bioland G-425-3 Blood Glucose Meter User Manual

June 9, 2024
bioland

Easy Trak II
Blood Glucose Monitoring System
Blood Glucose Meter
Model: G-425-3
User Manualbioland G-425-3 Blood Glucose Meter -

G-425-3 Blood Glucose Meter

Dear Blood Glucose Monitoring System Owner:
Thank you for choosing this system!
This manual contains important information you must know about the system. Please read it thoroughly and carefully.
The system provides No Code function which means that you don’t need to calibrate your meter, making it easier for the user to monitor blood glucose at home.

IMPORTANT SAFETY INSTRUCTIONS

  • The meter and lancing device are for single patient use. Do not share them with anyone including other family members! Do not use on multiple patients!
  • All parts of the kit are considered biohazardous and can potentially transmit infectious diseases, even after you have performed cleaning and disinfection.
  • Changes or modifications not expressly approved by the party responsible for compliance could void the users authority to operate the equipment.
  • If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
    — Reorient or relocate the receiving antenna.
    — Increase the separation between the equipment and receiver.
    — Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
    — Consult the dealer or an experienced radio/TV technician for help.

READ THIS BEFORE USING
DANGER – Misuse of this electrical device can cause electrocution, burns, fire and other HAZARDS.
Basic safety precautions should always be taken, including all those listed below.
Close supervision is necessary when equipment is used by, on, or near children, handicapped persons or invalids.

The following basic safety notice should always be taken:

  1. Disregard of these notices will void warranty.
  2. Do not place the device in liquid, nor put it where it could fall into liquid.
  3. Use the device only for the intended use described in this manual.
  4. Do not use adjunct which are not supplied by the producer.
  5. Do not make the device come into contact with surfaces which are too hot to touch.
  6. Do not use the device where aerosol sprays are being used.
  7. Do not use the device if it is not working properly, or if it has been damaged.
  8. Read all the manual, and practice the test, and then use the product to test your blood glucose. Do all quality control checks as directed and consult with a diabetes expert.
  9. Do not use the expired blood glucose test strips and sterile lancet.
    KEEP THE INSTRUCTIONS IN A SAFE PLACE

BEFORE YOU BEGIN

Installing the battery

  1. Open the battery cover on the back of meter.
  2. If replacing the batteries, please remove the used batteries.
  3. Insert 2 new (AAA) batteries with correct +/- direction.
  4. Close the battery cover.
    **** Environment Protection – Waste electrical products should not be disposed of with household waste. Please recycle where facilities exist. Check with your local Authority or retailer for recycling advice.

ABOUT YOUR SYSTEM

INTRODUCTION TO THE SYSTEM
Intended Use

  • The system is intended for use outside the body (in vitro diagnostic use). It is for single-patient use. It should be used only for testing glucose (sugar) and only with fresh capillary whole blood samples taken from the finger and the alternative sites including the palm, the forearm. The system is intended for use at home. It should not be used for the diagnosis of diabetes or for the testing of newborns.
  • The System is to be used by a single person and should not be shared.
  • The System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing should be done only during steady—state times (when glucose is not changing rapidly).
  • The “Easy Trak II” Test Strips are for use with the Easy Trak II Meter to quantitatively measure glucose in fresh capillary whole blood samples drawn from the fingertips, palm and forearm.
  • The control solutions is intended for use with Easy Trak II System, to check that the glucose meter and test strips are working properly and that you are doing the test correctly. These solutions contain a known range of glucose, as indicated on the strips vial bottles.

AST in this system can be used only during steady-state blood glucose conditions described in the section of “About AST”.
Test principle
The test is based on the measurement of electrical current generated by the reaction of glucose with the reagent of the strip.
Themeter measures the current and displays the  corresponding blood glucose level.
The strength of the current produced by the reaction depends on the amount of glucose in the blood sample.

ABOUT ALTERNATIVE SITE TESTING (AST)

Important: There are limitations for doing AST.
Please consult your healthcare professional before you do AST.
What is AST ?
Alternative site testing (AST) means that people use parts of the body other than the fingertips to check their blood glucose levels.
This system allows you to test on the palm, the forearm with the equivalent results to fingertips testing.bioland G-425-3 Blood Glucose Meter -
Alternative site testing

What’s the advantage?
Fingertips feel pain more readily because they are full of nerve endings (receptors). At other body sites, since nerve endings are not so condensed, you will not feel as much pain as at the fingertip.
When to use AST?
Food , medication , illness, stress and exercise can affect blood glucose levels. Capillary blood at fingertip reflects these changes faster than capillary blood at other sites. Therefore when testing blood glucose during or immediately after meal, physical exercise, or any other event, take a blood sample from your finger only. You do AST ONLY in the following instances:

  • In a pre-meal or fasting state (more than 2 hours since the last meal).
  • Two hours or more after taking insulin.
  • Two hours or more after exercising.

NOTE:

  • Alternative site testing(AST) should not be used to calibrate continuous glucose monitoring systems (CGMs).
  • Results from alternative site testing should not be used in insulin doses calculations.

Do NOT use AST if:

  • You think your blood glucose is low.
  • You are unaware of hypoglycemia.
  • You are testing for hyperglycemia.
  • Your AST results do not match the way you feel.
  • Your routine glucose results are often fluctuating.

CONTENT

  • User Manual
  • A Meter

If any items are missing from your kit or opened prior to use, please contact local customer services or place of purchase for assistance.

Meter Screen Display Message

bioland G-425-3 Blood Glucose Meter - assistance

  1. BLOOD DROP SYMBOL
    Flashes when it is ready to apply the blood sample

  2. TEST RESULT AREA
    Displays glucose results

  3. LOW BATTERY SYMBOL
    Appears when the battery power is low

  4. MEASUREMENT UNIT
    Appears with the test result in mg/dL

  5. DAY AVERAGE
    Indicate that the displayed test result is an average value

  6. CTL SYMBOL
    Appears when doing a control test and indicates that the result won’t be stored in the memory.

Blood Glucose Meter

  1. STRIP PORT
    Where you insert the test strip. The meter will turn on automatically

  2. LCD DISPLAY
    It guides you through the test using symbols and simple messages

  3. STRIP EJECTOR
    Slide forward to eject the test strip after test

  4. “M” BUTTON
    It is used to turn on the meter, enter the memory mode

Test Strip
Each strip can be used only once. The test strip consists of the following parts:

bioland G-425-3 Blood Glucose Meter - Absorbent Hole

Attention!
Test results might be wrong if the contact Bar is not fully inserted into the test slot.
bioland G-425-3 Blood Glucose Meter - contactThe front side of the test strip should face up when inserting the test strip.

SETTING THE METER AND DELETING THE MEMORY

  1. Memory Function: Short press “M” button, then device will appear the most recent test result after full display for 1 Sec.;Short press “M” button to appear the next most recent test result, you can use such way to review all 20 memories.
  2. Deleting the memory Press and hold “M” button for 5 seconds, device will appear “Del”, release “M” button, “Del” will flash, press “M” button within 5 seconds to delete all memories, otherwise device will turn off.

THE TWO MEASURING MODES

The meter provides you with two modes for measuring, General and CTL. The CTL mode should be used when a control solution is being tested. See the table below.

Modes Use when
General tests (not display) any time of day without regard to time since last

meal
CTL| testing with the control solution

You can switch between each mode by:

  1. Start with the meter turned off. Insert a test strip to turn on the meter, the screen will dis play a flashing blood drop.
  2. Press “M” to select General or CTL mode.

bioland G-425-3 Blood Glucose Meter - select Genera

TESTING WITH BLOOD SAMPLE

CAUTION: To reduce the chances of infection:

  • Choose a clean, dry work surface.
  • Never share a lancing device and lancet with another person.
  • Always use a new and sterile lancet. Lancets are for single use only.
  • Always use a new test strip. Test strips are for single use only.
  • Avoid getting lotion. oils. dirt or debris in or on the lancet and lancing device.
    Note: Before testing, please read the following steps thoroughly and carefully, please relax and do the test step by step. Simplif-ying or changing the test procedure may produce inaccurate results.

Overview of a generic Lancig Device

A.Device Cap
B.Handle
C.Needle Cap
D.Needle Holder
E.Device Body| F.Connecting Collar
G.Depth Adjuster
H.Release Button
I .Ejection Button
---|---

On the lancets and the lancing device.
If your lancing device differs from the one shown here, please refer to the manufacturer’s manual to ensure proper usage.

NOTE
To reduce the chance of infection:

  • Never share a lancet or the lancing device.
  • Always use a new, sterile lancet. Lancets are for single use only.
  • Avoid getting hand lotion, oils, dirt, or debris when handling the lancets.

Setting Up the Lancing Device

1. Screw the device cap.| 2. Put the needle into the needle holder until the bottom.
---|---
|
3. Screw the needle cap and put it in the recycling dustbin.| 4. Cover the cap.
|
5. Adjust the proper level (usually level 3)| 6. Aiming at the sterilized part for blood-taking, press the blue release key to finish bllodtaking.
|
7. Screw the device cap.| 8. Push the ejection button and discard the used lancet in appropriate container
|
9. Cover the device cap.|

Preparing the Puncture Site
Stimulating blood perfusion by rubbing the puncture site prior to blood extraction has a significant influence on the glucose value obtained. Blood from a site that has not been rubbed exhibits a measurably different glucose concentration than blood from the finger. When the puncture site was rubbed prior to blood extraction, the difference was significantly reduced.

Please follow the suggestions below before obtaining a drop of blood:

  • Rub the puncture site for about 20 seconds before penetration.
  • Select the puncture site either at the fingertips or another body part (please see section “Alternative Site Testing” (AST) on how to select the appropriate sites).
  • Clean the puncture site using cotton moistened with 75% alcohol and let it air dry.

Fingertip testing
Hold the lancing device firmly against the side of your finger. Press the release button. You will hear a click, indicating that the puncture is complete.

NOTE

  • Choose a different spot each time you test. Repeated punctures at the same spot may cause soreness and calluses.
  • Please consult your health care professional before you begin AST.
  • It is recommended to discard the first drop of blood as it might contain tissue fluid, which may affect the test result.

Performing a Blood Glucose Test
Step 1.
Wash hands with soap and warm water.
Step 2.
Remove a test strip from its vial. With clean, dry hands, you may touch the test strip anywhere on its surface. Do not bend, cut or modify the test strips in any way. Remove the test strip from the vial and use it instantly.
Step 3.
Insert the test strip into the meter’s test port and the meter is turned on.
NOTE: Be sure to use “ G-425-3S ” test strips, other test strips may not produce accurate results.
Step 4
When the blood drop symbol flashes on the screen, you may select the appropriate measurement mode, for selecting the measurement mode of Control(CTL) or General, please refer to the “THE TWO MEASURING MODES” section. Step 5
Apply your blood to the absorbent hole of the test strip.

Step 6
As soon as enough blood has filled the confirmation window of the test strip, you will hear a beep letting you know the test has begun.
A countdown of 5 seconds starts.

Note: make sure the confirmation window of the Test Strip is completely filled with your blood sample.
Step 7
Your blood glucose level, along date and time appears on the display. Blood glucose results are automatically stored in the memory.

NOTE
The results obtained from this meter are plasma-calibrated. This helps you and your physician or other qualified healthcare provider to compare your meter results with laboratory tests.
Reference to the instructions given by your physician or other qualified healthcare provider, do not deviate from these instructions on the basis of the result without first consulting your physician.
Step 8
Eject the used test strip and remove the lancet. To eject the test strip, point the strip at a disposal container for sharp objects. The meter will switch itself off automatically after the test strip is ejected.
Please wash hands thoroughly with soap and water after handling the meter, lancing device, or test strips.

Reference value:

Time of day People without diabetes
Before breakfast (fasting) <100 mg/dL
Two hours after meals <140 mg/dL

Source: American Diabetes Association. Standards of Medical Care in Diabetes-2012.Diabetes Care 2012,35(Suppl.1)S11-S63.
Please work with your doctor to determine a target range that works best for you.

Comparing Meter and Laboratory Results
The meter provides you with plasma-equivalent results. The result you obtain from your meter may differ somewhat from your laboratory result due to normal variation. Meter results can be affected by factors and conditions that do not affect laboratory results in the same way. To make an accurate comparison between meter and laboratory results, follow the guidelines below.

Before going to the lab:

  • Do a control solution test to make sure that the meter is working properly.
  • Fast for at least eight hours and then do comparison tests, if possible.
  • Take your meter with you to the lab.

While staying at the lab:
Make sure that the samples for both tests test are taken and tested within 15 minutes of each other.

  • Wash your hands and then obtain a blood sample
  • Never use your meter with blood that has been collected in a gray top test tube.
  • Use fresh capillary blood only.

CONTROL SOLUTION TEST

Material needed:

  • Glucose Meter
  • Test strips
  • TaiDoc Control solution

When to do a control solution test:

  • When you suspect that the meter or test strips are not working properly.
  • If your blood glucose test results are not consistent with how you feel.
  • After the test strips are exposed to extreme conditions.
  • After dropping the meter.
    The control solution is to check that the meter and the test strips can work together as a system and work correctly.

Step 1
Insert the test strip to turn on the meter. Insert the test strip into the meter. Wait for the meter to display the drop symbol.
Step 2
You may press “M” button to select CTL mode, for selecting the CTL mode, please refer to the “THE TWO MEASURING MODES” section. Step 3
Apply the control solution
Shake the control solution vial thoroughly before use.
Squeeze out a drop and wipe it off, then squeeze another drop and place it on the tip of the vial cap.
Hold the meter to move the absorbent hole of test strip so it touches the drop. Once the confirmation window fills completely, the meter will begin counting down.
To avoid contaminating the control solution, do not directlyapply control solution onto a strip.
Step 4
The result of the control solutions test appears on the display.
The result will not be stored under CTL mode.
The result of the control solutions test should be within the range printed on the test strip vial label. If the test results falls outside the specified range printed on test strip vial label, repeat the test.

Results falling outside the specified range may be caused by:

  • Error in performing the test
  • Expired or contaminated Control Solution
  • Expired or contaminated Test Strip
  • Malfunction of your Meter

If you continue to get control solution test results that fall outside of the range printed on the vial, the system may not be working properly, Do not use the meter. Please contact your customer support immediately

NOTE:
Do not use the expired control solution.
Only use the TaiDoc control solution, you can purchase this product through visits the website: www.taidoc.com

CARE AND STORAGE FOR YOUR SYSTEM

  • Handle the meter with care. Dropping or throwing the meter may cause damage to the device.
  • Don’t expose the Blood Glucose Monitoring System, Test strip, and Control Solution, to extreme conditions, such as high humidity, heat, freezing cold or dust. The torage conditions for the test strips are 4°C~30°C, Humidity 10~80%. Storage conditions for the meter are -20°C~55°C, Humidity ≤93%.
  • Always wash your hands with soap and water and rinse and dry completely before handling the Blood Glucose Monitoring system and Test strip.
  • Use CaviWipes (see below) to clean the Blood Glucose Monit oring System.
  • Discard the used lancet into a container for sharp objects.

Cleaning and Disinfection
The cleaning and disinfection is absolutely necessary for the test procedure, because cleaning can insure the meter works well (for example, display will be clear to see after cleaning); and disinfection can avoid the infection to you or to the other people, and the cross-infection. We strongly warn that the device should be cleaned and disinfected following use on every patient. And we suggest you use the following product: CaviWipes (EPA registration number :
46781-8). You can purchase this product according to the following information: visit the web site: www.metrex.com or contact Metrex at 800-841-1428 for product or technical information. CaviWipes are also available at www.amazon.com or via a Google search.

The meter for single-patient use and the lancing devices may be cleaned 5-6 times per day (Test 5-6 times every day), so the meter will be cleaned and disinfect about 11000 cycles that include cleanings plus disinfections (1 cycle = 1 cleaning + 1 disinfection) during its 5 years life cycle.

  1. Before the test, the user should clean and wash hands. If you are a caregiver, (not the patient), the please use disposable gloves.
  2. After test, take one wipe to clean the meter’s housing, buttons, lens, mark, usb cover and support block, including its front and back for 10 seconds for cleaning until there is no soil on the surface. Then disinfect the meter’s housing, buttons, lens, mark, usb cover and support block use another wipe for 2 min, keep the surface wet for 2 min.
  3. Take off the glove; discard carefully according to your health care provider’s instruction.

Note:

  1. Each disinfection step requires a pre-cleaning step.
  2. If the meter is being operated by a second person who is pro viding testing assistance to the user, the meter and lancing device shall be cleaned and disinfected first.
  3. The users should wash hands thoroughly with soap and water after handling the meter, lancing device, or test strips.

METER SPECIFICATION

  1. Model: G-425-3 (with “ G-425-3S ” test strips and TaiDoc control solution)
  2. Machine size: 91mm (L)×57mm (W)×22mm(H)
  3. Weight: 63g (including batteries)
  4. Measuring method: Amperometric technology using glucose oxidase
  5. Result range: 20 mg/dL ~600 mg/dL
  6. Glucose Units: mg/dL
  7. Memory : 20 blood glucose tests
  8. Display: LCD display
  9. Power source: DC 3V (AAA×2 alkaline batteries)
  10. Battery Life: Approx. 500 normal tests
  11. Storage Condition: Blood glucose Monitoring system (meter and test strips): 4 ℃~30 ℃ , Humidity 10~80%
  12. Transport and storage Condition (meter): -20 ℃~55 ℃ , Humidity≤93%
  13. Operating conditions: 10 ℃~40 ℃ , Humidity≤80%
  14. Blood Source: Fresh capillary whole blood
  15. Blood Volume: Minimum 0.7 micro liter.
  16. Use life: five years

Limitations of use:
The blood Monitoring System is not intended for use on neo nates.
※ The blood Monitoring System is not intended for use on artery blood, neonates’ serum and plasma.
※ The Blood Glucose Meter should be used with “ G-425-3S ” test strip.
※ The system can be used up to an altitude of 10744 feet.
※ The following substances at levels greater than normal or therapeutic levels may cause significant interference (affect the result by greater than 10%) resulting in an
inaccurate
※ result: ascorbic acid, uric acid, hemoglobin,acetaminophen, Dopamine, L-dopa, L-dopa and Tolbutamide etc. These sub stances do not affect test results in normal concentration but may affect test results in high concentration. Do not use hae molysis sample, icterus sample or high lipemia samples.
※ The blood icodextrin concentration higher than 13mg/dL would interference to the test result.
※ The following compounds, when determined to in excess of their limitation, may produce inaccurate results:

Compound Limitation
Acetaminophen 8.0mg/dL
Ascorbic Acid 5.0mg/dL
Bilirubin 90mg/dL
Cholesterol 500mg/dL
Creatinine 5.0mg/dL
Dopamine 2.0mg/dL
EDTA 360mg/dL
Galactose 900mg/dL
Gentisic Acid 5.0mg/dL
Glutathione 53mg/dL
Haemoglobin 500mg/dL
Heparin E3000U/dL
Lbuprofen 50mg/dL
Icodextrin 13mg/dL
L-dopa Omg/dL
Maltose 900mg/dL
Methyldopa 3.0mg/dL
Pralidoxime Iodide 25mg/dL
Salicylate 60mg/dL
Tolazamide 100mg/dL
Tolbutamide c400mg/dL
Triglycerides 2000mg/dL
Uric Acid E3.0mg/dL
Xylose 100mg/dL

※ Patients undergoing oxygen therapy may yield falsely lower results.
※ Not for use for patients in a hyperglycemic-hyperosmolar state, with or without ketosis.
※ Not for use on critically ill patients.
※ Not to be used for patients who are dehydrated, hypertensive, hypotensive or in shock.
※ Very low (less than 20%) or very high (more than 60%) red blood cell count (hematocrit) can lead to incorrect test results.
If you do not know your hemarocrit level, please consult your health care provider.

DISPLAY MESSAGE

MESSAGE WHAT IT MEANS ACTION

bioland G-425-3 Blood Glucose Meter - icon| Blood glucose level is lower than 20mg/dL| The message indicates very low blood sugar. You should consult with your healthcare professional.
| Blood glucose level is higher than 600mg/dL| This indicates sever hyperglycemia (high blood glucose).
You should seek immediate medical assistance.
| Meter is ready to test control solution under the CTL state.| Please place a drop of control solution onto the test well of test strip.
| Battery power is low.| Please change the battery.
| Use a used test strip.| Please use a new strip and re-test.
| Error massage indicates that you may remove the strip after applying blood to the absorbent hole.| Try again with a new test strip.
| Pull out strip during the testing.| Re-test with a new test strip.
| The meter self-inspection is failure.| Turn on the meter after calibration again.
| Environmental temperature is lower than 10 (50 ) or higher
than 40 (104 )| Please test at an envi-ronmental temperature within 10-40 °C ( 50- 104 °F) operation range.

Problems

Problems Causes Solutions

Display remains blank after the test strip has
been inserted  into the Meter.| 1. Battery may be installed improperly.
2. Battery power is too low for use.
3. Too much time has passed between inserting the Test Strip and perfor ming the test.
4. Test Strip has not been fully inserted into the Meter.| 1.Install the Battery correctly.
2.Replace with new Battery.
3.Reinsert the  Test Strip into the Meter.
4. Reinsert the Test Strip into the Meter.
Test results are incon sistent or Control Solution test results are not within the specified range.| 1. Not enough sample in the Test Strip.
2. Test Strip or Control Solution has expired.
3. Test Strip has been damaged due to heat or humidity so that sample cannot be applied or the speed of applying  sample is too slow.
4. System is not per forming due to the environment being above or below room temperature.| 1. Redo test with new test strip and make sure that enough sample has been added.
2. Redo test with new test strip.
3. Perform a Control Solution test using a new Test Strip. If results are still out of range, replace with  new vial of Test Strips.
4. Bring System to a room temperature environment and wait approximately 30 minutes before per orming a new test.
The Meter co-untdown did not start.| Test Strip has not been inserted correctly.| Use a new Test Strip and redo the test.

Performance Characteristics

Mean (mg/dL) 44.7 102 138 210 330
SD (mg/dL) 3.0 4.4 6.2 7.2 13.1
CV (%) 4.3 4.5 3.4 4.0

Intermediate precision

Mean (mg/dL) 47.0 108 348
SD (mg/dL) 3.30 4.2 9.3
CV (%) 3.9 2.7

System Accuracy
For glucose concentrations<100mg/dL

Within±5mg/dL Within±10mg/dL Within±15mg/dL
104/180 (57.78%) 168/180 (93.33%) 178/180 (98.89%)

For glucose concentrations≥100mg/dL

Within±5% Within±10% Within±15%
195/420 (46.43%) 327/420 (77.86%) 411/420 (97.86%)

For glucose concentrations between 47mg/dL and 548mg/dL

Within±15 mg/dL or ±15%
589/600 (98.17%)

Traceability of glucose monitoring system
The results of G-425-3/G-425-3S glucose monitoring system were compared to parallel results obtained on YSI-2300, which is the manufacturer’s standard measurement procedure.
User Performance
A study evaluating glucose values from fingertip, palm and Forearm capillary blood samples obtained by 100 lay persons showed the following results:Fingertip 100% , palm 98.1% and forearm 98.1% within ±15mg/dL of the medical laboratory values at glucose concentrations below 100 mg/dL, and fingertip 98.6% , palm 97.3% and forearm 97.3% within ±15% of the medical laboratory values at glucose concentrations at or above 100mg/dL.
The device was approved and met EN 1SO 15197: 2015.

ELECTRIICAL AND SAFETY STANDARDS

This meter complies with CISPR 11: Class B (Radiated Only).
Emissions of the energy used are low and not likely to cause interference in nearby electronic equipment. The meter has been tested for immunity to Level 3 electrostatic discharge as specified in IEC 61326.This meter complies with immunity to radio frequency interference over the frequency range 80MHz to 2.5GHz at 3V/m as specified in IEC 61326-1 or 61326-2.
The meter meets the requirements for immunity to electrical interference at the frequency range and test level specified in international standard ISO 15197.
CAUTION: Do not use the equipment where aerosol sprays are being used, or when oxygen is being administered.

SYMBOLS INFORMATION

Symbol Referent
Use by

Consult instructions
for use
CAUTION
Keep away from
sunlight
Keep dry
Temperature limitation
Authorised representative in the european community
| Date of manufacture
Batch code
In vitro diagnostic
medical device
Catalogue number
Serial number
Control
Manufacturer

Distributor:
HOME AIDE DIAGNOSTICS, INC.
1072 S. Powerline Rd.
Deerfield Beach, FL 33442
www.homeaide.us
1-800-915-0116

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals