MAJOR TECH MTD70T 600A AC True RMS Digital Clamp Meter Instruction Manual

June 9, 2024
MAJOR TECH

MAJOR -logo INSTRUCTION MANUAL
MTD70T
600A AC TRUE RMS
DIGITAL CLAMP METER 

MAJOR TECH MTD70T 600A AC True RMS Digital Clamp
Meter- TRUE RMS

SAFETY

1.1. lnternational Safety Symbols

**** This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present.
Icon Double insulation.
Application around and removal from uninsulated hazardous live conductors is permitted.

1.2. Safety Notes

  • Do not exceed the maximum allowable input range of any function.
  • Do not apply voltage to meter when resistance function is selected.
  • Set the function switch OFF when the meter is not in use.
  • Remove the battery if meter is to be stored for longer than 60 days.

1.3. Warnings

  • Set function switch to the appropriate position before measuring.
  • When measuring volts do not switch to current/resistance modes.
  • Do not measure current on a circuit whose voltage exceeds 600V.
  • When changing ranges always disconnect the test leads from the circuit under test.

1.4. Cautions

  • Improper use of this meter can cause damage, shock, injury or death.
    Read and understand this user manual before operating the meter.

  • Always remove the test leads before replacing the battery or fuses.

  • Inspect the condition of the test leads and the meter itself for any damage before operating the meter. Repair or replace any damage before use.

  • Use great care when making measurements if the voltages are greater than 25VAC RMS or 35VDC. These voltages are considered a shock hazard.

  • Always discharge capacitors and remove power from the device under test before performing Diode, Resistance or Continuity tests.

  • Voltage checks on electrical outlets can be difficult and misleading because of the uncertainty of connection to the recessed electrical contacts. Other means should be used to ensure that the terminals are not “live”.

  • If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.

Function Maximum Input
A AC 600A AC
V DC, V AC 1000V DC/AC
Resistance, Capacitance, Frequency,  Diode Test, Temperature. 250V DC/AC

Descriptions

2.1. Meter Descriptions

  1. NCV Test
  2. Current clamp
  3.  LED Flashlight
  4. Non-contact AC voltage indicator light
  5. Clamp trigger
  6. Rotary Function switch
  7. SELECTOR mode range button
  8. Data Hold button
  9. REL button for Relative Measurement
  10. Flashlight and LCD backlight button
  11. LCD display
  12.  COM input jack
  13. V Ω Diode Continuity CAP TEMP Hz% jack
  14.  Battery Cover

MAJOR TECH MTD70T 600A AC True RMS Digital Clamp Meter-
fig1

2.2. Display Icons

Button Operation Performed
HOLD Data Hold
Minus sign Negative reading display
0 to 3999 Measurement display digits
NCV Non-Contact Voltage
AUTO Auto Range mode
DC Direct Voltage
AC Alternating Current/Voltage

| Low battery
mV or V| Milli-volts or Volts (Voltage)
Ω| Ohms (Resistance)
A| Amperes (Current)
F| Farad (Capacitance)
Hz/%| Hertz (Frequency)/Percent(duty ratio)
°C| Celsius units (Temperature)
n, m, μ, M, k| Unit of measure prefixes: nano, milli, micro, mega, and kilo
| Continuity test
| Diode test

General Specifications

Basic Functions            Range

Clamp jaw opening| 28mm approx.
Display| 2000 Count (2-1/2 digits backlit LCD)
Low Battery indication| is displayed
Over-range indication| ‘OL’ display
Measurement rate| 3 readings per second, nominal
Input Impedance| 10M (VDC and VAC)
AC response| Average Reading (AAC and VAC)
ACV Bandwidth| 2KHZ
Operating Temperature| 5°C to 40°C (41°F to 104°F)
Storage Temperature| -20°C to 60°C (-4°F to 140°F)
Operating Humidity| Max 80% up to 31°C (87°F) Decreasing

linearly to 50% at 40°C (104°F)

Storage Humidity| <80%
Operating Altitude| 2000meters (7000ft.) maximum.
Battery| 2 x 1.5V AAA Battery
Battery life| Approx. 30h (backlight ON), Approx. 100h (backlight OFF)
Auto power OFF| After approx. 15 minutes
Dimensions| 185 x 65 x 37mm
Weight| 180g
Safety| For indoor use and in the requirement For double insulation IEC1010-1(2001): EN61010-2-030, EN61010-2-032, EN61010-2-033 Over voltage Category III 600V, Pollution Degree 2.

Specifications

4.1. AC True RMS Current

Range Resolution Accuracy (% of reading + digits)
2.000A 1mA ±3.0% of rdg±5 digits
20.00A 10mA **** ±2.0% of rdg±5 digits
200.0A 100mA
600A 1A

Over range protection: Maximum input 600A
Accuracy specified from 5% to 100% of the measuring range
Frequency Response: 50Hz to 60Hz True RMS

4.2. AC Voltage True RMS

Range Resolution Accuracy (% of reading + digits)
200mV 0.1mV ±1.5% of rdg±10 digits
2.00V 1mV **** ±1.0% of rdg±5 digits
20.0V 10mV
200V 0.1V
600V 1V ±1.2% of rdg±5 digits

Input impedance 10MΩ. protection 600V DC or RMS of AC.
Frequency response 60V at 40Hz to 1kHz (sine), other ranges 50/60Hz (all wave).

4.3. DC Voltage

Range Resolution Accuracy (% of reading + digits)
200mV 0.1mV **** ±0.5% of rdg±5 digits
2.00V 1mV
20.0V 10mV
200V 0.1V
600V 1V ±1.2% of rdg±5 digits

Input impedance on 200mV range >40MΩ all other ranges 10MΩ.
Overload protection 600V DC or RMS of AC.

4.4. Resistance (Auto Ranging)

Range Resolution Accuracy (% of reading + digits)
200.0Ω 0.1Ω ±0.8% of rdg±5 digits
2.000kΩ **** ±0.8% of rdg±1 digits
20.00kΩ 10Ω
200.0kΩ 100Ω
2.000MΩ 1kΩ
20.00MΩ 10kΩ ±1.2% of rdg±5 digits

Input Protection: 250V DC or Peak value of AC.
Note: a. On 200Ω range, short the test leads to measure lead resistance, then subtract the value from the real measurement.
b. When measuring resistance over 1MΩ, it is normal that the reading reacts slowly, wait until reading is stable.

4.5. Diode & Continuity

Function Testing Condition Reading
Diode Forward DCA is approx. 0.8mA, open circuit Voltage MAX. 2.2V Forward

voltage drop of Diode
Continuity| Test current MAX. 1.5mA| Buzzer makes a continuous sound while resistance is less than (50Ω)

Input Protection: 250V DC or 250V AC RMS
Warning DO NOT apply voltage to Continuity range

Operation

NOTES: Read and understand all Warning and Caution statements in this manual prior to using this meter. Set the function select switch to the OFF position when the meter is not in use.

5.1. AC Current Measurements

WARNING: Ensure that the test leads are disconnected from the meter before making current clamp measurements.

  1. Set the Function switch to the 200A/600A range if the approximate range of the measurement is not known.
  2. Set Function switch to 2A/20A for low current readings.
  3. Press the trigger to open jaw. Fully enclose only one conductor. For optimum results, center the conductor in the jaw.
  4. The clamp meter LCD will display the reading.

MAJOR TECH MTD70T 600A AC True RMS Digital Clamp Meter-
fig2

5.2. AC Voltage Measurements

  1. Insert the black test lead into the negative COM terminal and the red test lead into the positive V/Ω terminal.
  2. Set the function switch to the V~ position.
  3. Connect the test leads in parallel to the circuit under test.
  4. Read the voltage measurement on the LCD display.

5.3. DC Voltage Measurements

  1. Insert the black test lead into the negative COM terminal and the red test lead into the positive V/Ω terminal.
  2. Set the function switch to the V position.
  3. Connect the test leads in parallel to the circuit under test.
  4. Read the voltage measurement on the LCD display.

5.4. Resistance Measurements

  1. Insert the black test lead into the negative COM terminal and the red test lead into the positive V/Ω terminal.
  2. Set the function switch to the Ω position.
  3. Touch the test probe tips across the circuit or component under test. It is best to disconnect one side of the device under test so the rest of the circuit will not interfere with the resistance reading.
  4. For Resistance tests, read the resistance on the LCD display.

5.5. Continuity Measurements

  1. Insert the black test lead into the negative COM terminal and the red test lead into the positive V/Ω terminal.
  2. Set the function switch to the position.
  3. Use SELECT Button to select . The display icons will change when the SELECT button is pressed.
  4. Touch the Test Probe tips across the circuit or component under test.
  5. If the resistance is <50Ω, the buzzer tone will sound.

5.6. Diode Test

  1. Insert the black test lead into the negative COM terminal and the red test lead into the positive V/Ω terminal.
  2. Set the function switch to position
  3. Use SELECT Button to select . The display icons will change when the SELECT button is pressed.
  4. Touch the test probe tips to the diode or semiconductor junction under test. Note the meter reading.
  5. Reverse the test lead polarity by reversing the red and black leads. Note the reading
  6. The diode or junction can be evaluated as follows:
    • If one reading displays a value (typically 0.400V to 0.900V) and the other reading displays OL, the diode is good.
    • If both readings display OL device is open.
    • If both readings are very small or “0”, the device is shorted.

5.7. Non-Contact AC Voltage Measurements

WARNING: Risk of Electrocution. Before use, always test the Voltage Detector on a known live circuit to verify proper operation

  1. Touch the probe tip to the hot (LIVE) conductor or insert into the hot (LIVE) side of the electrical outlet.
  2. If AC voltage is present, the detector red LED light will illuminate.
    a. NOTE: The conductors in electrical cord sets are often twisted. For best results, rub the probe tip along a length of the cord to assure placing the tip in close proximity to the live conductor.
    b. NOTE: The detector is designed with high sensitivity. Static electricity or other sources of energy may randomly trip the sensor. This is normal operation.

5.8. Data HOLD
To freeze the LCD reading, press the HOLD button. While data hold is active, the HOLD icon appears on the LCD. Press the HOLD button again to return to normal operation.

5.9. Flashlight & LCD Backlight
The LCD is equipped with backlighting for easier viewing, especially in dimly lit areas. Press the button for 3 seconds to turn the Flashlight and LCD backlight on. Press once again to turn the Flashlight and backlight off. The both work simultaneously.

5.10. Range Function
Press the RANGE key to activate the manual mode and to disable the Auto range function. The symbol “AUTO” disappears from the upper left part of the display. In manual mode, press the RANGE key to change measuring range: the relevant decimal point will change its position. The RANGE key is not active in positions , . In Auto range mode, the instrument selects the most appropriate ratio for carrying out measurement. If a reading is higher than the maximum measurable value, the indication “OL” appears on the display. Press and hold the RANGE key for more than 1 second to exit the manual mode and restore the Auto range mode.

5.11. Automatic Power OFF

In order to conserve battery life, the meter will automatically turn off after approximately 15 minutes. To turn the meter on again, turn the rotary function switch to the OFF position and then to the desired function range. To cancel the auto power off function, press SELECT button and turn on the power with the rotary switch at the same time.

Maintenance

WARNING: To avoid electrical shock, disconnect the meter from any circuit, remove the test leads from the input terminals, and turn OFF the meter before opening the case. Do not operate the meter with an open case.

6.1. Cleaning and Storage
Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or solvents. If the meter is not to be used for 60 days or more, remove the battery and store it separately.

6.2. Battery Replacement

  1. When the LCD displays , please replace the battery to ensure correct readings
  2. Remove the Phillips head screw that secures the rear battery door
  3. Open the battery compartment
  4. Replace the 2x 1.5V AAA battery, recommend Alkaline batteries for longer use
  5. ecure the battery compartment

MAJOR -logo

MAJOR TECH (PTY) LTD
South Africa
www.major-tech.com
[email protected]
Australia
www.majortech.com.au
[email protected]

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

MAJOR TECH User Manuals

Related Manuals