EG1 Electric Surgical Drill
User Manual
ANSPACH™ EG1 WIRELESS HAND CONTROL SYSTEM
General Information
Device Description
The eG1 Wireless Hand Control System consists of the non-sterile reusable
Receiver for Wireless Hand Control, EG1A (RECEIVER-HC), and the disposable
Wireless Hand Control, EG1A (WIRELESS-HC) which is provided in a sterile
package configuration. The Wireless Hand Control, EG1A communicates wirelessly
through the Receiver for
Wireless Hand Control, EG1A to allow the user to regulate the speed of the eG1
High-Speed System.
Intended Use
The eG1 High-Speed System is intended for cutting and shaping bone
including the cranium and spine.
Indications for Use
The eG1 High-Speed System is indicated for cutting and shaping bone including
the cranium and spine.
Contraindications
The eG1 High-Speed System does not have any known product-specific
contraindications.
Potential Adverse Events
As with all major surgical procedures, risks, side effects, and adverse events
can occur. The following potential adverse events can occur:
- Infection
- Adverse Tissue Reaction
- Injury to Patient
- Injury to User
Intended User
This device is intended to be used by qualified health care professionals e.g.
surgeons, physicians, operating room staff, and individuals involved in the
preparation of the device.
Patient Target Group
The eG1 Wireless Hand Control System is for use with patients undergoing
surgery where the cutting and shaping of bone, including bones of the cranium
and spine, is desired.
Catalog number & Compatibility
Catalog Number | Consoles | Handpiece |
---|---|---|
SC3001 | SC3002 | EG1A |
RECEIVER-I-IC | • | • |
WIRELESS-HC |
Note: DePuy Synthes has not tested compatibility with devices provided by
other manufacturers and assumes no liability in such instances.
Cleaning and Sterilization
Receiver for Wireless Hand Control, EG1A (RECEIVER-HC)
Receivers are non-sterile capital equipment and shall always be maintained
outside the Operating Room Sterile Field.
Warning: Do not immerse or sterilize the receiver.
- Disconnect the receiver from the console.
- Clean the receiver by wiping with a non-abrasive cloth and neutral pH detergent and deionized, distilled, or purified water after each case.
- Dry thoroughly with a non-abrasive cloth.
Wireless Hand Control, EG1A (WIRELESS-HC)
Warning: Do not reprocess the Hand Control. Hand Controls are disposable
and SINGLE USE ONLY. Reprocessing hand controls may result in loss of function
causing a delay in surgery.
The Hand Control is provided sterile. Remove the product from the package
using an established operating room procedure for sterile transfer.
Store sterile devices in their original protective packaging, and do not
remove them from the packaging until immediately before use.
Prior to use, check the product expiration date and verify the integrity of
the sterile packaging. Do not use if the package is damaged.
Explanation of Symbols Used
General Symbols
| Direction of motion| | Reference or Catalogue Number (Ref. 5.1.6 ISO
15223-1)
---|---|---|---
| Consult Instructions for Use (Ref. 5.4.3 ISO 15223-1)| | Materials
| Caution: United States Federal law restricts this device to sale by or on
the order of a physician or other licensed healthcare provider(21 CFR
801.109)| | Date of Manufacture Manufacturer (Ref. 5.1.1 and 5.1.3 IS015223-1)
| Product is ETL listed to the requirements of US and Canada| | Keep dry (Ref.
5.3.4 ISO 15223-1)
| Locked| | Do not use it if the package is damaged. (Ref. 5.2.8 ISO 15223-1)
| Unlocked| |
| Non-ionizing electromagnetic radiation. (IEC 60417-5140) Transmitter
Interference may occur in the vicinity of equipment marked with this symbol.|
| CAUTION (Ref. 5.4.4 ISO 15223-1)
---|---|---|---
| Non-Sterile Ref. 5.2.7 ISO 15223-1| | Made in
| Single sterile barrier system with protective packaging inside (Ref. 5.2.13
ISO 15223-1)| | Packaging Unit – Indicates the number of pieces in the package
(ISO 7000-2794)
| Sterilized using ethylene oxide (Ref. 5.2.3 ISO 15223-1)| | Unique Device
Identifier
| Use-by Date (Ref. 5.1.4 ISO 15223- 1)| | Lot or Batch Number (Ref. 5.1.5 ISO
15223-1)
| Temperature Limit ((Ref. 5.3.7 ISO 15223-1)| | Humidity Limitation (Ref.
5.3.8 ISO 15223-1)
| Ingress protection rating according to IEC 60529| | Do not re-use (Ref.
5.4.2 ISO 15223-1)
| FCC Symbol| | Follow instructions for use (ISO 7010- M002)
Warnings
•The surgeon is responsible for learning proper techniques in the use of
equipment; improper use may cause serious injury to the user or patient or
damage to the system.
• Instrument operator and all operating room personnel must wear eye
protection.
• Do not use if the product package has been damaged or opened.
• Do not use if the product sterilization baRrier or its packaging is
compromised.
• Visually inspect device/components for damage before using; do not use if
damage or wear is evident (unrecognizable markings, missing or removed part
numbers, corrosion, etc.).
• MR Unsafe—Not for use in Magnetic Resonance Imaging (MRI) environments.
• Do not use excessive force.
• Wireless Hand Control, EG1A is disposable and intended for single patient
use only.
• Do not re-sterilize or reuse Wireless Hand Control, EG1A: reuse and/or re-
sterilization
of the hand, control could result in reduced performance, material
degradation, and/or patient contamination.| • Do not modify. Modifications
could result in loss of electrical safety.
• Faulty devices must not be used. Refer to the section “Disposal of Waste”.
• Do not operate in an explosive flammable environment.
• Wireless Hand Control, EG1A is only compatible with eG1 Handpiece and
Console (SC3001, SC3002). Using other than indicated high-speed systems may
lead to injury.
• Receiver for Wireless Hand Control, EG1A, and the console must be maintained
outside the sterile field.
• Keep Wireless Hand Control, EG1A safety cover in the locked position when
not in use
• Medical Electrical Equipment needs special precautions regarding
electromagnetic Compatibility (EMC) and needs to be installed and put into
service according to the EMC information provided in this accompanying
documentation.
• The emissions characteristics of this equipment make it suitable for use in
industrial areas and
---|---
• Do not immerse or sterilize Receiver for Wireless Hand Control, EG1A.
• Do not use in an oxygen-rich environment.
• Dispose of items contaminated with body fluids with other biohazardous
waste.
• At end of life recycle or dispose of devices in accordance with local and
national regulations.
• Verify that the Wireless Hand Control, EG1Ais connected to the intended
Receiver for Wireless Hand Control, EG1A. If the receiver controlling the
console is paired to unintended hand control, disconnect the receiver from the
EG1 console to prevent unintended motion of the handpiece.| hospitals (CISPR
11 class A). This equipment generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications.
• Operation of this equipment in a residential area (for which CISPR 11 class
B is normally required) this equipment might not offer adequate protection
to radio frequency communication services.
• In the event of interference, the user might need to take mitigation
measures such as relocating or reorienting the equipment.
• Portable and Mobile Radio Frequency (RF) communications equipment can affect
Medical Electrical Equipment.
• Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12
inches) to any part of the eG1 system. Otherwise, degradation of the
performance of this equipment could result, including intermittent loss of
function or inability to pair the device.
Precautions
• United States Federal law restricts this device to sale by or on the order
of a physician or other licensed healthcare provider.
• Do not use accessories other than those provided by DePuy Synthes Power
Tools and specified for use with Anspach Systems
• To ensure equipment operates as designed, read and follow manufacturer’s
instructions.| • Use care to protect when handling, cleaning, and during
system use.
• Do not push receiver connector into console connector port when out of
alignment.
---|---
Latex Information
Not made with natural rubber latex.
Wireless Hand Control Set Up
Prior to first use, the equipment (EG1A) must be processed as per the included EG1 High-Speed System Cleaning and Sterilization instructions.
-
Inspect the equipment prior to use
Warning : Wireless Hand Control System is only compatible with eG1 Handpiece (EG1A) and Console (SC3001, SC3002). Use with other than the indicated high-speed system may lead to injury or damage to the product.EG1A| Visually inspect for damage to the silicone hose or to the electrical connector. Connect to console and operate.
The handpiece should operate smoothly. There is no requirement to operate with an attachment or dissection tool.
---|---
SC3001| Visually inspect for damage or cracks to the housing. Visually inspect for damage to the electrical power cord. Power the system and inspect for illumination of the LEDs. If there is an irrigation pump present, check for proper function.
---|---
SC3002 -
nsert Receiver for Wireless Hand Control, EG1A into foot control connector port on the front of the console (SC3001 or SC3002) with a red dot on the connector facing up. See Figure 1.
Warning: Receiver and console must be outside sterile field.
Caution: Do not push the receiver connector into the console connector port when out of alignment.
-
Remove the Wireless Hand Control, EG1A from the packaging and attach the hand control to the eG1A handpiece. See Figure 2.
Warning: Keep hand control safety cover in the locked position when not in use.
-
Pair Handpiece. Press and release the button on the receiver (See Figure 3) and the hand control (See Figure 4) within 30 seconds to complete the pairing. When pairing is complete you will see the blue light go from flashing to solid on the receiver.
-
Start operation: Unlock the hand control, grab the battery pull tab and remove it by pulling in the direction indicated by the arrow, and press the button on the hand control to start the handpiece. Increase pressure on the button to increase the speed of the handpiece and release pressure on the button to decrease the speed of the handpiece.
Warning: Keep hand control safety cover in the locked position when not in use.
-
Stop Operation: Lock the wireless hand control and unpair the device. To unpair, the device simply presses the receiver button or unplug the receiver from the console.
Warning: Do not re-sterilize or reuse hand control: reuse and/or re- sterilization of hand control could result in reduced performance, material degradation, and/or patient contamination.
End of Life
Wireless Hand Control, EG1A (WIRELESS-HC)
The hand control is a SINGLE USE-only device. The product expiration date can
be found on the device package label.
Receiver for Wireless Hand Control, EG1A (RECEIVER-HC)
The lifetime requirements of the receiver are event-related depending on the
use and inspection of the device in clinical practice. The lifetime of the
device may be compromised following normal conditions of use such as, but not
limited to:
- Damage, including but not limited to corrosion (e.g. rust, pitting), discoloration, scratches, flaking, cracks and wear.
- The excessive temperature during operation.
- Improper function, including but not limited to inability to connect to console or failure to pair.
- Missing, faded, or removed device markings.
- Made obsolete or no longer in use.
- Damage impeding the ability to inspect or use the device.
Service Lifetime
These devices do not require periodic service.
Warranty & Return Policies
Warranty and return policy are available upon request.
All Instrument System components returned should be properly cleaned as
applicable prior to shipping to the manufacturer.
Warning: Transmissible Spongiform Encephalopathies (TSE)
DePuy Synthes Power Tools will not authorize or accept the return of products
that directly contact patients or are contaminated with a patient’s body
fluids who is suspected or confirmed with a Transmissible Spongiform
Encephalopathies/ Creutzfeldt-Jakob Disease (TSE/CJD) diagnosis. DePuy Synthes
Power Tools recommends that all ANSPACH High-Speed System products used on a
patient confirmed with a TSE/CJD diagnosis be incinerated. ANSPACH High-Speed
System Dissection Tools used on a patient suspected of TSE/CJD diagnosis must
be incinerated.
Contact your Sales Representative for replacement of products incinerated
under this policy or for temporary equipment while original equipment is
quarantined. Contact the DePuy Synthes Power Tools Customer Service Department
regarding TSE/CJD contamination for additional information.
Receiver for Wireless Hand Control, EG1A Displays
Display | Result | Description |
---|---|---|
Not Paired | Press the pairing button on the receiver and wait until the |
light flashes blue to initiate pairing.
| Pairing| The blue light indicator will “blink” with a slow flash during
parring.
| Paired| The blue light indicator will be solid blue.
| Out of Range| The blue light indicator will “blink” with a fast flash when
the device is out of range, or the signal is lost due to interference.
| Hand Control Battery is at FULL Capacity| The battery indicator will not be
ON.
| Hand Control Battery is at <50% Capacity| The battery indicator will be
solid red.
| Hand Control Battery is at <25% Capacity| The Battery indicator will start
flashing red.
Trouble Shooting
Wireless Hand Control, EG1A (WIRELESS-HC)
Receiver for Wireless Hand Control, EG1A (RECEIVER-HC)
Trouble Shooting Chart
Problem| Possible Cause| Possible Solution(s)
The receiver LED does not turn on when plugged into Console.| • Receiver not
fully inserted
• Console not powered ON| • Ensure receiver is fully inserted in foot pedal
connector
• Ensure Console is Power ON
Cannot pair the device| • Battery tab not removed from hand ontrol
• RF interference
• Dead battery on hand control
• Hand control was previously
paired with another receiver| • Ensure the battery pull tab is removed on
• Repeat “Pairing Instructions” found on setting up a section or ensure the
device maintains a minimum distance between portable and mobile RF
communications equipment (transmitters) and the eG1 System as recommended in
the ” Recommended separation distances between portable and mobile RF
communications equipment and the eG1 System ” table
• Discard and replace the receiver
• Discard and replace hand control (can only be paired with one receiver)
Paired, but does not actuate| • 1.EG1 Handpiece is unlocked
(load) position
• Interference| • Ensure EG1 handpiece is locked (run) position
• Move the equipment away from the source of interference
Device becomes unpaired| • Lost connection due to possible
RF interference
• Battery failure| • Ensure the device maintain a minimum distance between
portable and mobile RF communications equipment
(transmitters) and the eG1 System as recommended in the ” Recommended
separation distances between portable and mobile RF communications equipment
and the eG1 System ” table
• Discard and replace hand control
Loss of Power| • Console power failure| • Ensure the console in plugged in and
repeat the “pairing instructions” found in the setup section.
Technical Specifications
FCC ID:
WIRELESS-HC: 2AYUW-WHC1
RECEIVER-HC: 2AYUW-RHC1
IC:
WIRELESS-HC: 27015-WHC1
RECEIVER-HC: 27015-RHC1
The devices comply with the following standards:
FCC 47CFR PT 15.247 Issued:2004/10/01
Operation within the bands 902-928 MHz, 2400-2483.5 MHz, and 5725-5850 MHz.
All measurements were performed in 10m chambers.
FCC 47CFR: (Part 15 Subpart B)
Title 47 CFR Part 15 Subpart B: Unintentional Radiators
Concurrently with FCC Part 15.247
RSS-247
Issue: 2015/05/22 Issue 1 Digital Transmission Systems (DTSs),
Frequency Hopping Systems (FHS) and License-Exempt Local Area Network (LE-LAN)
Devices
Concurrently with FCC Part 15.247
RSS 102:2015 Ed.5
Radio Frequency (Rf) Exposure Compliance of Radiocommunication
Apparatus (All Frequency Bands)
Concurrently with FCC Part 15.247
IC ICES-003:2016Ed.6
Information Technology Equipment (Including Digital Apparatus) – Limits and
Methods of Measurement
Concurrently with FCC Part 15.247
ETSI EN 300 328:2019 Ed.2.2.2
Wideband Transmission Systems; Data Transmission Equipment Operating in the
2,4 GHz band; Harmonized
Standard for Access to Radio Spectrum
ETSI EN301489-1:2017Ed. V2.1.1
Electromagnetic Compatibility (EMC) Standard For Radio Equipment And Services;
Part 1: Common Technical
Requirements; Covering The Essential Requirements Of Article 3.1(B) Of
Directive 2014/53/Eu And The
Requirements Of Article 6 Of Directive 2014/30/Eu
Includes:
ETSI EN 301 489-17 V3.1.1 (2017-02)
ETSI EN 303 446-1 V1.2.1 (2019-07)
EN IEC 62311:2020 Testing and/or Evaluation
Receiver for Wireless Hand Control, EG1A, and Wireless Hand Control, EG1A intentionally receives electromagnetic energy for the purpose of its operation:
- The frequency band of reception: 2405MHz to 2475MHz
- Receiving bandwidth: 20dB bandwidth: 2.8MHz
Receiver for Wireless Hand Control, EG1A and Wireless Hand Control, EG1A include RF transmitters:
- The frequency band of transmission: 2405MHz to 2475MHz
- Type and frequency characteristics of the modulation: DSSS (direct sequence spread spectrum)/ QPSK (offset quadrature phase shift keying)
- Effective radiated power: 4dBm
FCC Compliance Statement:
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
ETL Authorization Mark:
Technical Specifications for Devices Compliant with IEC-60601-1-2 4 the
Edition
The device complies with the following standards:
AAMI ES60601-1 Issued: 2006/03/09 Medical Electrical Equipment – Part 1:
General Requirements for Basic Safety
and Essential Performance, Amendment 1 – 2012
CSA C22.2#60601-1 Issued: 2014/03/01 Ed: 3 Medical Electrical Equipment – Part
1: General Requirements for
Basic Safety and Essential Performance
IEC 60601-1-6 Issued: 2013/10/29 Ed: 3 Medical Electrical Equipment – Part
1-6: General Requirements for Basic
Safety and Essential Performance – Collateral Standard: Usability; Amendment 1
IEC 60601-1: 2005 + CORR. 1:2006 + CORR. 2:2007 + AM1:2012 (or IEC 60601-1:
2012 reprint) EN 60601-1:
2005 + A1:2013, IEC 60601-1 / EN 60601-1, Medical electrical equipment Part 1:
General requirements for basic
safety and essential performance (CB Scheme)
IEC 60601-1-6:2010 (Third Edition) + A1:2013 for use in conjunction with
IEC62366:2007 (First Edition) + A1:2014
and IEC 60601-1:2005 (Third Edition) + Corr.1 (2006) + Corr.2 (2007) + A1:
2012 or equivalent consolidated
version IEC 60601-1:2012 (Ad.3.1), Medical electrical equipment Part 1-6
General requirements for safety –
Collateral Standard: Usability
IEC 60601-1-2:2014 (Edition 4) Collateral Standard: Electromagnetic
Compatibility- Medical Electrical Equipment.
Environmental Conditions
Temperature| Wireless Hand Control, EG1A (WIRELESS-HC)| Receiver for
Wireless Hand Control, EG1A (RECEIVER-HC )
Operating: 18°-30°C (65°-85°F)| Operating: 18°-30°C (65°-85°F)
Transportation and Storage: -29° — +60°C (-20° – +140°F)| Transportation and
Storage: -29° — +60°C (-20° – +140°F)
Relative Humidity| Operating: 30% to 70%| Operating: 30% to 70%
Transportation and Storage: 10% to 90%| Transportation and Storage: 50% to 90%
Atmospheric Pressure| Operating: 0.7–1.06bar| Operating: 0.7–1.06bar
---|---|---
Transportation and Storage: Not applicable| Transportation and Storage: Not
applicable
Receiver for Wireless Hand Control, EG1A powered through Console
(SC3001/SC3002) Specifications
Primary: 100-240VAC, 50/60 Hz, 250 VA
Class I: Protective Earth
Wireless Hand Control, EG1A Specifications
Powered through: CR1620 lithium coin battery, nonreusable – disposable
Fluid Ingress Protection: IPX2
Guidance and manufacturer’s declaration —electromagnetic emissions
The eG1 System is intended for use in the electromagnetic environment
specified below. The customer or the user of the eG1 System should assure that
it is used in such an environment.
Emissions test| Compliance| Electromagnetic environment — guidance
RF emissions CISPR 11| Group 1| The eG1 System uses RF energy only for its
initial function. Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions CISPR 11| Class A| The eG1 System is suitable for use in all
establishments other than domestic establishments and those directly connected
to the public low-voltage power supply network which supplies buildings used
for domestic purposes.
Harmonic emissions IEC 61000-3-2| Class A
Voltage fluctuations / flicker emissions
IEC 61000-3-3| Complies
Guidance and manufacturer’s declaration —electromagnetic Immunity
The eG1 System is intended for use in the electromagnetic environment
specified below. The customer or the user of the eG1 System should assure that
it is used in such an environment.
Immunity test| IEC 60601 test level| Compliance level| Electromagnetic
environment
— guidance
Electrostatic discharge (ESD)
IEC 61000-4-2| +8 kV contact ±15 kV air| -±8 kV contact ±15 kV air| Floors
should be wood, concrete or ceramic tile. If floors are covered with synthetic
material, the relative humidity should be at least 30 %.
Electrical fast transient/burst
IEC 61000-4-4| ±2 kV for AC lines ±1 kV for I/O lines
100kHz Pulse Repetition Frequency| ±2 IN for AC lines ±1 IN for I/O lines
100kHz Pulse Repetition
Frequency| The main power quality should be that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5| ±0.5, 1kV line to line ±0.5, 1, 2 kV line to earth| ±0.5, 1kV
line to line ±0.5, 1, 2 kV line to earth| The main power quality should be
that of a typical commercial or hospital environment.
Voltage dips, short interruptions, and voltage variations on power supply
input lines IEC 61000-4-11| 0% UT (100% dip in UT) for 0.5 periods) at 0°,
45°, 90°,135°, 180°, 225°, 270°, 315° 0% UT (100% dip in UT) for 1 period) 70%
UT (30% dip in UT) for 25 periods (50Hz) 30periods (60Hz)
Interruptions 0% UT (100% dip in UT) for 250 periods (50Hz) 320 periods (60Hz)
225°, 270°, 315°| 0% UT (100% dip in UT) for 0.5 periods) at 0°, 45°,
90°,135°, 180°, 0% UT (100% dip in UT) for 1 period) 70 UT (30% dip in UT) for
25
periods (50Hz) 30 periods (60Hz) Interruptions 0% UT (100% dip in UT) for 250
periods (50Hz) 320 periods (60Hz) •| The main power quality should be that of
a typical commercial or hospital environment.
(*) If the user of the eG1 System requires continued operation during power
mains
interruptions, it is recommended that the eG1 System be powered from an
uninterru table
power supply or a battery.
---|---|---|---
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8| 30 A / m| 30 A / m|
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
NOTE UT is the A.C. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration –electromagnetic immunity
The eG1 System is intended for use in the electromagnetic environment
specified below. The customer or the user of the eG1 System should assure that
it is used in such an environment.
Immunity test| IEC 60601 test level| Compliance level| Electromagnetic
environment – guidance
---|---|---|---
Conducted RF IEC 61000-4-6| 3 Vrms150 kHz to 80 MHz
6 Vrms in ISM bands 150 kHz to 80 MHz| 3 Vrms 150 kHz to 80 MHz
6 Vrms in ISM bands 150 kHz to 80 MHz| Portable and mobile RF communications
equipment should
be used no closer to any part of the eG1 System, including
cables, then the recommended separation distance
calculated from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, should be less than the
compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked
with the following symbol:
Radiated RF IEC 61000-4-3| 3 V/m 80 MHz to 2,7 GHz80% at 1kHZ (AM modulation)|
3 V/m 80 MHz to 2,7 GHz 80% at 1kHZ (AM modulation)
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects,
and people.
- Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the eG1 System is used exceeds the applicable RF compliance level above, the eG1 System should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the eG1 System.
- Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
eG1 System Enclosure Port Immunity to RF Wireless Communications Equipment (IEC 61000-4-3)
Test Frequency (MHz)| Band a) (MHz)| Service a)| Modulation b)| Maximum Power
(W)| Distance (m)| Immunity Test Level (V/m)
385| 380 -390| TETRA 400| Pulse
modulation b)
18 Hz| 1,8| 0.3| 27
450| 430 – 470| GMRS 460, FRS460| FM c)
± 5 kHz deviation
1 kHz sine| 2| 0.3| 28
710| 704 – 787| LTE Band 13,17| Pulse modulation b) 217HZ| 0.2| 0.3|
745
780
810| 800 – 960| GSM 800/900. TETRA 800, iDEN 820, CDMA 850. LTE Band 5| Pulse
modulation b) 18Hz| 2| 0.3| 28
870
930
1720| 1 700-
1 990| GSM 1800;
COMA 1900;
GSM 1900;
DECT;
LTE Band 1. 3.
4.25; UMTS| Pulse modulation b) 217 Hz| 2| 0.3| 28
1845
1970
2450| 2 400 –
2 570| Bluetooth.
WLAN, 802.11 b/g/n.
RFID 2450, LTE Band 7| Pulse Modulation b) 217 Hz| 2| 0.3| 28
5240| 5 100 –
5 800| WLAN 802.11 a/n| Pulse modulation b) 217 Hz| 0,2| 0,3| 9
5500
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the
transmitting antenna and the eG1 System may be reduced to 1 m. The 1 m test
distance is permitted by IEC 610004-3.
a)For some services, only the uplink frequencies are included.
b)The carrier shall be modulated using a 50 % duty cycle square wave signal.
c)M an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be
used because while it does not represent actual modulation, it would be the
worst case.
Recommended separation distances between portable and mobile RF
communications equipment and the eG1 System
The eG1 System is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the eG1
System can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment
(transmitters) and the eG1 System as recommended below, according to the
maximum output power of the communications equipment.
The rated maximum output power of transmitter
w| Separation distance according to the frequency of the transmitter m
150 kHz to 80 MHz| 80 MHz to 800 MHz| 800 MHz to 2,5 GHz
| |
| |
0,01| 0.12| 0.12| 0.23
0,1| 0.38| 0.38| 0.73
1| 1.| 1.| 2.
10| 4.| 4.| 7.
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects,
and people.
Customer Service
Contact your local Depuy Synthes representative for Troubleshooting assistance. To request a paper copy of the IFU call 800-255-2500.
Disposal of Waste
Dispose of items contaminated with body fluids or with other biohazardous
waste. Please send devices that are no longer used to your local DePuy Synthes
representative. Devices must be disposed of as healthcare medical devices in
accordance with hospital procedures. The device contains electronics/battery
and should be disposed off
in accordance with local and/or national regulations.
www.e-IFU.com
US / 800-255-2500
The Anspach Effort, Inc 4500 Riverside Drive
Palm Beach Gardens, FL 33410 USA
Tel: (800) 327 6887 / +1 561 627 1080
Not all products are currently available in all markets. Data are available on
request. Please contact your DePuy Synthes Representative for more
information.
All DePuy Synthes Implant Instructions for Use as well as general Instructions
for Use are available as PDF files at
www.e-ifu.com
www.depuysynthes.com
Pat. www.jnjmedicaldevices.com/en-
US/patentmarking
© DePuy Synthes 2021. All rights reserved.
0000314957, A.5, 01/ 2022