aidacare A320 Aluminium Bariatric Lifter
IMPORTANT:
Do not use the ASPIRE A320 Bariatric Lifter without having first read this
manual.
OVERVIEW
The ASPIRE A320 Lifter is an ergonomically designed heavy duty mobile patient lifter with a safe working load (SWL) of 320kg. This lifter is a safe care solution for the most demanding institutional and aged care environments. The ASPIRE A320 Lifter allows patients to be transferred from floor positions, beds, chairs and other support surfaces with comfort and safety. The hinged rotating yoke with 4 attachment points generates more efficient patient space allowing a patient to be rotated through 360° at full lift height. This facilitates safe and easy transfers for both attendant and patient. The ASPIRE range of patient lifting solutions features industry-leading LINAK Jumbo Care electronics and have been designed and tested to comply with AS/NZ ISO 10535: 2011.
SAFETY PRECAUTIONS
INTENDED USE OF THE PRODUCT
The ASPIRE A320 Lifter has been designed to lift and transfer human patients
with the assistance of one carer or attendant. All ASPIRE patient lifters
should ONLY be used with Aidacare approved patient slings, which have been
chosen and assessed by a competent trained health professional. For a complete
list of all compliant lifting slings please contact Aidacare on 1300 133 120.
IMPORTANT – only approved slings and accessories are compatible with this
lifter. If you are unsure please contact Aidacare on 1300 133 120.
The correct and appropriate use of the ASPIRE A320 Lifter reduces the risks
associated with manual patient handling. The lifting and transferring of
patients should always be assessed and planned by a trained health
professional.
This is the direction of travel for which this lifter has been designed.
SAFETY CONSIDERATIONS
- ALWAYS ensure that attendants have read this guide before operating this patient lifter.
- ALWAYS use slings that are assessed and considered suitable by a trained healthcare professional.
- ALWAYS ensure the area around the lifter and patient is clear of obstructions.
- ALWAYS minimize the distance over which the patient is being transferred by positioning support surfaces as close as possible.
- DO NOT operate the lifter without training in the safe operation of lifters and slings. IF IN DOUBT DO NOT USE .
- DO NOT use the lifter where any doubt exists regarding the safety of the lifter including but not limited to wear or damage with particular importance to be placed on the yoke and lifter sling attachment points.
- DO NOT use the lifter with slings showing any signs of wear or damage, including fraying.
- DO NOT exceed the 320kg safe working load of the ASPIRE A320 Lifter.
- DO NOT use the lifter as a means of transportation over distances in excess of 3 metres. It is not designed as a transportation device.
INTENDED USE OF THE PRODUCT
WARNING: DO NOT manually lift the boom of the lifters.
How to Use Correctly:
The boom of the lifter is designed to be lifted only by an electric actuator.
Aidacare Lifters and Standers do not pull downwards during lowering. This
safety spline feature is designed to avoid entrapment and crush injuries. This
safety feature means that it is physically possible to raise the lifter boom
by hand on some models with LA31 actuators fitted. NEVER raise the lifter boom
by hand or any other means except under direct drive from the electric
actuator fitted to the lifter. 1300 133 120 Raising the boom can cause it to
become stuck in an elevated position. Under load this can result in patient
free fall and risk catastrophic injury to the patient and/or carer. ALWAYS use
the handset or control unit buttons to raise and lower the boom of the lifter.
This will ensure correct and safe functioning at all times.
PARTS GLOSSARY
- Boom
- Quick Release Yoke Attachment
- Yoke
- Battery
- Control Unit
- Emergency Lowering
- Emergency Stop
- Handset
- Push Handle
- Lift Actuator
- Mast Height Adjustment
- Base Fastening Lever Bolt
- Cover for Leg Actuator
- Leg Spread Actuator
- Rear Castor with Brake
- Front Castor
ASSEMBLY
Unpack
- Remove all parts from carton.
Set up mast
- Three alternative mounting holes are present for adjusting the height of the mast. Select the one most suited to your application.
- Secure mast to base mount using base fastening lever bolt.
Apply thread locker / loctite
- Locate actuator mounting hardware.
- Install actuator hardware. A few drops of medium-strength thread locker / loctite must be applied to the bolt threads.
NOTE: the A320 has threads on both sides of mounting bolt, so thread locker / loctite needs to be applied on both sides
Mount handle
- Mount push handle onto mast in position shown and fasten screws with Allen key.
Attach the quick release yoke
- Pull down quick release lever, engage with attachment point, release lever. Check security latch is closed.
Connecting the actuator
- Connect grey actuator lead into port (1) located on base of the control unit.
ASSEMBLY CONT.
- Connect the other end into the female lead from the actuator.
Connecting the handset
Connect grey handset lead into port located on base of control unit.
Connecting the leg spread actuator
Connect the black actuator lead port (2) on the base of the control unit.
PRELIMINARY CHECKS
After charging the battery these checks should be completed prior to initially using the lifter:
- Check handset functionality is correct.
- Check to see that all indicator lights are functioning.
- Check that the emergency stop button causes the lifter to stop.
- Check emergency lowering function.
- Check that all leads are inserted, fastened and clear of the rear castors.
LIFTER OPERATION
Handset
- Raise / Lower function buttons
- Leg spread function buttons (if applicable)
- Battery charge indicators
- Service indicator
- Press and hold the raise / lower or leg spread (if applicable) function buttons to activate. Release to deactivate. Beware of obstacles at all times to ensure boom and legs can move freely.
- Battery charge indictors will flash when battery reaches maximum recommended discharge of 18 volts, while the lifter is under load.
- Service indictor will illuminate when lifter has reached recommended service interval, either 12 months since last service or 8000 lift cycles. At this time the lifter needs to be serviced for safety and to maintain applicable warranty. A full load test should be completed at least every 12 months by an Aidacare service technician.
LIFTER OPERATION CONT
Emergency Lowering
To activate emergency lowering, rotate the sleeve located at the top of the
boom. This allows the patient to be fully lowered in the event of a fault in
the lifter or discharged batteries.
Charging the battery
The ASPIRE A320 Lifter is supplied with a LINAK external wall mount charger
and is compatible with standard Australian power outlets.
The following steps should be completed to recharge the battery pack correctly:
- Remove the battery pack from the lifter via the top mounted release lever.
- Connect the battery pack to the charger.
Charger will automatically activate the charging sequence and a light will illuminate to show charging has commenced.
CHARGING INDICATOR DISPLAY
CHARGING DOS AND DON’TS
- Charge the battery whenever possible, to maximize battery life. Allowing batteries to fully discharge can reduce the useful life of a battery.
- Always inspect all cables particularly the mains power cable on the charger for any damage; replace if any signs of damage or wear are visible.
- Always stow the handset, charger and mains power cable when transporting the lifter to prevent loss and or damage.
- Always clean the actuators, control box, charger, battery and handset at regular intervals to remove dust/dirt.
NEVER
- Never allow the battery to fully discharge before recharging. Allowing batteries to fully discharge can reduce the useful life of a battery.
Castors
To engage the castor brake place your foot on the brake pedal and apply
sufficient downward force to move it into the lock position. To disengage the
castor brake simply place the point of your toe underneath the brake pedal and
lift-up.
Emergency Stop
To engage the emergency stop press the red button located on the control unit
until it sits flush and remains in position. To disengage the emergency stop
function twist the red button clockwise until it pops out
LIFTING GUIDE
It is important that carers are familiar with manual handling techniques along with any sling instructions prior to completing any patient lifts. All patient’s should have the correct sling and transfer techniques assessed by a qualified health professional. Explaining the lift process to the patient is important to ensure patient comfort and confidence and whenever possible two carers should conduct patient lifts for additional safety.
BED
Patient sling positioning
After rolling the patient on their side, position the sling along the full
length of the back and head (if a head support model is being used).
The patient should be rolled into a supine position so that the sling can be
retrieved and pulled out so that the patient is positioned centrally (see
image)
Attaching the sling to the yoke
The lifter should be positioned with the legs sufficiently under the bed to
allow the boom and yoke to be positioned over the patient’s chest. Once in
place engage the rear castor brakes and attach the loops of the sling to the
hooks on the yoke.
Lifting the patient
Before commencing release the rear castor brakes so that the lifter can adjust
its centre of gravity during the lift. As the loops pull tight check that the
patient is still in a comfortable position and that the head support (if
present) is providing correct support. If the patient sling position remains
correct, the lift can be commenced. Always try to position the two support
surfaces that the lift is taking place between as close together as possible
to avoid moving the patient over distances in the sling.
CHAIR
Patient sling positioning
Ensure that the patient is ready, then slowly lean them forward to allow the
sling to be slid down behind the back until the bottom of the sling is
positioned under the coccyx. The sling leg straps can now be slid underneath
each thigh, remaining careful to avoid creasing or bunching of the fabric.
(Check with a health professional if leg straps should be crossed over for
additional support or separated for enhanced accessibility for hygiene)
Attaching the sling to the yoke
Ensure that the lifter legs are open to the maximum width and slowly position
it as close to the front of the chair as possible – be careful to keep the
boom and yoke clear of the patient during this process. Once the yoke is at
chest height on the patient the sling loops can be attached For chair lifts
the rear brake castors should not be used to ensure the lifter can locate its
centre of gravity easily
Lifting the patient
Raise the yoke until the loops of the sling pull tight. Check that the patient
is still comfortably positioned in the sling and that the head support is in
place
(if a head support model is being used). Lift the patient only to a height
needed to successfully transfer to the desired support surface. Always try to
position the two support surfaces that the lift is taking place between as
close together as possible to avoid moving the patient over distances in the
sling.
FLOOR
Generally, a floor lift is only required in the event of a fall or emergency.
Therefore it is important that a medical professional has assessed the patient
and given the all clear for a lift to be carried out.
Patient Sling Positioning
After rolling the patient on their side position the sling along the full
length of the back and head (if a head support model is being used). The
patient should be rolled into a supine position so that the sling can be
retrieved and pulled out so that the patient is positioned centrally (see
image).
Attaching the sling to the yoke
The lifter should be positioned with the legs open to full width to ensure
maximum area for the patient to be lifted. One leg should be positioned
adjacent to the patient’s head and the other underneath the knees of the
patient once their legs have been raised (as shown in the middle image on this
page). The rear castors should be locked. It is important to continually
communicate with a patient during floor lifts as they may often be in shock or
disoriented and need reassurance.
Lifting the patient
When carrying out a floor lift, the castors should be locked throughout the
lifting procedure. As you begin your lift and the loops pull tight, check that
the patient is still in a comfortable position and that the head support (if
present) is providing correct support. If the patient sling position remains
correct, the lift can be commenced. Lift the patient only to a height needed
to successfully transfer to the desired support surface. Always try to
position the two support surfaces that the lift is taking place between as
close together as possible to avoid moving the patient over distances in the
sling.
SLING COMPATIBILITY
AIDACARE has a range of ASPIRE slings designed for use with the ASPIRE lifters. This lifter may also be suitable for use with other brands of slings provided that the design is compatible with a hook yoke/spreader bar design. To confirm if a sling is compatible with any of the ASPIRE range of lifters contact AIDACARE on 1300 133 120.
CARE & MAINTENANCE
To prolong the working life of the lifter and for trouble free operation it is recommended that periodic inspections should be performed routinely by a person who is suitably qualified and well acquainted with the design, use and care of patient lifters. If inspection reveals that the safety of the lifter is jeopardized in any way the lifter should be IMMEDIATELY WITHDRAWN FROM SERVICE, tagged as such and the owner notified.
MAINTENANCE GUIDE & CHECK LIST
- Check castors are fastened securely to the frame
- Castors should run and swivel smoothly – if hair or other fibres have built up within the castor this should be removed
- Check that the brakes on the rear castors function normally
- Clean surfaces with a pH neutral detergent – chorine and phenol based detergents should not be used as these may damage lifter components
- All moving parts and pivot points should be checked for fatigue and lubricated if necessary
- All electrical components should be checked for normal function including battery and battery charger
- The actuator shaft should only be wiped with a cloth.
DO NOT lubricate, contact Aidacare maintenance 1300 133 120
AS A MINIMUM A FULL SERVICE SHOULD BE COMPLETED EVERY 6 MONTHS AND THE LOAD TEST EVERY 12 MONTHS TO MAINTAIN WARRANTY. ELECTRICAL TEST & TAG – WHERE LIFTERS ARE IN USE IN A COMMERCIAL ENVIRONMENT THEN THE TESTING REGIME APPLIED SHOULD ALIGN WITH THE RISK ASSESSMENT MADE BY THE OWNER FOR SAFETY AND OTHER COMPLIANCE REQUIREMENTS. TO MAINTAIN AUSTRALIAN STANDARDS COMPLIANCE UNDER AS ISO 10535-2011 A FULL SWL TEST MUST BE COMPLETED EVERY 12 MONTHS.
WARRANTY & SERVICE
LIFTER AND ELECTRONICS
AIDACARE warrants its products to be free from defects in materials and
workmanship under normal use and service and will within the period stated
below, from the date of purchase, repair or replace without cost to the
original customer, any part assembly or portion thereof which shall be
returned to AIDACARE and from OUR inspection shows to be defective.
- Patient Lifter Frame – 5 years
- LINAK Actuator and LINAK Controller – 3 years (excluding Battery Pack)
- LINAK Battery Pack – 12 month manufacturer’s warranty subject to stringent fair wear and tear clauses (please contact AIDACARE for a copy of this information)
WARRANTY IS CONTINGENT UPON ANNUAL SERVICING BEING COMPLETED FOR THE LENGTH OF
THE WARRANTY PERIOD.
AIDACARE cannot be held responsible for any personal injury, damage to the
lifter or damage to property as a result of the improper or unsafe use of the
product. No warranty claim shall apply where the product or any part thereof
has been modified, varied, altered or damaged either accidentally or through
improper or negligent use. No warranty claim shall apply where the lifter is
repaired or serviced by parties other than those accredited and or approved by
Aidacare. Warranty does not extend to items or components which may require
replacement due to normal wear and tear (e.g. castors, plastic molded
components, foam padding (where fitted) and painted surfaces) for which
AIDACARE nor its distributors can be held responsible.
BATTERIES
Batteries carry a limited warranty from the original manufacturer which is
subject to a stringent wear and tear clause. Battery faults due to defect in
original manufacture typically become apparent within the first two months of
use. Any gradual deterioration in performance after this period is typically
associated with fair wear and tear, misuse and accidental damage, and as such
is not covered by the LINAK manufacturer’s warranty. ALL WARRANTIES MADE ARE
BACK TO BASE WARRANTIES MEANING THAT ANY PARTS OR WHOLE LIFTERS REQUIRING
REPAIR NEED TO BE RETURNED TO THE NEAREST AIDACARE SERVICE CENTRE AT THE
CUSTOMER’S EXPENSE – PLEASE CONTACT AIDACARE FOR A CURRENT LIST OF SERVICE
CENTRE LOCATIONS.
SPARE PARTS
SPARE PARTS LIST AVAILABLE UPON REQUEST FROM AIDACARE ON 1300 130 122 or
www.aidacarelifters.com.au
TROUBLESHOOTING
PROBLEM POTENTIAL FAULT | ACTION |
---|---|
Warning alarm can be heard in control unit | Battery is discharged Battery may |
be faulty | Recharge battery Replace battery |
Power LED on charger does not illuminate | Not connected to power Fuse blown |
Faulty Charger
| Check plugged in to power point and switched on Replace fuse
Repair or replace charger
Actuator not running/no click heard from control unit when buttons pressed|
Faulty control unit Faulty handset| Repair or replace control unit
Actuator not running but control unit clocks when buttons pressed| Actuator
plug not inserted fully into control unit Faulty actuator
Faulty control unit
| Check plugs are completely inserted into control unit ports Repair or
replace actuator and or control unit
Actuator not lifting properly or running slowly| Discharged or faulty battery
Weight in excess of safe working load (SWL)
Faulty actuator
| Recharge or replace battery Repair or replace actuator
No clicking in control box when buttons pressed| Emergency stop button pressed
Battery completely discharged Battery fuse faulty
Faulty control unit
| Release emergency stop button Recharge battery
Repair or replace control unit
Actuator only works in one direction| Faulty handset Faulty control unit Emergency lower release activated| Repair or replace handset Repair or replace control unit Check emergency actuator release is reset
WARNING: TROUBLESHOOTING SHOULD ONLY BE CARRIED OUT BY AIDACARE AUTHORISED SERVICE TECHNICIANS WHO ARE TRAINED IN SERVICING THE ELECTRICAL AND MECHANICAL FEATURES OF THE PATIENT LIFTER. The following is a guideline only to general troubleshooting. Before carrying out any troubleshooting:
- CHECK TO ENSURE the RED emergency stop button is NOT pushed in.
- CHECK TO ENSURE maintenance procedures have been followed.
- CHECK TO ENSURE all lead connections are secure.
- CHECK TO ENSURE the battery is fully charged.
PRODUCT SPECIFICATIONS
Product Code | LSS390630 |
---|---|
Safe Working Load (SWL) | 320kg |
Base Length | 1500mm |
External Base Width | 765-1190mm |
Internal Base Width | 640-1065mm |
Base Width Adjustment | Electric |
Base Height / Under Bed Clearance | 155mm |
Minimum Height of Spreader Bar ** | 540-640mm |
Maximum Height of Spreader Bar** | 1840-1940mm |
Lifter Weight | 63kg |
Warranty | 3yrs |
Under Base Height | 155mm |
Maximum Overall Height | 2155mm |
Front Castor Diameter | 100mm |
Rear Locking Castor Diameter | 125mm |
Turning Diameter | 1595mm |
Weight Of Heaviest Part (Base Assembly) | 38.9kg |
Handle Height | 1200 – 1300mm |
Sound Level | <40Db@ 1m |
Electrical Approval
| EN 60601-1-2:2007; EN 61000-3-2:2006;
EN 6100-3-3:2008
Box Contains: Base with 4 castor wheels and LINAK electric leg opening actuator, mast, and boom with yoke, LINAK handset, LINAK battery charger, LINAK battery, LINAK controller, push handle. Screws, bolts, nuts, washers, plastic caps, and tool kit.
- Three mast height settings are possible hence the range of minimum and maximum
SPECIFICATIONS (CONT.)
SERVICE LOG BOOK
YEAR 1
- Date:
- Service Type:__
- Condition Report:_
- Inspected by:_____
- Signature:
- Action taken