required.
To temporarily stop Auto-Printing, press PRINT.
If there is data in the print buffer the printer will finish printing this data.
A second press will resume Auto-Printing.

SOFTWARE UPDATES

Ohaus is continuously improving its balance software. To obtain the latest release, please contact your Authorized Ohaus Dealer or Ohaus Corporation.

COMPLIANCE

Compliance to the following standards is indicated by the corresponding mark on the product.

Mark Standard
**** This product complies with the applicable harmonized standards of EU

Directives 2011/65/EU (RoHS), 2014/30/EU (EMC), 2014/35/EU (LVD) and 2014/31/EU (NAWI). The EU Declaration of Conformity is available online atwww.ohaus.com/ce.
| This product complies with the EU Directive 2012/19/EU (WEEE). Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment.
For disposal instructions in Europe, refer to www.ohaus.com/weee.
****| This product complies with the applicable statutory standards of the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment Regulations 2012, UK Electromagnetic Compatibility Regulations 2016, Electrical Equipment (Safety) Regulations 2016
and Non-Automatic Weighing Instruments Regulations 2016. The UK Declaration of Conformity is available online at www.ohaus.com/uk-declarations.
| EN 61326-1
| CAN/CSA-C22.2 No. 61010-1
UL Std. No. 61010-1

Important notice for EX… M verified weighing instruments in the EU
When the instrument is used in trade or a legally controlled application it must be set up, verified and sealed in accordance with local weights and measures regulations. It is the responsibility of the purchaser to ensure that all pertinent legal requirements are met.
Weighing Instruments verified at the place of manufacture bear the following supplementary metrology marking on the descriptive plate.

Weighing Instruments to be verified in two stages have no supplementary metrology marking on the descriptive plate. The second stage of conformity assessment must be carried out by the applicable weights and measures authorities.
If national regulations limit the validity period of the verification, the user of the weighing instrument must strictly observe the re-verification period and inform the weights and measures authorities
As verification requirements vary by jurisdiction, the purchaser should contact their local weights and measures office if they are not familiar with the requirements.
Industry Canada Note
CAN ICES-003(A) / NMB-003(A)
ISO 9001 Registration
The management system governing the production of this product is ISO 9001 certified.
FCC Supplier Declaration of Conformity
Unintentional Radiator per 47CFR Part B
Trade Name: OHAUS CORPORATION
Model: EX
Party issuing Supplier’s Declaration of Conformity:
Ohaus Instruments (Changzhou) Co., Ltd.
2F, 22 Block, 538 West Hehai Road, Xinbei District, Changzhou
Jiangsu 213022,
China
Phone: +86 519 85287270
Responsible Party – U.S. Contact Information:
Ohaus Corporation
7 Campus Drive, Suite 310
Parsippany, NJ 07054
United States
Phone: +1 973 377 9000
Web: www.ohaus.com
FCC Compliance Statement:
Note:
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense. Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.

LIMITED WARRANTY

Ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period. During the warranty period Ohaus will repair, or, at its option, replace any component(s) that proves to be defective at no charge, provided that the product is returned, freight prepaid, to Ohaus.
This warranty does not apply if the product has been damaged by accident or misuse, exposed to radioactive or corrosive materials, has foreign material penetrating to the inside of the product, or as a result of service or modification by other than Ohaus. In lieu of a properly returned warranty registration card, the warranty period shall begin on the date of shipment to the authorized dealer. No other express or implied warranty is given by Ohaus Corporation. Ohaus Corporation shall not be liable for any consequential damages.
As warranty legislation differs from state to state and country to country, please contact Ohaus or your local Ohaus dealer for further details.

Ohaus Corporation
7 Campus Drive
Suite 310
Parsippany, NJ 07054 USA
Tel: 973-377-9000
Fax: 973-944-7177
With offices worldwide.
www.ohaus.com

P/N 83021171 N © 2021 Ohaus Corporation, all rights reserved.

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals