ATi Q46S/81 Auto Chem Sulfide Monitor User Manual
- June 10, 2024
- ATi
Table of Contents
- ATi Q46S/81 Auto Chem Sulfide Monitor
- O & M Manual
- Product Information
- Product Usage Instructions
- Introduction
- Mechanical Installation
- Auto-Chem Startup
- Configuration
- Calibration
- PID Controller Details
- System Maintenance
- Troubleshooting
- Spare Parts
- PRODUCT WARRANTY
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
ATi Q46S/81 Auto Chem Sulfide Monitor
O & M Manual
Model Q46S/81 Auto-Chem Sulfide Monitor
Thank you for choosing the Q46S/81 Auto-Chem Sulfide Monitor. This manual
will provide you with all the necessary information to operate and maintain
your product effectively.
Product Information
- Home Office: Analytical Technology, Inc. 6 Iron Bridge Drive Collegeville, PA 19426 Phone: 800-959-0299 610-917-0991 Fax: 610-917-0992 Email: sales@analyticaltechnology.com
- European Office: ATI (UK) Limited Unit 1 & 2 Gatehead Business Park Delph New Road, Delph Saddleworth OL3 5DE Phone: +44 (0)1457-873-318 Fax: + 44 (0)1457-874-468 Email: sales@atiuk.com
- Table of Contents:
- Introduction
- General
- Q46S/81 System Specifications
- Q46S/81 Performance Specifications
- Installation
- General
- Wall Mount Bracket
- Panel Mount, AC Powered Monitor
- Auto-Chem Mounting
- Reagent Bottle Bracket and Pickup Assembly
- Inlet and Drain Connections
- Operation
- General
- Reagent
- Reagent Pump Priming
- Operation
- Internal Gas & Liquid Flow
- Sample Line Considerations
- Cleaning System Remote Activation
- User Interface
- User Interface
- Keys
- Display
- Software Navigation
- Measure Menu [MEASURE]
- Calibration Menu [CAL]
- Configuration Menu [CONFIG]
- Control Menu [CONTROL]
- Diagnostics Menu [DIAG]
- Calibration
- Calibration
- Zero Cal
- Span Cal
- Mixing a Sulfide Standard
- Temperature Calibration
- PID Control
- PID Description
- PID Algorithm
- Classical PID Tuning
- manual PID Override Control
- Common PID Pitfalls
- Maintenance
- General
- Analyzer Maintenance
- Sensor Maintenance
- Sample Pump Tube Replacement
- Reagent Pump Tube Replacement
- Troubleshooting
- General
- External Sources of Problems
- Analyzer Tests
- Display Messages
Product Usage Instructions
Before using the Q46S/81 Auto-Chem Sulfide Monitor, please read through the user manual and familiarize yourself with the product information provided. Follow the installation instructions provided in section 2 of the manual to set up the monitor properly. To operate the monitor, refer to section 4 for general instructions and section 5 for user interface details. Calibration instructions can be found in section 6. For maintenance and troubleshooting help, consult sections 8 and 9 of the manual respectively. If you encounter any issues while using the Q46S/81 Auto-Chem Sulfide Monitor, refer to section 9 for troubleshooting tips or contact the manufacturer’s support team for assistance. Thank you for choosing the Q46S/81 Auto-Chem Sulfide Monitor. We hope it meets your needs and exceeds your expectations.
Introduction
General
The Q46S/81 is an on-line monitoring system designed for the continuous
measurement of sulfide in solution. The full scale operating range of the
system may be selected by the user for 0-2.000 PPM or 0-20.00 PPM, and the
sensing system will operate on water streams with temperatures from 0 to 50C.
Sulfide ion measurement is made in the Q46S system by pumping a small amount
of sample out of the inlet overflow assembly, mixing the sample with acid to
convert any HS- or S= ions to hydrogen sulfide gas, and then air stripping the
H2S for measurement in the gas phase. A hydrogen sulfide gas sensor located in
the flow block next to the sample conditioning module measures the
concentration of H2S in the gas stream and is connected to an electronic
module with converts sensor signals to a direct concentration display. The
measured sulfide concentration is displayed on a backlit LCD on the front of
the instrument. Depending on the full scale operating range selected, the unit
will display sulfide with a resolution of either 0.01 PPM (10 PPB) or 0.001 (1
PPB). A 4-20 mA output is provided for recording or data logging.
The output can be programmed for a variety of ranges as is described later in this manual, with a minimum range of 0-0.1 PPM and a maximum range of 0-20 PPM. Alarm contacts are also standard in the electronic package and may be used either for simple control schemes or for signaling operators of abnormal operating conditions. The Q46S/81 system includes three components, a wall mounted NEMA 4X electronics unit, an Auto-Chem chemistry module housed in a wall-mount enclosure, and a reagent bottle mounting bracket. The electronics enclosure is supplied with a wall mounting bracket but may be panel mounted using the optional 05-0068 panel mount bracket kit. For connection of the chemistry module to the electronics, a 20′ sensor cable is supplied, and the cable length can be increased to 100′ maximum. Cable #31-0038 is available for extended cable runs. The typical system is shown in Figure 1 on the next page. Please note that the Auto-Chem chemistry module is designed for use in protected environments. It is not suitable for use outdoors without the addition of another enclosure that is temperature controlled.
Q46S/81 System Specifications
- Displayed Parameters
- Main input, 0.001 ppm to 20.00 ppm dissolved sulfide
- Sensor Current, 0.0 to 999.9 nA, 0.000 to 99.99 uA
- Sample Temperature
- Loop current, 4.00 to 20.00 mA
- Sensor slope & zero offset
- Model number and software version
- PID Controller Status
- Main Parameter Ranges Manual selection of one of the following ranges,
- 0.0 to 2.000 ppm
- 0.00 to 20.00 ppm
- Display 0.75” (19.1 mm) high 4-digit main display with sign
- 12-digit secondary display, 0.3″ (7.6 mm) 5×7 dot matrix.
- Integral LED back-light for visibility in the dark.
- Ambient Temperature Analyzer Service, 5 to 40 °C (41 to 104 ºF) Storage, -5 to 70 °C (-22 to 158 ºF)
- Ambient Humidity 0 to 95%, non-condensing.
- EMI/RFI Influence Designed to EN 61326-1
- Output Isolation 600 V galvanic isolation
- Filter Adjustable 0-9.9 minutes additional damping to 90% step input
- Sensor Electrochemical type for hydrogen sulfide gas
- Sensor Materials Noryl and PVC
- Interconnect Cable 20 ft. (6.15 meter) standard, 100 ft. (30 m) maximum
- Q46S Power 100-240 VAC ±10%, 50/60 Hz
- Auto-Chem Power 115 or 230 VAC, 50/60 Hz
- Q46S Enclosure NEMA 4X, IP66, polycarbonate, stainless steel hardware
- HWD: 4.9″ (124 mm) x 4.9″ (124 mm) x 5.5″ (139 mm)
- Flammability rating : UL 94 V-0
- Auto-Chem Enclosure NEMA 1, IP11, Kydex with Polycarbonate Clear Lid
- HWD : 4.9″ (124 mm) x 4.9″ (124 mm) x 5.5″ (139 mm)
- Flammability rating: UL 94 V-0
- Mounting Options Wall mount bracket standard. Panel mount adapter optional for Q46S Only
- Relays, Electromechanical Three SPDT, 6 amp @ 250 VAC, 5 amp @ 24 VDC contacts. Software selection for setpoint, phase, delay, deadband, hi-lo alarm, and failsafe. A-B indicators on main LCD.
- Analog Outputs Two 4-20 mA outputs. Output one programmable for PPM H2S or PID. Output 2 programmable for PPM or temperature. Maximum load 500 Ohms for each output. Outputs ground isolated and isolated from each other.
- Digital Communications Profibus-DP, Modbus-RTU. More versions pending. (Optional) See Q46 Digital Communications Manual for Specifications.
- Sample Flowrate 5 GPH Minimum (315 cc/min) 25 GPH Maximum (1.5 LPM)
- Weight 20 lbs. (9 Kg.)
Q46S/81 Performance Specifications
- Accuracy 1% of selected range or 0.05 PPM
- Repeatability 0.5% of selected range or 0.05 PPM
- Sensitivity 0.005 PPM
- Non-linearity 0.1% of selected range
- Warm-up Time 3 seconds to rated performance (electronics only) Sensor requires 8 hour stabilization at start-up
- Supply Voltage Effects ± 0.05% span
- Instrument Response Time 60 seconds to 90% of step input at lowest damping
Equipment bearing this marking may not be discarded by traditional methods in the European community after August 12 2005 per EU Directive 2002/96/EC. End users must return old equipment to the manufacturer for proper disposal.
Mechanical Installation
General
Mechanical installation of a Q46S/81 sulfide system involves mounting of the
Q46S electronic assembly, mounting of the Auto-Chem chemistry module, and
mounting of the reagent bottle bracket. Proper planning of the installation
will benefit operation and maintenance of the system. Here are a few
considerations
- Locate the Q46S electronics where personnel can easily access the front panel control keys. Calibration of the system requires access to these controls.
- Locate the chemistry module high enough above the floor so that servicing of the pumps is not difficult. Pump tubing must be replaced every 6 months so access to the chemistry system is required.
- Do not mount the reagent bottle closer than about 18” from the bottom of the chemistry module. Mounting too close will make removal of the bottle difficult.
- Water sample and drain lines connect to an inlet overflow assembly at the bottom of the chemistry module. Locate the chemistry module so that the drain line is short and can flow freely.
The Q46S electronics is wall mounted using a PVC plate supplied with the unit. The bracket kit contains 4 screws for attaching the plate to the back of the enclosure. A paper template is supplied to ease of locating anchors in the wall. The Auto-Chem chemistry module is supplied with 4 stainless steel mounting feet and 8 attachment screws (10-32 x ¼” long). Inserts in the back of the enclosure provide for attachment of the mounting feet.
Wall Mount Bracket
A PVC mounting bracket with attachment screws is supplied with each
transmitter (see Figure 3 for dimensions). The multi-purpose bracket is
attached to the rear of the enclosure using the four flat head screws. The
instrument is then attached to the wall using the four outer mounting holes in
the bracket. These holes are slotted to accommodate two sizes of U-bolt that
may be used to pipe mount the unit. Slots will accommodate U-bolts designed
for 1½ “or 2” pipe. The actual center to center dimensions for the U-bolts are
shown in the drawing. Note that these slots are for U-bolts with ¼-20 threads.
The 1½” pipe U-bolt (2” I.D. clearance) is available from ATI in type 304
stainless steel under part number 47-0005
Panel Mount, AC Powered Monitor
Panel mounting of the Q46S monitor uses the panel mounting flange molded into
the rear section of the enclosure. Figure 6 provides dimensions for the panel
cutout required for mounting. The panel bracket kit must be ordered separately
(part number 05-0094). This kit contains a metal retainer bracket that
attaches to the rear of the enclosure, 4 screws for attachment of this
bracket, and a sealing gasket to ensure that the panel mounted monitor
provides a water tight seal when mounted to a panel. The sealing gasket must
first be attached to the enclosure. The gasket contains an adhesive on one
side so that it remains in place on the enclosure. Remove the protective paper
from the adhesive side of the gasket and slide the gasket over the back of the
enclosure so that the adhesive side lines up with the back of the enclosure
flange. Once in place, mount the monitor in the panel.
Auto-Chem Mounting
The chemistry module should be mounted on a wall or flat plate. A template is
provided for marking the location of mounting screws. The stainless steel
mounting feet must first be screwed to the back of the enclosure using the
screws supplied. Tape the mounting template to the wall and mark the 4 screw
locations. Mount any wall anchors first and then screw in the two bottom
mounting screws about halfway. Place the two bottom mounting feet of the
enclosure on to the screws and tilt the enclosure into place. Install the two
top screws to secure the enclosure and then tighten the bottom screws.
Reagent Bottle Bracket and Pickup Assembly
The Auto-Chem system requires approximately 1 gallon of reagent every 25 days.
A one gallon bottle holder is supplied and mounts to the wall. It is
sufficient to mount the bracket using only the upper two mounting holes. A
reagent pick-up assembly is supplied that fits into the mouth of the bottle
and keeps the pick-up tubing from floating to the top of the liquid.
Inlet and Drain Connections
Note : To avoid shipping damage, the inlet assembly is shipped separately and must be installed on the bottom of the chemistry module after the enclosure has been mounted to the wall. The inlet assembly is supplied with 4 mounting screws for attachment to the enclosure. The photo below shows the inlet assembly attached to the enclosure.
The water sample containing dissolved sulfide must be connected to the inlet overflow assembly mounted on the bottom of the chemistry module. For convenience, a 10 ft. piece of I.D. PVC tubing is supplied. This tubing mates with the inlet hose barb located on the left side of the inlet assembly. The drain connection is a I.D. hose barb on the right side of the overflow assembly. A 5 ft. length of PVC drain tubing is also supplied. The drain should be a free flowing gravity drain. No backpressure is allowed on the drain line.
Electrical Installation
General
The Q46 electronics contains a universal power supply operating on voltages
between 90 and 265 VAC, 50/60 cycle. Auto-Chem Chemistry modules are not
universal and must be ordered for operation from EITHER 120 VAC or 230 VAC.
Please verify the type of unit before connecting any power. The Auto-Chem
assembly contains two DIN rail mounted power supplies, one of which is AC
power dependent. The 24 VAC supply located next to the AC power input
terminals is the power supply that must be compatible with the installation
site. If there is any question, remove the power supply from the DIN rail and
check the label on the bottom of the supply. Connection of 230 VAC power to a
115 VAC unit will result in damage to the power supply assembly.
Important Notes:
- Use wiring practices that conform to all national, state and local electrical codes. For proper safety as well as stable measuring performance, it is important that the earth ground connection be made to a solid ground point from terminal 12 (Figure 10).
- Do NOT run sensor cables or instrument 4-20 mA output wiring in the same conduit that contains AC power wiring. AC power wiring should be run in a dedicated conduit to prevent electrical noise from coupling with the instrumentation signals.
- This analyzer must be installed by specifically trained personnel in accordance with relevant local codes and instructions contained in this operating manual. Observe the analyzer’s technical specifications and input ratings. Proper electrical disconnection means must be provided prior to the electrical power connected to this instrument, such as a circuit breaker – rated 250 VAC, 2 A minimum. If one line of the line power mains is not neutral, use a double-pole mains switch to disconnect the analyzer.
- Repeated problems with lightning strikes damaging sensitive instrumentation are often attributed to poorly bonded earth grounds in the instrument power source. The protection schemes incorporated into this analyzer cannot operate to maximum efficiency unless the ground connection is at its’ absolute lowest impedance.
There is no standard ground resistance universally recognized. Many agencies
recommend a ground resistance value of 5 ohms or less. The NEC recommends an
impedance to ground of less than 25 ohms, and less than 5 ohms where sensitive
equipment is installed. Power sources feeding sensitive instruments like the
Q46H/79PR should have the lowest possible impedance to ground.
Q46 electronic units are supplied with 5 NPT ports, two on each side and one
on the bottom. Red plugs are provided for each port. Five cord grips are
supplied separately for use during installation. A cord grip should be
installed in the bottom port for sealing the sensor cable connection to the
Auto-Chem unit. AC power should enter through the lower right hand port as the
AC terminals are closest to this entry.
WARNING
Disconnect line power voltage BEFORE connecting line power wires to Terminal
TB7 of the power supply. The power supply accepts only standard three-wire
single phase power. The power supply is configured for 100-240 VAC ±10%
operation.. Do NOT connect voltages other than the labeled requirement to the
input.
Connect HOT, NEUTRAL, and GROUND to the matching designations on terminal strip TB7. The analog outputs from the system are present at terminals TB1. The loop-load limitation in this configuration is 500 Ohms maximum for each output. Also note that these two outputs are completely isolated from each other to ensure that ground loops do not result from the connection of both outputs to the same device such as a PLC or DCS. A ribbon cable connects the power supply assembly with the microprocessor assembly located in the front section of the enclosure. This cable can be disconnected from the front section during installation to facilitate wiring. The ribbon cable has a marking stripe on one edge that is used to indicate proper orientation.
The power strip, TB5, allows up to 12 AWG wire. A wire gauge of 16 AWG is recommended to allow for an easy pass-through into the NPT ports when wiring.
Auto-Chem Power Panel
The chemistry module contains terminals and power supplies on a DIN rail
located behind the front panel. During installation, it is often convenient to
completely remove the front panel, allowing access to terminals inside.
Disconnect the inlet and outlet fittings at the top of the inlet overflow
assembly. To remove the front panel, open the latch at the top of the panel
and allow the panel to hinge forward. A retaining strap will stop the panel
from moving too far. The two pumps are plugged into the 24 VAC power supply on
the right. The sample conditioner cooling element is plugged into the 6 VDC
power supply on the left. It is easiest to unplug all 4 items at the power
supplies. The RTD temperature element wires have a molex connector that can be
unplugged. Also, remove the air line from the lower fitting of the rotameter.
Once everything is disconnected, open the Velcro on the retaining strap and
slide the strap through the slots on the front panel. At this point, the
entire front panel can be removed from the enclosure. Below is a photo of the
interior showing power supplies and terminals.
Q46S to Auto-Chem Connection
A 10-conductor cable connects the chemistry module to the Q46 monitor and
includes 6 conductors for the sensor and 4 for the automated sample line
cleaning function. The chemistry module connections for this cable are done at
the factory and 10 ft. length of cable is supplied for routing to the monitor.
This cable can be cut to the length required by a particular installation.
Eliminating excess cable can reduce the potential for electrical interference.
When stripping cables, leave adequate lead lengths for connection in the
transmitter enclosure.
Figure 12 provides detail of the internal connections and the location of the
AC power connection for the Auto-Chem Module. Figure 13 shows the
interconnecting wiring between the Auto-Chem and Q46 monitor. Route signal
cable away from AC power lines, adjustable frequency drives, motors, or other
noisy electrical signal lines. Do not run sensor or signal cables in conduit
that contains AC power lines or motor leads.
Relay Wiring
Three SPDT relays are provided on the power supply board. None of the relay
contacts are powered. The user must supply the proper power to the contacts.
For applications that require the same switched operating voltage as the Q46
(115 or 230 V), power may be jumpered from the power input terminals at TB7.
Relay wiring is connected at TB4,TB5, and TB6 as shown below. Note that the
relay contact markings are shown in the NORMAL mode. Programming a relay for
“Failsafe” operation reverses the NO and NC positions in this diagram (Figure
14).
External Line Cleaning Control
As noted previously, the Auto-Chem chemistry unit contains a relay that can be
used to control an external chemical feed system for automating sample line
cleaning. The relay is mounted on the power panel and wiring is shown in
Figure 15 below.
Auto-Chem Startup
General
Prior to operating the chemistry module, connect inlet sample and drain tubing
to the hose barbs provided on inlet overflow assembly. Remove the clear cover
from the chemistry module and check the inlet and outlet tube connections on
the inside cover of the overflow assembly. Also, check to be sure that the
reagent pickup tube is connected to the inlet fitting of the regent pump.
Two sensors for the system are shipped in a separate container. One of these
sensors contains a keeper plug on the end and the other does not. The keeper
plug maintains the sensor in stable condition ready for immediate use.
Disconnect the keeper plug and plug the sensor cable into the back of the
sensor. Remove the protective cap from the end of the sensor and slide it into
the sensor flow block. Plug the keeper into the spare sensor and store it in a
convenient place. Storage in a sealed container in a cool location is best.
Reagent
Q45S/81 Sulfide systems require acid to convert all sulfides to H2S. The acid
required for operation must be supplied by the user. Either sulfuric acid or
acetic acid will generally be suitable for use in this analyzer. A one gallon
container with marking placard is supplied, along with a bottle holder. The
following solutions may be used. Note that acetic acid is not as strong an
acid as sulfuric acid, but it is safer to use. If using vinegar (5% acetic
acid), verify that the water coming from the sample pump drain tube has a pH
below 5.
- Sulfuric Acid : Use a 5% solution
- Acetic Acid : Use a 5% solution (vinegar is 5% acetic acid)
Reagent Pump Priming
Acid is pulled from the gallon container by a peristaltic pump in the Auto-
Chem. The pump rate for chemical feed is about 0.1 cc/min. It can take up to
20 minutes for acid to reach the sample injection point when power is first
turned on. If desired, the acid feed tube can be manually filled to reduce the
waiting time. A 10 cc syringe with plastic tip is supplied for this purpose.
To fill the feed tube, release the pressure on the reagent pump tube by
loosening the clamp plate on the left side of the pump. Then disconnect the
1/16” I.D. tube from the injection tee below the sample pump. Insert the tip
of the syringe into the tube and use the syringe to draw reagent from the
bottle into the syringe. When reagent has reached the syringe, replace the
pressure plate on the reagent pump. Once the pressure plate is properly
seated, disconnect the syringe and connect the tubing to the injection tee.
Carefully discard any acid in the syringe and rinse.
Operation
As noted in the “Electrical Installation” section, the chemistry module
contains a lever fuse holder that can be used to turn the chemistry system on
and off. Once the steps outlined earlier in this section are complete and
power has been connected to the proper terminals, simply close the lever fuse
to activate the chemistry module. Once power is applied, the two peristaltic
pumps will start to turn and the rotameter on the panel should begin to
indicate the air flowrate. Use the air flow control valve to adjust the
rotameter to 400 cc. per minute. The air is discharged into the reactor and
you will see bubbling in that chamber. The final discharge of the air is into
the drain chamber of the inlet overflow assembly, and you should also see
bubbling there. Assuming that sample is flowing to the overflow assembly,
sample will reach the reactor in just a few seconds. If the reagent feed has
been manually primed, the system will begin to function almost immediately. It
is best to allow the system to run undisturbed for 1 hour after initial
startup. Note that the electronics module should also be powered up at this
time. The sensor will not stabilize unless the power to the monitor is turned
on.
Internal Gas & Liquid Flow
In the Auto-Chem chemistry module there are two flow paths, one for gas and
one for liquid. Understanding how these two flow paths should function will
allow you to quickly verify that the system in operating normally. The gas
flow is used to strip hydrogen sulfide gas from the acidified sample and carry
it to the sensor. An air pump behind the front panel pulls ambient air from
inside the enclosure and pumps it through the rotameter and into the clear
acrylic stripping chamber. The air exits the stripping chamber through the
chamber plug at the top and flows through the Teflon gas conditioning block
where excess moisture is removed. From the conditioning block, the air flows
through the sensor chamber and then out through a tube connected to the drain
chamber of the inlet overflow at the bottom of the enclosure. During normal
operation, bubbling should be observed in both the stripping chamber and in
the overflow outlet chamber. The liquid flow originates in the inlet chamber
of the overflow block. Sample is pulled up from that chamber by vacuum and
reagent is injected into the sample just below the sample pump. The treated
water is pumped into the stripping chamber where it is stripped of hydrogen
sulfide. The water overflows into the drain side of the stripping chamber
where it is pumped out. The water from the drain side is then pumped into the
rear chamber of the gas conditioning block and from there it flows to the
drain chamber of the overflow assembly. One other liquid flow path exists but
does not carry much flow. One channel of the reagent pump is connected to the
bottom of the gas conditioning block and is used to pump condensate out of the
block and into the main drain line.
Sample Line Considerations
Sulfide monitor sampling systems can suffer from biological contamination that
changes the sulfide concentration in the sample during transport to the
monitor. This occurs because sulfur reducing bacteria grow in the sample
lines. When the amount of bacterial fouling is high, the sulfide in the sample
is consumed by the bacteria during the time it is flowing through the system.
To eliminate this problem, sample lines should periodically be cleaned with
high levels of chlorine. Injecting sodium hypochlorite solution to provide
50-100 PPM of chlorine in the sample line for a few minutes every few days
will usually control growth in the system. The Q46S/66 monitor provides
software to automate this cleaning procedure. To implement this function, the
user must provide a chemical feed pump that will inject the required
hypochlorite solution. Starting and stopping of that pump is controlled by the
monitor. The Auto-Chem chemistry module contains two relays which are
controlled by the Q46 monitor. One of these relays is used to turn the
internal air pump on and off during sample line cleaning. The other is an
unpowered contact used to signal a chemical feed pump to start and stop. The
relay in the chemistry module should be used as a pilot relay to signal a
power relay to switch the pump on and off. Switching inductive electrical
loads inside the chemistry module can result in interference with the sulfide
measurement and should be avoided. Details on programming the cleaning
function are found later in this manual. A schematic of how a typical line
cleaning system might be configured is shown on the following page.
Cleaning System Remote Activation
The automatic sample line cleaning function is not required for operation of
the sulfide monitor but can be useful in reducing manual cleaning. This
function can be activated in one of two ways, either with the internal timer
in the Q46 or by a contact closure input controlled remotely. If the internal
Q46 timer is used, the cleaning cycle will occur at whatever preset interval
is desired. The interval is programmed in the Control Menu of the Q46. Timer
function T1 is selected when the Q46 monitor controls the start of the
cleaning cycle. Activating the cleaning cycle remotely requires a normally
open contact input controlled by the user. The contact input must be a dry
contact and must close for a minimum of 2 seconds. There can be no power on
the contacts used to start the cycle. A contact input carrying voltage is
likely to cause severe damage to the Q46. The contact input is wired to a
terminal block located in the chemistry module. That terminal block contains
terminals marked 1 and 2 as shown below. When programming the Q46 Timer
Functions, be sure that the timer function T2 is selected.
Configuration
User Interface
The user interface for the Q46 Series instrument consists of a custom display
and a membrane keypad. All functions are accessed from this user interface (no
internal jumpers, pots, etc.).
Keys
All user configurations occur through the use of four membrane keys. These
keys are used as follows:
- MENU/ESC To scroll through the menu section headers or to escape from anywhere in software. The escape sequence allows the user to back out of any changes in a logical manner. Using the escape key aborts all changes to the current screen and backs the user out one level in the software tree. The manual will refer to this key as either MENU or ESC, depending upon its particular function.
- UP (arrow) To scroll through individual list or display items and to change number values.
- LEFT (arrow) To move the cursor from right to left during changes to a number value.
- ENTER To select a menu section or list item for change and to store any change.
Display
The large custom display provides clear information for general measurement
use and user configuration. There are three main areas of the display: the
main parameter display, the secondary message line, and the icon area.
Main Parameter
During normal operation, the main parameter display indicates the present
process input with sign and units. This main display may be configured to
display any of the main measurements that the system provides. During
configuration, this area displays other useful set-up information to the user.
Lower Line
During normal operation, the lower line of the display indicates
userselected secondary measurements that the system is making. This also
includes calibration data from the last calibration sequence and the
transmitter model number and software version. During configuration, the lower
line displays menu items and set-up prompts to the user. Finally, the lower
line will display error messages when necessary. For a description of all
display messages, refer to Section 9.4.
Icon Area
The icon area contains display icons that assist the user in set-up and
indicate important states of system functions. The CAL, CONFIG, and DIAG icons
are used to tell the user what branch of the software tree the user is in
while scrolling through the menu items. This improves software map navigation
dramatically. Upon entry into a menu, the title is displayed (such as CAL),
and then the title disappears to make way for the actual menu item. However,
the icon stays on.
HOLD
The HOLD icon indicates that the current output of the transmitter has been
put into output hold. In this case, the output is locked to the last input
value measured when the HOLD function was entered. HOLD values are retained
even if the unit power is cycled.
FAIL
The FAIL icon indicates that the system diagnostic function has detected a
problem that requires immediate attention. This icon is automatically cleared
once the problem has been resolved.
Relay Area A/B
The relay area contains two icons that indicate the state of the system relays
(if the relay card is installed). Relay C is normally configured for FAIL, so
it is only displayed on the lower MEASURE display line.
Software
The software of the Q46S is organized in an easy to follow menu-based system.
All user settings are organized under five menu sections: Measure, Calibration
[CAL], Configuration [CONFIG], Control [CONTROL] and Diagnostics [DIAG].
Note: The default Measure Menu is display-only and has no menu icon.
Software Navigation
Within the CAL, CONFIG, CONTROL, and DIAG menu sections is a list of
selectable items. Once a menu section (such as CONFIG) has been selected with
the MENU key, the user can access the item list in this section by pressing
either the ENTER key or the UP arrow key. The list items can then be scrolled
through using the UP arrow key. Once the last item is reached, the list wraps
around and the first list item is shown again. The items in the menu sections
are organized such that more frequently used functions are first, while more
permanent function settings are later in the list. See Figure 199 for a visual
description of the software. Each list item allows a change to a stored system
variable. List items are designed in one of two forms: simple single variable,
or multiple variable sequences. In the single variable format, the user can
quickly modify one parameter – for example, changing the display range from
2.000 to 20.00.
In the multiple variable sequence, variables are changed as the result of some process. For example, the calibration of hydrogen sulfide generally requires more than one piece of information to be entered. The majority of the menu items in the software consist of the single variable format type. Any data that may be changed will be flashing. This flashing indicates user entry mode and is initiated by pressing the ENTER key. The UP arrow key will increase a flashing digit from 0 to 9. The LEFT arrow key moves the flashing digit from right to left. Once the change has been completed, pressing ENTER again stores the variable and stops the flashing. Pressing ESC aborts the change and also exits user entry mode. The starting (default) screen is always the Measure Menu. The UP arrow key is used to select the desired display. From anywhere in this section the user can press the MENU key to select one of the four Menu Sections. The UP arrow icon next to all list items on the display is a reminder to scroll through the list using the UP arrow key. To select a list item for modified
Measure Menu [MEASURE]
The default menu for the system is the display-only menu MEASURE. This menu is
a display-only measurement menu and has no changeable list items. When left
alone, the instrument will automatically return to this menu after
approximately 30 minutes. While in the default menu, the UP arrow allows the
user to scroll through the secondary variables on the lower line of the
display. A brief description of the fields in the basic transmitter version is
as follows:
TRANSMITTER MEAS SCREENS
- 25.7C Temperature display. Can be displayed in °C or °F, depending on user selection. A small “m” on the left side of the screen indicates the transmitter has automatically jumped to a manual 25C setting due to a failure with the temperature signal input.
- 32.0 Na Raw sensor current. Useful for diagnosing problems.
- 100% 20.00 mA PID Status screen (if enabled.) Shows the present controller output level on left, and actual transmitter current on the right. The controller can be placed in manual while viewing this screen by pressing and holding the ENTER key for 5 seconds until a small flashing “m” appears on the screen. At that point the controller output can be adjusted up or down using the UP and LEFT arrow keys. To return to automatic operation, press and hold the ENTER key for 5 seconds and the “M” will disappear.
- 4.00 mA Transmitter output current # 1
- 20.00 mA Transmitter output current # 2 Slope = 100% Sensor output response vs. ideal calibration. This value updates after each calibration. As the sensor ages, the slope reading will decay indicating sensor aging. Useful for resolving sensor problems.
- Offset = 0.0 nA Sensor output current at a zero ppm input. This value updates after a zerocalibration has been performed. Useful for resolving sensor problems.
- Q46S VX.XX Transmitter software version number.
- T1-cyc 24.0h Time remaining until next cleaning cycle. Appears only if the cleaning function is enabled.
- T2-wait… Indicates that the timer function is waiting for a contact input. Appears only if the T2 timer is enabled.
Note: A display test (all segments ON) can be actuated by pressing and holding the ENTER key while viewing the model/version number on the lower line of the display.
The MEASURE screens are intended to be used as a very quick means of looking up critical values during operation or troubleshooting.
Calibration Menu [CAL]
The calibration menu contains items for frequent calibration of user
parameters. There are three
items in this list: Cal Sulfide, Cal Temp. and Set Range.
- Cal Allows the user to adjust the transmitter span reading to match a reference solution, or to set the sensor zero point. See Part 6 – Calibration for more details.
- Cal Temp Allows the user to adjust the offset of the temperature response by a small factor of ±5 °C. The temperature input is factory calibrated. Long cable lengths and junction boxes may degrade the accuracy in some extreme situations. See Part 6 – Calibration for more details.
- Set Range Provides for user selection of the display range. Once set, all output functions use this display range to establish configuration settings. Press ENTER to initiate user entry mode, and the value will flash. Use the arrow keys to modify value; available ranges include 2.000 ppm, 20.00 ppm, and 200.0 ppm. Press ENTER to store the new value.
Configuration Menu [CONFIG]
The Configuration Menu contains all of the general user settings:
- Entry Lock Used to inhibit unauthorized tampering with instrument settings. All settings may be viewed while the instrument is locked but cannot be modified. The Entry Lock is a toggle-type setting; that is, entering the correct code will lock the transmitter and entering the correct code again will unlock it. The code is preset at a fixed value. Press ENTER to initiate user entry mode and the first digit will flash. Use arrow keys to modify value. See Page 64 for the Q46S lock/unlock code. Press ENTER to toggle lock setting once code is correct.
- Set Delay Sets the amount of damping on the instrument. Both the display and the output value are affected by the degree of damping. Functions such as calibration are not affected by this parameter. The calibration routines contain their own filtering and stability monitoring functions to minimize the calibration timing. Press ENTER to initiate user entry mode, and the value will flash. Use the arrow keys to modify value; range is 0.2 to 9.9 minutes. Press ENTER to store the new value. Default is 0.2.
- Contrast Sets the contrast level for the display. The STN display provides the highest possible contrast and widest viewing angle under all conditions. Contrast control will rarely be adjusted, but extreme operating conditions could result in the need for a small change. Press ENTER to initiate user entry mode, and the value will flash. Use arrow keys to modify the value; range is 0 to 8 (0 being lightest). Press ENTER to update and store the new value. Default is 8.
- Main Units This function allows the user to select either PPM or mg/l. Default is PPM
- Zero Filter This function forces the reading to zero when reading is below the entered value. For example, if the entered value were 0.0020 the display at 0- 0019 would indicate 0.000. This feature is useful in blanking out zero noise. Default is 0.
- Main Display Allows the user to change the primary display. The selection is either sulfide concentration or output current. Press ENTER and the display will flash. Use the UP arrow key to modify the desired display value. Press ENTER to store the new value. Default is sulfide concentration.
- Timer Funcs Activates the automatic sample line cleaning feature and allows selection of either internal or remote activation. The choices are OFF, SENS T1 for internal timing, or SENS T2 for remote activation. Default setting is OFF.
- H2S Mode This function is not used in the Q46S/81 Sulfide Monitor and must always remain OFF.
- Com Mode Sets digital communication mode. A digital communication card must be plugged into the power supply slot for this function to work. Press ENTER and display will flash. Use the UP arrow key to modify the desired value; selections include 1-None, 2- P-DP for Profibus DP, 3 – Modbus, 4 – Ethernet IP. Press ENTER to store the new value.
- Com Address Sets bus address for digital communication. Press ENTER and the display will flash. Use the UP arrow key to modify the desired value. Range is 1- 125. Press ENTER to store the value.
- Com Baud Sets the data transmission speed. Press ENTER and the display will flash. Use the UP arrow key to modify the desired value. The selections are 300 Hz, 600 Hz, 900 Hz, 1.2 kHz, 2.4 kHz, 4.8 kHz, 9.6 kHz, 14.4 kHz, 19.2 kHz, 28.8 kHz, 38.4 kHz, 57.6 kHz, and 115.2 kHz. Press ENTER to store the value.
- Com Parity Sets data parity for digital communication. Press ENTER and the display will flash. Use the UP arrow key to modify the desired value. Selections are 1-None, 2-Odd, or 3-Even. Press ENTER to store the value
- Iout#1 Mode Sets analog output #1 for either sulfide concentration or PID control. Press ENTER and the display will flash. Use the UP arrow key to select. Selections include 1-ppm for sulfide concentration or 2-PID control. Press ENTER to store the new value. Default is 1-Sulfide.
- Iout#2 Mode Sets analog output #2 for either temperature or sulfide. Press ENTER and the display will flash. Use the UP arrow key to select. Selections include: 1-°C for temperature, or 2-PPM for sulfide. Press ENTER to store the new value. Default is 1-°C.
- Rly A Mode Relay A can be used in three different ways: as a setpoint control, as a fail alarm, or as a HI-LO alarm band. The three settings for Rly A Mode are CON, FAIL and AL. The CON setting enables normal control operation for Relay A, with settings for setpoint, hysteresis, delay and phasing appearing in the CONFIG menu automatically. See Figure 20 for further details. The FAIL setting enables the fail alarm mode for Relay A. Relay A will then trip on any condition that causes the FAIL icon to be displayed on the LCD. Using this mode allows the User to send alarm indications to other remote devices. The AL setting allows two setpoints to be selected for the same relay, producing a HI-LO alarm band. In this mode, Relay A will trip inside or outside of the band, depending upon the Phase selected. See Figure 21 for further details.
- Relay B Mode Relay B and C can be used in two ways: as a setpoint control, or as an
- Relay C Mode alarm. The two settings for Relay B Mode are CON and FAIL. The CON setting enables normal setpoint operation for Relay B/C. Relay B/C then operates identically to Relay A, with settings for setpoint, hysteresis, delay and phasing appearing in the CONFIG menu automatically. The FAIL setting enables the fail alarm mode for Relay B/C. Relay B/C will then trip on any condition that causes the FAIL icon to be displayed on the LCD. Note that the Relay C indicator shows up only on the lower screen of the display next to the temperature reading. This is because the default setting for relay C is the FAIL setting. Using this mode allows the User to send alarm indications to other remote devices.
Control Menu [CONTROL]
The Control Menu contains all of the output control user settings:
- Set 4 mA Allows adjustment of analog output #1. The units displayed depend on Set 20 mA the selection made in the CONFIG menu for Iout #1 Mode. [Iout1=PPM] The value stored for the 4 mA point may be higher or lower than the value stored for the 20 mA point. The entry values are limited to values within the range specified in “Set Range” in the Config Menu, and the 4 mA and the 20 mA point must be separated by at least 1% of this range Use the LEFT arrow key to select the first digit to be modified. Then use the UP and LEFT arrow keys to select the desired numerical value. Press ENTER to store the new value.
- Set PID 0% If the PID is enabled, on Output #1, this function sets the minimum and Set PID 100% maximum controller end points. Unlike the standard 4-20 mA output, the [Iout1=PID] controller does not “scale” output values across the endpoints. Rather, the endpoints determine where the controller would normally force minimum or maximum output in an attempt to recover the setpoint. If the 0% point is lower than the 100% point, then the controller action will be “reverse” acting. That is, the output of the controller will increase if the measured value is less than the setpoint, and the output will decrease if the measured value is larger than the setpoint. Flipping the stored values in these points will reverse the action of the controller to “direct” mode. The entry value is limited to a value within the range specified in “Set Range”, and the 0% and the 100% point must be separated by at least 1% of this range Use the LEFT arrow key to select the first digit to be modified. Then use the UP and LEFT arrow keys to select the desired numerical value. Press ENTER to store the new value.
- PID Setpnt The value which the controller is attempting to maintain by adjusting [Iout1=PID] output value.
- PID Prop Proportional gain factor. The proportional gain value is a multiplier on [Iout1=PID] the controller error (difference between measured value and setpoint value.) Increasing this value will make the controller more responsive.
- PID Int Integral is the number of “repeats-per-minute” of the action of the [Iout1=PID] controller. It is the number of times per minute that the controller acts on the input error. At a setting of 2.0 rpm, there are two repeats every minute. If the integral is set to zero, a fixed offset value is added to the controller (manual reset.) Increasing this value will make the controller more responsive.
- PID Deriv Derivative is a second order implementation of Integral, used to suppress [Iout1=PID] “second-order” effects from process variables. These variables may include items like pumps or mixers that may have minor impacts on the measured value. The derivative factor is rarely used in water treatment process, and therefore, it is best in most cases to leave it at the default value. Increasing this value will make the controller more responsive.
- Set 4 mA #2 Sets the span of analog output #2. The default setting for this output is Set 20 mA #2 temperature, but it may be set for PPM H2S if preferred. The values [Temp / PPM] stored for the 4 mA point may be higher or lower than the value stored for the 20 mA point. The entry value is limited to a value between 0 and 55 °C if it is set for temperature. The 4 mA and the 20 mA point must be at least 20 units away from each other. Press ENTER to initiate user entry mode, and the value will flash. Use arrow keys to modify value. Press ENTER to store the new value.
- A Setpoint Programs the setpoint for relay A. The entry value is limited to a value within the range specified in “Set Range”. Use the LEFT arrow key to select the first digit to be modified. Then use the UP and LEFT arrow keys to select the desired numerical value. Press ENTER to store the new value.
- A Hysteresis Programs the hysteresis or “deadband” for Relay A. Hysteresis is most often used to control relay chattering; however, it may also be used in control schemes to separate the ON/OFF trip points of the relay. Press ENTER and the display will flash. Use the arrow keys to modify value. Press ENTER to store the new value.
- A Delay Programs the amount of time delay on the trip point for relay A. The entry value is limited to a value between 0 and 999 seconds. Press ENTER and the display will flash. Use arrow keys to modify value; range is 0 to 999 seconds. Press ENTER to store the new value.
- A Phasing Establishes the direction of the relay trip. When phase is HI, the relay activates when the measured value exceeds setpoint A. When the phase is LO, the relay energizes and the LCD indicator illuminates when the measured value drops below the setpoint. The failsafe setting does have an impact on this logic. The description here assumes the failsafe setting is OFF. Press ENTER and the display will flash. Use the UP arrow key to select either HI for direct operation or LO for reverse operation. Press ENTER to store the new value.
NOTE : If the temperature units are changed between °C and °F (see Temp Units in this section), the default setting for this output will change between 100 °C and 212 °F accordingly.
- When value rises to ≥ 1.000 ppm, relay closes.
- When value rises to ≥ 1.050 ppm, relay opens.
Settings:
- Setpoint : 1.000 ppm
- Hyst : 0.050
- Delay : 000
- Failsafe : OFF
- Setpnt A-HI
- Hyst A-HI
- Delay A-HI
- Setpnt A-LO
- Hyst A-LO
- Delay A-LO
If Relay A Mode is set to Alarm Mode, AL, then the following settings will appear in the Config Menu list automatically. In this mode, two setpoints can be selected on the same relay, to create an alarm band. Phase HI selection causes the relay to energize outside of the band, and Phase LO causes the relay to energize inside of the band. This feature enables one relay to be used as a control relay while the other is used as a HI-LO Alarm relay at the same time. Setpoint A-LO must be set lower than Setpoint AHI. When AL mode is first selected, Setpoint A-LO is defaulted to 0.
- When value rises to ≥ 1.000 ppm, relay closes, until value falls back to < 0.950 ppm.
- When value falls to < 1.000 ppm, relay closes, until rises back to > 1.050 ppm
- When value falls to < 0.500 ppm, relay closes, until rises back to > 0.550 ppm.
- When value rises to ≥ 0.500 ppm, relay closes, until value falls back to < 0.450 ppm.
Settings:
- Setpoint A-HI : 1.000 ppm
- Setpoint A-LO : .500 ppm
- Hyst A-HI : 0.050
- Hyst A-LO : .0.050
- Delay A -HI: 000
- Delay A-LO : 000
- B Setpoint
- B Hysteresis
- B Delay
- B Phasing
- C Setpoint
- C Hysteresis
- C Delay
- C Phasing
If Relay B Mode is set to CON (see Relay B Mode), then Relay B will function identically to Relay A. Relay B settings appear in the CONFIG menu list automatically If Relay C Mode is set to CON (see Relay C Mode), then Relay C will function identically to Relay A. Relay C settings appear in the CONFIG menu list automatically.
- Timer Cycle Sets the frequency of the sample line cleaning function, if enabled in Config Menu. Press Enter to adjust the displayed interval to the desired number of hours between cleaning. Press Enter again to accept the new setting. Do not set the interval for less than 12 hours.
- Timer Hold Sets the output holding time after a cleaning cycle is complete. Recommended holding time is 2-4 minutes.
- Timer Clean Sets the amount of time that the relay controlling an external chemical feed pump will be energized. This is the amount of time that hypochlorite would be fed into the sample line to control biofouling. A period of 3-5 minutes is likely to be adequate, but experience will determine actual requirement.
- Timer Chem Sets the time period in which the Auto-Chem system is inhibited AFTER the cleaning period set in the previous step. During this period, the air pump in the system is shut off so that the sensor is not exposed to chlorine gas. A minimum of 3 minutes is recommended
Diagnostics Menu [DIAG]
The diagnostics menu contains all of the user settings that are specific to
the system diagnostic functions, as well as functions that aid in
troubleshooting application problems.
- Set Hold Locks the analog output values on the present value and halts operation of the PID controller. This function can be used prior to calibration, or when removing the sensor from the process to hold the output. Once HOLD is released, the outputs return to their normal function. The transfer out of HOLD is bumpless on both analog outputs – that is, the transfer occurs in a smooth manner rather than as an abrupt change. An icon on the display indicates the HOLD state, and the HOLD state is retained even if power is cycled. Press ENTER and the display will flash. Use the UP arrow key to select. Selections are ON for engaging the HOLD function, and OFF to disengage the function. Press ENTER to store the new value. The Set Hold function can also hold at an output value specified by the user. To customize the hold value, first turn the HOLD function on. Press the ESC key to go to the DIAG Menu and scroll to Sim Output using the UP arrow key. Press ENTER. Follow the instructions under Sim Output (see following page).
- Fault List The Fault List screen is a read-only screen that allows the user to display the cause of the highest priority failure. The screen indicates the number of faults present in the system and a message detailing the highest priority fault present. Note that some faults can result in multiple displayed failures due to the high number of internal tests occurring. As faults are corrected, they are immediately cleared. Faults are not stored; therefore, they are immediately removed if power is cycled. If the problem causing the faults still exists, however, faults will be displayed again after power is re-applied and a period of time elapses during which the diagnostic system re-detects them. The exception to this rule is the calibration failure. When a calibration fails, no corrupt data is stored. Therefore, the system continues to function normally on the data that was present before the calibration was attempted. After 30 minutes or if power to the transmitter is cycled, the failure for calibration will be cleared until calibration is attempted again. If the problem still exists, the calibration failure will re-occur. Press ENTER to initiate view of the highest priority failure. The display will automatically return to normal after a few seconds.
- Sim Out Allows the user to simulate the ammonia level of the instrument in the user selected range. Enter a ppm value directly onto the screen and the output responds as if it were actually receiving the signal from the sensor. This allows the user to check the function of attached monitoring equipment. Escaping this screen returns to normal operation. Press ENTER to initiate the user entry mode, and the right-most digit of the value will flash. Use arrow keys to modify desired value.
- PID Timer Sets a timer to monitor the amount of time the PID controller remains at 0% or 100%. This function only appears if the PID controller is enabled. If the timer is set to 0000, the feature is effectively disabled. If the timer value is set to any number other zero, a FAIL condition will occur if the PID controller remains at 0% or 100% for the timer value. If one of the relays is set to FAIL mode, this failure condition can be signaled by a changing relay contact. Press ENTER to initiate user entry mode, and the entire value will flash. Use the UP arrow key to modify desired value; range of value is 0-9999 seconds. Press ENTER to store the new value.
- Fail Out #1 Enables the user to define a specified value that the main current output will go to under fault conditions. When enabled to ON, the output may be forced to the current value set in Fail Val (next item.) With the Fail Out setting of ON, and a Fail Val setting of 6.5 mA, any alarm condition will cause the current loop output to drop outside the normal operating range to exactly 6.5 mA, indicating a system failure that requires attention. Press ENTER and the display will flash. Use the UP arrow key to modify desired value; selections are ON, OFF. Press ENTER to store the new value.
- Fail Val #1 Sets the output failure value for Iout#1. When Fail Out above is set to ON, this function sets value of the current loop under a FAIL condition. When the Relay Option Board is installed, the display will read Fail Out #1. The output may be forced to any current value between 4-20 mA. Press ENTER to initiate user entry mode, and the entire value will flash. Use the UP arrow key to modify desired value; selections are between 4mA, and 20mA. Press ENTER to store the new value.
- Fail Out #2 This function sets the fail-mode of current loop output #2 under a FAIL condition. The settings and operation are identical to Fail Out for output #1.
- Fail Val #2 This function sets the value of current loop output #2 under a FAIL condition. The settings and operation are identical to Fail Out for output #1.
- Backlight Selects the function of the LCD backlight. There are three options. ON – On all the time, OFF – Off all the time, AL – Alarm (Default). This function flashes the backlight on and off whenever the Fail icon is displayed.
- Failsafe Allows the user to set the relays to a failsafe condition. In a failsafe condition, the relay logic is reversed so that the relay coil is electrically energized in a normal operating state. By doing this, the relay will not only change state when the concentration limit is exceeded, but also when power is lost to the controller.
- Start Delay Minimizes control or alarm issues arising from temporary power loss. On power outage, the monitor records analog output values and relay status. When power is restored, the analog values and relays will be held at the pre-power loss values for a defined period of time. This “start delay” may be programmed for periods from 0-9.9 minutes.
- Set Default Allows the user to return the instrument to factory default data for all user settings (All) or for just the calibration default (Cal). It is intended as a last resort troubleshooting procedure. All user settings or the calibration settings are returned to the original factory values. Hidden factory calibration data remains unchanged. Press ENTER to initiate user entry mode and select either All or CAL with the UP arrow key.
Calibration
After the chemistry module has been running for an hour, the system should be zeroed using distilled water. To do this, disconnect the sample pickup line from the inlet overflow and connect the extension tube (supplied with the system). The extension tube is long enough to place into a bottle of distilled water placed near the unit
Zero Cal
Allow the system to run for about 30 minutes on distilled water. By the end of
that time, the sensor output should be at its lowest value. On the bottom line
of the display, you can see the actual sensor current if desired. This is the
“nA” display. If you watch that display, you should see a current value that
is normally below 50 nA. The value may be changing some up and down but will
not be trending in any one direction. If you look at the large numerical PPM
display, you will normally see a value below 0.3 PPM. Hydrogen Sulfide
sensors have fairly low offset currents at zero. In some cases, it is
sufficient to simply leave the zero at the factory default of 0.0 nA. If
measurements are being made that are normally above 5 PPM, leaving the unit at
electronic zero is satisfactory. If the units is to be used for low level
measurements, zeroing the sensor should be done at startup. An electronic zero
can be set by disconnecting the sensor from the cable and performing steps 1-3
below. The steps below assume that the system has been operating on distilled
water for 30 minutes after the initial 8 hour stabilization time
- Scroll to the CAL menu section using the MENU key and press ENTER or the UP arrow key. Cal H2S will be displayed.
- Press the ENTER key. The screen will display a flashing 1-Ref for span calibration or a 2- Zer for zero calibration. Using the UP arrow key, set for a 2-Zer zero calibration and press ENTER.
- The system now begins acquiring data for the sensor zero calibration value. As data is gathered, the units for sensor current in nanoamps (nA) will be displayed. Once the software has identified a stable value, the zero offset current will automatically be recorded. If you don’t wish to wait for the monitor to analyze the data, you can override by pressing ENTER. If the data remains unstable for 10 minutes, the calibration will fail and the message Cal Unstable will be displayed.
- If accepted, the screen will display the message PASS with the new sensor zero reading (offset), then it will return to the main measurement display. If the calibration fails, a message indicating the cause of the failure will be displayed and the FAIL icon will be turned on. The range of acceptable value for sensor offset is -999 to +999 nA. Should a FAIL occur, carefully inspect the sensor for membrane fouling. Should the offset value remain high and result in calibration failures, review the Service section of this manual, and then contact the service dept. at ATI for further assistance. The sensor zero offset value in nA from the last zero calibration is displayed on the lower line of the Default Menus for information purposes.
Span Cal
Calibration of the Sulfide Monitor can be done in one of two ways. If an
accurate laboratory measurement can be made on the same sample that the sensor
is measuring, the monitor can be adjusted to the laboratory result. A sulfide
standard can also be prepared in accordance with the procedure on the
following page, but the final value of the standard should be verified if
possible by laboratory test. A sample should be collected from the inlet line
feeding the flowcell and quickly analyzed for sulfide ion. When calibrating,
it is best to have a reasonably high concentration of sulfide in the system.
The higher the value, the smaller will be the calibration errors caused by
errors in the laboratory analytical procedure. It is generally preferable to
calibrate at values above 0.5 PPM to reduce calibration errors.
To calibrate the system, follow the procedure below
-
If a mixed standard is to be used, place the sample pickup tube into the container of standard. If you plan to us actual sample, collect this sample in a 500 cc container and have a portion of this sample tested for sulfide content.
-
Allow either the known sample or mixed sample flow for 5 minutes.
-
Scroll to the CAL menu section using the MENU key and press ENTER or the UP arrow key. Cal H2S is displayed.
-
Press the ENTER key. The screen will display a flashing 1-Ref for span calibration or a 2-Zer for zero calibration. Using the UP arrow key, set for a 1-Ref span calibration and press ENTER.
-
The system now begins acquiring data for the calibration value. As data is gathered, the units for ppm may flash. Flashing units indicate that this parameter is unstable. The calibration data point acquisition will stop only when the data remains stable for a pre-determined amount of time. This can be overridden by pressing ENTER. If the data remains unstable for 10 minutes, the calibration will fail and the message Cal Unstable will be displayed.
6. If stable, the screen will display the last measured ppm value and a message will be displayed prompting the user to adjust the value. Enter the sample or standard value depending on what you are using. Each digit must be adjusted separately using the key. Move to the left using the key. When adjusted, press Enter. -
If accepted, the screen will display the message PASS with the new sensor slope reading. It will return to the main measurement display. If the calibration fails, a message indicating the cause of the failure will be displayed and the FAIL icon will be turned on. The range of acceptable values for sensor slope is 25% to 500%.
Mixing a Sulfide Standard
While calibration against a lab analysis is preferred, it is possible to mix a
sulfide standard that can be used for calibration. The following procedure
will produce a standard of about 1.6 PPM. The actual value may be somewhat
lower, but this procedure should be within about 10-15% of that value. You
will require reagent grade sodium sulfide, an analytical balance, a volumetric
flask, and distilled water This standard requires 2 steps. First, accurately
weight out 1.000 grams of sodium sulfide and place into a 1 liter volumetric
flask. Fill to 1 liter with distilled water. This will provide a concentrated
standard of about 410 PPM and is stable for a few days. Use of a hydrated
sulfide salt will result in a different value that can be calculated using the
formula weight for that salt. The dilute standard is only stable for about an
hour, so this step should be taken just before use. Place 500 cc. of water in
a container and then add exactly 2 cc. of concentrated standard using a
pipette. Mix and feed into the sulfide monitor as a standard. The easiest way
to feed this standard is through the inlet port normally used for distilled
water. Simply use the same inlet tubing used for zeroing to introduce the
mixed standard. Once the monitor has stabilized, adjust the display to 1.64
PPM using the calibration procedure in the previous sections
Temperature Calibration
The temperature calibration sequence is essentially a 1-point offset
calibration that allows adjustments of approximately ±5 °C. The temperature
element is measuring the temperature of the sample inside the air stripping
chamber. This temperature may, or may not, be close to the temperature of the
water in the inlet overflow. Heat generated inside the chemistry module can
sometimes cause the displayed temperature to be above the actual sample
temperature. While temperature calibration is possible, it is not recommended.
If you wish to calibrate the temperature system, proceed as follows:
- Scroll to the CAL menu section using the MENU key and press ENTER or the UP arrow key.
- Press the UP arrow key until Cal Temp is displayed.
- Press the ENTER key. The message Place sensor in solution then press ENTER will be displayed. Press ENTER to begin the calibration sequence.
- The calibration data gathering process will begin. The message Wait will flash as data is accumulated and analyzed. The °C or °F symbol may flash periodically if the reading is too unstable.
- The message Adjust value – press ENTER will be displayed, and the right-most digit will begin to flash, indicating that the value can be modified. Using the UP and LEFT arrow keys, modify the value to your measured value. Adjustments up to ± 5 °C from the factory calibrated temperature are allowed. Press ENTER.
Once completed, the display will indicate PASS or FAIL. If the unit fails, the temperature adjustment may be out of range, the sensor may not have achieved complete temperature equilibrium, or there may be a problem with the temperature element. In the event of calibration failure, it is recommended to attempt the calibration again immediately.
PID Controller Details
PID Description
PID control, like many other control schemes, are used in chemical control to
improve the efficiency of chemical addition or control. By properly tuning the
control loop that controls chemical addition, only the amount of chemical that
is truly required is added to the system, saving money. The savings can be
substantial when compared to a system which may be simply adding chemical at a
constant rate to maintain some minimal addition under even the worst case
conditions. The PID output controller is highly advantageous over simple
control schemes that just utilize direct (proportional only) 4-20 mA output
connections for control, since the PID controller can automatically adjust the
“rate” of recovery based on the error between the setpoint and the measured
value – which can be a substantial efficiency improvement.. The PID controller
is basically designed to provide a “servo” action on the 4-20 mA output to
control a process. If the user requires that a measured process stay as close
as possible to a specific setpoint value, the controller output will change
from 0% to 100% in an effort to keep the process at the setpoint. To affect
this control, the controller must be used with properly selected control
elements (valves, proper chemicals, etc.) that enable the controller to add or
subtract chemical rapidly enough. This is not only specific to pumps and
valves, but also to line sizes, delays in the system, etc. This section is
included to give a brief description of tuning details for the PID controller,
and is not intended to be an exhaustive analysis of the complexities of PID
loop tuning. Numerous sources are available for specialized methods of tuning
that are appropriate for a specific application.
PID Algorithm
As most users of PID controllers realize, the terminology for the actual
algorithm terms and even the algorithms themselves can vary between different
manufacturers. This is important to recognize as early as possible, since just
plugging in similar values from one controller into another can result in
dramatically different results. There are various basic forms of PID
algorithms that are commonly seen, and the implementation here is the most
common version; The ISA algorithm (commonly referred to as the “ideal”
algorithm.)
Where:
- Output = controller output
- P = proportional gain
- I = integral gain
- D = derivative gain
- t = time
- e(t) = controller error (e=measured variable – setpoint)
The most notable feature of the algorithm is the fact the proportional gain term affects all components directly (unlike some other algorithms – like the “series” form.) If a pre-existing controller utilizes the same form of the algorithm shown above, it is likely similar settings can for made if the units on the settings are exactly the same. Be careful of this, as many times the units are the reciprocals of each other (i.e. reps-per-min, sec-per-rep.) PID stands for “proportional, integral, derivative.” These terms describe the three elements of the complete controller action, and each contributes a specific reaction in the control process. The PID controller is designed to be primarily used in a “closed-loop” control scheme, where the output of the controller directly affects the input through some control device, such as a pump, valve, etc. Although the three components of the PID are described in the setting area (section 5.25), here are more general descriptions of what each of the PID elements contribute to the overall action of the controller.
- P Proportional gain. With no “I” or “D” contribution, the controller output is simply a factor of the proportional gain multiplied by the input error (difference between the measured input and the controller setpoint.) Because a typical chemical control loop cannot react instantaneously to a correction signal, proportional gain is typically not efficient by itself – it must be combined with some integral action to be useful. Set the P term to a number between 2-4 to start. Higher numbers will cause the controller action to be quicker.
- I Integral gain. Integral gain is what allows the controller to eventually drive the input error to zero – providing accuracy to the control loop. It must be used to affect the accuracy in the servo action of the controller. Like proportional gain, increasing integral gain results in the control action happening quicker. Set the I term to a number between 3-5 to start (1-2 more than P). Like proportional gain, increasing the integral term will cause the controller action to be quicker.
- D Derivative gain. The addition of derivative control can be problematic in many applications, because it greatly contributes to oscillatory behavior. In inherently slow chemical control process’, differential control is generally added in very small amounts to suppress erratic actions in the process that are non-continuous, such as pumps and valves clicking on and off. However, as a starting point for chemical process control, its best to leave the “D” term set to 0.
Based on these descriptions, the focus on tuning for chemical applications really only involves adjustment of “P” and “I” in most cases. However, increasing both increases the response of the controller. The difference is in the time of recovery. Although combinations of high “P’s” and low “I” will appear to operate the same as combinations of low “P’s” and high “I’s”, there will be a difference in rate of recovery and stability. Because of the way the algorithm is structured, large “P’s” can have a larger impact to instability, because the proportional gain term impacts all the other terms directly. Therefore, keep proportional gain lower to start and increase integral gain to achieve the effect required. Many of the classical tuning techniques have the user start with all values at 0, and then increase the P term until oscillations occur. The P value is then reduced to ½ of the oscillatory value, and the I term is increased to give the desired response. This can be done with the Q46S controller, with the exception that the I term should start no lower than 1.0. If it appears that even large amounts of integral gain (>20) don’t appreciably increase the desired response, drop I back to about 1.0, and increase P by 1.00, and start increasing I again. In most chemical control schemes, I will be approximately 3 times the value of P.
Classical PID Tuning
Unlike many high speed position applications where PID loops are commonly
used, the chemical feed application employed by this instrument does not
require intense mathematical exercise to determine tuning parameters for the
PID. In fact, the risk of instability is far greater with overly tuned PID
control schemes. In addition, many of the classical mathematical exercises can
be damaging or wasteful in the use of chemicals when the process is bumped
with large amounts of input error to seek a response curve. Because of this,
the general adjustment guidelines described in section 7.2 are sufficient for
almost all application tuning for this instrument. Beyond this, many sources
are available for classical tuning methods.
Manual PID Override Control
The Q46 PID output function allows the user to take manual control of the PID
control signal. This is often useful when starting up a control loop, or in
the event that you wish to bump the system manually to measure system response
time. To access the manual PID control, you must be in the MEASURE mode of
operation and you must have the PID output displayed on the lower line. This
line will indicate “XX.X% XX.X mA” with the X values simply indicating the
current values. With this display on the screen, press and hold the ENTER key
for about 5 seconds. You will see a small “m” show up between the % value and
the mA value. This indicates you are now in manual mode. Once in manual, you
may increase the PID output by pressing the UP arrow or you may decrease the
output by pressing the LEFT arrow. This will allow you to drive the PID output
to any desired setting. To revert to normal PID control, press and hold the
ENTER key again until the “m” indicator disappears.
Common PID Pitfalls
The most common problem occurring in PID control applications involves the
false belief that proper settings on only the PID controller can balance any
process to an efficient level.
Close-loop control can only be effective if all elements in the loop are
properly selected for the application, and the process behavior is properly
understood. Luckily, the nature of simple
chemical control process’ are generally slow in nature. Therefore, even a de-
tuned controller (one that responds somewhat slowly) can still provide
substantial improvements to setpoint control. In fact, damaging oscillatory
behavior is far more likely in tightly tuned controllers where the user
attempted to increase response too much.
When deciding on a PID control scheme, it is important to initially review all elements of the process. Sticking valves, undersized pumps, or delays in reaction times associated with chemical addition can have a dramatic effect on the stability of the control loop. When controlling a chemical mix or reaction, the sensor should be placed in a location that ensures proper mixing or reaction time has occurred. The easiest process’ to control with closed- loop schemes are generally linear, and symmetrical, in nature. For example, controlling level in tank where the opening of valve for a fixed period of time corresponds linearly to the amount that flows into a tank. Chemical control process’ can be more problematic when the nature of the setpoint value is non-linear relative to the input of chemical added. For example, pH control of a process may appear linear only in a certain range of operation, and become highly exponential at the extreme ranges of the measuring scale. In addition, if a chemical process is not symmetrical, that means it responds differentially to the addition and subtraction of chemical. It is important in these applications to study steady-state impact as well as step-change impact to process changes. In other words, once the process has apparently been tuned under normal operating conditions, the user should attempt to force a dramatic change to the input to study how the output reacts. If this is difficult to do with the actual process input (the recommended method), the user can place the control in manual at an extreme control point such as 5% or 95%, and release it in manual. The recovery should not be overly oscillatory. If so, the loop needs to be de-tuned to deal with that condition (reduce P and/or I.)
System Maintenance
General
The Q46S/81 Auto-Chem Sulfide System requires regular maintenance for reliable
operation. Keep in mind that preventive maintenance on a regular schedule is
much less troublesome than emergency maintenance that always seems to come at
the wrong time.
Analyzer Maintenance
No maintenance of the electronic monitor is required if installed according to
the guidelines of this operating manual. During operation, it is best to use
only your finger to operate front panel keys. Using tools, especially sharp
objects, to press front panel keys will result in damage to the panel keys.
Sensor Maintenance
Sensor maintenance is limited to periodic inspection to be sure that the
membrane area is clean. It is unusual for contaminants to make it all the way
through the air system and reach the sensor, but it is good practice to
inspect the sensor every 2 or 3 months. The sensor used in this system will
have an operational life that is dependent on the amount of H2S that is being
routinely measured. Continuous operating below 1 PPM will normally result in
sensor life of 9-12 months. Constant operation above 10 PPM will result in
sensor life of about 6 months. After calibration, the “slope” value for a
sensor is calculated. When you see a slope value below 40% after a
calibration, the sensor is nearing the end of its life. It is good practice to
always have a spare sensor on hand. Two sensors are supplied with the unit.
When the first sensor is used up, it is best to order a replacement.
Sample Pump Tube Replacement
Peristaltic pump tubing replacement is the most common maintenance required
for the Auto-Chem system. The sample pump used in this system contains two
different tubes, one for the inlet (left) side of the pump and one for the
outlet (right) side of the pump. These tubes MUST be installed in the proper
location or the system will not work. The inlet tube has a slightly smaller
I.D. than the outlet tube. This is necessary to ensure that the air stripping
chamber does not fill with sample and force sample through the rest of the
system.
Both inlet and outlet tubes are black. One spare tube of each size is provided
in the spare parts kit. Spare tubing is available in packages of 5, and spare
tubes are cut to the proper length at the factory. Be very careful to ensure
that the I.D. of the tube used on the outlet is larger than the I.D. of the
inlet tube. The difference is clear when comparing the two tubes side-by-side.
Reversing the tube sizes on the pump will result in sample water being pumped
through the parts of the system that should only see gas. Also note that the
tube fittings on the inlet side are smaller than those on the outlet side. The
spare parts kit contains 2 of each size.
To replace the sample pump tubing, follow these steps.
- Remove the reagent pickup assembly from the reagent bottle and place it in a container of tap water. Allow the monitor to run for 30 minutes. This will ensure that all the acid is out of the system prior to maintenance.
- Release the front panel latch and tilt the front panel forward. Raise the lever on the fuse holder to turn off power to the chemistry system. Then, return the front panel to normal position.
- Release the pressure plates on each side of the pump. This will allow sample and drain tubes to drain into the overflow. Then disconnect the inlet and outlet tubing connections from both sides of the pump.
- Slide the tube holders out of the pump body and then pull the old tubes from barb end of the luer fittings.
- Place one inlet tube and one outlet tube onto the proper sized fittings. Place the tube holders with the inlet tube onto the left side of the pump and place the tube holders with the outlet tube on the right side.
- Once tube holders are in place, replace the pressure plates on each side of the pump.
Reagent Pump Tube Replacement
Replacement of the reagent pump tubes is similar to the procedure used for the
sample pump. However, the replacement reagent tubes are supplied with the
plastic tube fittings already attached. Do not pull the microbore tubing off
of the fittings as it is difficult to put back on. The reagent tubing
assemblies snap into the tube holders at the middle of the plastic tube
fittings. Observe the position of these assemblies prior to removing the old
ones. One tube assembly is required on the left side of the pump for delivery
of acid into the sample. A second tube assembly is required on the right side
of the pump to remove condensate from the gas sample conditioning module. The
tubes assemblies are the same for both sides of the pump.
Troubleshooting
General
The information included in this section is intended to be used in an attempt
to quickly resolve an operational problem with the system. During any
troubleshooting process, it will save the most time if the operator can first
determine if the problem is related to the analyzer, sensor, or some external
source. Therefore, this section is organized from the approach of excluding
any likely external sources, isolating the analyzer, and finally isolating the
sensor. If these procedures still do not resolve the operational problems, any
results the operator may have noted here will be very helpful when discussing
the problem with the factory technical support group.
External Sources of Problems
To begin this process, review the connections of the system to all external
connections.
- Verify the analyzer is earth grounded. An earth ground connection MUST be present for the shielding systems in the electronics to be active. Grounded conduit provides no earth connection to the plastic enclosure, so an earth ground wiring connection must be made at the power input terminal strip.
- Verify the proper power input is present (115/230 VAC).
- Verify the loads on any 4-20 mA outputs do not exceed the limits in the Instrument Specifications. During troubleshooting, it is many times helpful to disconnect all these outputs and place wire-shorts across the terminals in the instrument to isolate the system and evaluate any problems which may be coming down the analog output connections.
- Do not run sensor cables or analog output wiring in the same conduits as power wiring. If low voltage signal cables must come near power wiring, cross them at 90° to minimize coupling.
- If rigid conduit has been run directly to the Q45 enclosure, check for signs that moisture has followed conduit into the enclosure.
- Check for ground loops. Although the sensor is electrically isolated from the process, high frequency sources of electrical noise may still cause erratic behavior in extreme conditions.
- On systems where relays are in use, check the relay load to verify the load is within the contact rating of the relays. Relay contacts which have been used for higher power AC current loads may become unsuitable for very low signal DC loads later on because a small amount of pitting can form on the contacts. If the load is highly inductive (solenoids, motor starters, large aux relays), note that the contact rating will be de-rated to a lower level. Also, due to the large amount of energy present in circuits driving these types of loads when they are switched on an off, the relay wiring placement can result in electrical interference for other devices. This can be quickly resolved by moving wiring, or by adding very inexpensive snubbers (such As Quencharcs) to the load.
Analyzer Tests
- Disconnect power and completely disconnect all output wiring coming from the analyzer. Remove sensor wiring, relay wiring, and analog output wiring. Re-apply power to the analyzer.
- Using the Simulate feature, check operation of analog outputs and relays with a DMM.
- Check cell drive circuit. With a digital voltmeter (DVM), measure the voltage between the ORANGE and BLUE terminals with the BLUE jumpered to the WHITE on the monitor to verify that the millivolt value is actually 0 mV. (REMOVE JUMPER AFTER VERIFICATION)
- Check TC drive circuit. Place a wire-short between the GREEN and BLACK terminals. With a digital voltmeter (DVM), measure the voltage between the BLACK and RED terminals on the monitor to verify that the TC drive circuit is producing about 4.8-5.1 Vdc open circuit. Remove DVM completely and connect a 100 Ohm resistor from the BLACK to RED terminals. The temperature reading should display approximately 0°C.
Display Messages
The Q46 Series instruments provide a number of diagnostic messages which
indicate problems during normal operation and calibration. These messages
appear as prompts on the secondary line of the display or as items on the
Fault List.
MESSAGE | DESCRIPTION | POSSIBLE CORRECTION | |
---|---|---|---|
Max is 200 | Entry failed, maximum user value allowed is 200. | Reduce |
value to ≤ 200
Min is 200| Entry failed, minimum value allowed is 200.| | Increase
value to ≥ 200
Cal Unstable| Calibration problem, data too unstable to calibrate. Icons
will not stop flashing if data is too unstable. User can bypass by pressing
ENTER.| | Check to see that a stable standard is in use. If sample is stable,
verify that sample from the drain
tube is pH less than 4. Replace sensor is necessary
Out of Range| Input value is outside selected range of the specific list
item being configured.| | Check manual for limits of the function to be
configured.
Locked!| Transmitter security setting is locked.| | Enter security code
to allow modifications to settings.
Unlocked!| Transmitter security has just been unlocked.| | Displayed
just after security code has been entered.
Offset High| The sensor zero offset point is out of the acceptable range
of –999 to +999 nA.| | Check wiring connections to sensor. Allow sensor to
operate powered a minimum of 4 hours prior to first zero cal.
Sensor High| The raw signal from the sensor is too high and out of
instrument range.| | Check wiring connections to sensor. Be sure that the
blue and white wires are not reversed.
Sensor Low| The raw signal from the sensor is too low.| | Check wiring
connections to sensor.
Sulfide High| The sulfide reading is greater than the maximum of the
User-selected range.| | The reading is over operating limits. Set measuring
range to the next highest level.
Cal Fail| Failure of sulfide calibration. FAIL icon will not
extinguish until successful calibration has been
performed, or 30 minutes passes with no keys being pressed.
| | Clean sensor redo zero and span calibration. If still failure, sensor slope may be less than 25% or
greater than 500%. Replace sensor if failure recurs.
EPROM Fail| Internal nonvolatile memory failure| | System failure,
consult factory.
Chcksum Fail| Internal software storage error.| | System failure,
consult factory.
Display Fail| Internal display driver fail.| | System failure, consult
factory.
Range Cal Fail| Failure of factory temperature calibration.| | Consult
factory.
Spare Parts
Part No. Description Q46 Electronic Assembly
- 03-0431-2 Q46S front lid assembly
- 07-0365 100-240 VAC monitor electronics assembly
- 07-0367 100-240 VAC monitor electronics assembly with Profibus
- 03-0409 P/S Assy, 100-240 VAC with 3 relay exp. Board
- 23-0029 Fuse, 630mA, 250V, TR-5
- 38-0072 Terminal block plug, 3 position (relays)
- 38-0073 Terminal block plug, 4 position (outputs)
- 38-0074 Terminal block plug, 3 position (ground)
- 38-0081 Terminal block plug, 3 position (power)
- 38-0084 Terminal block plug, 3 position (power) – VDC version*
NOTE: prior to Dec 2018, VDC (power) Terminal block used the (38-0081)
- 44-0311 Conduit entry plug, red PE
- 44-0274 Cord grip, ½” NPT
Auto-Chem Assembly
- 00-1651 Hydrogen Sulfide Sensor
- 31-0177 Sensor cable with connector, 2 ft.
- 00-1654 RTD Assembly, Pt100
- 03-0428 RTD wiring harness
- 23-0002 Fuse, 2 A
- 03-0420 Thermoelectric cooler with cable
- 03-0421 Sample pump assembly (requires parameter label)
- 34-0392 Sulfide label for sample pump
- 03-0422 Reagent pump assembly
- 36-0051 Sample pump motor, 24RPM
- 36-0049 Reagent pump motor, 6 RPM
- 03-0427 Pump motor wiring harness
- 05-0103 Sample pump inlet tubing, pkg. of 5
- 05-0104 Sample pump drain tubing, pkg. of 5
- 05-0102 Reagent pump tubing assembly, pkg. of 10
- 00-1652 24 VAC power supply, 115 VAC input
- 00-1657 24 VAC power supply, 230 VAC input
- 00-1653 6 VDC power supply, 100-240 VAC Input
- 36-0053 Air pump assembly, 6 VDC
- 55-0014 Rotameter
Part No. Description
- 44-0116 Check valve, 1/8” I.D. tubing
- 45-0299 Air stripping chamber, Acrylic
- 45-0302 Air stripping chamber plug
- 42-0024 Stripping chamber plug o-ring
- 45-0304 Sensor flow block, Acrylic
- 42-0003 Sensor flow block o-ring
- 45-0317 Gas conditioner front block
- 45-0324 Gas conditioner rear block
Tubing Fittings
- 44-0004 Plug, 1/8” NPT
- 44-0036 1/8” NPT x 1/8” I.D. right angle barb fitting
- 44-0090 1/8” NPT x ¼” I.D. right angle barb fitting
- 44-0091 10-32 x 1/8” I.D. barb fitting
- 44-0148 10-32 x 1/8” I.D. right angle barb fitting
- 44-0159 Luer fitting, female, 1/8” barb
- 44-0164 Luer fitting, male, 1/8” barb
- 44-0167 1/8” NPT x 1/8” I.D. barb fitting
- 44-0168 1/8” I.D. tee fitting
- 44-0184 Luer adapter, 1/8” NPT x male luer
- 44-0185 Luer cap, female
- 44-0194 Luer adapter, female, 1/6” I.D. barb
- 44-0253 Luer adapter, female, panel mount, ¼-28 x 1/8” barb
- 44-0336 1/8” I.D. tube “L” fitting
- 44-0337 3/8” NPT x ½” I.D. barb fitting
- 44-0338 3/8” NPT plug
- 44-0348 Luer adapter, 5/32” tubing, sample pump inlet
- 44-0394 Luer adapter, 3/16” tubing, sample pump outlet
Tubing
- 44-0334 Tygon tubing, 1/8” I.D. x ¼” O.D.
- 44-0333 PVC tubing, 1/6” I.D. x 1/8” O.D.
- 03-0419 Reagent Feed Assembly (includes 00-1340)
- 44-0114 Buffer pickup tubing connector
- 47-0022 Reagent Bottle Wall Bracket
PRODUCT WARRANTY
Analytical Technology, Inc. (Manufacturer) warrants to the Customer that if any part(s) of the Manufacturer’s products proves to be defective in materials or workmanship within the earlier of 18 months of the date of shipment or 12 months of the date of start-up, such defective parts will be repaired or replaced free of charge. Inspection and repairs to products thought to be defective within the warranty period will be completed at the Manufacturer’s facilities in Collegeville, PA. Products on which warranty repairs are required shall be shipped freight prepaid to the Manufacturer. The product(s) will be returned freight prepaid and allowed if it is determined by the manufacturer that the part(s) failed due to defective materials or workmanship. This warranty does not cover consumable items, batteries, or wear items subject to periodic replacement including lamps and fuses. Gas sensors, except oxygen sensors, are covered by this warranty, but are subject to inspection for evidence of extended exposure to excessive gas concentrations. Should inspection indicate that sensors have been expended rather than failed prematurely, the warranty shall not apply. The Manufacturer assumes no liability for consequential damages of any kind, and the buyer by acceptance of this equipment will assume all liability for the consequences of its use or misuse by the Customer, his employees, or others. A defect within the meaning of this warranty is any part of any piece of a Manufacturer’s product which shall, when such part is capable of being renewed, repaired, or replaced, operate to condemn such piece of equipment. This warranty is in lieu of all other warranties (including without limiting the generality of the foregoing warranties of merchantability and fitness for a particular purpose), guarantees, obligations or liabilities expressed or implied by the Manufacturer or its representatives and by statute or rule of law. This warranty is void if the Manufacturer’s product(s) has been subject to misuse or abuse, or has not been operated or stored in accordance with instructions, or if the serial number has been removed. Analytical Technology, Inc. makes no other warranty expressed or implied except as stated above
WATER QUALITY MONITORS
- Dissolved Oxygen
- Free Chlorine
- Combined Chlorine
- Total Chlorine
- Residual Chlorine Dioxide
- Potassium Permanganate
- Dissolved Ozone
- pH/ORP
- Conductivity
- Hydrogen Peroxide
- Peracetic Acid
- Dissolved Sulfide
- Residual Sulfite
- Fluoride
- Dissolved Ammonia
- Turbidity
- Suspended Solids
- Sludge Blanket Level
- MetriNet Distribution Monitor
GAS DETECTION PRODUCTS
- NH3 Ammonia
- CO Carbon Monoxide
- H2 Hydrogen
- NO Nitric Oxide
- O2 Oxygen
- CO Cl2 Phosgene
- Br2 Bromine
- Cl2 Chlorine
- ClO2 Chlorine Dioxide
- F2 Fluorine
- I2 Iodine
- HX Acid Gases
- C2H4O Ethylene Oxide
- C2H6O Alcohol
- O3 Ozone
- CH4 Methane (Combustible Gas)
- H2O2 Hydrogen Peroxide
- HCl Hydrogen Chloride
- HCN Hydrogen Cyanide
- HF Hydrogen Fluoride
- H2S Hydrogen Sulfide
- NO2 Nitrogen Dioxide
- NOx Oxides of Nitrogen
- SO2 Sulfur Dioxide
- H2Se Hydrogen Selenide
- B2H6 Diborane
- GeH4 Germane
- AsH3 Arsine
- PH3 Phosphine
- SiH4 Silane
- HCHO Formaldehyde
- C2H4O3 Peracetic Acid
- DMA Dimethylamine
Home Office
Analytical Technology, Inc. 6 Iron Bridge Drive Collegeville, PA 19426
- Phone : 800-959-0299 610-917-0991
- Fax : 610-917-0992
- Email : sales@analyticaltechnology.com
European Office
ATI (UK) Limited Unit 1 & 2 Gatehead Business Park Delph New Road, Delph
Saddleworth OL3 5DE
- Phone : +44 (0)1457-873-318
- Fax : + 44 (0)1457-874-468
- Email : sales@atiuk.com
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>